The Ultimate rTMS is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medications in the current episode.
Device Story
Ultimate rTMS is a computerized repetitive transcranial magnetic stimulation (rTMS) system; produces non-invasive, repetitive pulsed magnetic fields to induce neural action potentials in the left dorsolateral prefrontal cortex. Principle of operation: high-voltage capacitor discharge through a figure-of-eight stimulation coil; generates rapidly alternating magnetic fields penetrating neuromuscular tissues to induce cortical electrical currents. Used in patient care institutions, diagnostic centers, neurosurgical hospitals, and research labs; operated by trained personnel. Healthcare providers use the system to deliver predefined stimulation protocols (iTBS, 19-min, 37-min) to treat MDD. Output affects clinical decision-making by providing a non-pharmacological treatment option for patients resistant to antidepressant medication. Benefits include potential symptom improvement for MDD patients.
Clinical Evidence
Bench testing only. Performance testing included electrical field strength decay, magnetic field linearity, spatial distribution, gradient (dB/dt), and pulse width/amplitude waveform analysis. Intensity variation of pulses within a burst was verified to be <1%. Biocompatibility of patient caps is exempt per ISO 10993-1 guidance.
Technological Characteristics
System consists of a TMS stimulator, electromagnetic figure-of-eight coils (BY90A, BF90A, BY90B), and a flexible positioning arm. Coil windings: copper tubing (GIS-C1100); housing: polycarbonate (Covestro Makrolon PC 2805). Energy source: high-voltage capacitor discharge. Waveform: biphasic sinusoid. Connectivity: standalone. Standards: IEC 60601-1, IEC 60601-1-2, ISO 13485, ISO 14971. Cooling: liquid or forced air depending on coil model.
Indications for Use
Indicated for treatment of Major Depressive Disorder in adults who failed to receive satisfactory improvement from prior antidepressant medications in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
Predicate Devices
Mag Vita TMS Therapy System w/Theta Burst Stimulation (K173620)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 17, 2025
Brain Ultimate, Inc.
℠ Barry Ashar
President
Makromed, Inc.
88 Stiles Road
Salem, New Hampshire 03079
Re: K243460
Trade/Device Name: Ultimate rTMS
Regulation Number: 21 CFR 882.5805
Regulation Name: Repetitive transcranial magnetic stimulation system
Regulatory Class: Class II
Product Code: OBP
Dated: March 20, 2025
Received: March 20, 2025
Dear Barry Ashar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243460 - Barry Ashar
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243460 - Barry Ashar
Page 3
Sincerely,
ROBERT KANG -S
for Pamela Scott
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243460
Device Name
Ultimate rTMS
Indications for Use (Describe)
The Ultimate rTMS is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medications in the current episode.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) #: 510(k) Summary
Prepared on: 2024-11-06
| Contact Details | 21 CFR 807.92(a)(1) |
| --- | --- |
| Applicant Name | Brain Ultimate, Inc. |
| Applicant Address | 5910 Shiloh Road East Alpharetta GA 30005 United States |
| Applicant Contact Telephone | (770) 378-6785 |
| Applicant Contact | Mr. Frank Ge |
| Applicant Contact Email | Frank.Ge@brainultimate.com |
| Device Name | 21 CFR 807.92(a)(2) |
| --- | --- |
| Device Trade Name | Ultimate rTMS |
| Common Name | Repetitive transcranial magnetic stimulation system |
| Classification Name | Transcranial Magnetic Stimulator |
| Regulation Number | 882.5805 |
| Product Code(s) | OBP |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
| --- | --- |
| Predicate #
Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K173620 | Mag Vita TMS Therapy System w/Theta Burst Stimulation |
| K171481 | Mag Vita TMS Therapy System |
| | OBP |
| | OBP |
| Device Description Summary | 21 CFR 807.92(a)(4) |
| --- | --- |
| The Ultimate rTMS is a repetitive transcranial magnetic stimulation (rTMS) system. This computerized medical device produces non-invasive, repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex for the treatment of Major Depressive Disorder.
It can be used in patient care institutions, diagnostics centers, neurosurgical hospitals and experimental laboratories of research institutions.
The Ultimate rTMS principle of operation is based on the discharge of high voltage capacitor through stimulation coil; the pulsed magnetic field generated by the discharge current penetrates through neuromuscular tissues nearby to induce electrical currents in cortical neurons.
Area of the brain to be simulated for MDD treatment is Left Dorsolateral Prefrontal Cortex. | |
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) |
| --- | --- |
| The Ultimate rTMS is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medications in the current episode. | |
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
Page 1 of 13
{5}
Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device of this 510(k) submission has the exact same Indication for Use as that of the predicate device.
# Technological Comparison
21 CFR 807.92(a)(6)
## Introduction
The subject device, the Ultimate rTMS, and the predicate device, Mag Vita TMS Therapy System, (K173620) are computerized electromechanical medical devices that produce and deliver non-invasive, repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex for the treatment of Major Depressive Disorder. Both devices have identical intended use / indication for use, identical treatment target and use identical treatment parameters of 19-minutes, 37-minutes and iTBS protocols.
## Design
The design of the subject Ultimate rTMS device is similar to that of the predicate Mag Vita TMS Therapy System, as both systems are based on applying transcranial magnetic stimulation by means of repetitive pulse trains at a predetermined frequency. Both systems use the same mechanism of action, i.e., an electromechanical instrument that produces and delivers brief duration, rapidly alternating (pulsed) magnetic fields to induce electrical currents in localized regions of the prefrontal cortex. Coils for subject Ultimate rTMS and the predicate device are of similar nature and share the same transducer design (figure-of-eight).
## Operational Characteristics
For the subject Ultimate rTMS and its predicate device, their basic operational procedures including system setup, patient preparations, motor threshold determination, coil positioning and treatment with predefined treatment stimulation parameters are the same.
## Technological Characteristics
Both the subject and predicate devices have similar components consisting of TMS stimulator with software, electromagnetic coil and a flexible arm for positioning of the treatment coil. The reliability of the positioning method used by the subject device is based on the direct relationship of the underlying cortical brain anatomy to the patient's scalp, as is the method used in the predicate devices. The method for identifying the correct treatment position in the subject device is at least as effective as the method employed by the predicate device.
## Non-clinical Performance Characteristics
1. Electrical/mechanical/thermal safety and electromagnetic compatibility of both the subject and predicate devices are in compliance with the following standards:
- IEC 60601-1:2005/(R)2012
- IEC 60601-1-2:2014
2. Patient caps are the only accessory of this device that comes in contact with the patient's intact skin for a duration of less than 24 hours. As they are made of commonly used textile materials they are exempt from ISO 10993-1:2018 Biocompatibility testing per FDA's Guidance Document 'Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", dated September 8, 2023.
3. The characteristics of the electric and magnetic fields generated by the subject device's coils (BY90A, BF90A, and BY90B), including linearity of output, biphasic sinusoidal waveform, pulse width and intensity, magnetic field spatial distribution and magnetic field strength gradient, are equivalent to that of the predicate device.
4. The subject device meets the same requirement as the predicate device that the actual individual stimuli of a theta burst are of equal intensity to ensure that a constant dose of stimuli is delivered during iTBS treatment (maximum 1% drop in machine output among the three stimuli in a burst).
## Conclusion
The subject device is equivalent in its performance and does not introduce any new safety considerations in comparison to the predicate device. There are no identified differences between the subject device and the predicate that impact on safety or efficacy.
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{6}
| Criteria | (Subject Device) Ultimate rTMS | (Primary Predicate) MagVita TMS Therapy System (K173620) | (Secondary Predicate) MagVita TMS Therapy System (K171481) | Equivalence Comments |
| --- | --- | --- | --- | --- |
| Intended Use (Indication for Use) | The Ultimate rTMS is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. | The MagVita TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. | The MagVita TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. | Identical Indication for Use |
| Recommended Standard Treatments | | | | |
| Magnetic Field Intensity | 120% of the MT (iTBS, 19/37 mins protocols) | 120% of the MT (iTBS protocol) | 120% of the MT (19/37 mins protocol) | Identical recommended treatment parameters. The subject device provides iTBS, 19-min and 37-min protocols. The primary predicate device provides iTBS protocol. The secondary predicate device |
| Repetition Rate | 50 Hz / 10 Hz (iTBS, 19/37 mins protocols) | 50 Hz (iTBS protocol) | 10 Hz (19/37 mins protocols) | |
| Train duration | 2 sec / 4 sec (iTBS, 19/37 mins protocols) | 2 sec (iTBS protocol) | 4 sec (19/37 mins protocols) | |
| Inter-train interval | 8 sec / 11-26 sec (iTBS, 19/37 mins protocols) | 8 sec (iTBS protocol) | 11-26 sec (19/37 mins protocols) | |
| iTBS Burst | Burst Pulses: 3 No. of Bursts: 200 Inter-pulse: 20 msec | Burst Pulses: 3 No. of Bursts: 200 Inter-pulse: 20 msec | N/A | |
| Number of trains | 20 / 75 (iTBS, 19/37 mins protocols) | 20 (iTBS protocol) | 75 (19/37 mins protocols) | |
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{7}
| Criteria | (Subject Device) Ultimate rTMS | (Primary Predicate) MagVita TMS Therapy System (K173620) | (Secondary Predicate) MagVita TMS Therapy System (K171481) | Equivalence Comments |
| --- | --- | --- | --- | --- |
| Magnetic Pulses per Session | 600 / 3000 (iTBS, 19/37 mins protocols) | 600 (iTBS protocol) | 3000 (19/37 mins protocols) | provides 19-min and 37-min protocols. |
| Treatment Session Duration | 3 min 09 sec / 19/37 min (iTBS, 19/37 mins protocols) | 3 min 09 sec (iTBS protocol) | 19/37 min (19/37 mins protocols) | |
| Treatment Schedule | 5 Sessions/Week for 6 Weeks | 5 Sessions/Week for 6 Weeks | 5 Sessions/Week for 6 Weeks | |
| Area of brain to be stimulated | Left Dorsolateral Prefrontal Cortex | Left Dorsolateral Prefrontal Cortex | Left Dorsolateral Prefrontal Cortex | |
| Coils | | | | |
| Coils | BY90A
BF90A
BY90B | Cool-B70 | C-B60
C-B65 | N/A |
| Configuration | BY90A : Figure-of-eight coil
BF90A : Figure-of-eight coil
BY90B : Figure-of-eight coil-With angle 120° | Figure-of-eight coil | Figure-of-eight coil | Same figure-of-eight configuration in all coils. See Note A. |
| Materials of Construction | Coil windings:
copper tubing (commercial brand GIS-C1100)
Coil housing:
polycarbonate (commercial brand Covestro Makrolon PC 2805) | Coil windings:
copper tubing (commercial brand unknown)
Coil housing:
polycarbonate (commercial brand unknown) | Coil windings:
copper tubing (commercial brand unknown)
Coil housing:
polycarbonate (commercial brand unknown) | Equivalent, with no additional concerns for safety and effectiveness. |
| Core material | Air core | Air core | Air core | Same core material design. |
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{8}
| Criteria | (Subject Device) Ultimate rTMS | (Primary Predicate) MagVita TMS Therapy System (K173620) | (Secondary Predicate) MagVita TMS Therapy System (K171481) | Equivalence Comments |
| --- | --- | --- | --- | --- |
| Cooling | BY90A:Liquid cooling | Cool-B70: Liquid cooling | C-B60: None | Equivalent.
The subject and secondary predicate devices offer liquid-cooled and forced air-cooled or uncooled coils. Primary predicate device offers liquid-cooled coil. |
| | BF90A:Forced Air cooling | | C-B65: Liquid cooling | |
| | BY90B:Liquid cooling | | | |
| Coil parameters | BY90A
Inner diameter - 34 mm
Outer diameter - 85 mm
N = 2 wings x 2 layers x 5 turns | Coil Cool-B70
Inner diameter - 35 mm
Outer diameter - 75 mm
N = 2 wings x 2 layers x 5 turns | Coil C-B60
Inner diameter - 35 mm
Outer diameter - 75 mm
Winding height - 12 mm
N = 2 wings x 2 layers x 5 turns | Equivalent, with no additional concerns for safety and effectiveness.
See Note A for equivalence explanation of coil parameters such as dimensions and windings. |
| | BF90A
Inner diameter - 34 mm
Outer diameter - 85 mm
N = 2 wings x 2 layers x 5 turns | | Coil Cool-B65
Inner diameter - 35 mm
Outer diameter - 75 mm
Winding height - 12 mm
N = 2 wings x 2 layers x 5 turns | |
| | BY90B
Inner diameter - 30 mm
Outer diameter -90 mm
N = 2 wings x 2 layers x 6 turns | | | |
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{9}
| Criteria | (Subject Device) Ultimate rTMS | (Primary Predicate) MagVita TMS Therapy System (K173620) | (Secondary Predicate) MagVita TMS Therapy System (K171481) | Equivalence Comments |
| --- | --- | --- | --- | --- |
| | | | | |
| Machine Output Parameters | | | | |
| Amplitude in Standard Motor Threshold (SMT) units | 0 – 2.4
41% Intensity Setting → 1 SMT | 0 – 2.1
47% Intensity Setting → 1 SMT | 0 - 1.7
58.8% Intensity Setting → 1 SMT | Equivalent, with no additional concerns for safety and effectiveness. See Note B for equivalence explanation of Intensity/Amplitude. |
| Waveform | Biphasic sinusoid | Biphasic sinusoid | Biphasic sinusoid | Identical waveform. |
| Active pulse width (μs) | 320 | 290 | 290 | Equivalent, with no additional concerns for safety and effectiveness. See Note C for equivalence explanation of Pulse Width and Amplitude. |
| Pulse amplitude (V) | BY90A: 1.35
BF90A: 1.26
BY90B: 1.22 | C-B70: 1.88 | C-B65: 1.84 | |
| Max magnetic field strength 2 cm from coil (T) | At 25% Intensity:
BY90A: 0.172
BF90A: 0.182
BY90B: 0.117
At 50% Intensity:
BY90A: 0.291 | At 25% Intensity:
C-B70: 0.144
At 50% Intensity:
C-B70: 0.283 | At 25% Intensity:
C-B65: 0.133
At 50% Intensity:
C-B65: 0.272 | Equivalent, with no additional concerns for safety and effectiveness. |
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{10}
| Criteria | (Subject Device) Ultimate rTMS | (Primary Predicate) MagVita TMS Therapy System (K173620) | (Secondary Predicate) MagVita TMS Therapy System (K171481) | Equivalence Comments |
| --- | --- | --- | --- | --- |
| | BF90A: 0.287
BY90B: 0.244
**At 75% Intensity:**
BY90A: 0.403
BF90A: 0.395
BY90B: 0.369
**At 100% Intensity:**
BY90A: 0.498
BF90A: 0.494
BY90B: 0.481 | **At 75% Intensity:**
C-B70: 0.418
**At 100% Intensity:**
C-B70: 0.549 | **At 75% Intensity:**
C-B65: 0.413
**At 100% Intensity:**
C-B65: 0.542 | See Note D for equivalence explanation of Max magnetic field strength. |
| Max initial dB/dt (kT/s) near the coil surface (z = 0 cm) | BY90A: 13.44
BF90A: 13.76
BY90B: 12.10 | C-B70: 13.27 | C-B65: 13.36 | Equivalent, with no additional concerns for safety and effectiveness. |
| Max initial dB/dt (kT/s) 2 cm from coil surface (z = 2 cm) | BY90A: 3.59
BF90A: 4.27
BY90B: 6.78 | C-B70: 5.25 | C-B65: 5.29 | See Note E for equivalence explanation of dB/dt. |
| Output Intensity Stability among Pulses in a Burst (Max Change %) | BY90A: 0.563
BF90A: 0.580
BY90B: 0.679 | C-B70: 0.615 | N/A | Equivalent, with no additional concerns for safety and effectiveness.
As established in the predicate device’s K173620, it is essential that Output Intensity of |
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{11}
| Criteria | (Subject Device) Ultimate rTMS | (Primary Predicate) MagVita TMS Therapy System (K173620) | (Secondary Predicate) MagVita TMS Therapy System (K171481) | Equivalence Comments |
| --- | --- | --- | --- | --- |
| | | | | pulses in a burst remain stable with max change of < 1%. |
| The system will automatically be disabled when the coil temperature exceeds: | 40 °C (104 °F) | 41 °C (106 °F) | 41 °C (104 °F) | Regardless of the intensity setting (at Maximum output or otherwise), both the subject and predicate device have a coil temperature safety feature that shuts down the system when the threshold temperature is reached. Ultimate rTMS system provides a slightly enhanced safety by not letting the coil temperature exceed 40 °C compared to 41 °C for MagVita TMS system. |
| Frequency range (Hz) | 0.1 - 100 | 0.1 - 30 or 0.1 - 100, depending on model | 0.1 - 100 | The differences in these parameters are simply the differences in the overall capabilities of these machines. These capabilities encompass the recommended |
| Pulse train duration range (s) | Rep Rate: 0.1 ...100Hz
Pulses in Train: 1,2,3,4 ... 2000
Train duration = Pulses in Train / Rep Rate | Rep Rate: 0.1 ...100Hz
Pulses in Train: 1,2,3,4 ... 1000
Train duration = Pulses in Train / Rep Rate | Rep Rate: 0.1 ...100Hz
Pulses in Train: 1,2,3,4 ... 1000
Train duration = Pulses in Train / Rep Rate | |
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{12}
| Criteria | (Subject Device) Ultimate rTMS | (Primary Predicate) MagVita TMS Therapy System (K173620) | (Secondary Predicate) MagVita TMS Therapy System (K171481) | Equivalence Comments |
| --- | --- | --- | --- | --- |
| Inter-train interval range (s) | 0 - 60 | 1 - 120 | 1 - 120 | treatment parameters for MDD listed above. In other words, these variations among different manufacturers’ models do not impact their ability to deliver the treatment parameters recommended for MDD. All machines use identical treatment parameters. |
| Maximum trains per session | 250 | 500 | 500 | |
| Maximum number of pulses per session | 500,000 = 2000 x 250
(Pulses In Train:2000
Maximum Trains:250 | 500,000 = 1000 x 500
(Pulses in Train:1000
Maximum Trains: 500 | 500,000 = 1000 x 500
(Pulses in Train:1000
Maximum Trains: 500 | |
| Standards | | | | |
| Electrical safety | Complies with IEC 60601-1, IEC 60601-1-2. | Complies with IEC 60601-1, IEC 60601-1-2. | Complies with IEC 60601-1, IEC 60601-1-2. | N/A |
| ISO Standards met | Company complies with ISO 13485:2016. ISO14971 | Company complies with ISO 13485:2012. | Company complies with ISO 13485:2012. | N/A |
Note A:
- Applicator Diameter: ID/OD = 30-34 mm/85 mm for subject and 35 mm/75 mm for predicate device. These differences are typical of variations among different manufacturers of TMS devices or even among different models of the same manufacturer. As shown in Section18 Appendix 18-1 Coils Test Report, they do not lead to significant differences in the electric and magnetic fields generated by these applicators.
- Applicator Windings: Both the subject and the predicate device have the exact same windings of 2 wings x 2 layers x 5 turns in the windings, as illustrated below:
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{13}

Coil BY90B has $2 \times 2 \times 6$ windings to better accommodate its geometric shape. These differences are typical of variations among different manufacturers of TMS devices or even among different models of the same manufacturer. As a comparison, a TMS device from another manufacturer (Rapid Therapy System, K143531) that is cleared for MDD treatment has a vastly different windings design of $3 \times 19 \times 2$ windings. As shown in Section18 Appendix 18-1 Coils Test Report, they do not lead to significant differences in the electrical and magnetic fields generated by these applicators.
Note B:
- Subject Device (Brain Ultimate TMS):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 41% → 1 SMT
- Intensity/Amplitude Setting = 100% → 2.4 SMT
- Predicate Device (MagVita TMS):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 47% → 1 SMT
- Intensity/Amplitude Setting = 100% → 2.1 SMT
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{14}
- These are comparable values and within the variations seen among previously cleared TMS devices by the FDA. These values are in slight favor of Brain Ultimate TMS as they indicate that the subject device will need a slightly lower intensity setting to achieve the same level of induced current as the predicate device.
# Note C:
The difference in pulse width (320 +/- 10% vs. 290 +/- 5%) between Brain Ultimate TMS and MagVita TMS is comparable to the differences in pulse widths of several TMS devices cleared by the FDA for MDD treatment, as shown in the table below:
| TMS Device | Pulse Width Spec (us) |
| --- | --- |
| Brain Ultimate | 320 +/- 10% (288-352) |
| MagVita | 290 +/- 5% (275-305) |
| NeuroStar | 185 +/- 10% (166-204) |
| Magstim | 330 +/- 10% (297-363) |
| Nexstim | 235 +/- 5 (230-240) |
In the above table:
- Subject = Brain Ultimate, Predicate = MagVita
Inherent differences in the design and construction of TMS devices of different manufacturers, or even different models of the same manufacturer, lead to the variations in pulse widths shown above. However, since a patient is always treated at $120\%$ MT for MDD on any TMS device, and MT is a function of both the intensity/amplitude and pulse width (higher pulse width will require lower intensity/amplitude to reach MT), differences in pulse widths are easily compensated by the changes in intensity settings needed to achieve $120\%$ MT stimulation.
Slightly higher pulse width of Ultimate rTMS compared to MagVita TMS would require slightly lower intensity/amplitude setting on Ultimate rTMS compared to Magvita TMS to achieve the same stimulation level. This is in alignment with Note B above that states that the subject device will need a slightly lower intensity setting to achieve the same level of induced current as the predicate device.
# Note D:
- The Max magnetic field values for the subject and predicate device at $2\mathrm{cm}$ are in a narrow range and exhibit equivalent slope and intercept values when measured at $25\%$ , $50\%$ , $75\%$ and $100\%$ intensity settings, as shown here:
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{15}

(See more details in Section18 Appendix 18-1 Coils Test Report, pages 7-8).
- These measurements were made using the same intensity setting on both devices. For a treatment given to the same patient, each device will be set to $120\%$ of the patient's MT, which does not translate to the exact same $\%$ power setting on each device. Different power setting on each device for the same MT will compensate for the observed differences in magnetic field strength values in the table.
- These differences are typical of variations among different manufacturers of TMS devices or even among different models of the same manufacturer.
# Note E:
- The dB/dt values for the subject and predicate device, at both $z = 0$ mm and $z = 20$ mm, are sufficiently close to each other to be considered equivalent. They were measured using the same intensity setting on both devices. For a treatment given to the same patient, each device will be set to $120\%$ of the patient's MT, which does not translate to the exact same % power setting on each device. Different power setting on each device for the same MT will compensate for the observed differences in dB/dt values in the table.
- These differences are typical of variations among different manufacturers of TMS devices or even among different models of the same manufacturer.
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
{16}
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The following special control is applicable to the subject device (Product Code: OBP) of this 510(k) submission:
"Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems."
The following performance tests specified in this guidance document were performed and the results were compared for a determination of substantial equivalence between the subject device (Ultimate rTMS System) and the predicate device (MagVita TMS System):
Electric Field:
a. SMT Comparison
b. Electrical Field Strength Decay over z-axis distance from coil center (in the direction of head center)
Magnetic Field:
a. Output Waveform for pulse width and amplitude
b. Magnetic Field Strength Linearity as a function of machine intensity setting
c. Magnetic Field Spatial Distribution
d. Magnetic Field Gradient (dB/dt)
Also, an additional test (Intensity Variation of Pulses within a Burst) not specified in the guidance document but essential for establishing substantial equivalence for iTBS protocol was performed.
Not Applicable
CONCLUSION STATEMENT
The Ultimate rTMS and the predicate device have identical intended use/indication for use, target population, treatment procedure, treatment position and all recommended standard treatment protocol parameters (intensity, frequency, number of pulses in a train, number of trains in a session, number of treatment sessions).
Both the subject device and the primary predicate device share the same Figure-of-eight coil design. The tested magnetic properties of the Ultimate rTMS and the primary predicate device are substantial equivalent for the coils. Variation in intensity of individual pulses in a burst are equivalent for the subject and predicate device as both show less than 1% variation.
The reliability of the positioning method used by the Ultimate rTMS is based on the direct relationship of the underlying cortical brain anatomy to the patient's scalp, as is the method used in the predicate device. The method for identifying the correct treatment position in the Ultimate rTMS is at least as effective as the method employed by the predicate device.
The Ultimate rTMS does not introduce any new safety considerations in comparison to the predicate device. All other identified differences between the subject device and the predicate device are minor and without any known impact on safety or efficacy.
Based on the information and supporting documentation provided in the premarket notification, the Ultimate rTMS is substantially equivalent to the cited primary predicate device. Testing demonstrates that the Ultimate rTMS fulfills prospectively defined design and performance specifications.
Brain Ultimate, Inc., Traditional 510 (k)
Ultimate rTMS
Page 13 of 13
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.