K243429 · Hjy Smart Medical Device Co., Ltd. · GWG · May 21, 2025 · Neurology
Device Facts
Record ID
K243429
Device Name
HJY VisualNext 3D Endoscopic Vision System
Applicant
Hjy Smart Medical Device Co., Ltd.
Product Code
GWG · Neurology
Decision Date
May 21, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1480
Device Class
Class 2
Indications for Use
HJY VisualNext 3D Endoscopic Vision System is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of structures within the brain during neurological surgical procedures as well as for viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy.
Device Story
System captures internal surgical site images via rigid endoscope with dual CMOS sensors and integrated LED illumination; transmits data via 5m BNC coaxial cable to Endoscope Control Unit (ECU) for processing; outputs 3D/2D video to external monitor. Used in OR by surgeons for visualization during neurosurgical and spinal procedures. Provides enhanced depth perception via 3D imaging to assist surgical navigation and decision-making. Images are recordable/markable for analysis. ECU is standalone, non-networked.
Clinical Evidence
No human clinical data. Bench testing evaluated image quality (FOV, DOF, MTF, SNR) on aged/non-aged devices. Porcine animal study compared 3D visualization performance against predicate 2D visualization, confirming clear/stable imaging and compatibility with external 3D monitors. All parameters met predefined acceptance criteria.
Technological Characteristics
Rigid endoscope with dual CMOS sensors, 2 integrated LEDs, 7.0mm outer diameter, 180mm working length. 0° direction of view, 120° FOV, 5-100mm depth of field. Connectivity: 5m BNC coaxial cable to ECU. Standalone system, non-networked. Single-use. Sterilization per FDA guidance. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18. Photobiological safety per IEC 62471.
Indications for Use
Indicated for patients undergoing general surgical, neurological, or spinal procedures (e.g., nucleotomy, discectomy, foraminotomy) requiring visualization of internal surgical sites or brain structures.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 21, 2025
HJY Smart Medical Device Co., Ltd.
John Chen, Ph.D.
Chairman & CEO
12F., No. 415, Sec. 4, Xinyi Rd., Xinyi Dist.,
Taipei, TW 11051
Taiwan
Re: K243429
Trade/Device Name: HJY VisualNext 3D Endoscopic Vision System (HDSES01 / HDSES301)
Regulation Number: 21 CFR 882.1480
Regulation Name: Neurological Endoscope
Regulatory Class: Class II
Product Code: GWG
Dated: April 21, 2025
Received: April 21, 2025
Dear Dr. John Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243429 - John Chen, Ph.D.
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243429 - John Chen, Ph.D.
Page 3
Sincerely,
Adam D. Pierce -S
Digitally signed by Adam D. Pierce -S
Date: 2025.05.21
14:20:13 -04'00'
Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243429 | |
| Device Name HJY VisualNext 3D Endoscopic Vision System | |
| Indications for Use (Describe) HJY VisualNext 3D Endoscopic Vision System is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of structures within the brain during neurological surgical procedures as well as for viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K243429 Page 1 of 5
# 510(k) Summary
1 Date of summary: May 21, 2025
2 Submitter: HJY Smart Medical Device Co., Ltd.
Address: 12F., No. 415, Sec. 4, Xinyi Rd., Xinyi Dist., Taipei City 11051, Taiwan
Phone: +886-933-869246
Fax: +886-2-2298-3287
Contact: John Jiannyuh Chen, M.S., Ph.D. (john.chen@hjy-med.com)
3 Identification of the device:
Proprietary/Trade name: HJY VisualNext 3D Endoscopic Vision System
Product code: GWG
Regulation number: 882.1480
Regulation description: Endoscope, Neurological
Device class: II
4 Identification of the predicate device:
Predicate device name: HJY VisualNext Endoscopic Vision System
Manufacturer: HJY Smart Medical Device Co., Ltd.
Product code: GWG
Regulation number: 882.1480
Device class: II
510(k) number K222735
5 Indications for Use
HJY VisualNext 3D Endoscopic Vision System is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of structures within the brain during neurological surgical procedures as well as for viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy.
{5}
K243429 Page 2 of 5
# 6 Device Description
The HJY VisualNext 3D Endoscopic Vision System is a system used for viewing internal surgical sites during neurosurgical procedures. The system consists of the following components:
- Endoscope Control Unit (ECU) (Model number: HDSES01)
- Endoscope (Model number: HDSE301)
The endoscope is physically connected via a 5m BNC coaxial cable to the Endoscope Control Unit (ECU). The Endoscope consists of 2 LED lamps and a CMOS camera with 2 sensors, embedded in the proximal end of a rigid metal arthroscope, which captures the image and transmits to and is processed by the Endoscope Control Unit (ECU), subsequently output to and presented on an external monitor. Images are recordable and markable for further analysis. The Endoscope Control Unit (ECU) is not connectable to intranet or internet.
# 7 Comparison of Technological Characteristics
| Item | Subject device | Predicate device | Comment |
| --- | --- | --- | --- |
| Proprietary name | HJY VisualNext 3D Endoscopic Vision System | HJY VisualNext Endoscopic Vision System | |
| 510(k) No. | K243429 | K222735 | |
| Indications for Use | HJY VisualNext 3D Endoscopic Vision System is intended for Viewing internal surgical sites during general surgical procedures and for use in visualization of structures within the brain during neurological surgical procedures as well as for viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy. | HJY VisualNext Endoscopic Vision System is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of structures within the brain during neurological surgical procedures as well as for viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy. | Same |
| Type of use | Prescription Use | Prescription Use | Same |
| System components | Rigid endoscope, ECU | Rigid endoscope, ECU | Same |
| Light transmission | Light emitted from integrated LED at proximal end of endoscope insertion tube | Light emitted from integrated LED at proximal end of endoscope insertion tube | Same |
| Light source | Integrated LED | Integrated LED | Same |
| Image | Rigid rod lenses + CMOS imaging sensor in endoscope main body | Rigid rod lenses + CMOS imaging sensor in endoscope main body | Same |
{6}
K243429 Page 3 of 5
| transmission | | | |
| --- | --- | --- | --- |
| Direction of view | 0° | 0° | Same |
| Field of view | 120° | 120° | Same |
| Depth of field | 5-100 mm | 5-100 mm | Same |
| Image resolution | Optical resolution: 800x800 pixels (HD imager)
TV Lines: 667 LW/PH @ 15% MTF | Optical resolution: 800x800 pixels (HD imager)
TV Lines: 667 LW/PH @ 15% MTF | Same |
| Image display | External monitor | External monitor | Same |
| 2D / 3D Imaging | 2D & 3D | 2D only | Different
The subject device has 2 sensors while the predicate device has only 1 sensor. |
| Recording attribute | Via USB-port | Via USB-port | Same |
| Insertion tube working length | 180 mm | 180 mm | Same |
| Insertion tube outer diameter | 7.0 mm | 5.0 mm | Different
The subject device is wider than the predicate device. |
| Single use or Reusable | Single use | Single use | Same |
# 8 Non-Clinical Testing
A series of performance tests were performed to assess the safety and effectiveness of HJY VisualNext 3D Endoscopic Vision System. All test results demonstrate that subject device meets the requirements of its pre-defined acceptance criteria and intended use.
- Sterility testing was completed in accordance with the FDA guidance document “Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile”.
- Biocompatibility testing – The endoscope (Model number: HDSE301) is categorized as an externally communicating device with limited contact duration (<24 hours) with neural tissue/bone/dentin and blood (indirect contact through reabsorption of CSF into the venous system) in accordance with the FDA guidance document “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.” The following biocompatibility endpoints were assessed in accordance with the methods specified in the FDA
{7}
K243429 Page 4 of 5
biocompatibility guidance: cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, neurotoxicity, and hemocompatibility (indirect hemolysis). Test results demonstrated that all acceptance criteria were met, supporting a favorable biocompatibility profile for the subject device.
- Software validation (Enhanced Documentation Level) was completed in accordance with requirements specified in the FDA guidance document "Content of Premarket Submissions for Device Software Functions"
- Electromagnetic compatibility and thermal safety testing were completed in accordance with IEC60601-1, IEC60601-1-2 and IEC60601-2-18
- Photobiological safety testing of the lamp was completed in accordance with IEC 62471.
- Non-clinical performance testing was conducted to evaluate image quality and establish substantial equivalence to the predicate device throughout the subject device's entire shelf life. This comprehensive assessment included the following tests:
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Bench Testing | The study assessed image quality and overall performance of both aged and non- aged subject devices with direct comparison to the predicate device to ensure long-term stability and comparability to the predicate device.
Key parameters evaluated included:
• Field of View (FOV)
• Direction of View (DOV)
• Depth of Field (DOF)
• Optical Magnification
• Distortion
• Image Intensity Uniformity
• Signal-to-Noise Ratio
• Sensitivity
• Resolution (MTF - Modulation Transfer Function). | Pass
Both aged and non-aged subject devices met the predefined acceptance criteria, demonstrating consistent image quality metrics comparable to the predicate device. |
{8}
K243429 Page 5 of 5
| Animal Study Testing | The study was conducted in a porcine animal model to evaluate the 3D visualization performance of the subject device compared to the 2D visualization of the predicate device in brain and spine procedures. Testing was performed with aged and non-aged subject devices and the predicate device to assess image quality to the following parameters: • Resolution • Illumination • Color Representation • Contrast • Noise The subject device was connected to the Sony LMD-2451MT 3D Monitor (K113203) to assess 3D visualization performance in comparison to the predicate device (2D visualization). | Pass |
| --- | --- | --- |
| | | The subject device provided clear and stable 3D visualization of brain and spine tissues across all tested conditions. Image quality parameters, including resolution, color representation, contrast, and noise, met the predefined acceptance criteria when compared to the predicate device. Testing also validated compatibility with the Sony LMD-2451MT 3D Monitor. |
| | | |
# 9 Conclusion
The results of non-clinical testing, demonstrate that the technological differences, between the subject and predicate device, do not raise different questions of safety or effectiveness. Based on the demonstrated identical intended use, similar technological characteristics and the evidence from all the performance testing, the subject VisualNext 3D Endoscopic Vision System is substantially equivalent to the predicate device HJY VisualNext Endoscopic Vision System (K222735).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.