Guided DAS Surgical Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Guided DAS Surgical Kit is intended to allow sterilization of the enclosed medical devices. Guided DAS Surgical Kit requires the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kit is to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the following cycle: dynamic air removal steam sterilization- Exposure at 132 °C for 4 minutes and 30 minutes dry time. Guided DAS Surgical Kit is intended for sterilization of non-porous loads. Guided DAS Surgical Kit is recommended not to be stacked during sterilization. The combined weight of the Guided DAS Surgical Kit and the associated instruments is 760 grams. The weight of the empty Guided DAS Surgical Kit is 600 grams.
Device Story
Guided DAS Surgical Kit is a reusable instrument sterilization tray system designed to organize and protect dental surgical drills and tools during steam sterilization. The device consists of a rigid, perforated base, inner tray, and lid made of polyphenylsulfone (PPSU) with silicone instrument holders. The tray is loaded with instruments, enclosed in an FDA-cleared sterilization wrap, and processed in a healthcare facility using dynamic air removal steam sterilization. The perforated design allows sterilant penetration. Healthcare providers use the tray to maintain instrument organization and sterility during the sterilization cycle and subsequent storage. The device is intended for non-porous loads and is not to be stacked. It provides a protective, reusable container that facilitates effective sterilization and safe handling of surgical instruments.
Clinical Evidence
No clinical data were included in this submission. Evidence is based on bench testing, including manual cleaning validation, sterilization validation (sterilant penetration and dry time), simulated use-life validation (50 cycles), and biocompatibility (cytotoxicity per ISO 10993-5).
Technological Characteristics
Materials: Polyphenylsulfone (PPSU) base, lid, and tray; medical-grade silicone holders. Design: Perforated rigid polymer tray. Sterilization: Moist heat (steam) via dynamic air removal cycle (132 °C, 4 min exposure, 30 min dry). Connectivity: None. Reusable: Up to 50 cycles.
Indications for Use
Indicated for use by healthcare providers to enclose and sterilize non-porous medical devices via dynamic air removal steam sterilization (132 °C for 4 minutes, 30 minutes dry time). Requires use of an FDA-cleared sterilization wrap. Not for stacked sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 22, 2025
Talladium España, SL
% Rebecca Kattan
Regulatory Specialist
PaxMed International, LLC
12264 El Camino Real
Suite 400
San Diego, California 92130
Re: K243425
Trade/Device Name: Guided DAS Surgical Kit
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap
Regulatory Class: Class II
Product Code: KCT
Dated: June 3, 2025
Received: June 3, 2025
Dear Rebecca Kattan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K243425 - Rebecca Kattan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K243425 - Rebecca Kattan
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen A. Anisko -S
Digitally signed by
Stephen A. Anisko -S
Date: 2025.07.22
16:53:22 -04'00'
Stephen Anisko
Acting Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243425
Device Name
Guided DAS Surgical Kit
Indications for Use (Describe)
Guided DAS Surgical Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Guided DAS Surgical Kit is intended to allow sterilization of the enclosed medical devices.
Guided DAS Surgical Kit requires the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.
The kit is to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the following cycle: dynamic air removal steam sterilization- Exposure at 132 °C for 4 minutes and 30 minutes dry time.
Guided DAS Surgical Kit is intended for sterilization of non-porous loads.
Guided DAS Surgical Kit is recommended not to be stacked during sterilization.
The combined weight of the Guided DAS Surgical Kit and the associated instruments is 760 grams.
The weight of the empty Guided DAS Surgical Kit is 600 grams.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{5}
510(k) Summary
Guided DAS Surgical Kit
# 510(k) Summary
K243425
Talladium España, SL
Guided DAS Surgical Kit
July 22, 2025
In accordance with 21 CFR 807.92 the following summary of information is provided:
1. ADMINISTRATIVE INFORMATION
Manufacturer Name
Talladium España, SL
Virginia Woolf, 17
Lleida, Lleida, ES 25005
Telephone +34 973-289-580
Official Contact
Xavier Soca Filella, General Manager
Representative/Consultant
Rebecca E. Kattan, PhD
Kevin A. Thomas, PhD
Floyd G. Larson, MS, MBA
PaxMed International, LLC
12264 El Camino Real, Suite 400
San Diego, CA 92130
Telephone: +1 858-792-1235
Fax: +1 858-792-1236
Email: rkattan@paxmed.com
kthomas@paxmed.com
flarson@paxmed.com
2. DEVICE NAME AND CLASSIFICATION
Trade/Device Name
Guided DAS Surgical Kit
Common Name
Instrument sterilization trays
Regulation Number
21 CFR 880.6850
Regulation Name
Sterilization Wrap Containers, Trays, Cassettes & Other
Accessories
Regulatory Class
Class II
Product Code
KCT
Classification Panel
General Hospital
Reviewing Office
Office of Health Technology 4
(OHT 4: Surgical and Infection Control Devices)
Reviewing Division
Division of Health Technology 4 C (Infection Control Devices)
3. PREDICATE DEVICE INFORMATION
Primary predicate device is:
K201878, ST-Z5, Z-Systems AG.
K243425
{6}
510(k) Summary
Guided DAS Surgical Kit
# 4. INDICATIONS FOR USE STATEMENT
Guided DAS Surgical Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Guided DAS Surgical Kit is intended to allow sterilization of the enclosed medical devices.
Guided DAS Surgical Kit requires the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.
The kit is to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the following cycle: dynamic air removal steam sterilization-Exposure at 132 °C for 4 minutes and 30 minutes dry time.
Guided DAS Surgical Kit is intended for sterilization of non-porous loads.
Guided DAS Surgical Kit is recommended not to be stacked during sterilization.
The combined weight of the Guided DAS Surgical Kit and the associated instruments is 760 grams.
The weight of the empty Guided DAS Surgical Kit is 600 grams.
# 5. DEVICE DESCRIPTION
The subject device tray is to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit (the complete tray) which contains and protects the interior components during sterilization. The lid latches to the base to secure the tray into a single unit. The base, inner tray, and lid components are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam) and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polyphenylsulfone (PPSU), and holders of various geometries to position instruments in the kits are manufactured from silicone.
The subject device comes in one (1) size and one (1) configuration, while the predicate device K201878 is provided in one (1) size and two (2) configurations. The subject device and the predicate device have similar overall dimensions, enclose similar volumes, and have similar vent to volume ratios. Differences in the dimensions and vent to volume ratios between the subject device and the predicate device are mitigated by the sterilization validation performed.
# 6. TECHNOLOGICAL CHARACTERISTICS COMPARISON
The subject device is compared in intended use to the predicate device K201878. The subject device and the predicate device K201878 have the same intended use, the same product classification and product code (KCT) and have similar Indications for Use statements. Components of the subject device and the predicate device K201878 are perforated to allow for penetration of the sterilant, are to be used with the same moist heat (steam) sterilization cycle, and require the use of an FDA cleared wrap to maintain sterility.
The basis for the belief of Talladium España, SL that the subject device compares favorably to the predicate device is summarized in the following Table of Technological Characteristics Comparison.
K243425
{7}
510(k) Summary
Guided DAS Surgical Kit
Table of Technological Characteristics Comparison
| Attribute | Subject Device | Primary Predicate Device | Comparison |
| --- | --- | --- | --- |
| | K243425
Guided DAS Surgical Kit
Talladium España, SL | K201878
ST-Z5
Z-Systems AG | |
| Indications for Use Statement | Guided DAS Surgical Kit is intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Guided DAS Surgical Kit is intended to allow sterilization of the enclosed medical devices.
Guided DAS Surgical Kit requires the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.
The kit is to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the following cycle: dynamic air removal steam sterilization- Exposure at 132 °C for 4 minutes and 30 minutes dry time.
Guided DAS Surgical Kit is intended for sterilization of non-porous loads.
Guided DAS Surgical Kit is recommended not to be stacked during sterilization.
The combined weight of the Guided DAS Surgical Kit and the associated instruments is 760 grams.
The weight of the empty Guided DAS Surgical Kit is 600 grams. | ST-Z5 surgical trays are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. ST-Z5 surgical trays are intended to allow sterilization of the enclosed medical devices.
ST-Z5 surgical trays require the use of an FDA cleared wrap to maintain the sterility of the enclosed devices.
The trays are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the following cycle:
Pre-vacuum steam: 132 °C (270 °F) for 4 minutes with 20 minutes drying time.
ST-Z5 surgical trays are intended for sterilization of non-porous loads.
ST-Z5 surgical trays are recommended not to be stacked during sterilization.
The combined weight of the ST-Z5_c_m_c(t)_m(t) tray and the associated instruments is 441 grams. The weight of the empty ST-Z5_c_m_c(t)_m(t) tray is 361 grams.
The combined weight of the Z5-BL/TL tray and the associated instruments is 427 grams. The weight of the empty Z5-BL/TL tray is 363 grams. | Similar |
| Intended Use | This tray is designed to hold various dental surgical drills and tools to organize and protect the instruments that are sterilized in the trays by the healthcare provider. | This tray is designed to hold various dental surgical drills and tools to organize and protect the instruments that are sterilized in the trays by the healthcare provider. | Same |
| Product Code | KCT | KCT | Same |
| Design | Rigid polymer base, lid, and removable inner tray | Rigid polymer base, lid, and removable inner tray | Same |
| Materials | Polyphenylsulfone (Radel®) [lid, base, tray]
Medical grade silicone [grommets/holders] | Polyphenylsulfone (Radel® R-5000) [lid, base, tray] Medical grade silicone [grommets/holders] | Same |
| Materials compatible with Sterilization Method | Yes | Yes | Same |
| Perforated | Yes; allows moist heat (steam) penetration to achieve sterilization | Yes; allows moist heat (steam) penetration to achieve sterilization | Same |
| Reusable | Yes | Yes | Same |
| Number of Overall Sizes | 1 | 1 | Same |
| Number of Configurations | 1 | 2 | Similar |
| Number of uses | 50 | 101 | Similar |
| Overall Dimensions | 264.67 mm Length x 145.85 mm Width x 61.02 mm Height | 185.1 mm Length x 133.6 mm Width x 61.5 mm Height | Similar |
| Vent to Volume Ratio | 0.01233 cm³/cm³ (0.001233 mm³/mm³) | 0.01596 cm³/cm³ | Similar |
| Reusable | Yes | Yes | Same |
| Use Life Testing | Reusable up to 50 cycles
Assembled, sterilized
Visual inspection
Component dimensional fit verification Functional closure (lid-base latch) verification | Reusable up to 101 cycles
Assembled, sterilized
Visual inspection
Component dimensional fit verification Functional closure (lid-base latch) verification | Similar |
| Sterilization Method | | | |
| Sterilant | Moist heat (steam) | Moist heat (steam) | Same |
| Cycles | Fractionated vacuum (pre-vacuum)
Exposure at 132 °C (270 °F) for 4 minutes with 30 minutes drying time. | Fractionated vacuum (pre-vacuum)
Exposure at 132 °C (270 °F) for 4 minutes with 20 minutes drying time. | Similar |
| Sterile Barrier | Sterilization wrap, FDA cleared for indicated method and cycle | Sterilization wrap, FDA cleared for indicated method and cycle | Same |
K243425
{8}
510(k) Summary
Guided DAS Surgical Kit
# 7.0 SUMMARY OF NONCLINICAL TESTING
Provided below are the nonclinical test methodologies performed to demonstrate that the subject device meets the acceptance criteria of the standard.
Summary of Nonclinical Testing Table
| Test Methodology | Purpose | Results |
| --- | --- | --- |
| Manual Cleaning Validation FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (issued March 2015) | The purpose of this test is to validate that the cleaning instructions provided in the Instructions for Use appropriately clean the tray, and to ensure the sterilization cycle will be effective. | Pass |
| Sterilization Validation including sterilant penetration and dry time validation ANSI/AAMI/ISO 17665-1 ANSI/AAMI/ISO 17665-2 | The purpose of this test is to validate that the sterilization instructions listed in the Instructions for Use appropriately sterilize the tray and contents. | Pass |
| Dry time | The purpose of this test is to validate that the sterilization instructions listed in the Instructions for Use appropriately dry the wrapped tray for storage. | Pass |
| Life Cycle / Simulated Use-life Validation FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (issued March 2015) | The purpose of this test is to validate the service life of the trays as stated in the Instructions for Use. | Pass |
| Biocompatibility of Subject Device Cytotoxicity testing ANSI/AAMI/ISO 10993-5 ANSI/AAMI/ISO 10993-12 | The purpose of this test is to evaluate the cytotoxicity potential of the test article using an in vitro cell culture assay. | Pass |
In summary, the nonclinical testing provided for these devices met the acceptance criteria for each standard and test methodology used to evaluate the devices as shown in the table above.
No clinical data were included in this submission.
# 8.0 CONCLUSION
The conclusions drawn from the nonclinical tests demonstrate that the subject device in this 510(k) submission, Guided DAS Surgical Kit, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K201878.
K243425
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.