Ventris anterior cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris anterior cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris 2-screw anterior cages may be used as a stand alone device only when two (2) vertebral body bone screws are used. Ventris 4-screw anterior cages may be used as a stand alone device only when at least two (2) vertebral body bone screws are inserted in the two medial fixation holes with one inferior and one superior screw trajectory. If the physician chooses to use Ventris anterior cages with fewer than two (2) screws in the two medial fixation holes with one inferior and one superior screw trajectory, then an additional supplemental spinal fixation system cleared for use in the lumbosacral spine must be used. Ventris anterior cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. When used to treat ASD, or when cages with more than 20° of lordosis are used, Ventris anterior cages must be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine and cannot be used as a stand alone. Ventris anterolateral cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris anterolateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. Ventris anterolateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris anterolateral cages are intended to be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine. Ventris lateral cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris lateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. Ventris lateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris lateral cages are intended to be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine.
Device Story
Ventris system comprises titanium alloy intervertebral body fusion devices for ALIF, AOLIF, and LLIF surgical approaches. Devices support spinal stability and fusion when used with autograft or allogenic bone graft. Components are manufactured via laser powder bed fusion (L-PBF) to create solid structures with porous lattice surfaces for bone in-growth; optional hydroxyapatite (HA) coating enhances osseointegration. Devices are secured to vertebral bodies using bone screws and cover plate assemblies to prevent screw back-out. Used in clinical settings by surgeons; devices are provided sterile. Output is a stabilized intervertebral space facilitating fusion. Benefits include improved spinal stability and potential for reduced vessel interference due to zero-profile design in anterior/anterolateral constructs.
Clinical Evidence
Bench testing only. Mechanical performance validated via static/dynamic compression and compression-shear testing (ASTM F2077) and subsidence testing (ASTM F2267). Additional testing included tensile, microstructure, and chemical composition assessment (ASTM F3001). No clinical data provided.
Indicated for skeletally mature patients with DDD (back pain, disc degeneration) and up to Grade I spondylolisthesis, or severe symptomatic ASD (ODI >40), requiring intervertebral body fusion at L2-S1 (1-2 levels). Requires supplemental fixation for ASD or lordosis >20°. Contraindicated if non-operative treatment <6 months.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Osseus Fusion Systems Pisces-SA Standalone ALIF Interbody S (K223413)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 22, 2025
Acuity Surgical Devices LLC
Mr. Chuck Forton
Director of Engineering and Regulatory Affairs
8710 N Royal Lane
Irving, Texas 75063
Re: K243386
Trade/Device Name: Ventris IBFD
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: OVD, MAX
Dated: March 21, 2025
Received: March 21, 2025
Dear Mr. Forton:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243386 – Mr. Chuck Forton
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243386 – Mr. Chuck Forton
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243386
Device Name
Ventris (Various PNs)
Indications for Use (Describe)
{4}
Ventris anterior cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris anterior cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Ventris 2-screw anterior cages may be used as a stand alone device only when two (2) vertebral body bone screws are used. Ventris 4-screw anterior cages may be used as a stand alone device only when at least two (2) vertebral body bone screws are inserted in the two medial fixation holes with one inferior and one superior screw trajectory. If the physician chooses to use Ventris anterior cages with fewer than two (2) screws in the two medial fixation holes with one inferior and one superior screw trajectory, then an additional supplemental spinal fixation system cleared for use in the lumbosacral spine must be used. Ventris anterior cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. When used to treat ASD, or when cages with more than 20° of lordosis are used, Ventris anterior cages must be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine and cannot be used as a stand alone.
Ventris anterolateral cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris anterolateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. Ventris anterolateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris anterolateral cages are intended to be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine.
Ventris lateral cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris lateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. Ventris lateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris lateral cages are intended to be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}
| 510(k) #: K243386 | 510(k) Summary | Prepared on: 2025-03-21 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Acuity Surgical Devices LLC | |
| Applicant Address | 8710 N Royal Lane Irving TX 75063 United States | |
| Applicant Contact Telephone | 512-585-3537 | |
| Applicant Contact | Mr. Chuck Forton | |
| Applicant Contact Email | cforton@acuitysurgical.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Ventris (Various PNs) | |
| Common Name | Intervertebral body fusion device | |
| Classification Name | Intervertebral Fusion Device With Integrated Fixation, Lumbar | |
| Regulation Number | 888.3080 | |
| Product Code(s) | OVD, MAX | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K201671 | A-Link Z | OVD |
| K222561 | Align | OVD |
| K232282 | aprevo® anterior and lateral lumbar interbody fusion device; apre | MAX |
| K230894 | NuVasive Modulus ALIF System | OVD |
| K223413 | Osseus Fusion Systems Pisces-SA Standalone ALIF Interbody S | OVD |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
{7}
The Ventris system consists of intervertebral body fusion devices intended for lumbar interbody fusion using an anterior lumbar interbody fusion surgical approach (ALIF), anterolateral (i.e., oblique) lumbar interbody fusion surgical approach (AOLIF), or a lateral lumbar interbody fusion surgical approach (LLIF). The devices are intended to improve stability of the spine while supporting fusion. The Ventris constructs are intended for use at one or two contiguous levels in the lumbar spine (L2-S1). The components are offered in different shapes and sizes to meet the requirements of the individual patient's anatomy and are provided sterile.
Ventris cages are available in six configurations: Ventris Ti Interbody anterior four-hole constructs for ALIF approach, Ventris Ti FRA Interbody anterior fully round ALIF (FRA) constructs for ALIF approach, Ventris Ti Open Interbody anterior two-hole constructs for ALIF approach, Ventris TI AL Interbody anterolateral (i.e., oblique) constructs for OLIF approach, Ventris TI Lateral FX two-screw lateral constructs for LLIF approach, and Ventris TI Lateral Interbody lateral constructs for LLIF approach. All cages are also available in increased surface area options that provide additional endplate surface area. For the increased surface area cages, the outer footprint remains the same, but the volume of the internal graft window is reduced, creating more endplate surface area.
Ventris cages are secured on the vertebral bodies using bone screws. A cover plate assembly prevents the screws from backing out after insertion. The cages and cover plates are made of titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) with Powder Bed Fusion. The bone screws and cover plate screws are made from titanium alloy (Ti-6Al-4V ELI) per ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications. All anterior and anterolateral constructs are zero profile, reducing potential for vessel interference with the anterior column.
All Ventris cages and cover plates are manufactured using the L-PBF (laser powder bed fusion) additive manufacturing method. L-PBF allows for the formation of solid, non-porous cages with a layered porous lattice structure on the surfaces of the components, including the surfaces of the interior graft window. This intricate structure facilitates bone in-growth by providing a larger surface of implant/bone contact than a buffed surface. Interbody cages and bone screws may also be titanium anodized to allow for identification of various heights/sizes by color.
Ventris Ti Interbody cages, Ventris Ti FRA Interbody cages, Ventris Ti Open Interbody cages, Ventris Ti AL Interbody cages, Ventris Ti Lateral FX Interbody cages, Ventris Ti Lateral Interbody cages, and Ventris bone screws are also available with a Promimic HAnano Surface (hydroxyapatite (HA) coating). Cages and bone screws are coated with a 20nm HA layer composed of crystalline hydroxyapatite particles that mimic human bone tissue through shape, composition, and structure. This surface treatment increases implant anchoring by facilitating osseointegration and enhancing early bone growth. All Ventris devices are only available sterile packaged.
Non-sterile, reusable surgical instruments to support implantation of the system are provided for use with Ventris devices are provided in steam sterilization trays.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
{8}
Ventris anterior cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris anterior cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Ventris 2-screw anterior cages may be used as a stand alone device only when two (2) vertebral body bone screws are used. Ventris 4-screw anterior cages may be used as a stand alone device only when at least two (2) vertebral body bone screws are inserted in the two medial fixation holes with one inferior and one superior screw trajectory. If the physician chooses to use Ventris anterior cages with fewer than two (2) screws in the two medial fixation holes with one inferior and one superior screw trajectory, then an additional supplemental spinal fixation system cleared for use in the lumbosacral spine must be used. Ventris anterior cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. When used to treat ASD, or when cages with more than 20° of lordosis are used, Ventris anterior cages must be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine and cannot be used as a stand alone.
Ventris anterolateral cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris anterolateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. Ventris anterolateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris anterolateral cages are intended to be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine.
Ventris lateral cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris lateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. Ventris lateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris lateral cages are intended to be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine.
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| --- | --- |
| Ventris has similar indications for use as the devices cleared in K201671, K222561, K232282, K230894, and K223413 The difference in indications for use between the subject device and the primary predicate device (K201671) is the addition of indications for treatment of severe symptomatic adult spinal deformity (ASD) conditions, and the usage of cages with more than 20° of lordosis, with both new indications requiring the use of a supplemental spinal fixation system cleared for use in the lumbosacral spine. | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
{9}
The Ventris system has similar indications for use, design and dimensions, and uses identical materials as the devices cleared in primary predicate device K201671. The subject device also has identical or similar technological characteristics as the primary predicate device.
Differences in technological characteristics from the predicate devices include introduce a change in the device name, additional implant height and lordosis offerings, additional implant configurations with increased surface area options, additional bone screw and cover plate options, an update to instrument designs, and additional instrument options. Furthermore, the additive manufacturing process was changed to a new supplier.
The subject device has identical or similar technological characteristics as the reference predicate devices cleared in K222561, K230894, K232282, and K223413 with an extended range of implant sizes and lordosis. The subject device also has similar indications for use as the devices cleared in K230894, K232282, and K223413. The subject device comes with an optional HA coating, which is the same material as the device cleared in K222561.
In order to ensure that these technological characteristics do not affect the safety and effectiveness of the subject device, mechanical testing and a worst-case analysis was conducted on the new implant sizes and configurations. Based on the testing conducted, including mechanical performance testing, design validation analysis, biocompatibility testing, and sterilization and packaging validation, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices K201671, K222561, K230894, K232282, and K223413.
The Ventris system is substantially equivalent to and is as safe and effective as the legally marketed predicate device, A-Link Z (K201671) under regulation 21 CFR 888.3080, product codes OVD and MAX.
## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The subject device underwent mechanical performance testing to validate that the performance of the worst-case subject device is substantially equivalent to the previously cleared predicate devices. Worst-case testing also took into account the effect of all process steps, including HIP and Non-HIP test articles. Test methods include static compression, static compression shear, dynamic compression, and dynamic compression shear per ASTM F2077 Test Methods for Intervertebral Body Fusion Devices, and subsidence testing per ASTM F2267 Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Devices Under Static Axial Compression. Mechanical testing ensured that the design features of the subject device met the required mechanical strength criteria for their intended use. Additional testing included tensile testing, microstructure assessment, and chemical composition assessment per ASTM F3001. The performance testing demonstrated substantial equivalence between the subject and predicate devices. The results of the non-clinical testing did not identify any new or increased risks associated with the change in additive manufacturing supplier. Performance equivalence demonstrated that the subject device met the acceptance criteria of the standards and is substantially equivalent to the predicate devices.
Ventris devices are substantially equivalent to the previously cleared devices with respect to their indications for use, design, function, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.