The WiTOF is a neuromuscular transmission station (nerve stimulator) that helps to pilot the neuromuscular blockade of adult patients in the operating room, recovery room or intensive care unit.
Device Story
WiTOF is a neuromuscular transmission (NMT) monitor for adult patients in OR, recovery, or ICU. Operated by anesthesiologists or nurse anesthetists. System includes a station and a wireless 3D accelerometer sensor (thumb splint). Device delivers electrical nerve stimulation (constant current, 0-60 mA, 200μsec pulse width) and measures resulting muscle contraction via the accelerometer. Supports multiple stimulation modes: TOF, PTC, ATP, DBS, ST, and TET. Station displays stimulation settings, elapsed time, and measurement results (e.g., TOF, PTC, T4/T1, T4/Tref). Clinicians use output to assess effectiveness of neuromuscular blocking agents or patient curare levels to guide clinical decision-making regarding blockade management.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1, IEC 60601-2-10), EMC (IEC 60601-1-2), and software verification and validation.
Technological Characteristics
Acceleromyography-based NMT monitor. Components: station, wireless 3D accelerometer sensor, AC power supply. Stimulation: constant current 0-60 mA, monophasic 200μsec pulse width, max 300V. Connectivity: wireless sensor link. Biocompatibility: thumb splint (intact skin, <24h). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-10, ISO 10993-1.
Indications for Use
Indicated for neuromuscular transmission monitoring in adult patients receiving neuromuscular blocking agents in operating rooms, recovery rooms, or intensive care units. No known contraindications.
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 23, 2025
Idmed
Perrine Moelle
Quality & Regulatory Affairs Director
3 rue John Maynard Keynes
Marseille, 13013
France
Re: K243339
Trade/Device Name: WiTOF
Regulation Number: 21 CFR 868.2775
Regulation Name: Electrical Peripheral Nerve Stimulator
Regulatory Class: Class II
Product Code: KOI
Dated: May 21, 2025
Received: May 21, 2025
Dear Perrine Moelle:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K243339 - Perrine Moelle
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K243339 - Perrine Moelle
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243339
Device Name
WiTOF
Indications for Use (Describe)
The WiTOF is a neuromuscular transmission station (nerve stimulator) that helps to pilot the neuromuscular blockade of adult patients in the operating room, recovery room or intensive care unit.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
idmed
an eye on your patient
# 5. 510 (k) Summary
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92.
## 1.1. Submitter
Submitter:
IDMED
Hotel Technoptic
2 rue Marc Donadille
13013 Marseille France
+33 (0)491 118 784
Establishment Registration Number: N° DUNS 260233256
Contact person:
Perrine MOLLE
Quality & Regulatory Affairs Director
E-mail : p.molle@idmed.fr
Date Summary Prepared: 2025/06/13
## 1.2. Device
Trade Name: WiTOF (reference WiTOF-MU)
Common/Usual Name: NeuroMuscular Transmission (NMT) station
Regulation Name: Electrical peripheral nerve stimulator
Classification Regulation: 21 CFR 868.2775
Product Code: KOI
Regulatory Class: Class II
## 1.3. Predicate device
Predicate Device: ToFscan, manufactured by Idmed, K172690
This predicate has not been subject to a design-related recall.
510(k) – WiTOF
{6}
idmed
an eye on your patient
# 1.4. Device description
The WiTOF is a neuromuscular transmission station (nerve stimulator) that helps to pilot the neuromuscular blockade of adult patients in the operating room, recovery room or intensive care unit. It is used by health professionals (anaesthesiologists, doctors, or fully qualified nurse anaesthetists) for:
1. Objective neuromuscular transmission monitoring
2. Subjective neuromuscular transmission monitoring
The continuous monitoring of NMT blocking involves electrical nerve stimulation and the measurement of acceleration in the corresponding muscle (with a three-dimensional acceleration sensor). Based on the force of contraction resulting from the stimulation, it is possible to draw conclusions about the effectiveness of an injected neuromuscular blocking agents or the patient's curare level.
The device and all of the settings associated with it are designed for use on adult patients in hospital or health institutions so that the patient's curare level can be monitored.
The WiTOF system is composed of the station, the wireless sensor (hand), and the US AC power supply unit. The wireless sensor itself consists of the 3D accelerometer in the splint.
The WiTOF can perform several different modes of electrical stimulation in accordance with usual clinical practice:
- TOF (Train Of Four)
- PTC (Post Tetanic Count)
- ATP (Automated TOF PTC)
- DBS (Double Burst) (3,3) (3,2)
- ST (Single Twitch) 0.1 Hz and 1 Hz
- TET (Tetanus) 50 Hz and 100 Hz
The station displays the various stimulation settings, the time elapsed between stimulations, and the results for the measurements made in response to TOF, PTC, and ATP stimulations. The responses to other types of stimulation are gaged visually by the health professional. The device and stimulation settings are chosen using a display screen.
The accessories that can be used with the device include:
- Optic-serial (RS232) cable to transfer data
- Fixation clamp – regular size (10–40 mm / 0.4–1.6 inches)
- Fixation clamp – large size (20–60 mm / 0.8–2.4 inches)
At a high level, the comparison of the subject and predicate devices is based on the following table:
| Features | WiTOF | ToFscan |
| --- | --- | --- |
| Product code | KOI | KOI |
| Regulation number | 868.2775 | 868.2775 |
| Class | II | II |
| Indication for use | | |
| Intended use | NMT monitor | NMT monitor |
510(k) - WiTOF
{7}
idmed
an eye on your patient
| Features | WiTOF | ToFscan |
| --- | --- | --- |
| Product code | KOI | KOI |
| Regulation number | 868.2775 | 868.2775 |
| Class | II | II |
| Environment of use | Hospital or health institutions: in the operating room, in recovery, or in the intensive care unit | Hospital or health institutions: in the operating room, in recovery, or in the intensive care unit |
| User | Health professionals | Health professionals |
| Patient population | Adult patients with neuromuscular blocking agent | Adult and pediatric patients with neuromuscular blocking agent |
| Contraindication | None known | None known |
| **Principle of operation and performances** | | |
| Stimulation Current Output | Constant current, 0-60 mA, monophasic 200μsec pulse width | Constant current, 0-60 mA, monophasic 200μsec pulse width |
| Maximum stimulation voltage | 300V (60 mA, 5kΩ) | 300V (60 mA, 5kΩ) |
| **Stimulation** | | |
| TOF | Yes | Yes |
| TOF auto | Yes | Yes |
| PTC | Yes | Yes |
| ST 0.1Hz and 1Hz | Yes | Yes |
| TET 50 Hz | Yes | Yes |
| TET 100 Hz | Yes | No |
| DBS 3.3 and 3.2 | Yes | Yes |
| ATP | Yes | Yes |
| **Calculations and measurements** | | |
| TOF calculation | Yes | Yes |
| PTC count | Yes | Yes |
| T4/T1 | Yes | Yes |
| T4/Tref | Yes | Yes |
| Displays amplitude of responses / Bars | Yes | Yes |
| **Technological characteristics** | | |
| Sensor position | Thumb – accelerometer | Thumb – accelerometer |
| Type of sensor | Accelerometer | Accelerometer |
| Type of electrodes | Surface | Surface |
| Portable device | Yes | Yes |
| General safety | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 |
| Connection Data | Wireless connection | Cable connection |
510(k) - WiTOF
5-3
{8}
idmed
an eye on your patient
## 1.5. Indications for use
The WiTOF is a neuromuscular transmission station (nerve stimulator) that helps to pilot the neuromuscular blockade of adult patients in the operating room, recovery room or intensive care unit.
The following difference exist between subject and predicate device:
- The subject device is not indicated to be used on pediatric population of patient.
However, this difference does not raise questions of safety and effectiveness, because the population not applicable for the subject device (pediatric) is not taken into account for the comparison. The subject device and its predicate device are both used in the same clinical conditions or purposes (adult patients under anesthesia for whom the effect of neuromuscular blocking agents is monitored) and are used by the same kind of users (Health professional).
## 1.6. Comparison of technological characteristics with the predicate device
The measuring the muscle movement in response to well-defined electrical nerve stimulation patterns in order to monitor the neuromuscular blocking of a patient is the technological principle for both the subject and predicate devices (acceleromyography).
The following technological differences exist between the subject and predicate device:
- The subject device does perform 100Hz TET stimulation
- The subject device is wireless connected to sensor
However, the differences presented above are supported by performance data (Software verification or safety test reports). Furthermore, these differences do not raise different questions of safety and effectiveness and do not negatively affect the clinical performances for the following reasons:
- Difference of connection to sensor (wireless for subject device): this difference is not clinically significant because the cable-free connection does not influence the results (performance evaluated on the SW verification and validation (see section 1.7 below). Furthermore, the safety of this technology has been tested according to the applicable standards (see section 1.7 below).
- Difference of TET stimulation 100Hz: this performance has been evaluated on the SW verification and validation (see section 1.7 below). Furthermore, this difference is not clinically significant because clinical literatures show that 100Hz stimulation is a common clinical practice used.
## 1.7. Performance data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biological evaluation for the WiTOF device was conducted in accordance with the FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
510(k) - WiTOF
{9}
idmed
an eye on your patient
The thumb splint is the only parts of the WiTOF in contact with the patient: intact skin for a duration of less than 24 hours.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the WiTOF device. The system complies with the IEC 60601-1, IEC 60601-1-6 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.” The software for this device was considered as a “moderate” level of concern.
## Animal and Clinical Studies
No animal or clinical testing was required to demonstrate the substantial equivalence of this device to its predicate, nor its safety and effectiveness.
## 1.8. Conclusions
Based on its intended use, design principles, and technological characteristics, the WiTOF device was found to be as safe, as effective, and performs comparably to the predicate device.
The technological differences identified do not raise new questions of safety and effectiveness as the non-clinical and clinical literature data support the safety of the device and the hardware and software verification and validation demonstrate that the WiTOF device should perform as intended in the specified use conditions.
510(k) – WiTOF
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.