Endoscopic Video Image Processor (RP-IPD-V1000F)

K243321 · Guangzhou Red Pine Medical Instrument Co., Ltd. · FET · Feb 7, 2025 · Gastroenterology, Urology

Device Facts

Record IDK243321
Device NameEndoscopic Video Image Processor (RP-IPD-V1000F)
ApplicantGuangzhou Red Pine Medical Instrument Co., Ltd.
Product CodeFET · Gastroenterology, Urology
Decision DateFeb 7, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Endoscopic Video Image Processor is used in conjunction with the Single-Use Video Flexible Cysto-Nephroscope (Models: RP-U-C01F, RP-U-C01FS) to process the images collected by the video endoscope and send them to the display, and provide power for the endoscope.

Device Story

Endoscopic Video Image Processor (RP-IPD-V1000F) functions as a reusable video processing unit for endoscopic procedures. It receives image signals from compatible Single-Use Video Flexible Cysto-Nephroscopes (RP-U-C01F, RP-U-C01FS), processes these signals, and outputs live video to a display. The device also serves as the power supply for the connected endoscope. Operated by clinicians in a clinical setting, the processor enables visualization of the internal anatomy during cystoscopy or nephroscopy. It features image adjustment capabilities including white balance, color tone, brightness, zoom, and image freezing. The device does not contact the patient directly but requires cleaning and disinfection between uses.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing of the Cysto-Nephroscope system, including direction of view, field of view, depth of field, resolution, signal-to-noise ratio, geometric distortion, image intensity uniformity, dynamic range, color performance, image frame frequency, and system delay. Testing complied with ISO 8600-1:2015 and ISO 8600-3:2019.

Technological Characteristics

Reusable video processing unit; provides power to endoscope. Features: white balance, color tone adjustment, image enhancement, zoom, brightness control, image freezing. Complies with ISO 8600-1:2015 and ISO 8600-3:2019. Requires cleaning and disinfection; no direct patient contact.

Indications for Use

Indicated for use in conjunction with Single-Use Video Flexible Cysto-Nephroscope (Models: RP-U-C01F, RP-U-C01FS) to process and display endoscopic images and provide power to the endoscope during endoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 7, 2025 Guangzhou Red Pine Medical Instrument Co., Ltd. Ping Yang Regulatory Affairs Manager 12 F, No.87 Luoxuan Avenue Guangzhou International Bioisland, Huangpu District Guangzhou, Guangdong 510000 CHINA Re: K243321 Trade/Device Name: Endoscopic Video Image Processor (RP-IPD-V1000F) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FET Received: January 7, 2025 Dear Ping Yang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {2}------------------------------------------------ # K243321 - Ping Yang (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {3}------------------------------------------------ the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Mark J. Antonino -S Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243321 Device Name Endoscopic Video Image Processor ( RP-IPD-V1 000F) Indications for Use (Describe) The Endoscopic Video Image Processor is used in conjunction with the Single-Use Video Flexible Cysto-Nephroscope (Models: RP-U-C01F, RP-U-C01FS) to process the images collected by the video endoscope and send them to the display, and provide power for the endoscope. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # K243321 Page 1 / 3 #### 510(k) Summary #### 510(k) Number: K243321 #### I. Contact Details Submitter Name: Guangzhou Red Pine Medical Instrument Co., Ltd. Submitter Address: 12 F, No.87 Luoxuan Avenue, Guangzhou International Bioisland, Huangpu District Guangzhou Guangdong 510000 China Submitter Contact Telephone: +86 13902971205 Submitter Contact: Mr. Weihua Yang Submitter Contact Email: regulation@gzredpine.com Correspondent Name: Guangzhou Red Pine Medical Instrument Co., Ltd. Correspondent Address: 12 F, No.87 Luoxuan Avenue, Guangzhou International Bioisland, Huangpu District Guangzhou Guangdong 510000 China Correspondent Contact Telephone: +86 18565651640 Correspondent Contact: Ms. Ping Yang Correspondent Contact Email: regulation(@gzredpine.com Date prepared: January 22, 2025 #### II. Subject Device Device Trade Name: Endoscopic Video Image Processor ( RP-IPD-V1000F) Common Name: Endoscope and accessories Classification Number: 21 CFR 876.1500 Classification Name: Endoscopic Video Imaging System/Component, Gastroenterology-Urology Product Code: FET #### III. Legally Marketed Predicate Device Predicate 510(k) Number: K241500 Device name: Endoscopic Video Image Processor (RP-IPD-V2000EF); Single-Use Video Flexible Cysto-Nephroscope (RP-U-C01F, RP-U-C01FS) Product Code: FAJ Note: This predicate device has not been subject to a design-related recall. {6}------------------------------------------------ K243321 Page 2 / 3 #### IV. Device Description Summary The Endoscopic Video Image Processor is a video processing system intended for use during endoscopic procedures. It receives and processes image signals from a compatible video endoscope and produces live video images during endoscopic procedures. Apart from the image processing functions, it also provides the power supply for the endoscope. The Endoscopic Video Image Processor is a reusable device. It does not require sterilization before use, as there is no direct/indirect patient contact. The device needs to be cleaned and disinfected before use. and the cleaning and disinfection method is outlined in the Instructions for Use. #### V. Indications for use The Endoscopic Video Image Processor is used in conjunction with the Single-Use Video Flexible Cysto-Nephroscope (Models: RP-U-C01F, RP-U-C01FS) to process the images collected by the video endoscope and send them to the display, and provide power for the endoscope. #### VI. Indications for Use Comparison The Indications for Use of the subject device is nearly identical to the Indications for Use of the predicate device. The only change is the addition of the compatible endoscope to the Indications for Use of the subject device. #### VII. Technological Comparison The principles of operation, white balance, color tone adjustment, image enhancement, zoom, adjusting brightness, image freezing and compatible endoscope of the subject device are identical to the predicate device. The product code, main configuration, dimension and weight of the subject device are different from or similar to the predicate device. The differences in product code, main configuration, video signal output, dimension and weight do not raise new question of safety and effectiveness. #### VIII. Non-Clinical Tests Summary #### 1. Performance Testing The Endoscopic Video Image Processor is compatible with the Single-Use Video Flexible Cysto-Nephroscope which was cleared under K241500 and is manufactured by Red Pine. The Cysto-Nephroscope system includes the Endoscopic Video Image Processor and the Single-Use Video Flexible Cysto-Nephroscope. The following performance testing was conducted on the Cysto-Nephroscope system: - 1. Direction of view - 2. Field of view - 3. Depth of field - 4. Resolution - 5. Signal-to-noise ratio - 6. Geometric distortion - 7. Image intensity uniformity {7}------------------------------------------------ 8. Dynamic range - 9. Color performance. The performance testing complied with the following standards: ISO 8600-1:2015 Endoscopes - Medical endoscopes and endotherapy devices. General requirements ISO 8600-3:2019 Endoscopes. Medical endoscopes and endotherapy devices. Part 3: Determination of field of view and direction of view of endoscopes with optics 2. Performance Testing-Image Frame Frequency and System Delay Image Frame Frequency and System Delay testing was also conducted on the Cysto-Nephroscope System. The Performance Testing demonstrated that the subject device and the predicate device have similar performance, and the subject device is as safe and effective as the predicate device. ### IX. Clinical Evidence N/A. ## X. Conclusion The results of nonclinical performance testing demonstrate that the subject device is as safe and effective as the predicate device to support a substantial equivalence determination.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...