K243290 · BTL Industries, Inc. · NUW · May 9, 2025 · Physical Medicine
Device Facts
Record ID
K243290
Device Name
BTL-785MJ
Applicant
BTL Industries, Inc.
Product Code
NUW · Physical Medicine
Decision Date
May 9, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BTL-785MJ device with the controller and single-use applicators has the following indications for use: 1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain 2. Muscle re-education 3. Increasing blood flow 4. Maintain or increase mandibular range of motion
Device Story
BTL-785MJ is an electrostimulation platform for dental applications. System comprises main unit, handpiece, holding arm, connection cables, single-use applicators, and therapy discomfort button. User operates via large color touch screen; on-screen interface provides step-by-step guidance for therapy procedures and parameter setting. Device delivers electromagnetic field to human tissues to induce muscle stimulation. Used in clinical settings by healthcare professionals. Output includes therapy type, remaining time, and parameters. Clinical benefit includes symptom relief for muscle spasm and TMJ dysfunction, muscle re-education, improved blood flow, and enhanced mandibular range of motion.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1, IEC 60601-2-10), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), software life cycle (IEC 62304), risk management (ISO 14971), and biocompatibility (ISO 10993-1, -5, -10).
Technological Characteristics
Electrostimulation platform; Class II, BF electrical safety. Output: rectangular pulse shape, 2 channels, max 80 mA current. Electrodes: silver conductive ink with hydrogel, 10-5000 Ohm impedance. Dimensions: 53.94" x 26.38" x 26.38". Connectivity: touch screen interface. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 62304, ISO 14971, ISO 10993.
Indications for Use
Indicated for patients requiring relief of muscle spasm symptoms, treatment of TMJ dysfunction and associated pain, muscle re-education, increased blood flow, and maintenance or increase of mandibular range of motion.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 9, 2025
BTL Industries, Inc.
David Chmel
CEO North America
362 Elm Street
Marlborough, Massachusetts 01752
Re: K243290
Trade/Device Name: BTL-785MJ
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: NUW, NFO
Dated: October 18, 2024
Received: April 10, 2025
Dear David Chmel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243290 - David Chmel
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243290 - David Chmel
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243290
Device Name
BTL-785MJ
Indications for Use (Describe)
The BTL-785MJ device with the controller and single-use applicators has the following indications for use:
1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain
2. Muscle re-education
3. Increasing blood flow
4. Maintain or increase mandibular range of motion
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
BTL Industries
# 510(k) Summary – K243290
## General Information
Sponsor: BTL Industries, Inc.
362 Elm Street
Marlborough, MA 01752
Tel: +1-866-285-1656
Fax: +1-888-499-2502
Applicant: BTL Industries, Inc.
362 Elm Street
Marlborough, MA 01752
Tel: +1-866-285-1656
Fax: +1-888-499-2502
Contact Person: David Chmel
BTL Industries, Inc.
chmel@btlnet.com
Summary Preparation
Date: May 08, 2025
## Device
Trade/Proprietary Name: BTL-785MJ
Common Name: Stimulator, muscle, powered, dental
Primary Classification Name: Powered muscle stimulator
Classification Regulation: 21 CFR 890.5850, Class II
Primary Product Code: NUW
Secondary Product Code: NFO
510(k) Summary
Page 1 of 6
{5}
BTL Industries
# Legally Marketed Predicate Device
The BTL-785MJ is substantially equivalent to the following product(s) that are already cleared for distribution in the USA under the following 510(k) Premarket Notification numbers:
## Primary predicate:
Device name: D function
Original 510(k) Sponsor: ITO CO., LTD.
510(k) Number: K203525
## Reference device:
Device name: BTL-785BNF-E
Original 510(k) Sponsor: BTL Industries, Inc.
510(k) Number: K242532
# Product Description
The BTL-785MJ is a state-of-the-art electrostimulation and radiofrequency platform with accessories that enables the application of therapy by electrical stimulation.
The control unit of the system is equipped with a large color touch screen that facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.
The generated electromagnetic field is intended to interact with the tissues of the human body in order to relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain, muscle re-education, increase blood flow and maintain or increase mandibular range of motion.
The BTL-785MJ device consists of the following main components:
- Main unit
- Handpiece
- Holding arm
- Connection cables
- Single-use applicators
- Therapy discomfort button
510(k) Summary
{6}
BTL Industries
# Indications for Use
The BTL-785MJ device with the controller and single-use applicators has the following indications for use:
1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain
2. Muscle re-education
3. Increasing blood flow
4. Maintain or increase mandibular range of motion
# Non-clinical Testing (Performance Data)
The BTL-785MJ device has been thoroughly evaluated for electrical safety. The device has been found to comply with applicable medical device safety standards:
IEC 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances – Requirements and tests
IEC 60601-2-10 Medical electrical equipment – Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
IEC 60601-1-6 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
IEC 62304 Medical device software – Software life cycle processes
ISO 14971 Medical devices – Application of risk management to medical devices
ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
ISO 10993-5 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
ISO 10993-10 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
# Clinical performance data
A clinical study was not conducted.
510(k) Summary
{7}
BTL Industries
Comparison with the Predicate Device
| 510(k) number
Device name
Company name: | K243290
BTL-785MJ
BTL Industries, Inc. | K203525
D function
ITO CO., LTD. | K242532
BTL-785BNF-E
BTL Industries, Inc. |
| --- | --- | --- | --- |
| Type | Subject device | Predicate device | Reference device |
| Product Code and Regulation | Physical medicine
Dental
21 CFR 890.5850
NUW – Stimulator, Muscle, Powered, Dental | Physical medicine
Dental
21 CFR 890.5850
NUW – Stimulator, Muscle, Powered, Dental | Neurology
Neurology
21 CFR 882.5890
NFO – Stimulator, Transcutaneous Electrical, Aesthetic Purposes |
| Indications for Use | The BTL-785MJ device with the controller and single-use applicators has the following indications for use:
1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain
2. Muscle re-education
3. Increasing blood flow
4. Maintain or increase mandibular range of motion | 1) To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain
2) Muscle re-education
3) Increasing blood flow
4) Maintain or increase mandibular range of motion | The BTL-785BNF-E device has the following indications for use:
The device is indicated for aesthetic use including facial and neck or body skin stimulation. |
| Principle of Action | Electrostimulation | Electrostimulation | Electrostimulation |
| Clinical Use | Prescription use | Prescription use | Prescription use |
510(k) Summary
{8}
BTL Industries
| Energy Source | 100 – 120 V AC, 50/60 Hz
200 – 240 V AC, 50/60 Hz | AC 100-240V, 50/60Hz
DC 12 V (AC adaptor)
DC 7.4 (Lithium ion battery) | 100 – 120 V AC, 50/60 Hz
200 – 240 V AC, 50/60 Hz |
| --- | --- | --- | --- |
| Electrical Safety & EMC Testing | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-10 | ANSI AAMI 60601-1
IEC 60601-1-2
IEC 60601-2-10 | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-10 |
| Energy Type | Electric stimulation | Electric stimulation | Electric stimulation |
| Accessory Attachment Method | Pad – single-use applicator | Pad – single-use applicator | Pad – single-use applicator |
| Therapy time | 20-60 minutes | 60 minutes (Up to 480 min) | 20 minutes |
| Pulse Width | Preset 1: Up to 150 μs
Preset 2: 488 μs | PAIN: Up to 150μs
CARE: 488 μs
MCR: 150ms | 160 μs |
| Output Frequency (Hz) | Preset 1: Up to 200
Preset 2: 0.67 | PAIN: Up to 200
CARE: 0.67
MCR: 400 | 250 |
| Output Current | Preset 1: Up to 80 mA
Preset 2: Up to 80 mA | PAIN: Up to 80 mA
CARE: 24 mA
MCR: 750 μA | Up to 130 mA |
| Max. Output Voltage [V] | BTL-785-7-2, -8
48,5V @ 500 Ω
48,5V @ 2000 Ω
0 @ 10 000 Ω | Not publicly available | BTL-785-7-2, -8
48,5V @ 500 Ω
48,5V @ 2000 Ω
0 @ 10 000 Ω |
| Max. Current density at 500 Ω [μA/mm²] | BTL-785-7-2 – 200 μA/mm²
BTL-785-7-8 -276 μA/mm² | Not publicly available | BTL-785-7-2 – 243 μA/mm²
BTL-785-7-8 - 335 μA/mm² |
| Max average power density at 500 Ω [W/cm²] | BTL-785-7-2 – 0,058 W/cm²
BTL-785-7-8 -0.082 W/cm² | Not publicly available | BTL-785-7-2 – 0,047 W/cm²
BTL-785-7-8 - 0,065 W/cm² |
| Pulse Shape | Rectangular | Rectangular | Rectangular |
| Channel Numbers | 2 | 2 | Up to 3 |
510(k) Summary
Page 5 of 6
{9}
BTL Industries
| Electrical Safety Class | Class II, BF | Class II, BF | Class II, BF |
| --- | --- | --- | --- |
| Timer | Yes | Yes | Yes |
| Buzzer | Yes | Yes | Yes |
| Display | Yes | Yes | Yes |
| (a) Electrodes materials
(b) Electroconductive media
(c) electrode-to-skin impedance range
(d) conductive surface area | a) Conductive ink (Silver) | a) Not publicly available | a) Conductive ink (Silver) |
| | b) Conductive hydrogel | b) Not publicly available | b) Conductive hydrogel |
| | c) 10-5000 Ohm | c) Not publicly available | c) 10-5000 Ohm |
| | d) BTL-785-7-2 - 4 cm²
BTL-785-7-8 - 2.9 cm² | d) Not publicly available | d) BTL-785-7-2 - 4 cm²
BTL-785-7-8 - 2.9 cm² |
| System Weight | 65 kg
(143 lb) | 230 g
(0.5 lb) | 65 kg
(143 lb) |
| System Dimensions (W×H×D) | 53.94" × 26.38" × 26.38" | 3.3" × 0.93" × 5.9" | 53.94" × 26.38" × 26.38" |
## Substantial Equivalence
The BTL-785MJ device has similar technological characteristics and the same intended use compared to the primary predicate device and comparable mode of action and intended use compared to reference device. Any differences between the predicate device and BTL-785MJ device have no significant influence on safety or effectiveness of the BTL-785MJ device.
Therefore, the BTL-785MJ device is substantially equivalent to the predicate device.
## Conclusion
Based upon the intended use, comparison of technical characteristics and performance testing provided in this premarket notification, the BTL-785MJ device has been shown to be substantially equivalent to the currently cleared predicate device for requested intended use.
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.