← Product Code [OLO](/productcode/OLO) · K243285

# TMINI Miniature Robotic System (K243285)

_THINK Surgical, Inc. · OLO · Nov 15, 2024 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K243285

## Device Facts

- **Applicant:** THINK Surgical, Inc.
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Nov 15, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

TMINI® Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate of knee implant components. The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools. The targeted population has the same characteristics as the population that is suitable for the implant(s) compatible with the TMINI® Miniature Robotic System. The TMINI® Miniature Robotic System is to be used with the following knee replacement systems in accordance with the indications and contraindications: - Enovis™ EMPOWR Knee System® - Ortho Development® BKS® and BKS TriMax® Knee System - Total Joint Orthopedics Klassic® Knee System - United® U2™ Total Knee System - Medacta® GMK® Sphere / SpheriKA Knee Systems - Zimmer Biomet Anterior & Posterior Referencing Persona® Knee - b-ONE MOBIO® Total Knee System - Maxx Orthopedics Freedom® Total & Titan Knee - LINK® LinkSymphoKnee System

## Device Story

TMINI® Miniature Robotic System assists surgeons in TKA procedures. Inputs: preoperative CT scans for 3D bone modeling; intraoperative anatomical landmark registration via optical tracking. Operation: TPLAN workstation enables surgeon-approved surgical planning; TNav console tracks bone position; hand-held robotic tool articulates in two degrees-of-freedom to guide bone pin placement. Output: physical bone pins defining planar cuts for implant components. Used in OR by surgeons. Benefits: provides software-defined spatial boundaries for accurate implant placement relative to patient anatomy. Affects clinical decision-making by enabling precise, pre-planned bone preparation and intraoperative gap balancing.

## Clinical Evidence

Bench testing only. Verification and validation activities included full system run-throughs, cutting accuracy, pin/block placement accuracy, cadaver lab validation, and system gap balance accuracy. All testing passed and met pre-determined acceptance criteria. No new clinical data was required for this modification.

## Technological Characteristics

System comprises TPLAN planning workstation, TNav optical tracking console, and hand-held robotic tool. Uses CT-based 3D modeling, six-camera overhead optical tracking, and active bone marker arrays. Materials have established history in orthopedics or ISO 10993 biocompatibility data. Connectivity involves workstation-to-robot data transfer for surgical plan execution. Software-defined spatial boundaries guide mechanical cut guides.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- TMINI® Miniature Robotic System ([K242264](/device/K242264.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 15, 2024

Think Surgical, Inc. Anand Patel Manager, Regulatory Affairs 47201 Lakeview Blvd Fremont, California 94538

Re: K243285

Trade/Device Name: TMINI® Miniature Robotic System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: October 17, 2024 Received: October 18, 2024

Dear Anand Patel:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

{2}------------------------------------------------

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shumaya Ali -S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K243285

Device Name TMINI® Miniature Robotic System

#### Indications for Use (Describe)

TMIN® Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate of knee implant components.

The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools.

The targeted population has the same characteristics as the population that is suitable for the implant(s) compatible with the TMIN® Miniature Robotic System. The TMIN® Miniature Robotic System is to be used with the following knee replacement systems in accordance with the indications and contraindications:

- · Enovis™ EMPOWR Knee System®
- · Ortho Development® BKS® and BKS TriMax® Knee System
- · Total Joint Orthopedics Klassic® Knee System
- · United® U2™ Total Knee System
- · Medacta® GMK® Sphere / SpheriKA Knee Systems
- · Zimmer Biomet Anterior & Posterior Referencing Persona® Knee
- b-ONE MOBIO® Total Knee System
- · Maxx Orthopedics Freedom® Total & Titan Knee
- · LINK® LinkSymphoKnee System

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|-----------------------------------------------|
|-----------------------------------------------------------------------------------------|-----------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the alphanumeric string 'K243285' in a simple, sans-serif font. The characters are uniformly sized and spaced, creating a clear and legible sequence. The string appears to be a code or identifier, possibly a serial number or product key. The black text on a white background enhances readability.

Image /page/4/Picture/1 description: The image contains the logo for Think Surgical. The logo features a stylized letter "T" in blue, with a pointed base resembling a surgical instrument. Below the "T" is the word "THINK" in bold, black letters, and below that is the word "SURGICAL" in smaller, black letters.

# 510(k) SUMMARY

# Applicant Information:

| Owner Name:                        | THINK Surgical, Inc.                    |
|------------------------------------|-----------------------------------------|
| Address:                           | 47201 Lakeview Blvd., Fremont, CA 94538 |
| Phone number:                      | 443-756-9392                            |
| Fax number:                        | 510-249-2396                            |
| Establishment Registration Number: | 3000719653                              |
| Contact Person:                    | Anand Patel                             |
| Date Prepared:                     | 17 October 2024                         |
| Device Information:                |                                         |
| Device Classification:             | Class II                                |
| Trade Name:                        | TMINI® Miniature Robotic System         |
| Common name:                       | Orthopedic Stereotaxic Instrument       |
| Classification name:               | Stereotaxic Instrument                  |
| Regulation number:                 | 882.4560                                |
| Product Code:                      | OLO                                     |

# Predicate Device:

The TMINI® Miniature Robotic System (Additional Knee System) is substantially equivalent in intended use, Indications for Use, design, materials, technology, operational principles and performance to the predicate, TMINI® Miniature Robotic System, cleared via K242264.

# Device Modification:

The purpose of this submission is to add three additional FDA cleared knee implant systems that are compatible with the TMINI® Miniature Robotic System to the existing implant systems that are already cleared for use with the device. The three new implant systems are: the b-ONE MOBIO® Total Knee System, the Maxx Orthopedics Freedom® Total & Titan Knee, and the LinkSymphoKnee System. As part of this change the labeling has been modified to show that the Indications for Use of the device has been updated to include compatibility with these three additional knee implant systems.

#### Device Description:

Like its predicate, the TMINI® Miniature Robotic System (Additional Knee System) consists of three primary components: a three-dimensional, graphical, Preoperative Planning Workstation (TPLAN® Planning Station), an Optical Tracking Navigation Console (TNav) and a robotically controlled hand-held tool (TMINI Robot) that assists the surgeon in preparing the bone for implantation of TKA components.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Think Surgical. The logo features the word "THINK" in bold, black letters, with a blue, three-dimensional "T" at the top. Below the word "THINK" is the word "SURGICAL" in smaller, black letters. The logo is clean and modern, with a focus on the company name.

The TPLAN Planning Station uses preoperative CT scans of the operative leg to create 3D surface models for case templating and intraoperative registration purposes. The Planning Workstation contains a library of 510(k) cleared knee replacement implant(s) available for use with the system. The surgeon can select an implant model from this library. The planner/surqeon can manipulate the 3D representation of the implant in relation to the bone model to optimally place the implant. The surgeon reviews and approves the case plan once the surgeon is satisfied with the implant selection and orientation. The data from the approved plan is written to a file that is used to guide the robotically controlled hand-held tool.

The hand-held robotic tool is optically tracked relative to optical markers placed in both the femur and tibia and articulates in two degrees-of-freedom, allowing the user to place bone pins in a planar manner in both bones. Mechanical quides are clamped to the bone pins, resulting in subsequent placement of cut slots and drill guide holes such that the distal femoral and proximal tibial cuts can be made in the pre-planned positions and orientations, and such that the implant manufacturer's multi-planer cutting block can be placed relative to drilled distal femoral pilot holes. If the surgeon needs to change the plan during surgery, it can be changed intraoperatively using TNav.

#### Intended Use:

Like the predicate, the TMIN® Miniature Robotic System (Additional Knee System) is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

#### Indications for Use:

The differences between the indications for use of the current device and the predicate are the addition of the three additional knee systems. See Table-1 below.

The table below identifies the substantial equivalence of TMIN® Miniature Robotic System (Additional Knee System) to the predicate, TMIN® Miniature Robotic System cleared via K242264.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Think Surgical. The logo features the word "THINK" in a bold, sans-serif font, with the letter "I" replaced by a blue, three-dimensional sphere with a pointed tip extending downwards. Below the word "THINK" is the word "SURGICAL" in a smaller, sans-serif font. The trademark symbol is located to the right of the letter "K".

| Product             | TMINI® Miniature Robotic<br>System (Additional Knee<br>System)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | TMINI® Miniature Robotic<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Conclusion                                                                                                          |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| 510(k) number       | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K242264                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                     |
| Manufacturer        | THINK Surgical, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | THINK Surgical, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                     |
| Product Code        | OLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | OLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | SAME                                                                                                                |
| Regulation          | 21 CFR 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 21 CFR 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | SAME                                                                                                                |
| Intended Use        | Intended to assist the surgeon<br>in providing software defined<br>spatial boundaries for<br>orientation and reference<br>information to anatomical<br>structures during orthopedic<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                         | Intended to assist the surgeon<br>in providing software defined<br>spatial boundaries for<br>orientation and reference<br>information to anatomical<br>structures during orthopedic<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                         | SAME                                                                                                                |
| Indications for Use | The TMINI® Miniature Robotic<br>System is indicated as a<br>stereotaxic instrumentation<br>system for total knee<br>replacement (TKA) surgery. It<br>is to assist the surgeon by<br>providing software-defined<br>spatial boundaries for<br>orientation and reference<br>information to identifiable<br>anatomical structures for the<br>accurate placement of knee<br>implant components.<br>The robotic device placement<br>is performed relative to<br>anatomical landmarks as<br>recorded using the system<br>intraoperatively and based on<br>a surgical plan-determined<br>preoperatively using CT based<br>surgical planning tools. | The TMINI® Miniature Robotic<br>System is indicated as a<br>stereotaxic instrumentation<br>system for total knee<br>replacement (TKA) surgery. It<br>is to assist the surgeon by<br>providing software-defined<br>spatial boundaries for<br>orientation and reference<br>information to identifiable<br>anatomical structures for the<br>accurate placement of knee<br>implant components.<br>The robotic device placement<br>is performed relative to<br>anatomical landmarks as<br>recorded using the system<br>intraoperatively and based on<br>a surgical plan-determined<br>preoperatively using CT based<br>surgical planning tools. | Substantially<br>Equivalent<br>(Adds<br>Compatibility<br>with three<br>additional total<br>knee implant<br>systems) |
|                     | The targeted population has<br>the same characteristics as<br>the population that is suitable<br>for the implant(s) compatible<br>with the TMINI® Miniature<br>Robotic System.<br>The TMINI® Miniature Robotic<br>System is to be used with the<br>following knee replacement<br>system(s) in accordance with<br>the indications and<br>contraindications:<br>- Enovis™ EMPOWR Knee<br>System®<br>- Ortho Development BKS®<br>and BKS TriMax® Knee                                                                                                                                                                                         | The targeted population has<br>the same characteristics as<br>the population that is suitable<br>for the implant(s) compatible<br>with the TMINI® Miniature<br>Robotic System.<br>The TMINI® Miniature Robotic<br>System is to be used with the<br>following knee replacement<br>system(s) in accordance with<br>the indications and<br>contraindications:<br>- Enovis™ EMPOWR Knee<br>System®<br>- Ortho Development BKS®<br>and BKS TriMax® Knee                                                                                                                                                                                         |                                                                                                                     |
| Product             | TMINI® Miniature Robotic<br>System (Additional Knee<br>System)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | TMINI® Miniature Robotic<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Conclusion                                                                                                          |
|                     | - Total Joint Orthopedics<br>- Klassic® Knee System<br>- United U2™ Knee Total<br>Knee System<br>- Medacta® GMK® Sphere /<br>SpheriKA Knee Systems<br>- Zimmer Biomet Anterior &<br>Posterior Referencing<br>Persona® Knee<br>- b-ONE MOBIO® Total<br>Knee System<br>- Maxx Orthopedics<br>- Freedom® Total & Titan<br>Knee<br>- LINK® LinkSymphoKnee<br>System                                                                                                                                                                                                                                                                            | - Total Joint Orthopedics<br>- Klassic® Knee System<br>- United U2™ Knee Total<br>Knee System<br>- Medacta® GMK® Sphere /<br>SpheriKA Knee Systems<br>- Zimmer Biomet Anterior &<br>Posterior Referencing<br>Persona® Knee                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                     |

### Table 1: Comparison of Intended Use and Indications for Use

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Think Surgical. The logo features a stylized letter "T" in blue, with a needle-like shape extending downwards from the center of the "T". Below the "T", the word "THINK" is written in a bold, sans-serif font, with the word "SURGICAL" written in a smaller font size below it.

#### Substantial Equivalence:

Both the TMINI® Miniature Robotic System (Additional Knee System), the subject of this submission, and the predicate device have the same intended use. Both are indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components during orthopedic procedures.

The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools. The difference between the new device and the predicate is that the new device includes compatibility with three additional knee systems. The addition of the three knee systems do not alter the intended use, Indications for Use (other than the addition of the three new implant systems), design, materials, technology, or operational principles of the TMINI® Miniature Robotic System and no new questions of safety or effectiveness resulted from the changes.

The Indications for Use of the new device and its predicate are identical except for the addition of the two new implant systems to the list of compatible implant systems and the removal of the sentence listing the system components.

Verification and validation activities were performed to ensure that the TMINI® Miniature Robotic System (Additional Knee System) fulfilled system requirements, ensure that the product design conforms to the user needs and intended uses and performs substantially equivalent to the predicate, TMINI® Miniature Robotic System, cleared via K242264. Throughout all testing performed, test samples were representative of the production product. The verification and validation activities for the TMINI® Miniature Robotic System (Additional Knee System) followed test protocols that were previously established and reviewed in 510(k) submissions for the predicate device. The verification and validation activities were successfully completed, and all pre-determined acceptance criteria were met.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Think Surgical. The logo features a stylized letter 'T' with a blue sphere at the top and a sharp, elongated point at the bottom, resembling a surgical instrument. Below the 'T', the word 'THINK' is written in bold, sans-serif font, and below that, the word 'SURGICAL' is written in a smaller, sans-serif font.

Biocompatibility information for patient contacting materials and testing for the TMINI® Miniature Robotic System were presented in submission K232802. There are no material changes to any of the direct patient contact components of the TMINI® Miniature Robotic System (Additional Knee System) included in this submission; therefore, no additional biocompatibility testing was required.

Substantial equivalence in technological characteristic and performance of the TMINI® Miniature Robotic System to the predicate device is outlined in Table-2 below:

| Product                                                      | TMINI® Miniature Robotic<br>System (Additional Knee<br>System)                                                                                             | TMINI® Miniature Robotic<br>System                                                                                                                         | Substantial<br>Equivalence<br>Conclusion |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| 510(k) number                                                | Subject Device                                                                                                                                             | K242264                                                                                                                                                    |                                          |
| Manufacturer                                                 | THINK Surgical Inc.                                                                                                                                        | THINK Surgical Inc.                                                                                                                                        |                                          |
| Materials                                                    |                                                                                                                                                            |                                                                                                                                                            |                                          |
| -Materials Used                                              | Uses materials with a long<br>history of use in orthopedic<br>procedures or provided<br>biocompatibility data<br>consistent with ISO 10993<br>requirements | Uses materials with a long<br>history of use in orthopedic<br>procedures or provided<br>biocompatibility data<br>consistent with ISO 10993<br>requirements | SAME                                     |
| Technological<br>Characteristics                             |                                                                                                                                                            |                                                                                                                                                            |                                          |
| -Major System<br>Components                                  | Planning and robot control<br>software, robotic positioning<br>device, navigation system,<br>reusable and disposable<br>instrumentation                    | Planning and robot control<br>software, robotic positioning<br>device, navigation system,<br>reusable and disposable<br>instrumentation                    | SAME                                     |
| -Patient Imaging                                             | CT images used to create a<br>3D model of the bone for<br>surgical planning                                                                                | CT images used to create a<br>3D model of the bone for<br>surgical planning                                                                                | SAME                                     |
| -Preoperative planning<br>workstation                        | TPLAN three-dimensional<br>preoperative planning<br>workstation                                                                                            | TPLAN three-dimensional<br>preoperative planning<br>workstation                                                                                            | SAME                                     |
| -Surgical planning<br>system                                 | Technician guided surgical<br>planning with surgeon<br>review and approval on a<br>desktop planning station                                                | Technician guided surgical<br>planning with surgeon<br>review and approval on a<br>desktop planning station                                                | SAME                                     |
| -Bone Marker Arrays for<br>bone registration and<br>tracking | Active markers on femur<br>and tibia mounted onto the<br>bones via an attachment<br>assembly                                                               | Active markers on femur<br>and tibia mounted onto the<br>bones via an attachment<br>assembly                                                               | SAME                                     |
| -Surgical Exposure                                           | Similar to traditional<br>surgical exposure                                                                                                                | Similar to traditional<br>surgical exposure                                                                                                                | SAME                                     |
| -Patient/Robot<br>Registration                               | Preoperatively determined<br>landmarks are compared to<br>intraoperatively identified<br>landmarks to complete<br>patient bone registration                | Preoperatively determined<br>landmarks are compared to<br>intraoperatively identified<br>landmarks to complete<br>patient bone registration                | SAME                                     |
| -Camera Tracking<br>Technology                               | Six camera overhead<br>tracking with a wide-angle<br>field of view                                                                                         | Six camera overhead<br>tracking with a wide-angle<br>field of view                                                                                         | SAME                                     |

# Table-2: Substantial Equivalence

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Think Surgical. The logo features a stylized letter "T" with a blue sphere at the top, followed by the text "THINK" in a bold, sans-serif font. Below the word "THINK" is the word "SURGICAL" in a smaller, sans-serif font.

# TMINI® Miniature Robotic System

| Product                                                                                                                          | TMINI® Miniature Robotic<br>System (Additional Knee<br>System)                                                                                                                                                                                                                                                                                                                                                                                | TMINI® Miniature Robotic<br>System                                                                                                                                                                                                                                                                            | Substantial<br>Equivalence<br>Conclusion                                     |
|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| 510(k) number                                                                                                                    | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                | K242264                                                                                                                                                                                                                                                                                                       |                                                                              |
| Manufacturer                                                                                                                     | THINK Surgical Inc.                                                                                                                                                                                                                                                                                                                                                                                                                           | THINK Surgical Inc.                                                                                                                                                                                                                                                                                           |                                                                              |
| -Cut guide positioning                                                                                                           | Robotic device places bone<br>pins in the correct plane,<br>then cutguide or drill block<br>is attached to the pins and<br>bone                                                                                                                                                                                                                                                                                                               | Robotic device places bone<br>pins in the correct plane,<br>then cutguide or drill block<br>is attached to the pins and<br>bone                                                                                                                                                                               | SAME                                                                         |
| -Intraoperative planning<br>changes                                                                                              | Implant position can be fully<br>adjusted, allowing deviation<br>from the intended implant<br>positioning philosophy and<br>implant size                                                                                                                                                                                                                                                                                                      | Implant position can be fully<br>adjusted, allowing deviation<br>from the intended implant<br>positioning philosophy and<br>implant size                                                                                                                                                                      | SAME                                                                         |
| -Bone Preparation<br>Technique                                                                                                   | A surgical saw is used to<br>cut the bone through a cut<br>guide                                                                                                                                                                                                                                                                                                                                                                              | A surgical saw is used to<br>cut the bone through a cut<br>guide                                                                                                                                                                                                                                              | SAME                                                                         |
| -Intraoperative Anatomic<br>Measurements                                                                                         | The tracked bone arrays<br>and bone registration data<br>are used to determine the<br>knee flexion angle and<br>varus/valgus laxity                                                                                                                                                                                                                                                                                                           | The tracked bone arrays<br>and bone registration data<br>are used to determine the<br>knee flexion angle and<br>varus/valgus laxity                                                                                                                                                                           | SAME                                                                         |
| -Gap Balancing                                                                                                                   | Displays the maximum<br>space in the medial and<br>lateral compartments in<br>millimeters with the knee in<br>extension and in flexion<br>allowing the surgeon to<br>perform gap balancing, if<br>desired                                                                                                                                                                                                                                     | Displays the maximum<br>space in the medial and<br>lateral compartments in<br>millimeters with the knee in<br>extension and in flexion<br>allowing the surgeon to<br>perform gap balancing, if<br>desired                                                                                                     | SAME                                                                         |
| -TKA Component<br>Implantation Technique                                                                                         | Implants are secured to the<br>bone, either with or without<br>cement using standard<br>surgical technique provided<br>by the implant manufacturer                                                                                                                                                                                                                                                                                            | Implants are secured to the<br>bone, either with or without<br>cement using standard<br>surgical technique provided<br>by the implant manufacturer                                                                                                                                                            | SAME                                                                         |
| Product                                                                                                                          | TMINI® Miniature Robotic<br>System (Additional Knee<br>System)                                                                                                                                                                                                                                                                                                                                                                                | TMINI® Miniature Robotic<br>System                                                                                                                                                                                                                                                                            | Substantial<br>Equivalence<br>Conclusion                                     |
| 510(k) number                                                                                                                    | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                | K242264                                                                                                                                                                                                                                                                                                       |                                                                              |
| Manufacturer                                                                                                                     | THINK Surgical Inc.                                                                                                                                                                                                                                                                                                                                                                                                                           | THINK Surgical Inc.                                                                                                                                                                                                                                                                                           |                                                                              |
| -Compatible Knee<br>Implant Systems                                                                                              | - Enovis™ EMPOWR Knee<br>System®<br>-Ortho Development® BKS®<br>and BKS TriMax® Knee<br>System<br>-Total Joint Orthopedics<br>Klassic® Knee System<br>-United® U2TM Knee<br>System<br>-Medacta® GMK® Sphere /<br>SpheriKA Knee Systems<br>-Zimmer Biomet Anterior &<br>Posterior Referencing<br>Persona® Knee<br>-b-ONE MOBIO® Total<br>Knee System<br>-Maxx Orthopedics<br>Freedom® Total & Titan<br>Knee<br>-LINK® LinkSymphoKnee<br>System | - Enovis™ EMPOWR Knee<br>System®<br>-Ortho Development® BKS®<br>and BKS TriMax® Knee<br>System<br>-Total Joint Orthopedics<br>Klassic® Knee System<br>-United® U2TM Knee<br>System<br>-Medacta® GMK® Sphere /<br>SpheriKA Knee Systems<br>-Zimmer Biomet Anterior &<br>Posterior Referencing<br>Persona® Knee | Substantially<br>Equivalent<br>(addition of three<br>new implant<br>systems) |
| Performance Testing                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                               |                                                                              |
| -Full System Run<br>Through Testing                                                                                              | Passed                                                                                                                                                                                                                                                                                                                                                                                                                                        | Passed                                                                                                                                                                                                                                                                                                        | SAME                                                                         |
| -Cutting Accuracy<br>-Pin & Block Placement<br>Accuracy<br>-Cadaver Lab Validation<br>Testing<br>-System Gap Balance<br>Accuracy | Passed                                                                                                                                                                                                                                                                                                                                                                                                                                        | Passed                                                                                                                                                                                                                                                                                                        | SAME                                                                         |
| *Biocompatibility<br>Testing                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                               |                                                                              |
| -Cytotoxicity                                                                                                                    | *Passed                                                                                                                                                                                                                                                                                                                                                                                                                                       | Passed                                                                                                                                                                                                                                                                                                        | SAME                                                                         |
| -Sensitization                                                                                                                   | *Passed                                                                                                                                                                                                                                                                                                                                                                                                                                       | Passed                                                                                                                                                                                                                                                                                                        | SAME                                                                         |
| -Intracutaneous<br>Reactivity                                                                                                    | *Passed                                                                                                                                                                                                                                                                                                                                                                                                                                       | Passed                                                                                                                                                                                                                                                                                                        | SAME                                                                         |
| -Acute Systemic Toxicity                                                                                                         | *Passed                                                                                                                                                                                                                                                                                                                                                                                                                                       | Passed                                                                                                                                                                                                                                                                                                        | SAME                                                                         |

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Think Surgical. The logo features a large blue sphere with a white "T" inside it. Below the sphere, the word "THINK" is written in bold, black letters. Underneath "THINK", the word "SURGICAL" is written in smaller, lighter letters.

## TMINI® Miniature Robotic System Special 510(k) Submission

* There are no material changes to any of the direct patient contact components of the TMINI® Miniature Robotic System (Additional Knee System) as a result of Additional Knee System included in this submission; therefore, no additional biocompatibility testing was required.

Risk assessment was performed on the device in accordance with ISO 14971:2019 and THINK Surgical Risk Management procedures. The risk assessment was comprised of

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Think Surgical. The logo features a stylized letter 'T' with a blue sphere at the top and a sharp point extending downwards. Below the 'T', the word 'THINK' is written in bold, sans-serif font, and below that, the word 'SURGICAL' is written in a smaller font.

analysis and mitigation of the risks associated with the addition of three compatible knee implant systems. The risk assessment resulted in the identification of no new clinical hazards. Each of the clinical hazards identified through this risk assessment is a previously documented hazard associated with the use of the TMINI® Miniature Robotic System. The addition of three new implant systems does not increase the likelihood or severity of these hazards; therefore, the risks associated with the use of the device remain unchanged as compared to the predicate.

#### Conclusion

The TMIN® Miniature Robotic System (Additional Knee System) is substantially equivalent to the predicate, TMINI® Miniature Robotic System cleared in K242264, in the following ways: it has the same intended use, the same technological characteristics and operating principles, and incorporates the same design and materials.

Performance testing and risk analysis has demonstrated that the performance and risk profile of the TMINI® Miniature Robotic System (Additional Knee System) is substantially equivalent to that of the predicate device and does not raise any new question of safety and effectiveness.

THINK Surgical Inc. respectfully submits that this filing contains adequate information and data to demonstrate the substantial equivalence of the TMIN® Miniature Robotic System (Additional Knee System) to the legally marketed TMINI® Miniature Robotic System cleared via K242264.

---

**Source:** [https://fda.innolitics.com/device/K243285](https://fda.innolitics.com/device/K243285)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).