← Product Code [KNT](/productcode/KNT) · K243228

# Flexi-Seal AIR (with ENFit Connector) (K243228)

_Convatec · KNT · Mar 28, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K243228

## Device Facts

- **Applicant:** Convatec
- **Product Code:** [KNT](/productcode/KNT.md)
- **Decision Date:** Mar 28, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Flexi-Seal AIR™ (with ENFit™ Connector) is an indwelling fecal management catheter intended to manage fecal incontinence through the collection of liquid stool and to provide access to administer medications. The device is intended for use in adult patients.

## Device Story

Indwelling fecal management catheter for adult patients; manages fecal incontinence by collecting liquid/semi-liquid stool; provides access for medication administration. System includes silicone catheter, foam-filled rectal retention balloon, auto-inflate valve, collection bag, and gelling/odor control sachets. Device operates by inserting catheter into rectum; auto-inflate valve automatically inflates balloon with air and regulates pressure to <22 mmHg; prevents over-inflation during patient movement. Healthcare providers use device in clinical settings; output is stool collection and medication delivery. Benefits include effective fecal management and reduced risk of balloon over-inflation via automated pressure relief.

## Clinical Evidence

Bench testing only. Performance evaluated for functional and structural parameters (balloon, auto-inflate valve, joint strength, mechanical properties, odor barrier). Biocompatibility testing performed per ISO 10993-1:2018.

## Technological Characteristics

Materials: 100% silicone catheter, foam-filled silicone retention balloon, polyurethane outer tube (tube-in-tube design). Energy: None (mechanical auto-inflate valve). Connectivity: None. Sterilization: Non-sterile. Form factor: Indwelling catheter with ENFit connector (ISO 80369-3).

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Flexi-Seal PROTECT Plus Fecal Management System ([K190114](/device/K190114.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 28, 2025

Convatec Michael Chitiyo Regulatory Affairs Specalist First Avenue, Deeside Industrial Estate Deeside Flintshire, CH5 2NU United Kingdom

Re: K243228

Trade/Device Name: Flexi-Seal AIR (with ENFit Connector) Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube And Accessories Regulatory Class: Class II Product Code: KNT, PIF Dated: October 24, 2024 Received: February 28, 2025

Dear Michael Chitiyo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Anthony Lee -S

Anthony C. Lee, Ph.D., M.B.A. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

Submission Number (if known)

K243228

Device Name

Flexi-Seal AIR (with ENFit ™ Connector)

Indications for Use (Describe)

Flexi-Seal AIR™ (with ENFit™ Connector) is an indwelling fecal management catheter intended to manage fecal incontinence through the collection of liquid stool and to provide access to administer medications. The device is intended for use in adult patients.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### 510(k) Summary

This summary of 510(k) safety and effectiveness information us submitted in accordance with the requirements of 21 CFR §807.92:

#### 1. Applicant:

ConvaTec Limited, GDC Building First Avenue Deeside Flintshire, GB CH5 2NU Establishment Registration number 3008806809

- Contact: Courtney Smith ConvaTec, Inc. Senior Director Regulatory Affairs Ostomy Care & Continence Care Tel: 239-7762947 Email: courtney.smith@convatec.com
#### 2. Device Identification:

| Trade Name                  | Flexi-Seal AIR (with ENFit Connector)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |  |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Common Name                 | Fecal Management System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |  |
| Regulation Number           | 21 CFR §876.5980                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |  |
| Classification Name         | Gastrointestinal tube and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |  |
| Regulatory Class            | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |  |
| Product Classification Code | KNT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |  |
| Device Description          | The Flexi-Seal Air™ (supplied non-sterile) is comprised of a low<br>pressure foam-filled 100% silicone rectal retention balloon fixed to<br>the distal end of a soft collapsible silicone catheter tube with a<br>labelled white inflation port and an auto Inflate valve, Luer lock<br>inflation syringe, an ENFit connector for irrigation/medication port<br>with an ENFit compatible syringe, cinch clamp, date formatted labels,<br>a sub-assembled self-closing quick connector to the catheter end, a<br>matching collection bag, and "Diamonds" gelling and odor control<br>sachets (4 sachets per kit). The components are contained in a rigid<br>thermoformed plastic clamshell.<br>The following is a list of design changes in the subject device, Flexi-<br>Seal Air™ versus the primary predicate device, Flexi-Seal® PROTECT<br>Plus Fecal Management System ref K190114. |  |  |
|                             | (1)<br>A new foam filled retention balloon which is similar to the<br>existing balloon in function and use. A single pilot balloon<br>indicator is used to assess if the balloon is fully deflated or                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |  |

{5}------------------------------------------------

|                     | approximately filled. Air is used instead of water for balloon<br>inflation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (2)                 | Addition of an auto-valve to automatically inflate the balloon,<br>self-regulate the balloon pressure to 0 mm Hg, and release of<br>excessive balloon pressure to prevent balloon over-inflation<br>during bowl/patient movement.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| (3)                 | Addition of an odor barrier outer polyurethane tube to the<br>external of the current silicone tube using the tube-in-tube<br>design except the first 10cm of the balloon end which is still a<br>100% silicone tube without polyurethane.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| (4)                 | A revised blue color silicone cover with fingertip grips and a<br>taller key at the bag flange to mitigate use errors in coupling<br>the bag upside down and trapping the bag film and minor<br>print edits in the catheter connector. These print<br>modifications include a blue arrow (catheter side) that<br>corresponds with the taller blue key (bag side) and a red 'no-<br>entry' symbol to show when the connector is in the upside-<br>down orientation.                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                     | Image: Product diagram with labels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                     | 1 Catheter Irrigation and Medication Port – ENFit™<br>2 Balloon Inflation/Deflation Port with Pilot Balloon<br>Indicator<br>3 Auto Inflate Valve<br>4 Sampling Port<br>5 Catheter (including Tube-in-Tube after Black Mark)<br>6 Position Indicator Line<br>7 Low-Pressure Retention Balloon with Finger Pocket<br>8 Self-Closing Connector<br>9 Hanging Strap<br>10 Luer-Lock Syringe<br>11 ENFit™ Syringe<br>12 Collection Bag<br>13 Blue Key<br>14 Fingertip Grips<br>15 Cinch Clamp<br>16 4 ConvaTec Diamonds™ gelling and odor control sachets                                                                                                                                                                                                                                                                                                                                                                           |
| Indications for Use | Flexi-Seal AIR™ (with ENFit™ Connector) is an indwelling fecal<br>management catheter intended to manage fecal incontinence<br>through the collection of liquid to semi-liquid stool and to provide<br>access to administer medications. The device is intended for use in<br>adult patients.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Performance Data    | Performance testing for Flexi-Seal Air™ was conducted for functional<br>and structural parameters. In this testing, the device's performance has<br>been found to be substantially equivalent to the predicate device both<br>functionally and structurally (balloon, auto inflate valve, joint strength,<br>mechanical properties, odor barrier, etc).<br>The device has also been evaluated for biocompatibility in accordance<br>with the US Food and Drug Administration's guidance entitled <i>Use of<br/>International Standard ISO 10993-1:2018, ‘Biological Evaluation of<br/>Medical Devices Part 1: Evaluation and Testing within a Risk<br/>Management Process’</i> , issued September 8, 2023, and has been found<br>safe in such respect for its intended use.<br>In conclusion, the subject device has been demonstrated as safe and<br>effective and substantially equivalent to the predicate device K190114. |

{6}------------------------------------------------

| Technological Comparison | Fecal management system is the intended technological principle for both the subject and predicate devices.<br>At a high level Flexi-Seal Air™ and the primary predicate Flexi-Seal Protect Plus are based on the following same technological elements:  Indications for use Low profile retention balloon A finger pocket to allow for easy balloon insertion. A silicone catheter A pilot balloon at the inflation/deflation port to indicate the fill status of the retention balloon. A quick connector that self-closes when bag is connected Ability to administer the medication through the use of a cinch clamp.              The following is a list of design changes in the subject device, Flexi-Seal Air™ versus the predicate device, Flexi-Seal® PROTECT Plus Fecal Management System ref K190114:  A new foam filled retention balloon, which is inflated with air. Addition of an auto-valve to automatically inflate the balloon. Addition of an odor barrier outer polyurethane tube A revised blue color silicone cover with fingertip grips and a taller key at the bag flange and minor print edits in the catheter connector, a matching blue arrow that corresponds with the taller blue key and a red 'no-entry' symbol to show the connector is upside down. |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

{7}------------------------------------------------

## Device Comparison Table

| Characteristic                                      | Subject Device                                                                                                                                                                                                                                                            | Primary Predicate Device (K190114)                                                                                                                                                                                                                                                                                               |                                                                   | regulate balloon pressure to less<br>than 22 mm Hg                   | corresponding pressure above 50 -<br>70 mm Hg)                    |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|----------------------------------------------------------------------|-------------------------------------------------------------------|
| Name                                                | Flexi-Seal Air™ ENFit                                                                                                                                                                                                                                                     | Flexi-Seal™ PROTECT Plus FMS                                                                                                                                                                                                                                                                                                     | Balloon Insertion Depth<br>Marking                                | Yes                                                                  | Yes                                                               |
| Intended Use                                        | Fecal Management                                                                                                                                                                                                                                                          | Fecal Management                                                                                                                                                                                                                                                                                                                 | Odor Barrier Catheter                                             | Yes                                                                  | Yes                                                               |
| Indication for Use                                  | Flexi-Seal Air™ is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients. | The Flexi-Seal™ PROTECT Plus Fecal Management System is an indwelling fecal management catheter with odor barrier intended for use to manage fecal incontinence through the collection of liquid or semi-liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients. | Catheter Material                                                 | Tube-in-Tube (polyurethane<br>outer tube and silicone inner<br>tube) | Silicone with Zeolite                                             |
| Sterility                                           | Non-Sterile                                                                                                                                                                                                                                                               | Non-Sterile                                                                                                                                                                                                                                                                                                                      | Medication Delivery                                               | Yes                                                                  | Yes                                                               |
| Max Usage of Device                                 | 29 days                                                                                                                                                                                                                                                                   | 29 days                                                                                                                                                                                                                                                                                                                          | ENFitTM Port for<br>irrigation<br>and medication<br>(ISO 80369-3) | Yes                                                                  | Yes                                                               |
| Inflated Balloon Height for anchoring in rectum, mm | 36                                                                                                                                                                                                                                                                        | 26                                                                                                                                                                                                                                                                                                                               | Self-Closing ENFitTM Port                                         | Yes                                                                  | Yes                                                               |
| Retention Balloon Diameter for anchoring in rectum  | 60 - 61                                                                                                                                                                                                                                                                   | 57                                                                                                                                                                                                                                                                                                                               | ENFitTM Syringe in the kit                                        | Yes                                                                  | Yes                                                               |
| Substance to inflate balloon                        | Air                                                                                                                                                                                                                                                                       | Water/Saline                                                                                                                                                                                                                                                                                                                     | Self-Closing Catheter<br>Connector                                | Yes                                                                  | Yes                                                               |
| Balloon fill volume per IFU                         | NA – Auto inflate by connecting Auto inflate Valve                                                                                                                                                                                                                        | 35 - 45 ml based on fill indicator                                                                                                                                                                                                                                                                                               | Collection Bag                                                    | Close with a cap<br>(not self-closing),<br>1 Liter                   | Close with a cap<br>(not self-closing),<br>1 Liter                |
| Balloon fill volume customized for each patient     | Yes, automatically based on auto inflate valve                                                                                                                                                                                                                            | Yes, passively using fill indicator by HCP providers                                                                                                                                                                                                                                                                             | Odor Control Sachet or<br>Spray                                   | DiamondsTM Gelling and odor<br>control sachets for collection bag    | DiamondsTM Gelling and odor<br>control sachets for collection bag |
| Marked Balloon inflation visual indicator           | Yes White dome shows inflation or deflation                                                                                                                                                                                                                               | Yes Green dome shows the optimal fill volume, and the red dome indicates over inflation.                                                                                                                                                                                                                                         |                                                                   |                                                                      |                                                                   |
| Active Balloon Pressure Management                  | Yes - Balloon pressure less than 22 mm Hg                                                                                                                                                                                                                                 | No – Only passive balloon pressure management using indicator                                                                                                                                                                                                                                                                    |                                                                   |                                                                      |                                                                   |
| Automatic Excessive Pressure Relief                 | Yes                                                                                                                                                                                                                                                                       | No                                                                                                                                                                                                                                                                                                                               |                                                                   |                                                                      |                                                                   |
| Means to mitigate balloon over-inflation            | The device has an auto inflate valve with pressure relief to self-                                                                                                                                                                                                        | Red dome signals over-inflation beyond 45 ml and above (or                                                                                                                                                                                                                                                                       |                                                                   |                                                                      |                                                                   |

{8}------------------------------------------------

* ENFit™ is a trademark of Global Enteral Device Supplier Association (GEDSA). Flexi-Seal™, Diamonds™– Trademark of ConvaTec Inc., NJ

#### Overall Conclusions

There are no differences in the intended use or performance of the subject and the predicate device.

Based on non-clinical testing of the subject device for functional and structural parameters, the modifications in the subject device do not affect its safety and effectiveness. Flexi-Seal Air™ described in this submission is substantially equivalent to the predicate device of Flexi-Seal™ PROTECT Plus Fecal Management System K190114.

---

**Source:** [https://fda.innolitics.com/device/K243228](https://fda.innolitics.com/device/K243228)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
