Mammotome AutoCore™ Single Insertion Core Biopsy System
K243095 · Devicor Medical Products, Inc. · KNW · Oct 29, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K243095
Device Name
Mammotome AutoCore™ Single Insertion Core Biopsy System
Applicant
Devicor Medical Products, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Oct 29, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Mammotome AutoCore™ Single Insertion Core Biopsy System is intended to obtain tissue samples from the breast or lymph nodes for diagnostic analysis of breast abnormalities. This instrument is for diagnostic use only and is not indicated for therapeutic use. The extent of a histologic abnormality cannot always be reliably determined from the palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Device Story
Automated core needle biopsy system; single insertion, multiple sample capability. Device comprises holster, charging base, and disposable probes (12G or 14G). Operation involves spring-loaded inner piercing stylet and outer cutting cannula to acquire soft tissue samples. Used in clinical settings for diagnostic breast/lymph node biopsy. Physician-operated. Output is physical tissue sample for histological analysis; assists in diagnosis of breast abnormalities. Benefits patient by enabling minimally invasive tissue acquisition for diagnostic evaluation.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing (static penetration force, lesion displacement, bond strength, bend load) and animal testing (sample weight, acquisition, and quality).
Technological Characteristics
Automated spring-loaded core biopsy system. Available in 12G and 14G sizes. Materials evaluated per ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-11:2017, and ISO 10993-23:2021. Electrical safety per IEC 60601-1-6. Software development per IEC 62304.
Indications for Use
Indicated for obtaining tissue samples from breast or lymph nodes for diagnostic analysis of breast abnormalities.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Achieve® Programmable Automatic Biopsy System (K141552)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
October 29, 2024
Devicor Medical Products, Inc. % Prithul Bom Most Responsible person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K243095
Trade/Device Name: Mammotome AutoCore™ Single Insertion Core Biopsy System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: September 27, 2024 Received: September 30, 2024
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Carr -S
Jessica Carr, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
Device Name
Mammotome AutoCore™ Single Insertion Core Biopsy System
Indications for Use (Describe)
The Mammotome AutoCore™ Single Insertion Core Biopsy System is indicated to obtain tissue samples from the breast or lymph nodes for diagnostic analysis of breast abnormalities. This instrument is for diagnostic use only and is not indicated for therapeutic use.
The extent of a histologic abnormality cannot always be reliably determined from the palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benian, it is essential that the tissue marqins be examined for completeness of removal using standard surgical procedures.
Type of Use (Select one or both, as applicable)
scription Use (Part 21 CFR 801 Subpart D)
he-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) SUMMARY | | |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92. | | |
| SUBMITTER INFORMATION | | |
| Name | Devicor Medical Products, Inc. | |
| Address | 300 E. Business Way, Fifth Floor, Cincinnati, OH 45241<br>U.S.A | |
| Establishment Registration<br>Number | 3008492462 | |
| Name of contact person | Jamie Edenborg | |
| Date prepared | 23-JUL-2024 | |
| PRODUCT | | |
| Mammotome AutoCore™ Single Insertion Core Biopsy System | | |
| PRODUCT COMPONENTS | | |
| PRODUCT | DESCRIPTION | |
| Mammotome AutoCore® Holster and Charging<br>Base | Holster and Charging Base | |
| Mammotome AutoCore® Probe 12 Gauge | Probe 12g x 10cm | |
| Mammotome AutoCore® Probe 14 Gauge | Probe 14g x 10cm | |
| Mammotome AutoCore® Introducer 12 Gauge | Introducer 12g x 10cm | |
| Mammotome AutoCore® Introducer 14 Gauge | Introducer 14g x 10cm | |
| DESCRIPTION OF DEVICE | | |
| Trade or proprietary name | Mammotome AutoCore™ Single Insertion Core<br>Biopsy System | |
| Common or usual name | Core Biopsy System, Single Insertion | |
| Classification name | Instrument, biopsy | |
| Classification panel | Gastroenterology/Urology | |
| Regulation | 21 CFR §876.1075 | |
| Product code(s) | KNW | |
| Legally marketed device(s) to which<br>equivalence is claimed | K141552 Achieve® Programmable Automatic<br>Biopsy System | |
| Submission Type | Traditional 510(k) Premarket Notification | |
| Device description | The Mammotome AutoCore™ Single Insertion<br>Core Biopsy System is an automated core<br>needle biopsy system. It is a single insertion,<br>multiple sample device that includes<br>automated arming and automated sample<br>collection. It is available in two gauge sizes<br>(12G and 14G) and acquires soft tissue<br>samples using a spring-loaded inner piercing<br>stylet and outer cutting cannula in the probe<br>needle. | |
| Intended use of the device | The Mammotome AutoCore™ Single Insertion<br>Core Biopsy System is intended to obtain<br>tissue samples from the breast or lymph nodes<br>for diagnostic analysis of breast abnormalities | |
{5}------------------------------------------------
This instrument is for diagnostic use only and is not indicated for therapeutic use. The extent of a histologic abnormality cannot always be reliably determined from the palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
#### SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
| Characteristic | New Device | Predicate |
|------------------|-----------------------------------------|-----------------------------------------------------------------------------------------------------|
| Mode of Action | Single Puncture and Multiple<br>Samples | Single Puncture and Multiple<br>Samples |
| Firing Modes | Automatic | Automatic and Delay |
| Anatomical Sites | Breast tissue and lymph<br>nodes | Breast tissue, lymph nodes,<br>kidney, liver, prostate, spleen<br>and various soft tissue<br>masses |
#### CONCLUSION OF DEVICE COMPARISON
A comparison of the technological characteristics of the proposed device and predicate device supports a determination of substantial equivalence.
#### PERFORMANCE DATA
# SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE Performance Test Summary
| Characteristic | Standard/Test/FDA Guidance |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance- Bench | Functional testing was performed for<br>comparison of static force to penetrate, lesion<br>displacement, overmold bond strength and<br>bend load to prove substantial equivalence. |
| Performance- Animal | Animal testing was performed to verify<br>appropriate sample weight, acquisition, and<br>quality. |
| <b>Software and Electrical Test Summaries</b> | |
| Characteristic | Standard/Test/FDA Guidance |
{6}------------------------------------------------
| Software | IEC 62304 Edition 1.1 2015-06 Medical Device<br>Software |
|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical | IEC 60601-1-6 Edition 3.1 2013-10 Medical<br>electrical equipment - Part 1-6: General<br>requirements for basic safety and essential<br>performance. |
| Biocompatibility Data | |
| Characteristic | Standard/Test/FDA Guidance |
| Biocompatibility | ISO 10993-1:2018-Biological Evaluation of<br>Medical Devices - Part 1: Evaluation and<br>Testing Within a Risk Management Process |
| Biocompatibility- Cytotoxicity | ISO 10993-5:2009-Biological Evaluation of<br>Medical Devices – Part 5: Tests for In Vitro<br>Cytotoxicity |
| Biocompatibility- Sensitization | ISO 10993-10:2010-Biological Evaluation of<br>Medical Devices – Part 10: Tests for Irritation<br>and Skin Sensitization |
| Biocompatibility- Pyrogenicity/Acute System<br>Toxicity | ISO 10993-11:2017-Biological Evaluation of<br>Medical Devices - Part 11: Tests for Systemic<br>Toxicity |
| Biocompatibility- Irritation/Intracutaneous<br>Reactivity | ISO 10993-23:2021 Biological Evaluation of<br>Medical Devices – Part 23: Tests for Irritation |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL<br>EQUIVALENCE AND/OR OF CLINICAL INFORMATION [Per 807.92(b)(1)(2)(3)] | |
| N/A - Clinical testing was not required for subject device | |
| CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA | |
The verification results and a side-by-side comparison of the technological characteristics of design, indication for use / intended use, components, and materials of construction, supports a determination that the Devicor Medical Products, Inc. Mammotome AutoCore™ Single Insertion Core Biopsy System [subject device] is substantially equivalent to the Achieve® Programmable Automatic Biopsy System (K141552) [predicate device].
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.