← Product Code [DZE](/productcode/DZE) · K243094

# Quickdent Dental Implant System (K243094)

_Quickdent Devices Private , Ltd. · DZE · Dec 19, 2024 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K243094

## Device Facts

- **Applicant:** Quickdent Devices Private , Ltd.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Dec 19, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Quickdent Dental Implant System are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient's esthetics and chewing function. Implants are intended for single or multiple unit restorations on splinted or non-splinted applications. Korticale TPR IH, Ossi Classic IH, Acti Fix CC and Kortifix Pro CC are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period. Trabecos 3.3 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.3 implants adjacent to one another. Korticale SP one piece implant is indicated for support and retention of fixed single tooth and fixed partial denture restorations in premolar, cuspid and incisor regions of partially edentulous jaws. Korticale SP one piece implants may be loaded immediately in the anterior mandibular arch if four implants are splinted together with a bar. Korticale SP one piece implants may be immediately restored with a temporary prosthesis that is not in functional occlusion.

## Device Story

Quickdent Dental Implant System consists of endosseous dental implants and associated abutments; used by dentists/oral surgeons in clinical settings to restore esthetics and chewing function. Implants are surgically placed in jaw arches; support prosthetic devices like artificial teeth. System includes internal hex (IH) and conical connection (CC) designs, plus one-piece implants. Implants feature grit-blasted and acid-etched surfaces. Abutments include healing caps, straight, angled, multi-unit, ball, and Quickloc/Quicksnap attachments. Clinicians select components based on patient anatomy and restoration needs. Output is a stable foundation for dental prosthetics. Benefits include restoration of oral function and esthetics. System supports immediate loading protocols when primary stability is achieved.

## Clinical Evidence

Bench testing only. Conducted dynamic fatigue testing per ISO 14801; biocompatibility testing (cytotoxicity, skin sensitization, irritation) per ISO 10993; sterilization validation (steam per ISO 17665-1, gamma per ISO 11137-2); bacterial endotoxin testing per ANSI/AAMI ST72:2019/USP <161>; packaging/shelf-life testing per ASTM standards. Surface characterization (SEM/EDX) confirmed cleanliness. Non-clinical MRI safety assessment performed using scientific rationale.

## Technological Characteristics

Materials: Ti-6AL-4V ELI. Surface: Grit-blasted and acid-etched (SLA). Connection types: Internal hex and conical. Components: Implants, abutments (straight, angled, multi-unit, ball, UCLA), healing caps, cover screws. Sterilization: Steam and Gamma irradiation. Connectivity: N/A (mechanical device).

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- TOV Dental Implant System ([K240837](/device/K240837.md))

## Reference Devices

- Zimmer One Piece Implant ([K052997](/device/K052997.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 19, 2024

Quickdent Devices Private Ltd. % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172

Re: K243094

Trade/Device Name: Quickdent Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 27, 2024 Received: September 30, 2024

Dear Angela Blackwell:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{3}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known) K243094

Device Name Quickdent Dental Implant System

#### Indications for Use (Describe)

The Quickdent Dental Implant System are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient's esthetics and chewing function. Implants are intended for single or multiple unit restorations on splinted applications. Korticale TPR IH, Ossi Classic IH, Acti Fix CC and Kortifix Pro CC are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period. Trabeos 3.3 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.3 implants adjacent to one another.

Korticale SP one piece implant is indicated for support and retention of fixed single tooth and fixed partial denture restorations in premolar, cuspid and incisor regions of partially edentulous jaws. Korticale SP one piece implants may be loaded immediately in the anterior mandibular arch if four implants are splinted together with a bar. Korticale SP one piece implants may be immediately restored with a temporary prosthesis that is not in functional occlusion.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size: 16px;">☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </div>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

# 510k Summary K243094 December 13, 2024 Quickdent Dental Implant System

Name and address: Quickdent Devices Private Ltd. Unit No. 128, Kuber Complex New Link Road S N of Village Oshiwar Link Road, Andheri West, Mumbai, Mumbai Suburban. Maharashtra 400053 India Contact Person: Dr. Mayur Khairnar Email: quickdentindia@gmail.com Name of device: Quickdent Dental Implant System Classification Name: Endosseous dental implants CFR: 21 CFR 872.3640 Primary Product Code: DZE Secondary Product Code: NHA

#### Submission Contact:

Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704)450-9934 angela@blackwelldevice.com

Device Description: The Quickdent Dental Implant System contains 3 designs of internal hex implants and various types of abutments as described below as well as 2 designs of conical implant with NP and RP platforms and abutments corresponding to both platforms. Additionally, a one piece implant is available. All devices are made from Ti-6AL-4V ELI unless otherwise noted. The implants have a grit blasted and acid etched surface. Conical abutments need to match the implant platform of narrow or regular. No abutments other than UCLA abutments are intended to be modified by the user.

Ossi Classic IH implants are tapered internal hex implants with more tightly spaced flat edge threads at the top and wider spaced flat edge threads in the lower tapered section. Ossi Classic IH comes in 3.5, 3.75, 4.2, 5.0 and 6.0 mm diameter with lengths of 8, 10, 11.5, 13, and 16mm (no 16mm in 6.0mm diameter).

Trabecos IH implants are cylindrical internal hex implants with evenly spaced threads which are flat edged at the top and sharp edged in the lower section. Trabecos comes in 3.3, 3.75, 4.2, 5.0, and 6.0 mm diameter with lengths of 8, 10, 11.5, 13, and 16mm in 5.0 or 6.0mm diameter). 3.3mm diameter implants are not for use with angled abutments in the posterior region of the mouth.

{6}------------------------------------------------

Korticale TPR IH are spiral internal hex implants with widely spaced threads. Korticale TPR IH comes in 3.5 and 4.2 mm diameter with lengths of 10, 12, 14 and 16mm.

Internal Hex Healing Abutments come in 4.5mm diameter with cuff heights of 2, 3, 4, 5, 6,7 mm. They come in narrow with a diameter 3.8mm with cuff heights of 3, 4, 5,6, 7 mm and wide with a diameter 5.5mm with cuff heights of 2,3,4,5,6,7 mm. Extra wide healing caps are 6.3mm diameter and come in cuff heights of 2,3,4,5 mm. An internal hex cover screw is also available. A healing cap for multi-units is available.

Internal Hex Straight Abutments have a 3.75mm interface diameter and come in 4.5mm diameter with total heights above the platform of 5,7,9,11,13,15 mm, and in 5.5mm diameter with heights above the platform of 9, 11, and 13 mm.

Internal Hex Straight Narrow Abutments have a 3.75mm interface area and come 3.75mm diameter with heights above the platform of 5,7,9, 11 mm.

Internal Hex Straight Shouldered Abutments come in 3.75mm interface diameter in 4.5 and 5.4 mm diameter with gingival heights of 1,2,3,4 mm.

Internal Hex Angled Abutments come in 15° 3.75mm diameter with height above platform of 9, and 11.4mm, and 25° 3.75mm diameter with height above platform of 8.5 mm.

Internal Hex Angled Anatomic Abutments come in 15° 3.75mm diameter with cuff heights of 1,2,3,4 mm and height above low side of shoulder of 8,9,10,11 mm and 25° 3.75mm diameter with cuff heights of 1,2,3,4 mm and height above platform of 8.3, 9.2, 10.3, 10.3 mm.

Internal Hex UCLA abutment bases come in 3.75mm diameter and use a Delrin plastic sleeve. The UCLA are for making straight restorations which are 4mm above the gingival collar and have a post height of no more than 12mm. The minimum wall thickness of the cast abutments is 0.3mm. The angulation, wall thickness and diameter of the UCLA base component are not intended to be modified.

Internal Hex Ball attachments come in 4.0mm diameter with cuff heights of 2, 3, 4, 5, 6, or 7mm.The ball attachments snap into a stainless steel housing which has a polyamide or polyether retention cap. The Retention Caps come in the colors yellow, pink and clear which represent 0.5, 0.9, and 1.3 kg retention levels. The retention caps allow implants to be within 14° of vertical and still snap into place.

Quickloc IH attachments come in 3.85mm diameter with cuff heights 1,2,3,4,5,6 mm. The Quickloc IH attachments snap into a Ti6AL4V ELI housing which has a polyamide, polyether or polyoxymethylene retention cap. The Retention Caps come in the colors yellow, pink, clear, purple which represent 0.6, 0.8, 1.0, 1.5 kg retention levels. The retention caps allow implants to be within 20° of vertical and still snap into place.

Internal Hex Straight Multi-Unit come in 4.8mm diameter with cuff heights of 1, 2, 3, 4, 5 mm.

{7}------------------------------------------------

Internal Hex Angled Multi-Units come in 17° and 30° with cuff heights of 1, or 2 mm.

Acti Fix CC implants are very slightly tapered conical implants with more tightly spaced flat edge threads at the top and wider spaced sharp edge threads in the lower tapered section. Acti Fix CC comes in 3.5 (NP), 4.3 (RP) and, 5.0 (RP) mm diameter with lengths of 8, 10, 11.5, 13, and 16mm.

Kortifix Pro CC are spiral conical implants with widely space threads. Kortifix Pro CC comes in 3.5 (NP) and 4.2 (RP) diameters in lengths of 10, 12, 14, and 16mm.

Conical healing caps come in NP and RP with cuff heights of 2,3,4,5 mm. NP and RP conical cover screws are also available. A healing cap for multi-units is available.

Conical straight abutments come in NP and RP with heights above platform of 9 or 13 mm.

Conical shoulder abutments come in NP and RP with gingival heights of 1,2,3 mm.

Conical angled shoulder abutments 15° come in NP and RP with cuff heights of 1,2,3 mm and total heights of 10.7, 12.2, 13.7 mm for NP and 11, 12, 13 mm for RP.

Conical angled shoulder abutments 25° come in NP and RP with cuff heights of 1,2,3 mm and total heights of 10.7, 12.2, 13.7 mm for NP and 11, 12, 13 mm for RP.

Conical ball attachments come in 3.5mm diameter for NP and 5.0mm diameter for RP with cuff heights of 1,2,3,4,5,6 mm. The ball attachments snap into a stainless steel housing which has a polyamide or polyether retention cap. The Retention Caps come in the colors yellow, pink and clear which represent 0.5, 0.9, and 1.3 kg retention levels. The retention caps allow implants to be within 14° of vertical and still snap into place.

Quicksnap CC attachments in 3.5mm diameter for NP and 5.0mm diameter for RP with cuff heights of 1,2,3,4 mm. The Quicksnap CC attachments snap into a Ti6AL4V ELI housing which has a polyamide, polyether or polyoxymethylene retention cap. The Retention Caps come in the colors yellow, pink, clear, purple which represent 0.6, 0.8, 1.0, 1.5 kg retention levels. The retention caps allow implants to be within 20° of vertical and still snap into place.

Conical straight multi-units in NP and RP are 4.8mm in diameter with cuff heights of 1,2,3,4,5 mm.

Conical NP and RP Angled Multi-Units come in 17° and 30° with cuff heights of 1, or 2 mm.

Korticale SP one piece implant comes in diameters of 3.5 and 4.2mm with lengths of 8, 10, 12, 14, and 16mm. The abutment portion has a 2mm gingival collar and a 7mm abutment post height above the gingival collar. Only straight models of Korticale SP are available angular correction.

{8}------------------------------------------------

#### Indications for Use:

The Quickdent Dental Implant System are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient's esthetics and chewing function. Implants are intended for single unit restorations on splinted or non-splinted applications. Korticale TPR IH, Ossi Classic IH, Acti Fix CC and Kortifix Pro CC are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period. Trabecos 3.3 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.3 implants adjacent to one another.

Korticale SP one piece implant is indicated for support and retention of fixed single tooth and fixed partial denture restorations in premolar, cuspid and incisor regions of partially edentulous jaws. Korticale SP one piece implants may be loaded immediately in the anterior mandibular arch if four implants are splinted together with a bar. Korticale SP one piece implants may be immediately restored with a temporary prosthesis that is not in functional occlusion.

Testing Summary: Dynamic fatigue testing according to ISO 14801 was conducted to determine the abutments and implants are strong enough for their intended use. Cytotoxicity testing according to ISO 10993 was done on both implants and abutments. Skin sensitization testing according to ISO 10993-10:2021 was conducted on implants. Irritation testing according to ISO 10993-23:2021 was conducted on implants. Steam sterilization was conducted according to ISO 17665-1. Bacterial endotoxin testing was conducted according to ANSI/AAMI ST72:2019 and USP <161>. Gamma irradiation was conducted according to ISO 11137-2. Package testing was conducted according to ASTM D999-08, ASTM F3039-13, and ASTM D5276-98(2009) and then shelf life testing was conducted according to ASTM F1929-12, and ASTM F1980-07. Testing of the modified surface included testing for organic carbon, hydrocarbons, and SEM/EDX. All were within the limits based on relevant standards set in the cleaning validation protocol.

#### MR Environment Condition

Non-clinical worst-case MRI review was performed to evaluate the metallic QUickdent Dental Implant System devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.

Primary Predicate Device: TOV Dental Implant System K240837 Reference Predicates: Zimmer One Piece Implant K052997

{9}------------------------------------------------

## Substantial Equivalence:

The Quickdent Dental Implant System is substantially equivalent to TOV Dental Implants in indications for use, materials, design, and fatigue performance. Slight differences in implant design between the subject devices and the predicate devices were addressed by the reference devices.

| Company &<br>Device Name | Quickdent Dental<br>Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | TOV Dental<br>Implant System<br>K240837<br>Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Zimmer One Piece<br>Implant K052997<br>Predicate for<br>Korticale SP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use   | The Quickdent<br>Dental Implant<br>System are<br>endosseous<br>implants intended<br>to be surgically<br>placed in the<br>upper or lower<br>jaw arches to<br>provide support<br>for prosthetic<br>devices, such as<br>an artificial tooth,<br>in order to<br>restore patient's<br>esthetics and<br>chewing function.<br>Implants are<br>intended for<br>single or multiple<br>unit restorations<br>on splinted or<br>non-splinted<br>applications.<br>Korticale TPR IH,<br>Ossi Classic IH,<br>Trabecos IH, Acti<br>Fix CC and Kortifix<br>Pro CC are<br>intended for<br>immediate<br>loading when<br>good primary<br>stability is<br>achieved, and<br>with appropriate<br>occlusive loading | The TOV Dental<br>Implant System<br>are endosseous<br>implants intended<br>to be surgically<br>placed in the<br>upper or lower<br>jaw arches to<br>provide support<br>for prosthetic<br>devices, such as<br>an artificial tooth,<br>in order to restore<br>patient's esthetics<br>and chewing<br>function. Implanet<br>implants are<br>intended for<br>single or multiple<br>unit restorations<br>on splinted or<br>non-splinted<br>applications.<br>Maer, Ragil and<br>TCX are intended<br>for immediate<br>loading when<br>good primary<br>stability is<br>achieved, and<br>with appropriate<br>occlusive loading.<br>These implants<br>can also be used<br>for loading after a<br>conventional<br>healing period | Zimmer® One-<br>Piece implants are<br>indicated for the<br>support and<br>retention of<br>fixed single tooth<br>and fixed partial<br>denture<br>restorations in the<br>premolar, cuspid,<br>and<br>incisor regions of<br>partially<br>edentulous jaws.<br>Zimmer® One-<br>Piece implants<br>may be loaded<br>immediately in<br>the anterior<br>mandibular arch if<br>four are splinted<br>together with a<br>bar. The Zimmer®<br>One-Piece implant<br>may be<br>immediately<br>restored with a<br>temporary<br>prosthesis that is<br>not in functional<br>occlusion. |

{10}------------------------------------------------

| These implants                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Ragil 3.3 implants                                                                                                                                                                                                                                                       |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| can also be used<br>for loading after a<br>conventional<br>healing period.<br>Trabecos 3.3<br>implants are<br>intended to<br>replace a lateral<br>incisor in the<br>maxilla and/or a<br>central or lateral<br>incisor in the<br>mandible.<br>Mandibular<br>central and lateral<br>incisors must be<br>splinted if using<br>two or more 3.3<br>implants adjacent<br>to one another.<br><br>Korticale SP one<br>piece implant is<br>indicated for<br>support and<br>retention of fixed<br>single tooth and<br>fixed partial<br>denture<br>restorations in<br>premolar, cuspid<br>and incisor<br>regions of<br>partially<br>edentulous jaws.<br>Korticale SP one<br>piece implants<br>may be loaded<br>immediately in<br>the anterior<br>mandibular arch if<br>four implants are<br>splinted together<br>with a bar.<br>Korticale SP one<br>piece implants | are intended to<br>replace a lateral<br>incisor in the<br>maxilla and/or a<br>central or lateral<br>incisor in the<br>mandible.<br>Mandibular<br>central and lateral<br>incisors must be<br>splinted if using<br>two or more 3.3<br>implants adjacent<br>to one another. |

{11}------------------------------------------------

|                                                          | immediately<br>restored with a<br>temporary<br>prosthesis that is<br>not in functional<br>occlusion.                                                                                                                                                                   |                                                                                                                                                                              |                                 |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Implant<br>Diameters                                     | Ossi Classic IH 3.5,<br>3.75, 4.2, 5.0, 6.0<br>mm<br>Trabecos IH 3.3,<br>3.75, 4.2, 5.0, 6.0<br>mm<br>Korticale TPR IH<br>3.5, 4.2<br>Acti Fix CC 3.5,<br>4.3, 5.0 mm NP =<br>3.5 RP = 4.3, 5.0<br>Kortifix Pro CC 3.5<br>(NP), 4.2 (RP)<br>Korticale SP 3.5,<br>4.2mm | Maer 3.5, 3.75,<br>4.2, 5.0, 6.0 mm<br>Ragil 3.3, 3.75,<br>4.2, 5.0, 6.0 mm<br>TCX 3.5, 4.3, 5.0<br>mm NP = 3.5 RP =<br>4.3, 5.0                                             | One Piece 3.7, 4.7<br>mm        |
| Implant Lengths                                          | Ossi Classic IH 8,<br>10, 11.5,<br>13,16mm (no<br>16mm in 6.0mm<br>diameter)<br>Trabecos IH 8, 10,<br>11.5, 13, and<br>16mm (no 16mm<br>in 5.0 or 6.0mm<br>diameter)<br>Korticale TPR IH<br>10, 12, 14, 16mm<br>Acti Fix CC 8, 10,<br>11.5, 13,16mm                    | Maer 8, 10, 11.5,<br>13,16mm (no<br>16mm in 6.0mm<br>diameter)<br>Ragil 8, 10, 11.5,<br>13, and 16mm (no<br>16mm in 5.0 or<br>6.0mm diameter)<br>TCX 8, 10, 11.5,<br>13,16mm | One Piece 10,<br>11.5, 13, 16mm |
|                                                          | Kortifix Pro CC 10,<br>12, 14, 16mm<br><br>Korticale SP 8, 10,<br>12, 14, 16mm                                                                                                                                                                                         |                                                                                                                                                                              |                                 |
|                                                          |                                                                                                                                                                                                                                                                        |                                                                                                                                                                              |                                 |
| Material of<br>devices included<br>in the submission     | Ti-6AL-4V ELI                                                                                                                                                                                                                                                          | Ti-6AL-4V ELI                                                                                                                                                                | Ti-6AL-4V ELI                   |
| Type of abutment<br>and maximum<br>angulation            | Pre-manufactured<br>of no more than<br>30°                                                                                                                                                                                                                             | Pre-manufactured<br>of no more than<br>30°                                                                                                                                   |                                 |
| Interface<br>type/shape                                  | Internal hex,<br>conical                                                                                                                                                                                                                                               | Internal hex,<br>conical                                                                                                                                                     | N/A                             |
| ISO 14801 Fatigue<br>Testing                             | Sufficient run out<br>load for their<br>intended use                                                                                                                                                                                                                   | Sufficient run out<br>load for their<br>intended use                                                                                                                         |                                 |
| Surface<br>Treatment                                     | SLA                                                                                                                                                                                                                                                                    | SLA                                                                                                                                                                          | RBM and acid<br>wash            |
| Post Surface<br>Treatment<br>Cleanliness<br>Demonstrated | Yes                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                          | Yes                             |

{12}------------------------------------------------

| Cover screw                                                                                                         | Cover screw for IH<br>NP and RP                                                                                                                                                                                                                                                                                         | Cover screw for IH<br>NP and RP                                                                                                                                                                                                                                                                                         |
|---------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Multi-Unit<br>Abutments in<br>IH, NP and RP                                                                         | 4.8 mm diameter<br>multi-units in IH,<br>NP and RP with<br>cuff heights of<br>1,2,3,4,5mm                                                                                                                                                                                                                               | 4.8 mm diameter<br>multi-units in IH,<br>NP and RP with<br>cuff heights of<br>1,2,3,4,5mm                                                                                                                                                                                                                               |
| 17° and 30°<br>Angled Multi-<br>Unit<br>Abutments<br>IH, NP and RP                                                  | 17° and 30°<br>Angled Multi-Unit<br>Abutments IH, NP<br>and RP with<br>platform heights<br>of 1,2 mm                                                                                                                                                                                                                    | 17° and 30°<br>Angled Multi-Unit<br>Abutments IH, NP<br>and RP with<br>platform heights<br>of 1,2 mm                                                                                                                                                                                                                    |
| Locator<br>Abutments<br>with metal<br>housing and<br>retention cap<br>allowing 20°<br>divergence of<br>the implants | Quickloc<br>IH/Quicksnap CC<br>attachments IH<br>(3.85mm), NP and<br>RP (3.9mm)<br>Cuff heights of 1,<br>2, 3, 4, 5, 6 mm (5<br>and 6 only in IH)                                                                                                                                                                       | Double Loc /<br>Conical Retentor<br>attachments IH<br>(3.85mm), NP and<br>RP (3.9mm)<br>Cuff heights of 1,<br>2, 3, 4, 5, 6 mm (5<br>and 6 only in IH)                                                                                                                                                                  |
|                                                                                                                     | Ti6Al4V ELI housing                                                                                                                                                                                                                                                                                                     | Ti6Al4V ELI housing                                                                                                                                                                                                                                                                                                     |
|                                                                                                                     | Polyamide, polyether or polyoxymethylene retention cap with retention levels from 0.6to 1.5 kg                                                                                                                                                                                                                          | Polyamide, polyether or polyoxymethylene retention cap with retention levels from 0.6to 1.5 kg                                                                                                                                                                                                                          |
| Ball<br>attachments<br>with metal<br>housing and<br>retention cap<br>allowing 14°<br>divergence of<br>the implants  | ball attachments<br>IH (4.00mm diameter), NP (3.5mm diameter) and RP (5.0mm diameter)<br>Cuff heights of 2, 3, 4, 5, 6, 7 mm for IH and 1.2.3.4.5.6 mm for NP and RP                                                                                                                                                    | ball attachments<br>IH (4.00mm diameter), NP (3.5mm diameter) and RP (5.0mm diameter)<br>Cuff heights of 2, 3, 4, 5, 6, 7 mm for IH and 1.2.3.4.5.6 mm for NP and RP                                                                                                                                                    |
|                                                                                                                     | Stainless steel 316 housing                                                                                                                                                                                                                                                                                             | Stainless steel 316 housing                                                                                                                                                                                                                                                                                             |
|                                                                                                                     | Polyamide or polyether retention cap with retention levels from 0.5 to 1.3 kg                                                                                                                                                                                                                                           | Polyamide or polyether retention cap with retention levels from 0.5 to 1.3 kg                                                                                                                                                                                                                                           |
| Healing Caps<br>3.8mm                                                                                               | IH 3.8mm diameter healing cap in 3,4,5,6,7 mm cuff height<br><br>3.75 diameter NP healing cap in 2,3,4,5 mm cuff heights                                                                                                                                                                                                | IH 3.8mm diameter healing cap in 3,4,5,6,7 mm cuff height<br><br>3.75 diameter NP healing cap in 2,3,4,5 mm cuff heights                                                                                                                                                                                                |
| Healing Caps<br>4.6 diameter<br>standard                                                                            | IH 4.5mm diameter healing cap in 2,3,4,5,6,7 mm cuff height<br><br>4.5 mm diameter RP healing cap with cuff heights 2,3,4,5 mm                                                                                                                                                                                          | IH 4.5mm diameter healing cap in 2,3,4,5,6,7 mm cuff height<br><br>4.5 mm diameter RP healing cap with cuff heights 2,3,4,5 mm                                                                                                                                                                                          |
| Healing Caps<br>5.5 diameter wide                                                                                   | IH 5.5mm diameter healing<br>abutment in<br>2,3,4,5,6,7 mm cuff height                                                                                                                                                                                                                                                  | IH 5.5mm diameter healing<br>abutment in<br>2,3,4,5,6,7 mm cuff height                                                                                                                                                                                                                                                  |
| Healing Cap<br>6.3mm diameter                                                                                       | IH 6.3mm diameter healing<br>abutment in<br>2,3,4,5 mm cuff height                                                                                                                                                                                                                                                      | IH 6.3mm diameter healing<br>abutment in<br>2,3,4,5 mm cuff height                                                                                                                                                                                                                                                      |
| Straight<br>Abutment                                                                                                | IH 4.5mm diameter<br>abutment with<br>heights above<br>platform of 5,<br>7,9,11,13,15 mm<br><br>NP and RP 4.5mm diameter with<br>heights above<br>platform of 9 and<br>13 mm                                                                                                                                            | IH 4.5mm diameter<br>abutment with<br>heights above<br>platform of 5,<br>7,9,11,13,15 mm<br><br>NP and RP 4.5mm diameter with<br>heights above<br>platform of 9 and<br>13 mm                                                                                                                                            |
| Straight Narrow<br>Abutment                                                                                         | IH 3.75mm diameter<br>abutment with<br>total heights of<br>5,7,9, 11 mm                                                                                                                                                                                                                                                 | IH 3.75mm diameter<br>abutment with<br>total heights of<br>5,7,9, 11 mm                                                                                                                                                                                                                                                 |
| Straight Wide<br>Abutment                                                                                           | IH 5.5mm diameter<br>abutment with<br>total heights of<br>9,11 or 13 mm                                                                                                                                                                                                                                                 | IH 5.5mm diameter<br>abutment with<br>total heights of<br>9,11 or 13 mm                                                                                                                                                                                                                                                 |
| Shoulder<br>Abutment                                                                                                | 4.5 diameter IH,<br>NP, RP shoulder<br>abutment with<br>shoulder heights<br>of 1,2,3,4 mm (4<br>only in IH)<br>IH total heights of<br>10.9, 11.9, 12.9,<br>13.9 mm<br>NP and RP total<br>heights of 11.8,<br>12.6, and 13.6mm                                                                                           | 4.5 diameter IH,<br>NP, RP shoulder<br>abutment with<br>shoulder heights<br>of 1,2,3,4 mm (4<br>only in IH)<br>IH total heights of<br>10.9, 11.9, 12.9,<br>13.9 mm<br>NP and RP total<br>heights of 11.8,<br>12.6, and 13.6mm                                                                                           |
|                                                                                                                     | 5.4mm diameter<br>IH shoulder<br>abutment with<br>shoulder heights<br>of 1.2.3.4 mm and<br>total heights of<br>10.7, 11.7, 12.17<br>and 13.7 mm                                                                                                                                                                         | 5.4mm diameter<br>IH shoulder<br>abutment with<br>shoulder heights<br>of 1.2.3.4 mm and<br>total heights of<br>10.7, 11.7, 12.17<br>and 13.7 mm                                                                                                                                                                         |
| 15° Angled<br>Abutment                                                                                              | 3.75mm diameter<br>IH 15° angled<br>abutment with<br>total heights of 9<br>or 11.4 mm                                                                                                                                                                                                                                   | 3.75mm diameter<br>IH 15° angled<br>abutment with<br>total heights of 9<br>or 11.4 mm                                                                                                                                                                                                                                   |
| 25° Angled<br>Abutment                                                                                              | 3.75mm diameter<br>IH 25° angled<br>abutment with<br>total height of 8.5 mm                                                                                                                                                                                                                                             | 3.75mm diameter<br>IH 25° angled<br>abutment with<br>total height of 8.5 mm                                                                                                                                                                                                                                             |
| Shouldered<br>15° Angled<br>Abutment<br>Hex, Conical<br>NP, and RP                                                  | IH 3.75mm diameter 15°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3,4 mm and<br>height above low<br>side of shoulder<br>8,9,10,11 mm<br><br>NP and RP 15°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3 mm and<br>total heights of<br>10.7, 12.2 and<br>13.7 mm & 11, 12,<br>13 mm | IH 3.75mm diameter 15°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3,4 mm and<br>height above low<br>side of shoulder<br>8,9,10,11 mm<br><br>NP and RP 15°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3 mm and<br>total heights of<br>10.7, 12.2 and<br>13.7 mm & 11, 12,<br>13 mm |
| Shouldered<br>25° Angled<br>Abutment<br>Hex, Conical<br>NP and RP                                                   | IH 3.75mm diameter 25°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3,4 mm with<br>heights above<br>platform of<br>8.3,9.2,10.3,10.3 mm                                                                                                                                                                | IH 3.75mm diameter 25°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3,4 mm with<br>heights above<br>platform of<br>8.3,9.2,10.3,10.3 mm                                                                                                                                                                |
|                                                                                                                     | NP and RP 25°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3 mm with<br>total heights of<br>10.7, 12.2, and<br>13.7 mm & 11, 12,<br>13mm                                                                                                                                                               | NP and RP 25°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3 mm with<br>total heights of<br>10.7, 12.2, and<br>13.7 mm & 11, 12,<br>13mm                                                                                                                                                               |
| UCLA Base Ti                                                                                                        | 4.5mm diameter                                                                                                                                                                                                                                                                                                          | 4.5mm diameter                                                                                                                                                                                                                                                                                                          |
|                                                                                                                     | UCLA base in                                                                                                                                                                                                                                                                                                            | UCLA base in                                                                                                                                                                                                                                                                                                            |
|                                                                                                                     | hexed                                                                                                                                                                                                                                                                                                                   | hexed                                                                                                                                                                                                                                                                                                                   |

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

## Conclusion:

Quickdent Dental Implant System is substantially equivalent to TOV Dental Implant System. They both have the same indications for use, are of the same material, and have internal hex and conical connections. The abutments, healing caps, and angled abutments are offered in similar designs and heights. Quickdent and TOV have the same types of performance testing with similar results. Performance testing demonstrates substantial equivalence to the identified predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K243094](https://fda.innolitics.com/device/K243094)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
