Cellbrick Knee Spacer

K243024 · United Orthopedic Corporation · JWH · Dec 20, 2024 · Orthopedic

Device Facts

Record IDK243024
Device NameCellbrick Knee Spacer
ApplicantUnited Orthopedic Corporation
Product CodeJWH · Orthopedic
Decision DateDec 20, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This product is indicated for temporary use (maximum of 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. This product is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). During the implantation period, patients have to use traditional mobility assist devices (e.g. crutches, walkers, canes) for daily activities. With pre-clinical validations of antibiotic elution, tibial spacer wear test, and knee spacer wear test, only Palacos MV+G bone cement may be used for preparing the spacers.

Device Story

Cellbrick Knee Spacer is a temporary, multi-component implant system for two-stage revision knee surgery in septic patients. System includes femoral spacer, tibial spacer, and optional canal rod. Femoral and tibial spacers feature multi-aperture, scaffold-like designs to act as carriers for Palacos MV+G antibiotic-loaded bone cement, facilitating local antibiotic release at the infection site. Canal rod (Ti-6Al-4V) provides additional surface area for cement attachment in deeper infection cases. Device is implanted by orthopedic surgeons in a clinical/hospital setting. Patients must use mobility aids (crutches, walkers) during the 180-day maximum implantation period. After 180 days, the spacer must be explanted and replaced with a permanent prosthesis or alternative treatment. The device provides temporary joint function and infection management, bridging the gap between initial debridement and definitive reconstruction.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included fatigue testing for tibial spacer, range of motion, constraint performance, wear testing, antibiotic release/in vitro antibiotic testing, usability evaluation, and endotoxin testing.

Technological Characteristics

Components: Femoral and Tibial spacers (UHMWPE, ASTM F648/ISO 5832-2), Canal Rod (Ti-6Al-4V, ASTM F136/ISO 5832-3). Design: Multi-aperture scaffold for antibiotic-loaded bone cement (Palacos MV+G). Fixation: Cemented. Sterilization: Not specified. Connectivity: None.

Indications for Use

Indicated for skeletally mature patients undergoing a two-stage total knee replacement procedure due to a septic process. Temporary use only (max 180 days). Requires use of mobility assist devices during implantation.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the date December 20, 2024. The text is written in a clear, sans-serif font. The date is written in a standard format, with the month, day, and year clearly indicated. The image is simple and straightforward, with the focus on the date itself. United Orthopedic Corporation Lois Ho Regulatory Affairs Manager No 57, Park Ave 2, Science Park Hsinchu, 30075 Taiwan Re: K243024 Trade/Device Name: Cellbrick Knee Spacer Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: September 13, 2024 Received: September 27, 2024 Dear Lois Ho: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {2}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {3}------------------------------------------------ Sincerely, JESSE MUIR - Digitally signed by JESSE MUIR - MUIR - MUIR -s mo Date: 2024.12.20 16:14:33 -05'00' S Jesse Muir, PhD Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K243024 Device Name Cellbrick Knee Spacer #### Indications for Use (Describe) This product is indicated for temporary use (maximum of 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. This product is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). During the implantation period, patients have to use traditional mobility assist devices (e.g. crutches, walkers, canes) for daily activities. With pre-clinical validations of antibiotic elution, tibial spacer wear test, and knee spacer wear test, only Palacos MV+G bone cement may be used for preparing the spacers. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for United Orthopedic. The logo consists of a green letter U with a white circle inside of it. To the right of the U is the text "UNITED ORTHOPEDIC" in black, with a registered trademark symbol after the word "ORTHOPEDIC". ## 510(K) SUMMARY # Traditional 510(k) [as required by 21 CFR 807.92(c)] ## Contact Details | Applicant Name | United Orthopedic Corporation | |--------------------------------|-------------------------------------------------------------| | Applicant Address | No 57, Park Ave 2, Science Park, Hsinchu City 30075, Taiwan | | Applicant Contact<br>Telephone | +88635773351 | | Applicant Contact | Mrs. Lois Ho | | Applicant Contact Email | lois.ho@unitedorthopedic.com | ### Device Name | Device Trade Name | Cellbrick Knee Spacer | |---------------------|-----------------------------------------------------------------------------------------------| | Common Name | Prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer | | Classification Name | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis | | Requlation Number | 888.3560 | | Product Codes | JWH | ### Predicate Device Information | 510(k) Number | Predicate Trade Name | Product Code | |--------------------------------|-----------------------------|--------------| | Primary Predicate | | | | K222570 | COPAL® knee moulds | JWH | | Predicate or Reference Devices | | | | K101356 | Spacer-K | JWH | | K181732 | Spacer-K | JWH | | K183017 | REMEDY® Stemmed Knee Spacer | JWH | | K140073 | U2 Total Knee System | JWH | | K132752 | U2 Total Knee System | JWH | | K150829 | U2 Total Knee System | JWH | | K222700 | U2 Total Knee System | JWH | | Device Description | Cellbrick Knee Spacer is a temporary knee spacer product, it's<br>suitable for the patients with mature skeleton and required to<br>perform two-stage knee joint prosthesis procedure for infection<br>control. It consists of three components, which are (1) Femoral<br>Spacer, (2) Tibial Spacer, and (3) Canal Rod. The Femoral<br>Spacers and the Tibial Spacers are designed with multiaperture<br>features, which can act as carrier of antibiotic-loaded bone | |----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Summary | | | | cement to release antibiotic at infection site. The Canal Rod is<br>optional.<br>- Femoral Spacer<br>The Femoral Spacer is made of UHMWPE (ASTM<br>F648/ISO5834-2). The geometrical appearance of Femoral<br>Spacer is designed and modified based on the 510(k) cleared<br>cruciate-retaining femoral component (K140073). Their<br>articular surface geometry and curvature is identical. In<br>addition, the Femoral Spacer is designed with multiaperture<br>feature. Such scaffold-like features can use as a carrier of<br>antibiotic-loaded cement to release antibiotic at the infection<br>site for infection control.<br>The Femoral Spacer can be used for both of left and right knee<br>because of the symmetric condyle design. Five femoral spacer<br>sizes have been prepared for different anatomical demands<br>with anteroposterior (AP) and mediolateral (ML) dimensions.<br>The Femoral Spacer is designed “same/ one up" sizing options<br>with the tibial spacers.<br>- Tibial Spacer<br>The Tibial Spacer is made of UHMWPE (ASTM F648/ISO5834-<br>2). The bearing surface's curvature and geometrical design of<br>the tibial spacer is the same as the 510(k) cleared ultra-<br>congruent tibial insert (K132752, K150829, and K222700). The<br>Tibial Spacer is designed with multiaperture feature as well.<br>This scaffold-like feature is used as a carrier of antibiotic-<br>loaded cement to release antibiotic at the infection site for<br>infection control.<br>Five Tibial Spacer sizes have been prepared for different<br>anatomical demands with anteroposterior (AP) and<br>mediolateral (ML) dimensions.<br>- Canal Rod<br>The canal rod is made of Ti-6Al-4V alloy (ASTM F136/ISO<br>5832-3). As a carrier, the rod provides the area for cement to<br>attach so that to increase the surface area of cement. The<br>canal rod is an optional device, it could be applied to patient<br>with deeper infection problem, applying Canal Rod can make<br>surgeon placing antibiotic-loaded cement into the deeper<br>infection site. The Canal Rod is provided with one dimension:<br>Ø 4(diameter) ×80mm (length). | | Intended Use/<br>Indications for Use | This product is indicated for temporary use (maximum of 180<br>days) as a total knee replacement (TKR) in skeletally mature<br>patients undergoing a two-stage procedure due to a septic<br>process. | | | This product is not intended for use for more than 180 days, at<br>which time it must be explanted and a permanent device<br>implanted or another appropriate treatment performed (e.g.,<br>resection arthroplasty, fusion, etc.). During the implantation<br>period, patients have to use traditional mobility assist devices<br>(e.g. crutches, walkers, canes) for daily activities.<br>With pre-clinical validations of antibiotic elution, tibial spacer<br>fatigue test, and knee spacer wear test, only Palacos MV+G<br>bone cement may be used for preparing the spacers. | | Indications for Use<br>Comparison | The indication for use of the subject device is substantially<br>equivalent to the primary predicate.<br>Although the content of indication for use is not completely<br>identical, however, both of the subject device and primary<br>predicate are temporary implants (maximum of 180 days),<br>which is intended to apply for the patient with mature skeleton<br>and required a two-stage procedure due to a septic process. In<br>addition, both should use mobility assist devices during the<br>period of implantation.<br>Base on above mentioned, the indication for use of subject<br>device is considered as substantially equivalent to the primary<br>predicate. | | Technological<br>Comparison | The technological characteristics of the subject device are<br>substantially equivalent to those of the primary predicate<br>(K222570), as the comparison given below shows.<br>- Their intended use is the same.<br>- The intended users, intended surgical procedure, body<br>contact and duration, and fixation method are identical to the<br>primary predicate.<br>- For the difference in the material, appearance design,<br>specification, device X-ray visibility, and sterilization method,<br>evaluation and verification activities were conducted, and the<br>equivalent safety and effectiveness compared to the predicate<br>devices were validated.<br>It established that no new risks arise compared to those of the<br>predicate devices. | | Non-Clinical and/or<br>Clinical Tests Summary<br>& Conclusions | Based on the technological characteristics of the subject device,<br>the following evaluation and tests were conducted to validate<br>the safety and effectiveness of the subject device.<br>- Fatigue test for Tibial spacer<br>- Range of Motion<br>- Constraint performance on the tibiofemoral interface<br>- Wear test<br>- Antibiotic release test & In vitro antibiotic test<br>- Usability evaluation<br>- Endotoxin testing | | | Not Applicable. | | | Based upon equivalences in intended use, patient population,<br>site of application, conditions of use, and non-clinical<br>performance data, the Subject device has been shown to be<br>safe and effective and to perform equivalently as compared to<br>the legally marketed predicate devices.<br>Therefore, the Subject device is substantially equivalent to the<br>legally marketed predicate devices. | {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for United Orthopedic. The logo consists of a green letter U with a white circle inside, next to the words "UNITED ORTHOPEDIC" in black. The word "ORTHOPEDIC" has a registered trademark symbol next to it. {7}------------------------------------------------ Image /page/6/Picture/1 description: The image contains the logo for United Orthopedic. The logo consists of a green letter U with a white circle inside, followed by the words "UNITED ORTHOPEDIC" in black. The word "ORTHOPEDIC" has a registered trademark symbol next to it. {8}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for United Orthopedic. The logo consists of a green circle with a white circle inside, resembling the letter 'U'. To the right of the symbol, the words 'UNITED' and 'ORTHOPEDIC' are stacked on top of each other in a simple, sans-serif font. A small registered trademark symbol appears next to the word 'ORTHOPEDIC'.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...