K243016 · Spectrawave, Inc. · NQQ · Mar 20, 2025 · Radiology
Device Facts
Record ID
K243016
Device Name
Starlight Imaging Catheter
Applicant
Spectrawave, Inc.
Product Code
NQQ · Radiology
Decision Date
Mar 20, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
Starlight Imaging Catheter with HyperVue Imaging System is intended for imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Starlight Imaging Catheter is intended for use in vessels 2.0 to 5.2 mm in diameter. The Starlight Imaging Catheter is not intended for use in a target vessel which has undergone a previous bypass procedure.
Device Story
Starlight Imaging Catheter is a sterile, single-use, rapid-exchange catheter used with the HyperVue Imaging System for coronary artery imaging. The device features an internal rotating fiber optic core that delivers near-infrared light to tissue and collects reflections. The HyperVue Controller drives the rotation and translational pullback of the imaging core within the catheter sheath to generate 360-degree cross-sectional images. Used under fluoroscopic guidance by clinicians during interventional procedures. The output allows visualization of coronary artery anatomy to assist in clinical decision-making during interventions. Benefits include high-resolution imaging of vessel structures.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing (optical performance, deliverability, pullback, trackability, kink resistance, tensile strength) and an animal study (3 porcine models, 18 imaging passes) evaluating vascular injury, thrombogenicity, and device safety/performance. Biocompatibility testing conducted per ISO 10993-1.
Technological Characteristics
Sterile, single-use, rapid-exchange catheter; 141 cm length, 2.5 Fr imaging window; compatible with 0.014" steerable guidewires. Operates via near-infrared light reflection, rotating/translational fiber optic core. Biocompatibility per ISO 10993-1. System connectivity via HyperVue Controller.
Indications for Use
Indicated for patients who are candidates for transluminal interventional procedures requiring imaging of coronary arteries (vessel diameter 2.0-5.2 mm). Contraindicated for use in target vessels with previous bypass procedures.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
March 20, 2025
SpectraWAVE, Inc. Ankit Shah Senior Manager Regulatory Affairs 12 Oak Park Dr. Bedford, Massachusetts 01730
Re: K243016
Trade/Device Name: Starlight Imaging Catheter Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: NQQ, ORD, OGZ, IYO Dated: February 18, 2025 Received: February 18, 2025
Dear Ankit Shah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243016
Device Name Starlight Imaging Catheter
Indications for Use (Describe)
Starlight Imaging Catheter with Hyper Vue Imaging System is intended for imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures.
The Starlight Imaging Catheter is intended for use in vessels 2.0 to 5.2 mm in diameter.
The Starlight Imaging Catheter is not intended for use in a target vessel which has undergone a previous bypass procedure.
| Type of Use (Select one or both, as applicable) | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <table border="0"><tr><td><span style="font-size: 2em;">☑</span> Prescription Use (Part 21 CFR 201 Subpart D)</td><td><span style="font-size: 2em;">☐</span> Over-The-Counter Use (21 CFR 201 Subpart C)</td></tr></table> | <span style="font-size: 2em;">☑</span> Prescription Use (Part 21 CFR 201 Subpart D) | <span style="font-size: 2em;">☐</span> Over-The-Counter Use (21 CFR 201 Subpart C) |
| <span style="font-size: 2em;">☑</span> Prescription Use (Part 21 CFR 201 Subpart D) | <span style="font-size: 2em;">☐</span> Over-The-Counter Use (21 CFR 201 Subpart C) | |
|× | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) Summary
- 1. Basic Information-Submitter:
| 510(k) Owner: | SpectraWAVE, Inc. |
|------------------------|--------------------------------------------------------------------------------------------|
| Address: | 12 Oak Park Drive<br>Bedford, MA 01730<br>(323) 401-6480 |
| Official Contact: | Ankit Shah<br>Senior Manager Regulatory Affairs<br>(323) 401-6480<br>ashah@spectrawave.com |
| Date Summary Prepared: | 26 September 2024 |
| Trade Name: | Starlight Imaging Catheter |
| Common Name: | Optical Coherence Tomography Imaging Catheter |
| Classification Name: | Ultrasonic pulsed echo imaging system |
| Regulation Numbers: | 21 CFR 892.1560 |
- 3. Predicate Devices:
Product Code:
Classification:
2.
- K221257 SpectraWAVE Imaging System .
- 4. Device Description:
The Starlight Imaging Catheter is a sterile, single-use, non-pyrogenic device and consists of two main assemblies: the catheter body and the internal rotating fiber optic imaging core. The catheter has an insertable length of 141 cm and a 2.5 Fr imaging window. It is a rapid exchange design with monorail tip, designed for compatibility with 0.014" (0.355mm) steerable guidewires used during coronary interventional procedures.
NQQ, ORD, OGZ, IYO
Class II
The Starlight Imaging Catheter connects to the HyperVue Imaging System through the HyperVue Controller (Controller), a reusable catheter connection allowing
{6}------------------------------------------------
direct control of basic data acquisition. All fiber optic rotation and translational pullback is driven by the Controller and occurs inside the catheter.
#### 5. Indications for Use Statement:
Starlight Imaging Catheter with HyperVue Imaging System is intended for imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures.
The Starlight Imaging Catheter is intended for use in vessels 2.0 to 5.2 mm in diameter.
The Starlight Imaging Catheter is not intended for use in a target vessel which has undergone a previous bypass procedure.
#### 6. Technological Characteristics:
The subject Starlight Imaging Catheter has same intended use, indications for use, functionality, and operating principle as the previously cleared predicate device. Both the subject and the predicate devices are based on the following same technological elements:
- Used with HyperVue Imaging System for viewing the coronary artery
- Inserted into an artery under fluoroscopic guidance. .
- Distal tip delivers near infrared light onto the tissue and collects reflections
- lmaging core moves longitudinally inside a sheath within the vessel to obtain ● images across the length of the target vessel (pullback).
- . lmaging core rotates inside a sheath within the vessel to obtain full 360 degrees to deliver light to create a cross-sectional image.
- . Patient population and user remain the same
The following technological differences exist between the subject and predicate devices:
- Different materials on the monorail and optical window
- Different outer and inner diameter on the shaft .
These differences do not introduce a new worst case and do not impact safety and effectiveness of the device.
{7}------------------------------------------------
#### 7. Performance Testing:
#### 7.1 Summary of Performance Testing:
Confirmatory testing was performed to verify and validate the design of the subject Starlight Imaging Catheter. This testing was in accordance with the standards and well-established methods used for the predicate devices.
#### 7.2 Sterilization and Shelf Life:
The modifications made to the subject Starlight Imaging Catheter do not impact the Sterilization and Shelf Life of the device.
#### 7.3 Biocompatibility:
The subject device underwent biocompatibility testing in accordance International Standard ISO 10993-1 "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". The studies performed on the catheter were determined to support the biocompatibility of the device for use in its intended application. The following tests were conducted:
- Cytotoxicity Sensitization Irritation Acute Systemic Toxicity Material Mediated Pyrogenicity Hemolysis Hemocompatibility Partial Thromboplastin Time Hemocompatibility Platelet & Leukocyte Hemocompatibility Complement Activation 40x Surface Morphology comparison
#### 7.6 Bench testing:
Confirmatory bench tests were performed to demonstrate the modifications do not introduce new concerns of safety and effectiveness. The following tests were done using well-established methods used for the predicate devices:
Optical performance Catheter deliverability Pullback performance Trackability Kink resistance Tensile strength
#### 7.7 Animal Testing:
Confirmatory animal study was completed using the well-established methods used for the predicate devices. The animal study demonstrated equivalent performance for the subject device in 3 porcine models that underwent 18 imaging
{8}------------------------------------------------
passes to evaluate for any vascular injury, thrombogenicity, device safety and device performance.
#### 7.8 Summative Usability Testing:
Usability evaluation testing is not required for the modifications being made within this special 510(k).
#### Clinical Testing: 7.9
No clinical testing is provided in this pre-market notification.
#### 8. Conclusion and Statement of Equivalence:
The information presented in this 510(k) submission demonstrates that the Starlight Imaging Catheter is as safe and effective as the predicate device and is therefore considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.