OncoStudio (OS-01)

K242994 · Oncosoft. Co., Ltd. · QKB · Feb 24, 2025 · Radiology

Device Facts

Record IDK242994
Device NameOncoStudio (OS-01)
ApplicantOncosoft. Co., Ltd.
Product CodeQKB · Radiology
Decision DateFeb 24, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K242994 · Feb 24, 2025OncoStudio (OS-01)Oncosoft. Co., Ltd.Yonsei Severance Hospital (South Korea) clinical CT images; OneMedNet (USA) clinical CT dataset; University Hospital Basel (Switzerland) clinical CT images (TotalSegmentator dataset)Retrospective clinical CT images were used to train the deep-learning model and to conduct a standalone performance evaluation of the OncoStudio device.Retrospective clinical data; Deep learning training; Standalone performance evaluation; CT images

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Standalone performance evaluation; Retrospective performance evaluationAdult patients; 58% male, 42% female; diverse ethnic backgrounds; Sample Size: 310 CT images; Number of Sites: 3Not applicable for this studyMean Dice coefficient (DSC) of structures for anatomical regions (Head & Neck, Thorax, Abdomen, and Pelvis)

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Automatic contouring of organs at riskDeep learning——Training: 2,128 images (731 from Yonsei Severance Hospital, 194 from OneMedNet, 1203 from University Hospital Basel).3 (radiation oncologists)Standalone performance test: 310 CT images (140 from Yonsei Severance Hospital, 121 from OneMedNet, 49 from University Hospital Basel).3 (radiation oncologists)

Indications for Use

OncoStudio provides deep-learning-based automatic contouring to organs at risk in DICOM-RT format from CT images. This software could be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed.

Device Story

OncoStudio is a standalone software application that processes CT images to perform automatic segmentation of organs at risk (OARs). Using deep-learning algorithms, the device generates DICOM-RT structure sets. It is intended for use by radiation oncology departments and other medical professionals to provide initial contours, which must be reviewed, edited, and confirmed by a clinician before use in treatment planning. The software supports manual contouring and standard DICOM image management functions (receive, transmit, store, retrieve, display). By automating the initial contouring process, the device aims to assist clinicians in radiation therapy treatment planning workflows, potentially improving efficiency in clinical practice.

Clinical Evidence

Bench testing only. Performance evaluated on 310 independent CT images (140 Korea, 121 USA, 49 Europe). Ground truth established by three radiation oncologists following RTOG and clinical guidelines. Performance metrics included mean Dice Similarity Coefficient (DSC) across anatomical regions (Head & Neck, Thorax, Abdomen, Pelvis). Training data (2,438 total) included diverse ethnicities and CT manufacturers (Siemens, Philips, Toshiba, GE).

Technological Characteristics

Standalone software; deep-learning-based segmentation; DICOM 3.0 compliant; input: CT images; output: DICOM-RT structure sets; platform: Windows-based; connectivity: DICOM network; software class: deep learning algorithm.

Indications for Use

Indicated for adult patients requiring automatic contouring of organs at risk from CT images for radiation oncology treatment planning or other clinical workflows requiring segmented organ masks. Covers Head & Neck, Thorax, Abdomen, and Pelvis regions.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure. The FDA acronym is in a blue square, and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue text to the right of the square. February 24, 2025 OncoSoft. Co., Ltd. Boram Kim RA/QA Manager 37, Myeongmul-gil, Seodaemun-gu SEOUL, 03776 KOREA, SOUTH Re: K242994 Trade/Device Name: OncoStudio (OS-01) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QKB Dated: January 23, 2025 Received: January 23, 2025 Dear Boram Kim: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {2}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {3}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Locon Weidner Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K242994 Device Name OncoStudio (OS-01) Indications for Use (Describe) OncoStudio provides deep-learning-based automatic contouring to organs at risk in DICOM-RT format from CT images. This software could be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed. - · Deep learning contouring from Head & Neck, Thorax, Abdomen, and Pelvis - · Generates DICOM-RT structure of contoured objects - · Manual Contouring - · Receive, transmit, store, retrieve, display, and process medical images and DICOM objects Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ ## 510(k) Summary [As Required by 21 CFR 807.92] #### 1. Date Prepared [21 CFR 807.92(a)(a)] January 23, 2025 #### 2. Submitter's Information [21 CFR 807.92(a)(1)] - Name of Manufacturer: OncoSoft Co., Ltd. 37, Myeongmul-gil, Seodaemun-gu, Seoul, Republic of Korea ● Address: (03776) ● Contact Name: Boram Kim ● Telephone No.: +82-10-6305-7428 ● Email Address: kbrstar@oncosoft.io - 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] | 510(k) Number | K242994 | |-----------------------------|------------------------------------------------| | Trade/Device/Model Name | OncoStudio / OS-01 | | Device Classification Name | Medical Image Management and Processing System | | Regulation Number | 21 CFR 892.2050 | | Classification Product Code | QKB | | Device Class | Class II | | 510(k) Review Panel | Radiology | {6}------------------------------------------------ ### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)] The identified predicate device within this submission is shown as follow; ### Predicate Device | 510(k) Number | K230685 | |-----------------------------|------------------------------------------------| | Trade/Device/Model Name | AutoContour Model RADAC V3 | | Device Classification Name | Medical Image Management And Processing System | | Regulation Number | 21 CFR 892.2050 | | Classification Product Code | OKB | | Device Class | Class II | | 510(k) Review Panel | Radiology | #### Reference Device | 510(k) Number | K232899 | |-----------------------------|------------------------------------------------| | Trade/Device/Model Name | AI-Rad Companion Organs RT | | Device Classification Name | Medical Image Management And Processing System | | Regulation Number | 21 CFR 892.2050 | | Classification Product Code | OKB | | Device Class | Class II | | 510(k) Review Panel | Radiology | These predicate devices have not been subject to a design-related recall {7}------------------------------------------------ #### 5. Description of the Device [21 CFR 807.92(a)(4)] OncoStudio is a standalone software that provides deep-learning-based automatic contouring to organs at risk in DICOM-RT format from CT images. This software could be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed. - Deep learning contouring from Head & Neck, Thorax, Abdomen, and Pelvis - · Generates DICOM-RT structure of contoured objects - Manual Contouring - · Receive, transmit, store, retrieve, display, and process medical images and DICOM objects It also has the following general functions: - Patient management; - · Review of processed images; - Open and Save of files. #### 6. Indications for use [21 CFR 807.92(a)(5)] OncoStudio provides deep-learning-based automatic contouring to organs at risk in DICOM-RT format from CT images. This software could be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed. - Deep learning contouring from Head & Neck, Thorax, Abdomen, and Pelvis - · Generates DICOM-RT structure of contoured objects - Manual Contouring - · Receive, transmit, store, retrieve, display, and process medical images and DICOM objects {8}------------------------------------------------ ### 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)] There are no significant differences in the technological characteristics of these devices compared to the predicate devices which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the OncoStudio and the predicate devices: #### [Table 1. Comparison of Proposed Device to Predicate Device and Reference Device] | | Subject Device | Predicate Device | Reference Device1 | | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Item | OncoStudio | AutoContour Model RADAC<br>V3 | AI-Rad Companion<br>Organs RT | SE<br>Note | | Regulation<br>Name | Medical Image Management<br>And Processing System | Medical Image<br>Management And<br>Processing System | Medical Image<br>Management And<br>Processing System | - | | Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 | - | | Product<br>Code | QKB | QKB | QKB | - | | Class | II | II | II | - | | 510k<br>Number | K242994 | K230685 | K232899 | - | | Indication<br>for Use | OncoStudio provides deep-<br>learning-based automatic<br>contouring to organs at risk<br>in DICOM-RT format from CT<br>images. This software could<br>be used as an initial<br>contouring for the clinicians<br>to be confirmed by the<br>radiation oncology<br>department for treatment<br>planning or other professions<br>where a segmented mask of<br>organs is needed.<br>• Deep learning<br>contouring from Head &<br>Neck, Thorax, Abdomen,<br>and Pelvis<br>• Generates DICOM-RT<br>structure of contoured<br>objects<br>• Manual Contouring<br>• Receive, transmit, store | AutoContour is intended to<br>assist radiation treatment<br>planners in contouring and<br>reviewing structures within<br>medical images in<br>preparation for radiation<br>therapy treatment<br>planning | AI-Rad Companion<br>Organs RT is a post-<br>processing software<br>intended to<br>automatically<br>contour DICOM CT<br>and MR predefined<br>structures using<br>deep-learning-based<br>algorithms. Contours<br>that are generated<br>by AI-Rad<br>Companion Organs<br>RT may be used as<br>input for clinical<br>workflows including<br>external beam<br>radiation therapy<br>treatment planning.<br>AI-Rad Companion<br>Organs RT must be<br>used in conjunction<br>with appropriate | Same | {9}------------------------------------------------ # K242994 | | retrieve, display, and process<br>medical images and DICOM<br>objects | | | | software such as<br>Treatment Planning<br>Systems and<br>Interactive<br>Contouring<br>applications, to<br>review, edit, and<br>accept contours<br>generated by AI-Rad<br>Companion Organs<br>RT. The output of AI-<br>Rad Companion<br>Organs RT are<br>intended to be used<br>by trained medical<br>professionals. The<br>software is not<br>intended to<br>automatically detect<br>or contour lesions. | | |----------------------------------|-----------------------------------------------------------------------|--------------------|------------------|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | | • A_Aorta | • LN_Sclav_L_RTOG | • A_Aorta | • LN_Ax_L | (166 OARs) | | | | • A_Carotid_L | • LN_Sclav_R_ESTRO | • A_Aorta_Asc | • LN_Ax_L1_L | Full list of OAR is not provided | | | | • A_Carotid_R | • LN_Sclav_R_RTOG | • A_Aorta_Dsc | • LN_Ax_L1_R | | | | | • A_Coronary_R | • Larynx | • A_LAD | • LN_Ax_L2_L | | | | | • A_Iliac_L | • Lens_L | • A_Pulmonary | • LN_Ax_L2_L3_L | | | | | • A_Iliac_R | • Lens_R | • Bladder | • LN_Ax_L2_L3_R | | | | | • A_LAD | • Liver | • Bladder_F | • LN_Ax_L2_R | | | | | • A_Subclavian_L | • Lobe_Temporal_L | • Bone_Ilium_L | • LN_Ax_L3_L | | | | | • A_Subclavian_R | • Lobe_Temporal_R | • Bone_Ilium_R | • LN_Ax_L3_R | | | | | • Anus | • Lung_L | • Bone_Mandible | • LN_Ax_R | | | | | • Atrium_L | • Lung_LLL | • Bone_Pelvic | • LN_IMN_L | | | | Regions of<br>Interest(R<br>OIs) | • Atrium_R | • Lung_LUL | • Bone_Skull | • LN_IMN_R | | | | | • Autochthon_L | • Lung_R | • Bone_Sternum | • LN_IMN_RC_L | | Differ | | | • Autochthon_R | • Lung_RLL | • Bowel | • LN_IMN_RC_R | | ent | | | • Bag_Bowel | • Lung_RML | • Bowel_Bag | • LN_Inguinofem_L | | | | | • Bladder | • Lung_RUL | • Bowel_Large | • LN_Inguinofem_R | | | | | • Bone_Mandible | • OpticChiasm | • Bowel_Small | • LN_Neck_IA | | | | | • Bowel_Large | • OpticNrv_L | • BrachialPlex_L | • LN_Neck_IB-V_L | | | | | • Bowel_Small | • OpticNrv_R | • BrachialPlex_R | • LN_Neck_IB-V_R | | | | | • BrachialPlex_L | • Pancreas | • Brain | • LN_Neck_II_L | | | | | • BrachialPlex_R | • Parotid_L | • Brainstem | • LN_Neck_II_R | | | | | • Brachiocephalic_Trunk | • Parotid_R | • Breast_L | • LN_Neck_II-IV_L | | | | | • Brain | • Pharynx | • Breast_R | • LN_Neck_II-IV_R | | | | | • Brainstem | • Pituitary | • Bronchus | • LN_Neck_II-V_L | | | | | • Breast_L | • Prostate | • BuccalMucosa | • LN_Neck_II-V_R | | | | | • Breast_R | • Rectum | • Carina | • LN_Neck_III_L | | | {10}------------------------------------------------ ## K242994 | • Bronchus_L | • Rib01_L | • CaudaEquina | • LN_Neck_III_R | |---------------------|-----------------|-------------------|----------------------| | • Bronchus_R | • Rib01_R | • Cavity_Oral | • LN_Neck_IV_L | | • CaudaEquina | • Rib02_L | • Cavity_Oral_Ext | • LN_Neck_IV_R | | • Cavity_Oral | • Rib02_R | • Chestwall_L | • LN_Neck_V_L | | • Clavicle_L | • Rib03_L | • Chestwall_OAR | • LN_Neck_V_R | | • Clavicle_R | • Rib03_R | • Chestwall_R | • LN_Neck_VIA | | • Cochlea_L | • Rib04_L | • Chestwall_RC_L | • LN_Neck_VIIA_L | | • Cochlea_R | • Rib04_R | • Chestwall_RC_R | • LN_Neck_VIIA_R | | • Colon | • Rib05_L | • Cochlea_L | • LN_Neck_VIIB_L | | • Kidney_Cortex_L | • Rib05_R | • Cochlea_R | • LN_Neck_VIIB_R | | • Kidney_Cortex_R | • Rib06_L | • Colon_Sigmoid | • LN_Paraaortic | | • Costal_Cartilages | • Rib06_R | • Cornea_L | • LN_Pelvics | | • Duodenum | • Rib07_L | • Cornea_R | • LN_Pelvic_NRG | | • Esophagus | • Rib07_R | • Duodenum | • LN_Sclav_L | | • Eye_L | • Rib08_L | • Ear_Internal_L | • LN_Sclav_R | | • Eye_R | • Rib08_R | • Ear_Internal_R | • LN_Sclav_RADCOMP_L | | • Femur_Head_L | • Rib09_L | • Esophagus | • LN_Sclav_RADCOMP_R | | • Femur_Head_R | • Rib09_R | • External | • Lobe_Temporal_L | | • Femur_L | • Rib10_L | • Eye_L | • Lobe_Temporal_R | | • Femur_R | • Rib10_R | • Eye_R | • Lung_L | | • Gallbladder | • Rib11_L | • Femur_Head_L | • Lung_R | | • Glnd_Adrenal_L | • Rib11_R | • Femur_Head_R | • Macula_L | | • Glnd_Adrenal_R | • Rib12_L | • Femur_L | • Macula_R | | • Glnd_Submand_L | • Rib12_R | • Femur_R | • Marrow_Ilium_L | | • Glnd_Submand_R | • Sacrum | • Femur_RTOG_L | • Marrow_Ilium_R | | • Glnd_Thyroid | • Scapula_L | • Femur_RTOG_R | • Musc_Constrict | | • Gluteus_Maximus_L | • Scapula_R | • GallBladder | • Nipple_L | | • Gluteus_Maximus_R | • Colon_Sigmoid | • Genitals_F | • Nipple_R | | • Gluteus_Medius_L | • Skull | • Genitals_M | • OpticChiasm | | • Gluteus_Medius_R | • SpinalCord | • Glnd_Lacrimal_L | • OpticNrv_L | | • Gluteus_Minimus_L | • Spleen | • Glnd_Lacrimal_R | • OpticNrv_R | | • Gluteus_Minimus_R | • Sternum | • Glnd_Submand_L | • Pancreas | | • Heart | • Stomach | • Glnd_Submand_R | • Parotid_L | | • Hip_L | • Trachea | • Glnd_Thyroid | • Parotid_R | | • Hip_R | • VB_C1 | • HDR_Cylinder | • PenileBulb | | • Hippocampus_L | • VB_C2 | • Heart | • Pericardium | | • Hippocampus_R | • VB_C3 | • Hippocampus_L | • Pituitary | | • Humerus_L | • VB_C4 | • Hippocampus_R | • Prostate | | • Humerus_R | • VB_C5 | • Humerus_L | • Rectum | | • Iliopsoas_L | • VB_C6 | • Humerus_R | • Rectum_F | | • Iliopsoas_R | • VB_C7 | • Kidney_L | • Retina_L | | • Joint_TM_L | • VB_L1 | • Kidney_R | • Retina_R | | • Joint_TM_R | • VB_L2 | • Kidney_Outer_L | • Rib | | • Kidney_L | • VB_L3 | | • Rib_L | {11}------------------------------------------------ # K242994 | | • Kidney_R | • VB_L4 | • Kidney_Outer_R | • Rib_R | | | |---------------------|--------------------------------|------------------------|------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------|---------------| | | • LN_Ax_L1_L | • VB_L5 | • Larynx | • SeminalVes | | | | | • LN_Ax_L1_R | • VB_S1 | • Larynx_Glottic | • SpinalCanal | | | | | • LN_Ax_L2_L | • VB_T01 | • Larynx_NRG | • SpinalCord | | | | | • LN_Ax_L2_R | • VB_T02 | • Larynx_SG | • Spleen | | | | | • LN_Ax_L3_L | • VB_T03 | • Lens_L | • Stomach | | | | | • LN_Ax_L3_R | • VB_T04 | • Lens_R | • Trachea | | | | | • LN_IMN_L | • VB_T05 | • Lips | • UteroCervix | | | | | • LN_IMN_R | • VB_T06 | • Liver | • V_Venacava_I | | | | | • LN_Neck_IA | • VB_T07 | | • V_Venacava_S | | | | | • LN_Neck_IB_L | • VB_T08 | | • VB | | | | | • LN_Neck_IB_R | • VB_T09 | | • VB_C1 | | | | | • LN_Neck_III_L | • VB_T10 | | • VB_C2 | | | | | • LN_Neck_III_R | • VB_T11 | | • VB_C3 | | | | | • LN_Neck_II_L | • VB_T12 | | • VB_C4 | | | | | • LN_Neck_II_R | • V_Brachioceph_L | | • VB_C5 | | | | | • LN_Neck_IVA_L | • V_Brachioceph_R | | • VB_C6 | | | | | • LN_Neck_IVA_R | • V_Iliac_L | | • VB_C7 | | | | | • LN_Neck_IVB_L | • V_Iliac_R | | • VB_L1 | | | | | • LN_Neck_IVB_R | • V_Portal_And_Splenic | | • VB_L2 | | | | | • LN_Neck_VA_L | • V_Pulmonary | | • VB_L3 | | | | | • LN_Neck_VA_R | • V_Venacava_I | | • VB_L4 | | | | | • LN_Neck_VBC_L | • V_Venacava_S | | • VB_L5 | | | | | • LN_Neck_VBC_R | • Ventricle_L | | • VB_T01 | | | | | • LN_Sclav_L_ESTRO | • Ventricle_R | | • VB_T02 | | | | | | | | • VB_T03 | | | | | | | | • VB_T04 | | | | | | | | • VB_T05 | | | | | | | | • VB_T06 | | | | | | | | • VB_T07 | | | | | | | | • VB_T08 | | | | | | | | • VB_T09 | | | | | | | | • VB_T10 | | | | | | | | • VB_T11 | | | | | | | | • VB_T12 | | | | | | | | Windows based .NET<br>front-end application that | | | | Operating<br>System | Local deployment on<br>Windows | | | also serves as agent<br>Uploader supporting<br>Microsoft Windows 10 (64-<br>bit) and Microsoft<br>Windows Server 2016. | Edge & Cloud<br>Deployment | Differ<br>ent | | | | | | Cloud-based Server based<br>automatic contouring<br>application compatible | | | C {12}------------------------------------------------ | | | with Linux. Windows<br>python-based automatic<br>contouring application<br>supporting Microsoft<br>Windows 10 (64-bit) and<br>Microsoft Windows Server<br>2016. | | | |---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------| | Image<br>Format | DICOM | DICOM | DICOM | Same | | General<br>Functions | 1) Deep learning contouring<br>from Head & Neck, Thorax,<br>Abdomen, and Pelvis<br>2) Generates DICOM-RT<br>structure of contoured<br>objects<br>3) Manual Contouring<br>4) Receive, transmit, store,<br>retrieve, display, and process<br>medical images and DICOM<br>objects | 1) Automatically contour<br>various structures of<br>interest for radiation<br>therapy treatment<br>planning<br>2) Allow the user to review<br>and modify the resulting<br>contours<br>3) Generate DICOM-<br>compliant structure set<br>data the can be imported<br>into a radiation therapy<br>treatment planning system | CT or MR series of<br>images serve as<br>input for AI-Rad<br>Companion Organs<br>RT and are acquired<br>as part of a typical<br>scanner acquisition.<br>Once processed by<br>the AI algorithms,<br>generated contours<br>in DICOM-RTSTRUCT<br>format are reviewed<br>in a confirmation<br>window, allowing<br>clinical user to<br>confirm or reject the<br>contours before<br>sending to the target<br>system. Optionally,<br>the user may select<br>to directly transfer<br>the contours to a<br>configurable DICOM<br>node | Same | | Algorithm | Deep Learning | Deep Learning | Deep Learning | Same | | Compatibl<br>e Modality | CT Images. DICOM<br>RTSTRUCT for output | CT or MR input for<br>contouring or<br>registration/fusion. PET/CT<br>input for<br>registration/fusion only.<br>DICOM RTSTRUCT for<br>output | CT & MR Images | Equiv<br>alent | | Segmentat<br>ion of<br>Organ | Head & Neck, Thorax,<br>Abdomen, Pelvis | Head and Neck, Thorax,<br>Abdomen, Pelvis | Head & Neck,<br>Thorax, Abdomen &<br>Pelvis Head & Neck<br>lymph nodes | Same | | Workflow | Automatically processes input<br>image data contour organs | Automatically contour<br>various structures of | AI-Rad Companion<br>Organs RT | Same | | | and DICOM sends generated<br>RT Structure set | interest for radiation<br>therapy treatment<br>planning using machine<br>learning based contouring | automatically<br>processes input<br>image data and<br>sends the results as<br>DICOM-RT Structure<br>Sets to a user-<br>configurable target<br>node | | | Compatibl<br>e Scanner<br>Models | No Limitation on scanner<br>model, DICOM 3.0<br>compliance required. | No Limitation on scanner<br>model, DICOM 3.0<br>compliance required. | No Limitation on<br>scanner model,<br>DICOM 3.0<br>compliance required. | Same | | Compatibl<br>e<br>Treatment<br>Planning<br>System | No Limitation on TPS model,<br>DICOM 3.0 compliance<br>required. | No Limitation on TPS<br>model, DICOM 3.0<br>compliance required. | No Limitation on TPS<br>model, DICOM 3.0<br>compliance required. | Same | | Patient<br>Population | Adult only | Adult only | Adult only | Same | {13}------------------------------------------------ The intended use of the predicate device and the subject device are equivalent. Both devices are intended to aid users to contour the body structure using artificial intelligence algorithm that can be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed. A detailed comparison shows the subject device is substantially equivalent in indications for use, image format, general functions, algorithm, segmentation of organs, workflow, compatible scanner models, compatible treatment planning system and patient population to the predicate device. The differences between the subject and the predicate devices do not raise any new questions regarding safety and effectiveness. #### 8. Non-Clinical Test summary The following data were provided in support of the substantial equivalence determination: - 1) Software Validation The OncoStudio contains basic document level of concern software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA guidance: - "Content of Premarket Submissions for Device Software Functions," dated June 14, 2023. . - 2) Performance characteristics {14}------------------------------------------------ We collected training data from mainly three datasets of source : OneMedNet, Yonsei Severance Hospital, and University Hospital Basel, Switzerland. OneMedNet is a purchased set of CT data, mainly comprised of U.S.A. population. Yonsei Severance Hospital is located in South Korea, and we collected mainly Eastern population data from this source. The Basel data is known as TotalSegmentator dataset, which is open for public at this moment. The collected data comprises a total of 2,438 dataset, consisting of 315 datasets from the US, 871 from Korea, and 1,252 from Europe. The data was constructed with various ethnics (White, Black, Asian, Hispanic, Latino, African, American, etc.), and the training model can be obtained by performing generalization without differences according to ethnicity. We allocated 310 data as training dataset, which is more than one-tenth of the total 2,438 dataset, and used the remaining data(2128) for training. For evaluation, we created a test dataset that was not involved in any kind of training process. The splitting was performed at the patient level to ensure that images from the same patient were not present in more than one dataset. A comprehensive audit was conducted to confirm the integrity of the data-splitting process and ensure that no patient overlap occurred between datasets. Ground truth seqmentations were established by three radiation oncologists following international clinical quidelines. #### a) Ground Truthing The ground truth annotations for the dataset of Yonsei Severance Hospital(Korea) and OneMedNet(U.S) were established by three different radiation oncologists with 3-20 years of clinical practice following RTOG and clinical quidelines using manual annotation. The radiation oncologists included associate professor, assistant professor, and radiation oncologist resident from two institutions (Yonsei Cancer Center, Samsung Seoul Hospital) - . Ground Truthing process - First, the 1 radiation oncologist manually delineated the organs - Second, seqmentation results generated by 1 radiation oncologist are sequentially edited and confirmed by 2 radiation oncologists. In this editing process, the first radiation oncologist makes corrections, and the corrected results are received and finalized by another radiation oncologist. {15}------------------------------------------------ In case of University Hospital Basel (Europe) dataset is public data comprising 104 anatomical structures. A total of 1,368 CT images were randomly sampled from the years 2012, 2016, and 2020 from the University Hospital Basel through picture archiving and communication system (PACS). The Nora Imaging Platform was used for manual segmentation and further refinement of generated segmentations for ground truth. Segmentation was supervised by two physicians with 3 (M.S.) and 6 years (H.B.) of experience in body imaging, respectively. #### b) Training Out of a total of 2,438 images, 2,128 were allocated as training data. The allocated training data consists of 731 images from Yonsei Severance Hospital(Republic of Korea), 194 images from OneMedNet(U.S.A), and 1203 images from University Hospital Basel(Switzerland). The training datasets consist of 62% of Contrast CT and 38% of Non-Contrast CT. The study population comprises 62% males and 38% females, with 22% under 49 years old, 47% aged 50-70 years, and 31% over 70 years old. The data was constructed with various ethnics (White, Black, Asian, Hispanic, Latino, African, American, etc.), and the result can be obtained by performing generalization without performance differences according to ethnicity. The acquired data encompasses CT manufacturers such as GE (2%), Siemens (72%), Philips (9%), Toshiba (11%), and unknown manufacturer(6%) #### c) Segmentation Performance Test A standalone performance test was conducted to compare the contouring capabilities of OncoStudio. The dataset used in this test comprises a total of 310 CT images, with 140, 121, and 49 images collected from Yonsei Severance Hospital (Republic of Korea), OneMedNet (U.S.A.), and University Hospital Basel (Switzerland), respectively, each meeting the established inclusion criteria. All data used during the standalone performance evaluation was composed independently of product development training. The images consist of 54% of Contrast CT and 46% of Non-Contrast CT. The study population comprises 58% males and 42% females, with 13% under 49 years old, 54% aged 50-70 years, and 28% over 70 years old. The data was constructed with various ethnics (White, Black, Asian, Hispanic, Latino, African, American, etc.), and the result can be obtained by performing generalization without performance differences according to ethnicity. {16}------------------------------------------------ The acquired data encompasses CT manufacturers such as GE (2%), Siemens (56%), Philips (14%), Toshiba (22%), and unknown manufacturer(6%) Ground truth segmentations were established by three radiation oncologists following international clinical guidelines. For the structures being compared, the mean Dice coefficient (DSC) of structures for each anatomical region (Head & Neck, Thorax, Abdomen, and Pelvis) should meet the established criteria. 3) Cybersecurity · "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", on September 27, 2023 #### 9. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92] There are no significant differences between the subject, predicate and reference devices, K230685 and K232899 that would adversely affect the use of the product. It is substantially equivalent to the predicate device in indications for use and technology characteristics. #### 10. Conclusion [21 CFR 807.92(b)(3)] In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that the OncoStudio is substantially equivalent in safety and effectiveness to the predicate device as described herein.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...