Cristaline Aligners Z FLX are series of clear, lightweight, plastic appliances, indicated for treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental tooth movements, it sequentially positions teeth by way of continuous force.
Device Story
Cristaline Aligners Z FLX are clear, thin, thermoformed removable trays for correcting tooth malocclusions without wires or brackets. Input: digital dental scan provided by clinician. Process: clinician-approved digital treatment plan created using commercially available software; 3D-printed molds generated; aligners thermoformed from biocompatible thermoplastic (co-polyester/polyurethane composite). Output: series of customized aligners. Usage: prescribed by dentist/orthodontist; patient wears each tray for 1-2 weeks, ~22 hours/day. Mechanism: continuous gentle force for progressive tooth realignment. Benefit: non-invasive, removable orthodontic correction.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 (cytotoxicity, oral mucosal irritation, skin sensitization). Bench testing validated manufacturing process accuracy relative to digital scans. No animal or human clinical performance testing required for this device category.
Technological Characteristics
Material: thermoplastic co-polyester/polyurethane composite. Form factor: clear, thin, thermoformed removable trays. Energy source: none (mechanical force). Connectivity: N/A. Sterilization: non-sterile. Software: commercially available treatment planning software used for digital modeling.
Indications for Use
Indicated for treatment of tooth malocclusions in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
Predicate Devices
Blue Sky Plan Software, Blue Sky Bio Aligner Z and Blue Sky Bio Z FLX (K232653)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
December 19, 2024
Cristaline Aligners GmbH % Patsy Trisler Regulatory Consultant Trisler Consulting (dba) 306 Turnberry Court Lebanon, Indiana 46052
Re: K242892
Trade/Device Name: Cristaline Aligners Z FLX Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: September 18, 2024 Received: September 23, 2024
Dear Patsy Trisler:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242892
Device Name
Cristaline Aligners Z FLX
#### Indications for Use (Describe)
Cristaline Aligners A FLX are series of clear, lightweight, plastic appliances, indicated for treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental tooth movements, it sequentially positions teeth by way of continuous force.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
| Submitter: | Cristaline Aligners Z FLX | | |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Cristaline Aligners GmbH | K242892 | Traditional Premarket Notification 510(k) | |
| SUBMITTER Name: | Cristaline Aligners GmbH | | |
| Address: | Hanauer Strasse 1-5<br>75181 Pforzheim, Germany | | |
| Contact Person: | Dr. Boris Simnovski | | |
| Email: | dr.simanovoski@gmail.com | | |
| Date Prepared: | November 13, 2024 | | |
| DEVICE Trade<br>Name: | Cristaline Aligners Z FLX | | |
| Common Name: | Aligner, Sequential (Clear Braces) | | |
| Classification Name | Orthodontic Plastic Bracket | | |
| Number | 21 CFR 872.5470 | | |
| Product Code | NXC | | |
| Regulatory Class | 2 | | |
| Review Panel | Dental | | |
| PRIMARY<br>PREDICATE: | K232653, Blue Sky Plan Software, Blue Sky Bio Aligner Z and<br>Blue Sky Bio Z FLX; Blue Sky Bio, LLC | | |
| DEVICE<br>DESCRIPTION | Cristaline Aligners Z FLX are comprised of a series of clear,<br>thin, thermoformed removable aligner trays that are designed to<br>correct tooth malocclusions without the use of conventional<br>wire and bracket orthodontic technology. They are<br>manufactured using a biocompatible thermoplastic sheet<br>composed of a composite of co-polyester and polyurethane.<br>They are provided non-sterile and are customized for each<br>patient according to the dental clinician's prescription.<br>The dental health professional (dentist/orthodontist) provides a<br>digital file (scan) of the patient's teeth to Cristaline. Based on<br>the dental health professional's treatment plan, Cristaline<br>develops a digital plan using commercially available treatment<br>planning software. Upon approval by the dental health<br>professional, molds are created with 3D-printing technology and<br>the clear aligners are formed on the molds.<br>The finished, customized aligners are provided to the dental<br>health care professional who provides them to the patient,<br>confirming fit and design. | | |
| MECHANISM OF<br>ACTION | Each aligner in the set is used for the specified period of time,<br>usually 1-2 weeks, to exert gentle force to achieve progressive<br>realignment of the teeth until the final correction has been<br>achieved. The daily treatment time is approximately 22 hours,<br>or except during eating, based on the clinician's prescribed<br>treatment plan.<br>The aligner trays are held in place by pressure and can be<br>removed by the patients at any time. | | |
| SUMMARY OF<br>TECHNOLOGICAL<br>CHARACTERISTICS | The thermoplastic materials used for the manufacture of the<br>aligners are identical to the material used for making the<br>Predicate device aligner Z FLX. The treatment planning<br>software system used is the Predicate Blue Sky Plan software. | | |
| INDICATIONS FOR<br>USE STATEMENT | Cristaline Aligners Z FLX are series of clear, lightweight, plastic<br>appliances, indicated for treatment of tooth malocclusions in<br>patients with permanent dentition (i.e. all second molars).<br>Utilizing a series of incremental tooth movements,<br>sequentially positions teeth by way of continuous force. | | |
| SAFETY TESTING | Testing of the aligner material has been conducted according to<br>ISO 10993, Biological evaluation of medical devices – Part 1:<br>Evaluation and testing within a risk management system, was<br>performed by a GLP-certified contract research laboratory:<br>• In vitro Cytotoxicity (Part 5)<br>• Oral Mucosal Irritation (Part 10)<br>• Skin Sensitization (Part 10) | | |
| OTHER TESTING | Bench testing was performed to validate the manufacturing<br>process to ensure the accuracy of the final aligners compared<br>to the initial digital scans. Final reports of the testing of aligners<br>fabricated with the thermoplastic material was part of the 510(k)<br>package.<br>In vivo Animal and Human Clinical performance testing are not<br>required for this device category. | | |
| COMPARISON TO<br>THE PREDICATE<br>DEVICE | The Cristaline Aligners Z FLX devices have the same intended<br>use as the predicate device. The thermoplastic polymer<br>materials are the same as the Predicate and digital treatment<br>planning involves use of the Predicate's commercially available<br>software The processes for fabrication of the clear aligners,<br>while proprietary to each manufacturer, makes use of similar,<br>industry-standard processes using similar machines. Any<br>differences in the specific manufacturing processes do not raise<br>new questions of safety and effectiveness. | | |
| SUBSTANTIAL<br>EQUIVALENCE<br>CONCLUSION | The information and data provided in this 510(k) establish that<br>the Cristaline Aligners Z FLX devices are substantially<br>equivalent to the Predicate Blue Sky Bio device in the intended<br>use, design, principle of operation, technology (same treatment<br>planning software as the Predicate), and thermoplastic material<br>(same as aligner Z FLX of the Predicate device). See the<br>following SE Comparison table. | | |
| | Proposed Device | Predicate #1 | SE Comparison |
| 510(k) Number<br>Device Name | K242892<br>Cristaline Aligners Z FLX | K232653<br>Blue Sky Plan Software, Blue | N/A |
| | | Sky Bio Aligner Z and Blue Sky<br>Bio Z FLX | |
| Manufacturer | Cristaline Aligners GmbH | Blue Sky Bio, LLC | N/A |
| Classification<br>Regulation and<br>Name<br>Product Code<br>Class | 21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>NXC<br>2 | 21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>NXC<br>2 | Same |
| Indications for<br>Use | Cristaline Aligners Z FLX are<br>series of clear, lightweight,<br>plastic appliances, indicated for<br>treatment of tooth malocclusions<br>in patients with permanent<br>dentition (i.e. all second molars).<br>Utilizing a series of incremental<br>tooth movements, it sequentially<br>positions teeth by way of<br>continuous force. | Blue Sky Bio Aligner Z and Blue<br>Sky Bio Aligner Z FLX are series<br>of clear, lightweight, plastic<br>appliances, indicated for<br>treatment of tooth malocclusions<br>in patients with permanent<br>dentition (i.e. all second molars).<br>Utilizing a series of incremental<br>tooth movements, it sequentially<br>positions teeth by way of<br>continuous force. | Same for the Z FLX<br>Aligner. |
| Mode of Action | The removable appliance<br>applies continuous gentle forces<br>on teeth according to the plan<br>prescribed by the doctor. | The removable appliance<br>applies continuous gentle forces<br>on teeth according to the plan<br>prescribed by the doctor. | No differences |
| Description of<br>Use | Each removable preformed<br>plastic tray, prescribed by the<br>Dr, is worn by the patient usually<br>for 1-2 weeks, prior to using the<br>next sequential aligner tray<br>according to Dr's direction. | Each removable preformed<br>plastic tray, prescribed by the Dr,<br>is worn by the patient usually for<br>a couple weeks, prior to using the<br>next sequential aligner tray<br>according to Dr's direction. | No differences |
| Material | Thermoplastic co-polyester<br>polyurethane composite. | For Z FLX aligner: thermoplastic<br>co-polyester polyurethane<br>composite | No differences |
| Manufacturing<br>Process | Forming of plastic sheets on SD<br>printed dental models designed<br>with use of the treatment<br>planning software | Forming of plastic sheets on SD<br>printed dental models designed<br>with use of the treatment<br>planning software | Manufacturing<br>processes used are<br>standard in the field<br>and similar to<br>predicate's. |
| Software | Commercially available<br>treatment planning software is<br>used - same as Predicate | Included in 510(k) | Same software is used |
| Prescription<br>Use | Rx | Rx | Both are Rx |
| Biocompatibility | Testing to meet requirements for<br>Category C | Testing to meet requirements for<br>Category C | Relied on the same<br>testing performed by<br>thermoplastic material<br>manufacturer. |
{6}------------------------------------------------
{7}------------------------------------------------
## Substantial Equivalence Comparison Table
{8}------------------------------------------------
| Submitter: | | Cristaline Aligners Z FLX |
|--------------------------|---------|-------------------------------------------|
| Cristaline Aligners GmbH | K242892 | Traditional Premarket Notification 510(k) |
| <b>Process</b><br><b>Validation</b> | Validation to determine fit and<br>form of manufactured aligners. | Similar testing<br>performed. |
|-------------------------------------|-------------------------------------------------------------------|-------------------------------|
|-------------------------------------|-------------------------------------------------------------------|-------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.