K242834 · Stryker Instruments · KCT · Jan 16, 2025 · General Hospital
Device Facts
Record ID
K242834
Device Name
System 9 Sterile Battery Container
Applicant
Stryker Instruments
Product Code
KCT · General Hospital
Decision Date
Jan 16, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container intended to allow for sterilization, by hospitals, surgery centers, and healthcare facilities, of the enclosed, compatible Stryker Sterile Battery Packs and maintain sterility of the enclosed device until used. The container is validated for use with compatible battery packs in low temperature hydrogen peroxide sterilization cycles.
Device Story
System 9 Sterile Battery Container is a reusable rigid container for sterilizing Stryker battery packs via low-temperature hydrogen peroxide cycles. Device consists of a Radel (PPSU) base and stainless-steel lid with filter retention plate and single-use polypropylene filters. User loads battery pack into base, attaches lid, secures latching mechanism, and applies tamper-evident lock. Container maintains sterile barrier for up to 45 days post-sterilization. Used in healthcare facilities; operated by clinical staff. Output is a sterilized, protected battery pack ready for clinical use. Sterilization verification performed via indicator strips. Design allows battery charging while maintaining sterile barrier. Benefits include safe, efficient sterilization and storage of power components for surgical equipment.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via sterilization efficacy (ISO 14937, ISO 22441), toxicological risk assessment (ISO 10993-17), cleaning validation (AAMI ST98), maintenance of sterility (AAMI ST77), simulated use design validation, and human factors engineering (IEC 62366-1).
Technological Characteristics
Rigid, reusable container. Base: Radel (PPSU). Lid: Stainless steel. Filter: Single-use polypropylene. Sterilization: Low-temperature hydrogen peroxide (STERRAD/Steris V-PRO cycles). Dimensions: Small (3.2"H x 4.2"W x 12.9"L), Large (3.9"H x 4.2"W x 12.9"L). Standards: ISO 14937, ISO 22441, ISO 10993-17, ISO 17664-1, AAMI TIR 12, AAMI ST98, AAMI ST77, IEC 62366-1.
Indications for Use
Indicated for use by hospitals, surgery centers, and healthcare facilities to sterilize and maintain the sterility of enclosed, compatible Stryker Sterile Battery Packs until use. Not intended for direct patient contact.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 16, 2025
Stryker Instruments Megan Guilbault Staff Regulatory Affairs Specialist 1941 Stryker Way Portage, Michigan 49002
Re: K242834
Trade/Device Name: System 9 Sterile Battery Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: September 19, 2024 Received: September 19, 2024
Dear Megan Guilbault:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Stephen A. Anisko -S
Digitally signed by Stephen A. Anisko -S Date: 2025.01.16 20:37:15 -05'00'
for: Christopher K. Dugard Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242834
Device Name
System 9 Sterile Battery Container
Indications for Use (Describe)
The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container intended to allow for sterilization, by hospitals, surgery centers, and healthcare facilities, of the enclosed, compatible Stryker Sterile Battery Packs and maintain sterility of the enclosed device until used. The container is validated for use with compatible battery packs in low temperature hydrogen peroxide sterilization cycles.
Validated Low Temperature Hydrogen Peroxide Sterilization Cycles: Load Configuration Cycles
System 9 Sterile Battery Container Bundle, Small: STERRAD NX Standard cycle STERRAD 100NX Standard cycle STERRAD 100S short cycle Steris V-PRO maX Lumen cycle Steris V-PRO maX 2 Lumen cycle Steris V-PRO 60 Lumen cycle Steris V-PRO s2 Lumen cycle
System 9 Sterile Battery Container Bundle, Large: STERRAD NX Standard cycle STERRAD 100NX Standard cycle STERRAD 100S short cycle Steris V-PRO maX Lumen cycle
Steris V-PRO maX 2 Lumen cycle
Steris V-PRO 60 Lumen cycle
Steris V-PRO s2 Lumen cycle
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 1.0 Submitter Information
## 1.1 This Premarket Notification is submitted by:
Stryker Instruments 1941 Stryker Way Portage MI 49002 USA
## 1.2 Contact Information
| Contact name: | Megan Guilbault |
|---------------------|-----------------------------|
| Cellular Telephone: | (401) 241-6152 |
| Email: | megan.guilbault@stryker.com |
| Date Prepared: | January 15, 2025 |
### 1.3 Device Name
### Table 1: Subject Device Information
| Subject Device Information | |
|-----------------------------|---------------------------------------------------------------------|
| Trade/Proprietary Name | System 9 Sterile Battery Container |
| Common Name | Sterilization Container Wrap |
| Classification | Class II |
| Classification Product Code | KCT |
| Classification Name | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories |
| Classification Regulation | 21 CFR 880.6850 |
| Review Panel | General Hospital |
### 1.4 Predicate Device
| Table 2: Predicate Device Information | | | |
|---------------------------------------|---------------|--------------|----------------|
| Predicate Device Trade Name | 510(k) Number | Product Code | Manufacturer |
| Aesculap Aicon™ Container | K214041 | KCT | Aesculap, Inc. |
## 1.5 Device Description
The Stryker System 9 Sterile Battery Container is a rigid, reusable device intended to enclose compatible Stryker sterile battery packs for hydrogen peroxide sterilization. The container consists of a stainless-steel perforated lid with a stainless-steel filter retention plate and a plastic (Radel®) base. The Sterile Battery container utilizes single use polypropylene filters and indicator cards. The container base is designed to exclusively fit onto the Stryker sterile battery charger to allow the sterile battery pack to charge while maintaining a sterile barrier. The Stryker System 9 Sterile Battery Container encompasses one distinct configuration in two sizes, small and large.
{6}------------------------------------------------
# 1.6 Indications for Use
The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container intended to allow for sterilization, by hospitals, surgery centers, and healthcare facilities, of the enclosed, compatible Stryker Sterile Battery Packs and maintain sterility of the enclosed device until used. The container is validated for use with compatible battery packs in low temperature hydrogen peroxide sterilization cycles.
| Load Configuration | Cycles |
|--------------------------------------------------------|--------------------------------|
| System 9 Sterile Battery<br>Container Bundle,<br>Small | STERRAD NX Standard cycle |
| | STERRAD 100NX Standard cycle |
| | STERRAD 100S short cycle |
| | Steris V-PRO maX Lumen cycle |
| | Steris V-PRO maX 2 Lumen cycle |
| | Steris V-PRO 60 Lumen cycle |
| | Steris V-PRO s2 Lumen cycle |
| System 9 Sterile Battery<br>Container Bundle,<br>Large | STERRAD NX Standard cycle |
| | STERRAD 100NX Standard cycle |
| | STERRAD 100S short cycle |
| | Steris V-PRO maX Lumen cycle |
| | Steris V-PRO maX 2 Lumen cycle |
| | Steris V-PRO 60 Lumen cycle |
| | Steris V-PRO s2 Lumen cycle |
Validated Low Temperature Hydrogen Peroxide Sterilization Cycles:
# 1.7 Comparison of Technological Characteristics
A comparison of the technological characteristics of the subject devices included in the scope of this Traditional 510(k) is included in Table 3 below.
| Description | Subject Device | Predicate Device | Discussion |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container | A device intended to be used to enclose another medical device that is to | Similar |
| Description | Subject Device<br>System 9 Sterile Battery<br>Container | Predicate Device<br>Aesculap AICON TM<br>Container | Discussion |
| | intended to allow for<br>sterilization, by hospitals,<br>surgery centers, and healthcare<br>facilities, of the enclosed,<br>compatible Stryker Sterile<br>Battery Packs and maintain<br>sterility of the enclosed device<br>until used. | be sterilized by a health<br>care provider. It is<br>intended to allow<br>sterilization of the<br>enclosed medical device<br>and also to maintain<br>sterility of the enclosed<br>device until used. | |
| Classification | Class II | Class II | Identical |
| Product<br>Code | KCT | KCT | Identical |
| Condition of<br>Use | Professional Use Only,<br>Reusable, for use in a<br>healthcare facility | Professional Use Only,<br>Reusable, for use in a<br>healthcare facility | Identical |
| Patient<br>Population | Not intended for direct patient<br>contact | Not intended for direct<br>patient contact | Identical |
| Operating<br>Principle | The user loads Stryker Sterile<br>Battery Packs into the Sterile<br>Battery Container Base. The<br>Sterile Battery Container Lid<br>with installed filter and filter<br>retention plate is attached to the<br>Sterile Battery Container Base,<br>the external latching<br>mechanism is closed, and the<br>Tamper-Evident Lock is placed<br>on the outside latch. The lid,<br>filter, filter retention plate, and<br>base all generate a sterile<br>barrier. The loaded container<br>becomes sterile once it has been<br>run through a low-temperature<br>hydrogen peroxide sterilizer.<br>The indicator strips color-<br>change mechanism is used to<br>verify successful sterilization.<br>Once sterilized, the container<br>maintains sterility until opened,<br>up to 45 days. | The user places<br>instruments inside the<br>container in a manner to<br>allow proper sterilant<br>penetration. The<br>container lid is closed<br>securely with a locking<br>mechanism to ensure<br>airtight seal. The<br>container is placed in the<br>sterilization chamber<br>(under various<br>sterilization modalities).<br>After completion of<br>appropriate sterilization<br>cycle, verification of<br>sterilization is conducted<br>using indicator strips<br>color-change<br>mechanism. The<br>containers are then<br>stored in a clean, dry<br>environment to maintain<br>sterility and opened in a<br>sterile area when ready<br>to use. Once sterilized<br>the container maintains | Identical |
| | Subject Device | Predicate Device | Discussion |
| Description | System 9 Sterile Battery<br>Container | Aesculap AICON™<br>Container<br>sterility for 365 days or<br>until opened. | |
| Sterilization<br>Method | Low-temperature hydrogen<br>peroxide sterilization STERRAD NX Standard cycle STERRAD 100NX Standard cycle STERRAD 100S Short cycle Steris V-PRO maX Lumen cycle Steris V-PRO maX 2 Lumen cycle Steris V-PRO 60 Lumen cycle Steris V-PRO s2 Lumen cycle | Low-temperature<br>hydrogen peroxide<br>sterilization Steam Sterilization STERIS V-PRO maX/maX 2: Lumen,<br>Non-Lumen, Flex STERIS V-PRO 60:<br>Lumen, Non- Lumen, Flex STERRAD 100NX:<br>Standard, Express, Flex, Duo STERRAD NX:<br>Standard, Advanced STERRAD 100S STERIZONE VP4 EtO | Similar |
| Container<br>Base<br>Material | Radel® PPSU<br>Polyphenylsulfone | Anodized Aluminum | Different<br>The predicate<br>device is made<br>from material<br>that is<br>compatible with<br>a variety of<br>sterilization<br>methods. As the<br>subject device is<br>only being<br>cleared with<br>hydrogen<br>peroxide<br>sterilization,<br>Stryker has<br>chosen material<br>suitable for the<br>specific<br>sterilization<br>method and to<br>prevent<br>interference for<br>wireless charging |
| | Subject Device | Predicate Device | Discussion |
| Description | System 9 Sterile Battery<br>Container | Aesculap AICONTM<br>Container | |
| Container<br>Lid Material | Stainless Steel | Anodized Aluminum | Different<br>The predicate<br>device is made<br>from material<br>that is<br>compatible with<br>a variety of<br>sterilization<br>methods. As the<br>subject device is<br>only being<br>cleared with<br>hydrogen<br>peroxide<br>sterilization,<br>Stryker has<br>chosen material<br>suitable for the<br>specific<br>sterilization<br>method.<br>Verification and<br>Validation<br>Testing confirms<br>that the subject<br>device is safe and<br>effective for its<br>intended use and<br>the differences<br>do not raise<br>different types of<br>questions of |
| | | | battery and<br>charger.<br>Verification and<br>Validation<br>Testing confirms<br>that the subject<br>device is safe and<br>effective for its<br>intended use and<br>the differences<br>do not raise<br>different types of<br>questions of<br>Safety and<br>Effectiveness. |
| Description | Subject Device<br>System 9 Sterile Battery<br>Container | Predicate Device<br>Aesculap AICON™<br>Container | Discussion |
| Filter<br>Material | Polypropylene (Single Use) | Polypropylene (Single Use) | Identical |
| Maintenance<br>of Sterility | 45 Days | 365 Days | Different<br>The predicate<br>device maintains<br>a sterile barrier<br>on the shelf for<br>365 days,<br>whereas the<br>subject maintains<br>a sterile barrier<br>for 45 days.<br>Testing did not<br>raise any<br>different types of<br>questions of<br>Safety and<br>Effectiveness. |
| Device Size | Small: 3.2" H x 4.2" W x 12.9" L<br><br>Large: 3.9" H x 4.2" W x 12.9" L | Full size: 4" H x 11 ¾" W x 23 5/8" L<br><br>Full size: 6" H x 11 ¾" W x 23 5/8" L<br><br>Full size: 8" H x 11 ¾" W x 23 5/8" L<br><br>Full size:10" H x 11 ¾" W x 23 5/8" L<br><br>Three-Quarter: 4" H x<br>11 ¾" W x 18 3/8" L<br><br>Three-Quarter: 6" H x<br>11 ¾" W x 18 3/8" L<br><br>Three-Quarter: 8" H x<br>11 ¾" W x 18 3/8" L<br><br>Three-Quarter:10" H x<br>11 ¾" W x 18 3/8" L | Different<br>The predicate<br>offers a larger<br>range of sizes<br>while the subject<br>device is<br>designed to<br>specifically be<br>used with the<br>System 9 Sterile<br>Batteries, which<br>are available in<br>two sizes, small<br>and large.<br>Verification<br>testing confirms<br>that the subject<br>device is safe and<br>effective for its<br>intended use and<br>the differences<br>do not raise<br>different types of<br>questions of<br>Safety and<br>Effectiveness. |
| Description | Subject Device | Predicate Device | Discussion |
| | System 9 Sterile Battery Container | Aesculap AICON™ Container | |
| | | Half size: 4" H x 11 ¾" W x 11 7/8" L | |
| | | Half size: 6" H x 11 ¾" W x 11 7/8" L | |
| | | Half size: 8" H x 11 ¾" W x 11 7/8" L | |
| | | Half size: 10" H x 11 ¾" W x 11 7/8" L | |
| Container Design | Solid Bottom | Solid Bottom | Identical |
| | Perforated Lid with filter retention plates and single use polypropylene filters | Perforated Lid with filter retention plates and single use polypropylene filters | |
| | Gasket | Gasket | |
Table 3: Technological Comparison
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
# 1.8 Summary of Non-Clinical Testing
The intended use of the subject device and the predicate device are identical, and the technological characteristics are similar. The differences between the two devices do not raise any new or different types of questions of safety and effectiveness.
Risk management was conducted in accordance with ISO 14971.
A suite of non-clinical testing was performed in accordance with applicable standards. See table below:
| | | | | Table 4: Performance Testing | |
|--|--|--|--|------------------------------|--|
|--|--|--|--|------------------------------|--|
| Testing | Standard in<br>compliance with | Acceptance Criteria | Results |
|-------------------------------------|----------------------------------|-------------------------------|---------|
| Sterilization<br>Efficacy | ISO 14937:2009<br>ISO 22441:2022 | Confirms sterile efficacy | Pass |
| Toxicological<br>Risk<br>Assessment | ISO 10993-17 | Confirms acceptable residuals | Pass |
{12}------------------------------------------------
| Testing | Standard in<br>compliance with | Acceptance Criteria | Results |
|---------------------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Cleaning<br>Validation | AAMI ST98:2022 | Cleaning instructions were validated and<br>demonstrated that the subject devices<br>could be visually and quantifiably cleaned | Pass |
| Maintenance of<br>Sterility | AAMI<br>ST77:2013/(R)2018 | Demonstrates that the subject device is<br>capable of maintaining product sterility for<br>the duration of the shelf life | Pass |
| Simulated Use<br>Design<br>Validation | ISO 17664-1:2021<br>AAMI TIR 12:2020<br>ST77:2013/(R)2018 | Confirms that the device can undergo<br>repeated processing to support that the<br>device's<br>or<br>performance is maintained following a<br>minimum number of exposures to the<br>proposed sterilization methods | Pass |
| Human Factors<br>Engineering | IEC 62366-1 Ed 1.1 b<br>2020 | Confirms safety and effectiveness of the<br>subject device for the intended users, uses,<br>and use environments | Pass |
Performance testing was performed in compliance with the following standards:
- ISO 14937:2009 Sterilization of health care products Low temperature . vaporized hydrogen peroxide - Requirements for the development, valid…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.