OsteoCentric Technologies Cannulated Fasteners and Nuts
Applicant
OsteoCentric Technologies
Product Code
HWC · Orthopedic
Decision Date
Oct 9, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The OsteoCentric Cannulated Fasteners and Nuts is indicated for use in bone reconstruction, osteotomy, arthrodesis (joint fusion), and fixation of fractures and ligaments: The OsteoCentric 2.4mm and 3.0mm Cannulated Screw are intended for fixation of fractures and non-unions of small bones and small bone arthrodesis. Examples include scaphoid and other carpal fractures, metacarpal and phalangeal fusions, osteotomies, and bunionectomies. The OsteoCentric 3.5mm Cannulated Screws are intended for the arthrodesis or fixation of fractures and fragments of hand, midfoot, and forefoot bones, extremity osteochondritis dissecans lesions, and for extremity ligament fixation. The OsteoCentric 4.5mm to 8.0mm Cannulated Screws are intended for fixation of long bone fractures and fragments. Iong bone osteotomies, femoral neck fractures, slipped capital femoral epiphyses, as an adjunct to treatment with a dynamic hip screw (DHS) in basilar neck fractures, tibial plateau fractures, pediatric femoral neck fractures, intercondylar femur fractures, fixation of the pelvis and pelvic joints, and arthrodesis (e.g., ankle, subtalar). The OsteoCentric 7.0mm and 8.0mm Cannulated Nut is indicated for reduction and stabilization of fractures or inter-osseous distances (e.g. syndesmosis injuries).
Device Story
Cannulated fasteners and nuts provide mechanical support and stabilization for orthopedic procedures including bone reconstruction, osteotomy, arthrodesis, and fracture fixation. Device consists of metallic screws and nuts in various diameters and lengths to accommodate patient anatomy. Used by surgeons in clinical settings; fasteners provide rigid fixation to facilitate bone healing. Device is non-sterile; requires surgical implantation. Benefits include stabilization of bone fragments and joints to promote union and restore function.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological identity to the predicate device and established performance characteristics.
Technological Characteristics
Metallic bone fixation fasteners (screws and nuts). Materials: Stainless Steel (ASTM F138) or Titanium (ASTM F136 or F1295). Cannulated design for guidewire-assisted insertion. Provided non-sterile; reusable instrumentation; single-use guidewires. No software or electronic components.
Indications for Use
Indicated for bone reconstruction, osteotomy, arthrodesis, and fixation of fractures and ligaments in patients requiring orthopedic stabilization. Specific applications vary by screw diameter (2.4mm-8.0mm) for small bones (carpal, metacarpal, phalangeal), midfoot/forefoot bones, and long bones (femoral, tibial, pelvic).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
OsteoCentric Technologies Cannulated Fasteners and Nuts (K211290)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol of the Department of Health & Human Services. To the right, the FDA logo is displayed in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, reflecting the FDA's role in regulating and overseeing food and drug products in the United States.
October 9, 2024
OsteoCentric Technologies Todd Evans Vice President of Quality & Regulatory Affairs 75 West 300 North Suite 150 Logan, Utah 84321
Re: K242691
Trade/Device Name: OsteoCentric Technologies Cannulated Fasteners and Nuts Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, HTN Dated: September 5, 2024 Received: September 9, 2024
Dear Todd Evans:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image shows the name "Shumaya Ali-S" in a simple, sans-serif font. The text is black against a white background, making it easily readable. The name appears to be a person's name, with the "-S" possibly indicating an initial or a suffix.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K242691
Device Name
OsteoCentric Technologies Cannulated Fasteners and Nuts
Indications for Use (Describe)
The OsteoCentric Cannulated Fasteners and Nuts is indicated for use in bone reconstruction, osteotomy, arthrodesis (joint fusion), and fixation of fractures and ligaments:
The OsteoCentric 2.4mm and 3.0mm Cannulated Screw are intended for fixation of fractures and non-unions of small bones and small bone arthrodesis. Examples include scaphoid and other carpal fractures, metacarpal and phalangeal fusions, osteotomies, and bunionectomies.
The OsteoCentric 3.5mm Cannulated Screws are intended for the arthrodesis or fixation of fractures and fragments of hand, midfoot, and forefoot bones, extremity osteochondritis dissecans lesions, and for extremity ligament fixation.
The OsteoCentric 4.5mm to 8.0mm Cannulated Screws are intended for fixation of long bone fractures and fragments. Iong bone osteotomies, femoral neck fractures, slipped capital femoral epiphyses, as an adjunct to treatment with a dynamic hip screw (DHS) in basilar neck fractures, tibial plateau fractures, pediatric femoral neck fractures, intercondylar femur fractures, fixation of the pelvis and pelvic joints, and arthrodesis (e.g., ankle, subtalar).
The OsteoCentric 7.0mm and 8.0mm Cannulated Nut is indicated for reduction and stabilization of fractures or inter-osseous distances (e.g. syndesmosis injuries).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### K242691 510(k) Summary Prepared on: 2024-09-06 Contact Details 21 CFR 807.92(a)(1) OsteoCentric Technologies Applicant Name Applicant Address 75 West 300 North Suite 150 Logan UT 84321 United States 435-757-2253 Applicant Contact Telephone Applicant Contact Mr. Todd Evans Applicant Contact Email todd.evans@osteocentric.com Device Name 21 CFR 807.92(a)(2) OsteoCentric Technologies Cannulated Fasteners and Nuts Device Trade Name Common Name Screw, Fixation, Bone (primary); Washer, Bolt, Nut Smooth or threaded metallic bone fixation fastener (primary), Classification Name Single/multiple component metallic bone fixation appliances and accessories 21 CFR 888.3040 (primary), 888.3030 Regulation Number Product Code(s) HWC (primary), HTN Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K211290 OsteoCentric Technologies Cannulated Fasteners and Nuts HWC
### Device Description Summary
K151418
Monster Screw System™
The OsteoCentric Technologies Cannulated Fasteners and Nuts come in a variety of lengths and diameters to accommodate different anatomic sizes of patients. The fasteners are provided non- sterile and are manufactured from Stainless Steel per ASTM F138 or from Titanium per ASTM F136 or F1295. Instrumentation is also provided non-sterile to properly insert the OsteoCentric Technologies Cannulated Fasteners and Nuts. All instrumentation is intended to be reusable, with the exception of the guidewires are provided non-sterile and are intended for single-use only.
HWC
21 CFR 807.92(a)(4)
21 CFR 807.92(a)(5)
# Intended Use/Indications for Use
The OsteoCentric Cannulated Fasteners and Nuts is indicated for use in bone reconstruction, osteotomy, arthrodesis (joint fusion), and fixation of fractures and ligaments:
The OsteoCentric 2.4mm and 3.0mm Cannulated Screw are intended for fixation of fractures and mall bone arthrodesis. Examples include scaphoid and other carpal and phalangeal fusions, osteotomies, and bunionectomies.
The OsteoCentric 3.5mm Cannulated Screws are intended for the arthrodesis or fixation of fractures and fragments of hand, midfoot, and forefoot bones, extremity osteochondritis dissecans lesions, and for extremity ligament fixation.
{5}------------------------------------------------
#### K242691
The OsteoCentric 4.5mm to 8.0mm Cannulated Screws are intended for fixation of long bone fractures and fragments, long bone osteotomies, femoral neck fractures, slipped capital femoral epiphyses, as an adjunct to treatment with a dynamic hip screw (DHS) in basilar neck fractures, tibial plateau fractures, pediatic femoral neck fractures, fixation of the pelvis and pelvic joints, and arthrodesis (e.g., ankle, subtalar).
The OsteoCentric 7.0mm and 8.0mm Cannulated for reduction and stabilization of fractures or inter-osseous distances (e.g. syndesmosis injuries).
# Indications for Use Comparison
## 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
The OsteoCentric Technologies Cannulated Fasteners and Nuts provide support and stabilization for orthopedic procedures. It is substantially equivalent to the predicate device, OsteoCentric Technologies Cannulated Fasteners and Nuts (510(k) as well as Reference Device, Paragon 28, Inc. Monster Screw System™ (510(k) Number K151418).
Safety and Effectiveness:
The device adheres to the same safety standards and performance characteristics as the predicate. No new safety or effectiveness concerns have been identified.
Indications for Use:
The Indications for Use has been updated to indicate the use of OsteoCentric Cannulated Fasteners and Nuts as deemed appropriate for the size of the device, based on the following and what is currently cleared: bone reconstruction, osteotomy, arthrodesis, joint fusion, ligament fixation, fracture repair and fracture fixation.
The OsteoCentric Technologies Cannulated Fasteners and Nuts remain safe and effective with the updated indications for use, appropriate for the size of the device.
# Technological Comparison
The subject devices are physically and technologically identical to the previously cleared predicate device (K211290).
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The Purpose of this submission is to update the indicate the use of OsteoCentric Cannulated Fasteners and Nuts as deemed appropriate for the size of the device, based on the following applications and what is currently cleared: bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair and fracture fixation. The Indications for use regarding the Cannulated Nuts have not been updated and is identical to what was previously cleared.
Indications for Use statement is similar to OsteoCentric Technologies Cannulated Fasteners and Nuts cleared under K211290 (Primary Predicate) and Paragon 28, Inc. Monster Screw System™ cleared under K151418 (Reference Device).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.