AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K242683 · Mar 18, 2025
QP-Prostate® CAD
Quibim S.L.
Retrospective multi-center clinical cases; Pathology reports; Clinical follow-up records
Retrospective clinical cases were used to conduct standalone and MRMC performance testing to validate the device's sensitivity, specificity, and AUC-ROC in detecting clinically significant prostate cancer.
Training dataset: cases acquired in the US from multiple centers, multiple magnetic fields (1.5T, 3T) and vendors (Siemens, GE, Philips).
—
Clinical reader performance assessment: 228 cases interpreted by 9 readers in a fully crossed multi-reader multi-case study.
9 (readers)
Indications for Use
QP-Prostate® CAD is a Computed Aided Detection and Diagnosis (CADe/CADx) image processing software that automatically detects and identifies suspected lesions in the prostate gland based on bi-parametric prostate MRI. The software is intended to be used as a concurrent read by physicians with proper training in a clinical setting as an aid for interpreting prostate MRI studies. The results can be displayed in a variety of DICOM outputs, including identified suspected lesions marked as an overlay onto source MR images. The output can be displayed on third-party DICOM workstations and Picture Archive and Communication Systems (PACS). Patient management decisions should not be based solely on the results of QP-Prostate® CAD.
Device Story
AI-based CADe/CADx software; inputs bi-parametric prostate MRI (T2W and DWI sequences). Algorithms detect and identify suspected prostate lesions; provides visual overlays on source images with two suspicion levels (moderate/high). Used in clinical settings by trained physicians (radiologists/urologists) as a concurrent reading aid. Output displayed on standard DICOM workstations/PACS. Assists clinicians in identifying regions suspicious for clinically significant prostate cancer (Gleason score ≥7). Does not replace clinical interpretation; patient management decisions must not rely solely on device output. Benefits include improved diagnostic accuracy for prostate cancer detection.
Clinical Evidence
Standalone performance (N=247 lesions) showed AUC-ROC 0.732. Clinical reader study (N=228 cases, 9 readers) demonstrated statistically significant improvement in diagnostic performance: AUCunaided 0.849 vs. AUCaided 0.868 (ΔAUC 0.019, p=0.039).
Technological Characteristics
Software-based CADe/CADx; DICOM-compatible; inputs T2W and DWI MRI sequences. Uses Neural Networks and Machine Learning algorithms. Deployed as software for third-party DICOM workstations/PACS. No hardware components.
Indications for Use
Indicated for patients above 40 years undergoing prostate MR imaging. Intended for use by physicians (radiologists, urologists) qualified to interpret prostate MRI consistent with ACR recommendations and PI-RADS context.
Regulatory Classification
Identification
A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Special Controls
A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 18, 2025
Quibim S.L. % John J. Smith Partner Hogan Lovells US LLP 555 13th St NW Washington, District of Columbia 20004
Re: K242683
Trade/Device Name: QP-Prostate® CAD Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological Computer Assisted Detection And Diagnosis Software Regulatory Class: Class II Product Code: ODO Dated: February 14, 2025 Received: February 14, 2025
Dear John J. Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.R.K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242683
Device Name QP-Prostate® CAD
#### Indications for Use (Describe)
OP-Prostate® CAD is a Computed Aided Detection and Diagnosis (CADe/CADx) image processing software that automatically detects and identifies suspected lesions in the prostate gland based on bi-parametric prostate MRI. The software is intended to be used as a concurrent read by physicians with proper training in a clinical setting as an aid for interpreting prostate MRI studies. The results can be displayed in a variety of DICOM outputs, including identified suspected lesions marked as an overlay onto source MR images. The output can be displayed on third-party DICOM workstations and Picture Archive and Communication Systems (PACS). Patient management decisions should not be based solely on the results of QP-Prostate® CAD.
Intended patient population: Patients above 40 years with prostate MR imaging.
Intended users:
QP-Prostate® CAD is intended to be used by physicians with proper training, including radiologists, urologists and any physician qualified to read and interpret prostate MRI consistent with ACR recommendations in the context of PI-RADS.
Type of Use (Select one or both, as applicable)
| <div style="width:30px"></div> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------|
| <div style="width:30px"></div> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) SUMMARY Quibim's QP-Prostate® CAD (K242683)
Submitter Quibim S.L. Avenida Aragon 30, 13th floor, Office I-J, 46021 Valencia (Spain)
Phone:
- . +34 961 243 255
Contact Person:
- Ángel Alberich Bayarri, CEO and Founder of Quibim .
- Josep Hortigüela Zamora, VP of Quality Assurance and Regulatory Affairs .
Date Prepared: March 12, 2025
Name of Device: QP-Prostate® CAD
Common or Usual Name: CAD software for lesions suspicious for clinically significant prostate cancer
Classification Name: 892.2090 radiological computer assisted detection/diagnosis software for lesions suspicious for cancer Requlatory Class: Class II
Product Code: QDQ
# Predicate Device
- . Applicant: ScanMed, LLC
- Device's trade name: ProstatIDTM ●
- . 510(k) number: K212783
- Product code: QDQ .
# Device Description
QP-Prostate® CAD is an artificial intelligence-based Computed Aided Detection and Diagnosis (CADe/CADx) image processing software. QP-Prostate® CAD uses Al-based algorithms trained with pathology data to detect suspicious lesions for clinically significant prostate cancer. The device automatically detects and identifies suspected lesions in the prostate gland based on bi-parametric prostate MRI and provides marks over regions of the images that may contain suspected lesions. There are two possible markers that are provided in different colors suggesting different levels of suspicion of clinically significant prostate cancer (moderate or high suspicion level).
The software is intended to be used as a concurrent read by physicians with proper training in a clinical setting as an aid for interpreting prostate MRI studies. The results can be displayed in a variety of DICOM outputs, including identified suspected lesions marked as an overlay onto source MR images. The output can be displayed on third-party DICOM workstations and Picture Archive and Communication Systems (PACS). Based on biparametric input consisting of T2W and DWI series, the analysis is run automatically, and the output in standard DICOM formats is returned to PACS.
{6}------------------------------------------------
# Intended Use / Indications for Use
QP-Prostate® CAD is a Computed Aided Detection and Diagnosis (CADe/CADx) image processing software that automatically detects and identifies suspected lesions in the prostate gland based on bi-parametric prostate MRI. The software is intended to be used as a concurrent read by physicians with proper training in a clinical setting as an aid for interpreting prostate MRI studies. The results can be displayed in a variety of DICOM outputs, including identified suspected lesions marked as an overlay onto source MR images. The output can be displayed on third-party DICOM workstations and Picture Archive and Communication Systems (PACS). Patient management decisions should not be based solely on the results of QP-Prostate® CAD.
### Intended patient population:
Patients above 40 years with prostate MR imaging.
### Intended users:
QP-Prostate® CAD is intended to be used by physicians with proper training, including radiologists, urologists and any physician qualified to read and interpret prostate MRI consistent with ACR recommendations in the context of PI-RADS.
### Comparison to predicate device
QP-Prostate® CAD and the predicate device ProstatID™ have very similar indications statements. Both are medical image processing applications intended for concurrent reading for physicians interpreting prostate MRI identifying suspected lesions and assessing likelihood for prostate cancer. Both devices generate a 3D rendition. ProstatID™ provides the regional overlay scores on a continuous scale ranging from 0-1, presented as a colorized translucent overlay, and a suggested level of suspicion or overall PI-RADS score, whereas QP-Prostate® CAD provides overlay markings around the detected suspicious lesions, in two possible colors suggesting different levels of suspicion of clinically significant prostate cancer (moderate or high).
# Summary of Technological Characteristics
At a high level, the subject and predicate devices are based on the following same technological elements:
- Both devices are compatible with DICOM standard.
- Both devices use the same input sequences to run the analysis: T2-weighted and DWI acquired with specified protocols.
The following technological differences exist between the subject and predicate devices:
- · Both devices are artificial intelligence-based, but they differ in algorithm methodology (e.g. ProstatID™ is based on random forest and QP-Prostate® CAD is based on Neural Networks and Machine Learning). This difference could affect its safety or effectiveness but does not raise any different questions of safety or effectiveness, because a set of verification and
{7}------------------------------------------------
validation tests (performance testing) demonstrate the safety and effectiveness of QP-Prostate® CAD.
A table comparing the key features of the subject and predicate devices is provided below:
| Feature | Proposed device:<br>QP-Prostate® CAD | Predicate device:<br>ProstatID™ (K212783) | Comments |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| REGULATORY DATA | | | |
| Class | II | II | N/A |
| Regulation name | Radiological Computer<br>Assisted<br>Detection/Diagnosis<br>Software | Radiological Computer<br>Assisted<br>Detection/Diagnosis<br>Software | N/A |
| Regulation number | 21 CFR 892.2090 | 21 CFR 892.2090 | N/A |
| Classification Panel | Radiology | Radiology | N/A |
| Product Code | QDQ | QDQ | N/A |
| Applicant | Quibim S.L. | ScanMed, LLC. | N/A |
| | | Indications for Use | |
| Medical device<br>description | QP-Prostate® CAD is a<br>Computed Aided<br>Detection and<br>Diagnosis<br>(CADe/CADx) image<br>processing software<br>that automatically<br>detects and identifies<br>suspected lesions in the<br>prostate gland based<br>on bi-parametric<br>prostate MRI. The<br>software is intended to<br>be used as a<br>concurrent read by<br>clinicians with proper<br>training in a clinical<br>setting as an aid for<br>interpreting prostate<br>MRI studies. The<br>results can be<br>displayed in a variety of<br>DICOM outputs,<br>including identified<br>suspected regions of<br>abnormalities marked<br>as an overlay onto<br>source MR images. The<br>output can be displayed<br>on third-party DICOM<br>workstations and<br>Picture Archive and<br>Communication<br>Systems (PACS).<br>Patient management | ProstatIDTM is a<br>radiological computer<br>assisted detection<br>(CADe) and diagnostic<br>(CADx) software device<br>for use in a healthcare<br>facility or hospital to<br>assist trained<br>radiologists in the<br>detection, assessment,<br>and characterization of<br>prostate abnormalities,<br>including cancer lesions<br>using MR image data<br>with the following<br>indications for use.<br>ProstatID analyzes T2W,<br>DWI and ADC MRI data.<br>ProstatID does not<br>include DCE images in<br>its analysis.<br>ProstatID software is<br>intended for use as a<br>concurrent reading aid<br>for physicians<br>interpreting<br>prostate MRI exams of<br>patients presented for<br>high-risk screening or<br>diagnostic imaging, from<br>compatible MRI<br>systems, to identify<br>regions suspicious for<br>prostate cancer and<br>assess their likelihood of<br>malignancy.<br>Outputs of the device<br>include the volume of the<br>prostate and locations,<br>as well as the extent of<br>suspect lesions, with | Substantially<br>equivalent. Minor<br>differences in outputs,<br>exist, but both devices<br>are CADe/x devices<br>that characterize<br>prostate abnormalities<br>on MR image data.<br>The differences do not<br>raise different<br>questions of safety<br>and effectiveness. |
| Feature | Proposed device: | Predicate device: | Comments |
| | QP-Prostate® CAD | ProstatID™ (K212783) | |
| | decisions should not be<br>based solely on the<br>results of QP-Prostate®<br>CAD. | index scores indicating<br>the likelihood that cancer<br>is present, as well as an<br>exam score by way of<br>PI-RADS interpretation<br>suggestion. "Extent of<br>suspect lesions" refers to<br>both the assessment of<br>the boundary of a<br>particular abnormality,<br>as well as identification<br>of multiple abnormalities.<br>In cases where multiple<br>abnormalities are<br>present, ProstatID can<br>be used to assess each<br>abnormality<br>independently.<br>Outputs of this device<br>should be interpreted<br>with all available MR<br>data consistent with<br>ACR clinical<br>recommendations<br>(e.g., dynamic contract<br>enhanced images if<br>available) in context of<br>PI-RADS v2, and in<br>conjunction with bi-<br>parametric<br>MRI acquired with either<br>surface or endorectal<br>MRI accessory coils<br>from compatible MRI<br>systems. Analysis by<br>ProstatID<br>is not intended as a<br>replacement for<br>interpreting prostate<br>abnormalities using MR<br>image data consistent<br>with clinical<br>recommendations<br>(including DCE); nor<br>should patient<br>management decisions<br>be made solely on the<br>basis of ProstatID. | |
| Intended user<br>population | QP-Prostate® CAD is<br>intended to be used by<br>physicians with proper<br>training, including<br>radiologists, urologists<br>and any physician<br>qualified to read and<br>interpret prostate MRI<br>consistent with ACR<br>recommendations in the<br>context of PI-RADS. | Intended users of<br>ProstatID are physicians<br>qualified to read and<br>interpret prostate MRI<br>exams consistent with<br>ACR recommendations<br>in the context of PI-<br>RADS v2. | Substantially<br>equivalent. |
| Feature | Proposed device:<br>QP-Prostate® CAD | Predicate device:<br>ProstatID™ (K212783) | Comments |
| Intended patient<br>population | Patients above 40<br>years with prostate MR<br>imaging. | The device is intended to<br>be used in the<br>population of biological<br>adult males with a<br>prostate gland<br>undergoing screening or<br>clinical MRI exams. This<br>includes biological males<br>with clinical indicators<br>suggestive of possible<br>prostate cancer or<br>with family history of<br>prostate cancer. | Substantially<br>equivalent. |
| CHARACTERISTICS | | | |
| Clinical finding | Identification of location<br>of suspicious lesions,<br>with two suspicion<br>levels of clinically<br>significant prostate<br>cancer, moderate or<br>high | Extent and location of<br>identified suspected<br>lesions, assigned to a<br>level of suspicion (0-1).<br>A suggested level of<br>suspicion or overall PI-<br>RADS exam score. | Substantially<br>equivalent. While<br>minor differences exist<br>in that QP-Prostate<br>provides two suspicion<br>levels, while ProstatID<br>assigns a level of<br>suspicion on a scale of<br>0-1, the intended use<br>of the devices remain<br>the same.<br>Furthermore, the<br>safety and<br>effectiveness of QP-<br>Prostate are supported<br>through performance<br>assessments and do<br>not raise different<br>questions of safety<br>and effectiveness. |
| DICOM compatibility<br>Input for analysis | DICOM-in DICOM-out<br>T2W and DWI series,<br>DCE is not included for<br>analysis | DICOM-in DICOM-out<br>T2W, DWI and ADC MRI<br>data, DCE is not<br>included for analysis | Same<br>Substantially similar.<br>While QP-Prostate<br>takes a subset of the<br>input data used in the<br>predicate device. This<br>difference does not<br>raise different<br>questions of safety<br>and effectiveness. |
| Algorithm methodology | Neural Networks and<br>Machine Learning | Random Forest | AI-based, safety and<br>effectiveness are<br>supported<br>through performance<br>assessments. |
| Safety | The software is<br>intended to be used as<br>a concurrent read by<br>clinicians with proper<br>training in a clinical<br>setting as an aid for<br>interpreting prostate<br>MRI studies. Patient<br>management decisions<br>should not be based<br>solely on the results of<br>QP-Prostate® CAD. | ProstatID software is<br>intended for use as a<br>concurrent reading aid.<br>Analysis by ProstatID is<br>not intended as a<br>replacement for<br>interpreting prostate<br>abnormalities using MR<br>image data consistent<br>with clinical<br>recommendations<br>(including DCE); nor | Same |
| Feature | Proposed device:<br>QP-Prostate® CAD | Predicate device:<br>ProstatID™ (K212783) | Comments |
| | | should patient<br>management decisions<br>be made solely on the<br>basis of ProstatID. | |
# Table 1: Comparison of key features of QP-Prostate® CAD (Quibim) and the predicate device ProstatID™ (Bot Image, Inc.)
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
# Model development
QP-Prostate® CAD uses Al-based algorithms trained with biopsy outcomes as ground truth to detect suspicious lesions for csPCa (Gleason score ≥7). The training dataset included cases acquired in the US from multiple centers, with multiple magnetic fields (1.5T, 3T) and vendors (Siemens, GE Medical Systems, Philips Medical Systems). T2w and DWI sequences of these cases were used for training. The dataset contained representation of different Gleason score subgroups, was ethnically diverse, and the age ranged from 43 to 86 years. The fine-tuning of the model aimed to optimize the sensitivity and specificity at a case level for the high and moderate markers. To ensure generalizability and performance of QP-Prostate® CAD, standalone and MRMC performance testing studies have been performed with a dataset sourced from different institutions or sources in the US not part of the model training or development datasets. The description of this performance testing dataset and the results of the standalone and MRMC performance testing studies are presented below.
# Performance Data
QP-Prostate® CAD software was developed according to FDA recognized consensus standards for software development. Software verification and validation was performed following V&V plans and protocols verifying that product specifications were met.
# Dataset selection
The pivotal dataset included 228 cases collected retrospectively from multiple centers across the US. This dataset was completely independent from the training dataset that was acquired from different institutions. The pivotal dataset included both positive and negative cases in the same proportion. Positive cases were biopsy confirmed (Gleason score ≥ 7); negative cases were either biopsy-confirmed (Gleason score < 7) or non-biopsied with a clinical followup of at least one year.
Cases were acquired with multiple magnetic fields (1.5 T and 3T) and multiple vendors, and met the device's recommended acquisition parameters for the T2w and DWI sequences. The ethnicity distribution in the dataset is representative of the broader US population.
# Case demographics:
Age range: 43 to 81 Age mean: 64.7
Asian: 3/228 (1.3%) African American: 19/228 (8.3%) White: 194/228 (85.1 %) Other: 4/228 (1.8%) Declined/unavailable: 8/228 (3.5%)
Hispanic: 7/228 (3.1%)
{11}------------------------------------------------
Not Hispanic: 206/228 (90.4%) Declined/unavailable: 6/228 (6.6%)
Magnetic field strength:
1.5T: 32/228 (14.0%) 3T: 196/228 (86.0%)
Scanner vendor:
Philips Medical Systems: 46/228 (20.2 %) GE Medical Systems: 40/228 (17.5%) Siemens: 142/228 (62.3%)
### Standalone performance assessment
For the performance evaluation. QP-Prostate® CAD outputs were compared to ground truth diagnoses derived from associated pathology reports and radiologist interpretations, at lesion-level, based on targeted biopsy locations (both positives and negatives) and negative cases without biopsy in the database (N=247).
Diagnostic accuracy of QP-Prostate® CAD was assessed with the AUC-ROC at lesion level, as well as with the sensitivity and specificity at lesion level at the operating points determined by high suspicion and moderate suspicion markers.
The results are summarized in the following table:
| Metric (lesion level) | Result |
|------------------------------------------------------------------------------------------|--------------------------------|
| AUC-ROC | 0.732<br>(95% CI: 0.668-0.791) |
| Sensitivity<br>(high suspicion marker) | 0.677<br>(95% CI: 0.593-0.761) |
| False Positive Rate per Case<br>(high suspicion marker, any biopsy source) | 0.417<br>(95% CI: 0.313-0.522) |
| Sensitivity<br>(high and moderate suspicion markers) | 0.795<br>(95% CI: 0.722-0.861) |
| False Positive Rate per Case<br>(high and moderate suspicion markers, any biopsy source) | 0.855<br>(95% CI: 0.709-0.996) |
# Table 2: Summary of the standalone performance testing for QP-Prostate® CAD.
Additional subgroup analyses of patient age (≤65, >65), BMI (<25, ≥25), race/ethnicity, magnetic field (1.5T, 3T) and machine vendor (Siemens, Philips, GE) were performed. QP-Prostate® CAD's performance was consistent across all age. BMI, race/ethnicity and machine vendor subgroups. 1.5T subgroup showed a minor deviation in performance, attributed to the inherently lower image quality of 1.5T acquisitions over 3T; 3T acquisitions are preferable when available.
{12}------------------------------------------------
# Clinical reader performance assessment
For the clinical reader performance assessment, a fully crossed multi-reader multi-case study was performed, where each case (N=228) was interpreted by each of the 9 readers with and without QP-Prostate® CAD output in a random-reader random-case factorial design. The clinical reader performance assessment compares the clinical reader performance in identifying clinically significant prostate cancer (Gleason score ≥ 7) when using or not using the device output.
The primary endpoint of the clinical reader performance assessment was the performance advantage of using QP-Prostate® CAD output compared to not using its output for diagnosis of csPCa lesions by clinical readers, determined by the AUCaided versus AUCunaided at case level. The acceptance criterion was a statistically significant improvement.
The secondary endpoint was the performance advantage of using QP-Prostate® CAD output compared to not using its output for diagnosis of csPCa lesions by clinical readers, determined by sensitivity and specificity of the readers with and without QP-Prostate® CAD at case level.
| Metric | Result |
|-----------------------------|--------------------------------------------------|
| AUCunaided | 0.849<br>(95% CI: 0.814-0.884) |
| AUCaided | 0.868<br>(95% CI: 0.834-0.902) |
| ΔAUC (AUCaided- AUCunaided) | 0.019<br>(95% CI: 0.001-0.038)<br>p-value: 0.039 |
# Table 3: Summary of the clinical reader performance testing results for QP-Prostate® CAD.
The test results demonstrate that QP-Prostate® CAD functioned as intended and met its primary endpoint, is acceptable for clinical use, and is as safe and effective as its predicate device, without introducing new questions of safety and efficacy.
# Conclusions
QP-Prostate® CAD is as safe and effective as ProstatID™. QP-Prostate® CAD has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended diagnostic use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the QP-Prostate® CAD and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that QP-Prostate® CAD is as safe and effective as ProstatID™. Thus, QP-Prostate® CAD is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.