← Product Code [FER](/productcode/FER) · K242668

# Gastro Concepts Air Assist (K242668)

_Gastro Concepts, LLC · FER · Dec 13, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K242668

## Device Facts

- **Applicant:** Gastro Concepts, LLC
- **Product Code:** [FER](/productcode/FER.md)
- **Decision Date:** Dec 13, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Air Assist is intended to slide over the outer surface of an endoscope to facilitate endoscopic therapy by providing a manual, reversible method to control the insufflation (air/CO2) so the physician can better maintain an appropriate field of view.

## Device Story

Single-use device; cylindrical body with conical tip and open channel for endoscope insertion. Positioned against anus (not inserted into rectum) to form temporary seal; maintains air/CO2 insufflation during colonoscopy. Used by physicians in clinical settings to improve visualization by impeding pressure loss. Benefits patient by ensuring adequate colonic distension for therapy. No electronic components or software.

## Clinical Evidence

No clinical data. Bench testing only: dimensional testing, simulated use, mechanical properties (compression/tension strength), and leak rate/sealing capability. Testing confirmed device withstands clinically relevant forces and reduces air loss during simulated insufflation. Biocompatibility verified per ISO 10993-1.

## Technological Characteristics

Material: Silicone rubber. Form factor: Cylindrical body with conical tip, open channel for endoscope (11.5-13mm compatibility). Principle: Physical barrier to impede pressure loss. Non-sterile. No energy source. No software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- GelPOINT Path Transanal Access Platform ([K110792](/device/K110792.md))

## Reference Devices

- Embrella™ Endoscopic Distal Attachment ([K190030](/device/K190030.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.

December 13, 2024

Gastro Concepts, LLC % Sarah Moss Regulatory Consultant Prime Path Medtech 1321 Upland Dr. #6792 Houston, Texas 77043

Re: K242668

Trade/Device Name: Gastro Concepts Air Assist Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FER, FED Dated: November 13, 2024 Received: November 13, 2024

Dear Sarah Moss:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## Sivakami Venkatachalam -S

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

Submission Number (if known)

K242668

Device Name

Gastro Concepts Air Assist

#### Indications for Use (Describe)

Air Assist is intended to slide over the outer surface of an endoscope to facilitate endoscopic therapy by providing a manual, reversible method to control the insufflation (air/CO2) so the physician can better maintain an appropriate field of view.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

| Submitter:                                  | Gastro Concepts                                                                               |
|---------------------------------------------|-----------------------------------------------------------------------------------------------|
| Company Contact Person:<br>Phone:<br>Email: | Rick Brown, President<br>(787) 642-2532<br>rick.brown@gastroconcepts.com                      |
| Submission<br>Correspondent:                | Sarah Moss, Regulatory Affairs Consultant                                                     |
| Address:<br>Phone:<br>Email:                | 1321 Upland Dr. Suite 6792, Houston, TX 77043<br>(281) 795-1812<br>smoss@primepathmedtech.com |
| Date Prepared:                              | Sept 2024                                                                                     |
| Proprietary Name:                           | Gastro Concepts Air Assist                                                                    |
| Common Name:                                | Anoscope and accessories                                                                      |
| Product Code/CFR:                           | FER, FED; 876.1500 Endoscope and accessories                                                  |
| Device Classification:                      | Class II                                                                                      |
| Predicate Device:                           | GelPOINT Path Transanal Access Platform (K110792)                                             |

#### Device Description:

The Gastro Concepts Air Assist Device is a single-use device that is used to aid in maintaining air/CO2 and insufflation during a colonoscopy. The Air Assist, a single-use device, comprises a cylindrical body with a conical-shaped tip and an open channel for endoscope insertion. Positioned against the anus (not inserted into the rectum), it forms a temporary seal, maintaining air/CO2 insufflation leading to adequate colonic distension for enhanced visualization.

#### Indications for Use:

Air Assist is intended to slide over the outer surface of an endoscope to facilitate endoscopic therapy by providing a manual, reversible method to control the insufflation (air/CO2) so the physician can better maintain an appropriate field of view.

{5}------------------------------------------------

|                                      | Gastro Concepts<br>Air Assist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | GelPOINT Path Transanal<br>Access Platform<br>(Predicate Device)                                                                                                                                                                                                                              | Embrella™ Endoscopic<br>Distal Attachment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Comparison Result                                               |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Specification                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                 |
| 510(k) Number                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K110792                                                                                                                                                                                                                                                                                       | K190030                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar                                                         |
| Device<br>Description                | The Gastro Concepts Air<br>Assist Device is a single-use<br>device that is used to aid in<br>maintaining air/CO2 and<br>insufflation during a<br>colonoscopy. The Air<br>Assist, a single-use device,<br>comprises a cylindrical<br>body with a conical-shaped<br>tip and an open channel<br>for endoscope insertion.<br>Positioned against the anus<br>(not inserted into the<br>rectum), it forms a<br>temporary seal,<br>maintaining air/CO2<br>insufflation leading to<br>adequate colonic<br>distension for enhanced<br>visualization. | The GeIPOINT Path is a<br>sterile, single use,<br>disposable surgical<br>instrument designed to<br>provide multiple<br>instrument or camera<br>access into the rectal<br>cavity and lower<br>sigmoid colon. The<br>device is used in concert<br>with standard<br>laparoscopic<br>instruments. | The proposed device<br>Embrella™ Endoscopic<br>Distal Attachment is an<br>additional<br>device, made of medical<br>grade silicone rubber<br>designed to attach to the<br>distal end of the<br>endoscope. Its shape is<br>umbrella shaped, with six<br>flexible wings that<br>easily to fold backwards<br>and forwards. It can be<br>used to facilitate<br>endoscopic<br>therapy and maintain an<br>appropriate endoscopic<br>field of view. Embrella™<br>Endoscopic Distal<br>Attachment has two<br>models, one is EM-S-01,<br>which is compatible with<br>endoscopy distal diameter<br>12.8-14.5mm; the other is<br>EM-S-02, which is<br>compatible with endoscopy<br>distal diameter 11.5-<br>13.0mm. |                                                                 |
| Indication for<br>Use                | Air Assist is intended<br>to slide over the outer<br>surface of an<br>endoscope to facilitate<br>endoscopic therapy by<br>providing a manual,<br>reversible method to<br>control the insufflation<br>(air/CO2) so the<br>physician can better<br>maintain an<br>appropriate field of<br>view.                                                                                                                                                                                                                                               | The GeIPOINT Path is a<br>sterile, single use,<br>disposable surgical<br>instrument designed to<br>provide multiple<br>instrument or camera<br>access into the rectal<br>cavity and lower<br>sigmoid colon.                                                                                   | Embrella™ Endoscopic<br>Distal Attachment is<br>intended to be attached<br>to the distal end of the<br>endoscope to facilitate<br>endoscopic therapy and<br>maintain an appropriate<br>endoscopic field of view.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Similar/In-Scope for<br>Product Code                            |
| Product<br>Code                      | FER                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | FER                                                                                                                                                                                                                                                                                           | FED                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as predicate                                               |
|                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K242668<br>Page 3 of 4                                          |
| Specification                        | Gastro Concepts<br>Air Assist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | GelPOINT Path Transanal<br>Access Platform<br>(Predicate Device)                                                                                                                                                                                                                              | Embrella™ Endoscopic<br>Distal Attachment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Comparison Result                                               |
| Regulations                          | 21CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 21CFR 876.1500                                                                                                                                                                                                                                                                                | 21CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent                                                      |
| Class                                | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 2                                                                                                                                                                                                                                                                                             | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Equivalent                                                      |
| Single Use                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Equivalent                                                      |
| Scope<br>Compatibility<br>(diameter) | 11.5-13mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Not published                                                                                                                                                                                                                                                                                 | 11.5-14.5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Similar                                                         |
| Device<br>Functionality              | Provides improved<br>field of view by<br>maintaining proper<br>insufflation                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Provides improved<br>field of view by<br>maintaining proper<br>insufflation.                                                                                                                                                                                                                  | Provides improved field of<br>view by prohibiting the<br>surrounding tissues from<br>block the distal end of the<br>scope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Equivalent                                                      |
| Mechanism<br>of Action               | The device uses a<br>physical barrier to<br>impede pressure loss.<br>This allows the colon<br>to be insufflated, thus<br>enlarging the<br>operative space.                                                                                                                                                                                                                                                                                                                                                                                  | The device uses a<br>physical barrier to<br>impede pressure loss.<br>This allows the colon<br>to be insufflated, thus<br>enlarging the<br>operative space.                                                                                                                                    | Flexible wings that<br>easily fold backwards and<br>forwards. It can be used to<br>facilitate endoscopic therapy<br>and maintain an appropriate<br>endoscopic field of view.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Equivalent                                                      |
| Device<br>Placement                  | Device is placed<br>immediately adjacent<br>to the anus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Device is inserted into<br>the anus                                                                                                                                                                                                                                                           | Device is attached to the<br>distal end of the scope and<br>inserted into the body cavity.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Not equivalent<br>(Air assist is less invasive)                 |
| Supplied<br>Sterile/Non-<br>Sterile  | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Sterile                                                                                                                                                                                                                                                                                       | Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Not equivalent<br>(Air assist is not inserted<br>into the body) |
| Materials                            | Silicone Rubber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Gel material                                                                                                                                                                                                                                                                                  | Silicone Rubber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar to predicate<br>Equivalent to reference                 |
| Biocompatibil<br>ity: ISO<br>10993-1 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Equivalent                                                      |
| Shelf Life                           | 6 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | One Year                                                                                                                                                                                                                                                                                      | One Year                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Similar                                                         |

{6}------------------------------------------------

K242668

{7}------------------------------------------------

#### Performance Testing (Bench)

The following bench tests were performed on the Gastro Concepts Air Assist Device:

- Dimensional Testing
- Simulated Use
- Mechanical Properties: Structural integrity, compression and tension strength
- Leak Rate and Sealing Capability

Testing was conducted to establish that the Gastro Concepts Air Assist device can withstand clinically relevant mechanical forces. Leak testing was also completed to demonstrate a reduction in air loss during insufflation. The bench testing proved that the Air Assist device was able to withstand forces greater than expected during clinical use and that the device was successful in reducing the leak rate during simulated colonic insufflation.

The Air Assist device meets the requirements of ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within A Risk Management Process."

Therefore, the performance of the Air Assist device was shown to be at least equivalent to the predicate device and did not raise any new safety or effectiveness questions.

#### Clinical Testing

No Clinical Testing was required for this product.

#### Statement of Equivalence

Based on comparison of indications for user population, performance testing, mechanical and technological features, the Air Assist device has been shown to be substantially equivalent to the legally marketed predicate device does not raise any new safety or effectiveness questions as compared to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K242668](https://fda.innolitics.com/device/K242668)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
