← Product Code PHN · K242634

# PuraStat (K242634)

_3-D Matrix Europe SAS · PHN · Jan 15, 2025 · GU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K242634

## Device Facts

- **Applicant:** 3-D Matrix Europe SAS
- **Product Code:** PHN
- **Decision Date:** Jan 15, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** GU
- **Attributes:** Therapeutic

## Indications for Use

Purastat is indicated for the symptomatic management of rectal mucositis, such as radiation proctitis that may be caused by chemotherapy or radiotherapy.

## Device Story

PuraStat is a sterile, synthetic peptide-based hydrogel (2.5% peptide content) designed to provide a protective barrier over rectal mucosa. The device consists of a synthetic 16-amino acid (RADA-16) oligopeptide in sterile water. Upon contact with physiological fluids (blood or tissue), the peptide solution self-assembles into a transparent, non-expanding scaffold structure. It is supplied in prefilled cyclo-olefin polymer syringes and delivered endoscopically to the rectal mucosa via a commercially available catheter and polypropylene adapter. The device is intended for use by clinicians to manage symptoms of rectal mucositis. By forming a physical protective coating, it aims to alleviate symptoms associated with radiation proctitis or chemotherapy-induced mucosal damage.

## Clinical Evidence

Bench testing and biocompatibility testing only; no clinical data provided. Equivalence established through comparison of technological characteristics and performance testing following the sterilization process change.

## Technological Characteristics

Synthetic 16-amino acid (RADA-16) oligopeptide in sterile water. Delivered via cyclo-olefin polymer syringe with high-density polyethylene plunger and butyl rubber components. Self-assembling scaffold mechanism triggered by physiological pH. Terminally sterilized via gamma radiation.

## Regulatory Identification

Mucoadhesive application for the protective coating of the rectal mucosa.

## Predicate Devices

- PuraStat-RM ([K213552](/device/K213552.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 15, 2025

3-D Matrix Europe SAS Audrey Vion RAQA Manager 11 Chemin des Petites Brosses Caluire et Cuire, 69300 France

Re: K242634 Trade/Device Name: PuraStat Regulatory Class: Unclassified Product Code: PHN Dated: December 18, 2024 Received: December 18, 2024

Dear Audrey Vion:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming

product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Anthony Lee -S

Anthony Lee, Ph.D., M.B.A. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K242634

Device Name

PuraStat

Indications for Use (Describe)

Purastat is indicated for the symptomatic management of rectal mucositis, such as radiation proctitis that may be caused by chemotherapy or radiotherapy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Page 1 of 3

In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary is provided below.

#### SUBMITTER 1.

3-D Matrix Europe SAS 11 chemin des Petites Brosses 69300 Caluire et Cuire - FRANCE

Contact Person: Audrey VION Phone: +33 (0) 627 635 514 Email: avion@puramatrix.com

Date Prepared: September 2, 2024

#### 2. DEVICE

Name of Device: PuraStat Common Name: Mucoadhesive Application for the Protective Coating of the Rectal Mucosa Classification Regulation: Unclassified Regulatory Class: Unclassified Product Code: PHN Panel: Gastroenterology / Urology

#### PREDICATE DEVICE 3.

Predicate Device: 3-D Matrix, Inc.'s PuraStat-RM (K213552).

#### DEVICE DESCRIPTION 4.

PuraStat is a sterile gel composed of a synthetic peptide and sterile water for injection. It is provided as a prefilled syringe (2.5% peptide content) ready for use as a mucoadhesive hydrogel that provides a protective barrier over rectal mucosa. The gel is delivered to the intended application site(s) via a commercially available endoscopic catheter that is attached to the gel-filled syringe via the polypropylene adapter (Class 1 - 510k Exempt) also commercially available (Class 1. 510k Exempt).

PuraStat is completely non-animal and non-plant derived and contains no preservatives that might present a risk of allergic reaction or skin irritation.

Exposure to physiological fluids such as blood causes the peptide solution to quickly form a transparent gel without expansion in volume.

{5}------------------------------------------------

#### 5. INDICATIONS FOR USE

PuraStat is intended to be used for the symptomatic management of rectal mucositis, such as radiation proctitis that may be caused by chemotherapy or radiotherapy.

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6.

The subject device is identical to the PuraStat-RM predicate (K213552) except for the final sterilization, which is replaced by gamma radiation sterilization. The final product specifications are unchanged.

PuraStat comprises of a synthetic repeating 16-amino acid (RADA-16) oligopeptide in sterile water for injection. In the presence of physiological pH (blood or tissue), the peptides assemble into a scaffold structure that results in the formation of a protective coating over the rectal mucosa.. The solution is filtered and filled into 5-ml syringes made of cyclo-olefin polymers with a highdensity polyethylene plunger and a butyl rubber head cap and gasket. Each syringe is filled with either 1, 3, or 5 ml of gel. The device is terminally sterilized, and the resorbable gel is delivered to the intended application site(s) via a commercially available adapter connected to a commercially available catheter. The device is delivered endoscopically via a catheter to the patient.

### 7. PERFORMANCE DATA

The subject device is identical to the predicate device. The intended use of the subject device is identical to the intended use of the predicate device K213552. The indications for use of the subject device are identical to the predicate's indications for use.

The subject device is identical to the predicate device and has the same intended use, and an equivalence demonstration was performed following the final sterilization change. The bench testing and biocompatibility testing provided in this submission assess the equivalence demonstration.

Overall, the difference in technological characteristics (Final sterilization) of the subject and the predicate device 3-D Matrix Inc. - PuraStat-RM do not raise any new questions of safety and effectiveness.

{6}------------------------------------------------

#### 8. CONCLUSIONS

The subject PuraStat has identical intended use and indications as the predicate PuraStat-RM (K213552).

Technological characteristics between the predicate PuraStat-RM (K213552) and the subject device are identical, with the exception of the final sterilization. This difference do not raise any questions of safety or effectiveness. PuraStat is identical in material, formulation and manufacturing, and so final product specification to the predicate PuraStat-RM (K213552)

In conclusion, PuraStat is substantially equivalent to the predicate (K213552).

---

**Source:** [https://fda.innolitics.com/device/K242634](https://fda.innolitics.com/device/K242634)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
