K242627 · Ningbo Medsun Medical Co., Ltd. · FMK · Oct 30, 2024 · General, Plastic Surgery
Device Facts
Record ID
K242627
Device Name
Safety Lancet
Applicant
Ningbo Medsun Medical Co., Ltd.
Product Code
FMK · General, Plastic Surgery
Decision Date
Oct 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4850
Device Class
Class 2
Indications for Use
Safety Lancet is intended to be used to obtain capillary blood sample to perform medical testing, including blood glucose monitoring and for tests using small amounts of blood.
Device Story
Safety Lancet (Models XA Pro and ZF) is a single-use, sterile blood sampling device featuring an integrated sharps injury prevention mechanism. The device is used in clinical or home settings by patients or healthcare providers to obtain capillary blood samples. Operation involves pressing a button or slider, which releases a spring-loaded tri-bevel needle or blade to puncture the skin. Upon completion, the needle automatically retracts into the housing, preventing reuse and accidental needle sticks. The device is color-coded by specification. It provides a safe, disposable method for blood collection, reducing the risk of sharps injuries during and after the sampling process.
Clinical Evidence
Bench testing only. Performance testing included dimension, trigger force, corrosion resistance, retraction, penetration depth, and sharps injury prevention (per ISO 23908:2011). Biocompatibility testing (ISO 10993) included in vitro cytotoxicity, skin sensitization, intracutaneous reactivity, acute systemic toxicity, and pyrogen testing. All tests passed.
Technological Characteristics
Single-use, spring-loaded lancet with integrated sharps protection. Materials include plastic housing, metal needle/blade, and springs. Sterilized via Co-60 or e-beam irradiation (SAL 10^-6). Complies with ISO 10993 (biocompatibility) and ISO 9626:2016 (needle gauge). Features automatic inactivation system to prevent reuse.
Indications for Use
Indicated for patients requiring capillary blood sampling for medical testing, including blood glucose monitoring and other tests requiring small blood volumes.
Regulatory Classification
Identification
The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.
Special Controls
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
*Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
*Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
*Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class III (premarket approval).(3)
*Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 30, 2024
Ningbo Medsun Medical Co., Ltd. Liu Ping Regulation Affairs Manager No. 298 Huangjipu Road, Jiangbei Ningbo, Zhejiang 315031 China
Re: K242627
Trade/Device Name: Safety Lancet Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: August 28, 2024 Received: September 3, 2024
Dear Liu Ping:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
Digitally signed by Long H. Chen Long H. Chen-S s Date: 2024.10.30 10:36:21 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242627
Device Name
Safety Lancet
Indications for Use (Describe)
Safety Lancet is intended to be used to obtain capillary blood sample to perform medical testing, including blood glucose monitoring and for tests using small amounts of blood.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
### 1.Date Prepared
Auqust 28st, 2024
#### 2.Submitter
Ningbo Medsun Medical Co., Ltd.
No. 298 Huangjipu Road, Jiangbei, 315031,Ningbo, P.R.China
Contact Person: Liu Ping, Regulation Affairs Manager
Tel: +86-574-86301887/ Fax:+86-574-86301887
Date Prepared: August 28st, 2024
#### 3.Device
Trade Name: Safety Lancet
Common Name:Blood Lancets
Classification Name:Single Use Only Blood Lancet With An Integral Sharps Injury
Prevention Feature
Regulation Number:21 CFR 878.4850
Requlatory Class: II
Product Code: FMK
Review Panel: General & Plastic Surgery
#### 4.Predicate Device
Manufacturer: Ningbo Medsun Medical Co., Ltd.
Device Name: Safety Lancet
510(k) Number: K222090
#### 5.Device Description
Safety Lancet could be divided into Model XA Pro and Model ZF.Each model has sharps protection features that it can reduce the occurrence of accidental needle sticks. The device is sterilized by Co-60 or e-beam gamma ray and for single use.The shelf life is 5 years. Model XA Pro has four versions according to the different three penetration depths, version XA1 Pro,version XA3 Pro,version XA5 Pro and version XA11 Pro. Model ZF also
{6}------------------------------------------------
has four versions according to the different need type and material of slider, version ZF1,version ZF2,version ZF1 Blade and version ZF2 Blade.
Model XA Pro is composed of a protective cap, a knob (for depth settings), a front spring, housing, needle body with tri-bevel edge needle, a rear spring and a back cover (push button). The needle body with needle or blade is hidden inside the housing/protective cap. The device is colour coded for different versions.There is an integrated automatic inactivation system to prevent lancet reuse. When the back cover is pressed down, it drives the rear spring to compress downwards. The back cover support legs are flush with the inclined surface of the protruding line head of the housing. The needle is rotated and slided downwards under the force of the rear spring until it is disengaged from the inclined surface and fired forward for blood sampling, while compressing the front spring. The needle will be retracted automatically back into the housing after the puncture due to the force from the pressed front spring. The needle cannot go back to the starting position and stays between the relaxed rear spring and front spring. Therefore, the device cannot be re-used.
Model ZF lancet is composed of spring, three-facet needle or blade, housing, and slider. The needle body with needle or blade is concealed within housing/slider. The lancets of different specifications are in different colors. There is an integrated automatic inactivation system to prevent lancet reuse. The spring is fixed at the bottom of the housing, and when the slider is pressed down, which drives the spring to compress downwards.The two locking points(blocked the needle) on the slider will be opened along the slope of the housing, the needle will be released to fire under the force of the spring for blood sampling. After the puncture is completed, the spring retracts to a relaxed state and synchronously pulls the needle body back into the housing. Therefore, the device cannot be re-used.
#### 6.Indications for Use
Safety Lancet is intended to be used to obtain capillary blood sample to perform medical testing, including blood glucose monitoring and for tests using small amounts of blood.
| Description | Subject Device<br>(Model XA Pro) | Subject Device<br>(Model ZF) | Predicate Device<br>(K222090) | Remark |
|-----------------|----------------------------------|------------------------------|-------------------------------|--------|
| Product<br>Code | FMK | FMK | FMK | Same |
#### 7.Comparison of Technological Characteristics With the Predicate Device
{7}------------------------------------------------
| Regulation<br>Number | 21CFR 878.4850 | 21CFR 878.4850 | 21CFR 878.4850 | Same |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Indications<br>for use | Safety Lancet is<br>intended to be used to<br>obtain capillary blood<br>sample to perform<br>medical testing,<br>including blood glucose<br>monitoring and for tests<br>using small amounts of<br>blood. | Safety Lancet is<br>intended to be used to<br>obtain capillary blood<br>sample to perform<br>medical testing,<br>including blood glucose<br>monitoring and for tests<br>using small amounts of<br>blood. | Safety Lancet is<br>intended to be used to<br>obtain capillary blood<br>sample to perform<br>medical testing,<br>including blood glucose<br>monitoring and for tests<br>using small amounts of<br>blood. | Same |
| Prescription/<br>over-the<br>counter use | Over-the<br>counter use | Over-the<br>counter use | Over-the<br>counter use | Same |
| Design | There is an integrated<br>automatic inactivation<br>system to prevent lancet<br>reuse. The spring is fixed<br>at the bottom of the<br>housing, and when the<br>slider is pressed down,<br>which drives the spring<br>to compress<br>downwards.The two<br>locking points(blocked<br>the needle) on the slider<br>will be opened along the<br>slope of the housing, the<br>needle will be released<br>to fire under the force of<br>the spring for blood<br>sampling. After the<br>puncture is completed,<br>the spring retracts to a<br>relaxed state and<br>synchronously pulls the<br>needle body back into<br>the housing. Therefore,<br>the device cannot be<br>re-used. | There is an integrated<br>automatic inactivation<br>system to prevent lancet<br>reuse. When the back<br>cover is pressed down,<br>it drives the rear spring<br>to compress<br>downwards. The back<br>cover support legs are<br>flush with the inclined<br>surface of the protruding<br>line head of the housing.<br>The needle is rotated<br>and slided downwards<br>under the force of the<br>rear spring until it is<br>disengaged from the<br>inclined surface and<br>fired forward for blood<br>sampling, while<br>compressing the front<br>spring. The needle will<br>be retracted<br>automatically back into<br>the housing after the<br>puncture due to the<br>force from the pressed<br>front spring. The needle<br>cannot go back to the<br>starting position and<br>stays between the | There is an integrated<br>automatic inactivation<br>system to prevent<br>lancet reuse. When the<br>slider is pressed, the<br>slider pushes the two<br>bosses of the needle<br>holder backwards. The<br>bosses move<br>backwards over the<br>short rib on the inner<br>side of the housing, and<br>the rear spring is<br>pressed and activated<br>to launch the needle<br>forwards for blood<br>sampling. The needle<br>retracts automatically<br>back into the housing<br>after the puncture, due<br>to the force from the<br>pressed front spring.<br>The needle cannot go<br>back to the starting<br>position and stays<br>between the relaxed<br>rear spring and front<br>spring. Therefore, the<br>device cannot be<br>re-used. | Similar<br>(Note 1) |
{8}------------------------------------------------
| 510(k) Submission Documents-Safety Lancet - K242627 | | | | |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| | | relaxed rear spring and<br>front spring. Therefore,<br>the device cannot be<br>re-used. | | |
| Materials<br>and<br>composition | Safety lancet is<br>composed of a<br>protective cap, a knob<br>(for depth settings), a<br>front spring, housing,<br>needle body with<br>tri-bevel edge needle, a<br>rear spring and a back<br>cover (push button). | Safety lancet lancet is<br>composed of spring,<br>three-bevel edge needle<br>or blade, housing, and<br>slider. | The safety lancet are<br>composed needle/<br>blade,silicone oil,spring<br>and plastic part(back<br>cover,slider,housing<br>and protective cap). | Similar<br>(Note 2) |
| Version | XA1 Pro, XA3 Pro,<br>XA5 Pro, XA11 Pro | ZF1,ZF2,<br>ZF1 Blade, ZF2 Blade | XY1,XY1 Blade,<br>XH1,XH1 Blade,<br>XA1 | Different<br>(Note 3) |
| Performance | Dimension,Appearance,<br>Needle-tip,Drop<br>performance,Trigger<br>force, Corrosion<br>resistance feature,<br>Retractable,Penetration<br>Depth,Challenge Safe<br>Mode-Resistance,<br>Challenge Safe Mode-<br>Needle Tip Exposed,<br>Anti-activation test,<br>Self-destruct<br>performance test,Fall<br>performance test,<br>pH and total heavy metal | Dimension,Appearance,<br>Needle-tip,Drop<br>performance, Trigger<br>force, Corrosion<br>resistance feature,<br>Retractable,Penetration<br>Depth,Challenge Safe<br>Mode-Resistance,<br>Challenge Safe Mode-<br>Needle Tip Exposed,<br>Anti-activation test,<br>Self-destruct<br>performance test,Fall<br>performance test,<br>pH and total heavy<br>metal | Dimension, Appearance<br>,Needle-tip,Drop<br>performance,Trigger<br>force, Corrosion<br>resistance feature,<br>Retractable,Penetration<br>Depth,Challenge Safe<br>Mode- Resistance,<br>Challenge Safe Mode-<br>Needle Tip Exposed,<br>Anti-activation test,<br>Self-destruct<br>performance test,Fall<br>performance test,<br>pH and total heavy<br>metal | Same |
| Sterilization | Irradiation Sterilization<br>SAL:10-6 | Irradiation Sterilization<br>SAL:10-6 | Irradiation sterilized,<br>SAL: 10-6 | Same |
| Labeling | Conform with 21 CFR<br>801 | Conform with 21 CFR<br>801 | Conform with 21 CFR<br>801 | Same |
| | 510(k) Submission Documents-Safety Lancet - K242627 | | | |
| Biocom-<br>patibility | Meet the requirements of<br>ISO10993 series<br>standards,<br>bocompatibility has been<br>evaluated in K222090. | Meet the requirements<br>of ISO10993 series<br>standards, and the<br>following tests are<br>performed:In vitro<br>cytotoxicity, skin<br>sensitization,intracutane<br>ous reactivity, acute<br>systemic toxicity, and<br>pyrogen. | Meet the requirements<br>of ISO10993 series<br>standards, and the<br>following tests are<br>performed:In vitro<br>cytotoxicity, skin<br>sensitization,intracutan<br>eous reactivity, acute<br>systemic toxicity, and<br>pyrogen. | Same |
| Reuse or<br>single use | Single use | Single use | Single use | Same |
| Dimension/<br>gauge | 18G,19G,20G,21G,<br>21G,22G,23G,24G,<br>25G,26G,27G,28G,<br>29G,30G,31G,32G,<br>33G(Needle type) | 21G,22G,23G,24G,25G,<br>26G,27G,28G,29G, 30G<br>(Needle type)<br>0.8mm,0.9mm,1.1mm,<br>1.2mm,1.5mm(Blade<br>type) | Model XY:<br>21G,23G,25G,26G,28<br>G,30G(Needle type),<br>1.2mm(Blade type)<br>Model XH:<br>18G,21G,23G,28G<br>(Needle type), 1.5mm<br>(Blade type)<br>Model XA:<br>21G,23G,28G(Needle<br>type) | Different<br>(Note 4) |
| Penetration<br>depth | 1.4mm/1.8mm/2.2mm,<br>1.6mm/2.0mm/2.4mm,<br>1.3mm/1.8mm/2.3mm,<br>1.5mm/2.0mm/2.5mm | 1.2mm-2.8mm<br>(Needle type)<br>1.8 mm-2.0mm<br>(Blade type) | Model XY:<br>1.4mm-2.8mm(Needle<br>type), 1.6mm-2.0mm<br>(Blade type)<br>Model XH:<br>1.2mm-2.8mm(Needle<br>type),1.6mm-2.0mm(Bl<br>ade type)<br>Model XA:<br>1.3mm/1.8mm/2.3mm,<br>1.5mm/2.0mm/2.5mm(<br>Needle type) | Different<br>(Note 5) |
{9}------------------------------------------------
Note 1:The design of the subject device is similar as the predicate device. Although the designs of the Model XA Pro, Model ZF and the predicate device have some differences, all models are used the elasticity of spring to launch a needle puncture for blood sampling. The performance of the subject device has been tested in the bench testing.Therefore,the difference will not affect the safety and effectiveness of the product.
{10}------------------------------------------------
Note 2:The materials and composition of the subject device are similar as the predicate device. For Model XA Pro, the materials and composition is same as the Model XA,except that the three indicator bars(indicated the penetration depth). The three indicator bars is on the knob of Model XA Pro(subject device), the three indicator bars is on the housing of Model XA(predicate device). The materials and composition of the predicate device has one more spring than Model ZF, the performance of the Model ZF has been tested in the bench testing. Therefore,the difference will not affect the safety and effectiveness of the product.
Note 3:The version of the subject device is different as the predicate device. The version is named as the different penetration depth, different needle type/blade type and material of slider. The version do no affect the safety and effectiveness of the product.
Note 4:Model XA Pro only has needle type, Model ZF has needle type and blade type. The subject device has needle type and blade type. The dimension/gauge of Model XA Pro is 18G-33G, the predicate device is not including 19G/20G/22G/24G/27G/29G/31G/32G/33G. The dimension/gauge of Model ZF is 21G-30G(Needle type) and 0.8mm,0.9mm,1.1mm,
1.2mm,1.5mm(Blade type), the predicate device is not including 22G/24G/27G/29G(Needle type) and 0.8mm,0.9mm,1.1mm(Blade type).The needle gauge of subject device is a widely used in the market and meet the requirements ISO 9626:2016, the different blade width of subject device is smaller than the predicate device. The dimension/gauge of the subject device also has been tested in the bench testing.Therefore,the difference will not affect the safety and effectiveness of the product.
Note 5: Model XA Pro has four versions, which have three penetration depths for depth setting ,the range of penetration depth is 1.3mm-2.5mm. The penetration depth range of Model XA(subject device) is 1.3mm-2.5mm. Therefore, their extreme depth is the same. The penetration depth of Model ZF is 1.2mm-2.8mm(Needle type) and 1.8 mm-2.0mm (Blade type), the penetration depth range of the predicate device is 1.2mm-2.8mm(Needle type) and 1.6mm-2.0mm(Blade type). The range of penetration depth for the blade type subject device is getting shorter than the predicate device.It means the the length of puncture entering the skin is not longer, which will not increase patient's pain and there is no standard for regulating the penetration depth.Therefore,the difference will not affect the safety and effectiveness of the product.
6/8
{11}------------------------------------------------
# 8.Performance Testing Summary
# Performance Test
All non-clinical bench testing performed on the subject device is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with corporate standard, which is formulated according to relevant standards and product characteristics. Results of the testing are demonstrating the compliance to the standards and matching the performance of the subject device to the predicate devices.
The following data of non-clinical performance bench testing are provided in support of the substantial equivalence determination.
| Items | Requirements | Results |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimension | Dimension should meet the requirements. | Passed |
| Appearance | The needle holder should be smooth and<br>free of defects such as burrs, oil stains, rust<br>spots, and bending. The housing should be<br>smooth, free of burrs, bubbles, impurities<br>and cracks. With normal or corrected to<br>normal vision, there should be no visible<br>accumulation of lubricant on the outer<br>surface of the needle tip. | Passed |
| Needle-tip | The tip of the lancet should have no flat<br>head, burrs and hooks.Penetration Force<br>should meet the requirements. | Passed |
| Trigger force | The lancet should be activated against<br>2N~20N axial pressure. | Passed |
| Corrosion<br>resistance feature | Needle should have good corrosion<br>resistance feature. | Passed |
| Retractable | After use, the needle can be automatically<br>retracted. | Passed |
| Penetration Depth | Penetration depth should meet the<br>requirements. | Passed |
| Challenge Safe Mode-<br>Resistance | Apply a 15N force horizontally at the raised<br>platform of the lancet, and the back cover<br>must not be detached. | Passed |
| Challenge Safe Mode-<br>Needle Tip Exposed | After the needle tip of the lancet is<br>retracted, the needle tip cannot be<br>exposed. Observe the needle tip after<br>removing the protective cap, the needle tip<br>should not be exposed. | Passed |
| Anti-activation test | The lancet should not activate when it is<br>mistakenly pushed the button before use. | Passed |
{12}------------------------------------------------
| Self-destruct<br>performance test | The lancet should not be activated again<br>after use. | Passed |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Fall Performance | The product falls freely to a smooth hard<br>surface at a height of 1 meter, than test the<br>appearance,triggering force,retraction and<br>penetration depth should meet the<br>requirements. | Passed |
| pH and total heavy<br>metal content and Cd content | pH(difference of blank solution)≤1,Total<br>heavy metal content≤5mg/L and Cd<br>content ≤0.1mg/L | Passed |
| Sterility | After the safety lancet has been sterilized<br>by radiation, the needle tip should be<br>sterile. | Passed |
#### Biocompatibility Test
The subject device meets the requirements of ISO 10993 series standards, and the following items are evaluated:In vitro cytotoxicity, skin sensitization, intracutaneous reactivity, acute systemic toxicity and pyrogen. The results show that the product has good biocompatibility.
# Performance of Clinical Simulated Use Testing for Sharps Injury Protection
According to ISO 23908:2011 Sharps injury protection-Requirements and test methods-Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling and Guidance for Industry and FDA Staff-Medical Devices with Sharps Injury Prevention Features, we have completed the test of clinical simulated use testing for sharps injury protection.
Select the most widely sold in the market of each model as typical product for study.The results show that the product has well sharps injury prevention feature.
# 9.Conclusions
Based on device comparison information and non-clinical bench testing, the subject device is substantially equivalent to the predicate devices (K222090).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.