Immunoassay for the in vitro quantitative determination of cortisol in human urine. The determination of cortisol is used for the recognition and treatment of functional disorders of the adrenal gland. The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers.
Device Story
Elecsys Cortisol III is an in vitro diagnostic immunoassay for quantitative cortisol measurement in human urine; used on cobas e 801 analyzers in clinical laboratory settings. Principle of operation: competition test; sample incubated with biotinylated monoclonal anti-cortisol antibody; subsequent addition of streptavidin-coated microparticles and ruthenium-labeled cortisol derivative. Microparticles magnetically captured on electrode; voltage application induces chemiluminescent emission measured by photomultiplier. Results determined via instrument-specific calibration curve. Output provides quantitative cortisol concentration to assist clinicians in diagnosing and managing adrenal gland functional disorders.
Clinical Evidence
Bench testing only. Performance evaluated via precision (CLSI EP05-A3), linearity (CLSI EP06-Ed2), and interference/cross-reactivity (CLSI EP07). Method comparison against predicate (N=125) showed Passing Bablok slope 1.051, intercept -0.810, r=0.933. Reference range established using 144 healthy subjects (2.5th-97.5th percentile: 24.8–238 nmol/24h). No clinical prospective/retrospective studies performed.
Technological Characteristics
ECLIA-based immunoassay. Reagents: streptavidin-coated microparticles, biotinylated monoclonal anti-cortisol antibody, ruthenium-labeled cortisol derivative. Buffer: MES (pH 6.0). Detection: chemiluminescence via photomultiplier. Instrument: cobas e 801 analyzer. Traceable to NIST SRM 921a. Software: instrument-specific calibration and master curve via cobas link.
Indications for Use
Indicated for the in vitro quantitative determination of cortisol in human urine to assist in the recognition and treatment of functional disorders of the adrenal gland. No specific patient population age or gender restrictions are provided.
Regulatory Classification
Identification
A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.
K152227 — Elecsys Cortisol II, Cortisol II CalSet · Roche Diagnostics · Apr 27, 2016
K021218 — ELECSYS CORTISOL TEST SYSTEM · Roche Diagnostics Corp. · Sep 9, 2002
K984520 — NICHOLS ADVANTAGE CHEMILUMINESCENCE CORTISOL IMMUNOASSAY, MODEL 62-7005 · Nichols Institute Diagnostics · Feb 5, 1999
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 17, 2025
Roche Diagnostics
Chunhong (Emma) Tao
Regulatory Affairs Manager
9115 Hague Road
Indianapolis, Indiana 46250
Re: K242505
Trade/Device Name: Elecsys Cortisol III
Regulation Number: 21 CFR 862.1205
Regulation Name: Cortisol (Hydrocortisone And Hydroxycorticosterone) Test System
Regulatory Class: Class II
Product Code: JFT
Dated: July 17, 2025
Received: July 17, 2025
Dear Chunhong (Emma) Tao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242505 - Chunhong (Emma) Tao
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K242505 - Chunhong (Emma) Tao
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Paula V. Caposino -S
Paula Caposino, Ph.D.
Deputy Director
Division of Chemistry and
Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242505 | |
| Device Name Elecsys Cortisol III | |
| Indications for Use (Describe) Immunoassay for the in vitro quantitative determination of cortisol in human urine. The determination of cortisol is used for the recognition and treatment of functional disorders of the adrenal gland. The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” | |
{4}
510k Summary
Elecsys Cortisol III (K242505)
# 510k Summary
# Elecsys Cortisol III (K242505)
This summary of $510(\mathrm{k})$ safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
In accordance with 21 CFR 807.87, Roche Diagnostics hereby submits official notification as required by Section $510(\mathrm{k})$ of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Traditional $510(\mathrm{k})$ Premarket Notification.
The purpose of this Traditional 510(k) Premarket Notification is to obtain FDA review and clearance for Elecsys Cortisol III.
| Submitter Name: | Roche Diagnostics |
| --- | --- |
| Address: | 9115 Hague Road Indianapolis, IN 46250-0457 |
| Contact: | Emma Tao Phone: (317) 260-1427 Email: emma.tao@roche.com |
| Date Prepared: | July 15, 2025 |
| Proprietary Name: | Elecsys Cortisol III |
| Common Name: | Fluorometric, Cortisol |
| Classification Name: | Cortisol (hydrocortisone and hydroxycorticosterone) test system |
| Product Code: | JFT |
| Regulation Number: | 21 CFR 862. 1205 |
{5}
510k Summary
Elecsys Cortisol III (K242505)
| Predicate Device: | ARCHITECT Cortisol (K062204) |
| --- | --- |
| Establishment | Roche Diagnostics GmbH Mannheim, Germany: 9610126 |
| Registration | Roche Diagnostics GmbH Penzberg, Germany: 9610529 |
| | Roche Diagnostics Indianapolis, IN United States: 1823260 |
Page 2
{6}
510k Summary
Elecsys Cortisol III (K242505)
# 1. DEVICE DESCRIPTION
The Elecsys Cortisol III immunoassay employs a competitive test principle using a cortisol-specific biotinylated monoclonal antibody and a cortisol-derivative labeled with a ruthenium complex. The Elecsys Cortisol III immunoassay is intended for the *in vitro* quantitative determination of cortisol in human urine. The determination of cortisol is used for the recognition and treatment of functional disorders of the adrenal gland on the **cobas e** immunoassay analyzers.
Results are determined via a calibration curve which is instrument-specifically generated by a two-point calibration and a master curve provided via the **cobas** link.
The Elecsys Cortisol III immunoassay reagent Rack Pack comprises the following:
M Streptavidin-coated microparticles (transparent cap), 1 bottle, 12.4 mL:
Streptavidin-coated microparticles 0.72 mg/mL; preservative.
R1 Anti-cortisol-Ab~biotin (gray cap), 1 bottle, 21.0 mL:
Biotinylated monoclonal anti-cortisol antibody (mouse) 18 ng/mL; danazol 20 µg/mL; MES buffer 100 mmol/L, pH 6.0; preservative.
R2 Cortisol-peptide~Ru(bpy) (black cap), 1 bottle, 21.0 mL:
Cortisol derivative (synthetic), labeled with ruthenium complex, 5 ng/mL; danazol 20 µg/mL; MES buffer 100 mmol/L, pH 6.0; preservative.
MES = 2-morpholino-ethane sulfonic acid
# 2. INDICATIONS FOR USE
Immunoassay for the *in vitro* quantitative determination of cortisol in human urine. The determination of cortisol is used for the recognition and treatment of functional disorders of the adrenal gland.
{7}
510k Summary
Elecsys Cortisol III (K242505)
The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.
# 3. TECHNOLOGICAL CHARACTERISTICS
The following table compares Elecsys Cortisol III with its predicate device, ARCHITECT Cortisol (K062204).
Table 1: Technical Characteristics Comparison between Elecsys Cortisol III and ARCHITECT Cortisol (K062204)
| Item | Elecsys Cortisol III
(Candidate Device) | ARCHITECT Cortisol
(K062204, Predicate Device) |
| --- | --- | --- |
| Intended Use | Immunoassay for the in vitro quantitative determination of cortisol in human urine.
The determination of cortisol is used for the recognition and treatment of functional disorders of the adrenal gland. | The ARCHITECT Cortisol is a chemiluminescent microparticle immunoassay (CIMA) for the quantitative determination of cortisol in human serum, plasma or urine on the ARCHITECT / System. The ARCHITECT Cortisol assay is intended for use as an aid in the diagnosis and treatment of adrenal disorders. |
| Indications for Use | The results obtained are used for the recognition and treatment of functional disorders of the adrenal gland. | The results obtained are used as an aid in the diagnosis and treatment of adrenal disorders. |
| Detection Method | electrochemiluminescence immunoassay
"ECLIA" | chemiluminescent microparticle immunoassay (CMIA) |
| Instrument Platform | cobas e immunoassay analyzers | ARCHITECT / System |
| Sample Type | Urine | Serum, plasma or urine |
| Expected Range | 8.98 – 86.2 μg/24 h (24.8 – 238 nmol/24 h) | 4.3 – 176 μg/24 h |
| Reportable Range | 0.725 – 18.1 μg/dL (20.0 – 500 nmol/L) | 0.8 – 59.8 μg/dL |
# 4. NON-CLINICAL PERFORMANCE EVALUATION
The following performance data are provided in support of the substantial equivalence determination. All performance specifications were met.
{8}
510k Summary
Elecsys Cortisol III (K242505)
## 4.1. Precision
### 4.1.1. Repeatability and Intermediate Precision
Precision was evaluated with human urine samples (24-hour urine) and controls by testing two replicates per run, two runs per day for 21 days. Repeatability and Intermediate precision were calculated according to CLSI EP05-A3. All data met the predefined acceptance criteria.
| cobas e 801 analyzer | | | | | |
| --- | --- | --- | --- | --- | --- |
| Sample | Mean nmol/L | Repeatability | | Intermediate Precision | |
| | | SD nmol/L | CV % | SD nmol/L | CV % |
| Human Urine 1 | 88.7 | 2.41 | 2.7 | 3.34 | 3.8 |
| Human Urine 2 | 236 | 5.46 | 2.3 | 6.90 | 2.9 |
| Human Urine 3 | 296 | 5.86 | 2.0 | 9.37 | 3.2 |
| Human Urine 4 | 453 | 9.39 | 2.1 | 14.5 | 3.2 |
| Control 1 | 204 | 4.04 | 2.0 | 5.00 | 2.5 |
| Control 2 | 374 | 8.98 | 2.4 | 11.3 | 3.0 |
| cobas e 801 analyzer | | | | | |
| --- | --- | --- | --- | --- | --- |
| Sample | Mean μg/dL | Repeatability | | Intermediate Precision | |
| | | SD μg/dL | CV % | SD μg/dL | CV % |
| Human Urine 1 | 3.22 | 0.0873 | 2.7 | 0.121 | 3.8 |
| Human Urine 2 | 8.56 | 0.198 | 2.3 | 0.250 | 2.9 |
| Human Urine 3 | 10.7 | 0.212 | 2.0 | 0.340 | 3.2 |
| Human Urine 4 | 16.4 | 0.340 | 2.1 | 0.525 | 3.2 |
| Control 1 | 7.39 | 0.147 | 2.0 | 0.181 | 2.5 |
| Control 2 | 13.6 | 0.325 | 2.4 | 0.408 | 3.0 |
### 4.1.2. Reproducibility
Lot-to-lot reproducibility was performed using three reagent lots according to CLSI EP05-A3. All data met the predefined acceptance criteria.
Page 5
{9}
510k Summary
Elecsys Cortisol III (K242505)
## 4.2. Analytical Sensitivity
### 4.2.1. Limit of Blank (LoB)
The Limit of Blank (LoB) was determined according to CLSI EP17-A2. The LoB claim in the labeling will be set to 4.00 nmol/L (0.145 µg/dL).
### 4.2.2. Limit of Detection (LoD)
The Limit of Detection (LoD) was determined according to CLSI EP17-A2. The LoD claim in the labeling will be set to 7.50 nmol/L (0.272 µg/dL).
### 4.2.3. Limit of Quantitation (LoQ)
The Limit of Quantitation (LoQ) was determined according to CLSI EP17-A2. The LoQ claim in the labeling will be set to 10.0 nmol/L (0.363 µg/dL).
## 4.3. Linearity/Assay Reportable Range
Linearity was evaluated on the cobas e 801 analyzer according to CLSI EP06-Ed2. The dilution series contained a minimum of 9 concentrations throughout the measuring range. A measuring range of 20.0-500 nmol/L or 0.725-18.1 µg/dL will be claimed in the labeling.
## 4.4. Human Anti-Mouse Antibodies (HAMA)
The effect of the presence of human anti-mouse antibodies (HAMA) was assessed on the cobas e 801 analyzer. The differentiation between HAMA-negative and HAMA-positive samples was assessed. Data met the predefined acceptance criteria.
## 4.5. Endogenous Interferences
The effect on quantitation of Cortisol in the presence of thirteen endogenous interfering substances (lipemia, biotin, bilirubin, hemoglobin, rheumatic factor, human serum albumin, human IgG, human IgM, human IgA, Sodium Chloride, Urea, glucose, and creatinine) was tested using human urine samples (24-hour urine). No significant interference for the assay was observed up to the concentrations of the potential interfering substances tested as shown in the table below.
{10}
510k Summary
Elecsys Cortisol III (K242505)
| Compound | Concentration tested |
| --- | --- |
| Bilirubin | ≤ 1130 μmol/L or ≤ 66 mg/dL |
| Hemoglobin | ≤ 0.621 mmol/L or ≤ 1000 mg/dL |
| Intralipid | ≤ 2000 mg/dL |
| Biotin* | ≤ 204658 nmol/L or ≤ 50000 ng/mL |
| Rheumatoid Factor | ≤ 1200 IU/mL |
| Human IgG | ≤ 7.0 g/dL |
| Human IgA | ≤ 1.6 g/dL |
| Human IgM | ≤ 1.0 g/dL |
| Human serum albumin | ≤ 4.9 g/dL |
| Creatinine | ≤ 5 mmol/L |
| Glucose | ≤ 5 mmol/L |
| NaCl | ≤ 750 mmol/L |
| Urea | ≤ 350 mmol/L |
* Biotin was tested at 100,000 ng/mL. The data supports no interference up to 50,000 ng/mL.
# 4.6. Analytical Specificity/Cross-Reactivity
A cross-reactivity study was conducted on the cobas e 801 analyzer to evaluate the potential cross-reacting compounds using human urine (24-hour urine). Samples were measured in the presence and absence of the potential cross-reactants and cross-reactivity was calculated with one lot of reagents. The following cross-reactivities (in %) were found at the respective cross-reactant concentration, tested with a cortisol concentration of approximately $17\mathrm{nmol/L}$ (0.6 $\mu\mathrm{g/dL}$ ).
| Cross-Reactant | Concentration tested μg/dL | Cross reactivity % |
| --- | --- | --- |
| 11-Deoxycorticosterone | 100 | 0.174 |
| 11-Deoxycortisol | 50 | 24.3 |
| 17α-Hydroxyprogesterone | 1000 | 0.412 |
| 21-Deoxycortisol | 100 | 2.33 |
| Corticosterone | 750 | 0.368 |
| Cortisone | 500 | 1.49 |
| Androstenedione | 100 | n. d.d) |
| DHEAS | 1000 | n. d. |
| DHEA | 1000 | n. d. |
| Progesterone | 1000 | 0.00930 |
| Testosterone | 1000 | n. d. |
{11}
510k Summary
Elecsys Cortisol III (K242505)
| Cross-Reactant | Concentration tested μg/dL | Cross reactivity % |
| --- | --- | --- |
| Estradiol | 1000 | n. d. |
| Estriol | 1000 | n. d. |
| Estrone | 1000 | n. d. |
| Aldosterone | 1000 | n. d. |
| Pregnenolone | 1000 | n. d. |
| 17α-Hydroxypregnenolone | 1000 | 0.0417 |
| 11β-Hydroxyprogesterone | 1000 | 0.0173 |
| Pregnanetriol | 1000 | n. d. |
| 6α-Hydroxycortisol | 100 | n. d. |
| 6β-Hydroxycortisol | 100 | 0.0698 |
| Cortisol-21 glucuronide | 1000 | 0.0301 |
| Allotetrahydrocortisol | 10 | 11.3 |
| Cortisol-21-sulfate | 1000 | n. d. |
| β-Cortol | 1000 | n. d. |
| β-Cortolone | 1000 | n. d. |
| Pregnanediol | 1000 | n. d. |
| Tetrahydrocortisol | 10 | n. d. |
## 4.7. Exogenous Interferences – Drugs
In vitro tests were performed on 12 commonly used pharmaceuticals. No interference with the assay was found. At concentrations corresponding to the daily therapeutic dose, the special drugs prednisolone and hydrocortisone caused elevated concentrations of cortisol. For the special drug 6 methylprednisolone, no interference was observed for concentrations ≤ 0.157 mg/dL.
In addition, the following special drugs were tested. No interference with the assay was found.
| Drug | Concentration tested mg/dL |
| --- | --- |
| Amlodipine | 0.008 |
| Betamethasone | 0.0345 |
| Beclomethasone | 0.000631 |
| Budenoside | 0.00063 |
| Canrenone | 0.075 |
| Dexamethasone | 1.20 |
| Fludrocortisone | 0.120 |
| Fluticasone propionate | 0.0003 |
Page 8
{12}
510k Summary
Elecsys Cortisol III (K242505)
| Drug | Concentration tested mg/dL |
| --- | --- |
| HCT (hydrochlorothiazide) | 4.77 |
| Lisinopril | 27 |
| Losartan potassium | 0.092 |
| Metformin | 850 |
| Metoprolol | 0.150 |
| Mometasone | 0.000045 |
| Prednisone | 0.010 |
| Spironolactone | 0.0555 |
| Triamterene | 0.059 |
| Valsartan | 1.17 |
| Verapamil | 0.160 |
| Triamcinolone | 0.003 |
| Atorvastatin | 0.075 |
| Danazol | 0.030 |
| Diclofenac | 2.40 |
| β-Sitosterol | 1.00 |
## 4.8. Method Comparison
Method comparison was performed between the candidate device and the predicate device using native 24 h urine samples. Samples span the measuring range. Data was analyzed according to CLSI EP09-A3 and met all predefined acceptance criteria.
## 4.9. Stability
The stability studies were performed and the predefined acceptance criteria were met. The stability data supports the claims as reported in labeling.
| Stability: | |
| --- | --- |
| unopened at 2-8 °C | up to the stated expiration date |
| on the analyzer | 16 weeks |
## 5. REFERENCE RANGE STUDY
A study was performed to determine the expected values of the Elecsys Cortisol III assay in 24 h urine samples from an apparently healthy population in the United States. Samples were
Page 9
{13}
510k Summary
Elecsys Cortisol III (K242505)
collected across three study sites according to the inclusion and exclusion criteria. An additional study site did sample testing. The reference range was computed and reported as the 2.5th and 97.5th percentiles for pooled data from all study sites and with all sites combined according to CLSI EP28-A3c.
| 2.5th percentile | 97.5th percentile | Unite |
| --- | --- | --- |
| 24.8 | 238 | nmol/24h |
| 8.98 | 86.2 | μg/24h |
## 6. ADDITIONAL INFORMATION
Elecsys Cortisol III is intended to be used with the following calibrators and controls:
- CalSet Cortisol III Urine
- PreciControl Cortisol Urine
## 7. CONCLUSIONS
The information provided in this 510(k) Premarket Notification supports the determination that Elecsys Cortisol III is equivalent to the predicate device, ARCHITECT Cortisol (K062204).
Page 10
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.