The LMA® Fastrach™ ETT Single Use is indicated for tracheal intubation through the LMA® Fastrach™. The Stabiliser Rod is indicated for use during removal of the LMA® Fastrach™ and LMA® Fastrach™ Single Use after intubation to keep the LMA® Fastrach™ ETT Single Use in place.
Device Story
LMA Fastrach ETT SU is a straight, cuffed, wire-reinforced endotracheal tube designed for use with the LMA Fastrach airway. Features include a Murphy Eye, 15mm connector, pilot balloon with luer check valve, and a soft, molded tip for atraumatic vocal cord passage. Depth markers assist with positioning; radiopaque materials enhance X-ray visibility. Used by clinicians for airway management during intubation. The Stabiliser Rod accessory facilitates tube retention during LMA removal. Device is sterile, single-use, and MR Conditional. Benefits include facilitated oxygenation and ventilation in patients requiring intubation via the LMA Fastrach system.
Clinical Evidence
Bench testing only. Evaluations included visual inspection, dimensional testing, bonding strength, kink resistance, cuff herniation, resistance to collapse, and tracheal seal testing. Biocompatibility testing conducted per ISO 10993-1 and ISO 18562-1. All tests passed.
Indicated for patients requiring tracheal intubation through the LMA® Fastrach™ to facilitate oxygenation and ventilation. Contraindicated for patients undergoing procedures involving laser beams or electrosurgical active electrodes in the immediate device area.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 29, 2025
Teleflex Medical
Irma Govea
Regulatory Affairs Specialist
3015 Carrington Mills Blvd
Morrisville, North Carolina 27560
Re: K242495
Trade/Device Name: LMA Fastrach ETT SU
Regulation Number: 21 CFR 868.5730
Regulation Name: Tracheal Tube
Regulatory Class: Class II
Product Code: BTR
Dated: April 2, 2025
Received: April 2, 2025
Dear Irma Govea:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242495 - Irma Govea
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K242495 - Irma Govea
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K242495
Device Name
LMA Fastrach ETT SU
Indications for Use (Describe)
The LMA® Fastrach™ ETT Single Use is indicated for tracheal intubation through the LMA® Fastrach™.
The Stabiliser Rod is indicated for use during removal of the LMA® Fastrach™ and LMA® Fastrach™ Single Use after intubation to keep the LMA® Fastrach™ ETT Single Use in place.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) Submission
LMA Fastrach ETT SU
# 510(k) Summary
A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Inc.
3015 Carrington Mill Blvd
Morrisville, NC 27560 USA
Phone: 1 919-361-3973
B. Contact Person
Irma Govea
Regulatory Affairs Specialist
C. Date Prepared
March 31, 2025
D. Device Name
Subject Device 1
Trade Name: LMA Fastrach™ ETT SU
Common Name: Endotracheal Tube
Product Code: BTR
Regulation Number: 21 CFR 868.5730
Classification Name: Tube, Tracheal (W/Wo Connector)
Classification: II
Classification Panel: Anesthesiology
E. Predicate Device
K051993 LMA Fastrach ETT SU
F. Device Description
The LMA® Fastrach™ ETT Single Use has been developed specifically for use with the LMA® Fastrach™. It is a straight, cuffed, wire-reinforced tube with a Murphy Eye and a standard 15mm connector.
The LMA® Fastrach™ ETT Single Use has a pilot balloon with a luer check valve and a unique, soft, moulded tip for atraumatic passage through the vocal cords. As a reference during intubation, the LMA® Fastrach™ ETT Single Use has depth markers to indicate the distance to the distal tip of the LMA™ airway. The LMA® Fastrach™ ETT Single Use is radiopaque along its full length and its tip is made out of a radio opaque material to enhance its visibility in x-rays.
Accessory device: The Stabiliser Rod Single Use is an accessory and it is indicated for use during removal of the reusable LMA® Fastrach™ and LMA® Fastrach™ Single Use after intubation to keep the Endotracheal Tube (ETT) in place.
G. Indications for Use
See Substantial Equivalence Table Below
510(k) Summary - Page 1 of 4
{5}
510(k) Submission
LMA Fastrach ETT SU
# H. Contraindications
The LMA® Fastrach™ ETT Single Use should not be placed in patients eligible for procedures which involve the use of a laser beam or electrosurgical active electrode in the immediate area of the device.
There are no known contraindications associated with the Stabiliser Rod accessory device.
# I. Substantial Equivalence
The subject device is substantially equivalent to the predicate device with respect to intended use, technology, and design. The differences between the predicate and the subject devices are minor and any risks have been mitigated through testing. The table below summarizes the differences between the subject and predicate device.
Table 1. Substantial Equivalence Comparative Table LMA Fastrach ETT SU
| Features | Teleflex Medical LMA Fastrach ETT SU | Teleflex Medical LMA Fastrach ETT SU (Predicate K051993) | Comments |
| --- | --- | --- | --- |
| Classification Name | Tube, Tracheal (w/wo connector) | Same | Same |
| Product Code | BTR | Same | Same |
| Regulation Number | 868.5730 | Same | Same |
| Indications for Use | The LMA® Fastrach™ ETT Single Use is indicated for tracheal intubation through the LMA® Fastrach™. The Stabiliser Rod is indicated for use during removal of the LMA® Fastrach™ and LMA® Fastrach™ Single Use after intubation to keep the LMA® Fastrach™ ETT Single Use in place. | The LMA Fastrach™ ETT SU is indicated for airway management by oral intubation on the trachea. Reinforced ETT may be used to reduce the potential for kinking whenever an unusual positioning of the head or neck is required following intubation. | Indication is not changing; a rewording was made for better readability |
| Patient Population | LMA® Fastrach™ ETT Single Use is intended for patients who require intubation through the Fastrach LMA to facilitate oxygenation and ventilation. Clinical judgement should be used in the selection of the appropriate device size for an individual patient. | Not stated | Although predicate device does not clearly state the patient population in the IFU, it remains the same for both products |
| Contraindications | The LMA® Fastrach™ ETT Single Use should not be placed | LMA Fastrach ETT is contraindicated in procedures | Contraindications are not |
510(k) Summary - Page 2 of 4
{6}
510(k) Submission
LMA Fastrach ETT SU
| | in patients eligible for procedures which involve the use of a laser beam or electrosurgical active electrode in the immediate area of the device. There are no known contraindications associated with the LMA ® Fastrach™ Stabiliser Rod accessory device. | which involve the use of a laser beam or electrosurgical active electrode in the immediate area of the device. | changing; a rewording was made for better readability A statement has been added for the stabilizer rod, even though there are no known contraindications |
| --- | --- | --- | --- |
| Single Use | Yes | Same | Yes |
| Size Range | 6.0 mm – 8.0 mm | Same | Same |
| Cuffed | Yes | Same | Same |
| Radiopaque | Yes | Same | Same |
| Connection to ventilation source | 15 mm connector Polyamide | 15 mm connector Polypropylene (PP) | Materials have been tested per ISO 10993-1 |
| Method of Sterilization | Ethylene Oxide 10-6SAL | Same | Same |
| Biocompatibility | Materials have been tested per ISO 10993-1 | Materials biocompatible | Materials have been tested per ISO 10993-1 |
| Components & Materials | Main tube – Non-DEHP Polyvinylchloride (PVC) reinforced with Stainless Steel Cuff – PVC Boat tip – PVC Inflation line – PVC Ink – Black Ink + Reducer Stabilizer rod - PVC | Main tube – Polyvinylchloride (PVC) reinforced with Stainless Steel Cuff -PVC Boat tip – PVC Inflation line – PVC Ink – Black Ink Stabilizer rod - PVC | Materials were changed to non-DEHP |
| Sterile | Yes | Same | Same |
| Eye | Murphy | Same | Same |
| Graduations | Multiple cm markings | Same | Same |
| MR Safety | MR Conditional | Not declared | Testing has been conducted |
| Shelf-life | 1 year | 2 years | Testing has been conducted |
510(k) Summary - Page 3 of 4
{7}
510(k) Submission
LMA Fastrach ETT SU
| Packaging | Preformed pouch | Blister tray | Packaging performance testing was conducted |
| --- | --- | --- | --- |
| | Packaging film – PET/EP 12/60
Tyvex – Uncoated Tyvex Paper 1073B | Packaging film – Perfecseal 30652-EE (8 mil) nylon film
Tyvex – PerfecSeal 2FS Tyvex heat seal coated with SBP 2000 | |
J. Comparison to the Predicate
The table above illustrates the similarities and differences between the subject and predicate device. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are for airway management. Both the subject and predicate devices are disposable, sterile, single patient use devices. The subject devices are substantially equivalent to the predicate device. There are no significant differences that would affect safety and efficacy.
K. Performance Data
The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:
- Testing performed included visual inspection, dimensional testing, bonding strength (main tube with boat tip, side arm, 15mm connector and cuff bonding), kink resistance, cuff herniation, resistance to collapse, tracheal seal testing.
The testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device.
The LMA Fastrach ETT SU passed all evaluations and tests. The device is biocompatible and meets the requires pertaining to biocompatibility per 10993-1, ISO 18562-1, the US FDA Guidance on Application of 10993-1.
L. Conclusion
The subject device has the same intended use and technological characteristics and construction as the predicate. Test results demonstrate that the subject device meets it's intended use. It is for these reasons that the subject device can be found substantially equivalent to the predicate devices.
510(k) Summary - Page 4 of 4
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.