K242487 · Laminar Digital Health, Inc. · JAF · Dec 13, 2024 · Radiology
Device Facts
Record ID
K242487
Device Name
Laminar P1 (LDH-HW-001)
Applicant
Laminar Digital Health, Inc.
Product Code
JAF · Radiology
Decision Date
Dec 13, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1540
Device Class
Class 2
Indications for Use
The Laminar P1 device is a wearable and disposable, ultrasound-based (4.5 MHz, continuous-wave) patch sensor designed to detect blood flow in peripheral vessels and assist in the detection of peripheral vascular disease. The Laminar P1 device is intended to be used by medical professionals, such as physicians and nurses, in hospitals, clinics and private offices and healthcare facilities. It is to be used on adults, ages 18 years and older.
Device Story
Laminar P1 is a wearable, ultrasound-based sensor patch for peripheral blood flow detection. System comprises a single-use sensor patch, adhesive patch frame, and reusable battery-operated processing unit. Device utilizes 4.5 MHz piezoelectric transducers in continuous-wave Doppler mode to detect blood flow in veins and arteries. Medical professionals (physicians/nurses) apply the patch to the patient; the processing unit controls the transducer and provides audio output via a built-in speaker. Users can adjust volume levels and select between three receiver options (shallow, medium, deep) to optimize signal quality. Output is an audible Doppler signal, similar to traditional handheld Doppler devices, used by clinicians to assess peripheral vascular status and assist in diagnosing peripheral vascular disease. Benefits include a lightweight, body-worn form factor facilitating monitoring over time compared to traditional handheld probes.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), software (IEC 62304), electrical/EMC safety (IEC 60601-1/1-2), and acoustic output (IEC 60601-2-37, IEC 62359).
Technological Characteristics
Wearable ultrasound patch; 4.5 MHz continuous-wave piezoelectric transducer. Components: single-use sensor patch, adhesive frame, reusable battery-operated processing unit (3.7V Li-polymer). Dimensions: 52x38x22 mm; weight: 44g. Features adjustable volume and three receiver depth settings. Standards: ISO 10993-1 (biocompatibility), IEC 62304 (software), IEC 60601-1/1-2 (electrical/EMC), IEC 60601-2-37/62359 (acoustic).
Indications for Use
Indicated for adults 18+ in clinical settings to detect peripheral blood flow and assist in peripheral vascular disease diagnosis.
Regulatory Classification
Identification
A nonfetal ultrasonic monitor is a device that projects a continuous high-frequency sound wave into body tissue other than a fetus to determine frequency changes (doppler shift) in the reflected wave and is intended for use in the investigation of nonfetal blood flow and other nonfetal body tissues in motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 13, 2024
Laminar Digital Health, Inc. % Pierre Bounaud Principal Consultant Ram+ 2790 Mosside Blvd, Suite 800 Monroeville, Pennsylvania 15146
Re: K242487
Trade/Device Name: Laminar P1 Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW Dated: November 12, 2024 Received: November 12, 2024
#### Dear Pierre Bounaud:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K242487
Device Name
Laminar P1
Indications for Use (Describe)
The Laminar P1 device is a wearable and disposable, ultrasound-based (4.5 MHz, continuous-wave) patch sensor designed to detect blood flow in peripheral vessels and assist in the detection of peripheral vascular disease. The Laminar P1 device is intended to be used by medical professionals, such as physicians and nurses, in hospitals, clinics and private offices and healthcare facilities. It is to be used on adults, ages 18 years and older.
Type of Use (Select one or both, as applicable)
< | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY
#### DATE PREPARED
December 11, 2024
#### MANUFACTURER AND 510(k) OWNER
Laminar Digital Health, Inc. 440 N Wolfe Road, Sunnyvale, CA 94085, USA Telephone: Official Contact: Jess Lee, CEO
#### REPRESENTATIVE/CONSULTANT
Pierre Bounaud, Ph.D. Allison Komiyama, Ph.D., RAC ROM+ Telephone: +1 (412) 816-8139 Email: pbounaud@rqmplus.com.akomiyama@rgmplus.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | Laminar P1 |
|------------------------------|-------------------------------------------------------------------------------------------|
| Common Name: | Ultrasound Sensor Patch |
| Regulation Number: | 21 CFR 870.2100 |
| Class: | II |
| Product Code: | DPW |
| Premarket Review: | OHT2/Division of Cardiac Electrophysiology, Diagnostics<br>and Monitoring Devices (DHT2A) |
| Review Panel: | Cardiovascular |
#### PREDICATE DEVICE IDENTIFICATION
The Laminar P1 is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate |
|---------------|--------------------------------------|----------------------|
| K191388 | FloPatch (FP110) / Flosonics Medical | ✓ |
The predicate device has not been subject to a design related recall.
#### DEVICE DESCRIPTION
The Laminar P1 is an ultrasound-based battery-powered sensor patch designed to detect blood flow in veins and arteries to assist in the detection of peripheral vascular disease. The Laminar P1 consists of a single use sensor patch, a single use patch frame, and a reusable processing unit. The sensor patch is designed to have the same performance characteristics as traditional ultrasound transducer probes (4.5 MHz frequency) but in a patch form factor that is easy to place on the patient's body for a given time period. The patch frame is a sticky frame that aids in the positioning and alignment of the processing unit. The processing unit controls the piezoelectric transducers in the sensor patch and has a built-in speaker with volume-control, like the
{5}------------------------------------------------
traditional handheld Doppler devices, and operates in the Doppler continuous-wave mode (4.5 MHz frequency). The Processing Unit is battery-operated.
#### INDICATIONS FOR USE
The Laminar P1 device is a wearable and disposable, ultrasound-based (4.5 MHz, continuouswave) patch sensor designed to detect blood flow in peripheral vessels and assist in the detection of peripheral vascular disease.
The Laminar P1 device is intended to be used by medical professionals, such as physicians and nurses, in hospitals, clinics and private offices and healthcare facilities. It is to be used on patients, ages 18 years and older.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Laminar Digital Health, Inc. believes that the Laminar P1 is substantially equivalent to the predicate device based on the information summarized here:
Both devices have the same intended use, i.e., to detect blood flow in peripheral vessels for assisting in the detection of peripheral vascular disease. Both devices are intended for the same patient population and the same use environments.
The subject device shares similar technological characteristics with the predicate device. Both devices use the Doppler effect to detect blood flow in peripheral vasculature via piezoelectric transducers and one continuous wave mode.
Main differences in technological characteristics include:
- Smaller form factor The subject device is smaller and lighter than the predicate. This . difference has undergone testing to ensure the subject device is as safe and effective as the predicate.
- . Transducer frequency - The subject device operates at 4.5 MHz while the predicate device features a flat face probe at 4 MHz. This does not raise different questions of safety or effectiveness.
- Global maximum outputs – Global maximum outputs for the subject device are higher than the predicate. This difference has undergone testing to ensure the subject device is as safe and effective as the predicate.
- . Volume level and receiver options – The subject device offers a choice of volume levels and three ultrasound receiver options (shallow, medium, and deep) to boost signal quality. The user can cycle between the three options to pick the best audio sound quality and amplitude. This does not raise different questions of safety or effectiveness.
These differences have undergone testing to ensure the subject device is as safe and effective as the predicate device.
{6}------------------------------------------------
| | Subject Device | Predicate Device | |
|-----------------------------------------|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------|
| | | | |
| | Laminar P1 | FloPatch (FP110) | |
| | | | |
| | | K191388 | |
| Indications for Use | The Laminar P1 device is a wearable and | The FloPatch(FP110) is intended for the | |
| | disposable, ultrasound-based (4.5 MHz, | detection of blood flow in peripheral | |
| | continuous-wave) patch sensor designed | vasculature. | |
| | to detect blood flow in peripheral vessels | The device is intended to be used by | |
| | and assist in the detection of peripheral | medical professionals, such as | |
| | vascular disease. | physicians and nurses, in hospitals and | |
| | The Laminar P1 device is intended to be | professional environments. The device is | |
| | used by medical professionals, such as | intended for prescription use only. | |
| | physicians and nurses, in hospitals, clinics | | |
| | and private offices and healthcare | | |
| | facilities. It is to be used on patients, ages | | |
| | 18 years and older. | | |
| Product Code | DPW | DPW | |
| Use Environment | Healthcare settings | Healthcare settings | |
| Patient Population | 18 years and older | 18 years and older | |
| Installation and Use | Body Worn | Body Worn | |
| Applications | Peripheral vascular | Peripheral vascular | |
| Device Description | The Laminar P1 is an ultrasound-based | The FloPatch (FP110) is a non-invasive | |
| | battery-powered sensor patch designed to | blood flow detection device intended to | |
| | detect blood flow in veins and arteries to | be used in a medical/hospital setting for<br>use by medical professionals. The device | |
| | assist in the detection of peripheral<br>vascular disease. The sensor patch is | uses ultrasound and the Doppler effect to | |
| | designed to have the same performance | evaluate the flow of blood. The device | |
| | characteristics as traditional ultrasound | consists of signal processing box (main | |
| | transducer probes (4.5 MHz frequency) | unit) and an ultrasonic vascular flow | |
| | but in a patch form factor that is easy to | transducer. The device transmits | |
| | place on the patient's body for a given | ultrasonic waves from the vascular flow | |
| | time period. | transducer to a peripheral vessel such as | |
| | | the carotid artery. The Doppler shifted | |
| | | ultrasonic waves are reflected by moving | |
| | | blood cells back to the ultrasonic flow | |
| | | transducer. The reflected signal is | |
| | | received by the signal processing box | |
| | | (main unit) which outputs the doppler | |
| | | signal to the device speaker. | |
| Device Components | • Processing Unit | • Main Unit (signal processing box) | |
| | • Sensor Patch with Patch Frame | • Ultrasonic transducer | |
| | • Power Cable | | |
| Dimensions (Length x<br>Width x Height) | 52 x 38 x 22 mm | 135 x 108 x 43.3 mm | |
| Weight | 44 grams | <450 grams | |
| Principle of Operation | Doppler effect to detect blood flow in | Doppler effect to detect blood flow in | |
| | peripheral vasculature | peripheral vasculature | |
| Modes of Operation | One mode, continuous | One mode, continuous | |
| Transducers | 4.5 MHz probe | 4 MHz probe | |
| | | Subject Device | Predicate Device |
| | | Laminar P1 | FloPatch (FP110) |
| | | | K191388 |
| Global<br>Maximum | ISPTA, $\alpha$<br>(mW/cm²) | 204.89 | 21.47 |
| Outputs | MI | 0.0537 | 0.01 |
| Operational Modes | | • Volume levels (silent, low, medium, high)<br>• Receiver options for signal strength (shallow, medium, deep) | No Volume slider |
| Audio Bandwidth | | 150 Hz – 2 KHz | N/A |
| Audio Power | | 0.8 W | N/A |
| Power Source | | Rechargeable 3.7V lithium-polymer battery | Internal AA alkaline batteries |
| Maintenance | | • Reusable processing unit<br>• Single use sensor patch and patch frame | Single use transducer |
{7}------------------------------------------------
#### SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate substantial equivalence based on current industry and FDA recognized standards:
- Biocompatibility per ISO 10993-1 and FDA's guidance Use of International Standard . ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
- . Software per IEC 62304 and FDA's guidance Content of Premarket Submissions for Device Software Functions
- Electrical and EMC safety per IEC 60601-1 and IEC 60601-1-2 .
- . Acoustic output levels per IEC 60601-2-37 and IEC 62359
- . Bench performance testing (functional testing, audio performance testing, mechanical testing)
The results of these tests indicate that the Laminar P1 is substantially equivalent to the predicate device.
#### SUMMARY OF CLINICAL TESTING
No clinical testing was provided to support the demonstration of substantial equivalence.
#### CONCLUSION
Based on the testing performed, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The indications for use, technological characteristics, and performance characteristics for the proposed Laminar P1 are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.