K242437 · Cube Click, Inc. · MYN · May 14, 2025 · Radiology
Device Facts
Record ID
K242437
Device Name
Smile Dx®
Applicant
Cube Click, Inc.
Product Code
MYN · Radiology
Decision Date
May 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2070
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K242437 · May 14, 2025
Smile Dx®
Cube Click, Inc.
Retrospective clinical radiographs collected from multiple U.S. sites
Retrospective clinical radiographs were used to conduct standalone performance testing and a multi-reader, multi-case (MRMC) clinical evaluation to demonstrate the device's effectiveness in aiding dentists in the detection of caries, periapical radiolucencies, and bone level measurements.
Smile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy. Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated with each tooth. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment. Smile Dx® supports both digital and phosphor sensors.
Device Story
Smile Dx® is a web-based CADe software for dentists; it processes digital bitewing and periapical radiographs. Users manually upload images from local computers. The device uses computer vision machine learning to detect and segment caries, periapical radiolucencies (PARL), restorations, and normal anatomy; it also calculates mesial and distal bone levels. Outputs are presented as annotations (contours and color-coded lines) overlaid on the original radiograph within the software's user interface. Dentists use these outputs as a second reader to assist in clinical assessment and diagnosis. The device does not replace professional clinical judgment or comprehensive patient evaluation. It benefits patients by enhancing the accuracy and consistency of dental radiographic interpretation.
Clinical Evidence
Clinical performance was evaluated via a retrospective, fully-crossed MRMC study with 14 US licensed dentists using 352 cases. Ground truth was established by four dentists. Primary endpoint: wAFROC analysis showed statistically significant improvement for caries (+0.127, p<0.001) and PARL (+0.098, p<0.001). Secondary endpoints: Caries sensitivity improved from 64.3% to 83.9%; PARL sensitivity improved from 70.7% to 89.8%. Specificity also improved for both. Standalone testing on 867 cases showed caries sensitivity of 88.3% and PARL sensitivity of 86.1%. Bone level measurement mean absolute error was 0.30 mm for bitewings and 2.6% for periapicals.
Technological Characteristics
Web-based SaMD; compatible with Edge, Chrome, Firefox, Safari. Uses computer vision machine learning algorithms for image analysis. Inputs: digital bitewing/periapical radiographs (jpg, png, tif, tig, dex, gif, bmp, raw). Outputs: contour segmentations for caries/PARL/anatomy/restorations and linear/percentage bone level measurements. User interface supports image adjustment (brightness, contrast, opacity). No direct hardware connection; manual file upload.
Indications for Use
Indicated for licensed dentists to aid in the review of bitewing and periapical radiographs of permanent teeth in adult patients at least 22 years of age.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 14, 2025
Cube Click, Inc.
% Meritxell Martinez
Project Manager and Regulatory Specialist
Innolitics
1101 West 34th St #550
AUSTIN, TX 78705
Re: K242437
Trade/Device Name: Smile Dx®
Regulation Number: 21 CFR 892.2070
Regulation Name: Medical Image Analyzer
Regulatory Class: Class II
Product Code: MYN, LLZ
Dated: April 6, 2025
Received: April 7, 2025
Dear Meritxell Martinez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242437 - Meritxell Martinez
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K242437 - Meritxell Martinez
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242437 | |
| Device Name Smile Dx® | |
| Indications for Use (Describe) Smile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy. Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated with each tooth. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment. Smile Dx® supports both digital and phosphor sensors. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K242437
| | 510(k) Summary |
| --- | --- |
# 1. CONTACT INFORMATION
| Company Name | Cube Click, Inc. |
| --- | --- |
| Address | 105 East 34 Street, Suite 231, New York, NY 10016 |
| Phone Number | +1.646.224.5908 |
| Company Representative | Richard Ricci, D.D.S, M.S., F.A.G.D, President, Cube Click |
| Primary Correspondent(s) | Meritxell Martinez, Regulatory Specialist, Innolitics
Yujan Shrestha, MD, Partner and Co-Founder, Innolitics |
| Primary Correspondent Email | fda@innolitics.com |
| Date Prepared | May 13th, 2025 |
| Documentation Prepared Using | Innolitics Medtech OS Framework |
# 2. DEVICE INFORMATION
| Trade Name | Smile Dx® |
| --- | --- |
| Common Name | Computer-Assisted Detection Device |
| Classification Name | Medical Image Analyzer |
| Primary Product Code | MYN |
| Primary Regulation Number | 21 CFR 892.2070 |
| Secondary Product Code | LLZ |
# 3. PRIMARY PREDICATE DEVICE INFORMATION
| Predicate Device Name | Overjet Caries Assist |
| --- | --- |
| Predicate Device K Number | K212519 |
# 4. REFERENCE DEVICE(S) INFORMATION
| Predicate Device Name | Second Opinion |
| --- | --- |
{5}
CUBE CLICK
510(k) Summary
| Predicate Device K Number | K210365 |
| --- | --- |
| Predicate Device Name | Overjet Dental Assist |
| Predicate Device K Number | K210187 |
# 5. DEVICE DESCRIPTION
Smile Dx® is a computer assisted detection (CADe) device indicated for use by licensed dentists as an aid in their assessment of bitewing and periapical radiographs of secondary dentition in adult patients. Smile Dx® utilizes machine learning to produce annotations for the following findings:
- Caries
- Periapical radiolucencies
- Bone level measurements (mesial and distal)
- Normal anatomy (enamel, dentin, pulp, and bone)
- Restorations
# 6. INDICATIONS FOR USE
Smile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy.
Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated with each tooth.
The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.
Smile Dx® supports both digital and phosphor sensors.
# 7. INTENDED PATIENT POPULATION
The intended patient population of Smile Dx® are patients that have permanent adult dentition and are at least 22 years of age.
{6}
CUBE CLICK
510(k) Summary
# 8. SUBSTANTIAL EQUIVALENCE SUMMARY
Smile Dx® is similar to the predicate device and reference devices in the following ways:
- Intended use: All devices are intended to be used to identify suspected areas of interest in intra-oral radiographs.
- Technology characteristics: All devices employ machine learning algorithms to output detections.
- Safety: As all devices are CADe and/or SaMD systems, and don't pose any direct safety hazard to the patient.
- Clinical Performance: All devices have undergone clinical evaluations which demonstrate statistically significant improvement in aided reader performance.
# 8.1. Device Comparison Table
For technological characteristics comparison, the following table is provided:
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| Manufacturer | Cube Click, Inc. | Overjet, Inc. | Pearl Inc. | Overjet, Inc. | N/A |
| Regulation Number | 892.2070 | 892.2070 | 892.2070 | 892.2050 | Smile Dx® falls under the same regulation (892.2070) as primary predicate (Overjet Caries Assist). |
| Regulation Name | Medical image analyzer | Medical image analyzer | Medical image analyzer | Picture archiving and communications system | Smile Dx® falls under the same regulation (892.2070) as predicate (Overjet Caries Assist). |
| Regulatory Class | Class II | Class II | Class II | Class II | Same |
| Product Code | MYN; LLZ | MYN | MYN | LLZ | Smile Dx® has the same primary product MYN as predicate (Overjet Caries Assist) and reference device |
{7}
4
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | | | | | (Second Opinion). Additionally, Smile Dx®’s secondary product code LLZ is the same as reference device (Overjet Dental Assist). Thus, the subject device has the same product codes as the identified predicate and reference devices. |
| Device Property | SaMD/CADE | SaMD/CADE | SaMD/CADE | SaMD | Same |
| Indications for Use | Smile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy. Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated with | The Overjet Caries Assist (OCA) is a radiological, automated, concurrent read, computer-assisted detection software intended to aid in the detection and segmentation of caries on bitewing radiographs. The device provides additional information for the dentist to use in their diagnosis of a tooth surface suspected of being carious. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant | Second Opinion® is a computer aided detection ("CADe") software to identify and mark regions in relation to suspected dental findings which include Caries, Discrepancy at the margin of an existing restoration, Calculus, Periapical radiolucency, Crown (metal, including zirconia & non-metal), Filling (metal & non-metal), Root canal, Bridge and Implants. It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years | Overjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. Intended Patient Population: | Smile Dx® has similar indications for use to predicate K212519 for detecting caries, reference device K210365 for detecting PARLs, and reference device K210187 for measurement of bone levels. Thus, the subject device combines the indications for use of the three identified predicate and reference devices. All three devices share the same intended use to identify suspected areas of interest in intra-oral radiographs. |
{8}
5
CUBE CLICK
510(k) Summary
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | each tooth. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment. Smile Dx® supports both digital and phosphor sensors. | information from the image, patient history, and actual in vivo clinical assessment. | of age or older as a second reader. | The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition. | |
| Detections and/or measurements | • Normal anatomy
• Caries
• Periapical radiolucencies
• Bone level measurements (mesial and distal)
• Restorations | Caries | Caries, Margin discrepancy, Calculus, Periapical radiolucencies, Restorations | Bone level measurements (mesial and distal) | Smile Dx® detects caries similar to predicate Overjet Caries Assist K212519.
Smile Dx® also detects PARLs and restorations, similar to reference device Second Opinion K210365. Second Opinion also detects other features such as margin discrepancy and calculus, however, these are not supported by subject device. The absence of these features in Smile Dx® does not create a new intended use.
Smile Dx® also measures mesial and distal bone levels, same as reference device Overjet |
{9}
6
| | 510(k) Summary |
| --- | --- |
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | | | | | Dental Assist K210187. |
| Intended User | Dentists | Dentists | Dentists | Dental professionals including, but not limited to, dentists and dental hygienists. | Smile Dx® is intended for use by licensed dentists, same as all of the identified predicate and reference devices. The reference device Overjet Dental Assist is also intended for use by other dental professionals, such as dental hygienists. However, use of Smile Dx® has been limited to use by licensed dentists only. |
| Patient Population | Patients requiring dental services, all sexes, at least 22 years of age, and with permanent dentition. | Patients requiring dental services, all sexes, at least 18 years of age, and with permanent dentition. | Patients requiring dental services, all sexes, at least 12 years of age, and with permanent dentition. | Patients requiring dental services, all sexes, at least 22 years of age, and with permanent dentition. | The intended patient population for Smile Dx® is limited to patients with permanent dentition who are at least 22 years of age. This is identical to the reference device Second Opinion. This intended patient population age has been supported by use of representative patient data in testing. |
| Rx or OTC | Rx-only | Rx-only | Rx-only | Rx-only | Same |
| Technical Characteristics | | | | | |
| Platform | Web - Edge, Chrome, Firefox, Safari | Web - Edge, Chrome, Firefox | Local computer application (Cloud-based installation) | Web - Edge, Chrome, Firefox | The subject device is a web-based application, similar to Overjet Caries Assist and Overjet |
{10}
7
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | | | | | Dental Assist. It is compatible with similar browsers and is also compatible with Safari, which is a commonly used browser. |
| OS | Any | Any | Windows 7 or higher | Any | The subject device is compatible with any operating system, same as predicate Overjet Caries Assist and reference device Overjet Dental Assist. |
| User Interface | Mouse, Keyboard, Trackpad | Mouse, Keyboard, Trackpad | Mouse, Keyboard | Mouse, Keyboard, Trackpad | Same |
| Image Modality | Radiograph | Radiograph | Radiograph | Radiograph | Same |
| Radiograph Type | Bitewing and periapical | Bitewing and periapical | Bitewing and periapical | Bitewing and periapical | Same |
| Image Input Sources | Images imported from local computer, from a range of manufacturers | Images imported from the radiographic device, or from the practice management system, from Carestream or Schick sensors | Radiography devices from a range of manufacturers | Images imported from the radiographic device, or from the practice management system | Images are manually imported from the dentist user’s local computer into Smile Dx®. The subject device does not connect to practice management systems like Overjet Caries Assist and Overjet Dental Assist. However, this difference does not raise any new questions of safety and effectiveness.
Additionally, Smile Dx® is compatible with images from a range of radiography device |
{11}
CUBE CLICK
510(k) Summary
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | | | | | manufacturers, similar to Second Opinion. This has been supported by using data from a range of manufacturers during testing. |
| Image Format | jpg, png, tif, tig, dex, gif, bitmap/bmp, and raw | jpg, png, eop, jif, dicom | jpeg, png, rvg, dcm, tiff, and dic | jpg, png, jfif, eop, etp, jif | Smile Dx® supports image formats that are commonly used in dental radiographs. The difference in supported image formats between the subject and predicate devices is considered to be minor and does not interfere with the ability of Smile Dx® to achieve its intended use. |
| Processing Architecture | Three layers: - The Network layer contains a login module and a data ingestion module. Radiographs are manually uploaded from local computer to the server. - The Detection and Measurement layer processes the image and classifies it as either bitewing or periapical, and then annotates it via the algorithm - The visualization layer displays the annotated image via Smile Dx®’s | Three layers: - The Network layer works with the practice PACS or EMR to transmit the image and meta-data to Overjet. - The decision layer processes the image to ensure it is the correct data type, and then annotates it via the algorithm - The presentation layer displays the annotated image in a non diagnostic viewer. The dentist can filter, | The Second Opinion Client continuously monitors a local or networked resource for dental radiographs. When new images are found, it sends them to cloud-based Computer Vision Models (CV Models) for analysis. These models generate metadata describing the nature and location of detected features in the radiographs. The metadata is returned to the Second Opinion® Client, which | Unknown | Smile Dx® has a similar processing architecture to the predicate Overjet Caries Assist (K212519) and reference device Second Opinion (K210365). All devices obtain radiographs, annotate them, and present annotations to via a user interface. One of the main differences in processing architecture is that the predicate device can establish connections with external sources to upload images, whereas the subject |
8
{12}
CUBE CLICK
510(k) Summary
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | user interface. The dentist can adjust image settings, display and hide the presented annotations. | display, hide, create and edit the annotations presented. | displays it within its user interface. Detected features are highlighted with color-coded boundary boxes overlaid on the original radiograph. | | device exclusively uploads images from the local computer. Similarly, users of Smile Dx® are required to manually upload images to the software, whereas the reference device Second Opinion automatically uploads images. Nevertheless, these technological differences do not raise new questions of safety and effectiveness. The absence of these features in the subject device does not impede its ability to fulfill its intended use. |
| Algorithm | Utilizes computer vision machine learning algorithm(s). | Utilizes computer vision machine learning algorithm(s). | Utilizes computer vision neural network algorithms, developed from open-source models using supervised machine learning techniques. | Utilizes computer vision techniques. | Same. Smile Dx® utilizes computer vision techniques, specifically, machine learning algorithms to detect suspected findings, same as the identified predicate and reference devices. |
| Output | Suspected findings (i.e., caries, PARLs) and measurements (bone levels) are overlaid on the original radiograph | Caries detection and segmentation on radiograph resulting in outline of suspected caries | Detected features (e.g., caries, PARLs) are highlighted with color-coded boundary boxes overlaid on the original radiograph | Bone-level annotations on radiograph | The output of Smile Dx® is similar to the outputs of all of the identified predicate and reference devices, as they all overlay the findings on the original radiograph. |
| Marker Type/Size | Contour segmentations for caries and periapical | Presents suspected carious lesions as segmented | Color-coded bounding boxes / Fixed | Color-coded lines for bone levels | The subject device has similar marker types to its predicate and reference |
9
{13}
10
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | radiolucencies. Color-coded lines for bone levels | polygons outlining the prediction | | | devices.
For suspected caries, the subject device creates a contour of the suspected caries, similar to the predicate (Overjet Caries Assist).
For periapical radiolucencies, the subject device creates countours. This is different to reference device’s (Second Opinion) bounding boxes. The performance and accuracy of the segmentations has been supported with appropriate performance testing.
For bone levels, the subject device outputs color-coded lines depending on the level of measured boneloss, similar to reference device Overjet Dental Assist. |
| Image Measurement | Linear distance | N/A | N/A | Linear distance | Smile Dx® is similar to reference device Overjet Dental Assist in that it creates a linear distance for measured bone levels. |
| Image Viewing | Full, Thumbnail | Full, Thumbnail | Full, Thumbnail | Full, Thumbnail | Same |
| Image Manipulation | Image adjustment tools (e.g., brightness, contrast, opacity, | Annotations | Image adjustment tools (e.g., brightness, | Annotation (line) | Smile Dx® has similar image adjustment tools as reference devie Second Opinion |
{14}
510(k) Summary
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | rotation prior to upload) | | contrast, zoom, invert, rotate) | | (e.g., contrast and brightness). Smile Dx® introduces a new tool (i.e., opacity adjustment), however, this does not raise any new questions of safety and effectiveness and is a common image adjustment tool.
Smile Dx® lacks some of the features found in the predicate and reference devices (e.g., zoom, invert, and manual annotation tools). However, these tools are not needed to accomplish the intended use of the device and do not raise new questions of safety and effectiveness. |
| Performance Testing | • Standalone performance study for pathologic feature (i.e., caries and periapical radiolucency) detection, non-pathologic feature detection (i.e., normal anatomy and restorations) and bone level measurement performance.
• Multiple-Reader, Multiple-Case (MRMC) study for pathologic dental features | • Standalone study for pathologic feature detection performance.
• Multiple-Reader, Multiple-Case (MRMC) study for pathologic dental features
• Analysis included:
• Sensitivity and specificity evaluations
• Dice coefficient analysis | • Standalone study for pathological and non-pathologic feature detection performance.
• Multiple-Reader, Multiple-Case (MRMC) study for pathologic dental features
• Analysis included:
• wAFROC-FOM analysis for primary endpoints
• Determination of the changes in sensitivity and change in number | • Bench testing evaluated precision and recall against labeled keypoints within radiographs.
• Retrospective clinical performance testing that compared adjudicated measurements against Overjet Dental Assist’s predicted measurements.
Analysis included | Smile Dx® underwent similar evaluations as predicate and reference devices, following recommendations in 2022 FDA Guidance “Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Notification [510(k)] Submissions”. |
{15}
12
CUBE CLICK
510(k) Summary
| Characteristic | Smile Dx® | Predicate Device: Overjet Caries Assist (K212519) | Reference Device: Second Opinion (K210365) | Reference Device: Overjet Dental Assist (K210187) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | • Analysis included:
• wAFROC analysis for primary endpoint
• Sensitivity and specificity evaluations | | of false positive dental pathologies of a given type per image (FPPI) | sensitivity and specificity. | |
# 9. PERFORMANCE TESTING
## 9.1. Software Verification and Validation
As Smile Dx® is a SaMD, software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Device Software Functions”.
## 9.2. CADe Testing
Smile Dx® was subjected to clinical performance testing according to 2022 FDA Guidance “Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Notification [510(k)] Submissions” Testing involved both standalone performance testing and a fully-crossed multi-reader multi-case (MRMC) reader clinical evaluation. As CADe devices, predicate device Overjet Caries Assist (K212519) and Second Opinion (K210365) also underwent the recommended testing.
### 9.2.1. Standalone Testing
#### Caries and Periapical Radiolucency Detection
Smile Dx®'s caries and periapical radiolucency detection performance was evaluated using a test dataset of 867 cases collected from multiple U.S. sites. The test results were as follows:
| Metric | Caries | Periapical Radiolucency |
| --- | --- | --- |
| Dice | 0.74 [0.72 0.76] | 0.77 [0.74, 0.80] |
| Sensitivity | overall: 88.3% [83.5%, 92.6%] | 86.1% [80.2%, 91.9%] |
#### Bone Level Detection and Bone Loss Measurement
{16}
CUBE CLICK
510(k) Summary
Smile Dx®'s bone level detection and bone loss measurement performance was evaluated using a test dataset of 352 cases collected from multiple U.S. sites. The test results were as follows:
| Radiograph Type | Bone Level Detection - Sensitivity | Bone Level Detection - Specificity | Bone Loss Measurement (Mean Absolute Error [95% CI] |
| --- | --- | --- | --- |
| Bitewing | 95.5% [94.3%, 96.7%] | 94.0% [91.1%, 96.6%] | 0.30 mm [0.29mm, 0.32mm] |
| Periapical | 87.3% [85.4%, 89.2%] | 92.1% [89.9%, 94.1%] | 2.6% [2.4%, 2.8%] |
# Normal Anatomy and Restorations
Smile Dx®'s performance for normal anatomy and restoration detection was evaluated using a test dataset of 200 cases collected from different U.S. sites. The test results were as follows:
| Non-pathologic Region | Dice | Sensitivity (Pixel-level) | Sensitivity (Contour-level) | Specificity (Contour-level) |
| --- | --- | --- | --- | --- |
| Normal Anatomy | 0.84 [0.83, 0.85] | 86.1% [85.4%, 86.8%] | 95.2% [94.5%, 96%] | 93.5% [91.6%, 95.8%] |
| Restorations | 0.87 [0.85, 0.90] | 83.1% [80.3%, 86.4%] | 90.9% [88.2%, 93.9%] | 99.6% [99.3%, 99.8%] |
# 9.2.2. MRMC Clinical Evaluation - Reader Improvement
The clinical performance assessment for Smile Dx® used a fully-crossed, multiple-reader multiple-case (MRMC) evaluation method. The test dataset consisted of 352 cases collected from multiple U.S. sites. Smile Dx® underwent clinical evaluation using methods similar to those used for the primary predicate device Overjet Caries Assist (K212519):
Both devices were evaluated in a multi-reader, multi-case (MRMC) retrospective study with at least 13 US licensed dentists (Smile Dx® had 14 readers). Ground truth was established by the consensus labels of at least three US licensed dentists (the ground truth for Smile Dx®'s study was established by four US licensed dentists). Half of the data set contained unannotated images, and the second half contained radiographs that had been processed through the CADe device. The radiographs were presented to the readers in alternating groups throughout two different sessions, which were separated by a washout period. The results were compared against the consensus ground truth, and the sensitivity, specificity, and alternative free response receiver operating characteristic (AFROC) was evaluated to characterize the performance of the readers with and without viewing the model annotations.
# Primary Endpoint: wAFROC Analysis
The weighted alternative free-response receiver operating characteristic (wAFROC) figure of merit serves as the primary performance metric, giving equal importance to each case while reducing the influence of cases with multiple lesions.
{17}
CUBE CLICK
510(k) Summary
The wAFROC analysis revealed statistically significant improvements, with an increase of $\Delta \theta$ for both caries detection (+0.127 [0.081, 0.172]) and PARL detection (+0.098 [0.061, 0.135]). These results demonstrate that Smile Dx® enhances reader sensitivity without compromising specificity, thereby meeting the acceptance criteria for effective detection of caries and PARLs.
| | θcontrol | θCAD | Δθ [95% CI] | p-value |
| --- | --- | --- | --- | --- |
| Caries | 0.781 | 0.908 | +0.127 [0.081, 0.172] | < 0.001 |
| PARLs | 0.849 | 0.947 | +0.098 [0.061, 0.135] | < 0.001 |
# Secondary Endpoint: Sensitivity and Specificity
Sensitivity metrics were determined at the lesion level. Specificity metrics were determined at the case level.
# Sensitivity
For caries detections, reader sensitivity improved from $64.3\%$ without device to $83.9\%$ with device, an increase of $19.6\%$ [12.8%, 26.4%]. For PARL detections, reader sensitivity improved from $70.7\%$ without device to $89.8\%$ with device, an increase of $19.1\%$ [13.6%, 24.7%].
# Specificity
For caries detections, reader specificity improved from $73.6\%$ without device to $90.2\%$ with device, a difference of $16.7\%$ [13.5%, 19.9%]. For PARL detections, reader specificity improved from $92.6\%$ without device to $97.3\%$ with device, a difference of $4.7\%$ [3%, 6.4%].
# Sub-group Analysis
Sub-group analyses were performed for patient demographics (i.e., age, sex, race), imaging hardware, and primary vs. secondary classification (caries). No distinct differences in performance were observed among sub-groups that raise concerns about the effectiveness of Smile Dx®.
# 10. CONCLUSION
Smile Dx® is substantially equivalent to the identified predicate device. Specifically, the subject device has similar indications and technological characteristics when compared to the predicate device. Any differences in technological characteristics do not raise new questions of safety or effectiveness.
Additionally, the results of the performance testing (i.e., software verification and validation, standalone performance assessment and MRMC clinical evaluation) of Smile Dx® support that the subject device's performance is as effective as the predicate and reference devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.