SiOxC Cream

K242406 · Sioxmed, LLC · FRO · Mar 13, 2025 · SU

Device Facts

Record IDK242406
Device NameSiOxC Cream
ApplicantSioxmed, LLC
Product CodeFRO · SU
Decision DateMar 13, 2025
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

SiOxC Cream is a skin emulsion indicated for management of dry intact skin by maintaining a moist skin environment. Not to be used on Breached or Compromised skin or open sores.

Device Story

SiOxC Cream is a non-sterile, fragrance-free, preservative-protected topical skin emulsion. It functions by maintaining a moist environment on dry, intact skin. The cream is applied topically by the patient or caregiver, typically twice daily or as needed. It is provided as a prescription-only product. The formulation includes emollients, humectants, structural/stabilizing agents, pH adjusters, surfactants/emulsifiers, antioxidants, and a proprietary amorphous hydrated silica. By supporting a moist skin environment, the device aids in the management of dry skin conditions. It is intended for single-patient use.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-1, -5, -10, and -23. Preservative effectiveness testing conducted per USP <51>. Clinical data includes a Human Repeat Insult Patch Test (HRIPT) with 106 subjects (18-70 years old). Results showed 100% of subjects had no visible skin reaction during Induction or Challenge phases, demonstrating no potential for dermal irritation or sensitization.

Technological Characteristics

Water-based emulsion containing emollients, humectants, structural/stabilizing agents, pH adjusters, surfactants/emulsifiers, antioxidants, and amorphous hydrated silica. Preserved with phenoxyethanol (0.5% ± 1%). Non-sterile. Applied topically. Biocompatible per ISO 10993 standards.

Indications for Use

Indicated for management of dry intact skin in patients requiring a moist skin environment. Contraindicated for use on breached or compromised skin or open sores.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 13, 2025 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. SiOxMed, LLC % Justin Gracyalny Regulatory Affairs Program Manager Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock. Georgia 30188 Re: K242406 Trade/Device Name: SiOxC Cream Regulatory Class: Unclassified Product Code: FRO Dated: August 14, 2024 Received: August 14, 2024 Dear Justin Gracyalny: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. {2}------------------------------------------------ For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Yu-chieh Chiu -S Yu-Chieh Chiu, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K242406 Device Name SiOxC Cream Indications for Use (Describe) SiOxC Cream is a skin emulsion indicated for management of dry intact skin by maintaining a moist skin environment. Not to be used on Breached or Compromised skin or open sores. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ | Date Prepared | March 13, 2025 | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor | SiOxMed<br>2452 Salem Park Dr.<br>Winston-Salem, NC 27127<br>(336) 551-2209 | | 510(k) Contact | SiOxMed<br>Adam Jorgensen, MD, PhD<br>2452 Salem Park Dr.<br>Winston Salem, NC, 27127<br>(336) 551-2209<br>Regulatory@SiOxMed.com | | Trade Name | SiOxC Cream | | Common Name | Cream | | Code - Classification | FRO<br>Unclassified | | Predicate Device | K052643 Ceragenix Corporation EPICERAM® Skin Barrier Emulsion | | Reference Device | K241660 SiOxMed LLC SiOxD Wound Matrix | | Device Description | SiOxC Cream is a fragrance free, preservative protected, non-sterile, topical cream intended for management of dry skin. SiOxC Cream supports a moist environment. SiOxC Cream is provided prescription only in various sizes for single patient use. | | Indications for Use Statement | SiOxC Cream is a skin emulsion indicated for management of dry intact skin by maintaining a moist skin environment. Not to be used on Breached or Compromised skin or open sores. | #### 510(k) SUMMARY: SiOxMed SiOxC Cream #### Comparison of Technological Characteristics The subject device is intended for use for management of dry intact skin. As such, this difference does not represent a new intended use. There are no significant technological differences between the subject and predicate devices are intended for similar uses and have similar technological characteristics including promoting a moist environment as the principle of operation, application frequency, and sterility. Both the subject and predicate device utilize the same preservative. Any minor differences such as specific cream formulation and size offerings are addressed via biocompatibility and non-clinical performance testing. {5}------------------------------------------------ | Characteristic | Subject Device<br>SiOxMed LLC<br>SiOxC Cream K242406 | Primary Predicate<br>Oculus Innovative Sciences, Inc.<br>EPICERAM® Skin Barrier<br>Emulsion<br>K052643 | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Regulation | Unclassified | Unclassified | | Product Classification | FRO | FRO | | Common Name | Cream | Cream | | Principle of Operation | Promotes a moist environment | Promotes a moist environment | | Composition | A water-based formulation containing<br>emollients, humectants, structural and<br>stabilizing agents, pH adjusters,<br>surfactants/emulsifiers, antioxidants,<br>and preservatives. The formulation<br>includes a proprietary<br>amorphous<br>hydrated silica. | A water-based formulation with<br>emollients, humectants, lipid-<br>replenishing agents, stabilizers, pH adjusters,<br>surfactants/emulsifiers,<br>antioxidants, and preservatives. The<br>formulation includes a proprietary lipid<br>delivery system. | | Preservative | Phenoxyethanol (0.5% ± 1%) | Phenoxyethanol | | Available Offerings | 1 or 8 fl oz jar | 90g tube or 225g airless pump | | Reapplication | Twice daily, as needed | Twice daily, as needed | | Sterility | Non-Sterile | Non-Sterile | | Biocompatibility | Biocompatible | Biocompatible | ### Summary of Performance Testing Non-clinical performance testing for the SiOxC Cream included characterization of pH, viscosity, and macroscopic and microscopic appearance. Preservative effectiveness testing per USP <51> was also conducted to support the subject device shelf-life. The SiOxC Cream was found to be biocompatible in accordance with ISO 10993-1 for its intended use when tested in compliance with ISO 10993-5, ISO 10993-10, and ISO 10993-23. A Human Repeat Insult Patch Test (HRIPT) was also performed to determine the potential of the test material to elicit dermal irritation and/or induce sensitization following repeated patch applications in human subjects. The Induction Phase of the study is designed to assess the potential of the subject device to elicit an irritation reaction, whereas the Challenge Phase of the study is designed to assess the potential of the subject device to elicit a sensitization response. 120 male and female subjects ranging from 18 to 70 years old were enrolled in the subjects who completed the Induction Phase, 100% were categorized as "No visible skin reaction" at any time point. Of the subjects who completed the Challenge Phase, 100% were categorized as "No visible skin reaction" at any time point. No re-challenge testing was required for any subjects. {6}------------------------------------------------ Based on the test population of 106 subjects who completed the study, SiOxC Cream did not demonstrate a potential for eliciting dermal irritation or inducing sensitization. #### Conclusions Based on the similarities of the intended use / indications for use, technological and functional characteristics, and the results of the biocompatibility / non-clinical performance testing, the subject device is substantially equivalent to the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...