← Product Code [POV](/productcode/POV) · K242388

# LensHooke X12 PRO Semen Analysis System (K242388)

_Bonraybio Co., Ltd. · POV · May 9, 2025 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K242388

## Device Facts

- **Applicant:** Bonraybio Co., Ltd.
- **Product Code:** [POV](/productcode/POV.md)
- **Decision Date:** May 9, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** AI/ML

## Indications for Use

The LensHooke X12 PRO Semen Analysis System used with LensHooke Semen Test Slides is an optical device for human semen analysis which provides direct and calculated measurements for: - Sperm concentration (M/mL) - Total motility (PR+NP, %) - Progressive Motility (PR, %) - Sperm Morphology (normal forms, %) The LensHooke X12 PRO Semen Analysis System used with LensHooke R10 Plus Sperm DNA Fragmentation Rapid Test Kit (SCD Assay) and LensHooke R11 Plus Sperm Double Strand DNA Fragmentation Rapid Test Kit (SDFR Assay) is an optical device for human semen analysis which provides direct measurement for: - Sperm DNA Fragmentation Index (DFI, %) The LensHooke X12 PRO Semen Analysis System does not provide a comprehensive evaluation of a male's fertility status. It is an in-vitro diagnostic system intended for human semen analysis of individuals in clinical laboratories to evaluate male fertility.

## Device Story

Desktop optical system for human semen analysis; utilizes built-in video microscopy and AI-based image processing to automate assessment of sperm quality. Inputs: semen samples loaded into specific test slides (CS3 for traditional parameters; R10/R11 kits for DNA fragmentation). Operation: cameras capture images; software detects sperm locations; concentration calculated by unit density; motility via trajectory tracing; morphology via head/tail percentage; DFI via halo dispersion analysis (SCD/SDFR methods). Output: quantitative measurements of concentration, motility, morphology, and DFI. Used in clinical laboratories by professionals. Results assist clinicians in evaluating male fertility status; does not provide comprehensive fertility diagnosis.

## Clinical Evidence

User performance study conducted with professional users across educational backgrounds. Compared subject device results against reference method (LensHooke X1 PRO). Demonstrated user accuracy and ease of use via participant questionnaire scoring. No clinical diagnostic accuracy metrics (sensitivity/specificity) reported; focus on usability and performance verification.

## Technological Characteristics

Desktop unit; light sources; built-in video microscopy; internal computer. Consumables: CS3 test slides, R10/R11 DNA fragmentation kits. Connectivity: HDMI, USB, Ethernet. Software: AI-based image processing. Factory calibrated. Standards: CLSI EP05-A3, EP06-Ed2, EP07, EP25-A2, ISO 14971, IEC 60601-1, IEC 61326-1.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- LensHooke X1 PRO Semen Quality Analyzer ([K202089](/device/K202089.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY AND INSTRUMENT

## I Background Information:

A 510(k) Number

K242388

B Applicant

Bonraybio Co., LTD.

C Proprietary and Established Names

LensHooke X12 PRO Semen Analysis System

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  POV | Class II | 21 CFR 864.5220 - Automated Differential Cell Counter | HE - Hematology  |

## II Submission/Device Overview:

A Purpose for Submission:

Clearance of a new device

B Measurand:

Sperm concentration (M/mL), total motility (%), progressive motility (%), morphology (%), and DNA Fragmentation Index

C Type of Test:

Analysis of semen parameters

## III Intended Use/Indications for Use:

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K242388 - Page 2 of 15

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

The LensHooke X12 PRO Semen Analysis System used with LensHooke Semen Test Slides is an optical device for human semen analysis which provides direct and calculated measurements for:

- Sperm concentration (M/mL)
- Total motility (PR+NP, %)
- Progressive Motility (PR, %)
- Sperm Morphology (normal forms, %)

The LensHooke X12 PRO Semen Analysis System used with LensHooke R10 Plus Sperm DNA Fragmentation Rapid Test Kit (SCD Assay) and LensHooke R11 Plus Sperm Double Strand DNA Fragmentation Rapid Test Kit (SDFR Assay) is an optical device for human semen analysis which provides direct measurement for:

- Sperm DNA Fragmentation Index (DFI, %)

The LensHooke X12 PRO Semen Analysis System does not provide a comprehensive evaluation of a male's fertility status. It is an in-vitro diagnostic system intended for human semen analysis of individuals in clinical laboratories to evaluate male fertility.

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

D Special Instrument Requirements:

- LensHooke X12 PRO Semen Analysis System

IV Device/System Characteristics:

A Device Description:

LensHooke X12 PRO Semen System consists of the following components:

- LensHooke X12 PRO Semen Analysis System
- LensHooke CS3 10 µm Semen Test Slides
- LensHooke R10 Sperm DNA fragmentation test kit (SCD assay)
- LensHooke R11 Sperm double strand DNA fragmentation test kit (SDFR assay)
- LensHooke X QC Beads (For Semen) (cleared in K202089)
- LensHooke X QC Slide (For Semen)
- Self-Test Slide

{2}

- Power Cable

For CS3, R10, and R11 slides, there are two wells with 25 fields per well (5x5). Samples can be evaluated per well or across two wells. The fields evaluated are determined prior to analysis and can follow two automatic modes or customized modes, as shown in the following:

- Auto Mode 1 (CS3): six default center fields in well 1 are counted;
- Auto Mode 2 (CS3): six default center fields in well 2 are counted;
- Auto Mode 1 (DFI): all 25 fields in well 1 are counted;
- Auto Mode 2 (DFI): all 25 fields in well 2 are counted;
- Customized (either CS3 or DFI): end-users can pre-select any number or arrangement of fields they prefer. The end user may customize the mode to utilize both wells, if the concentration is expected to be low;
- For CS3, additional fields are also added by the software automatically if the count does not reach 200.

## B Principle of Operation:

LensHooke X12 PRO Semen Analysis System integrates optical design and image analysis, and is combined with artificial intelligence image processing method, to fully automate analysis of semen quality including sperm concentration, motility, morphology and DNA fragmentation index. The images are captured and recorded by cameras and with image processing methods, the locations of sperms are detected. The sperm concentration is analyzed by the sperm unit density; the sperm motility is calculated by tracing sperm trajectories and the sperm morphology is calculated by comparing head and tail percentage.

LensHooke R10 Plus Sperm DNA fragmentation test kit (SCD assay) is an assay using Sperm Chromatin Dispersion (SCD) method to evaluate DNA fragmentation in human spermatozoa and analyze sperm DNA fragmentation index. The SCD assay uses unfixed semen sample (fresh, diluted, or neat samples) embedded in a melted agarose microgel which is attached to a pretreated glass slide. After DNA denaturation, lysis of nuclear proteins and staining procedures, sperms with fragmented DNA do not form the characteristic DNA halo as seen in sperms with intact DNA.

LensHooke R11 Plus Sperm double strand DNA fragmentation test kit (SDFR assay) is an assay using Sperm DNA Fragmentation Release (SDFR) method to evaluate double-stranded DNA fragmentation in human spermatozoa and assess sperm DNA fragmentation index. The SDFR Assay uses unfixed semen samples (fresh, diluted, or neat samples) embedded in a hydrogel which is attached to a pretreated slide. After lysis of nuclear proteins and staining procedures, sperms with fragmented DNA form the dispersion DNA halos which are not seen in sperms with intact DNA.

Through staining by the DNA fragmentation test kit, the DNA fragmented and non-DNA fragmented sperms are categorized. The DNA fragmentation index is calculated as the ratio described in the formula below.

$$
\text{DNA fragmentation index} = \frac{\text{Number of sperms with fragmented DNA}}{\text{Number of sperms evaluated}} \times 100\%
$$

K242388 - Page 3 of 15

{3}

K242388 - Page 4 of 15

C Instrument Description Information:

1. Instrument Name:
LensHooke X12 PRO Semen Analysis System

2. Specimen Identification:
A patient ID number can be manually entered or barcode scanned on LensHooke X12 PRO Semen Analysis System.

3. Specimen Sampling and Handling:
The sample should be allowed to liquefy at room temperature for at least 30 minutes before testing. The test can be performed up to 1 hour after sample collection. It is recommended that users allow 2 to 7 days without ejaculation before collecting a semen sample. Condoms and lubricants should not be used when collecting a semen sample. Hands should be washed with soap and water before and after handling the semen sample.

4. Calibration:
The LensHooke X12 PRO Semen Analysis System is factory calibrated. User calibration is not required.

5. Quality Control:
External QC materials are provided for users of the LensHooke X12 PRO upon request by contacting customer service. Quality controls are provided as three different sperm concentration levels (negative, low positive (~25 × 10⁶/mL), high positive (~50 × 10⁶/mL)), which can be supplied in two formats: LensHooke X QC Beads (For Semen) or LensHooke X QC Slides. The QCs assess the accuracy and precision of sperm counting by providing a known target value and +/- range.

For DNA Fragmentation Index assays (R10 Plus, R11 Plus), the positive and negative control samples are prepared as per the recommended procedures in the R10 Plus Sperm DNA Fragmentation Rapid Test Kit (SCD Assay) user manual and R11 Plus Sperm Double Strand DNA Fragmentation Rapid Test Kit (SDFR0 Assay) user manual.

V Substantial Equivalence Information:

A Predicate Device Name(s):
LensHooke X1 PRO Semen Quality Analyzer, LensHooke X1 PRO SE Semen Quality Analyzer

B Predicate 510(k) Number(s):
K202089

{4}

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K242388 | K202089  |
| --- | --- | --- |
|  Device Trade Name | LensHooke X12 PRO
Semen Analysis System | LensHooke X1 PRO
Semen Quality Analyzer  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | The LensHooke X12 PRO
Semen Analysis System
used with LensHooke
Semen Test Slides is an
optical device for human
semen analysis which
provides direct and
calculated measurements
for:
Sperm concentration
(M/mL)
Total motility (PR+NP,
%)
Progressive Motility (PR,
%)
Sperm Morphology
(normal forms, %)

The LensHooke X12 PRO
Semen Analysis System
used with LensHooke R10
Plus Sperm DNA
Fragmentation Rapid Test
Kit (SCD Assay) and
LensHooke R11 Plus
Sperm Double Strand
DNA Fragmentation
Rapid Test Kit (SDFR
Assay) is an optical
device for human semen
analysis which provides
direct measurement for:
Sperm DNA
Fragmentation Index
(DFI, %)

The LensHooke X12 PRO
Semen Analysis System
does not provide a
comprehensive evaluation | The LensHooke X1 PRO
Semen Quality Analyzer
used with LensHooke
Semen Test Cassette is an
optical device for human
semen analysis which
provides direct and
calculated quantitative
measurements for:
(1) Sperm concentration
(10^6 per ml)
(2) Total motility
(PR+NP, %)
• Progressive motility (%)
• Non-Progressive
motility (%)
(3) Sperm morphology
(normal forms, %)
(4) pH value

The LensHooke X1 PRO
Semen Quality Analyzer
does not provide a
comprehensive evaluation
of a male's fertility status.
It is an in-vitro diagnostic
system intended for
human semen analysis of
individuals in healthcare
professional setting to
evaluate male fertility.  |

K242388 - Page 5 of 15

{5}

|   | of a male's fertility status. It is an in-vitro diagnostic system intended for human semen analysis of individuals in clinical laboratories to evaluate male fertility. |   |
| --- | --- | --- |
|  Technology | Desk-top unit consists of light sources, built-in video microscopy and an internal computer containing algorithms for the assessment of semen parameters. | Same  |
|  Intended User | Clinical laboratory professional | POC and Clinical laboratory professional  |
|  Parameters | Sperm concentration (M/mL), total motility (%), sperm morphology (%) | Same  |
|  General Device Characteristic Differences |  |   |
|  Transmission interface | HDMI/USB/Ethernet | HDMI/USB  |
|  Standard/Guidance Document | WHO 5^{th} and 6^{th} Edition | WHO 5^{th} Edition  |
|  Compatible/Consumable | Semen Test Slide (CS3) | Semen Test Cassette (CS0, CS1)  |
|  Test Kits | SCD Assay (R10), SDFR Assay (R11) | N/A  |
|  Different parameters | DNA Fragmentation Index (DFI) | pH  |
|  Control Material | X QC Beads, X QC Slide | X QC Beads, X QC Reticle  |
|  Sample prep | Liquefaction/ dilution | Liquefaction/ sample cup  |

VI Standards/Guidance Documents Referenced:

CLSI EP05-A3 (Reaffirmed: September 2019): Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition

CLSI EP06-Ed2 (2020): Evaluation of the Linearity of Quantitative Measurement Procedures - Second Edition

CLSI EP07 (2018): Interference Testing in Clinical Chemistry - Third Edition

CLSI EP25-A2 (2023): Evaluation of Stability of In Vitro Diagnostic Reagents - Second edition

K242388 - Page 6 of 15

{6}

ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices

IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 61326-1 Edition 3.0 2020-10 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements

IEC 61326-2-6 Edition 3.0 2020-10 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment

IEC 61000-3-2:2018/AMD1:2020 Amendment 1 - Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic current emissions (equipment input current ≤16 A per phase)

IEC 61000-3-3:2013+AMD1:2017 Electromagnetic compatibility (EMC) - Part 3-3: Limits - Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current ≤16 A per phase and not subject to conditional connection

## VII Performance Characteristics (if/when applicable):

### A Analytical Performance:

1. Precision/Reproducibility:

**Repeatability**

A precision study was conducted evaluating sperm concentration, motility, progressive motility, and morphology using seven native semen samples. A separate repeatability study was conducted to evaluate the precision of the DFI on the LensHooke device. Semen samples were prepared at seven DFI values spread across the measuring range as verified by the manual method (reference). For both studies, the analysis was carried out over the course of one day by three operators using two analyzer/three test slide lot combinations. Due to the limited stability of semen samples, each "day" in the statistical analysis represents different times of day (e.g., every one hour = one "day"). Measurements were separated into five separate 1-hour time periods or "days". The study includes three replicates per run, and two runs for every one hour, five times/day (three operators × three replicates (with three lots) × two runs (with two devices) × five times/day = 90 data points per sperm concentration/sperm motility/sperm morphology level). The sample test on well 1 at RUN 1 and test on well 2 at RUN 2. Data analyses were performed to provide repeatability, between-run, within-day, between-day, and within-laboratory. All results met the predefined acceptance criteria.

K242388 - Page 7 of 15

{7}

|  Concentration (10⁶/mL) |   |   | Repeatability |   | Between-Run |   | Between-Day |   | Within-Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 90 | 6.97 | 0.6 | 8.2% | 0.0 | 0.0% | 0.1 | 1.0% | 0.6 | 8.3%  |
|  2 | 90 | 75.50 | 6.0 | 7.9% | 0.0 | 0.0% | 1.1 | 1.5% | 6.1 | 8.0%  |
|  3 | 90 | 231.09 | 8.5 | 3.7% | 4.7 | 2.0% | 0.0 | 0.0% | 9.7 | 4.2%  |
|  4 | 90 | 74.27 | 4.0 | 5.3% | 1.3 | 1.8% | 0.0 | 0.0% | 4.2 | 5.6%  |
|  5 | 90 | 65.29 | 3.7 | 5.7% | 0.0 | 0.0% | 0.0 | 0.0% | 3.7 | 5.7%  |
|  6 | 90 | 93.32 | 7.5 | 8.1% | 0.0 | 0.0% | 3.2 | 3.4% | 8.2 | 8.8%  |
|  7 | 90 | 63.62 | 4.7 | 7.4% | 2.0 | 3.2% | 0.0 | 0.0% | 5.1 | 8.0%  |
|  Motility (%) |   |   | Repeatability |   | Between-Run |   | Between-Day |   | Within-Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 90 | 69.3 | 6.3 | 9.10% | 0.4 | 0.60% | 0.0 | 0.00% | 6.3 | 9.10%  |
|  2 | 90 | 4.8 | 0.4 | 9.00% | 0.1 | 1.90% | 0.0 | 0.00% | 0.4 | 9.20%  |
|  3 | 90 | 81.4 | 3.7 | 4.50% | 0.0 | 0.00% | 0.9 | 1.10% | 3.8 | 4.60%  |
|  4 | 90 | 42.8 | 2.6 | 6.00% | 0.0 | 0.00% | 0.0 | 0.00% | 2.6 | 6.00%  |
|  5 | 90 | 65.8 | 4.0 | 6.10% | 0.7 | 1.10% | 0.9 | 1.40% | 4.2 | 6.30%  |
|  6 | 90 | 54.4 | 2.5 | 4.70% | 0.6 | 1.00% | 0.0 | 0.00% | 2.6 | 4.80%  |
|  7 | 90 | 80.2 | 4.5 | 5.60% | 1.9 | 2.30% | 0.0 | 0.00% | 4.9 | 6.10%  |
|  Progressive Motility (%) |   |   | Repeatability |   | Between-Run |   | Between-Day |   | Within-Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 90 | 59.2 | 5.7 | 9.60% | 1.6 | 2.70% | 0.0 | 0.00% | 5.9 | 10.00%  |
|  2 | 90 | 4.1 | 0.4 | 9.70% | 0.0 | 0.00% | 0.1 | 2.00% | 0.4 | 9.90%  |
|  3 | 90 | 70.0 | 4.1 | 5.90% | 1.1 | 1.60% | 0.0 | 0.00% | 4.3 | 6.10%  |
|  4 | 90 | 36.8 | 2.6 | 6.90% | 0.0 | 0.00% | 0.3 | 0.90% | 2.6 | 7.00%  |
|  5 | 90 | 56.9 | 3.9 | 6.90% | 1.0 | 1.80% | 0.6 | 1.10% | 4.1 | 7.20%  |
|  6 | 90 | 46.9 | 3.1 | 6.60% | 0.0 | 0.00% | 0.0 | 0.00% | 3.1 | 6.60%  |
|  7 | 90 | 69.0 | 4.7 | 6.80% | 2.0 | 2.90% | 0.0 | 0.00% | 5.1 | 7.40%  |
|  Morphology (%) |   |   | Repeatability |   | Between-Run |   | Between-Day |   | Within-Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  1 | 90 | 7.3 | 0.7 | 9.60% | 0.0 | 0.00% | 0.0 | 0.70% | 0.7 | 9.60%  |
|  2 | 90 | 2.9 | 0.3 | 9.60% | 0.0 | 0.00% | 0.0 | 0.00% | 0.3 | 9.60%  |
|  3 | 90 | 8.4 | 0.5 | 5.70% | 0.1 | 1.00% | 0.1 | 1.10% | 0.5 | 5.90%  |
|  4 | 90 | 4.9 | 0.3 | 6.60% | 0.0 | 0.00% | 0.0 | 0.60% | 0.3 | 6.60%  |
|  5 | 90 | 12.1 | 0.9 | 7.40% | 0.1 | 0.80% | 0.0 | 0.00% | 0.9 | 7.40%  |
|  6 | 90 | 7.5 | 0.5 | 6.80% | 0.0 | 0.00% | 0.1 | 1.50% | 0.5 | 6.90%  |
|  7 | 90 | 15.0 | 1.1 | 7.30% | 0.3 | 1.90% | 0.3 | 1.90% | 1.2 | 7.80%  |

K242388 - Page 8 of 15

{8}

|  DFI (R10 Plus) (%) |   |   | Repeatability |   | Between-Run |   | Within-Day |   | Between-Day |   | Within-Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean | SD | CV % | SD | CV% | SD | CV% | SD | CV % | SD | CV%  |
|  1 | 90 | 7.92 | 0.6 | 7.90% | 0.0 | 0.00% | 0.6 | 7.90% | 0.0 | 0.00% | 0.6 | 7.90%  |
|  2 | 90 | 17.38 | 1.1 | 6.10% | 0.0 | 0.00% | 1.1 | 6.10% | 0.0 | 0.00% | 1.1 | 6.10%  |
|  3 | 90 | 24.35 | 0.7 | 3.00% | 0.2 | 0.90% | 0.8 | 3.10% | 0.0 | 0.00% | 0.8 | 3.10%  |
|  4 | 90 | 37.08 | 0.6 | 1.70% | 0.0 | 0.00% | 0.6 | 1.70% | 0.1 | 0.40% | 0.6 | 1.70%  |
|  5 | 90 | 46.90 | 0.7 | 1.50% | 0.3 | 0.70% | 0.8 | 1.70% | 0.1 | 0.20% | 0.8 | 1.70%  |
|  6 | 90 | 53.75 | 0.8 | 1.50% | 0.1 | 0.20% | 0.8 | 1.50% | 0.2 | 0.40% | 0.9 | 1.60%  |
|  7 | 90 | 67.8 | 0.7 | 1.10% | 0.1 | 0.20% | 0.7 | 1.10% | 0.1 | 0.20% | 0.7 | 1.10%  |
|  DFI (R11 Plus) (%) |   |   | Repeatability |   | Between-Run |   | Within-Day |   | Between-Day |   | Within-Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean | SD | CV % | SD | CV% | SD | CV% | SD | CV % | SD | CV%  |
|  1 | 90 | 5.74 | 0.4 | 7.60% | 0.1 | 2.10% | 0.5 | 7.90% | 0.0 | 0.00% | 0.5 | 7.90%  |
|  2 | 90 | 16.47 | 1.0 | 6.20% | 0.0 | 0.00% | 1.0 | 6.20% | 0.0 | 0.00% | 1.0 | 6.20%  |
|  3 | 90 | 22.06 | 0.6 | 2.50% | 0.2 | 1.10% | 0.6 | 2.70% | 0.0 | 0.00% | 0.6 | 2.70%  |
|  4 | 90 | 36.38 | 0.7 | 2.00% | 0.2 | 0.50% | 0.8 | 2.10% | 0.0 | 0.00% | 0.8 | 2.10%  |
|  5 | 90 | 45.28 | 0.7 | 1.50% | 0.0 | 0.00% | 0.7 | 1.50% | 0.0 | 0.00% | 0.7 | 1.50%  |
|  6 | 90 | 55.75 | 1.4 | 2.60% | 0.6 | 1.10% | 1.6 | 2.80% | 0.0 | 0.00% | 1.6 | 2.80%  |
|  7 | 90 | 67.40 | 1.1 | 1.60% | 0.0 | 0.00% | 1.1 | 1.60% | 0.1 | 0.20% | 1.1 | 1.70%  |

# Reproducibility

To evaluate the reproducibility performance of the LensHooke X12 PRO Semen Analysis System, a precision study was conducted using a total of three test slide lots and two analyzers at three clinical laboratory sites in the U.S. For sperm concentration, three levels of LensHooke X QC Beads control solution were evaluated. For the DNA fragmentation index (DFI) evaluation, three fresh native semen samples of various levels were utilized. Due to the limited stability of semen samples, for the DFI study each "day" in the statistical analysis represents different times of day (e.g. every one hour = one "day"). The studies included three operators  $\times$  six replicates (two devices and three slides lots)  $\times$  five days. For each of the three site, three lots of test slides and 2 devices were used. Each "replicate" is one well per slide. The sample test was performed on well 1 at RUN 1 and test on well 2 at RUN 2. Data analyses were performed for each site and overall to provide repeatability, between-day, between-laboratory, between-lot, and total reproducibility. All results met the predefined acceptance criteria.

|  Parameter [units] | N | Mean | Repeatability |   | Between-Day |   | Between Laboratory |   | Between Lot |   | Reproducibility  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  Concentration (10^6/mL) | 90 | 0.00 | 0.00 | N/A | 0.00 | N/A | 0.00 | N/A | 0.00 | N/A | 0.00 | N/A  |
|   |  90 | 26.41 | 0.91 | 3.4% | 0.05 | 0.2% | 0.00 | 0.00% | 0.07 | 0.3% | 0.91 | 3.4%  |
|   |  90 | 51.00 | 0.63 | 1.2% | 0.04 | 0.1% | 0.00 | 0.00% | 0.00 | 0.0% | 0.65 | 1.3%  |
|  R10 Plus DFI (%) | 90 | 15.92 | 1.08 | 6.80% | 0.19 | 1.20% | 0.36 | 2.30% | 0.00 | 0.00% | 1.17 | 7.40%  |
|   |  90 | 30.43 | 2.13 | 7.00% | 0.73 | 2.40% | 0.86 | 2.80% | 0.00 | 0.00% | 2.30 | 7.60%  |
|   |  90 | 68.45 | 1.30 | 1.90% | 0.14 | 0.20% | 0.72 | 1.00% | 0.00 | 0.00% | 1.56 | 2.30%  |
|  R11 Plus DFI (%) | 90 | 13.41 | 1.19 | 8.90% | 0.00 | 0.00% | 0.00 | 0.00% | 0.00 | 0.00% | 1.28 | 9.50%  |
|   |  90 | 37.21 | 2.51 | 6.80% | 0.81 | 2.20% | 1.76 | 4.70% | 0.00 | 0.00% | 3.13 | 8.40%  |
|   |  90 | 80.66 | 6.73 | 8.30% | 0.00 | 0.00% | 0.50 | 0.60% | 0.00 | 0.00% | 6.82 | 8.50%  |

K242388 - Page 9 of 15

{9}

Between-Instrument and Between-Operator Precision

Two concurrent studies were performed to evaluate the variability between instruments and the variability between operators independently. For the between-operator study, three operators were evaluated. For the between-instrument study, three instruments were evaluated. Both evaluations included three replicates for every 1 hour, five times/day, totaling 45 data points per level. Analysis was performed to calculate the overall repeatability  $\% \mathrm{CV}$ , between-instrument  $\% \mathrm{CV}$ , and between-operator  $\% \mathrm{CV}$ , which meet the pre-defined acceptance criteria.

|  Concentration (106/mL) | Operator Precision |   |   |   |   | Instrument Precision  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Mean | Between-Operator |   | Repeatability |   | Mean | Between-instrument |   | Repeatability  |   |
|  Sample | N |   | SD | CV% | SD | CV% |   | SD | CV% | SD | CV%  |
|  1 | 45 | 15.68 | 0.0 | 0.0% | 0.1 | 0.5% | 15.78 | 0.0 | 0.1% | 0.1 | 0.5%  |
|  2 | 45 | 89.39 | 0.8 | 0.8% | 1.5 | 1.7% | 52.68 | 0.0 | 0.0% | 0.3 | 0.5%  |
|  3 | 45 | 242.14 | 0.3 | 0.1% | 1.6 | 0.7% | 237.01 | 0.4 | 0.2% | 3.2 | 1.3%  |
|  4 | 45 | 84.03 | 0.0 | 0.0% | 1.1 | 1.3% | 75.65 | 0.1 | 0.1% | 0.3 | 0.3%  |
|  5 | 45 | 92.11 | 0.4 | 0.4% | 1.1 | 1.1% | 65.66 | 0.0 | 0.0% | 0.3 | 0.5%  |
|  6 | 45 | 90.18 | 0.0 | 0.0% | 0.6 | 0.7% | 80.43 | 0.0 | 0.0% | 0.5 | 0.6%  |
|  7 | 45 | 94.97 | 0.0 | 0.0% | 1.3 | 1.4% | 57.07 | 0.0 | 0.0% | 0.2 | 0.3%  |
|  Total Motility | Operator Precision |   |   |   |   | Instrument Precision  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Mean | Between-Operator |   | Repeatability |   | Mean | Between-instrument |   | Repeatability  |   |
|  Sample | N |   | SD | CV% | SD | CV% |   | SD | CV% | SD | CV%  |
|  1 | 45 | 66.1 | 0.0 | 0.0% | 0.9 | 1.4% | 60.4 | 0.0 | 0.0% | 1.2 | 2.0%  |
|  2 | 45 | 3.9 | 0.0 | 0.0% | 0.3 | 8.3% | 4.9 | 0.0 | 0.0% | 0.3 | 6.0%  |
|  3 | 45 | 82.4 | 0.2 | 0.3% | 1.2 | 1.5% | 80.9 | 0.1 | 0.1% | 1.4 | 1.7%  |
|  4 | 45 | 47.3 | 0.0 | 0.0% | 1.3 | 2.7% | 45.7 | 0.6 | 1.3% | 1.1 | 2.4%  |
|  5 | 45 | 72.5 | 0.4 | 0.5% | 1.2 | 1.6% | 76.9 | 0.0 | 0.0% | 1.5 | 2.0%  |
|  6 | 45 | 79.2 | 0.0 | 0.0% | 1.2 | 1.5% | 73.0 | 0.4 | 0.5% | 1.6 | 2.1%  |
|  7 | 45 | 80.7 | 0.0 | 0.0% | 1.5 | 1.8% | 79.0 | 0.0 | 0.0% | 1.3 | 1.6%  |
|  Progressive Motility | Operator Precision |   |   |   |   | Instrument Precision  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Mean | Between-Operator |   | Repeatability |   | Mean | Between-instrument |   | Repeatability  |   |
|  Sample | N |   | SD | CV% | SD | CV% |   | SD | CV% | SD | CV%  |
|  1 | 45 | 52.0 | 0.0 | 0.0% | 0.8 | 1.6% | 25.0 | 0.0 | 0.0% | 0.8 | 3.2%  |
|  2 | 45 | 0.0 | 0.0 | N/A | 0.0 | N/A | 1.0 | 0.0 | 0.0% | 0.0 | 0.0%  |
|  3 | 45 | 47.9 | 0.1 | 0.2% | 1.1 | 2.3% | 53.2 | 0.0 | 0.0% | 1.4 | 2.7%  |
|  4 | 45 | 26.1 | 0.0 | 0.0% | 0.8 | 3.3% | 21.6 | 0.0 | 0.0% | 1.1 | 5.0%  |
|  5 | 45 | 48.4 | 0.2 | 0.3% | 1.2 | 2.4% | 38.2 | 0.4 | 1.1% | 1.6 | 4.1%  |
|  6 | 45 | 50.4 | 0.0 | 0.0% | 1.8 | 3.5% | 34.6 | 0.0 | 0.0% | 1.1 | 3.2%  |
|  7 | 45 | 58.6 | 0.0 | 0.0% | 1.2 | 2.0% | 47.7 | 0.0 | 0.0% | 1.7 | 3.6%  |

K242388 - Page 10 of 15

{10}

|  Morphology | Operator Precision |   |   |   |   | Instrument Precision  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Mean | Between-Operator |   | Repeatability |   | Mean | Between-instrument |   | Repeatability  |   |
|  Sample | N |   | SD | CV% | SD | CV% |   | SD | CV% | SD | CV%  |
|  1 | 45 | 3.0 | 0.0 | 0.0% | 0.0 | 0.0% | 3.0 | 0.0 | 0.0% | 0.0 | 0.0%  |
|  2 | 45 | 2.0 | 0.0 | 0.0% | 0.0 | 0.0% | 2.0 | 0.0 | 0.0% | 0.0 | 0.0%  |
|  3 | 45 | 14.0 | 0.0 | 0.0% | 0.2 | 1.5% | 12.0 | 0.0 | 0.0% | 0.2 | 1.8%  |
|  4 | 45 | 8.1 | 0.0 | 0.2% | 0.3 | 3.1% | 6.0 | 0.0 | 0.0% | 0.0 | 0.0%  |
|  5 | 45 | 5.0 | 0.0 | 0.0% | 0.0 | 0.0% | 10.0 | 0.0 | 0.0% | 0.2 | 2.1%  |
|  6 | 45 | 7.0 | 0.0 | 0.0% | 0.2 | 3.0% | 7.0 | 0.0 | 0.0% | 0.0 | 0.0%  |
|  7 | 45 | 11.0 | 0.0 | 0.0% | 0.1 | 1.4% | 9.0 | 0.0 | 0.0% | 0.0 | 0.0%  |
|  DFI (R10 Plus) | Operator Precision |   |   |   |   | Instrument Precision  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Mean | Between-Operator |   | Repeatability |   | Mean | Between-instrument |   | Repeatability  |   |
|  Sample | N |   | SD | CV% | SD | CV% |   | SD | CV% | SD | CV%  |
|  1 | 45 | 18.07 | 0.0 | 0.0% | 0.5 | 3.0% | 16.15 | 0.0 | 0.0% | 0.3 | 2.1%  |
|  2 | 45 | 37.27 | 0.0 | 0.0% | 0.7 | 2.0% | 30.20 | 0.0 | 0.0% | 0.6 | 2.0%  |
|  3 | 45 | 61.27 | 0.4 | 0.6% | 1.6 | 2.7% | 54.78 | 0.0 | 0.0% | 0.7 | 1.4%  |
|  DFI (R11 Plus) | Operator Precision |   |   |   |   | Instrument Precision  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Mean | Between-Operator |   | Repeatability |   | Mean | Between-instrument |   | Repeatability  |   |
|  Sample | N |   | SD | CV% | SD | CV% |   | SD | CV% | SD | CV%  |
|  1 | 45 | 5.22 | 0.1 | 1.2% | 0.2 | 4.2% | 4.40 | 0.0 | 0.0% | 0.2 | 5.2%  |
|  2 | 45 | 37.22 | 0.1 | 0.1% | 0.4 | 1.0% | 24.51 | 0.0 | 0.0% | 0.3 | 1.0%  |
|  3 | 45 | 45.30 | 0.0 | 0.1% | 0.3 | 0.7% | 59.36 | 0.0 | 0.0% | 0.5 | 0.8%  |

2. Linearity:

Linearity for sperm concentration was evaluated using one analyzer and three slide lots. Semen samples were prepared at 11 sperm concentrations ranging from  $0 - 400 \times 10^{6} / \mathrm{mL}$ , verified using reference methods LensHooke X1 and manual microscope. The mean and SD of results were calculated. Regression analysis was used to verify the linear range. Sperm concentration was demonstrated to be linear from  $0.2 - 360 \times 10^{6} / \mathrm{mL}$ .

3. Analytical Specificity/Interference:

The potential interference of various substances on LensHooke X12 concentration results were evaluated by using three sperm concentration levels (0–20 M/mL, 50–100 M/mL, and 100–200 M/mL). The following 11 interfering substances were tested in the study: vitamin B, testosterone, yeast, E. Coli, RBC, WBC, urine, saliva, agglutination of semen sample, D-norgestrel, and  $\beta$ -estradiol. Samples were tested in five replicates on one analyzer using three lots of test slides. Results of the test group were compared to the control group. Study results showed that all tested interfering substances met the acceptance criteria, and no significant interference was caused by the tested substances.

A separate interference study was conducted to evaluate the interference on the DFI of the LensHooke X12 device. The potential interference of various substances on LensHooke X12 DFI results were evaluated by using two sperm DFI levels (R10 Plus:  $&lt; 20\%$  and  $\geq 20\%$ ; R11

K242388 - Page 11 of 15

{11}

Plus: &lt;14% and ≥14%). The following 20 interfering substances were tested in the study: WBC, pH 6.4, pH 7.8, pH 9.2, vitamin B12, testosterone, yeast, E. coli, RBC, urine, saliva, agglutination, D-norgestrel, B-estradiol, clomiphene citrate, follicle-stimulating hormone (FSH), coenzyme Q10, conjugated bilirubin, unconjugated bilirubin and yellow pigment (Sunset Yellow FCF, Merck Cat# 465224). Samples were tested in five replicates on one analyzer using three lots of test reagents. Results of the test group were compared to the control group. Study results showed that all tested interfering substances met the acceptance criteria, and no significant interference was caused by the tested substances.

4. Assay Reportable Range:

Sperm concentration: 0.4–300 M/mL

Total Motility: 0–100%

Progressive Motility: 0–100%

Morphology: 0–100%

Sperm DNA fragmentation index (DFI): 0–100%.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

## Sample Stability

Stability of semen samples was determined with three samples at different concentration levels. The semen samples were stored at room temperature. At time zero (0 hour), the samples were tested with nine replicates to establish the baseline. These samples were tested again (n=9) at various timepoints and for each time point, the results were compared to the baseline results, and the percent difference was calculated to meet the predefined acceptance criteria. The results support a 1-hour semen sample stability duration for traditional semen parameters (CS3) analysis and an 8-hour semen sample stability duration for DFI analysis.

## Reagent Stability

A stability study was conducted to determine the shelf-life of the LensHooke Semen Test Slide in real-time (at 2°C and 30°C). Three semen samples with varying sperm concentrations were tested in three replicates using one analyzer and three lots of test slides. All results were compared against the reference method LensHooke X1 PRO Analyzer. Results support an 18-month shelf-life claim at refrigerated and room temperature conditions (2°C and 30°C).

6. Detection Limit:

A study was conducted to determine the detection limits of the LensHooke X12 PRO Semen Analysis System. Semen samples were taken from volunteers and two interval ranges were prepared: sperm-free seminal plasma and low concentration seminal plasma. Semen was centrifuged to obtain sperm-free seminal plasma to a concentration of ~0 M/mL (blank sample) as verified by manual microscope. Low concentration seminal plasma was prepared by diluting semen to a concentration of ~0.4 M/mL as verified by manual microscope. Blank and low-level samples were divided into five aliquots and tested in three replicates (low-level sample) or four replicates (blank sample) once a day for three days, using two lots of test

K242388 - Page 12 of 15

{12}

cassettes and one analyzer. Results were calculated, and the detection limits were determined to be:

Limit of Blank (LoB) = 0 M/mL

Limit of Detection (LoD) = 0.07 M/mL

Limit of Quantitation  $(\mathrm{LoQ}) = 0.37\mathrm{M / mL}$

7. Assay Cut-Off:

See Reference Range study, section VII.E.

8. Accuracy (Instrument):

See method comparison, section VII.B.1.

9. Carry-Over:

Not applicable.

# B Comparison Studies:

1. Method Comparison with Predicate Device:

A method comparison study was conducted to evaluate the performance of the LensHooke X12 PRO Semen Analysis System when used by intended users following the instructions in the package insert. The study was performed at three US sites, with a total of ten operators for semen analysis and sperm DNA fragmentation analysis. Following collection of the specimen, the user analyzed the sample on the LensHooke X12 PRO Semen Analysis System. After recording the results, a different trained technologist at the study site performed semen analysis on the sample using either the predicate (i.e., the LensHooke X1 PRO Semen Quality Analyzer for concentration, motility and morphology) or the reference method (i.e., manual microscope for DNA fragmentation analysis as recommended in the WHO 6th edition). A total of 130 semen specimens spanning the analytical measuring range were analyzed by using the Passing-Bablok regression. All results met the predefined acceptance criteria.

|  Parameter [Units] | N | Result Range | Slope (95% CI) | Intercept (95% CI) | R²  |
| --- | --- | --- | --- | --- | --- |
|  Concentration [10^6/mL] | 130 | 0.0–264.1 | 0.96 (0.93, 1.00) | -0.22 (-1.16, 0.67) | 0.98  |
|  Total Motility [%] | 130 | 0.0–92.0 | 1.00 (0.98, 1.04) | -1.00 (-2.27, 0.10) | 0.98  |
|  Progressive Motility [%] | 130 | 0.0–88.0 | 1.00 (0.96, 1.28) | 0.50 (-0.15, 1.90) | 0.97  |
|  Morphology [%] | 130 | 0.0–13.0 | 1.00 (1.00, 1.00) | 0.00 (0.00, 0.00) | 0.95  |
|  R10 Plus DFI [%] | 130 | 0.0–55.18 | 1.01 (0.98, 1.04) | -0.40 (-0.84, 0.10) | 0.99  |
|  R11 Plus DFI [%] | 130 | 0.0–55.36 | 1.00 (0.96, 1.04) | -0.04 (-0.46, 0.51) | 0.97  |

2. Matrix Comparison:

The matrix comparison study was conducted to assess the comparison of fresh neat semen samples and diluted semen samples on the LensHooke X12 PRO. For the DFI evaluations, a

K242388 - Page 13 of 15

{13}

total of three semen samples were included in the study where paired neat semen samples were diluted with 0.01M PBS (control) or a sperm extender HTF (Human Tubal Fluid Medium) at 5x, 10x, 15x, 30x, and 60x dilution. For the traditional parameters' evaluation, a total of three semen samples were included in the study where paired neat semen samples were diluted with seminal plasma (control), or a sperm extender (SpermRinse, LifeGlobal HTF Total) at 2x and 4x dilution. Samples included low, near cutoff, and normal concentration levels. The results were compared to the control results, and the percent bias was calculated and met the predefined acceptance criteria. The study results support the use of 0.01M PBS or sperm extenders (HTF) as diluents during analysis on LensHooke X12 PRO (CS3, R10 Plus, R11 Plus), as described above.

## C Expected Values/Reference Range:

A reference interval study was performed to determine the reference intervals for sperm DNA fragmentation index using the R10 Plus and R11 Plus kits on the LensHooke X12 PRO device. Semen samples from 120 healthy male adult subjects were evaluated at five clinical sites for the study (1 OUS, 4 US). The reference intervals were established with the non-parametric bootstrap method following CLSI EP28-A3c, by computing the lower reference limit (2.5th percentile) and the upper reference limit (97.5th percentile). The 95% reference interval for the SCD (R10 Plus) method was 4.16 to 35.07%, and for the SDFR (R11 Plus) method was 1.90 to 28.57%.

Based on the 90% confidence intervals (CI) of the upper reference limits from the study data, the low reference DNA Fragmentation Index (DFI) cut-off is 32.10% (R10 Plus) and 24.07% (R11 Plus).

For traditional semen parameters (CS3), the lower limit of the reference range for parameters Sperm Concentration, Total Motility (%), Progressive Motility (%), and Morphology (%) were directly reported as the 5th percentile in the WHO Laboratory Manual for the Examination and Processing of Human Semen, 6th Edition (Table 8.3).

## D Other Supportive Instrument Performance Characteristics Data:

### 1. Capture Mode Evaluation:

An evaluation was conducted to assess the comparison of different capture modes on the LensHooke X12 PRO, to demonstrate the equivalence of Auto-modes to the Custom Mode. Semen samples were included in the study at various levels including above, below, and near the cut-off, including four samples for DFI, six samples for concentration, five samples for motility, and three samples for morphology evaluations. Results from Auto-mode were compared to eight Custom Mode designs, intended to represent the worst-case scenarios. The percent bias was calculated by comparing the results obtained from the Auto-modes to different Custom Modes. The results demonstrate that custom mode configurations produce equivalent results compared to auto-mode, validating use of custom mode on LensHooke X12 PRO (CS3, R10 Plus, and R11 Plus).

K242388 - Page 14 of 15

{14}

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K242388 - Page 15 of 15

---

**Source:** [https://fda.innolitics.com/device/K242388](https://fda.innolitics.com/device/K242388)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
