The Bluemoon is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photopolymerized in the 385~515nm waveband of visible light.
Device Story
Bluemoon is a cordless, pen-style dental curing light used by clinicians in hospital or clinical settings. It utilizes an LED light source to emit visible light in the 385-515nm wavelength range to photopolymerize dental materials. The device features four user-selectable operational modes (HYPER, TURBO, LOW, SOFT START) that vary in output intensity and irradiation time. The clinician operates the device by selecting a mode and directing the light probe toward the dental material. The device includes an internal rechargeable lithium-ion battery, a charger, and an anti-glare shield. By providing controlled light energy, the device ensures proper polymerization of restorative materials, which is essential for the structural integrity and clinical success of dental restorations and orthodontic treatments.
Clinical Evidence
Bench testing only. Performance evaluations included irradiation time, peak wavelength verification, polymerization strength (≥ 80MPa), depth of cure (≥ 2.0mm), light probe temperature (≤ 41°C), and power management features. Biocompatibility testing (cytotoxicity, skin sensitization, intracutaneous reactivity) was conducted per ISO 10993-1. All results met established specifications.
Technological Characteristics
Cordless pen-style dental curing light. LED light source; 385-515nm wavelength. Powered by 3.7V internal Li-ion battery. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 62133-2 (battery). Non-sterile. Biocompatibility per ISO 10993-1, -5, -10, -23.
Indications for Use
Indicated for polymerization of resinous dental materials, restorative composites, orthodontic brackets, and bonding/sealing materials requiring 385-515nm visible light.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
GENOSS Co., Ltd.
Jiyeon Lee
Regulatory affair
12F, 76, Changnyong-daero 256beon-gil, Yeongtong-gu,
Suwon-si, Gyeonggi-do, Republic of Korea
Re: K242386
Trade/Device Name: Bluemoon
Regulation Number: 21 CFR 872.6070
Regulation Name: Ultraviolet activator for polymerization
Regulatory Class: Class II
Product Code: EBZ
Dear Jiyeon Lee:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 5, 2025. Specifically, FDA is updating this SE Letter to correct the official correspondent as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Michael Adjodha, OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices, 301-796-6276, Michael.Adjodha@fda.hhs.gov.
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 5, 2025
GENOSS Co., Ltd.
Jiyeon Lee
Regulatory affair
12F, 76, Changnyong-daero 256beon-gil, Yeongtong-gu
Manufacturing site: 3F, 4F, 5F, D-factory, 56, Changnyong-da
Suwon-si, Gyeonggi-do
China
Re: K242386
Trade/Device Name: Bluemoon
Regulation Number: 21 CFR 872.6070
Regulation Name: Ultraviolet Activator For Polymerization
Regulatory Class: Class II
Product Code: EBZ
Dated: September 8, 2024
Received: April 29, 2025
Dear Jiyeon Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K242386 - Jiyeon Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K242386 - Jiyeon Lee
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Bobak
Shirmohammadi - S
For Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242386 | |
| Device Name Bluemoon | |
| Indications for Use (Describe) The Bluemoon is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photopolymerized in the 385~515nm waveband of visible light. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
GENOSS
510(k) Summary
K242386
04/28/2025
# 1. Company
| | Submitter |
| --- | --- |
| Name | GENOSS Co., Ltd. |
| Address | Head Office: 12F, 76, Changnyong-daero 256beon-gil, Yeongtong-gu, Suwon-si, Gyeonggi-do, Republic of Korea
Manufacturing site: D-Factory, 56, Changnyong-daero 256beon-gil, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea |
| Phone/Fax | |
| Contact person | +82-70-7098-6923/ +82-31-888-5595 |
| Summary Date | Jiyeon Lee
jylee3@genoss.com
04/28/2025 |
# 2. Device Name
Trade name: Bluemoon
Common name: Dental Curing Light Device
Classification name: Ultraviolet activator for polymerization, 21 CFR 872.6070, product code EBZ, Device Class II
# 3. Predicate Device
K200809
D-Lux+ (Primary predicate)
# 4. Description
The Bluemoon Dental Curing Light is an instrument for photopolymerizing dental materials and starting agents by irradiating them with visible light between 385-515nm wavelength. It offers four modes (HYPER, TURBO, LOW, and SOFT START) depending on the output intensity and irradiation time.
# 5. Indication for use
The Bluemoon is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photopolymerized in the 385~515nm wavelength of visible light.
510(k) Summary
{6}
GENOSS
# 6. Comparison Technological Characteristics with the Predicate Devices
Bluemoon was compared with the predicate device ‘D-Lux+’. The characteristics that differed from the predicate device were performed by gap analysis, which confirmed equivalence with the predicate device. Technological characteristics of Bluemoon and D-Lux+ are as following;
| Item | Proposed Device | Predicate Device K200809 | Remark |
| --- | --- | --- | --- |
| Device name | Bluemoon | D-Lux+ | - |
| Classification Regulation | 21 CFR 872.6070 | 21 CRF 872.6070 | Same |
| Classification | II | II | Same |
| Product Code | EBZ | EBZ | Same |
| Common Name | Activator, ultraviolet for polymerization | Activator, ultraviolet for polymerization | Same |
| Indications for use | The Bluemoon is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photopolymerized in the 385~515nm wavelength of visible light. | The D-Lux+ is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 385~515nm waveband of visible light | Same |
| Use | Prescription / Hospital | Prescription / Hospital | Same |
| Delivery form content | - Light probe
- Main body
- Battery pack
- Charger
- Anti-glare shield
- USB C to C cable
- Disposable vinyl cover
- User Manual | -D-Lux+ Handpiece
-C-Battery
-D-Lux+ Charger
-Adapter
-Power Cord
-Disposable sheath
-Light Probe
-Light Protector
-Instruction Manual | Gap (1) |
| Wavelength range | 385nm - 515nm | 385nm - 515nm | Same |
| Peak wavelength | Dual peak: 410nm, 460nm | Dual peak: 405nm, 460nm | Gap (2) |
| Operational modes | 4 programs:
-Hyper: 3s(3000mW/cm²)
-Turbo: 5s(2500mW/cm²)
-Low: 10s(1200mW/cm²)
-Soft Start: 10s(0~2000mW/cm²) | 5 modes
-STD (Standard) mode: 900mW/cm²
5, 10, 15 and 20 seconds
-SFT (Soft Start) mode: 1300mW/cm²
10, 15 seconds
-HIG (High Power) mode: 1300mW/cm²
5, 10 and 15 seconds
-ORT (Orthodontic) mode: 1800mW/cm²
3, 4, 5 seconds
-MAX (Max Power) mode: 2400mW/cm²
1, 2, 3 seconds | Gap (3) |
| Power Source | 3.7V DC with Lithium ion battery
5V DC with charger power | 3.6V DC with Lithium ion battery
6V DC with charger power | Gap (4) |
510(k) Summary
{7}
GENOSS
| Light source | LED light | LED light | Same |
| --- | --- | --- | --- |
| Electrical | IEC 60601-1 | IEC 60601-1 | Same |
| Safety | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Biocompatibility | ISO 10993-1
ISO 10993-5
ISO 10993-10
ISO 10993-23 | Direct contact with issue in not intended. Therefore ISO 10993-1 is not applicable. | Gap (5) |
## 6.1 Gap Analysis
Gap (1): The Main body (has a built-in battery), charger, Light probe, Anti-glare shield, adapter, Disposal vinyl cover of proposed device have the same functions with the D-Lux+ Handpiece, D-Lux+ Charger, Light Probe, Light Protector, C-Battery (Included in the handpiece), Adapter, Power Cord and Disposable Sheaths of the D-Lux+. Both devices adopt the cordless pen-style.
Gap (2): The peak wavelength values of the predecessor and application devices are very similar. The predicate device is 405, 460 nm and the proposed device is 410, 460 nm. Because the dual peak wavelengths of the two devices are so similar, they do not affect the safety or efficacy of the product.
Gap (3): The proposed device has several modes corresponding to the light output intensity and the available time. The light output safety and performance tests were performed according to IEC 60601-1, IEC 60601-1-2 and FDA guidance performance test requirements. The test results show that these differences do not affect safety and effectiveness. Therefore, it is determined that these mode differences do not cause substantial equivalence issues.
Gap (4): Both devices use the internal rechargeable lithium ion battery. Though the battery technical specification of proposed device is a bit different from that of the predicate device, the battery of proposed device meets the IEC 62133-2 standard. This difference does not affect safety and effectiveness.
Gap (5): Biocompatibility testing of the subject device was conducted because although the device does not come in contact with oral tissue on the chance that contact does occur. However, the predicate device is not evaluated. The testing results show that these differences do not affect safety and effectiveness.
510(k) Summary
{8}
GENOSS
# 7. Biocompatibility Data
Biocompatibility testing was conducted because although the device does not come in contact with oral tissue on the chance that contact does occur. The biocompatibility evaluation for the Bluemoon was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:
- Cytotoxicity
- Skin Sensitization
- Intracutaneous reactivity test report
# 8. Performance Data
Bluemoon was evaluated using the following performance bench testing to confirm the performance characteristics:
| No. | Title | Properties |
| --- | --- | --- |
| 1 | Irradiation time | 1) SOFT START: 10±1s
2) LOW: 10±1s
3) TURBO: 5±1s
4) HYPER: 3±1s |
| 2 | Peak wavelength | Check that the wavelength maxima occur within 5% of 400 nm and 455 nm respectively.
1) Purple: 400nm±5%
2) Blue: 455nm±5% |
| 3 | Polymerization strength measurement | ≥ 80MPa |
| 4 | Depth of cure | ≥ 2.0mm |
| 5 | Light Probe temperature | ≤ 41°C |
| 6 | Energy-saving mode | 60±2s |
| 7 | Turn off energy-saving mode | 1) Buzzer sounds once,
2) LCD on |
| 8 | Auto power off | 10 minutes ± 2 seconds |
All test results demonstrate that the materials chosen, the manufacturing process, and the design
510(k) Summary
{9}
GENOSS
utilized for the Bluemoon met the established specifications necessary for consistent performance according to its intended use.
## 9. Conclusion
Performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.
510(k) Summary
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.