The Ulspira TS Nitric Oxide Therapy System is intended for use by healthcare professionals for the delivery of nitric oxide (NO) and the monitoring of inspired NO. NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy. The Ulspira TS must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use. The Ulspira TS primary NO therapy system delivers NO gas in the 0-80 ppm range while in the Constant Rate or flow sensing modes. This includes: · Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system. · A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb. • Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO cylinders, and a low O2 pressure alarm when using an oxygen cylinder. · An automated emergency dosing algorithm activated by certain high-risk alarms, which impact patient dosing, to avoid sudden cessation of therapy. · Compatibility with a wide inspiratory flow rate range of 0.25-120 //min, utilizing an automatically detected low or high flow sensor. · An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access. The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment during transit within hospitals.
Device Story
System delivers physician-prescribed nitric oxide (NO) gas into respiratory device circuits; monitors inspired NO, NO2, and O2 concentrations. Used in NICU and during neonatal transport by healthcare professionals. Inputs include gas supply from cylinders and inspiratory flow data from connected ventilators. Operates via primary delivery system (Constant Rate or flow-sensing modes) or integrated pneumatic backup system for manual ventilation. Features touch-screen interface with waveform display, automated cylinder switching, and emergency dosing algorithm to prevent therapy cessation during high-risk alarms. Output includes controlled NO delivery and real-time gas monitoring; visual/audible alarms alert clinicians to parameter deviations. Benefits include continuous, accurate iNO therapy delivery and monitoring, ensuring patient safety during respiratory failure treatment.
Clinical Evidence
Bench testing only. Evaluated NO delivery accuracy, O2 dilution, NO2 formation, alarm functionality, and influence on ventilator performance across seven different ventilators at 0, 1, 5, 20, 40, and 80 ppm settings. Results confirmed performance within specifications.
Technological Characteristics
Nitric oxide administration apparatus; 0-80 ppm delivery range; 0.25-120 L/min flow compatibility. Integrated gas analyzers for NO (0-120 ppm), NO2 (0-30 ppm), and O2 (18-100%). Features touch-screen UI, internal battery (2h), 12V DC inlet, and pneumatic backup system. Connectivity via gas connectors (CGA 626). Software v1.6.
Indications for Use
Indicated for term and near-term (>34 weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
{0}------------------------------------------------
January 17, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Airgas Therapeutics LLC % Mikel Alberdi President Drug Device Consulting 13014 North Dale Mabry Highway, #326 Tampa, Florida 33618
Re: K242374
Trade/Device Name: ULSPIRA TSTM Nitric Oxide Therapy System Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: Class II Product Code: MRN, MRO, MRP, MRQ, CCL Dated: December 20, 2024 Received: December 20, 2024
Dear Mikel Alberdi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242374
Device Name
ULSPIRA TS™ Nitric Oxide Therapy System
## Indications for Use (Describe)
The Ulspira TS Nitric Oxide Therapy System is intended for use by healthcare professionals for the delivery of nitric oxide (NO) and the monitoring of inspired NO. NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy.
The Ulspira TS must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use.
The Ulspira TS primary NO therapy system delivers NO gas in the 0-80 ppm range while in the Constant Rate or flow sensing modes. This includes:
· Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system.
· A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb.
• Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO cylinders, and a low O2 pressure alarm when using an oxygen cylinder.
· An automated emergency dosing algorithm activated by certain high-risk alarms, which impact patient dosing, to avoid sudden cessation of therapy.
· Compatibility with a wide inspiratory flow rate range of 0.25-120 //min, utilizing an automatically detected low or high flow sensor.
· An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access.
The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment during transit within hospitals.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
#### 510(k) SUMMARY
The following summary is provided in accordance with 21 CFR 807.92:
#### A. DATE
January 17, 2025
#### B. APPLICANT
Airgas Therapeutics 6141 Easton Rd. P.O. Box 310 Plumsteadville PA 18949 United States
- Mr. Timothy Warren Contact Person: Airgas Therapeutics Senior Director of Quality and Regulatory Affairs (215) 766-7407 x3034 tim.warren@airgas.com
### C. CORRESPONDENT
Drug Device Consulting 13014 North Dale Mabry Highway, #326 Tampa FL 33618 United States
Contact Person: Mr. Mikel Alberdi Drug Device Consulting (813) 708-8303 malberdi@drugdeviceconsulting.com
### D. DEVICE
| Device Trade Name: | UISPIRA TS™ Nitric Oxide Therapy System |
|----------------------|-----------------------------------------|
| Common Name: | Nitric Oxide Administration Apparatus |
| Classification: | Class II - 21 CFR 868.5165 |
| Classification Name: | Apparatus, Nitric Oxide Delivery |
| Product Codes: | MRN (Primary), MRO, MRP, MRQ, CCL |
### E. PREDICATE DEVICE
K212409 - Ulspira TS Nitric Oxide Therapy System
{5}------------------------------------------------
#### ﺖ DEVICE DESCRIPTION
Ulspira TS Nitric Oxide Therapy System delivers physician-prescribed NO therapy gas and monitors inspired Nitric Oxide, Nitrogen Dioxide, and Oxygen gas in combination with a respiratory device.
The main device functionalities of the Ulspira TS Nitric Oxide Therapy System include:
- A primary delivery system to administer NO gas into a respiratory device circuit.
- Monitoring of NO, NO2, and O2 gas concentrations close to the patient interface.
- System includes a user interface that contains all controls used to set the NO delivery and monitoring parameters. All set parameters as well as other information are shown on the user interface screen.
- The system will produce visual and audible alarms if vital parameters vary beyond preset or default limits.
- The system includes an integrated pneumatic back-up system for manual hand bagging in order to deliver NO therapy in the event of a failure of the primary delivery system and during manual ventilation.
The Ulspira TS system consists of the base unit, the mobile cart and bedside rail holder, and various components and accessories, including the gas regulators and patient kits for use with validated respiratory devices.
### G. INDICATIONS FOR USE
The Ulspira TS Nitric Oxide Therapy System is intended for use by healthcare professionals for the delivery of nitric oxide (NO) and the monitoring of inspired NO, NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy.
The Ulspira TS must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use.
The Ulspira TS primary NO therapy system delivers NO gas in the 0-80 ppm range while in the Constant Rate or flow sensing modes. This includes:
• Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system.
• A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb.
• Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO cylinders, and a low O2 pressure alarm when using an oxygen cylinder.
{6}------------------------------------------------
- An automated emergency dosing algorithm activated by certain high-risk alarms, which impact patient dosing, to avoid sudden cessation of therapy.
• Compatibility with a wide inspiratory flow rate range of 0.25-120 //min, utilizing an automatically detected low or high flow sensor.
• An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access.
The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment during transit within hospitals.
## H. TECHNOLOGICAL COMPARISON
Ulspira TS Nitric Oxide Therapy System modifications include a modification of the labeling to include six additional respiratory devices. The Ulspira TS Nitric Oxide Therapy System with the additional respiratory devices has the identical intended use and uses the same device hardware as the cleared Ulspira TS Nitric Oxide Therapy System predicate device. All features are identical except for those described in the table below.
| Device name | Ulspira TS<br>(K212409) | Ulspira TS<br>(Proposed Device) | Comparison |
|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Comparison of general attributes, indication, patient population, operating environment, etc. | | | |
| Product code(s) | MRN | MRN | Identical |
| Regulation number | 21 CFR 868.5165 | 21 CFR 868.5165 | Identical |
| Regulation Description | - Nitric oxide administration apparatus (Nitric<br>oxide delivery/backup delivery)<br>- Nitric oxide analyzer<br>- Nitrogen dioxide analyzer | - Nitric oxide administration apparatus (Nitric<br>oxide delivery/backup delivery)<br>- Nitric oxide analyzer<br>- Nitrogen dioxide analyzer | Identical |
| Indications for Use | The Ulspira TS Nitric Oxide Therapy System is<br>intended for use by healthcare professionals<br>for the delivery of nitric oxide (NO) and the<br>monitoring of inspired NO, NO2 and O2<br>concentrations for a patient undergoing<br>inhaled Nitric Oxide (iNO) therapy and must<br>only be used in accordance with the<br>indications, contraindications, warnings and<br>precautions described in the nitric oxide<br>drug packaging inserts and labeling<br>(currently neonates). Refer to this material<br>prior to use.<br>The Ulspira TS primary NO therapy system<br>delivers NO gas in the 0-80 ppm range while<br>in the Constant Rate or flow sensing modes.<br>This includes:<br>• Continuous integrated monitoring for<br>inspired NO, NO2 and O2 and a<br>comprehensive alarm system.<br>• A touch-screen user interface with a<br>waveform display of the ventilation device<br>flow as measured in the inspiratory limb.<br>• Cylinder handling facilitated by manual or<br>an automatic cylinder switch which is<br>reactive to the detected gas supply state of<br>NO cylinders, and a low O2 pressure alarm<br>when using an oxygen cylinder. | The Ulspira TS Nitric Oxide Therapy System is<br>intended for use by healthcare professionals<br>for the delivery of nitric oxide (NO) and the<br>monitoring of inspired NO, NO2 and O2<br>concentrations for a patient undergoing<br>inhaled Nitric Oxide (iNO) therapy and must<br>only be used in accordance with the<br>indications, contraindications, warnings and<br>precautions described in the nitric oxide<br>drug packaging inserts and labeling<br>(currently neonates). Refer to this material<br>prior to use.<br>The Ulspira TS primary NO therapy system<br>delivers NO gas in the 0-80 ppm range while<br>in the Constant Rate or flow sensing modes.<br>This includes:<br>• Continuous integrated monitoring for<br>inspired NO, NO2 and O2 and a<br>comprehensive alarm system.<br>• A touch-screen user interface with a<br>waveform display of the ventilation device<br>flow as measured in the inspiratory limb.<br>• Cylinder handling facilitated by manual or<br>an automatic cylinder switch which is<br>reactive to the detected gas supply state of<br>NO cylinders, and a low O2 pressure alarm<br>when using an oxygen cylinder. | Identical |
{7}------------------------------------------------
# K242374 – 510(k) SUMMARY
| Device name | Ulspira TS<br>(K212409) | Ulspira TS<br>(Proposed Device) | Comparison | | | | |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-------------------------------------------------------------------|-----------------------------|-----------|--|
| | • An automated emergency dosing algorithm<br>activated by certain high-risk alarms, which<br>impact patient dosing, to avoid sudden<br>cessation of therapy.<br>• Compatibility with a wide inspiratory flow<br>rate range of 0.25-120 l/min, utilizing an<br>automatically detected low or high flow<br>sensor.<br>• An internal battery which provides at least<br>two hours of uninterrupted therapy and a<br>12V DC inlet for additional external battery<br>access.<br>The integrated Ulspira TS pneumatic backup<br>NO therapy system provides backup NO<br>delivery capability that is intended to deliver<br>a continuous flow of NO mixed with O2, for<br>iNO therapy which allows continuous<br>treatment during transit within hospitals. | • An automated emergency dosing algorithm<br>activated by certain high-risk alarms, which<br>impact patient dosing, to avoid sudden<br>cessation of therapy.<br>• Compatibility with a wide inspiratory flow<br>rate range of 0.25-120 l/min, utilizing an<br>automatically detected low or high flow<br>sensor.<br>• An internal battery which provides at least<br>two hours of uninterrupted therapy and a<br>12V DC inlet for additional external battery<br>access.<br>The integrated Ulspira TS pneumatic backup<br>NO therapy system provides backup NO<br>delivery capability that is intended to deliver<br>a continuous flow of NO mixed with O2, for<br>iNO therapy which allows continuous<br>treatment during transit within hospitals. | | | | | |
| Physical dimensions and<br>weight (excl. carrier/cart) | Weight: 7.0 kg<br>Width: 320 mm<br>Depth: 150 mm<br>Height: 300 mm | Weight: 7.0 kg<br>Width: 320 mm<br>Depth: 150 mm<br>Height: 300 mm | Identical | | | | |
| NO gas connectors | CGA 626 | CGA 626 | Identical | | | | |
| NO Injection location | NO injection connected between ventilator<br>and humidifier. | NO injection connected between ventilator<br>and humidifier. | Identical | | | | |
| Battery backup | Yes | Yes | Identical | | | | |
| Battery backup time | 2h | 2h | Identical | | | | |
| Power Supply - Main | Voltage: 100-240V, 50 - 60Hz | Voltage: 100-240V, 50 - 60Hz | Identical | | | | |
| Automated Pre-Use<br>check | Yes | Yes | Identical | | | | |
| | Alarms | | | | | | |
| NO Delivery/Flow Sensor<br>Alarms | Yes | Yes | Identical | | | | |
| Power Supply/Battery<br>Alarms | Yes | Yes | Identical | | | | |
| NO, NO2, O2 Monitoring<br>Alarms | Yes | Yes | Identical | | | | |
| | Primary NO administration system | | | | | | |
| NO administration<br>principle | NO delivery into the inspiratory limb of a<br>ventilation device's patient circuit. | NO delivery into the inspiratory limb of a<br>ventilation device's patient circuit. | Identical | | | | |
| Range of NO gas<br>concentration delivered | 0-80 ppm | 0-80 ppm | Identical | | | | |
| NO delivery accuracy | ±20% or 2 ppm, whichever is the greatest. | ±20% or 2 ppm, whichever is the greatest. | Identical | | | | |
| Operating Modes | The subject device incorporates two modes<br>(Autosense and Jet Sense) that provides a<br>user set dose of NO into the inspiratory limb<br>of the respiratory device circuit, based on<br>the measured respiratory device flow.<br>The subject device also includes a third<br>mode, Constant Rate, which provides flow<br>of NO (0.5 - 60 l/min) into the inspiratory<br>limb of the respiratory device circuit. | The subject device incorporates two modes<br>(Autosense and Jet Sense) that provides a<br>user set dose of NO into the inspiratory limb<br>of the respiratory device circuit, based on<br>the measured respiratory device flow.<br>The subject device also includes a third<br>mode, Constant Rate, which provides flow<br>of NO (0.5 - 60 l/min) into the inspiratory<br>limb of the respiratory device circuit. | Identical | | | | |
| Backup NO administration system | | | | | | | |
| Backup power source | Pneumatic system | Pneumatic system | Identical | | | | |
| Backup NO<br>Administration | Backup system delivering a user set<br>concentration of NO, via adjustment of O2<br>flow, to the patient via a resuscitator. | Backup system delivering a user set<br>concentration of NO, via adjustment of O2<br>flow, to the patient via a resuscitator. | Identical | | | | |
| Backup NO delivery<br>accuracy | accuracy | Within ±20% of set value or ±2 ppm,<br>whichever is the greatest. | Which ever is the greatest. | Within ±20% of set value or ±2 ppm,<br>whichever is the greatest. | Which ever is the greatest. | Identical | |
{8}------------------------------------------------
| Device name | Ulspira TS<br>(K212409) | Ulspira TS<br>(Proposed Device) | Comparison |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Gas analysis (NO, NO2, O2) - General characteristics | | | |
| Breathing circuit sample<br>source location | On the inspiratory limb of the breathing<br>circuit, after the humidifier. | On the inspiratory limb of the breathing<br>circuit, after the humidifier. | Identical |
| Sample flow rate | 150 ml/min | 150 ml/min | Identical |
| NO gas analysis | | | |
| Integrated NO Gas<br>Analyzer | Yes | Yes | Identical |
| NO measurement<br>accuracy | +- (0.5 ppm +20 % of actual concentration) in<br>the range 0-20 ppm<br>+- (0.5 ppm +10 % of actual concentration) in<br>the range 20-120 ppm | +- (0.5 ppm +20 % of actual concentration) in<br>the range 0-20 ppm<br>+- (0.5 ppm +10 % of actual concentration) in<br>the range 20-120 ppm | Identical |
| NO Measurement range | 0 - 120 ppm | 0 - 120 ppm | Identical |
| NO2 gas analysis | | | |
| Integrated NO2 Gas<br>Analyzer | Yes | Yes | Identical |
| NO2 measurement<br>accuracy | ±(20% or 0.5 ppm), whichever is the<br>greatest. | ±(20% or 0.5 ppm), whichever is the<br>greatest. | Identical |
| NO2 measurement range | 0 - 30 ppm | 0 - 30 ppm | Identical |
| O2 gas analysis | | | |
| Integrated O2 Gas<br>Analyzer | Yes | Yes | Identical |
| O2 measurement<br>accuracy | ±(2.5 % volume fraction + 2.5 % of gas<br>concentration) | ±(2.5 % volume fraction + 2.5 % of gas<br>concentration) | Identical |
| O2 measurement range | 18 - 100 % | 18 - 100 % | Identical…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.