The sponsor used data from a PMCF study (including both prospective and retrospective patient data) to demonstrate the safety and effectiveness of the Symani Surgical System for microsurgical anastomoses in mouth and scalp free-flap procedures.
Patients undergoing free tissue transfer procedures with mouth or scalp as recipient sites; Sample Size: 62 patients (55 prospective, 7 retrospective); Number of Sites: Multicenter (specific count not provided)
Not applicable for this study
Intraoperative patency at first attempt, operative/suturing time, rate of intra-operative change to manual suturing, freedom from Serious Adverse Device Effects (SADE), 3-day reoperation rate
Indications for Use
The Symani® Surgical System is intended for soft tissue manipulation to perform anastomosis, suturing, and ligation microsurgery techniques on small blood vessels and lymphatic ducts between 0.1 and 2.5 mm in open free-flap surgery of the breast, mouth, scalp, extremities, and in open lymphatic surgery of the extremities. The Symani® Surgical System is indicated for use during microsurgical procedures when use of a motion scaling function is deemed appropriate by the surgeon. The System is indicated for use in adults. It is intended to be used by trained physicians in an appropriate operating environment in accordance with the Instructions for Use.
Device Story
Symani Surgical System is an electromechanical robotic-assisted surgical device (RASD) for microsurgery. System comprises three main components: CMM (Cartesian Motion Manipulator), Console, and Instruments. Surgeon operates from console, controlling wristed end-effectors that provide increased dexterity for tissue manipulation and suturing. System features selectable motion scaling (7x to 20x) to filter physiological tremor and enhance precision. Used in open surgical environments by trained physicians. Output is real-time robotic-assisted movement of surgical instruments. Benefits include improved precision and dexterity for delicate microsurgical tasks on small vessels/ducts. Device is used for free-flap and lymphatic surgeries.
Clinical Evidence
Prospective, non-randomized, single-arm, multicenter PMCF study (N=62; 55 prospective, 7 retrospective). Evaluated safety/effectiveness for free-flap procedures in mouth (N=40) and scalp (N=20). Primary endpoints: intraoperative patency at first attempt (Mouth: 90.9%; Scalp: 69.2%), 3-day reoperation rate (Mouth: 5.0%; Scalp: 0.0%), and SADE rate (0.0% for both). All anastomoses achieved 100% patency prior to closure. Adverse event incidence (41.0%) consistent with literature for head/neck procedures.
Technological Characteristics
Electromechanical system for open microsurgery. Components: CMM, console, and wristed instruments. Features selectable motion scaling (7x-20x). Connectivity: Standalone robotic system. Software: Robotic control algorithm for motion scaling and tremor filtration. Sterilization: Not specified.
Indications for Use
Indicated for adults undergoing microsurgical procedures requiring soft tissue manipulation, anastomosis, suturing, or ligation of small blood vessels/lymphatic ducts (0.1-2.5 mm) in open free-flap surgery (breast, mouth, scalp, extremities) or open lymphatic surgery (extremities) when motion scaling is appropriate.
Regulatory Classification
Identification
An electromechanical system for open microsurgery is a software-controlled electromechanical system with bedside human/device interfaces and without an integrated visualization system which allows a qualified user to perform surgical techniques during open mircosurgical procedures using surgical instruments attached to an electromechanical arm.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym stacked above the words "U.S. FOOD & DRUG ADMINISTRATION". The seal on the left is a stylized image of an eagle.
February 7, 2025
Medical Microinstruments, Inc. Zainab Amini Sr. Regulatory Affairs Specialist Via Egidio Giannessi 54 Pisa, Italy 56121 Italy
Re: K242368
Trade/Device Name: Symani Surgical System Regulation Number: 21 CFR 878.4963 Regulation Name: Electromechanical System For Open Microsurgery Regulatory Class: Class II Product Code: SAQ, Dated: January 8, 2025 Received: January 8, 2025
Dear Zainab Amini:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898, and the April 5, 2024 De Novo classification order for this
{2}------------------------------------------------
type of device. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Per the April 5, 2024 De Novo classification order for this device type, you must demonstrate that the device performs as intended under anticipated conditions of use in the intended patient population and anatomical location. The special control requirements set forth in that order include initiation, enrollment, completion, and reporting requirements associated with any required postmarket surveillance. Within 30 days of receipt of this letter, you must submit a complete study protocol for a postmarket surveillance study consistent with the special control requirements. FDA expects to work with you to approve your study protocols within 60 days of this letter. Your submission should be clearly labeled as "Postmarket Study Protocol" and submitted to the Agency as specified below. Please reference the 510(k) number above to facilitate processing. If there are multiple protocols being finalized after clearance of this 510(k) submission, please submit each protocol as a separate submission, identified by their unique study name(s).
From the date of study protocol approval, you must meet the following timelines:
- First subject enrolled within 12 months
- 20% of subjects enrolled within 24 months ●
- 50% of subjects enrolled within 36 months ●
- 100% of subjects enrolled within 48 months
In addition, you must submit separate periodic reports on the progress of the study as follows:
- Postmarket surveillance progress reports every six (6) months until subject enrollment has been . completed, and annually thereafter, from the date of the protocol approval letter, unless otherwise specified by FDA.
- If any enrollment milestones are not met, you must begin submitting enrollment status reports every . three (3) months in addition to your annual postmarket study progress reports, until enrollment has been completed, or FDA notifies you otherwise.
- Submit the final postmarket study report three (3) months from study completion (i.e., last subject's ● last follow-up date).
Each postmarket surveillance report should be submitted to the Agency as specified below, identified as a "Postmarket Surveillance Report" in accordance with how the study is identified above, and bearing the applicable 510(k) reference number.
Be advised that failure to comply with any special control requirement, including the initiation, enrollment, completion, and reporting per the postmarket surveillance data requirements outlined above, may result in the adulteration and misbranding of your device.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510/k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming
{3}------------------------------------------------
product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
All required documents should be submitted, unless otherwise specified, to the address below and should reference the above 510(k) number to facilitate processing.
Postmarket Mandated Studies Program U.S. Food and Drug Administration Center for Devices and Radiological Health
{4}------------------------------------------------
Document Control Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
Alternatively, documents can be submitted electronically through the CDRH Portal. For more information on the CDRH Portal, please visit https://www.fda.gov/medical-devices/industry-medical-devices/send-andtrack-medical-device-premarket-submissions-online-cdrh-portal.
Sincerely,
Digitally signed by Mark Mark Trumbore Trumbore -S Date: 2025.02.07 14:17:23 -2 -05'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242368
Device Name Symani Surgical System
#### Indications for Use (Describe)
The Symani® Surgical System is intended for soft tissue manipulation to perform anastomosis, suturing, and ligation microsurgery techniques on small blood vessels and lymphatic ducts between 0.1 and 2.5 mm in open free-flap surgery of the breast, mouth, scalp, extremities, and in open lymphatic surgery of the extremities.
The Syman® Surgical System is indicated for use during microsurgical procedures when use of a motion scaling function is deemed appropriate by the surgeon. The System is indicated for use in adults. It is intended to be used by trained physicians in an appropriate operating environment in accordance with the Instructions for Use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
# 510(k) Summary
### 1. SUBMITTER INFORMATION
| Submitter: | MEDICAL MICROINSTRUMENTS, INC.<br>Via Egidio Giannessi 54<br>Pisa, IT 56121 |
|------------|-----------------------------------------------------------------------------|
| Contact: | Zainab Amini<br>(904) 896-6812<br>Zainab.amini@mmimicro.com |
| Date: | 4 February 2025 |
## II. SUBJECT DEVICE INFORMATION
| Device Trade Name: | Symani® Surgical System |
|----------------------|------------------------------------------------------------------|
| Classification Name: | Electromechanical System For Open Microsurgery (21CFR §878.4963) |
| Regulatory Class: | II |
| Product Code: | SAQ |
| Submission Type: | Traditional 510(k) |
# III. PREDICATE DEVICE INFORMATION:
Predicate Device: Symani Surgical System, DEN230032
## ıv. DEVICE DESCRIPTION:
This 510(k) is for a labeling modification only, the addition of the mouth and scalp to the indication for use among the existing open free-flap surgery. The technological characteristics remain identical as the predicate device, Symani Surgical System in DEN230023.
Symani is designed to provide surgeons with increased precision while leveraging the surgeon's own surgical experiences. The Symani Surgical System consists of three components, the CMM, Console, and Instruments, with selectable scaling factors (between 7x to 20x) to improve motion control and scale down any physiological tremor of the operator in a similar fashion to conventional robotic-assisted surgical device (RASD) systems, and the instruments with wristed end-effectors to improve dexterity for tissue manipulation and suturing.
## v. INDICATIONS FOR USE
The Symani® Surgical System is intended for soft tissue manipulation to perform anastomosis, suturing, and ligation microsurgery techniques on small blood vessels and lymphatic ducts between 0.1 and 2.5 mm in open free-flap surgery of the breast, mouth, scalp, extremities, and in open lymphatic surgery of the extremities. The Symani® Surgical System is indicated for use during microsurgical procedures when use of a motion scaling function is deemed appropriate by the surgeon. The System is indicated for use in adults. It is intended to be used by trained physicians in an appropriate operating environment in accordance with the Instructions for Use.
{7}------------------------------------------------
# vi. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technological characteristics of the subject and predicate devices remain the same in this submission.
# vii. PERFORMANCE DATA
The data used for this 510(k) comes from a prospective, non-randomized, single-arm, multicenter, post-market clinical follow-up (PMCF) study designed to evaluate the safety and effectiveness of the Symani System when used to assist microsurgical anastomoses in the real-world setting.
For this 510(k), a subset of patients has been analyzed, specifically the free tissue transfer procedures performed using Symani with mouth or scalp as the recipient sites. The endpoints considered and analysis performed follow a robust Statistical Analysis Plan from the predicate device.
Sixty-two patients were identified, of which 55 are prospective enrollments and 7 are retrospective enrollments.
### Clinical Study Demographics
| Characteristic | Statistics<br>Demographics | Mouth | Scalp |
|----------------------------------------------|----------------------------|-------------------|-------------------|
| Age (years) | Mean (SD); N | 62.1 (12.03); 40 | 63.5 (12.77); 20 |
| Gender | | | |
| Male | n (%); N | 25 (62.5%); 40 | 14 (70.0%); 20 |
| Female | n (%); N | 15 (37.5%); 40 | 6 (30.0%); 20 |
| Race | - | - | |
| Caucasian | n (%); N | 40 (100.0%); 40 | 20 (100.0%); 20 |
| Black | n (%); N | 0 (0.0%); 40 | 0 (0.0%); 20 |
| Asian | n (%); N | 0 (0.0%); 40 | 0 (0.0%); 20 |
| Other | n (%); N | 0 (0.0%); 40 | 0 (0.0%); 20 |
| Height (cm) | Mean (SD); N | 173.1 (9.92); 39 | 173.2 (10.34); 20 |
| Weight (kg) | Mean (SD); N | 75.4 (16.05); 39 | 71.0 (17.33); 20 |
| BMI | Mean (SD); N | 25.22 (5.292); 39 | 23.34 (4.001); 20 |
| Pre-operative co-existing conditions/disease | | | |
| Pre-operative radiotherapy | n (%); N | 8 (20.0%); 40 | 3 (15.0%); 20 |
| Pre-operative chemotherapy | n (%); N | 5 (12.8%); 39 | 4 (20.0%); 20 |
| Smoking now/past | n (%); N | 18 (45.0%); 40 | 11 (55.0%); 20 |
| Hypertension | n (%); N | 14 (35.0%); 40 | 11 (55.0%); 20 |
| Pulmonary disease | n (%); N | 7 (17.5%); 40 | 5 (25.0%); 20 |
| Diabetes | n (%); N | 4 (10.0%); 40 | 1 (5.0%); 20 |
| Ischaemic heart disease | n (%); N | 4 (10.0%); 40 | 1 (5.0%); 20 |
| Alcohol consumption over limits | n (%); N | 10 (25.0%); 40 | 5 (25.0%); 20 |
| Steroids | n (%); N | 1 (2.5%); 40 | 0 (0.0%); 20 |
| Extra-cardiac arteriopathy | n (%); N | 4 (10.0%); 40 | 2 (10.0%); 20 |
| BMI | | | |
| <20 | n (%); N | 5 (12.8%); 39 | 5 (25.0%); 20 |
| 20-<25 | n (%); N | 19 (48.7%); 39 | 9 (45.0%); 20 |
| 25 to <30 | n (%); N | 9 (23.1%); 39 | 5 (25.0%); 20 |
| 30 to <35 | n (%); N | 4 (10.3%); 39 | 1 (5.0%); 20 |
| >35 | n (%); N | 2 (5.1%); 39 | 0 (0.0%); 20 |
Table 1. Free-Flap Clinical Study Demographics in the Mouth and Scalp
{8}------------------------------------------------
| American Society of<br>Anesthesiologist Score (ASA) | - | - | |
|-----------------------------------------------------|----------|----------------|----------------|
| 1 | n (%); N | 18 (45.0%); 40 | 7 (35.0%); 20 |
| 2 | n (%); N | 13 (32.5%); 40 | 10 (50.0%); 20 |
| 3 | n (%); N | 9 (22.5%); 40 | 3 (15.0%); 20 |
| 4 | n (%); N | 0 (0.0%); 40 | 0 (0.0%); 20 |
### Analysis results - Effectiveness Endpoints:
#### Intraoperative patency at first attempt
The subset of data used for this effectiveness endpoint were anastomoses that were marked as first attempt using the robotic approach. The patency of all anastomoses (100%) was confirmed prior to closure. Results reported by site for intraoperative patency at first attempt are as follows:
Table 2. Intra-operative Anastomosis Patency at First Attempt
| Site | Number of<br>anastomoses | Number patent | Patency estimate | 95% Cl |
|-------|--------------------------|---------------|------------------|----------------|
| Mouth | 66 | 60 | 90.9% | [81.3%, 96.6%] |
| Scalp | 26 | 18 | 69.2% | [48.2%, 85.7%] |
### Operative time and suturing time
The average procedure time and average suturing time for robotic, manual, and hybrid anastomoses were analyzed and are summarized as follows:
Table 3. Operative Time and Suturing Time
| Site | Average<br>Operative<br>time (hours) | Average suturing time (minutes) | | |
|-------|--------------------------------------|---------------------------------|--------|--------|
| | | Robotic | Manual | Hybrid |
| | | Mouth | 11.66 | 37.4 |
| Scalp | 12.30 | 41.2 | 26.0 | 38.5 |
### Rate of intra-operative change from robotic to manual suturing:
The subset of data used for this effectiveness endpoint were anastomoses that used the robotic approach. There were five changes in the intraoperative suturing approach with the estimated change rate being 5.1% (5/98). Results reported by site for rate of intra-operative approach change from robotic to manual suturing are as follows:
#### Table 4. Rate of intra-operative change
| Site | Intraoperative<br>changes | Intraoperative<br>Change Rate | 95% Confidence<br>Interval |
|-------|---------------------------|-------------------------------|----------------------------|
| Mouth | 3/72 | 4.2% | [0.9%, 11.7%] |
| Scalp | 2/26 | 7.7% | [0.9%, 25.1%] |
{9}------------------------------------------------
### Analysis results - Safety Endpoints:
The safety endpoint in the study was freedom from Serious Adverse Device Effects (SADE), which was 100.0%. The freedom from SADE rate is the complement of the estimated SADE rate. The Serious Adverse Device Effect Rate is presented in Table 5 below.
| Site | SADE/Total patients | SADE Rate | 95% Confidence Interval |
|-------|---------------------|-----------|-------------------------|
| Mouth | 0/41 | 0.0% | [0.0%, 8.6%] |
| Scalp | 0/20 | 0.0% | [0.0%, 16.8%] |
Table 5. Serious Adverse Device Effect Rate
### Adverse events
There were a total of 27 adverse events (from 25 patients). The incidents of AE was 41.0%(25/61 patients). A literature search showed a high variability in AE reporting, leading to the conclusion that for head & neck recipient sites, the AE rate ranges between 21.5% and 61.3% (KM Estimate 54.2 % [95% C1: 38.6%, 69.1%]). The incidence of AE from the head & neck procedures of the subject device study at the time of this interim analysis is consistent with the range identified in the literature search.
There was a total of 5 (5/41; 12.2%) and 2 (2/20; 10%) serious adverse events for the mouth and scalp, respectively. This was lower than the predicate device. The rate of adverse events resulting in anastomosis revision of the mouth was 5% (2/40). The were no adverse events resulting in anastomosis revision for the scalp.
#### 3-Day Reoperation Rate
The 3-day anastomosis specific reoperation rate is a clinical risk to the anastomosis in free flap surgery (Table 6). This is the timeframe the anastomosis typically fails and needs a revision. In the mouth cohort, there were two reoperations (2/40; 5.00%) done within 3 days to revise the anastomosis.
| Site | Count/Total | Percentage | 95% CI |
|-------|-------------|------------|------------------|
| Mouth | 2/40 | 5.00% | [0.61% - 16.92%] |
| Scalp | 0/20 | 0.00% | [0.00% - 16.84%] |
#### Table 6. 3-day Reoperation Rate
## viii. Postmarket Surveillance
In summary, the premarket data demonstrates several probable benefits and risks when used in the creation of vascular anastomoses. The risks of the Symani Surgical System and clinical study limitations are mitigated by the special controls and labeling. Outstanding uncertainty regarding the generalizability of effectiveness of device use in the target U.S. patient population will be addressed by postmarket data collection associated with special controls requirements. Medical Microinstruments, Inc. will meet defined enrollment milestones following device clearance and will submit periodic postmarket surveillance progress reports until enrollment is complete, followed by annual updates.
{10}------------------------------------------------
# IX. CONCLUSION
Based on the information provided within this submission, and comparison of the subject vs. predicate device, the subject device Symani Surgical System has the same intended use and technological characteristics as the predicate device.
The clinical evidence provided relating to free flap procedures in the mouth and scalp anatomical regions demonstrated that use of the Symani Surgical System in these indications is as safe and effective as the predicate device. For the mouth, the patency at first attempt rate and 3-day reoperation rate are similar to the predicate device. The patency at first attempt rate for the scalp was lower than the predicate device. This may be a result of unique challenges in the anatomy. If the first attempt at creating an anastomosis is not be successful, the anastomosis may be redone intraoperatively. The patency of all anastomoses (100%) was confirmed prior to closure. There were no device related adverse events resulting in anastomosis revision.
Therefore, the subject device is deemed to be substantially equivalent to the legally commercialized predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.