The SignalNED Device is intended to record and store EEG signals and display Quantitative EEG (qEEG) (relative band power, e.g., alpha, beta, delta, theta), which is intended to help the EEG. The SignalNED does not provide any diagnostic conclusion about the patient's condition. The device is intended to be used on adults by qualified medical and clinical professionals. The SignalNED is intended to be used in a professional healthcare environment.
Device Story
SignalNED System (Model RE) is a portable EEG device for professional healthcare environments. It utilizes a sensor cap with 10 Ag/AgCl electrodes (8 channels) to acquire EEG signals. The system processes these signals via an analog-to-digital converter and calculates relative band power (alpha, beta, delta, theta). The device displays these qEEG measurements in real-time to assist clinicians in EEG analysis. It does not provide diagnostic conclusions or automated clinical alerts. The system is battery-powered and operates without wireless connectivity requirements. Clinicians use the output to support their clinical decision-making process. The device benefits patients by providing rapid, portable EEG recording and quantitative analysis.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including lead-off detection, signal acquisition noise levels, software ADC conversion accuracy, QEEG relative band power calculation accuracy, and electrical performance testing per EC12:2020 and IEC 80601-2-26. Biocompatibility testing was performed per ISO 10993-1, -5, -10, and -23.
Technological Characteristics
Portable EEG system using 10 Ag/AgCl cutaneous electrodes in a spandex/nylon cap. Features: 250Hz sampling rate, 0.5-100Hz frequency band, >2 Gohm differential input impedance, >1 Gohm common mode input impedance. Processor: NXP i.MX 8M Plus (1.8GHz Quad Cortex-A53). Connectivity: 2.4 GHz IEEE 802.11 b/g/n (integrated antenna). Power: Battery-operated with 100-240V AC charger. Biocompatibility: ISO 10993 compliant. Software: Non-diagnostic signal processing and qEEG calculation.
Indications for Use
Indicated for use on adults by qualified medical and clinical professionals in professional healthcare environments to record and store EEG signals and display qEEG (relative band power) to assist in EEG analysis. No diagnostic conclusions provided.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
September 4, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Forest Devices, Inc. % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K242306
Trade/Device Name: SignalNED System (Model RE) Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMC, OLT, GWQ Dated: August 5, 2024 Received: August 5, 2024
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image shows the name "Patrick Antkowiak -S" in black text. The text is positioned to the right of a faded blue FDA logo. The name is split into two lines, with "Patrick" on the first line and "Antkowiak -S" on the second.
for
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242306
Device Name SignalNED System (Model RE)
## Indications for Use (Describe)
The SignalNED Device is intended to record and display Quantitative EEG (qEEG) (relative band power, e.g., alpha, beta, delta, theta), which is intended to help the user analyze the EEG. The SignalNED does not provide any diagnostic conclusion about the patient's condition. The device is intended to be used on adults by qualified medical and clinical professionals.
The SignalNED is intended to be used in a professional healthcare environment.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K242306
# 510(k) Summary
# Traditional 510(k) Premarket Notification Summary of Safety and Effectiveness
This 510(k) summary was prepared on 5/28/2024 to provide an understanding of the basis for a determination of substantial equivalence in accordance with the requirements outlined in 21 CFR 807.92. The content is also contained with the eSTAR format within Submitter's eSTAR application.
# Submitter Information (Contact Details)
| Applicant Name: | Forest Devices, Inc. |
|--------------------|----------------------------------------------------------------------|
| Applicant Address: | 544 Miltenberger St 1st Floor, #4 Pittsburgh PA 15219 United States: |
| Applicant Contact: | Telephone: 412.522.0076 |
| Applicant Contact: | Email: carmelo.montalvo@forestdevices.com) |
Date Prepared: September 3, 2024
## Device Name
| Device Trade Name: | SignalNED System (Model RE) |
|----------------------|------------------------------------------------|
| Common Name: | Electroencephalograph |
| Classification Name: | Reduced-Montage Standard Electroencephalograph |
| Regulation Number: | 882.1400 |
| Regulatory Class: | Class II |
## Product Code(s)
Primary Product Code: OMC Secondary Product Codes: OLT and GXY
## Legally Marketed Predicate Devices
| Predicate | Predicate Trade Name<br>(Primary Predicate is listed first) | Product Code |
|-----------|-------------------------------------------------------------|------------------------------------------------------------------------------|
| K170363 | Ceribell Pocket EEG Device | OMC, Reduced-Montage Standard Electroencephalograph (Primary) |
| K143487 | Lifelines iEEG 2.0 | OLT, Non-Normalizing Quantitative Electroencephalograph Software (Secondary) |
| K171459 | Ceribell Instant EEG Headband | GXY, Cutaneous Electrode (Secondary) |
## Device Description Summary
The SignalNED Model RE machine uses 10 patient electrodes (4 left, 4 right, 2 midh are used to form the 8 channels. The SignalNED machine requires the use of the SignalNED Sensor Cap, and the system includes the following components:
- Portable EEG machine (Device)
- I Battery & External Battery Charger
- I SignalNED Sensor Cap
- I SignalNED Sensor Cap Cable
The primary function of the SignalNED Model RE is to rapidly record EEG and derive the Quantitative EEG (qEEG) measurement of Relative Band Power for multiple bands (e.g., alpha, beta, theta) at each electrode. These measurements are intended to help the user analyze the underlying EEG. The SignalNED Model RE (client) achieves its intended without relying on wireless comectivity. The SignalNED RE does not provide any diagnostic conclusion about the patient's condition.
## Intended Use/Indications for Use
The SignalNED Device is intended to record and store EEG signals and display Quantitative EEG (qEEG) (relative band power, e.g., alpha, beta, delta, theta), which is intended to help the EEG. The SignalNED does not provide any diagnostic conclusion about the patient's condition. The device is intended to be used on adults by qualified medical and clinical professionals. The SignalNED is intended to be used in a professional healthcare environment.
{5}------------------------------------------------
## Substantial Equivalence Comparison Summary
| Attribute | Subject Device<br>The SignalNED Model RE System<br>K Pending | Primary Predicate Device<br>Ceribell Pocket EEG<br>K170363 | Comparison |
|--------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------|
| Intended Use | To record and store EEG signals | To record and store EEG signals | Identical. |
| | Presents EEG signals in visual formats in device real-time | Presents EEG signals in visual formats in device real-time | Identical. |
| | Device does not have a diagnostic claim | Device does not have a diagnostic claim | Identical. |
| | Device does not provide any diagnostic conclusions about<br>the patient's condition | Device does not provide any diagnostic conclusions about<br>the patient's condition | Identical. |
| | Used on Adults | Used on Adults | Identical. |
| Technology | Records and stores EEG signals | Records and stores EEG signals | Identical. |
| | Uses specially designed cutaneous electrodes | Uses specially designed cutaneous electrodes | Identical. |
| | Uses specially designed cutaneous electrodes | Uses specially designed cutaneous electrodes | Identical. |
| | Uses analog to digital converter | Uses analog to digital converter | Identical. |
| | Uses frequency band of 0.5Hz - 100Hz | Uses frequency band of 0.5Hz - 100Hz | Identical. |
| | Device samples at 250Hz | Device samples at 250Hz | Identical. |
| | Displays calculated EEG measures | Displays calculated EEG measures | Identical. |
| | Use 2.4 GHz IEEE 802.11 b/g/n WIFI frequency | Use 2.4 GHz IEEE 802.11 b/g/n WIFI frequency | Identical. |
| | Uses Integrated Antenna | Uses Integrated Antenna | Identical. |
| | Features surface mount PCB module | Features surface mount PCB module | Identical. |
| | Battery powered | Battery powered | Identical. |
# Intended Use/Indications for Use Comparison
Primary Predicate: K170363 Ceribell Pocket EEG
The SignalNED Model RE System (Subject Device) is identical to the Primary Predicate (Predicate) in Intended to record and store EEG signals and present them in real-time. Neither the Subject Device nor Primary Predicate provides any diagnostic conclusion about the subject's condition or automated alerts of a clinical event. Additionally, neither device has a diagnostic claim, nor do they provide any diagnostic conclusions about the patient's condition. Both the Primary Predicate are intended to be used on adults by qualified medical and clinical professionals.
# Technology Comparison
| Device & Predicate Device(s): | K242306 | K170363<br>OMC<br>Predicate | K143487<br>OLT<br>Predicate | K171459<br>GWQ<br>Predicate | Discussion |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Device Characteristics | | | | | |
| EEG Data Acquisition | 10 electrodes to<br>form 8 recording<br>channels | 10 electrodes to<br>form 8 recording<br>channels | | | Similar |
| Data Acquisition Method | Analog to digital<br>converter | Analog to digital<br>converter | | | Similar |
| Frequency band | 0.5 Hz - 100 Hz | 0.5 Hz - 100 Hz | | | Similar |
| Sampling rate | 250 Hz | 250 Hz | | | Similar |
| Displays calculated EEG measures | Yes | Yes | | | Similar |
| WIFI Frequency Standard | 2.4 GHz IEEE<br>802.11 b/g/n | 2.4 GHz IEEE<br>802.11 b/g/n | | | Similar |
| Antenna Type | Integrated | Integrated | | | Similar |
| Form Factor / Dimensions | Surface mount<br>PCB module | Surface mount<br>PCB module | | | Similar |
| Integrated Processor and Memory | 1.8GHz Quad<br>CortexTM-A53 | STM32 ARM<br>Cortex-M3 | | | Similar; The<br>subject device |
| Device & Predicate Device(s): | K242306 | K170363<br>OMC<br>Predicate | K143487<br>OLT<br>Predicate | K171459<br>GWQ<br>Predicate | Discussion |
| | NXP i.MX 8M<br>Plus | processor with<br>64 KB RAM | | | includes<br>additional<br>processing and<br>memory<br>capability |
| Device Power | Battery Powered | Battery Powered | | | Similar |
| Battery | 100-240 V AC<br>power adapter to<br>battery charger | 100-240 V AC<br>power adapter to<br>battery charger | | | Similar |
| Differential input impedance | >2 Gohm | | >20 Mohm | | Similar;<br>electrical<br>specification<br>exceeds<br>predicate device.<br>This difference<br>does not impact<br>safety and<br>effectiveness. |
| Common mode input impedance | >1 Gohm | | >100 Mohm | | Similar:<br>electrical<br>specification<br>exceeds<br>predicate<br>device. This<br>difference does<br>not impact safety<br>and<br>effectiveness. |
| Channel equivalent input noise | 1.97 μVpp | | <3.5 µVpp | | Similar;<br>electrical<br>specification<br>exceeds<br>predicate<br>device. This<br>difference does<br>not impact safety<br>and<br>effectiveness. |
| Frequency band | 0.5 Hz - 30 Hz | | 0.16 Hz - 70 Hz<br>(-6 dB) | | Similar; within<br>range of<br>predicate. This<br>difference does<br>not impact safety<br>and<br>effectiveness. |
| Low filter | 0.5 Hz | | 0.5 Hz | | Same |
| High filter | 30 Hz | | 10 Hz - 100 Hz | | Similar; within<br>range of<br>predicate. This<br>difference does<br>not impact safety<br>and<br>effectiveness. |
| Sampling rate | 1 kHz | | 200,256 Hz | | Similar; |
| Device & Predicate Device(s): | K242306 | K170363<br>OMC<br>Predicate | K143487<br>OLT<br>Predicate | K171459<br>GWQ<br>Predicate | Discussion |
| | | | | | specification<br>exceeds<br>predicate<br>device. This<br>difference does<br>not impact safety<br>and<br>effectiveness. |
| Displays calculated EEG measures | Yes | | Yes | | Same |
| Calculated EEG measures displayed | Power Spectrum<br>Density | | Spectrum, Power<br>Spectrum,<br>Density, band<br>power, spectral<br>edge, peak<br>frequency | | Similar; subject<br>device is a<br>subset to the<br>predicate<br>device. |
| WIFI Frequency Standard | 2.4 GHz IEEE<br>802.11 b/g/n | | 2.4 GHz IEEE<br>802.11 b/g/n | | Similar |
| Antenna Type | Integrated | | Integrated | | Similar |
| Where used | On the head | | | On the head | Similar |
| Means of Connection | A cable<br>attached to the<br>Sensor Cap to<br>allow<br>connection to<br>SignalNED<br>Model RE device | | | A cable<br>attached to<br>the headband<br>to allow<br>connection<br>to an EEG<br>acquisition/recor<br>ding device | Similar |
| Electrode Type | Silver/silver-<br>chloride<br>(Ag/AgCl)<br>electrodes | | | Silver/silver-<br>chloride<br>(Ag/AgCl)<br>electrodes | Similar |
| Number of Electrodes | 10 electrodes | | | 10 electrodes | Similar |
| Electrode Form | Reservoir is<br>filled with<br>conductive<br>electrolyte gel<br>by separate<br>dispensing<br>container | | | Reservoir is<br>filled with<br>conductive<br>electrolyte gel<br>by separate<br>dispensing<br>container | Similar |
| Electrode Locations | The placement of<br>the electrodes is<br>according to the<br>International 10-<br>20 system of<br>electrode<br>placement or the<br>American<br>Electroencephalo<br>graphic Society<br>positioning<br>system. | | | The placement<br>of the electrodes<br>is according to<br>the International<br>10- 20 system of<br>electrode<br>placement or the<br>American<br>Electroencephal<br>ographic Society<br>positioning<br>system. The<br>number of the<br>electrodes in use<br>is according to | Similar; both<br>use the<br>International<br>10-20 system of<br>electrode<br>placement or<br>the American<br>Electroencephal<br>ographic<br>Society<br>positioning<br>system. The<br>Subject device<br>uses 10<br>electrodes |
| Device & Predicate Device(s): | K242306 | K170363<br>OMC<br>Predicate | K143487<br>OLT<br>Predicate | K171459<br>GWQ<br>Predicate | Discussion |
| | | | | whereas the<br>Predicate device<br>has<br>configuration<br>option from 9-<br>19<br>electrodes. | |
| Electrode Mounts | Thermoplastic<br>Elastomer | | | Thermoplastic<br>Elastomer | Same |
| Available Sizes | Various sizes<br>(overall head<br>size range 26 cm<br>–<br>66 cm) | | | Various sizes<br>(overall head<br>size range 26<br>cm –<br>66 cm) | Same |
| Cap Material | Spandex blend:<br>Nylon Knitted<br>Fabric, 82%<br>Nylon and 18%<br>Spandex | | | Spandex blend:<br>Nylon Knitted<br>Fabric, 82%<br>Nylon and 18%<br>Spandex | Same |
| Type of Connector | Custom Touch-<br>proof safety<br>socket | | | Touch-proof<br>safety socket<br>DIN 42- 802 | Similar; the<br>electrical<br>performance<br>testing completed<br>for both Subject<br>and Predicate met<br>the same<br>requirements per<br>FDA guidance.<br>document<br>"Cutaneous<br>Electrodes for<br>Recording<br>Purposes-<br>Performance<br>Criteria for Safety<br>and Performance<br>Based Pathway,"<br>August 2020. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# Intended Use/Indications for Use Comparison
# Secondary Predicate: K143487 Lifelines iEEG 2.0
The Subject is identical to the Secondary Predicate (K143487, Lifelines iEEG 2.0) in Intended to acquire, display, and store EEG signals, using cutaneous electrodes. Neither the Subject Device nor Secondary Predicate provides any diagnostic conclusion about the subject's condition or automated alerts of a clinical event. Additionally, neither device claim, nor do they provide any diagnostic conclusions about the patient's condition.
# Technology Comparison
## Secondary Predicate: K143487 Lifelines iEEG 2.0
Both the Subject Device and Secondary Predicate (K143487) are identical in that they display Quantitative Electroencephalograph (qEEG) measures that are derived from the acquired EEG. They are similar in differential input impedance, common mode input impedance, channel equivalent input noise, frequency band, low filter, sampling rate. The Subject exceeds the predicate in electrical specifications or is within the specified range of the Predicate.
## Intended Use/Indications for Use Comparison
{9}------------------------------------------------
# Secondary Predicates: K171459 Ceribell Instant EEG Headband
The Intended Use of the Subject Device and the Secondary Predicate (K171459, Cerbell Instant EEG Headband) is identical. Both use a disposable headpiece with an integrated array of 10 passive cutaneous electrodes that are applied to the patient's head to record EEG signals when connected to an EEG recording device.
## Technology Comparison
## Secondary Predicates: K171459 Ceribell Instant EEG Headband
The technology of the Subject Device and the Secondary Predicate (K171459, Cerbell Instant EEG Headband) is identical in type of patient contact, where used, means of connection, electrodes and electrode form. The Subject and Secondary Predicate both follow placement of the electrodes is according to the International 10-20 system of the American Electroencephalographic Society positioning system. The Subject uses 10 electrodes, and the Seconder 9-19 electrodes. The electrode mounts, sizes electrode cap materials. performance requirement, electrical compliance, designation for use are identical. Both are packaged non-sterile. Both demonstrate conformance to ISO 10993-1, ISO 10993-5, and ISO 10993-10.
| Testing | Objective and Outcome |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Lead Off Detection | ■ Test is designed to validate the ability to detect disconnected electrodes.<br>■ All testing passed acceptance criteria and details are contained in the test report. |
| Signal Acquisition Noise Levels | ■ Test is designed to assess noise levels in signal acquisition.<br>■ All testing passed acceptance criteria and details are contained in the test report |
| Software ADC Conversion Accuracy | ■ Test is designed to verify accuracy of software in ADC conversion.<br>■ All testing passed acceptance criteria and details are contained in the test report. |
| Quantitative Electroencephalogram<br>(QEEG) | ■ Test is designed to ensure accuracy of the QEEG Relative Band Power calculation.<br>■ All testing passed acceptance criteria and details are contained in the test report. |
| EC12:2020 Electrical Performance | ■ Test is designed to confirm compliance with EC12:2020 electrical standards.<br>■ All testing passed acceptance criteria and details are contained in the test report. |
| Essential Performance Tests<br>(IEC 80601-2-26) | ■ Test is designed to verify compliance with IEC 80601-2-26 essential performance requirements.<br>■ All testing passed acceptance criteria and details are contained in the test report. |
#### Summary of Non-clinical Performance Testing
## Electrical Performance
Electrical testing was conducted on the subject device complies with IEC 60601-1, IEC 60601-1-2. All testing passed.
# Biocompatible Performance
The Forest SignalNED System is an EEG device used by trained medical professionals. The EEG machine datient, but the Sensor Cap, which is part of the system, is placed on the patient's head (Limited Exposure less than 24 hours, Surface Contacting Medical Device that contacts intacts ). The SignalNED was assessed for biocompatibility under the requirements of ISO 10993 for the Biological Evaluation of Medical Devices for cytotoxicity, irritation within the Limited Contact duration (A), for Surface Medical Device for Intact Skin scope.
EN-ISO-10993-1:2020 the following test are the end points for our biocompatibility testing:
- I ISO 10993-1: Risk Management
- ISO 10993-5: Cytotoxicity
- 트 ISO 10993-10: Sensitization
- I ISO 10993-23: Irritation or intracutaneous reactivity
All testing passed.
# Conclusion
In conclusion, Forest Devices, Inc. has successfully demonstrated that the SignalNED is equivalent in safety and efficacy to the predicate in the intended use, technological characteristics and performance testing. The Forest Devices SignalNED Model RE System raises no additional questions of safety or effectiveness and is substantially equivalent to the cleared with the ences in the Subject Device and predicates do not raise additional questions of safety or effectiveness or exceeds all conditions that are required for establishing substantial equivalence and obtaining FDA clearance to market our medical device, the SignalNED Model RE System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.