← Product Code [ODP](/productcode/ODP) · K242195

# Gemini Cervical Fusion Cage System (K242195)

_ZheJiang Decans Medical Devices Co., Ltd. · ODP · Nov 4, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K242195

## Device Facts

- **Applicant:** ZheJiang Decans Medical Devices Co., Ltd.
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Nov 4, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Gemini Cervical Fusion Cage System is indicated for use in anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one levels from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The Gemini Cervical Fusion Cage System requires additional supplemental fixation cleared for the cervical spine. The Gemini Cervical Fusion Cage System is designed for use with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, to facilitate fusion and is to be implanted via an open, anterior approach. This cervical device is to be used in patients who have had six weeks of nonoperative treatment.

## Device Story

Intervertebral body fusion device; implanted via open, anterior approach to facilitate fusion in cervical spine. Device consists of PEEK-OPTIMA LT1 cage with tantalum markers for radiographic visualization. Requires supplemental fixation. Surgeon packs cage with autogenous or allogenic bone graft. Provides structural support to disc space; promotes fusion. Used in clinical/surgical settings by orthopedic or neurosurgeons.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated per ASTM F2077 and ASTM F2267, including static/dynamic axial compression, compression shear, torsion, and subsidence testing. Results met acceptance criteria per ISO 23089-2.

## Technological Characteristics

Material: PEEK-OPTIMA LT1 (ASTM F2026). Markers: Tantalum (ISO 13782). Design: Interbody cage with graft windows and serrated surfaces. Sterilization: Gamma irradiation (single use). Instruments: Steam sterilized.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- CORNERSTONE® PSR Cervical Fusion System ([K153373](/device/K153373.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol representing the Department of Health & Human Services on the left, and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in a sans-serif font, with "FDA" in a blue square and the rest of the text in blue as well.

November 4, 2024

ZheJiang Decans Medical Devices Co., Ltd. Chen Liu RA No.2836 Xincheng Avenue, Gaozhao Street, Xiuzhou District Jiaxing, Zhejiang 314031 China

Re: K242195

Trade/Device Name: Gemini Cervical Fusion Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: July 26, 2024 Received: October 16, 2024

Dear Chen Liu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Ethan R. Naylor -S

for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K242195

Device Name

Gemini Cervical Fusion Cage System

Indications for Use (Describe)

The Gemini Cervical Fusion Cage System is indicated for use in anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one levels from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The Gemini Cervical Fusion Cage System requires additional supplemental fixation cleared for the cervical spine. The Gemini Cervical Fusion Cage System is designed for use with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, to facilitate fusion and is to be implanted via an open, anterior approach. This cervical device is to be used in patients who have had six weeks of nonoperative treatment.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text 'K242195' on the first line and 'Page 1 of 2' on the second line. The text is centered and uses a serif font. The image appears to be a header or footer from a document.

Image /page/4/Picture/1 description: The image contains a logo for DECANS. The logo consists of a blue cube-like shape on the left and the word "DECANS" in blue on the right. Below the word "DECANS" are three Chinese characters, also in blue. The cube-like shape is made up of smaller cubes, creating a three-dimensional effect.

## 510K Summary

#### 1.SUBMITTER

ZheJiang Decans Medical Devices Co., Ltd. No.2836 Xincheng Avenue, Gaozhao Street, Xiuzhou District, Jiaxing City, Zhejiang Province,314031 P.R. China Contact Person:Chen Liu,RA Telephone:+86 18132030804 Email:cliu@decansmd.com Date Prepared:JUL/26/2024

#### 2.DEVICE

Trade Name:Gemini Cervical Fusion Cage System Classification :21 CFR 888.3080 Intervertebral body fusion device Class:II Product Code:ODP

#### 3.PREDICATE DEVICE

Primary Predicate:CORNERSTONE® PSR Cervical Fusion System(K153373)

#### 4.DEVICE DESCRIPTION

The Gemini Cervical Fusion Cage Systeme is an anterior cervical interbody device consisting of implants with various widths, heights and lengths to accommodate individual patient anatomy and graft material size. It is to be packed with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and is to be implanted via an open, anterior approach.

The Gemini Cervical Fusion Cage System is an implant constructed of medical grade Polyetheretherketone, (PEEK-OPTIMA® LT1) as described by ASTM F2026. The radiolucent PEEK-OPTIMA®material allows visualization of the defect site on radiography to assess bone growth and incorporates tantalum markers conforming to ISO13782 to permit verification of position. The Gemini Cervical Fusion Cage System is provided sterile via gamma irradiation for single use.Instruments are provided clean and non-sterile for steam sterilization at the user's facility.

#### 5.INDICATION FOR USE

The Gemini Cervical Fusion Cage System is indicated for use in anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one levels from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The Gemini Cervical Fusion Cage System requires additional supplemental fixation cleared for the cervical

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for DECANS. The logo consists of a blue cube-like shape on the left, with the text "DECANS" in blue to the right of the cube. Below the word "DECANS" is the text "德康医疗" in a smaller font size, also in blue. 

spine. The Gemini Cervical Fusion Cage System is designed for use with autografi bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, to facilitate fusion and is to be implanted via an open, anterior approach. This cervical device is to be used in patients who have had six weeks of nonoperative treatment.

## 6.COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject Gemini Cervical Fusion Cage System and the predicates share similar design features:

- · Graft windows for packing autogenous or allogenic bone
- · Serrations on the superior and inferior surfaces
- · Comparable heights, widths, depths

The indication for use,design features, materials used, manufacturing, and sterilization methods are identical to the previously cleared CORNERSTONE® PSR Cervical Fusion System.

There are insignificant differences between the subject Gemini Cervical Fusion Cage System and the predicates. Performance analysis shows that the subject Gemini Cervical Fusion Cage System is expected to perform as well as the predicate devices.

### 7.PERFORMANCE DATA

Performance testing was conducted per ASTM F2077 and ASTM F2267. Specifically, Implant performed static and dynamic axial compression testing, static and dynamic compression shear testing, static and dynamic torsional testing and subsidence testing, Test results meet the acceptance criteria described in ISO 23089-2 Implants for surgery - Preclinical mechanical assessment of spinal implants and particular requirements - Part 2: Spinal intervertebral body fusion devices

#### 8.CONCLUSION

Gemini Cervical Fusion Cage System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. Thus, the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K242195](https://fda.innolitics.com/device/K242195)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
