OTOPLAN is intended to be used by otologists and neurotologists as a software interface allowing the display, segmentation, and transfer of medical image data from medical CT, MR, and XA imaging systems to investigate anatomy relevant for the preoperative planning and postoperative assessment of otological and neurotological procedures (e.g., cochlear implantation).
Device Story
OTOPLAN is a standalone SaMD platform for otological/neurotological surgical planning. Inputs include DICOM-compliant CT, MR, and XA images. The software provides tools for image fusion, semi-automatic/automatic segmentation of ENT structures (temporal bone, inner ear, skin), and geometric measurements (distances, cochlear parameters). It allows virtual trajectory planning and visualization of cochlear implant components (electrodes, leads, housings). Audiogram data can be imported for visualization. The device is used by otologists/neurotologists in clinical settings to assist in surgical planning and postoperative assessment. It does not control medical devices; it provides assistive information requiring clinical judgment. The output is a visual report and 3D reconstruction, aiding surgeons in assessing electrode insertion and anatomical landmarks, potentially improving surgical precision and patient outcomes.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench-only software verification and validation, including performance testing of automatic segmentation algorithms (Temporal Bone, Skin, Inner Ear) and image fusion (CT-CT, CT-MR). Validation used pooled clinical datasets (e.g., 43 temporal bones, 450 CBCT datasets) compared against ground truth established by three surgeons. Metrics included DICE coefficients (range 0.65–0.89) and Mean Absolute Difference (MAD) for thickness measurements.
Technological Characteristics
Standalone software (SaMD) running on Windows 10/11. Functions include DICOM viewing, 3D reconstruction, and image fusion. Segmentation uses intensity-based masking, gradient-based edge detection, and statistical shape modeling (SSM). Image fusion uses rigid registration to optimize mutual information. No hardware components. Software documentation level: Basic.
Indications for Use
Indicated for otologists and neurotologists to display, segment, and transfer medical image data (CT, MR, XA) for preoperative planning and postoperative assessment of otological and neurotological procedures, such as cochlear implantation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K243933 — Ceevra Reveal 3+ · Ceevra, Inc. · Mar 4, 2025
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 11, 2025
Cascination AG
Gordana Salleles
Head of Regulatory and Clinical Affairs
Steigerhubelstrasse 3
Bern, BE 3008
Switzerland
Re: K242120
Trade/Device Name: Otoplan
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QQE
Dated: July 19, 2024
Received: July 19, 2024
Dear Ms. Gordana Salleles:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K242120 – Gordana Salleles
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K242120 – Gordana Salleles
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S
Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K242120
Device Name
OTOPLAN
Indications for Use (Describe)
OTOPLAN is intended to be used by otologists and neurotologists as a software interface allowing the display, segmentation, and transfer of medical image data from medical CT, MR, and XA imaging systems to investigate anatomy relevant for the preoperative planning and postoperative assessment of otological and neurotological procedures (e.g., cochlear implantation).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
CASCINATION
# 510(K) SUMMARY
K242120
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.
# I. SUBMITTER
Manufacturer: CASCINATION AG
Steigerhubelstrasse 3
CH-3008 Bern
Switzerland
Tel: +41 31 632 0440
Fax: +41 31 552 04 41
Contact Person: Gordana Salleles
Head of Regulatory and Clinical Affairs
Date Prepared: April 9, 2025
# II. SUBJECT DEVICE
Device Name: OTOPLAN
Classification Name: Medical Image Management and Processing System
Regulation: 892.2050
Regulatory Class: Class II
Product Code: QQE
The Subject Device (OTOPLAN version 3.1) is an updated version of the Predicate Device (OTOPLAN version 2.0).
# III. PREDICATE DEVICE
Company: CASCINATION AG
CASCINATION AG
{5}
Section 05 – 510(k) Summary - K242120
Page: 2/13
Steigerhubelstrasse 3
CH-3008 Bern
Switzerland
Device name: OTOPLAN
510(k) number: K203486
Product Code: QQE
# IV. INDICATIONS FOR USE
OTOPLAN is intended to be used by otologists and neurotologists as a software interface allowing the display, segmentation, and transfer of medical image data from medical CT, MR, and XA imaging systems to investigate anatomy relevant for the preoperative planning and postoperative assessment of otological and neurotological procedures (e.g., cochlear implantation).
# V. DEVICE DESCRIPTION
OTOPLAN is a Software as a medical Device (SaMD) which consolidates a DICOM viewer, ruler function, and calculator function into one software platform. The user can
- import DICOM-conform medical images, fuse supported images and view these images.
- navigate through the images and segment ENT relevant structures (semi-automatic/automatic), which can be highlighted in the 2D images and 3D view.
- use a virtual ruler to geometrically measure distances and a calculator to apply established formulae to estimate cochlear length and frequency.
- create a virtual trajectory, which can be displayed in the 2D images and 3D view.
- identify electrode array contacts, lead, and housing of a cochlear implant to assess electrode insertion and position.
- input audiogram-related data that were generated during audiological testing with a standard audiometer and visualize them in OTOPLAN.
OTOPLAN allows the visualization of third-party information, that is, cochlear implant electrodes, implant housings and audio processors.
The information provided by OTOPLAN is solely assistive and for the benefit of the user. All tasks performed with OTOPLAN require user interaction; OTOPLAN does not alter data sets but constitutes a software platform to perform tasks that are otherwise performed manually. Therefore, the user is required to have clinical experience and judgment.
CASCINATION AG
{6}
Section 05 - 510(k) Summary - K242120
Page: 3/13
# VI. SUBSTANTIAL EQUIVALENCE
The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence:
Table 1
Summary of the Substantial Equivalence Comparison to Predicate Device
| Item | Subject Device (OTOPLAN version 3.1) | Predicate Device (OTOPLAN version 2.0) | Conclusion |
| --- | --- | --- | --- |
| **Intended Use** | | | |
| Intended Use | Plan surgical procedures in the head and neck area by medical professionals | Plan surgical procedures in the head and neck area by medical professionals | ⇒ Same
Both the subject and predicate devices have the same intended use |
| Indications For Use Statement | OTOPLAN is intended to be used by otologists and neurotologists as a software interface allowing the display, segmentation, and transfer of medical image data from medical CT, MR, and XA imaging systems to investigate anatomy relevant for the preoperative planning and postoperative assessment of otological and neurotological procedures (e.g., cochlear implantation). | OTOPLAN is intended to be used by otologists and neurotologists as a software interface allowing the display, segmentation, and transfer of medical image data from medical CT, MR, and XA imaging systems to investigate anatomy relevant for the preoperative planning and postoperative assessment of otological and neurotological procedures (e.g., cochlear implantation). | ⇒ Same
Both the subject and predicate devices have the same indications for use statement. |
| **Technological Characteristics** | | | |
| Type | Standalone Software. Does not control the functions or parameters of any medical device | Standalone Software. Does not control the functions or parameters of any medical device | ⇒ Same |
| Operating System | Windows 10
Windows 11 | Windows 10 | ⇒ Same
Both Windows-Versions have the same technological characteristics |
CASCINATION AG
{7}
Section 05 – 510(k) Summary - K242120
Page: 4/13
| Item | Subject Device (OTOPLAN version 3.1) | Predicate Device (OTOPLAN version 2.0) | Conclusion |
| --- | --- | --- | --- |
| Functions (Same Functions) | • Data Management
• Cochlear Parametrization (based on established formula)
• Audiogram
• Electrode Visualization
• Trajectory Planning
• Postoperative Quality Checks
• Export Report
• 3D reconstruction
· Temporal bone
· Incus, Malleus
· Stapes
· Facial nerve
· Chorda tympani
· External ear canal
· Cochlea
· Sigmoid sinus
· Cochlear bony overhang
· Cochlear round window
· Electrode contacts | • Data Management
• Cochlear Parametrization (based on established formula)
• Audiogram
• Electrode Visualization
• Virtual Trajectory Planning
• Postoperative Quality Checks
• Export Report
• 3D reconstruction
· Temporal bone
· Incus, Malleus
· Stapes
· Facial nerve
· Chorda tympani
· External ear canal
· Cochlea
· Sigmoid sinus
· Cochlear bony overhang
· Cochlear round window
· Electrode contacts | ⇒ Same |
| Functions (New Functions with Same technological characteristic) | • DICOM Viewer (incl. Fluoroscopy Viewer and plain X-ray)
• Electrode contacts (manual identification on plain X-ray; manual and automatic on CT images)
• Implant Placement (for visualization only)
• Identify the cochlear implant lead and housing | • DICOM Viewer
• Electrode contacts (manual and automatic identification on CT images) | ⇒ Same
Technological characteristics as included in the predicate device |
CASCINATION AG
{8}
Section 05 – 510(k) Summary - K242120
Page: 5/13
| Item | Subject Device (OTOPLAN version 3.1) | Predicate Device (OTOPLAN version 2.0) | Conclusion |
| --- | --- | --- | --- |
| Functions (New functions with different technological characteristic) | • 3D reconstruction
· Automatic Temporal bone
· Temporal Bone Thickness
· Automatic Skin and Skin Thickness
· Automatic Inner Ear (cochlea, semicircular canals, internal auditory canal)
· Automatic cochlear parameter
· Automatic Scala tympani and Scala vestibuli
• Image Fusion | --- | ⇒ Different Technological characteristics which do not affect the safety and effectiveness. |
| Performance Testing | Software design verification and validation and documentation (the software for this device was considered requiring “Basic Documentation”.)
Formal Internal Testing Standards
Human Factors Testing | Software design verification and validation and documentation (the software for this device was considered a “moderate” level of concern.)
Formal Internal Testing Standards
Human Factors Testing | ⇒ Same |
## Substantial Equivalence Discussion
The Subject Device OTOPLAN (version 3.1) is an updated version of the Predicate Device OTOPLAN (version 2.0). Both the Subject Device OTOPLAN (version 3.1) and the Predicate Device OTOPLAN (version 2.0) have the same Intended Use.
The subject device introduces eleven new functions. Four of these functions have the same technological characteristics and seven have different technological characteristics from the predicate device.
The subject device has four new functions with the same technological characteristics:
- Fluoroscopy and plain X-Ray Viewer (as part of the DICOM Viewer): simple extension of the DICOM Viewer by standardized DICOM SOP classes
- Electrode contacts (manual identification on plain X-ray): similar to the predicate device's contact identification on CT images the subject device allows the user to manual identify electrode contacts on plain X-ray images.
CASCINATION AG
{9}
Section 05 – 510(k) Summary - K242120
Page: 6/13
- Implant Placement (for visualization only): overlay function for visualization purposes, similar to other overlay functions in the predicate device
- Identify the cochlear implant lead and housing: similar to the predicate device's identification of cochlear implant electrode contacts, the subject device also identifies cochlear implant lead and housing
Those functions have the same technological characteristics as functions in the predicate device.
Software verification and validation has been carried out to ensure proper performance of those functions.
## Discussion of Technological Differences:
The subject device and predicate device have seven different technological characteristics in the 3D reconstruction and image fusion module. The subject device introduces the following functions:
- New 3D reconstructions
- Automatic Temporal Bone: automatic segmentation of the temporal bone in CT images. Formal internal testing using DICE similarity coefficient has been carried out to verify the accuracy of the automatic CT temporal bone 3D reconstruction algorithm. The ground truth has been established by three qualified surgeons.
- Temporal Bone Thickness: calculation of the bone thickness based on temporal bone reconstruction. Formal internal testing: manual measurements at typical areas of interest for ENT surgery by three qualified surgeons were compared to the automated thickness calculation
- Automatic Skin and Skin Thickness: automatic segmentation of the skin in CT images and calculation of the skin thickness based on the skin reconstruction. Formal internal testing using DICE similarity coefficient has been carried out to verify the accuracy of the automatic Skin 3D reconstruction algorithm. The ground truth has been established by three qualified surgeons. Formal internal testing: manual measurements at typical areas of interest for ENT surgery by three qualified surgeons were compared to the automated thickness calculation.
- Automatic Inner Ear: automatic segmentation of the inner ear (cochlea, semicircular canals, internal auditory canal) using a pre-computed statistical shape model (SSM). Formal internal testing using CT and MR images was carried out using DICE similarity coefficient to verify the accuracy of the automatic inner ear reconstruction algorithm. The ground truth has been established by three qualified surgeons.
- Automatic cochlear parameters: based on the Automatic Inner Ear reconstruction this function allows to identify the cochlear landmark points (cochlear diameter (A) and width (B)). Formal internal testing using CT and MR images was carried out comparing the
CASCINATION AG
{10}
Section 05 – 510(k) Summary - K242120
Page: 7/13
automatic parameter calculation to the parameter calculation based on the manual measurements by three qualified surgeons.
- Automatic Scala tympani and Scala vestibuli: automatic segmentation of the Scala tympani. Scala tympani: Formal internal testing was carried out using DICE similarity coefficient and other parameters (diameter, width, cross-section, volume) to verify the accuracy of the automatic inner ear reconstruction algorithm, between ground truth and test dataset. The scala vestibuli reconstruction has not been validated and is intended solely for visualization purposes.
- Image Fusion: After loading at least two images (CT-CT or CT-MR), two of them can be aligned onto each other). Formal internal testing was carried out by three experienced surgeons marking cochlear landmark points on pre-op (CT) and post-op (CT and MR) images. After the fusion of the pre-op and post-op images the distances between the corresponding landmark points were compared to validate the accuracy of the image fusion function.
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## i. Software Verification and Validation Testing
Software verification and validation testing were conducted to demonstrate safety and effectiveness of the subject device. The testing confirmed that the subject device performs as intended and meets all predefined acceptance criteria. Software validation documentation was prepared according to the "Content of Premarket Submissions for Device Software Functions - Guidance for Industry and Food and Drug Administration Staff" (June 14, 2023). The required documentation level for the subject device has been determined to be "basic documentation level". All planned software verification and validation testing were successfully completed, thereby demonstrating the safety and effectiveness of the subject device.
## ii. Automatic Outputs Validation
A summary of the validation of the software's automatic outputs is provided in Table 2 and Table 3.
CASCINATION AG
{11}
Section 05 - 510(k) Summary - K242120
Table 2
Summary of Performance Tests - Temporal Bone, Skin, and Inner Ear
| Variable | Temporal Bone | Skin | Inner Ear | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Scala Tympany | Cochlea, Semi-circular canals, internal auditory canal | | Cochlea | | |
| | Thickness Mapping | 3D Reconstruction | Thickness Mapping | 3D Reconstruction | | 3D Reconstruction | | Cochlear Parameters | |
| | | | | | | CT | MR | CT | MR |
| Nature of algorithm | Thickness mapping is performed by analyzing spatial relationships within the 3D model to determine the shortest distances between internal and external surfaces. | The algorithm estimates bone structures by analyzing intensity patterns in the imaging data. It then reconstructs a 3D model based on image-derived features. | Thickness mapping is performed by analyzing spatial relationships within the 3D model to determine the shortest distances between internal and external surfaces. | The algorithm relies on intensity-based masking, gradient-based edge detection, and geometric filtering within an ROI defined by inflating the segmented temporal bone. | A rule-based separation of cochlear chambers is performed using a cochlear centerline derived from statistical anatomical data. This guides a voxel-wise chamber assignment and reconstruction. | The algorithm uses statistical shape modeling techniques combined with image-based feature extraction to identify and segment key inner ear structures from imaging data. | | Combines statistical shape modelling-based initial estimation with geometric optimization. Final anatomical landmarks are refined through projection onto the 3D cochlear surface mesh. | |
| Testing Summary | | | | | | | | | |
| Aim | Validate the Thickness Measurement | Validate the Reconstruction | Validate the Thickness Measurement | Validate the Reconstruction | Validate the Measurements and Reconstruction | Validate the Reconstruction | Validate the Reconstruction | Validate the Measurements | Validate the Measurements |
| Testing Image Data | ○ 43 temporal bones (29 patients)○ 6 different CT, CBCT models○ Slice spacing: 16 different values (mean: 0.45 mm) | ○ 31 temporal bones (23 patients)○ 2 different CT, CBCT models○ Slice spacing: 15 different values (mean: 0.34 mm)○ Pixel spacing: 14 values | ○ 43 temporal bones (29 patients)○ 6 different CT, CBCT models○ Slice spacing: 16 different values (mean: 0.45 mm)○ Pixel spacing: 19 values | ○ 31 temporal bones (23 patients)○ 2 different CT, CBCT models○ Slice spacing: 15 different values (mean: 0.34 mm)○ Pixel spacing: 14 values | ○ Test data: 450 clinical-resolution CBCT datasets derived from 75 cochleae with voxel sizes 0.1 mm to 0.6 mm and pixel spacing 0.1 mm to 0.6 mm○ Ground truth data from Synchrotron Radiation Phase- | ○ 44 ears (27 patients)○ 2 different CT models○ Slice spacing: 7 different values (mean: 0.42 mm)○ Pixel spacing: 16 | ○ 41 ears (24 patients)○ 4 different MR models○ Slice spacing: 7 different values (mean: 0.53 mm)○ Pixel spacing: 12 | ○ 61 ears (53 patients)○ 4 different CT, CBCT models○ Slice spacing: 10 different values (mean: 0.55 mm)○ Pixel spacing: 24 values (mean: 0.37 mm) | ○ 63 ears (52 patients)○ 5 different MR models○ Slice spacing: 7 different values (mean: 0.52 mm)○ Pixel spacing: 15 values (mean: 0.44 mm) |
CASCINATION AG
{12}
Section 05 – 510(k) Summary - K242120
Page: 9/13
| Variable | Temporal Bone | | Skin | | Inner Ear | | Cochlea | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Thickness Mapping | 3D Reconstruction | Thickness Mapping | 3D Reconstruction | Scala Tympany | Cochlea, Semi-circular canals, internal auditory canal | Cochlea | MR |
| | ○ Pixel spacing: 19 values (mean: 0.27 mm) | (mean: 0.31 mm) | (mean: 0.27 mm) | (mean: 0.31 mm) | Contrast Imaging (SR-PCI) with 9 μm resolution derived from same 75 cochleae | values (mean: 0.29 mm) | values (mean: 0.34 mm) | |
| Image Sites | Pooled (4 clinical sites) | | Pooled (4 clinical sites) | | Pooled (1 clinical site) | Pooled (4 clinical sites) | | |
| Ground Truth Process | Thickness manually measured at 5 locations on each CT image by three surgeons. | Three surgeons annotated each CT slice using 3D Slicer. For each ear, binary masks were generated per slice. | Thickness manually measured at 5 locations on each CT image by three surgeons. | Three surgeons annotated each CT slice using 3D Slicer. For each ear, binary masks were generated per slice. | Binary segmentation masks were created for each sample and independently reviewed for accuracy by three experienced otologists. | Three surgeons annotated the entire inner ear slice by slice using 3D Slicer, generating binary masks for each slice per ear. | The inner ear was annotated slice by slice by three surgeons using 3D Slicer, with binary masks exported for each slice per ear. | The cochlear parameters were manually measured in each ear by three experienced surgeons. |
| Test Dataset | Algorithm over the Testing Image Data | | Algorithm over the Testing Image Data | | Algorithm over the Testing Image Data | Algorithm over the Testing Image Data | | Algorithm over the Testing Image Data |
CASCINATION AG
{13}
Section 05 – 510(k) Summary - K242120
Page: 10/13
| Variable | Temporal Bone | | Skin | | Inner Ear | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Thickness Mapping | 3D Reconstruction | Thickness Mapping | 3D Reconstruction | Scala Tympany | Cochlea, Semi-circular canals, internal auditory canal | | Cochlea | |
| Acceptance Criteria and Test Results | Acceptance Criteria:
○ Mean Absolute Difference (MAD) ≤ 0.6 mm
○ 95% Confidence Interval (CI) upper limit ≤ 0.8 mm | Acceptance Criteria:
○ Mean DICE coefficient ≥ 0.85
○ 95% Confidence Interval (CI) lower limit ≥ 0.85 | Acceptance Criteria:
○ Mean Absolute Difference (MAD) ≤ 0.6 mm
○ 95% Confidence Interval (CI) upper limit ≤ 0.8 mm | Acceptance Criteria:
○ Mean DICE coefficient ≥ 0.68
○ 95% Confidence Interval (CI) lower limit ≥ 0.68 | 3D Reconstruction | | 3D Reconstruction | | Cochlear Parameters |
| | | | | | CT | MR | CT | MR | |
| Acceptance Criteria and Test Results | Acceptance Criteria:
○ Mean Absolute Difference (MAD) ≤ 0.6 mm
○ 95% Confidence Interval (CI) upper limit ≤ 0.8 mm | Acceptance Criteria:
○ Mean DICE coefficient ≥ 0.85
○ 95% Confidence Interval (CI) lower limit ≥ 0.85 | Acceptance Criteria:
○ Mean DICE coefficient ≥ 0.68
○ 95% Confidence Interval (CI) lower limit ≥ 0.68 | Acceptance Criteria:
○ Mean DICE coefficient ≥ 0.65
○ 95% CI lower limit ≥ 0.65 | Acceptance Criteria:
○ Mean DICE coefficient ≥ 0.80
○ 95% Confidence Interval (CI) lower limit ≥ 0.80 | Acceptance Criteria:
○ Mean DICE coefficient ≥ 0.80
○ 95% Confidence Interval (CI) lower limit ≥ 0.80 | Acceptance Criteria:
○ Mean absolute error (MAE) CDLoc measurement ≤ 1.5 mm | Acceptance Criteria:
○ Mean absolute error (MAE) CDLoc measurement ≤ 1.5 mm | Acceptance Criteria:
○ Mean absolute error (MAE) CDLoc measurement ≤ 1.5 mm |
| | Results:
○ MAD: 0.17–0.20 mm
○ CI: 0.19–0.22 | Results:
DICE coefficient:
○ R1: 0.88 [CI: 0.87–0.89]
○ R2: 0.86 [CI: 0.85–0.87]
○ R3: 0.89 [CI: 0.88–0.90] | Results:
○ MAD: 0.21–0.23 mm
○ CI: 0.23–0.26 | Results:
DICE coefficient:
○ R1: 0.89 [CI: 0.88–0.90]
○ R2: 0.87 [CI: 0.86–0.88]
○ R3: 0.86 [CI: 0.84–0.88] | Results:
DICE coefficient:
○ R1: 0.82 [CI: 0.81–0.83]
○ R2: 0.84 [CI: 0.83–0.85]
○ R3: 0.85 [CI: 0.84–0.86] | Results:
DICE coefficient:
○ R1: 0.81 [CI: 0.80–0.82]
○ R2: 0.83 [CI: 0.82–0.84]
○ R3: 0.84 [CI: 0.83–0.85] | Results:
MAE (±SD) for CDLoc:
○ R1: 0.59 ± 0.37 mm
○ R2: 0.64 ± 0.44 mm
○ R3: 0.62 ± 0.39 mm | Results:
MAE (±SD) for CDLoc:
○ R1: 0.56 ± 0.42 mm
○ R2: 0.70 ± 0.39 mm
○ R3: 0.64 ± 0.43 mm | Results:
MAE (±SD) for CDLoc:
○ R1: 0.65 ± 0.43 mm
○ R2: 0.70 ± 0.39 mm
○ R3: 0.64 ± 0.43 mm |
Note: * Algorithm not trained on a dataset. Use established segmentation methods that don't require training.
CASCINATION AG
{14}
Section 05 - 510(k) Summary - K242120
Table 3
Summary of Performance Tests - CT-CT and CT-MR Image Fusion
| Variable | Image Fusion | | | |
| --- | --- | --- | --- | --- |
| | CT-CT | | CT-MR | |
| Nature of Algorithm | A rigid registration algorithm is used to optimize mutual information between two image volumes, yielding a transformation matrix for alignment. | | | |
| Testing Summary | | | | |
| Aim | Validate Accuracy Using Semitones | Validate Accuracy Using Landmark Point Distances | Validate Accuracy Using Semitones | Validate Accuracy Using Landmark Point Distances |
| Testing Image Data | ○ 32 temporal bones (32 patients)○ 2 different CT models○ Pre-op CT:○ Slice spacing: 7 values (mean: 0.41 mm)○ Pixel spacing: 11 values (mean: 0.20 mm)○ Post-op CT:○ Slice spacing: 10 values (mean: 0.50 mm)○ Pixel spacing: 9 values (mean: 0.21 mm) | ○ 32 temporal bones (32 patients)○ 2 different CT models○ Pre-op CT:○ Slice spacing: 7 values (mean: 0.41 mm)○ Pixel spacing: 11 values (mean: 0.20 mm)○ Post-op CT:○ Slice spacing: 10 values (mean: 0.50 mm)○ Pixel spacing: 9 values (mean: 0.21 mm) | ○ 31 temporal bones (25 patients)○ CT: 4 models○ Slice spacing: 12 values (mean: 0.65 mm)○ Pixel spacing: 17 values (mean: 0.21 mm)○ MR: 4 models○ Slice spacing: 9 values (mean: 0.52 mm)○ Pixel spacing: 11 values (mean: 0.33 mm) | ○ 31 temporal bones (25 patients)○ CT: 4 models○ Slice spacing: 12 values (mean: 0.65 mm)○ Pixel spacing: 17 values (mean: 0.21 mm)○ MR: 4 models○ Slice spacing: 9 values (mean: 0.52 mm)Pixel spacing: 11 values (mean: 0.33 mm) |
| Ground Truth Process | Cochlear parameters were manually measured by three experienced surgeons. Electrode contact positions were defined, and the software calculated the insertion metrics and frequency allocation. | 3D coordinates of points were manually measured on each post-operative image by 3 experienced surgeons. | Cochlear parameters were manually measured by three experienced surgeons. Electrode contact positions were defined, and the software calculated the insertion metrics and frequency allocation. | 3D coordinates of points were manually measured on each post-operative image by 3 experienced surgeons. |
| Image Sites | Pooled (4 clinical sites) | | | |
| Test Dataset | Algorithm over the Testing Image Data | Algorithm over the Testing Image Data | Algorithm over the Testing Image Data | Algorithm over the Testing Image Data |
| Acceptance Criteria and Test Results | Acceptance Criteria:○ Maximum mean absolute semitone error per electrode contact < 7.0 semitones | Acceptance Criteria:○ Mean point distance error at each anatomical landmark per rater must be < 0.88 mm | Acceptance Criteria:○ Maximum mean absolute semitone error per electrode contact < 7.0 semitones | Acceptance Criteria:Mean point distance error at each anatomical landmark per rater must be < 1.25 mm |
CASCINATION AG
{15}
Section 05 – 510(k) Summary - K242120
Page: 12/13
| Variable | Image Fusion | | | |
| --- | --- | --- | --- | --- |
| | CT-CT | | CT-MR | |
| | Results:
Max semitone error (per rater):
○ R1: 5.34
○ R2: 4.43
○ R3: 4.20 | Results:
○ RWP: 0.49–0.51 mm
○ LWP: 0.53–0.66 mm
○ IWP: 0.47–0.52 mm
○ SWP: 0.42–0.53 mm | Results:
Max semitone error (per rater):
○ R1: 3.94
○ R2: 3.90
○ R3: 3.97 | Results:
○ RWP: 0.82–0.84 mm
○ LWP: 0.68–0.85 mm
○ IWP: 0.63–0.74 mm
SWP: 0.63–0.76 mm |
## Development (Training) and Test Dataset
The data from the different sites were pooled based on a prior review to confirm consistency in key image acquisition parameters per validated feature. The data was then separated into a development dataset and validation. The dataset used for algorithm development is entirely separate from the dataset used for performance testing. Prior development, the available data was systematically divided into distinct development and test datasets. At no point was data from the test dataset used during algorithm development.
CASCINATION AG
{16}
Section 05 – 510(k) Summary - K242120
Page: 13/13
## iii. Internal Test Standards
Internal test protocols were executed and documented in test reports to demonstrate performance characteristics of OTOPLAN. This included tests with data sets with known dimensions which were loaded into OTOPLAN and results compared to the know dimension. All test results were in the expected range. The internal tests demonstrate that the subject device can fulfill the expected performance characteristics, and no questions of safety or performance were raised.
## iv. Human Factors and Usability Validation
Human factors and usability validation was carried out with the predicate device according to the FDA guidance “Applying Human Factors and Usability Engineering to Medical Devices – Guidance for Industry and Food and Drug Administration Staff (2016-02)” and international standard “AAMI / ANSI / IEC 62366-1:2015, Medical Devices - Part 1: Application of Usability Engineering to Medical Devices”. An Impact assessment, comparing the subject device to the predicate device, was carried out to identify hazard-related use scenarios that require additional summative evaluation testing. The additional summative usability testing included 20 and 15 U.S.-based participants from each user group. The additional summative usability testing was successfully completed, demonstrating that the subject device is safe and effective for its intended users, uses, and use environments.
## v. Clinical Studies
Clinical testing was not required to demonstrate the safety and effectiveness of OTOPLAN. This conclusion is based upon a comparison of intended use, technological characteristics, and non-clinical performance data (Software Verification and Validation Testing, Human Factors and Usability Validation, and Internal Test Standards).
## VIII. CONCLUSIONS
The subject device is substantially equivalent to the predicate device. This conclusion is based upon a comparison of the intended use, technological characteristics, and benchtop testing. An impact assessment on usability, comparing the subject device to the predicate device, and summative usability testing were carried out and demonstrated that the subject device is as safe and effective for the intended users, uses, and use environments as the cited predicate device. Each different technological characteristic has been addressed by a specific performance test. The performance testing demonstrates that the different technological characteristics do not adversely affect the performance of the subject device; therefore, it can be concluded that the subject device is substantially equivalent to the predicate device.
CASCINATION AG
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.