The Xpert® Xpress CoV-2 plus test, performed on the GeneXpert® Xpress System, is a rapid real-time RT-PCR test intended for the qualitative detection of SARS-CoV-2 RNA in nasopharyngeal and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. The Xpert Xpress CoV-2 plus test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens. Negative results do not preclude SARS-CoV-2 infection. The results of this test should not be used as the sole basis for diagnosis and patient management decisions.
Device Story
The Xpert Xpress CoV-2 plus test is an automated in vitro diagnostic assay for qualitative detection of SARS-CoV-2 viral RNA. Input: nasopharyngeal or anterior nasal swab specimens in VTM/UTM or eNAT media. Operation: user transfers sample to a single-use, self-contained cartridge; cartridge is loaded into the GeneXpert Xpress System. The system performs automated RNA extraction, reverse transcription, and real-time RT-PCR amplification of three SARS-CoV-2 targets (E, N2, RdRP). The system includes an Early Assay Termination (EAT) function for high-titer specimens, reducing time to result to ~19 minutes. Output: qualitative detection results displayed on the system touchscreen. Used in clinical settings by healthcare providers to aid in COVID-19 diagnosis. Benefits: rapid, automated, integrated sample-to-result workflow minimizing cross-contamination risk.
Clinical Evidence
Prospective clinical study of 1,783 symptomatic subjects (883 NPS, 900 ANS). Performance compared to an FDA-cleared molecular respiratory panel. For NPS: PPA 98.2% (95% CI: 93.8-99.5%), NPA 99.1% (95% CI: 98.1-99.6%). For ANS: PPA 99.0% (95% CI: 94.8-99.8%), NPA 99.1% (95% CI: 98.2-99.6%). Final non-determinate rate was 0.6%.
Technological Characteristics
Real-time RT-PCR assay; targets SARS-CoV-2 E, N2, and RdRP genes. Single-use disposable cartridges with integrated sample processing, nucleic acid extraction, and amplification. Internal controls: Sample Processing Control (SPC) and Probe Check Control (PCC). Connectivity: GeneXpert Xpress System with integrated barcode scanner and touchscreen. Software: GeneXpert Xpress software v6.4a.
Indications for Use
Indicated for qualitative detection of SARS-CoV-2 RNA in nasopharyngeal and anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Intended as an aid in COVID-19 diagnosis in conjunction with other clinical, epidemiologic, and laboratory findings.
Regulatory Classification
Identification
A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.
Predicate Devices
Xpert Xpress CoV-2 plus, Performed on the GeneXpert Instrument Systems (K230440)
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K242109
B Applicant
Cepheid
C Proprietary and Established Names
Xpert Xpress CoV-2 plus
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QQX | Class II | 21 CFR 866.3981 - Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The SARS-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
To obtain 510(k) clearance for the Xpert Xpress CoV-2 plus test for use on the GeneXpert Xpress System.
### B Measurand:
Conserved RNA sequences within the genes encoding the nucleocapsid protein (N), envelope protein (E) and RNA-dependent RNA polymerase protein (RdRP) of SARS-CoV-2 viruses.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
K242109 - Page 2 of 20
C Type of Test:
A multiplexed, real-time reverse transcriptase polymerase chain reaction (RT-PCR) assay for the qualitative detection and differentiation of SARS-CoV-2 viral RNA from nasopharyngeal swab (NPS) specimens and anterior nasal swab (ANS) specimens using the GeneXpert Instrument Systems platform.
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Xpert Xpress CoV-2 plus test, performed on the GeneXpert Xpress System, is a rapid real-time RT-PCR test intended for the qualitative detection of SARS-CoV-2 RNA in nasopharyngeal and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection.
The Xpert Xpress CoV-2 plus test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens.
Negative results do not preclude SARS-CoV-2 infection. The results of this test should not be used as the sole basis for diagnosis and patient management decisions.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in vitro diagnostic use only
D Special Instrument Requirements:
GeneXpert Xpress System
IV Device/System Characteristics:
A Device Description:
The Xpert Xpress CoV-2 plus test, performed on the GeneXpert Xpress System, is an automated in vitro diagnostic test for the qualitative detection and differentiation of SARS-CoV-2 viral RNA in nasopharyngeal swab (NPS) and anterior nasal swab (ANS) specimens collected from individuals showing signs and symptoms of respiratory viral infection. The test is performed with three steps: 1) transfer liquid sample to the test cartridge with the transfer pipette (provided in the test kit), 2) run the test on the GeneXpert Xpress System, and 3) read the results.
{3}
The Hub configuration of the GeneXpert Xpress System includes a GeneXpert IV Instrument, an integrated barcode scanner, and a computer with a touchscreen and preloaded GeneXpert Xpress software designed for running tests and viewing results. The system requires the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences.
The Xpert Xpress CoV-2 plus test cartridge includes all the necessary reagents for the detection of SARS-CoV-2 viral RNA from NPS and ANS specimens collected with nylon flocked swabs and placed into either 3 mL viral transport medium (VTM)/universal transport medium (UTM) or 2 mL eNAT. With a supplied transfer pipette, eluted NPS and ANS swab specimens are loaded into the sample chamber of a single-use, self-contained Xpert Xpress CoV-2 plus test cartridge. Each Xpert Xpress CoV-2 plus test cartridge contains separate chambers for sample loading, sample processing, and target amplification by real-time RT-PCR, and includes all the reagents necessary to carry out these processes.
The primers and probes in the Xpert Xpress CoV-2 plus test are designed to amplify and detect gene sequences for the following:
- SARS-CoV-2 (envelope small membrane protein [E], a sequence in the N nucleocapsid protein [N2] and RNA-dependent RNA polymerase [RdRP]).
A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functioning as expected. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability. Because the cartridges are self-contained, and specimens never contact working parts of the instrument modules, cross-contamination between samples is minimized. The test cartridge containing the patient sample is loaded onto one of four randomly accessible GeneXpert instrument modules, which performs fully automated and integrated sample preparation and real-time RT-PCR for the Xpert Xpress CoV-2 plus test in approximately 30 minutes.
The Xpert Xpress CoV-2 plus test includes an early assay termination (EAT) function which will provide earlier time to results in high titer specimens if the signal from the target nucleic acid reaches a predetermined threshold before all assay cycles have been completed. When the EAT feature is activated, the earliest time to receive a positive result is calculated to be approximately 19 minutes.
## B Principle of Operation:
The Xpert Xpress CoV-2 plus test is a nucleic acid-based test using real-time RT-PCR. After addition of the specimen to the Sample Chamber of the Xpert Xpress CoV-2 plus test cartridge, the cartridge is loaded onto the GeneXpert Xpress System platform. The instrument then performs automated sample processing including RNA extraction, followed by reverse
K242109 - Page 3 of 20
{4}
transcription, amplification, detection, and reporting of results. The results are interpreted automatically by the GeneXpert System and are shown in the View Results window.
## C Instrument Description Information:
1. Instrument Name:
GeneXpert Xpress System
2. Specimen Identification:
Sample ID is a unique identifier that links the sample being processed to the patient that provided the specimen. The Sample ID can be entered into the GeneXpert Xpress System either by scanning a barcode, entering the Sample ID manually using the virtual keyboard or having the system assign a random Sample ID.
To perform a test, the user selects "NEW TEST" icon from the Home Screen. Either patient information must be entered if configured by an administrator or the Sample ID screen appears. If the Patient Information screen appears, manually enter or scan the Patient ID barcode. If the Sample ID screen appears, scan the Sample ID barcode or manually enter the Sample ID for the patient specimen. The user is then instructed to scan the cartridge barcode and select the test to run. The prepared test cartridge is loaded into an available instrument module with the flashing green light to initiate the test.
3. Specimen Sampling and Handling:
The Xpert Xpress CoV-2 plus test cartridge includes reagents for the detection of viral RNA from SARS-CoV-2 in nasopharyngeal swab (NPS) and anterior nasal swab (ANS) specimens. The specimens are collected with a nylon flocked swab and placed into a transport tube containing 3 mL of viral transport medium (VTM)/Universal Transport Medium (UTM) or 2 mL of eNAT. At the test facility, the operator mixes the specimen by rapidly inverting the collection tube 5 times. A disposable transfer pipette supplied with the Xpert Xpress CoV-2 plus is used to transfer an aliquot of the specimen into the Sample Chamber of the Xpert Xpress CoV-2 plus test cartridge. After closing the cartridge lid, the GeneXpert cartridge is loaded onto the GeneXpert Xpress instrument, which automates and integrates sample preparation, nucleic acid extraction and amplification, and detection of target sequences in simple or complex samples using real-time PCR and RT-PCR assays.
4. Calibration:
Calibration of the GeneXpert Xpress instrument is performed at the factory, and recalibration in the field at customer sites is not required during the initial system startup. A GeneXpert operator or Field Service Engineer with Administrator user permissions can perform calibration checks during annual maintenance.
5. Quality Control:
Internal Controls
K242109 - Page 4 of 20
{5}
The GeneXpert Xpress System relies on the built-in internal controls included in each Xpert Xpress CoV-2 plus test cartridge to ensure the test system is performing as expected. These internal controls consisted of Sample Processing Control (SPC) and Probe Check Control (PCC).
Sample Processing Control (SPC) – The SPC is an encapsulated RNA pseudovirus that ensures adequate lysis of target virus, monitors the presence of PCR inhibitors, and verifies the use of proper PCR conditions. The SPC passes if it meets the validated acceptance criteria. The SPC result should be PASS in a sample that is negative for all three SARS-CoV-2 target analytes and can be NEGATIVE or POSITIVE in a sample containing detectable levels of one or more of the SARS-CoV-2 target analytes.
Probe Check Control (PCC) – Before the start of the PCR reaction, the GeneXpert System measures the fluorescence signal from the probes to monitor bead rehydration, reaction tube filling, probe integrity, and dye stability. The PCC passes if it meets the validated acceptance criteria. If any of the PCC conditions fails, the result is reported as NO RESULT – REPEAT TEST and the test must be repeated using a new test cartridge.
## External Controls
External controls (ECs) are not required for the end users to obtain a valid Xpert test result. External controls are not provided with the Xpert Xpress CoV-2 plus test and are available from Zeptometrix. Specifically, the following external controls have been validated for use with the Xpert Xpress CoV-2 plus test:
- External Positive Control: NATtrol SARS-Related Coronavirus 2 (SARS-CoV-2); Catalog# NATSARS(COV2)-ERC-IVD
- External Negative Control: NATtrol SARS Associated Coronavirus 2 (SARS-CoV-2); Catalog# NATSARS(COV2)-NEG-IVD
All external controls must be used in accordance with local, state, and federal accrediting organizations, as applicable.
It is recommended that external controls be tested following the manufacturer’s instruction at the frequency noted below.
- Each time a new lot of Xpert Xpress CoV-2 plus kits is received.
- Each time a new shipment of Xpert Xpress CoV-2 plus kits is received even if it is the same lot as previously received.
- Each time a new operator is performing the test (i.e., operator who has not performed the test before).
- When problems (storage, operator, instrument, or other) are suspected or identified.
- If otherwise required by your institution’s standard Quality Control (QC) procedures.
## V Substantial Equivalence Information:
K242109 - Page 5 of 20
{6}
K242109 - Page 6 of 20
A Predicate Device Name(s):
Xpert Xpress CoV-2 plus, performed on the GeneXpert Instrument System
B Predicate 510(k) Number(s):
K230440
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K242109 | K230440 |
| --- | --- | --- |
| Device Trade Name | Xpert Xpress CoV-2 plus, Performed on the GeneXpert Xpress System | Xpert Xpress CoV-2 plus, Performed on the GeneXpert Instrument Systems |
| Intended Use/Indications for Use | The Xpert Xpress CoV-2 plus test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens. Negative results do not preclude SARS-CoV-2 infection. The results of this test should not be used as the sole basis for diagnosis and patient management decisions. | The Xpert Xpress CoV-2 plus test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens. Negative results do not preclude SARS-CoV-2 infection. The results of this test should not be used as the sole basis for diagnosis and patient management decisions. |
| Regulation Name | Same | 21 CFR 866.3981 Devices to detect and identify nucleic acid targets in respiratory samples from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-analyte test |
| Product Code | Same | QQX Respiratory Specimen Nucleic Acid SARS-CoV-2 Test |
| Technology/ Detection | Same | Real-time reverse transcription polymerase chain reaction (RT-qPCR) |
| Assay Targets | Same | SARS-CoV-2 (E, N2, RdRP) |
| Specimen Type | Same | Nasopharyngeal swab (NPS) Anterior nasal swab (ANS) |
| Transport Media | Same | Universal Transport Medium (UTM)/Viral Transport Medium (VTM) eNAT |
| Test Format | Same | Single Use |
| Automation | Same | Automated Nucleic Acid Extraction, Detection and Results Interpretation |
| Assay Results | Same | Qualitative |
{7}
| Non-Determinate Results Test | NO RESULT - REPEAT TEST INSTRUMENT ERROR | INVALID ERROR NO RESULT |
| --- | --- | --- |
| Internal Control | Same | • Sample Processing Control (SPC)
• Probe Check Control (PCC) |
| Time to Result | Same | ~30 minutes |
| Instrument Systems | Cepheid GeneXpert Xpress System | Cepheid GeneXpert Instrument Systems (Dx and Infinity) |
# VI Standards/Guidance Documents Referenced:
Consensus Standards
| FDA Recognition Number | Document Number | Title | Application of Standards |
| --- | --- | --- | --- |
| 5-125 | ISO 14971:2019 | Medical devices — Application of risk management to medical devices | General Use |
| 7-233 | CLSI EP17-A2 | Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline | General Use |
| 7-251 | CLSI EP05-A3 | Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition | General Use |
| 7-275 | CLSI EP07 - A3 | Interference Testing in Clinical Chemistry; Approved Guideline – Third Edition. | General Use |
| 7-152 | CLSI EP12-A2 | User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition | General Use |
| 19-42 | IEC 61326-1 Edition 3.0 2020-10 | Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements | General Use |
| 19-43 | IEC 61326-2-6 Edition 3.0 2020-10 | Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment | General Use |
| 13-79 | IEC 62304 Edition 1.1 2015-06 | Medical device software - Software life cycle processes | Declaration of Conformity |
| 19-36 | IEC 60601-1-2 Edition 4.1 2020-09 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | General Use |
FDA Special Controls
K242109 - Page 7 of 20
{8}
The Xpert Xpress CoV-2 plus test was developed in accordance with the special controls for 21 CFR 866.3981 as detailed in the reclassification order DEN200031.
## Guidance Documents
| Document |
| --- |
| Guidance for Industry and Staff – Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices, issued on February 26, 2020. |
| Guidance for Industry and FDA Staff – Recommendations for Dual 510(k) and CLIA Waiver by Application Studies, issued February 26, 2020. |
| Guidance for Industry and FDA Staff – Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices, issued September 14, 2018 |
| Guidance for Industry and FDA Staff – Design Considerations and Premarket Submission Recommendations for Interoperable Medical Devices, issued September 6, 2017. |
| Guidance for Industry and FDA Staff, Content of Premarket Submissions for Device Software Functions, issued June 14, 2023 |
| Guidance for Industry and FDA Staff, Electromagnetic Compatibility (EMC) of Medical Devices, issued on June 6, 2022 |
| Guidance for Industry – Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software, issued January 14, 2005. |
| Guidance for Industry and FDA Staff, Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued on September 27, 2023 |
| Guidance for Industry and FDA Staff, General Principles of Software Validation, issued January 11, 2002 |
| Guidance for Industry and FDA Staff – Electronic Submission Template for Medical Device 510(k) Submission, issued October 2, 2023. |
| Guidance for Test Developers and FDA Staff – Policy for Evaluating Impact of Viral Mutations on COVID-19 Tests (Revised); issued January 12, 2023. |
| Guidance for Industry and FDA Staff – Transition Plan for Medical Devices Issued Emergency Use Authorizations (EUAs) Related to Coronavirus Disease 2019 (COVID-19), issued March 2023 |
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
The studies conducted to support the analytical performance of the Xpert Xpress CoV-2 plus test, performed on the GeneXpert Instrument Systems, were previously reviewed and described in K230440. In this submission, analytical studies data generated on the GeneXpert Instrument Systems were reanalyzed using the latest 6.4a GeneXpert Xpress software to assess the analytical performance on the GeneXpert Xpress System.
The results from the analysis are described in sections below.
1. Precision / Reproducibility
a. Within-laboratory Precision:
K242109 - Page 8 of 20
{9}
Within-laboratory precision was previously evaluated as described in K230440. A total of 281 test results (240 samples and 41 controls) were processed using the GeneXpert Xpress 6.4a software. In all cases, the reanalysis gave the same results as described in K230440.
The reanalysis of the within-laboratory precision study data from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System.
# b. Reproducibility:
A blinded, multi-site reproducibility study was conducted to assess the total variability of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System across operators, study sites, testing days, runs and instruments. The reproducibility study design incorporated contrived panel members as indicated in Table 1.
Table 1. Reproducibility Panel
| Panel Member | Target | Level |
| --- | --- | --- |
| 1 | Negative | Negative |
| 2 | SARS-CoV-2 | ~1.5x LoD |
| 3 | SARS-CoV-2 | ~3x LoD |
Panel members were contrived using inactivated NATtrol SARS-CoV-2 (ZeptoMetrix, Buffalo, NY, catalog number NATSARS(COV2)-ERC) in simulated nasal/NP swab matrix. The "Negative" sample did not contain any target microorganism or target RNA. The study was evaluated at three (3) external CLIA-Waived sites with three (3) operators at each site. The study was conducted with one (1) lot of Xpert Xpress CoV-2 plus cartridges over five (5) independent days of testing per operator (may be non-consecutive) with one (1) run per operator per day, with a run consisting of two (2) replicates per panel member. Study results are summarized in Table 2.
Table 2. Summary of Reproducibility Results by Site and Operator
| Sample | Site 1 | | | | Site 2 | | | | Site 3 | | | | Agreement (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | OP1 | OP2 | OP3 | Site | OP1 | OP2 | OP3 | Site | OP1 | OP2 | OP3 | Site | |
| Negative | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% (90/90; 95.9-100) |
| | 10/10 | 10/10 | 10/10 | 30/30 | 10/10 | 10/10 | 10/10 | 30/30 | 10/10 | 10/10 | 10/10 | 30/30 | |
| SARS-CoV-2 Low Pos | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% (90/90; 95.9-100) |
| | 10/10 | 10/10 | 10/10 | 30/30 | 10/10 | 10/10 | 10/10 | 30/30 | 10/10 | 10/10 | 10/10 | 30/30 | |
| SARS-CoV-2 Mod Pos | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% (90/90; 95.9-100) |
| | 10/10 | 10/10 | 10/10 | 30/30 | 10/10 | 10/10 | 10/10 | 30/30 | 10/10 | 10/10 | 10/10 | 30/30 | |
$\mathrm{OP} =$ Operator
The Xpert Xpress CoV-2 plus test demonstrated $100\%$ agreement with expected results for all panel members. The results of the study demonstrate acceptable assay reproducibility for the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System.
K242109 - Page 9 of 20
{10}
# Non-determinate Rate
Of the 270 samples included in the Reproducibility study, 12 samples yielded non-determinate results upon initial test. Table 3 shows the list of samples with test results from both initial test and re-test. All non-determinate results were resolved upon retest.
Table 3. Samples with Non-determinate Results
| Reason for Failure | Number of Samples | Number of Samples Resolved on Retest |
| --- | --- | --- |
| SPC failed | 1 | 1 |
| Module was reset | 2 | 2 |
| Insufficient volume | 5 | 5 |
| Probe check failure | 2 | 2 |
| Cartridge integrity test failure | 2 | 2 |
| Total | 12 | 12 |
# 2. Analytical Reactivity (Inclusivity):
# In silico Analysis of Inclusivity
In silico SARS-CoV-2 analysis was previously reviewed and described in K230440. Additionally, the inclusivity of Xpert Xpress CoV-2 plus was evaluated using in-silico analysis of the assay amplicons and probes against the 2,691,091 SARS-CoV-2 sequences available in the GISAID gene database, for variants that were circulating between August $1^{\text{st}}$ , 2022 and September $4^{\text{th}}$ , 2023 for three SARS-CoV-2 gene targets, E, N2 and RdRP. The 2,691,091 SARS-CoV-2 sequences were separated into the lineages of interest based on the Pango Lineage assigned to each genome by GISAID, and those with ambiguous nucleotides were removed. The inclusivity analyses focused on the combined, non-ambiguous sequences from the variants of interest and variants of concern between August $1^{\text{st}}$ , 2022 and September $4^{\text{th}}$ , 2023. These constituted 2,519,206 sequences for the E target, 2,551,690 sequences for the N2 target and 2,476,229 sequences for the RdRP target. Table 4 shows the number of sequences in each amplicon alignment and summarizes the effective predicted inclusivity for E, N2 and RdRP amplicons for the variants of interest and concern.
Table 4. Predicted Inclusivity for E, N2 and RdRP Amplicons for SARS-CoV-2 Variants of Interests and Concern
| SARS-CoV-2 Target Amplicon | Exact Match | 1 Mismatcha | 2 or More Mismatches | Predicted Inclusivity |
| --- | --- | --- | --- | --- |
| E | 2,503,319 of 2,519,206 total (99.4%) | 15,694 (0.6%) | 193 (0.007%) | 100% |
| N2 | 2,495,212 of 2,551,690 total (97.8%) | 55,765 (2.2%) | 713 (0.028%) | 99.95% |
| RdRP | 2,448,754 of 2,476,229 total (98.9%) | 27,322 (1.1%) | 153 (0.006%) | 100% |
a. Single-nucleotide mismatches are predicted to not impact the performance of the test.
The in-silico inclusivity of the Xpert Xpress CoV-2 plus probe oligonucleotides for E, N2 and RdRP was also assessed against the top 20 most frequent matches in the GISAID EpiCoV sequence database between August $1^{\text{st}}$ , 2022 and September $4^{\text{th}}$ , 2023, which constituted 2,519,206 for the E target, 2,551,690 for the N2 target and 2,476,229 for the
K242109 - Page 10 of 20
{11}
RdRP target. For each of the probe oligonucleotides used in the Xpert Xpress CoV-2 plus test, Table 5 summarizes the number sequences as well as the corresponding percentage of sequences from this dataset with exact match, 1 mismatch/insertion, and 2 or more mismatches/insertions in the alignment.
Table 5. Predicted Inclusivity for E, N2 and RdRP Probes for SARS-CoV-2 Variants of Interests and Concern
| SARS-CoV-2 Target Probe | Exact Match | 1 Mismatch/ Insertiona | 2 or More Mismatches/ Insertions | Predicted Inclusivity |
| --- | --- | --- | --- | --- |
| E | 2,516,153 of 2,519,206 total (99.9%) | 2997 (0.1%) | 0 (0.0%) | 100% |
| N2 | 2,528,096 of 2,551,690 total (99.1%) | 12728 (0.5%) | 0 (0%) | 99.6% |
| RdRP | 0 | 2,465,482 of 2,476,229 total (99.6%) | 10418 (0.4%) | 99.6% |
a Single-nucleotide mismatches are predicted to not impact the performance of the test.
Based on the built-in redundancy of the Xpert Xpress CoV-2 plus test's SARS-CoV-2 amplification system (i.e., three independent targets, only one of three must be detected to assign a positive result), all evaluated sequences are predicted to be detected by the Xpert Xpress CoV-2 plus test.
# Wet-Testing
Inclusivity wet-testing was previously evaluated as described in K230440. Two hundred twenty-seven (227) runs were previously evaluated for the inclusivity study for K230440. Out of the 227 runs, 20 tests were not able to be reanalyzed due to software incompatibility. Two hundred and seven tests from the analytical inclusivity study were reanalyzed. There were 206 valid results and one non-determinate result. All 206 valid tests gave the exact same test results after reanalysis. The number of non-determinate (ND) GeneXpert results was unchanged after reanalysis and the ND test results descriptions remained the same.
The reanalysis of the data from the inclusivity study from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System
# 3. Analytical Specificity/Interference
Analytical Specificity (Cross-Reactivity)
# Cross-Reactivity Wet-Testing
Analytical specificity (cross-reactivity) was previously evaluated as described in K230440. One hundred twenty-five (125) tests from the analytical specificity study were reanalyzed using the latest 6.4a GeneXpert Xpress software. There were 125 valid results and zero (0) non-determinate (ND) results. All 125 valid tests gave exactly the same test results after reanalysis.
The reanalysis of the analytical specificity (cross-reactivity) study data from K230440 demonstrates equivalent performance of the Xpert Xpress Cov-2 plus test on the GeneXpert Xpress System.
K242109 - Page 11 of 20
{12}
K242109 - Page 12 of 20
# Interference Testing
## Microbial Interference
Microbial interference was previously evaluated as described in K230440. Three hundred twenty (320) tests were reanalyzed using the latest 6.4a GeneXpert Xpress software. There were 312 valid results and eight (8) non-determinate results. All 312 valid tests gave the exact same test results after reanalysis. The number of non-determinate (ND) GeneXpert results was unchanged after reanalysis. The ND test results descriptions remained the same.
The reanalysis of the analytical microbial interference study data from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System.
## Potentially Interfering Substances
The impact of potentially interfering substance on the Xpert Xpress CoV-2 plus test was previously evaluated as described in K230440. Four hundred twenty-four (424) tests from the potentially interfering substances study were reanalyzed using the latest 6.4a GeneXpert Xpress software. There were 421 valid results and three (3) non-determinate results. All 421 valid tests gave the exact same test results after reanalysis. The number of non-determinate GeneXpert results was unchanged after reanalysis and the ND test results descriptions remained the same.
The reanalysis of the data from the interfering substances study from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System.
## 4. Assay Reportable Range:
Not Applicable; this is a qualitative assay.
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### a. Controls
#### External Control Evaluation
Two hundred and fifty (250) tests from the external control evaluation study in K230440 were reanalyzed. Of the 250 runs tested there were 248 valid tests that gave the exact same test results after reanalysis. The number of non-determinate test results remained unchanged.
The reanalysis of the data from the external control validation study from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System.
### b. Specimen Stability
#### NPS in UTM/VTM Matrix
{13}
Two hundred ninety-two (292) tests from the NPS Specimen Stability in the UTM/VTM Matrix Study in K230440 were reanalyzed using the latest 6.4a GeneXpert Xpress software. There were 290 valid results and two (2) non-determinate results. All 290 valid tests gave the exact same test results after reanalysis while the number of non-determinate results was unchanged.
This reanalysis supports specimen storage conditions in NPS UTM/VTM matrix at refrigerated (2-8 °C) temperatures for up to seven (7) days and at room temperature (15-30°C) for up to 48 hours until testing is performed on the GeneXpert Xpress Systems.
## NPS in eNAT Matrix
Two hundred ninety-three (293) tests from the NPS Specimen Stability in eNAT Matrix Study in K230440 were reanalyzed using the latest 6.4a GeneXpert Xpress software. There were 290 valid results and three (3) non-determinate results. All 290 valid tests gave the exact same test results after reanalysis, the number of non-determinate results remained unchanged. The ND test results descriptions remained the same after reanalysis.
This reanalysis supports specimen storage conditions in NPS eNAT matrix at refrigerated (2 - 8 °C) temperatures for up to seven (7) days and at room temperature (15 - 30 °C) for up to 48 hours until testing is performed on the GeneXpert Xpress Systems.
## c. Specimen Stability – Freeze-Thaw
Three hundred seventy-four (374) tests from the NPS-UTM/VTM and NPS-eNAT specimens freeze/thaw study conducted at -80°C in K230440 were reanalyzed using the latest 6.4a GeneXpert Xpress software. There were 366 valid results and eight non-determinate (ND) results. All 366 valid tests gave the exact same test results after reanalysis. The number of non-determinate GeneXpert results was unchanged after reanalysis. The ND test results descriptions remained the same.
This reanalysis supports storage and freeze-thaw stability claims for NPS and NS samples collected in VTM and eNAT that can undergo one cycle of freeze-thaw (-80°C) before testing with the Xpert Xpress CoV-2 plus test.
## d. Prepared Cartridge Hold Time
Two hundred and sixty-six (266) tests from the cartridge hold time study in K230440 were reanalyzed using the latest 6.4a GeneXpert Xpress software. There were 258 valid results and eight non-determinate results. All 258 valid tests gave the exact same test results after reanalysis and the number of non-determinate GeneXpert results was unchanged. The ND test results descriptions remained the same.
The reanalysis of the data from the cartridge hold time study from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System and supports a cartridge hold time of 4.5 hours.
## 6. Detection Limit:
Analytical Sensitivity/ Limit of Detection (LoD)
K242109 - Page 13 of 20
{14}
# LoD in Clinical NPS in UTM/VTM and eNAT Matrix
The LoD in clinical NPS in UTM/VTM was previously established as described in K230440. Eight hundred fifty-two (852) tests from the LoD study for clinical NPS in UTM/VTM matrix and in eNAT were reanalyzed using the latest 6.4a GeneXpert Xpress software. There were 836 valid tests and 16 non-determinate results (ND). All 836 valid tests gave the exact same test results after the reanalysis. The number of non-determinate GeneXpert results was unchanged after reanalysis and the ND descriptions remained the same.
The reanalysis of the data from the LoD in clinical NPS in UTM/VTM matrix studies from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System.
The LoD concentrations and mean Ct values at the verified LoDs for clinical NPS in UTM/VTM or in eNAT specimens tested using two reagent lots are shown in Table 6.
Table 6. Results for Verification of SARS-CoV-2 LoD in Clinical NPS-UTM/VTM and Clinical NPS-eNAT
| Virus | Reagent Lot | Matrix | LoD (copies/ mL)^{a} | Positive Results out of # Valid Replicates | % Positive | Mean E Ct | Mean N2 Ct | Mean RdRP Ct |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| SARS-CoV-2 | Lot 1 | Clinical NPS-UTM/VTM | 403 | 20/20 | 100% | 34.3 | 38.3 | 36.6 |
| | Lot 2 | | | 20/20 | 100% | 34.1 | 37.7 | 36.5 |
| | Lot 1 | Clinical NPS-eNAT | 403 | 20/20^{b} | 100% | 33.4 | 36.8 | 35.5 |
| | Lot 2 | | | 20/20 | 100% | 33.5 | 36.8 | 35.5 |
a Genomic copies/ mL
b One of 20 replicates reported INVALID. The run was successfully repeated to obtain 20 valid replicates.
# LoD in Clinical ANS in UTM/VTM Matrix
The LoD in clinical ANS in UTM/VTM was previously established as described in K230440. Six hundred sixty-six (666) tests from the LoD in clinical NS in UTM/VTM matrix were reanalyzed. There were 652 valid results and 14 non-determinate (ND) results. All 652 valid tests gave the exact same test results after reanalysis and the number of ND results was unchanged. The ND test result descriptions remained the same.
The LoD concentrations and mean Ct values at the verified LoDs for clinical ANS in UTM/VTM specimens tested using two reagent lots are shown in Table 7.
K242109 - Page 14 of 20
{15}
Table 7. Results for Verification of SARS-CoV-2 LoD in Clinical ANS-UTM/VTM
| Virus | Reagent Lot | LoD (copies/mL) | Positive Results out of # of Valid Replicates | % Positive | Mean E Ct | Mean N2 Ct | Mean RdRP Ct |
| --- | --- | --- | --- | --- | --- | --- | --- |
| SARS-CoV-2 | Lot 1 | 462 | 20/20 | 100% | 34.4 | 38.9 | 36.8 |
| | Lot 2 | 462 | 20/20 | 100% | 34.1 | 37.6 | 36.1 |
## LoD in Clinical ANS Matrix – WHO Study
One hundred twenty-two (122) tests from the LoD in clinical ANS matrix study testing the 1st WHO International Standard for SARS-CoV-2 RNA Virus in K230440 were reanalyzed. There were 116 valid results and six (6) non-determinate results. All 116 valid tests gave the exact same test results after reanalysis and the number of non-determinate GeneXpert results was unchanged after reanalysis. The ND test results descriptions remained the same. Table 8 presents the Mean Ct values for gene targets within the WHO international standard for the SARS-CoV-2 virus.
Table 8. LoD Concentration and Mean Ct Values for E, N2 and RdRP for the 1st WHO International Standard for SARS-CoV-2 RNA Virus in Clinical NS-UTM/VTM Matrix
| Virus | Strain/Isolate | LoD Concentration | Mean E Ct | Mean N2 Ct | Mean RdRP Ct |
| --- | --- | --- | --- | --- | --- |
| 1st WHO International Standard for SARS-CoV-2 RNA Virus | England/02/2020 isolate of SARS-CoV-2 | 1000 IU/mL (3.0 Log_{10} IU/mL) | 34.3 | 39.5 | 36.5 |
7. Assay Cut-Off:
The Xpert Xpress CoV-2 plus test includes defined cycle threshold (Ct) ranges for the SARS-CoV-2 targets and SPC. The Ct cut-offs are hard coded values and are included as automatic calculations in the assay definition file (ADF) of the Xpert Xpress CoV-2 plus test. The cut-off values are identical to those reported in K230440.
8. Accuracy (Instrument):
Not Applicable
9. Carry-Over:
Ninety (90) tests from the carry-over contamination study in K230440 were reanalyzed. There were 88 valid results and two (2) non-determinate (ND) results. All 88 valid tests gave the exact same test results after reanalysis with GeneXpert Xpress software version 6.4a. The number of ND GeneXpert results was unchanged after reanalysis. The ND test results descriptions remained the same.
The reanalysis of the data from the carry-over contamination study from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System.
K242109 - Page 15 of 20
{16}
10. Matrix Comparison:
Clinical NPS UTM/VTM, Clinical ANS UTM/VTM and Simulated NPS/ANS UTM/VTM Matrices
Equivalent Xpert Xpress CoV-2 plus performance for the detection of viral targets (SARS-CoV-2 – E, N2, RdRP) between clinical ANS-UTM/VTM matrix, clinical NPS- UTM/VTM matrix and the simulated NPS/ANS UTM/VTM matrix was previously evaluated in K230440.
Five hundred thirteen (513) tests from the specimen matrices equivalency study in K230440 were reanalyzed. There were 503 valid results and ten (10) non-determinate (ND) results. All 503 valid tests gave the exact same test results after reanalysis with GeneXpert Xpress software version 6.4a. The number of non-determinate GeneXpert results remained unchanged after reanalysis and the ND test result descriptions remained the same.
The reanalysis of the data from the specimen matrices equivalency study with clinical NPS and clinical ANS in UTM/VTM, and simulated NPS/ANS in UTM/VTM matrices from K230440 demonstrates equivalent performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System when testing natural NPS or ANS collected in UTM/VTM matrix and simulated NPS/ANS in UTM/VTM matrices.
B Clinical Studies:
1. Clinical Sensitivity:
Prospective Clinical Study
The performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System was established in a multi-site prospective clinical study that included 22 geographically diverse locations in the United States in specimens collected from individuals showing signs and symptoms of respiratory infection. Of the 22 sites, 21 sites performed GeneXpert Xpress System testing and specimen collection while one (1) site performed comparator and discrepant investigation.
Prospectively collected, fresh nasopharyngeal swab (NPS) and anterior nasal swab (ANS) specimens in UTM/VTM, from individuals with signs and symptoms of respiratory viral infection, were enrolled from January 24, 2022 to May 16, 2022 in the US in an all-comers fashion. Xpert Xpress CoV-2 plus testing on the GeneXpert Xpress System was performed from January 24, 2022 to May 16, 2022, at 21 testing sites.
There were 2200 prospective subjects/ specimens from symptomatic individuals enrolled for evaluating the performance of the Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System (1094 NPS and 1106 ANS). Of these, ten (10) subjects/ specimens (five (5) NPS and five (5) ANS) were found to be ineligible as not meeting inclusion/exclusion criteria or having incomplete Informed Consent leaving 2190 eligible samples (1089 NPS and 1101 ANS). From these specimens, an additional 256 (128 NPS and 128 ANS) were excluded due
K242109 - Page 16 of 20
{17}
to protocol deviations leaving 1934 subjects/ specimens (961 NPS and 973 ANS) in the final dataset. There were an additional 151 specimens excluded (NPS = 78 and ANS = 73) due to: unresolved repeat ND for Xpert testing (n = nine (9), unresolved repeat ND and procedural deviation with comparator testing (n = two (2) invalid and 138 non-evaluable comparator results (n = 140). There 1783 samples/ specimens eligible for performance evaluation (NPS = 883 and ANS = 900). Table 9 provides demographic data for the prospective fresh specimens available for performance analyses.
Table 9. Clinical Study Demographics
| Prospectively Collected Fresh Specimens from 2022 (Category I) | NPS (N=961) | ANS (N=973) | Overall (N=1934) |
| --- | --- | --- | --- |
| Gender | | | |
| Female | 569 (59.2%) | 616 (63.3%) | 1185 (61.3%) |
| Male | 392 (40.8%) | 357 (36.7%) | 749 (38.7%) |
| Age Group (Years) | | | |
| ≤5 | 8 (0.8%) | 54 (5.5%) | 62 (3.2%) |
| 6-21 | 275 (28.6%) | 285 (29.3%) | 560 (29.0%) |
| 22-59 | 537 (55.9%) | 505 (51.9%) | 1042 (53.9%) |
| ≥60 | 141 (14.7%) | 129 (13.3%) | 270 (14.0%) |
| Race | | | |
| American Indian or Alaska Native | 0 (0%) | 1 (0.1%) | 1 (0.1%) |
| Asian | 27 (2.8%) | 25 (2.6%) | 52 (2.7%) |
| Black or African American | 225 (23.4%) | 217 (22.3%) | 442 (22.9%) |
| White | 656 (68.3%) | 661 (67.9%) | 1317 (68.1%) |
| Black or African American, White | 4 (0.4%) | 4 (0.4%) | 8 (0.4%) |
| Other Mixed (each N≤5) | 3 (0.3%) | 2 (0.2%) | 5 (0.3%) |
| Participant Declined to Answer or Unknown | 46 (4.8%) | 63 (6.5%) | 109 (5.6%) |
| Ethnicity | | | |
| Hispanic | 75 (7.8%) | 81 (8.3%) | 156 (8.1%) |
| Non-Hispanic | 861 (89.6%) | 857 (88.1%) | 1718 (88.8%) |
| Participant Declined to Answer or Unknown | 25 (2.6%) | 35 (3.6%) | 60 (3.1%) |
| COVID-19 Vaccination Status | | | |
| Vaccinated | 721 (75.0%) | 717 (73.7%) | 1438 (74.4%) |
| Not Vaccinated | 222 (23.1%) | 239 (24.6%) | 461 (23.8%) |
| Unknown | 18 (1.9%) | 17 (1.7%) | 35 (1.8%) |
The Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System was evaluated by comparing to a U.S. FDA-cleared molecular respiratory panel that includes SARS-CoV-2. Positive Percent Agreement (PPA) was calculated as $100\% \times (\mathrm{TP} / (\mathrm{TP} + \mathrm{FN}))$ . True positive (TP) indicates that both the Xpert Xpress CoV-2 plus and the comparator method had a positive result, and false negative (FN) indicates that the Xpert Xpress CoV-2 plus was negative while the comparator result was positive. Negative Percent Agreement (NPA) was calculated as $100\% \times (\mathrm{TN} / (\mathrm{TN} + \mathrm{FP}))$ . True negative (TN) indicates that both the Xpert Xpress CoV-2 plus and the comparator method had negative results, and false positive (FP)
K242109 - Page 17 of 20
{18}
indicates that the Xpert Xpress CoV-2 plus was positive while the comparator result was negative. Specimens that obtained discordant SARS-CoV-2 results underwent additional testing with a U.S. FDA EUA SARS-CoV-2 molecular test.
A summary of the Xpert Xpress CoV-2/ plus test on the GeneXpert Xpress System prospective clinical study performance is provided in Table 10 for both NPS and ANS specimens.
Table 10. Xpert Xpress CoV-2 plus test on the GeneXpert Xpress System Prospective Clinical Performance (Category I/Fresh NPS and ANS specimens)
| Analyte | Number of Specimens | True Positives | False Negatives | False Positive | True Negative | PPA (%) | 95% CI | NPA (%) | 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| SARS-CoV-2 | NPS Samples | | | | | | | | |
| | 883 | 112 | 2 | 7 | 762 | 98.2 | 93.8-99.5 | 99.1 | 98.1-99.6 |
| | ANS Samples | | | | | | | | |
| | 900 | 103 | 1 | 7 | 789 | 99.0 | 94.8-99.8 | 99.1 | 98.2-99.6 |
a. Discrepant test results based on U.S. FDA EUA SARS-CoV-2 molecular test: 1/14 SARS-CoV-2 positive; 12/14 SARS-CoV-2 negative; 1/14 invalid.
b. Discrepant test results based on U.S. FDA EUA SARS-CoV-2 molecular test: 1/3 SARS-CoV-2 positive; 2/3 SARS-CoV-2 negative.
## Non-determinate (ND) rate
Of the total 1934 prospective specimens evaluated, 81 specimens yielded an initial ND result. After repeat testing, 70 specimens yielded a valid result such that the initial ND rate was 4.2% (81/1934) and final ND rate was 0.6% (11/1934).
- Of the 961 NPS specimens evaluated, 39 specimens yielded an initial ND result. After repeat testing, 32 specimens yielded a valid result. The initial ND rate using NPS specimens was 4.1% (39/961) and final ND rate was 0.7% (7/961).
- Of the 973 ANS specimens evaluated from the symptomatic population, 42 specimens yielded an initial ND result. After repeat testing, 38 specimens yielded a valid result. The initial ND rate using ANS specimens was 4.3% (42/973) and final ND rate was 0.4% (4/973).
2. Clinical Specificity:
See Section “Clinical Sensitivity” above.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not Applicable.
## C Clinical Cut-Off:
Not applicable
K242109 - Page 18 of 20
{19}
# D Expected Values/Reference Range:
Expected values as determined by the Xpert Xpress CoV-2 plus test during the prospective clinical study, stratified by specimen type, gender, age, and vaccination status are presented in Table 11.
Table 11. Xpert Xpress CoV-2 plus Positivity Rates Stratified by Specimen Type, Gender, Age and Vaccination Status
| Subgroup | Category | Total | Black or African American | | White | | | American Indian or Alaska Native | Asian | Black or African American, White | Other Mixed | Participant Declined to Answer or Unknown |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Hispanic | Non-Hispanic | Hispanic | Non-Hispanic | Unknown | | | | | |
| Specimen Type | NPS | 13.5% (119/883) | 0% (0/3) | 10.8% (21/195) | 9.5% (6/63) | 15% (82/546) | 0% (0/1) | NA (0/0) | 4.2% (1/24) | 0% (0/3) | 33.3% (1/3) | 17.8% (8/45) |
| | ANS | 12.2% (110/900) | NA (0/0) | 9.5% (19/200) | 5.6% (4/71) | 13.6% (73/537) | 0% (0/1) | 0% (0/1) | 8.7% (2/23) | 0% (0/4) | 0% (0/1) | 19.4% (12/62) |
| Gender | Female | 12.9% (141/1097) | 0% (0/2) | 11.1% (29/262) | 3.4% (3/87) | 14.4% (94/653) | 0% (0/2) | 0% (0/1) | 7.7% (2/26) | 0% (0/3) | 100% (1/1) | 20.0% (12/60) |
| | Male | 12.8% (88/686) | 0% (0/1) | 8.3% (11/133) | 14.9% (7/47) | 14.2% (61/430) | NA (0/0) | NA (0/0) | 4.8% (1/21) | 0% (0/4) | 0% (0/3) | 17.0% (8/47) |
| Age Group (Years) | ≤5 | 6.6% (4/61) | 0% (0/2) | 4.2% (1/24) | NA (0/0) | 18.2% (2/11) | NA (0/0) | NA (0/0) | 0% (0/3) | 0% (0/1) | NA (0/0) | 5.0% (1/20) |
| | 6 -21 | 9% (46/509) | NA (0/0) | 8.2% (12/146) | 4.9% (3/61) | 9.3% (22/237) | 0% (0/1) | 0% (0/1) | 16.7% (2/12) | 0% (0/5) | 33.3% (1/3) | 14.0% (6/43) |
| | 22-59 | 13.7% (132/962) | 0% (0/1) | 12.1% (25/206) | 7.4% (5/68) | 14.6% (90/617) | 0% (0/1) | NA (0/0) | 3.3% (1/30) | 0% (0/1) | 0% (0/1) | 29.7% (11/37) |
| | ≥60 | 18.7% (47/251) | NA (0/0) | 10.5% (2/19) | 40% (2/5) | 18.8% (41/218) | NA (0/0) | NA (0/0) | 0% (0/2) | NA (0/0) | NA (0/0) | 28.6% (2/7) |
| Vaccination Status | Vaccinated | 13.4% (179/1334) | 0% (0/1) | 11.6% (29/249) | 7.8% (8/102) | 14.4% (124/860) | 0% (0/2) | 0% (0/1) | 7.5% (3/40) | 0% (0/2) | 25% (1/4) | 19.2% (14/73) |
| | Not Vaccinated | 11.0% (46/417) | 0% (0/2) | 7.7% (11/143) | 6.2% (2/32) | 14.4% (30/209) | NA (0/0) | NA (0/0) | 0% (0/4) | 0% (0/5) | NA (0/0) | 13.6% (3/22) |
| | Unknown | 12.5% (4/32) | NA (0/0) | 0% (0/3) | NA (0/0) | 7.1% (1/14) | NA (0/0) | NA (0/0) | 0% (0/3) | NA (0/0) | NA (0/0) | 25.0% (3/12) |
| Overall | | 12.8% (229/1783) | 0% (0/3) | 10.1% (40/395) | 7.5% (10/134) | 14.3% (155/1083) | 0% (0/2) | 0% (0/1) | 6.4% (3/47) | 0% (0/7) | 25.0% (1/4) | 18.7% (20/107) |
# E Other Supportive Instrument Performance Characteristics Data:
Not Applicable
# VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
K242109 - Page 19 of 20
{20}
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K242109 - Page 20 of 20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.