K242102 · Cerrozone, LLC · FRF · Nov 19, 2024 · General Hospital
Device Facts
Record ID
K242102
Device Name
CerroZone Mini
Applicant
Cerrozone, LLC
Product Code
FRF · General Hospital
Decision Date
Nov 19, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5045
Device Class
Class 2
Indications for Use
The CerroZone Mini is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.
Device Story
CerroZone Mini is a medical recirculating air cleaner for medical facilities. Device uses centrifugal fan to pull air through an inline flow-through system. Inactivation processes include reactive oxidizing species (ozone) and ultraviolet radiation; particulate and activated carbon filters trap resulting debris. Ozone is converted back to oxygen via catalyst substrate and carbon filter. Device features redundant external ozone sensors; automatic shutdown if ozone exceeds 0.03 ppm. Output is purified air. Benefits include reduction of airborne pathogens in medical environments. Operated by facility staff; free-standing unit.
Clinical Evidence
No clinical data provided. Bench testing only. Efficacy demonstrated via MS2 bacteriophage aerosolization in environmental chamber (4.0 log reduction target; achieved 5.76 ± 0.20 log reduction in 45 mins). Single-pass testing against MRSE demonstrated >3 log reduction.
Technological Characteristics
Medical recirculating air cleaner; free-standing. Uses centrifugal fan, UV radiation, reactive oxidizing species (ozone), particulate filter, and activated carbon filter. Safety features: catalyst substrate for ozone removal, redundant ozone sensors. Complies with UL 867 and CSA C22.2 No.187. Input: 120V.
Indications for Use
Indicated for use as a room recirculating air cleaner to filter and inactivate airborne bacteria and viruses for medical purposes. Suitable for use in medical facilities.
Regulatory Classification
Identification
A medical recirculating air cleaner is a device used to remove particles from the air for medical purposes. The device may function by electrostatic precipitation or filtration.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 19, 2024
CerroZone LLC % Erin Gontang Senior Consultant ROM+ 2790 Mosside Blvd., Suite 800 Monroeville, Pennsylvania 15146
Re: K242102
Trade/Device Name: CerroZone Mini Regulation Number: 21 CFR 880.5045 Regulation Name: Medical Recirculating Air Cleaner Regulatory Class: Class II Product Code: FRF Dated: July 18, 2024 Received: October 18, 2024
## Dear Erin Gontang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katharine Segars -S
Katharine Segars, Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242102
Device Name CerroZone Mini
Indications for Use (Describe)
The CerroZone Mini is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) Summary
## DATE PREPARED
November 19, 2024
#### MANUFACTURER AND 510(k) OWNER
CerroZone LLC 113975 Riverport Place Drive, Suite 102 Maryland Heights, MO 63043 Telephone: Official Contact: Marlon E. Robinson, Director of Operations
## REPRESENTATIVE/CONSULTANT
Erin A. Gontang, Ph.D. Allison C. Komiyama, Ph.D., RAC RQM+ Telephone: +1-412-816-8290 Email: egontang@rqmplus.com, akomiyama@rqmplus.com
## DEVICE INFORMATION
| Proprietary Name/Trade Name: | CerroZone Mini |
|------------------------------|-----------------------------------|
| Common Name: | Medical Recirculating Air Cleaner |
| Regulation Number: | 21 CFR 880.5045 |
| Class: | II |
| Product Code: | FRF |
| Premarket Review: | Infection Control Devices (DHT4C) |
| Review Panel: | General Hospital |
## PREDICATE DEVICE IDENTIFICATION
The CerroZone Mini is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|--------------------------------------|-------------------|
| K220298 | CerroZone Mobile / CerroZone LLC | ✓ |
#### DEVICE DESCRIPTION
The CerroZone Mini is a medical recirculating air cleaner that uses multiple inactivation processes, including use of reactive oxidizing species (ozone) and ultraviolet radiation, to inactivate bacteria and viruses in the CerroZone Mini may be used in medical facilities. Once turned on, inlet air is passed through CerroZone Mini's inline flow-through design which inactivates bacteria and viruses in a single pass flow-through. The device generates 110 cubic feet of air per minute (CFM), or 2.00 air changes per hour (ACH) in a standard 5,600 cubic foot
{6}------------------------------------------------
room. The process of air being inlet, purified, then outlet, lasts approximately 1.2 seconds. A 4 LOG reduction, or 99.99% reduction, of airborne particles in a standard 579 cubic foot room is achieved in 45 minutes or less.
#### INDICATIONS FOR USE
The CerroZone Mini is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The following table summarizes the similarities and differences between the subject and predicate device.
| | Subject Device | Predicate Device | |
|-------------------------|--------------------------------|--------------------------------|------------|
| | CerroZone LLC | CerroZone LLC | |
| | CerroZone Mini | CerroZone Mobile | Comparison |
| | K242102 | K220298 | |
| Indications for Use | The CerroZone Mini is | The CerroZone Mobile is | Identical |
| | intended as a room | intended as a room | |
| | recirculating air cleaner. The | recirculating air cleaner. The | |
| | system is used for filtering | system is used for filtering | |
| | out and inactivating airborne | out and inactivating airborne | |
| | particles (i.e., viruses and | particles (i.e., viruses and | |
| | bacteria) from the air for | bacteria) from the air for | |
| | medical purposes. | medical purposes. | |
| Product Code / | FRF/ 21 CFR 880.5045 | FRF/ 21 CFR 880.5045 | Identical |
| Regulation Number | | | |
| Regulation Description | Medical recirculating air | Medical recirculating air | Identical |
| | cleaner | cleaner | |
| Type of Use | Over-the-Counter Use | Over-the-Counter Use | Identical |
| Mechanism of Action | Microorganisms including | Microorganisms including | Identical |
| | viruses and bacteria, are | viruses and bacteria, are | |
| | inactivated by the device via | inactivated by the device via | |
| | damage by multiple | damage by multiple | |
| | inactivation processes: | inactivation processes: | |
| | | | |
| | Reactive Oxidizing Species | Reactive Oxidizing Species | |
| | (ozone) | (ozone) | |
| | Ultraviolet Radiation | Ultraviolet Radiation | |
| | | | |
| | Also, a filter to trap the | Also, a filter to trap the | |
| | resulting virus/bacteria | resulting virus/bacteria | |
| | particulates: | particulates: | |
| | | | |
| | Particulate Filter | Particulate Filter | |
| | Activated Carbon Filter | Activated Carbon Filter | |
| Reduction of Biological | MS2 phage reduced by 4 log | MS2 phage reduced by 4 log | Similar |
| Agents | reduction in 45 minutes or | reduction in 30 minutes or | |
| | less when operating at full | less and 5.67 log reduction in | |
| | fan speed in a standard room | 45 minutes or less when | |
{7}------------------------------------------------
| | of 579 ft³ (16 m³) | operating at full fan speed in a std. room of 579 ft³ (16 m³) | |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Ozone Emitted | Within the maximum acceptable level of ozone given in 21 CFR 801.415 | Within the maximum acceptable level of ozone given in 21 CFR 801.415 | Identical |
| Installation | Free standing | Free standing | Identical |
| Reactive Oxidizing Species (Ozone) Source | Activated Carbon Filter<br>Proprietary Catalyst<br>Substrate | Activated Carbon Filter<br>Proprietary Catalyst<br>Substrate | Identical |
| Reactive Oxidizing Species (Ozone) Removal | Activated Carbon Filter<br>Proprietary Catalyst<br>Substrate | Activated Carbon Filter<br>Proprietary Catalyst<br>Substrate | Identical |
| Air Source | Centrifugal Fan | Centrifugal Fan | Identical |
| Device Air Changes Per Hour (ACH) | 2.00 device air changes per hour in a 5,600 ft³ room | 2.36 device air changes per hour in a 5,600 ft³ room | Similar |
| UV Light or Reactive Oxidative Species (Ozone) Exposure Safety Features | A catalyst substrate converts any ozone generated back into oxygen. An activated carbon filter absorbs any remaining ozone.<br><br>If > 0.03 ppm of ozone is measured the external ozone sensors, the CerroZone Mini will automatically shut down. The CerroZone Mini also utilizes a second and redundant external ozone sensor that runs on a different circuit. As such, if one external ozone sensor were to stop functioning, another sensor will be active.<br><br>Safety feature confirmed by UL 867 & CSA C22.2 No.187. | A catalyst substrate converts any ozone generated back into oxygen. An activated carbon filter absorbs any remaining ozone.<br><br>If > 0.03 ppm of ozone is measured the external ozone sensors, the CerroZone Mini will automatically shut down. The CerroZone Mini also utilizes a second and redundant external ozone sensor that runs on a different circuit. As such, if one external ozone sensor were to stop functioning, another sensor will be active.<br><br>Safety feature confirmed by UL 867 & CSA C22.2 No.187. | Identical |
| Input Voltage | 120 Volt | 120 Volt | Identical |
| Dimensions | Unit Dimensions:<br>Height: 36 in<br>Width: 21 in<br>Depth: 17 in | Unit Dimensions:<br>Height: 59 in<br>Width: 29 in<br>Depth: 17 in | Different |
## SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate safety based on current industry standards:
Electromagnetic Compatibility and Electrical Safety: The subject device is equivalent to CerroZone Mobile that tested in compliance to EN/IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests and UL 867 & CSA C22.2 No.187 Electrostatic Air Cleaners.
{8}------------------------------------------------
Ozone Emissions: Ozone Emissions of the CerroZone Mini have been assessed per Section 40 of UL 867 & CSA C22.2 No.187 Electrostatic Air Cleaners. Results of the test show that the CerroZone Mini emits between 0.000 and 0.002 ppm of Ozone with a measured absolute max of 0.002 ppm. All Ozone emissions measured fall within the maximum acceptable level of ozone, 0.050 ppm, given in 21 CFR 801.415.
| Test Methodology | Purpose | Acceptance Criteria | Results |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| MS2 bacteriophage were<br>aerosolized into a sealed<br>environmental bioaerosol<br>chamber containing the<br>CerroZone Mini. | To evaluate the efficacy of<br>the CerroZone Mini at<br>reducing viability of<br>aerosolized MS2<br>bacteriophage by a<br>combination of entrainment<br>and destruction. | Greater than 4 log reduction<br>(99.99%). | Average net log reduction /<br>time, MS2 bacteriophage,<br>$5.76 \pm 0.20$ / 45 mins. |
| Methicillin Resistant<br><i>Staphylococcus epidermidis</i><br>(MRSE) were aerosolized into<br>a single-pass flow through<br>chamber that modeled the<br>CerroZone Mini. | To evaluate the efficacy of<br>the CerroZone Mini at<br>reducing viability of<br>aerosolized bacteria by a<br>combination of entrainment<br>and destruction. | Greater than 3 log reduction<br>(99.99%) in single-pass<br>testing. | Demonstrated effectiveness<br>against <i>Staphylococcus epidermidis</i> . |
#### SUMMARY OF CLINICAL TESTING
No clinical data were provided in order to demonstrate substantial equivalence.
## CONCLUSION
CerroZone concludes that the non-clinical tests demonstrate that the CerroZone Mini is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.