VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700; VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700; 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA
K242067 · Olympus Corporations of the Americas · FET · Apr 2, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K242067
Device Name
VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700; VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700; 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA
Applicant
Olympus Corporations of the Americas
Product Code
FET · Gastroenterology, Urology
Decision Date
Apr 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 is intended to process electronic signals transmitted from a video endoscope/camera head and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, light sources, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 is intended to provide light to an endoscope/video endoscope in order to process electronic signals transmitted from them and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, video system centers, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA is intended to be used with endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.
Device Story
The VISERA ELITE III Surgical Imaging System is a modular endoscopic visualization platform. It consists of the OTV-S700 Video System Center, CLL-S700 LED Light Source, and CH-S700-XZ-EA 4K Camera Head. The system captures optical images via the camera head, which are processed by the video system center and displayed on a monitor for the clinician. The light source provides illumination to the endoscope. The system supports 4K resolution, High Dynamic Range (HDR) observation, and 3D imaging (via the MAJ-2511 accessory). It is used in clinical settings by physicians for endoscopic diagnosis and treatment. The output allows real-time visualization of internal organs, aiding clinical decision-making during surgical or diagnostic procedures. The system improves visualization through high-definition 4K imaging and automated focus control, potentially enhancing diagnostic accuracy and procedural efficiency.
Clinical Evidence
No clinical data. Bench testing only. Performance verified through testing of field of view, resolution, image noise, dynamic range, brightness, color performance, latency, HDR function, and auto-focus capabilities. Electrical safety and EMC testing performed per IEC 60601 series standards.
Technological Characteristics
Modular endoscopic imaging system. Components: OTV-S700 (Video System Center), CLL-S700 (LED Light Source), CH-S700-XZ-EA (4K Camera Head). Sensing: 1 CMOS sensor (4K). Connectivity: Optical digital signal transmission. Power: AC 100-240V. Sterilization: Camera head is autoclavable/V-PRO compatible; system center/light source are not sterilized. Standards: ISO 14971, IEC 60601-1, IEC 60601-2-18.
Indications for Use
Indicated for endoscopic diagnosis, treatment, and observation in patients requiring endoscopic procedures. Used with endoscopes, video system centers, and ancillary equipment.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
OLYMPUS OTV-S300 Video System Center w/ Light Source (K201200)
OLYMPUS CH-S200-XZ-EA HD 3CMOS AUTOCLAVABLE CAMERA HEAD (K190449)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 2, 2025
Olympus Corporations of the Americas
Elizabeth Greene
Program Manager
800 West Park Drive
Westborough, Massachusetts 01581
Re: K242067
Trade/Device Name: Visera Elite III Video System Center Olympus Otv-s700; Visera Elite III Led Light Source Olympus Cll-s700; 4k Camera Head Olympus Ch-s700-xz-ea
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: Class II
Product Code: FET, NWB
Dated: March 4, 2025
Received: March 4, 2025
Dear Elizabeth Greene:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K242067 - Elizabeth Greene
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{3}
K242067 - Elizabeth Greene
Page 3
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2025.04.02
22:57:33 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242067 | |
| Device Name VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700; VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700; 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA | |
| Indications for Use (Describe) The VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 is intended to process electronic signals transmitted from a video endoscope/camera head and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, light sources, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 is intended to provide light to an endoscope/video endoscope in order to process electronic signals transmitted from them and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, video system centers, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA is intended to be used with endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
K242067
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
510(k) Summary
For
VISERA ELITE III Surgical Imaging System
General Information
Applicant: OLYMPUS MEDICAL SYSTEMS CORP.
2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
Phone: (+81) 42-642-2111
Fax: (+81) 42-642-2307
Establishment Registration Number: 8010047
Manufacturer: Shirakawa Olympus Co., Ltd.
3-1 Okamiyama, Odakura, Nishigo-mura,
Nishishirakawa-gun, Fukushima 961-8061,
Japan
510(k) Submitter: Olympus Corporation of the Americas
800 West Park Drive
Westborough, MA 01581
Establishment Registration Number: 2429304
Contact Person: Elizabeth Greene
Program Manager
Mobile: (772) 254-3793
Email: elizabeth.greene@olympus.com
Date Prepared: July 12, 2024
Device Description
Device Name: VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700
Model Name: OLYMPUS OTV-S700
Generic/Common Name: Endoscopic Video Imaging System
Regulation Number: 876.1500
Regulatory Class: Class II
Classification Name: Endoscope and accessories
Product Codes: FET, NWB
Review Panel: Gastroenterology/Urology
510(k) Summary
{6}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
Device Name: VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700
Model Name: OLYMPUS CLL-S700
Generic/Common Name: Endoscopic Video Imaging System
Regulation Number: 876.1500
Regulatory Class: Class II
Classification Name: Endoscope and accessories
Product Codes: FET, NWB
Review Panel: Gastroenterology/Urology
Device Name: 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA
Model Name: OLYMPUS CH-S700-XZ-EA
Generic/Common Name: Endoscopic Video Imaging System
Regulation Number: 876.1500
Regulatory Class: Class II
Classification Name: Endoscope and accessories
Product Codes: FET, NWB
Review Panel: Gastroenterology/Urology
## Predicate Device
| Device Name | 510(k) Submitter | 510(k) No. |
| --- | --- | --- |
| OLYMPUS OTV-S300 Video System Center w/ Light Source | Olympus Medical Systems Corp. | K201200 |
| OLYMPUS CH-S200-XZ-EA HD 3CMOS AUTOCLAVABLE CAMERA HEAD | Olympus Medical Systems Corp. | K190449 |
## Indications for Use
Table 1 details the Indications for Use for the devices in the VISERA ELITE III Surgical Imaging System.
Table 1. Indications for Use for VISERA ELITE III Surgical Imaging System
| Device | Indications for Use |
| --- | --- |
| VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 | The VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 is intended to process electronic signals transmitted from a video endoscope/camera head and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, light sources, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation. |
| VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 | The VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 is intended to provide light to an endoscope/video endoscope in order to process electronic signals transmitted from them and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, video system centers, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation. |
510(k) Summary
{7}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
Device | Indications for Use
---|---
4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA | The 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA is intended to be used with endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.
## Device Description
The VISERA ELITE III Surgical Imaging System is intended to be used with ancillary equipment for endoscopic diagnosis, treatment, and observation and supports the function of high definition (HD) videoscopes and is Camera Head (CH) compatible.
The following devices of the VISERA ELITE III Surgical Imaging System are the subject of this premarket notification submission:
- VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 (Model: OLYMPUS OTV-S700) - A video system center that processes electronic signals transmitted from a video endoscope or a camera head and outputs the image signal to a monitor.
- VISERA ELITE III 3D Upgrade Pack (Model: MAJ-2511) - A function activation portable memory key accessory that unlocks the 3D software function when connected with OTV-S700 to enable the observation of 3D mode.
- VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 (Model: OLYMPUS CLL-S700) - A LED light source provides examination light to a video endoscope and a camera head.
- 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA (Model: OLYMPUS CH-S700-XZ-EA) - A 4K Inline camera head is intended to be used with Olympus endoscopes, the video system center, and other ancillary equipment for the visualization of internal organs (endoscopic diagnosis), treatment and observation.
## Comparison of Technological Characteristics
Table 2 compares VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 to the predicate device with respect to intended use, and technological characteristics, providing detailed information regarding the basis for the determination of substantial equivalence.
Table 2. Comparison of the VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 to the predicate device
| Feature/Technological Characteristics | Subject Device | Predicate Device |
| --- | --- | --- |
| Regulatory | | |
| Device Name | VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 | VISERA ELITE II Video System Center w/Light Source |
| Model Number | OLYMPUS OTV-S700 | OLYMPUS OTV-S300 |
| Legal Manufacturer | Same as the predicate | OLYMPUS MEDICAL SYSTEMS CORP. |
| Regulatory Decision | This submission | K201200 |
| Regulation Number | Same as predicate | 876.1500 |
510(k) Summary
Page 3 of 11
{8}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
| Feature/Technological Characteristics | Subject Device | Predicate Device |
| --- | --- | --- |
| Regulation Name | Same as predicate | Endoscope and accessories |
| Product Code | Same as predicate | FET, NWB |
| Regulatory Class | Same as predicate | II |
| Classification Panel | Gastroenterology/Urology | Obstetrics/Gynecology |
| Indications for Use | The VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 is intended to process electronic signals transmitted from a video endoscope/camera head and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, light sources, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation. | This video system center is intended to be used with OLYMPUS camera heads, endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation. |
| Specifications | | |
| Main Function | Video System Center | Video System Center and Light Source |
| Rated Voltage | AC 100-240V/ 50-60Hz | AC 100V/ 50-60Hz |
| Rated Input | <170VA | 400VA |
| Dimensions (Max.) | W: 390 x H: 198 x D: 500 mm | W: 383 x H: 199 x D: 506 mm |
| Weight | 12.1 kg | 19.3 kg |
| 2D Observation | Same as predicate | Available |
| 3D Observation | With the accessory MAJ-2511 | Available |
| WLI Observation | Same as predicate | Available |
| NBI Observation | Same as predicate | Available |
| HDR Observation | Available | Not Available |
| Iris Area Observation | Same as predicate | Available |
| Laser Mode Observation | Same as predicate | Available |
| Automatic Gain Control | Same as predicate | Available |
| Front Panel (Operation) | Same as predicate | Touch Panel |
| Fog Free Function | Same as predicate | Available |
| Compatible Endoscopes | Same as predicate | Legacy 2D and 3D endoscopes |
Table 3 compares VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 to the predicate device with respect to intended use, technological characteristics, and principle of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
510(k) Summary
Page 4 of 11
{9}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
Table 3. Comparison of the VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 to the predicate device
| Feature/Technological Characteristics | Subject Device | Predicate Device |
| --- | --- | --- |
| Regulatory | | |
| Device Name | VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 | VISERA ELITE II Video System Center w/Light Source |
| Model Number | OLYMPUS CLL-S700 | OLYMPUS OTV-S300 |
| Legal Manufacturer | Same as the predicate | OLYMPUS MEDICAL SYSTEMS CORP. |
| Regulatory Decision | This submission | K201200 |
| Regulation Number | Same as predicate | 876.1500 |
| Regulation Name | Same as predicate | Endoscope and accessories |
| Product Code | Same as predicate | FET, NWB |
| Regulatory Class | Same as predicate | II |
| Classification Panel | Gastroenterology/Urology | Obstetrics/Gynecology |
| Indications for Use | The VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 is intended to provide light to an endoscope/video endoscope in order to process electronic signals transmitted from them and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, video system centers, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation. | This video system center is intended to be used with OLYMPUS camera heads, endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation. |
| Specifications | | |
| Main Function | Light Source | Video System Center and Light Source |
| Rated Voltage | AC 100-240V/ 50-60Hz | AC 100V/ 50-60Hz |
| Rated Input | 215VA | 400VA |
| Dimensions (Max.) | W: 390 x H: 162 x D: 502 mm | W: 383 x H: 199 x D: 506 mm |
| Weight | 13.5 kg | 19.3 kg |
| 2D Observation | Same as predicate | Available |
| 3D Observation | Same as predicate | Available |
| WLI Observation | Same as predicate | Available |
| NBI Observation | Same as predicate | Available |
| Iris Area Observation | Same as predicate | Available |
| Laser Mode Observation | Same as predicate | Available |
| Front Panel (Operation) | Not Available | Touch Panel |
| Examination Lamp | Same as predicate | LED |
| Average Lamp Life | Same as predicate | 10,000 hours |
| Emergency Lamp | Same as predicate | LED |
| Average Emergency Lamp Life | Same as predicate | 10,000 hours |
| Compatible Endoscopes | Same as predicate | Legacy 2D and 3D endoscopes |
510(k) Summary
{10}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
Table 4 compares 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA to the predicate device with respect to intended use, technological characteristics, and principle of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
Table 4. Comparison of the 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA to the predicate device
| Feature/Technological Characteristics | Subject Device | Predicate Device |
| --- | --- | --- |
| Regulatory | | |
| Device (Model) Name | 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA | HD 3CMOS AUTOCLAVABLE CAMERA HEAD |
| Model Number | OLYMPUS CH-S700-XZ-EA | OLYMPUS CH-S200-XZ-EA |
| Legal Manufacturer | Same as the predicate | OLYMPUS MEDICAL SYSTEMS CORP. |
| Regulatory Decision | This submission | K190449 |
| Regulation Number | Same as predicate | 876.1500 |
| Regulation Name | Same as predicate | Endoscope and accessories |
| Product Code | Same as predicate | FET, NWB |
| Regulatory Class | Same as predicate | II |
| Classification Panel | Same as predicate | Gastroenterology/Urology |
| Indications for Use | The 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA is intended to be used with endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. | The camera head has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and observation. |
| Specifications | | |
| Observation Function | Same as predicate | 2D Function |
| 4K Compatible | Available | Not Available |
| Head Dimensions (Max.) | W: 44 x H: 54 x L:133 mm | W: 44 x H: 52 x L: 120 mm |
| Cable | Diameter: 5.1 mm x L: 3 m | Diameter: 6.8 mm x L: 3 m |
| Connector Surface | Card edge Connector with Optical Connection | Card Edge Connector |
| Image Module | 1 CMOS | 3CMOS |
| Pixel Count | 4,152(H) x 3,062(V) = 12,713,424 pixels | 1,944(H) x 1,092(V) = 2,122,848 pixels (Per CMOS) |
| Pixel Size | 1.55μm(H) x 1.55μm(V) | 2.75μm(H) x 2.75μm(V) |
| Focus Control | Electrical manual focus with focus control switches | Electrical manual focus with focus control switches |
| | Electrical automatically focus with one-touch auto focus/Continuous auto focus switch | Not Available |
| | Continuous electrical automatically focus | Not Available |
| Optics Zoom | Not Available | Available |
| Field of View | 20.7°88° with representative rigid scope | 10.90° (Low magnification) to 22.42° (High magnification) |
510(k) Summary
{11}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
| Feature/Technological Characteristics | Subject Device | Predicate Device |
| --- | --- | --- |
| NBI Observation | Same as predicate | Available |
| Image Signal Transmission | Optical transmission (Digital) | Electro transmission (Digital) |
| Degree of protection against electric shock | Type CF | Type BF |
| Number of remote switches | Same as predicate | 3 |
| Compatible endoscopes | Same as predicate | Legacy 2D endoscopes |
| Head Weight | 270 g (excluding cable) | 295 g (excluding cable) |
| Reprocessing | End User Sterilized
Autoclaving/V-PRO | End User Sterilized
Autoclaving/V-PRO/STERRAD |
## Compliance to Voluntary Standards
The following voluntary standards have been applied to the subject devices:
- ISO 15223-1
- ISO 7000
- ISO 7010
- ANSI AAMI ST98:2022
- ISO 17664-2 First Ed. 2021-02
- ISO 17665-1 First Ed. 2006-08
- ANSI AAMI IEC 62304:2006/A1: 2016
- IEC 81001-5-1 Ed. 1.0 2021-12
- AAMI TIR57:2016
- ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012
- IEC 60601-1-2 Ed. 4.1 2020-09
- IEC 60601-1-6 Ed. 3.2 2020-07
- IEC 60601-4-2 Ed. 1.0 2016-05
- IEC 60601-2-18 Ed. 3.0 2009-08
- ISO 8600-1:2015
- ISO 8600-3:2019
- IEC 62471:2006-07
- IEC 60825-1 Edition 2.0 2007-03
- ISO 14971:2019
## Summary of Performance Testing
The following performance testing was conducted in support of the substantial equivalence determination.
510(k) Summary
Page 7 of 11
{12}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
# Reprocessing, Sterilization, and Shelf Life Validation
The VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 and VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 are not intended to be sterilized or reprocessed. Therefore, sterilization and reprocessing testing and data was not applicable for these devices. The VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 and VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 are reusable and the associated instruction manuals for these devices includes cleaning and disinfection instructions with the following:
- Endozime AW
- 70% Isopropyl Alcohol
- Sodium Hypochlorite
The 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA is distributed non-sterile to the end user. Before using the instrument for the first time and after each use, the device must be reprocessed according to the instructions in the companion Reprocessing Manual. The 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA is validated as safe and effective for reprocessing as detailed in the Reprocessing Manual with the following:
- Manual Cleaning with Endozime AW
- Delayed Manual Cleaning with Endozime AW
- Sterilization with:
- V-PRO maX
- Autoclave
- Drying time of 80 minutes
The VISERA ELITE III Surgical Imaging System has a low likelihood of time-dependent product degradation. Therefore, shelf life testing and data was not applicable for the VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700, VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700, and 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA.
# Software Verification and Validation Testing
Software verification and validation testing of the VISERA ELITE III Surgical Imaging System has been performed and documented in compliance with the FDA guidance "Guidance for the Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions."
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC performance testing for VISERA ELITE III Surgical Imaging System is confirmed to be in compliance with the relevant requirements as noted below:
- ANSI AAMI ES 60601-1:2005+A1:2012 [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
510(k) Summary
Page 8 of 11
{13}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
- IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2 Edition 4.1: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 4)
- IEC 60601-2-18: Edition 3.0 2009-08: Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- IEC 60601-1-6 Edition 3.1: Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
## Bench Testing
Bench testing as listed below was conducted for VISERA ELITE III Surgical Imaging System to ensure that the subject device performs as intended and meets design specifications for the following:
- Field of View and Direction of View
- Resolution
- Image Noise and Dynamic Range
- Brightness
- Image Intensity Uniformity
- Color Performance
- Latency
- Iris Area
- Laser Mode
- Magnification
- Distortion
- Depth of Field
- Photobiological Safety
- Laser Product Safety
- Auto Focus Function
- High Dynamic Range (HDR) Function
- Durability
## Risk Analysis
Risk analysis for the VISERA ELITE III Surgical Imaging System was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2019. The design verification tests, and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
In the risk management process, Olympus performed preliminary analysis and evaluation to identify user tasks, user interface components, and use issues of the VISERA ELITE III Surgical Imaging System in accordance with the FDA Guidance, "Applying Human Factors
510(k) Summary
Page 9 of 11
{14}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
and Usability Engineering to Medical Devices" issued on February 2, 2016. Olympus determined that human factors validation testing was not required for the VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 and VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700. For 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA, user tasks associated with reprocessing procedures were identified as critical tasks. Therefore, Olympus performed human factors validation and determined that risks have been mitigated effectively.
## Animal and Clinical Testing
Animal and Clinical testing was not applicable and not performed.
## Substantial Equivalence
Olympus has determined that the VISERA ELITE III Surgical Imaging System is substantially equivalent to the legally marketed predicate devices, VISERA ELITE II Video System Center OLYMPUS OTV-S300 (K201200) and the VISERA ELITE II HD 3CMOS Autoclavable Camera Head OLYMPUS CH-S200-XZ-EA (K190449) for the following reasons:
- same intended use;
- technological characteristics (design, materials, and operations) are similar or identical to the predicate devices; and
- does not introduce any new or novel treatments or standard of care that differs from predicate devices in commercial use.
## VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700
The intended use, principles of operation, fundamental technology of VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 are identical to the predicate device. The differences in specifications include separation of processing and light source functions, rated voltage and input, device dimensions, and weight. New features include HDR Observation and 3D functionality availability via the accessory, MAJ-2511 which have been verified as safe and effective. The difference in indications for use between the VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 and the VISERA ELITE II Video System Center with Light Source is not a change from single use labeling to reusable and is not a change from prescription (Rx) use to over the counter (OTC) use. Further, this change does not describe a new disease, condition, or patient population that the device is intended in diagnosing, treating, preventing, curing, or mitigating. A risk-based assessment of these differences did not identify any new risks or significantly modified existing risks or raise new or different questions with respect to safety and effectiveness.
## VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700
The intended use, principles of operation, fundamental technology of VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 are identical to the predicate device. The differences in specifications include rated voltage and input, device dimensions, weight, and absence of a front panel. The difference in indications for use between the VISERA ELITE III LED LIGHT
510(k) Summary
Page 10 of 11
{15}
OLYMPUS
TRADITIONAL 510(K) NOTIFICATION
VISERA ELITE III Surgical Imaging System
SOURCE OLYMPUS CLL-S700 and the VISERA ELITE II Video System Center with Light Source is not a change from single use labeling to reusable, is not a change from prescription (Rx) use to over the counter (OTC) use. Further, this change does not describe a new disease, condition, or patient population that the device is intended in diagnosing, treating, preventing, curing, or mitigating. A risk-based assessment of these differences did not identify any new risks or significantly modified existing risks or raise new or different questions with respect to safety and effectiveness.
## 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA
The intended use, principles of operation, fundamental technology of 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA are identical to the predicate device. The differences in specifications include 4K compatibility, device dimensions, image module, increased pixel count, decreased pixel size, additional focus control, digital zoom, increased field of view, image signal transmission, degree of protection against electric shock, and weight. The difference in indications for use between the 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA and the HD 3CMOS AUTOCLAVABLE CAMERA HEAD is not a change from single use labeling to reusable, is not a change from prescription (Rx) use to over the counter (OTC) use. Further, this change does not describe a new disease, condition, or patient population that the device is intended in diagnosing, treating, preventing, curing, or mitigating. A risk-based assessment of these differences did not identify any new risks or significantly modified existing risks or raise new or different questions with respect to safety and effectiveness.
The VISERA ELITE III Surgical Imaging System has been verified and validated to be equivalent in electrical performance for use with video endoscopes/camera head to output image signals to a monitor for endoscopic diagnosis, treatment, and observation when compared to the predicates. As the electrical safety and electromagnetic compatibility test results demonstrate equivalent performance, Olympus has determined there are no new concerns or modified existing risks regarding safety and effectiveness of the subject device.
## Conclusion
In summary, the VISERA ELITE III Surgical Imaging System (VISERA ELITE III VIDEO SYSTEM, CENTER OLYMPUS OTV-S700, VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700, and 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA) is substantially equivalent to the predicate devices and raises no new questions of safety or effectiveness.
510(k) Summary
Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.