← Product Code [NHA](/productcode/NHA) · K242064

# Omni-Directional Multi-unit Abutment System (Trade Name: Omnibut™) (K242064)

_Smart Denture Conversions, LLC · NHA · Jan 17, 2025 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K242064

## Device Facts

- **Applicant:** Smart Denture Conversions, LLC
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Jan 17, 2025
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K242064 · Jan 17, 2025 | Omni-Directional Multi-unit Abutment System (Trade Name: Omnibut™) | Smart Denture Conversions, LLC | Published clinical literature (Woods, Delfino, and Rajan, 2021) | Published literature was used to evaluate the subject device in the MR environment, specifically addressing magnetically induced displacement force and torque for metal alloys used in medical devices. | MR safety; Literature review; Metal alloy behavior |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices; Literature review / Scientific analysis; Follow-up/Duration: Not applicable; Study Period: 2021 | Not applicable; Sample Size: Not applicable; Number of Sites: Not applicable | Not applicable for this study | Magnetically induced displacement force and torque in MR environment |

## Indications for Use

Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function with four or more endosseous implants, are indicated for multiple unit reconstructions when screw-retained prosthetics are preferred.

## Device Story

Omnibut is a transmucosal, pre-manufactured prosthetic component connecting directly to endosseous dental implants. It supports screw-retained prostheses on four or more implants. The system features a ball abutment attached to the implant, with a retention attachment allowing up to 30° of angle correction via an orientation screw. Prostheses connect via titanium cylinders incorporated into resin or ceramic structures and are secured by prosthetic screws. Used in clinical settings by dental professionals. The device aids in prosthetic rehabilitation by restoring chewing function. It is provided non-sterile and requires user sterilization.

## Clinical Evidence

Bench testing only. Evidence includes reverse engineering dimensional analysis for implant compatibility, dynamic load testing per ISO 14801:2016, retention force testing, simulated use testing, and simulated cleaning validation. Biocompatibility was confirmed per ISO 10993-1 and ISO 10993-5. Sterilization was validated to a 10^-6 SAL using the overkill method per ANSI/AAMI/ISO 17665-1. MR environment safety was assessed via scientific literature and analysis.

## Technological Characteristics

Material: Ti-6Al-4V alloy (ASTM F136). Design: Transmucosal ball abutment with orientation screw for up to 30° angulation. Connection: Internal interface compatible with specific Neodent, NobelActive, Straumann BLX, and Zimmer Tapered Screw-Vent platforms. Sterilization: Non-sterile, user-sterilizable via moist heat. Single-use.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- NobelActive Multi Unit Plus Abutment ([K161416](/device/K161416.md))

## Reference Devices

- Meg-Ball Attachment System ([K192614](/device/K192614.md))
- DESS Dental Smart Solutions ([K240208](/device/K240208.md))
- Neodent Implant System – GM Line ([K163194](/device/K163194.md))
- Neodent Implant System – GM Line ([K180536](/device/K180536.md))
- Neodent Implant System – GM Helix Implants 7.0 ([K201225](/device/K201225.md))
- Nobelactive Internal Connection Implant ([K071370](/device/K071370.md))
- Nobelactive 8.5 Mm & 18.0 Mm ([K083205](/device/K083205.md))
- NobelActive ([K142260](/device/K142260.md))
- Straumann BLX Implant System ([K173961](/device/K173961.md))
- Straumann BLX Line Extension ([K181703](/device/K181703.md))
- Straumann BLX Ø3.5 mm Implants ([K191256](/device/K191256.md))
- Straumann BLX Implant System ([K210855](/device/K210855.md))
- BLX WB Ø5.0, Ø5.5 and Ø6.5 mm Implants ([K212533](/device/K212533.md))
- Tapered Screw-Vent Implant ([K072589](/device/K072589.md))
- Screw-Vent Implant System ([K061410](/device/K061410.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 17, 2025

Smart Denture Conversions, LLC % Azary Joseph Regulatory Consultant Aztech Regulatory & Quality LLC 543 Long Hill Avenue Shelton, Connecticut 06484

Re: K242064

Trade/Device Name: Omni-Directional Multi-unit Abutment System (Trade Name: Omnibut™) Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 12, 2024 Received: December 19, 2024

Dear Azary Joseph:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K242064

## Device Name

Omni-Directional Multi-unit Abutment System (Trade Name: Omnibut™)

Indications for Use (Describe)

The Omnibut is a pre-manufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

| Compatible Implant System<br>(Connection)        | Implant Body Diameter, mm         | Implant Platform, mm |
|--------------------------------------------------|-----------------------------------|----------------------|
| Neodent Helix<br>(Morse taper GM)                | 3.5,3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | GM (3.0)             |
| NobelActive®<br>(Conical Connection)             | 3.5                               | NP (3.5)             |
|                                                  | 4.3, 5.0                          | RP (3.9)             |
| Straumann BLX<br>(TorcFit™ Internal Hexalobular) | 3.5, 3.75, 4.0, 4.5               | RB                   |
|                                                  | 5.0, 5.5, 6.5                     | WB                   |
| Tapered Screw-Vent®<br>(Zimmer Internal Hex)     | 3.7, 4.1                          | 3.5                  |
|                                                  | 4.7                               | 4.5                  |
|                                                  | 6                                 | 5.7                  |

| Table 1 - Compatible Implant Systems |  |  |
|--------------------------------------|--|--|

| Type of Use (Select one or both, as applicable)  |                                             |
|--------------------------------------------------|---------------------------------------------|
| ×   Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## 510(k) Summary

#### K242064

#### Smart Denture Conversions, LLC

#### Omni-Directional Multi-Unit Abutment System

January 14, 2025

#### ADMINISTRATIVE INFORMATION

| Manufacturer Name         | Smart Denture Conversions, LLC  |
|---------------------------|---------------------------------|
|                           | 1800 N. Salem St., Suite 104    |
|                           | Apex, NC 27523                  |
| FDA Registration          | 3015527825                      |
| Telephone                 | (855) 550-0707                  |
| Representative/Consultant | Joseph Azary                    |
|                           | Aztech Regulatory & Quality LLC |
|                           | 543 Long Hill Ave               |
|                           | Shelton, CT 06484               |
|                           | Telephone (203) 242-6670        |
|                           | Email jazary@rcn.com            |

#### DEVICE NAME AND CLASSIFICATION

| Proprietary Name:     | Omni-Directional Multi-unit Abutment System (Trade Name: Omnibut™)                             |
|-----------------------|------------------------------------------------------------------------------------------------|
| Common/Usual Name:    | Dental Implant Abutment                                                                        |
| Regulation Number:    | 21 CFR 872.3630                                                                                |
| Regulation Name:      | Endosseous dental implant abutment                                                             |
| Regulatory Class:     | Class II                                                                                       |
| Product code:         | NHA                                                                                            |
| Classification Panel: | Dental                                                                                         |
| Reviewing Office:     | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory, ENT, and<br>Dental Devices |
| Reviewing Division:   | Division of Dental and ENT Devices                                                             |

#### PREDICATE AND REFERENCE DEVICE INFORMATION

Primary Predicate: K161416, NobelActive Multi Unit Plus Abutment, Nobel Biocare AB

Reference Devices:

K192614, Meg-Ball Attachment System, Meg-Magnet Abutment, Meg-Magnet Abutment, MegaGen Implant CO., Ltd. K240208, DESS Dental Smart Solutions, Terrats Medical SL

Reference Devices for OEM Implant Body Clearances K163194, Neodent Implant System – GM Line, JJGC Indústria e Comércio de Materiais Dentários SA K180536, Neodent Implant System – GM Line, JJGC Indústria e Comércio de Materiais Dentários S.A. K201225, Neodent Implant System – GM Helix Implants 7.0, JJGC Indústria e Comércio de Materiais S.A. K071370, Nobelactive Internal Connection Implant, Nobel Biocare AB K083205, Nobelactive 8.5 Mm & 18.0 Mm, Nobel Biocare AB K142260, NobelActive®, Nobel Biocare AB K173961, Straumann® BLX Implant System, Institut Straumann AG

{5}------------------------------------------------

510(k) Summary Page 2 of 8

K181703, Straumann® BLX Line Extension - Implants, SRAs and Anatomic Abutments, Institut Straumann AG

K191256, Straumann BLX Ø3.5 mm Implants, Institut Straumann AG

K210855, Straumann BLX Implant System, Institut Straumann AG

K212533, BLX WB Ø5.0 (L18), Ø5.5 and Ø6.5 mm (L14 and L16) Implants, Institut Straumann AG

K072589, Tapered Screw-Vent Implant, 4.1mmD, Zimmer Dental, Inc.

K061410, Screw-Vent Implant System, SV, Tapered Screw-Vent Implant System TSV. Advent Implant System AV, Zimmer One Piece Implant, Zimmer Dental (includes 3.7, 4.7 and 6.0mm diameters)

## INTENDED USE/INDICATIONS FOR USE STATEMENT

Indications for Use:

The Omnibut is a pre-manufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

| Compatible Implant System<br>(Connection)        | Implant Body Diameter, mm         | Implant Platform, mm |
|--------------------------------------------------|-----------------------------------|----------------------|
| Neodent Helix<br>(Morse taper GM)                | 3.5,3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | GM (3.0)             |
| NobelActive®<br>(Conical Connection)             | 3.5                               | NP (3.5)             |
| NobelActive®<br>(Conical Connection)             | 4.3, 5.0                          | RP (3.9)             |
| Straumann BLX<br>(TorcFit™ Internal Hexalobular) | 3.5, 3.75, 4.0, 4.5               | RB                   |
| Straumann BLX<br>(TorcFit™ Internal Hexalobular) | 5.0, 5.5, 6.5                     | WB                   |
| Tapered Screw-Vent ®<br>(Zimmer Internal Hex)    | 3.7, 4.1                          | 3.5                  |
| Tapered Screw-Vent ®<br>(Zimmer Internal Hex)    | 4.7                               | 4.5                  |
| Tapered Screw-Vent ®<br>(Zimmer Internal Hex)    | 6                                 | 5.7                  |

#### Table 1 - Compatible Implant Systems

Intended Use:

Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function with four or more endosseous implants, are indicated for multiple unit reconstructions when screw-retained prosthetics are preferred.

## SUBJECT DEVICE DESCRIPTION

The Omnibut™ is a transmucosal abutment used to support screw-retained prostheses on four or more implants. The subject device has a premanufactured connection for the platforms listed in Table 1 Compatible Implant Systems.

The system involves a ball abutment attached to an implant. A retention attachment allows for angle corrections of up to 30° off the implant axis. The ball abutment is inserted into the attachment is adjusted to the desired angle using an orientation screw. The abutment supports prostheses that connect via titanium cylinders, which are incorporated into resin or ceramic prostheses. Finally, the prostheses are retained to the abutment by prosthetic screws.

The subject device abutments and system components are manufactured from Ti-6Al-4V alloy conforming to ASTM F136. The subject device is a single use device is provided nonsterile and intended to be sterilized by the user prior to placement in the patient.

{6}------------------------------------------------

## PERFORMANCE DATA

Non-clinical testing performed on the subject device to demonstrate substantial equivalence included:

- provided in this submission is reverse engineering dimensional analysis (of OEM implant bodies, OEM abutments, and OEM abutment screws) to demonstrate the subject device abutments are compatible with the Neodent GM Helix (Morse Taper GM connection) implants, the Nobel Internal (Conical Connection) implants, Straumann BLX (TorcFit™ Internal Hexalobular connection) implants, and Zimmer Internal (Internal Hex connection) implants;
- provided in this submission is dynamic load testing conducted according to ISO 14801:2016 to support the performance of the subject device abutments with the Neodent GM Helix (Morse Taper GM connection) implants, the Nobel Internal (Conical Connection) implants, Straumann BLX (TorcFit™ Internal Hexalobular connection) implants, and Zimmer Internal Hex connection) implants;
- provided in this submission was non-clinical analysis to evaluate the subject device in the MR environment using scientific rationale and published literature (TO Woods, JG Delfino, and S Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices," Journal of Testing and Evaluation, Volume 49, No. 2, 2021, pp. 783-795); the analysis addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (issued May 2021) including magnetically induced displacement force and torque;
- . provided in this submission is moist heat sterilization for subject devices provided non-sterile to the end user, validated to a sterility assurance level of 10° by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2;
- Verification of biocompatibility of the final device in accordance with ISO 10993-1 and ISO 10993-5;
- Bench testing included Retention Force Testing and Simulated Use Testing.
- Simulated cleaning testing to show the cleanability of the Omnibut vs a predicate device.

| Table 2 – Summary of Performance Data                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Standard                                                                                                                                                                                                | Description of Test                                                                                                                                                                                                                                                                                                        | Results                                                                                                                                                                                                                                                                                                                                               |
| Biological Risk Assessment<br>ISO 10993-1:2018<br>FDA Guidance on the Use of<br>ISO 10993-1, 2023                                                                                                       | Biological Evaluation of Medical<br>Devices – Part 1: Evaluation and<br>Testing within a Risk<br>Management Process                                                                                                                                                                                                        | All biological endpoint testing performed<br>on the device, along with the analysis on<br>the physical and chemical information,<br>returned passing results. All biological<br>endpoint testing suggests that the Omnibut<br>is biocompatible and does not present a<br>foreseen biological risk to those patient<br>populations it is intended for. |
| Cytotoxicity<br>ISO 10993-5 (2009)                                                                                                                                                                      | Biological Evaluation of Medical<br>Devices – Part 5: Tests for in<br>vitro cytotoxicity                                                                                                                                                                                                                                   | The test article showed no evidence of<br>causing cell lysis or toxicity. The test article<br>met the requirements of the test since the<br>grade was grade 0 (no reactivity).                                                                                                                                                                        |
| Sterilization Validation<br>Testing per standards:<br>• AAMI TIR12:2020<br>• ANSI/AAMI/ISO 17665-<br>1:2026/(R)2013<br>• ANSI/AAMI ST79:2017                                                            | AAMI TIR12:2020 Designing,<br>Testing, And Labeling Medical<br>Devices Intended For Processing<br>By Health Care Facilities: A Guide<br>For Device Manufacturers,<br>Overkill method according to<br>Section 5.7                                                                                                           | PASS – Results from testing have<br>demonstrated that the Omnibut was able<br>to achieve a 10-6 SAL when using the<br>recommended parameters in the<br>Instructions for Use (IFU).                                                                                                                                                                    |
| Dynamic Fatigue Testing                                                                                                                                                                                 | Dynamic Fatigue testing per:                                                                                                                                                                                                                                                                                               | PASS – The results conclude that when                                                                                                                                                                                                                                                                                                                 |
| Testing per ISO 14801:2016<br><br>FDA Guidance Document,<br>"Class II Special Controls<br>Guidance Document: Root-form Endosseous Dental<br>Implants and Endosseous<br>Dental Abutments" (May 12, 2004) | ISO 14801:2016: Dentistry —<br>Implants — Dynamic loading test<br>for endosseous dental implants<br><br>Root-form Endosseous Dental<br>Implants and Endosseous Dental<br>Abutments - Class II Special<br>Controls Guidance Document for<br>Industry and FDA Staff Section 8                                                | evaluated in a manner consistent with ISO<br>14801:2016, the Omnibut met all<br>predetermined acceptance criteria.                                                                                                                                                                                                                                    |
| Compatibility                                                                                                                                                                                           | Reverse engineering dimensional<br>analysis of OEM implant bodies,<br>OEM abutments, and OEM<br>abutment screws were<br>performed to demonstrate that<br>the Omnibut abutments are<br>compatible with the noted<br>implant systems.<br><br>Compatibility must be<br>demonstrated to ensure proper<br>device functionality. | PASS – The engineering and dimensional<br>analysis concluded that each Omnibut<br>design is compatible with the applicable<br>implant connection.<br><br>PASS - Omnibut prosthetic side<br>components were demonstrated to be<br>compatible.                                                                                                          |
| FDA Guidance Document,<br>"Class II Special Controls<br>Guidance Document: Root-form Endosseous Dental<br>Implants and Endosseous<br>Dental Abutments" (May 12, 2004)                                   |                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                       |
| Retention Force Testing                                                                                                                                                                                 | The Omnibut has a retention<br>attachment. Retention Force<br>testing was performed via<br>tensile push-out to ensure that<br>the attachment will not detach<br>during clinical use.                                                                                                                                       | PASS - The Omnibut retention attachment<br>did not detach at a predetermined<br>acceptable force.                                                                                                                                                                                                                                                     |
| No standard                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                       |
| Simulated-Use Testing                                                                                                                                                                                   | Simulated Use of four Omnibuts<br>with components connected to a<br>Titanium Bar under a clinically<br>relevant cyclic load.                                                                                                                                                                                               | PASS - The Omnibut and components did<br>not yield, deform, or fracture after fatigue<br>testing.                                                                                                                                                                                                                                                     |
| No standard                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                       |
| Simulated Cleaning Testing                                                                                                                                                                              | Simulated cleaning of Omnibuts<br>in a fixture with a clinically worst<br>case cleansibility construction.                                                                                                                                                                                                                 | PASS - All parts of the Omnibuts were<br>clean of soil indicators after six soilage and<br>cleaning cycles.                                                                                                                                                                                                                                           |
| No standard                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                       |

#### of Dorform

{7}------------------------------------------------

Clinical performance data is not required to establish substantial equivalence for the subject devices.

## EQUIVALENCE TO MARKETED DEVICES

The subject device abutments are substantially equivalent in intended use to the primary predicate device K161416, and the reference devices K192614 and K240208. All are intended for use with endosseous dental implants in the maxilla and mandible to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible.

{8}------------------------------------------------

The primary predicate device K161416 and the reference devices K192614 and K240208 are in support of the substantial equivalence of the subject device designs, material, manufacturing, and biocompatibility.

All subject device abutments have identical materials as the predicate device is different in technology compared to the predicate, as it has a variable angulation and differently designed components. The device has many common characteristics with the predicate and reference devices as found in 510(k)sK161416, K192614, and K240208. The subject submission, K161416, K192614, and K240208 all include abutments manufactured from Ti-6Al-4V alloy conforming to ASTM F136. The scientific data described above demonstrates that the subject device has equivalent performance compared to the predicate device.

The subject device abutments have similar or identical ranges of abutment-implant platform diameter, prosthetic platform diameter, and angulation as the components cleared in K161416, K192614, and K240208.

The subject device abutments and the reference device K240208 include abutments compatible with the Reference Devices for OEM Implant Body Clearances above.

The subject device abutments and the reference device K240208 include non-sterile provided devices.

The risks associated with the use of the subject device abutment angulation with the compatible implants are mitigated by the mechanical testing as described in the performance testion of this summary.

## CONCLUSION

Any differences in the technological characteristics between the subject device, the predicate device, and reference devices do not raise different questions of safety or effectiveness. The data included in this submission demonstrate substantial equivalence to the predicate and reference devices listed above.

Overall, the subject devices have the following similarities to the primary predicate and reference devices:

- Have similar intended use,
- Use similar operating principles,
- Incorporate identical materials,
- Are sterilized using similar materials and processes,
- And have similar packaging

The basis for the belief of Smart Denture Conversions, LLC that the subject device is substantially equivalent to the predicate devices is summarized in the following Table 3, Substantial Equivalence.

{9}------------------------------------------------

## Table 3 – Substantial Equivalence

| Descriptive<br>Information              | Subject Device                                                                                                                                                                                                                                                                                                                                  | Primary Predicate                                                                                                                                                                                                                                                                                                                             | Reference Device #1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Reference Device #2                                                                                                                                                                                                                                                       | Comparison                                                                                                                                                                          |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                         | Omni-Directional Multi-<br>Unit Abutment System<br><br>Smart Denture<br>Conversions                                                                                                                                                                                                                                                             | K161416<br>Multi-unit Abutment Plus<br><br>Nobel Biocare AB                                                                                                                                                                                                                                                                                   | K192614<br>Meg-Ball Attachment System,<br>Meg-Loc Abutment, Meg-Magnet<br>Abutment<br><br>MegaGen Implant CO., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                | K240208<br>DESS Dental Smart Solutions<br><br>Terrats Medical SL                                                                                                                                                                                                          | N/A                                                                                                                                                                                 |
| Regulatory Classification               |                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                           |                                                                                                                                                                                     |
| Registration #                          | 21 CFR 872.3630                                                                                                                                                                                                                                                                                                                                 | 21 CFR 872.3630                                                                                                                                                                                                                                                                                                                               | 21 CFR 872.3630                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 21 CFR 872.3630                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                           |
| Registration Title                      | Endosseous Dental<br>Implant Abutment                                                                                                                                                                                                                                                                                                           | Endosseous Dental Implant<br>Abutment                                                                                                                                                                                                                                                                                                         | Endosseous Dental Implant<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Endosseous Dental Implant<br>Abutment                                                                                                                                                                                                                                     | Identical                                                                                                                                                                           |
| Regulation Class                        | II                                                                                                                                                                                                                                                                                                                                              | II                                                                                                                                                                                                                                                                                                                                            | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | II                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                           |
| Product Code                            | NHA                                                                                                                                                                                                                                                                                                                                             | NHA                                                                                                                                                                                                                                                                                                                                           | NHA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | NHA                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                           |
| Reason for<br>predicate or<br>reference | N/A                                                                                                                                                                                                                                                                                                                                             | Indications, function, system<br>compatibility, retention testing,<br>general technological<br>characteristics                                                                                                                                                                                                                                | Technological comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Implant System Compatibility                                                                                                                                                                                                                                              | N/A                                                                                                                                                                                 |
|                                         | Subject Device                                                                                                                                                                                                                                                                                                                                  | Primary Predicate                                                                                                                                                                                                                                                                                                                             | Reference Device #1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Reference Device #2                                                                                                                                                                                                                                                       | Comparison                                                                                                                                                                          |
| 510(k) Number                           | N/A                                                                                                                                                                                                                                                                                                                                             | K161416                                                                                                                                                                                                                                                                                                                                       | K192614                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | K240208                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                 |
| Device Name                             | Omni-Directional Multi-Unit<br>Abutment<br>For Omni-Directional Multi-Unit<br>Abutment System                                                                                                                                                                                                                                                   | Multi-unit Abutment Plus<br>For NobelActive                                                                                                                                                                                                                                                                                                   | Meg-Ball Abutment<br>For Meg-Ball Attachment System                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Multi-Unit Abutment<br>For DESS Dental Smart Solutions                                                                                                                                                                                                                    | N/A                                                                                                                                                                                 |
| Intended Use<br>Statement               | Dental implant abutments are<br>intended to be used in the<br>upper or lower jaw and used for<br>supporting tooth replacements<br>to restore chewing function.<br>Omnibuts, in combination with<br>four or more endosseous<br>implants, are indicated for<br>multiple unit reconstructions<br>when screw-retained<br>prosthetics are preferred. | Dental implant abutments are<br>intended to be used in the<br>upper or lower jaw and used<br>for supporting tooth<br>replacements to restore<br>chewing function. Multi-unit<br>Abutment in combination with<br>endosseous implants are<br>indicated for multiple unit<br>reconstructions when screw<br>retained prosthetics is<br>preferred. | Meg-Ball Attachment System is a<br>superstructure of a dental implant<br>system to provide support for<br>prosthetic restorations. It is<br>intended to be used in implant-<br>retained and removable<br>overdenture restorations where<br>the patient is fully edentulous.                                                                                                                                                                                                                                        | Functional and esthetic rehabilitation<br>of the edentulous mandible or maxilla.                                                                                                                                                                                          | Similar Intended<br>Use expressed<br>through a similar<br>choice of words.                                                                                                          |
| Indications for<br>Use                  | The Omnibut is a pre-<br>manufactured prosthetic<br>component directly connected<br>to the endosseous dental<br>implant and is intended for use<br>as an aid in prosthetic<br>rehabilitation.<br>[Table of Compatible Implant]                                                                                                                  | The Multi-unit Abutment Plus<br>is a pre-manufactured<br>prosthetic component directly<br>connected to the endosseous<br>dental implant and is intended<br>for use as an aid in prosthetic<br>rehabilitation.                                                                                                                                 | The Meg-Ball Attachment System<br>is intended to be used in the upper<br>or lower jaw and used for<br>supporting tooth replacements to<br>restore chewing function. Intended<br>for fully edentulous jaw retaining a<br>tissue supported overdenture.<br>The abutments in combination<br>with endosseous implants are used<br>as the foundation for anchoring<br>tooth replacements in either jaw.<br>The attachments are used in fixed<br>overdenture restorations that can<br>be attached with a snap-in system. | DESS Dental Smart Solutions<br>abutments are intended to be used in<br>conjunction with endosseous dental<br>implants in the maxillary or<br>mandibular arch to provide support for<br>prosthetic restorations.<br>[Table of Compatible Implant]                          | Similar Indications<br>for Use expressed<br>through a similar<br>choice of words.<br>Subject, Predicate<br>and Reference<br>Device #2 used for<br>multi-unit fixed<br>restorations. |
| Design                                  | Image: Omni-Directional Multi-Unit Abutment Design                                                                                                                                                                                                                                                                                              | Image: Multi-unit Abutment Plus Design                                                                                                                                                                                                                                                                                                        | Image: Meg-Ball Abutment Design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Image: Multi-Unit Abutment Design                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                 |
| Compatibility                           | Neodent Grand Morse,<br>NobelActive®, Straumann BLX<br>TorcFit™, Zimmer Tapered<br>Screw-Vent®                                                                                                                                                                                                                                                  | NobelActive, NobelParallel and<br>NobelReplace Conical<br>Connection                                                                                                                                                                                                                                                                          | MegaGen AnyRidge Internal<br>Fixture, AnyRidge Octa 1 Fixture,<br>ExFeel Internal Fixture                                                                                                                                                                                                                                                                                                                                                                                                                          | Astra Tech EV, Astra Tech OsseoSpeed,<br>BioHorizons Internal, Biomet 3i<br>OSSEOTITE®, Dentium Superline, DESS<br>Active, PrimaConnex, Genesis, Molaris,<br>Paltop, MIS Seven, MIS M4, MegaGen<br>AnyRidge, Neodent Grand Morse,<br>NobelActive®, NobelParallel Conical. | Similar connections<br>as Predicate and<br>Reference Device<br>#2.                                                                                                                  |
| Prosthesis<br>Attachment                | Screw Retained                                                                                                                                                                                                                                                                                                                                  | Screw Retained                                                                                                                                                                                                                                                                                                                                | Snap Retained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Screw Retained                                                                                                                                                                                                                                                            | Similar to Primary<br>Predicate                                                                                                                                                     |
| Restoration                             | Multi-Unit                                                                                                                                                                                                                                                                                                                                      | Multi-Unit                                                                                                                                                                                                                                                                                                                                    | Multi-unit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Single and Multi-Unit                                                                                                                                                                                                                                                     | Similar to Primary<br>Predicate                                                                                                                                                     |
| Abutment<br>Collar Height               | 2.3 mm                                                                                                                                                                                                                                                                                                                                          | 1.5, 2.5, 3.5, 4.5 mm                                                                                                                                                                                                                                                                                                                         | 0-6.0mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 1.0mm-5.5mm                                                                                                                                                                                                                                                               | Similar.<br>Height is within<br>cleared range of<br>Primary Predicate<br>and Reference<br>Devices.                                                                                  |
| Platform<br>Diameter                    | 4.8mm                                                                                                                                                                                                                                                                                                                                           | 4.8mm                                                                                                                                                                                                                                                                                                                                         | 5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 4.8mm                                                                                                                                                                                                                                                                     | Identical to Primary<br>Predicate                                                                                                                                                   |
| Abutment Angle                          | Straight to 30°                                                                                                                                                                                                                                                                                                                                 | Straight to 30°                                                                                                                                                                                                                                                                                                                               | 0°, 5°, 10°, 15°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Straight to 30°                                                                                                                                                                                                                                                           | Similar to Primary<br>Predicate                                                                                                                                                     |
| Abutment<br>Material                    | Ti-6Al-4V ELI                                                                                                                                                                                                                                                                                                                                   | Ti-6Al-4V ELI                                                                                                                                                                                                                                                                                                                                 | Ti-6Al-4V ELI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Ti-6Al-4V ELI                                                                                                                                                                                                                                                             | Identical to Primary<br>Predicate                                                                                                                                                   |
| Abutment<br>Fixation                    | Abutment fixation with an<br>integral screw                                                                                                                                                                                                                                                                                                     | Abutment fixation with an<br>integral screw or retaining<br>screw                                                                                                                                                                                                                                                                             | Abutment fixation with an integral<br>screw                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Abutment fixation with integral screw<br>or retaining screw                                                                                                                                                                                                               | Identical to Primary<br>Predicate                                                                                                                                                   |
| Abutment/<br>Implant Interface          | Internal                                                                                                                                                                                                                                                                                                                                        | Internal                                                                                                                                                                                                                                                                                                                                      | Internal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Internal, External                                                                                                                                                                                                                                                        | Identical to Primary<br>Predicate                                                                                                                                                   |
| Sterility                               | Non-sterile                                                                                                                                                                                                                                                                                                                                     | Sterile                                                                                                                                                                                                                                                                                                                                       | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Non-sterile                                                                                                                                                                                                                                                               | Identical to<br>Reference Devices<br>#1 and #2                                                                                                                                      |
| Usage                                   | Single patient, single use                                                                                                                                                                                                                                                                                                                      | Single patient, single use                                                                                                                                                                                                                                                                                                                    | Single patient, single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Single patient, single use                                                                                                                                                                                                                                                | Identical to Primary<br>Predicate                                                                                                                                                   |
| Performance Testing                     |                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |…

---

**Source:** [https://fda.innolitics.com/device/K242064](https://fda.innolitics.com/device/K242064)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
