1CMR Pro

K242062 · Mycardium AI Limited · LLZ · Nov 15, 2024 · Radiology

Device Facts

Record IDK242062
Device Name1CMR Pro
ApplicantMycardium AI Limited
Product CodeLLZ · Radiology
Decision DateNov 15, 2024
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

Intended Use

1CMR Pro is software that displays, analyses and transfers DICOM cardiovascular images acquired in Cardiovascular Magnetic Resonance (CMR) scanners, specifically structure, function and flow in the heart and major vessels using multi-slice, multi-parametric and velocity encoded CMR images. It is compatible with 1.5T and 3T CMR acquisitions. The intended patient population is both known healthy patients in whom an underlying cardiac disease is suspected. The standard viewing tools are indicated for all patients. The Al analysis components are not intended for use in patients with a known congenital cardiac abnormality, children (Age<18), or individuals with pacemakers (even if MRI compatible).

Device Story

1CMR Pro is a software application for displaying, analyzing, and transferring DICOM cardiovascular magnetic resonance (CMR) images. It processes multi-slice, multi-parametric, and velocity-encoded images from 1.5T and 3T scanners to assess cardiac structure, function, and flow. The device utilizes AI-based analysis to perform automated segmentation and quantification of cardiac parameters, including volumes (LV, RV, LA, RA), LV mass, and functional metrics (LVEF, RVEF, TAPSE, MAPSE, GLS). Clinicians use the software to review automated outputs, which assist in clinical decision-making regarding cardiac health. The device is intended for use in clinical settings by healthcare professionals. By providing automated, precise measurements, the device aims to improve the accuracy and consistency of cardiac image analysis compared to manual clinician assessment, potentially benefiting patients through more reliable diagnostic information.

Clinical Evidence

Non-clinical performance testing on independent datasets. Accuracy assessed in 64 adults (18-84 years) across various cardiac pathologies; AI achieved mean DICE scores of 0.90 (vs. 0.89 for human truthers). Precision assessed in 110 adults scanned twice; 1CMR Pro demonstrated superior precision (lower coefficient of variation) compared to clinicians for LVEF (4.3% vs 7.0%, p<0.001), LV mass (3.8% vs 4.6%, p<0.001), and LVEDV (4.9% vs 6.2%, p<0.001). Comparison against Circle CVI v 5.13 showed 1CMR Pro exceeded performance in all measured variables (e.g., LVEF 4.2% vs 10.4%, p<0.001).

Technological Characteristics

Software-based medical image management and processing system. Compatible with 1.5T and 3T CMR acquisitions. Implements AI-based automated segmentation and quantification algorithms. Complies with EN ISO 14971 (risk management), IEC 62304 (software lifecycle), and IEC 62366-1/2 (usability).

Indications for Use

Indicated for adults (age >=18) with suspected cardiac disease. Standard viewing tools are for all patients. AI analysis components are contraindicated for patients with known congenital cardiac abnormalities, children (<18 years), or individuals with pacemakers (including MRI-compatible models).

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 15, 2024 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". Mycardium AI Limited % Michael Pogose Director of Quality Assurance and Regulatory Affairs Hardian Ltd t/a Hardian Health c/o Galloways, 3rd Floor 21 Perrymount Road Haywards Heath, RH16 3TP United Kingdom Re: K242062 Trade/Device Name: 1CMR Pro Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: October 18, 2024 Received: October 18, 2024 Dear Michael Pogose: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, D. R. K. Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) #### K242062 Device Name 1CMR Pro Indications for Use (Describe) 1CMR Pro is software that displays, analyses and transfers DICOM cardiovascular images acquired in Cardiovascular Magnetic Resonance (CMR) scanners, specifically structure, function and flow in the heart and major vessels using multi-slice, multi-parametric and velocity encoded CMR images. It is compatible with 1.5T and 3T CMR acquisitions. The intended patient population is both known healthy patients in whom an underlying cardiac disease is suspected. The standard viewing tools are indicated for all patients. The Al analysis components are not intended for use in patients with a known congenital cardiac abnormality, children (Age<18), or individuals with pacemakers (even if MRI compatible). Type of Use (Select one or both, as applicable) | X | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ 510(k) #: K242062 510(k) Summary Prepared on: 2024-11-15 ### Contact Details 21 CFR 807.92(a)(1) | Applicant Name | Mycardium Al Limited | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------| | Applicant Address | The Spine, 2 Paddington Village Liverpool L7 3FA United Kingdom | | | Applicant Contact Telephone | +44 7812167512 | | | Applicant Contact | Mr. Michael Walker | | | Applicant Contact Email | michael.walker@mycardium.com | | | Correspondent Name | Hardian Ltd t/a Hardian Health | | | Correspondent Address | c/o Galloways, 3rd Floor 21 Perrymount Road Haywards Heath<br>RH16 3TP United Kingdom | | | Correspondent Contact Telephone | +44 7958656882 | | | Correspondent Contact | Mr. Michael Pogose | | | Correspondent Contact Email | mike@hardianhealth.com | | | Device Name<br>21 CFR 807.92(a)(2) | | | | Device Trade Name | 1CMR Pro | | | Common Name | Medical image management and processing system | | | Classification Name | Automated Radiological Image Processing Software | | | Regulation Number | 892.2050 | | | Product Code(s) | QIH, LLZ | | | Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K220624 | AI4CMR | LLZ | | K213998 | Cvi42 auto | QIH | | Device Description Summary | | 21 CFR 807.92(a)(4) | | 1CMR Pro is software that displays, analyses and transfers DICOM cardiovascular images acquired in Cardiovascular Magnetic Resonance<br>(CMR) scanners, specifically structure, function and flow in the heart and major vessels using multi-slice, multi-phase, multi-parametric<br>and velocity encoded CMR images. It is compatible with 1.5T and 3T CMR acquisitions. | | | | Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | 1CMR Pro is software that displays, analyses and transfers DICOM cardiovascular Magnetic Resonance {5}------------------------------------------------ (CMR) scanners, specifically structure, function and flow in the heart and major vessels using multi-phase, multi-parametric and velocity encoded CMR images. It is compatible with 1.5T and 3T CMR acquisitions. The intended patient population is both known heathy patients in whom an underlying cardiac disease is suspected. The standard viewing tools are indicated for all patients. The standard viewing tools are indicated for all patients. The Al analysis components are not intended for use in patients with a known congenital cardiac abnormality and children (Age 18) or pacemakers (even if MRI compatible). ## Indications for Use Comparison Age: It is acknowledged that the CDRH typically defines adults as individuals greater than or equal to 22 years of age, and those between the ages of 18-21 as older adolescents, which may require special consideration. However, numerous SCMR and American College of Cardiology consensus statements and clinical practice quidelines for a number of cardiac pathologies define children as anyone below the age of 18 years of age, and adults greater than or equal to 18 years of age, with no special consideration or expected differences between those aged 18-21 vs. >= 22 years of age. As such, this subgroup was not independently evaluation of 1CMR Pro: adults have been considered to be people >= the age of 18 years. # Technological Comparison There is a minor difference with the primary predicate - 1CMR Pro provides a user interface to allow users the predicate, AI4CMR, requires the user incorporating it into their DICOM application of choice to be responsible for implementing a user interface - however the intended use for Cardiac MRI image analysis is the same. There is a minor difference with the secondary predicate - the secondary predicate device can also be used on CT images and can be used for calcium scoring and to confirm the presence of physician identified lesions - however the use with Cardiac MRI (CMR) images is the same. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non-clinical testing included the use of data obtained through the internal risk management process, software verification and validation testing, usability engineering, internal validation. Testing was conducted to verify compliance with specified requirements in accordance with EN ISO 14971, IEC 62304, IEC 62366-1 and IEC 62366-2. Verfication and validation testing were conducted to ensure specifications and performance of the device and were performed per the FDA Guidance documents "Guidance for the Content of Premarket Submissions for Software in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices". 1CMR Pro was clinically validated through non-clinical performance testing on independent validation datasets, to confirm acceptable accuracy and precision of the output measurements and segmentations. The accuracy of 1CMR Pro was tested on a dataset of 64 adults (ranging between 18 and 84 vears, 39% female, with a mixture of cardiac pathologies [healthy, dilated cardiomyopathy, hypertrophic cardiomyopathy, myocardial infarction, left ventrioular hypertrophy]), using images collected on either 1.5T or 3T scanners (range of Siemens, Philips, and GE). This comprised 3 types of assessment: DICE sores, Accuracy and Precision. This was done by 3 independent US based truthers, all with >5 years experience. Results: DICE scores were used for the LV short axis contours. The Al was superior to truthers (In brief, Al vs. Truther overall DICE scores averaged 0.90 compared to Truther DICE scores average 0.89). Both humans and Al were more variable at the base of the heart. Results: Accuracy: Accuracy was assessed for 14 variables: volumes (LV and RV: EDV, ESV, SV: LA, RA), LV mass, function (LVEF, RVEF, TAPSE, MAPSE, GLS). The Al passed all assessments and the Al exceeded that the Al exceeded that of the truthers. There were variations in the performance: the LV was more accurate than the RV, analysis in health more accurate than in disease. There were no differences by age or ethnicity. Performance was higher for Siemens scanners – note the sample size was smaller for non-Siemens scanners (53 vs. 11 [8 Philips + 3 GE]). The precision of 1CMR was tested on a dataset of 110 adults (ranging from 22 to 72 years, 35% female, with a mixture of cardiac pathologies, scanned twice. Results: Precision: Precision was performed for LV variables. The Al passed was superior to humans for all measurements and prior FDA cleared software. Results include: Al vs Clinician coefficient of variation (CoV) for LVEF: 4.3±0.3% vs 7.0±0.6%, p<0.001. Al vs Clinician CoV for LVmass: 3.8±0.3% vs 4.6±0.3%, p<0.001, Al vs Clinician CoV for LVEDV: 4.9±0.4% vs 6.2±0.5% p=0<0.001, Al vs Clinician CoV for LVESV: (5.4 (4.3-6.4%) vs 11.4 (6.5-15.6%, p=0.008). ## 21 CFR 807.92(a)(5) ### 21 CFR 807.92(a)(6) {6}------------------------------------------------ In addition, 1CMR Pro was assessed against other cleared software (Circle CVI v 5.13). The testing here used no human editing. 1CMR Pro exceeded the performance in every measured variable: LVEF: 4.2% (95% Cl: 3.5-5.0%) vs 10.4% (95% Cl: 6.8-14.0%), p<0.001; LV mass: 4.2% (95% Cl: 3.5-5.0%) vs 10.4% (95% Cl: 6.8-14.0%), p<0.001; LVEDV: 5.4% (95% Cl: 4.3-6.4%) vs 11.4 (95% Cl: 6.5-15.6%), p=0.008). Based on the information submitted in this premarket notifications for use, technological characteristics and performance testing, 1CMR Pro raises no new questions of safety and is substantially equivalent to the predicate device in terms of safety and effectiveness.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%