K242023 · Implant Microdent System S.L.U. · KCT · Feb 19, 2025 · General Hospital
Device Facts
Record ID
K242023
Device Name
Microdent Sterilization Cassette
Applicant
Implant Microdent System S.L.U.
Product Code
KCT · General Hospital
Decision Date
Feb 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Microdent Sterilization Cassette is intended to organize, protect, and store various dental surgical drills and tools in order to organize and facilitate the sterilization process by allowing steam penetration and air removal. When used in conjunction with FDA-cleared sterilization accessories (wrap, biological indicators, and chemical indicators) in an FDA-cleared sterilizer, sterility of the enclosed medical devices is maintained until used. The sterilization cassette is intended for sterilization in a pre-vacuum steam sterilizer utilizing this recommended cycle: Temperature: 270°F/132°C Exposure time: 4 minutes. Drying time: 20 minutes. The cassettes are not to be stacked during sterilization.
Device Story
Microdent Sterilization Cassette is a reusable container system for dental surgical instruments. It consists of a plastic base and cover with an evenly distributed hole pattern to facilitate steam penetration and air removal during sterilization. Accessories like silicone mats may be used to organize or separate contents. The device is used in clinical settings by dental staff to prepare instruments for sterilization. The cassette is placed in an FDA-cleared pre-vacuum steam sterilizer (132°C for 4 minutes, 20-minute dry time). The cassette, when wrapped with FDA-cleared sterilization wrap, maintains the sterility of the enclosed instruments until use. It benefits patients by ensuring dental tools are properly organized and sterilized, reducing the risk of contamination during surgical procedures.
Clinical Evidence
No clinical data. Bench testing only. Validation included sterilization cycle efficacy (SAL ≤ 10^-6), drying validation (no visible moisture), cleaning validation (protein < 6.4 µg/cm², hemoglobin < 2.2 µg/cm²), and durability testing (100 reprocessing cycles without deformation or corrosion).
Technological Characteristics
Materials: Plastic and metal. Design: Base and cover with evenly distributed hole pattern; includes silicone mats. Sterilization: Pre-vacuum steam (132°C, 4 min exposure, 20 min dry). Standards: ISO 17665-2:2009, ANSI/AAMI/ISO 17665-1:2006, ANSI/AAMI ST77:2013, AAMI TIR30:2011. Reusable. Connectivity: None.
Indications for Use
Indicated for organizing, protecting, and storing dental surgical drills and tools to facilitate steam sterilization in pre-vacuum steam sterilizers. Intended for use with FDA-cleared sterilization accessories (wrap, biological/chemical indicators).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
PolyVac Surgical Instrument Delivery System (K012105)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 19, 2025
Implant Microdent System S.L.U. Joan Muñoz COO and R&D Director C/Carles Buïgues, 1 1 Pol. Ind. Can Magre Santa Eulalia de Ronçana, Barcelona 08187 Spain
Re: K242023
Trade/Device Name: Microdent Sterilization Cassette Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: January 23, 2025 Received: January 23, 2025
Dear Joan Muñoz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Stephen A. Anisko -S
Digitally signed by Stephen A. Anisko -S Date: 2025.02.19 13:34:26 -05'00'
for: Christopher K. Dugard Division Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242023
Device Name Microdent Sterilization Cassette
#### Indications for Use (Describe)
Microdent Sterlization Cassette is intended to organize, protect, and store various dental surgical drills and tools in order to organize and facilitate the sterilization process by allowing steam penetration and air removal.
When used in conjunction with FDA-cleared sterilization accessories (wrap, biological indicators, and chemical indicators) in an FDA-cleared sterilizer, sterility of the enclosed medical devices is maintained until used.
The sterilization cassette is intended for sterilization in a pre-vacuum steam sterilizer utilizing this recommended cycle:
- Temperature: 270°F/132°C
- Exposure time: 4 minutes.
- Drying time: 20 minutes.
- The cassettes are not to be stacked during sterilization.
- KTF and KBE references represent the worst case validated load due to the total weight of 509 grams and 220 grams respectively.
- Implant Microdent System S.L. does not make any lumen claims for the Microdent Sterilization Cassette.
The following list of sterilization cassette models are available and the corresponding instrument load weight that can be sterilized in each model is as listed.
| REFERENCE | WEIGHT | DIMENSIONS | VOLUME TO VENT RATIO |
|-----------|--------|----------------|---------------------------|
| KCIGN | 481g | 190x140x6.5mm | 3.38 mm<br>(0.133 inch) |
| KCIEK | 481g | 190x140x6.5mm | 3.38 mm<br>(0.133 inch) |
| KIO | 450g | 200x105x34mm | 27.737 mm<br>(1.092 inch) |
| KFM | 450g | 190x140x61.5mm | 3.38 mm<br>(0.133 inch) |
| KTF | 509g | 184x136x31mm | 131.98 mm<br>(5.196 inch) |
| KEO | 464g | 190x140x61.5mm | 3.38 mm<br>(0.133 inch) |
| KBE | 220g | 143x79x23mm | 24.36 mm<br>(0.959 inch) |
| KCF | 262g | 143x99.5x61mm | 3.302 mm<br>(0.130 inch) |
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Microdent Implant System. The logo features the letters "MD" inside of a circle, with three additional circles trailing off to the right. Below the circle graphic, the word "MICRODENT" is written in a large, bold, sans-serif font, with the words "IMPLANT SYSTEM" written in a smaller font below it.
# 510(k) Summary
K242023
| DATE PREPARED: | 2025-02-10 |
|-------------------------|--------------------------------------------------------------------------------------|
| SUBMITTER NAME: | Implant Microdent Systems S.L.U. |
| SUBMITTER ADDRESS: | C/ Carles Buigas, 1 - Can Magre<br>08187, Sta Eulalia de Ronçana, Barcelona<br>SPAIN |
| CONTACT: | Marcela Gil |
| TELEPHONE: | +34 93 544 2003 |
| e-mail: | marcela@microdent.com |
| DEVICE TRADE NAME: | Microdent Sterilization Cassette |
| COMMON NAME: | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories |
| REGULATION DESCRIPTION: | Sterilization wrap |
| REGULATION NUMBER: | 21 CFR 880.6850 |
| PRODUCT CODE: | KCT |
| PREDICATE DEVICE(S): | PolyVac Inc (PolyVac Surgical Instrument Delivery System) (K012105) |
# DEVICE DESCRIPTION:
Microdent Sterilization Cassette consists of different sizes of the same basic configuration (maximum weight 509 gr). All systems consist of a minimum of a plastic base and cover. Each cover can be fastened to its corresponding base by means of locking tab or guides. Accessories may be used with systems to organize or separate contents to be placed in them for use.
The Delivery Systems are designed using plastic and metal materials that can be reused with steam sterilization methods. Each tray and cover have an evenly distributed hole pattern in relation to its size.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Microdent Implant System. The logo features the text "MICRODENT IMPLANT SYSTEM" in bold, black letters. Above the text is a graphic of overlapping circles with the letters "MD" inside the first circle. Below the text is the phrase "INDICATIONS FOR USE:" in bold, black letters.
Microdent Sterilization Cassette is intended to organize, and store various dental surgical drills and tools in order to organize and facilitate the sterilization process by allowing steam penetration and air removal.
When used in conjunction with FDA-cleared sterilization accessories (wrap, biological indicators, and chemical indicators) in an FDA-cleared sterilizer, sterility of the enclosed medical devices is maintained until used.
The sterilization cassette is intended for sterilization in a pre-vacuum steam sterilizer utilizing this recommended cycle:
- -Temperature: 270°F/132°C
- -Exposure time: 4 minutes.
- -Drying time: 20 minutes.
- -The cassettes are not to be stacked during sterilization.
KTF and KBE references represent the worst case validated load due to the total weight of 509 grams and 220 grams respectively.
Implant Microdent System S.L. does not make any lumen claims for the Microdent Sterilization Cassette.
The following list of sterilization cassette models are available and the corresponding instrument load weight that can be sterilized in each model is as listed.
| REFERENCE | WEIGHT | DIMENSIONS | VOLUME<br>TO VENT<br>RATIO |
|-----------|--------|----------------|----------------------------|
| KCIGN | 481g | 190x140x6.5mm | 3.38 mm<br>(0.133 inch) |
| KCIEK | 481g | 190x140x6.5mm | 3.38 mm<br>(0.133 inch) |
| KIO | 450g | 200X105X34mm | 27,737 mm<br>(1.092 inch) |
| KFM | 450g | 190x140x61.5mm | 3.38 mm<br>(0.133 inch) |
| KTF | 509g | 184x136x31mm | 131,98 mm<br>(5.196 inch) |
| KEO | 464g | 190x140x61.5mm | 3.38 mm<br>(0.133 inch) |
| KBE | 220g | 143x79x23mm | 24.36 mm<br>(0.959 inch) |
| KCF | 262g | 143x99.5x61mm | 3,302 mm<br>(0.130 inch) |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Microdent Implant System. The logo includes a stylized "MD" in a circle with curved lines emanating from it. Below the logo, the words "MICRODENT" and "IMPLANT SYSTEM" are written in a bold, sans-serif font. The text "SUMMARY OF COMPARISON WITH PREDICATE DEVICE:" is written in a smaller, sans-serif font below the words "IMPLANT SYSTEM".
In the establishment of a technology comparison, Microdent Sterilization Cassettes are compared with the following previously cleared devices:
| Summary of<br>comparison<br>with predicate<br>devices | Proposed Device<br>(K242023) | Predicate Device<br>(K012015) | Comments |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Appearance | Microdent Sterilization Cassette<br><br>Image: (KBE) | PolyVac Surgical<br>Instrument Delivery<br>System<br><br>Image: (Minitainer II) | Similar |
| Classification /<br>PROCODE | 880.6850 (KCT) | 880.6850 (KCT) | Same |
| Intended Use | Microdent Sterilization Cassette is<br>intended to organize, protect, and<br>store various dental surgical drills<br>and tools in order to organize and<br>facilitate the sterilization process<br>by allowing steam penetration and<br>air removal.<br>When used in conjunction with<br>FDA-cleared sterilization<br>accessories (wrap, biological<br>indicators, and chemical indicators)<br>in an FDA-cleared sterilizer,<br>sterility of the enclosed medical<br>devices is maintained until used.<br>The sterilization cassette is<br>intended for sterilization in a pre-<br>vacuum steam sterilizer utilizing<br>this recommended cycle:<br>Temperature: 270°F/132°C<br>Exposure time: 4 minutes.<br>Drying time: 20 minutes.<br>The cassettes are not to be<br>stacked during sterilization.<br>KTF and KBE references represent<br>the worst case/best case validated<br>load due to the total weight of 509<br>grams and 220 grams respectively.<br>Implant Microdent System S.L.<br>does not make any lumen claims<br>for the Microdent Sterilization<br>Cassette. | PolyVac's delivery<br>systems are intended to<br>protect medical device<br>instrumentation and to<br>facilitate the sterilization<br>process by allowing steam<br>penetration and air<br>removal. When used in<br>conjunction with an<br>approved sterilization<br>wrap, sterility of the<br>enclosed medical device<br>is maintained until used.<br>PolyVac's delivery<br>systems are to be<br>sterilized in one of the<br>following cycles:<br>Prevacuum Steam: 132°C<br>- 4 minutes minimum<br>Dry for 20 - 40 minutes as<br>needed<br>Gravity Steam: 132°C -<br>30 minutes minimum<br>Gravity Steam: 121°C -<br>55 minutes minimum<br>Dry for 20 - 50 minutes as<br>needed | Similar; the subject device<br>does not have a gravity steam<br>sterilization cycle claim. |
| Summary of<br>comparison | Proposed Device<br>(K242023) | Predicate Device<br>(K012105) | Comments |
| with predicate<br>devices | Microdent Sterilization<br>Cassette | PolyVac Surgical<br>Instrument Delivery<br>System | |
| Description | - Base, modular insert trays and<br>covers<br>- Evenly distributed<br>perforated hole pattern.<br>- Silicone mats | - Base, modular insert<br>trays and lids<br>- Evenly distributed<br>perforated hole<br>pattern.<br>- Silicone mats | Similar |
| Dimensions<br>(l x w x h) | Largest: 184 x 136 x 31 mm<br>Smallest: 43 x 79 x 23 mm. | 17.3 x 7.25 x 4 mm | Different |
| Maximum weight | 509 kg | Not stated in summary | N/A |
| Sterilant<br>Penetration (air<br>permeance /<br>perforations) | Evenly distributed hole pattern | Evenly distributed hole<br>pattern | Same |
| Microbial<br>Barrier<br>Properties | To be used in conjunction with an<br>FDA cleared sterilization wrap. | To be used in<br>conjunction with an FDA<br>cleared sterilization<br>wrap. | Same |
| Shelf Life | Reusable | Reusable | Same |
| Steam<br>Sterilization<br>Parameters | - Temperature: 270°F/132°C<br>- Exposure time: 4 minutes.<br>- Drying time: 20 minutes. | Prevacuum Steam:<br>132°C - 4 minutes<br>minimum.<br>Dry for 20-40 minutes as<br>needed<br>Gravity Steam: 132°C -<br>30 minutes minimum<br>Gravity Steam: 121°C -<br>55 minutes minimum<br>Dry for 20 - 50 minutes<br>as needed | Similar |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains a logo for Microdent Implant System. The logo features the word "MICRODENT" in a bold, sans-serif font, with the words "IMPLANT SYSTEM" in a smaller font underneath. Above the word "MICRODENT" is a circular graphic with the letters "MD" inside, followed by a series of curved lines that resemble sound waves or ripples.
Microdent Sterilization Cassettes are similar in fundamental scientific technology to the predicate devices in that they all have been designed, manufactured, and tested in compliance with FDA'S Class II device (product code KCT).
Microdent Sterilization Cassettes are technologically similar in materials, intended use, packaging, labeling and performance to the predicate devices currently marketed in the U.S. The only differences the subject devices and the predicates are in design and dimensions.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains a logo for a company called MICRODENT IMPLANT SYSTEM. The logo features the letters "MD" inside of a circle, with a series of curved lines extending to the right of the circle. The words "MICRODENT" are written in a bold, sans-serif font, and the words "IMPLANT SYSTEM" are written in a smaller, sans-serif font below the word "MICRODENT". The logo is simple and modern, and it is likely used to represent a dental implant company.
## SUMMARY DISCUSSION OF NON-CLINICAL DATA:
The proposed device has been subject to bench testing to determine conformance to performance specifications and requirements taking account of its intended use.
Non-clinical testing listed in the table below and performed in foreseeable operating conditions showed correct operation of the device as per its intended use.
| Summary of Non-Clinical Testing | | | | | |
|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|---------|--|
| Test Name | Test<br>Methodology | Purpose | Acceptance Criteria | Results | |
| Sterilization Cycle<br>Validation | ISO 17665-2:2009<br>ANSI/AAMI/ISO<br>17665-1:2006/<br>(R)2013<br>ANSI / AAMI<br>ST77:2013 | To validate that the trays<br>can be sterilized via moist-<br>heat sterilization as<br>specified on labeling (132°<br>C for 4 minutes) | Sterility assurance level<br>(SAL) of ≤ 10-6 | Pass | |
| Drying Validation | Internal Test Method | To validate that the trays<br>can be dried as specified<br>on labeling (drying time of<br>20 minutes) | No visible moisture | Pass | |
| Cleaning<br>(Manual<br>Pre- Cleaning and<br>Automated<br>Cleaning) | AAMI TIR30:2011 | To validate that the trays<br>can be cleaned as<br>specified on labeling | No visible soil.<br>Protein: < 6.4 µg/cm²<br>Hemoglobin: < 2.2 µg/cm² | Pass | |
| Reprocessing<br>of<br>Trays (cleaning and<br>sterilization) | Internal Test Method | To confirm that the trays<br>can be reprocessed as<br>specified on labeling (up to<br>100 reprocessing cycles<br>without any signs<br>of<br>abrasion) | No signs of flush rust,<br>rust, corrosion,<br>deformation or damage | Pass | |
### SUMMARY DISCUSSION OF CLINICAL DATA
Non-clinical test data are submitted to support this premarket notification. No clinical studies are submitted.
### CONCLUSIONS
The conclusions drawn from the non-clinical tests demonstrate that the proposed device, K242023, is as safe, as effective, and performs as well as or better than the predicate device, K012105.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.