VitruvianScan is a software medical device intended to assist trained healthcare professionals in the noninvasive evaluation of body fat and muscle composition in the general population.
Device Story
VitruvianScan is a standalone, post-processing software medical device used to analyze MRI data. Input consists of DICOM 3.0 compliant MR image datasets (1.5T or 3T) acquired from GE, Siemens, or Philips scanners. Within Perspectum’s secure infrastructure, trained analysts use the software to process images, generating quantified metrics and composite images of visceral fat, subcutaneous fat, and muscle area in the abdomen. Output is provided to healthcare professionals to assist in clinical decision-making regarding patient health and wellness. The device does not perform direct diagnosis; it serves as an adjunct to other clinical information. Benefits include non-invasive, standardized quantification of body composition to support clinical management.
Clinical Evidence
Bench testing only. Performance evaluated across Siemens, GE, and Philips scanners (1.5T and 3T). Metrics assessed for repeatability (same subject/scanner/day), reproducibility (same subject/manufacturer/different field strength/day), and inter/intra-operator variability. Results benchmarked against OSIRIX MD and validated against a gold standard (mean of 3 radiologists' results). All tests met defined acceptance criteria, confirming robust performance across scanners and patient characteristics.
Technological Characteristics
Standalone post-processing software. Inputs: DICOM 3.0 MR image datasets. Compatibility: GE, Siemens, Philips MRI systems (1.5T and 3T). Deployment: Secure cloud-based infrastructure managed by Perspectum. Software functions as a diagnostic aid for image quantification. No hardware components; software-only device.
Indications for Use
Indicated for use as a magnetic resonance diagnostic device software application for non-invasive fat and muscle evaluation that enables the generation, display and review of magnetic resonance medical image data. VitruvianScan produces quantified metrics and composite images from magnetic resonance medical image data which when interpreted by a trained healthcare professional, yield information that may assist in clinical decisions. Intended for the general population (all ages suitable for MRI).
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 2, 2024
Perspectum Ltd Bhaskar Chikkanna Director of Regulatory Affairs Gemini One. 5520 John Smith Drive Oxford, OX4 2LL United Kingdom
Re: K241925
Trade/Device Name: VitruvianScan (v1.0) Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: September 11, 2024 Received: September 11, 2024
### Dear Bhaskar Chikkanna:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241925
Device Name
VitruvianScan (v1.0)
### Indications for Use (Describe)
VitruvianScan is indicated for use as a magnetic resonance diagnostic device software application for non-invasive fat and muscle evaluation that enables the generation, display and review of magnetic resonance medical image data.
VitruvianScan produces quantified metrics and composite images from magnetic resonance medical image data which when interpreted by a trained healthcare professional, yield information that may assist in clinical decisions.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Date Prepared:
25th June 2024
10056574
# 1 Applicant Details
Applicant Name and Address:
25th June 2024
Perspectum Ltd Gemini One, 5520 John Smith Drive, Oxford Business Park, Oxford, OX4 2LL
Owner/Operator Number: Establishment Registration Number: Applicant Contact Name: Applicant Contact Email: Applicant Contact Phone:
3014232555 Bhaskar Chikkanna bhaskar.chikkanna@perspectum.com +44 (0) 1865 655329
# 2 Subject and Predicate Device
| | Subject Device | Predicate Device | Reference Device |
|--------------------|-----------------------------------------|-----------------------------------------|------------------------------------------------|
| 510(k) number | K241925 | K211983 | K101342 |
| Legal Manufacturer | Perspectum Ltd. | AMRA Medical AB | PIXMEO SARL |
| Device Trade Name | VitruvianScan (v1.0) | AMRA Profiler | OSIRIX MD |
| Common Name | Software as a Medical<br>Device | Software as a Medical<br>Device | Picture Archiving<br>Communications System |
| Panel | Radiology | Radiology | Radiology |
| Regulation | 892.1000 | 892.1000 | 892.2050 |
| Risk Class | Class II | Class II | Class II |
| Product Class code | LNH | LNH | LLZ |
| Regulation Name | Magnetic Resonance<br>Diagnostic Device | Magnetic Resonance<br>Diagnostic Device | Picture archiving and<br>communications system |
#### Subject Device Description ന
#### 3.1 General Description
VitruvianScan is a standalone, post processing software medical device. VitruvianScan enables the generation, display and review of magnetic resonance (MR) medical image data from a single timepoint (one patient visit).
When a referring healthcare professional requests quantitative analysis using VitruvianScan, relevant images are acquired from patients at MRI scanning clinics and are transferred to the Perspectum portal through established secure gateways. Perspectum trained analysts use the VitruvianScan software medical device to process the MRI images and produce the quantitative metrics and composite images. The device output information is then sent to the healthcare professionals for their clinical use.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Perspectum" in a bold, blue font. To the right of the word is a circular logo with four different colored sections: light blue, green, pink, and dark blue. The dark blue section is at the bottom of the circle, and the other three colors are arranged in a clockwise direction.
The metrics produced by VitruvianScan are intended to provide insight into the composition of muscle and fat of a patient. The device is intended to be used as part of an overall assessment of a patient's health and wellness and should be interpreted whilst considering the device's limitations when reviewing or interpreting images. Therefore, VitruvianScan presents a low level of concern with respect to patient safety since its metrics only augment the clinical information utilized by trained Healthcare Professionals for the management of patients.
VitruvianScan is only used within Perspectum's secure infrastructure by trained analysts to process clinical images and generate metrics. The metrics and composite images output by VitruvianScan are then sent to trained Healthcare Professionals who then utilize these to make clinical decisions in combination with other relevant information obtained from other clinical sources.
#### 3.2 Intended Use
VitruvianScan is a software medical device intended to assist trained healthcare professionals in the noninvasive evaluation of body fat and muscle composition in the general population.
### 3.3 Indications for Use
VitruvianScan is indicated for use as a magnetic resonance diagnostic device software application for noninvasive fat and muscle evaluation that enables the generation, display and review of magnetic resonance medical image data.
VitruvianScan produces quantified metrics and composite images from magnetic resonance medical image data which when interpreted by a trained healthcare professional, yield information that may assist in clinical decisions.
#### 3.4 Intended Patient Population
General population - VitruvianScan has no demographic or population restrictions.
#### 3.5 Contraindications
- None Software only device .
### 3.6 Warnings/Precautions
- VitruvianScan must not be used for direct diagnosis of patients. o
- o VitruvianScan generated metrics must only be interpreted by trained healthcare professionals.
- VitruvianScan output must not be used on its own to make any treatment decisions relating o to any specific diseases or health conditions.
#### Substantial Equivalence ব
Table below provides a comparison of attributes between the subject device and the predicate device to demonstrate substantial equivalence. Differences between the devices are commented.
| Attributes | Subject Device | Predicate Device | Comments |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device trade name | VitruvianScan | AMRA Profiler | N/A |
| Manufacturer | Perspectum Ltd | AMRA Medical AB | N/A |
| 510(k) number (if<br>assigned) | TBD | K211983 | N/A |
| Regulation | 892.1000 | 892.1000 | No difference |
| Product Code | LNH | LNH | No difference |
| Intended use | VitruvianScan is a<br>software medical device | Not defined | Subject device has a separate<br>intended use statement. |
| | intended to assist trained<br>healthcare professionals<br>in the non-invasive<br>evaluation of body fat<br>and muscle composition<br>in the general population | | |
| Indications for use | VitruvianScan is indicated<br>for use as a magnetic<br>resonance diagnostic<br>device software<br>application for non-<br>invasive fat and muscle<br>evaluation that enables<br>the generation, display<br>and review of magnetic<br>resonance medical image<br>data.<br>VitruvianScan produces<br>quantified metrics and<br>composite images from<br>magnetic resonance<br>medical image data which<br>when interpreted by a<br>trained healthcare<br>professional, yield<br>information that may<br>assist in clinical<br>decisions. | Indicated for use as a<br>magnetic resonance<br>diagnostic device<br>software application for<br>non- invasive fat and<br>muscle evaluation that<br>enables the generation,<br>display and review of 2D<br>magnetic resonance<br>medical image data.<br>Designed to utilize<br>DICOM 3.0 compliant<br>magnetic resonance<br>image datasets, acquired<br>from compatible MR<br>Systems, to display the<br>internal structure of the<br>body including the liver.<br>Other physical<br>parameters derived from<br>the images may also be<br>produced. Provides a<br>number of quantification<br>tools, such as Region of<br>Interest (ROI)<br>placements, to be used<br>for the assessment of<br>regions of an image to<br>quantify liver tissue<br>characteristics, including<br>the determination of fat<br>fraction in the liver, T2 *<br>and muscle volume.<br>These images and the<br>physical parameters<br>derived from the images,<br>when interpreted by a<br>trained clinician, yield<br>information that may<br>assist in diagnosis. | Subject device does not<br>quantify liver tissue<br>characteristics, liver fat<br>fraction, T2* and muscle<br>volume. Other than these<br>differences, both the subject<br>device and the predicate<br>device are indicated for non-<br>invasive fat and muscle<br>evaluation. |
| | | | |
| Intended users | Perspectum trained<br>analysts use the device to<br>analyze MRI data and<br>produce output. | AMRA operators use the<br>device to analyze MRI<br>data and produce<br>results. | Similar users |
| | Trained healthcare<br>professionals interpret<br>the device output for<br>additional support in their<br>clinical decision<br>pathways. | Trained clinicians<br>interpret the results. | |
| Application | Health and wellness | Health and wellness | No difference |
| Anatomy and<br>measurements | Abdomen. | Abdomen and Thighs. | Subject device anatomy and<br>measurements are a subset of<br>the predicate device anatomy<br>and measurements |
| | Visceral Fat | Visceral Fat | |
| | Subcutaneous Fat | Subcutaneous Fat | |
| | Muscle Area | Muscle Fat | |
| | | Muscle Volume | |
| | | Liver Fat | |
| Target population | General population | Adults | Subject device is intended for<br>all ages who are suitable for<br>MRI scans |
| Contraindications | None, Software only | None, Software only | No difference |
| Imaging modality | Magnetic resonance<br>imaging systems | Magnetic resonance<br>imaging systems | No difference |
| Data format | DICOM 3.0 compliant MR<br>image datasets from<br>compatible MR scanners | DICOM 3.0 compliant MR<br>image datasets from<br>compatible MR scanners | No difference |
| MRI Scanners | GE, Siemens and Philips | GE, Siemens and Philips | No difference |
| MR field strength | 1.5T and 3T | 1.5T and 3T | No difference |
{6}------------------------------------------------
## Perspectu D
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Perspectum. The word "Perspectum" is written in a bold, sans-serif font in blue color. To the right of the word is a circular graphic made up of different colored segments. The segments are blue, light blue, green, and pink.
## 5 Software Testing
VitruvianScan is a software medical device. The documentation document justifies only basic documentation is applicable to VitruvianScan as per FDA guidance, Content of Premarket Submissions for Device Software Functions for Industry and Food and Drug Administration Staff Document issued on June 14, 2023. The device has undergone the following tests and successfully passed the acceptance criteria with no residual anomalies.
- a. Unit testing
- b. Integration testing
- c. Software system verification
- d. Software system validation
- e. User acceptance tests
## 6 Performance Testing
VitruvianScan underwent performance testing under controlled conditions to corroborate that it is safe and effective when used as intended. The performance testing conducted demonstrates that the subject device is at least as safe and effective as the predicate device.
| Scanners Assessed |
|-------------------|
| Siemens 1.5T |
| Siemens 3T |
| GE 1.5T |
| GE 3T |
| Philips 1.5T |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Perspectum" in a bold, dark blue font. To the right of the word is a circular logo with four colored sections: light blue, green, pink, and dark blue. The dark blue section is at the bottom of the circle and has a white line running through it.
Philips 3T
The performance was tested for the following aspects:
- Repeatability of metrics for the same subject, on the same manufacturer and field strength, a. acquired on the same day
- b. Reproducibility of metrics for the same subject, on the same manufacturer but a different field strength, acquired on the same day
- c. Characterization of inter-operator variability
- d. Characterization of intra-operator variability
- e. Results from testing using the subject device were compared with the results from testing using reference requlated device `OSIRIX MD' for benchmarking performance
- f. Comparative testing between the operators' results and the gold standard (mean of 3 radiologists results)
All aspects of the performance tests met the defined acceptance criteria, thereby assuring robust clinical performance of the device across different scanners, field strengths and patient characteristics.
## 7 Conclusion
The subject device and the predicate device have similar indications for use and are meant for the same application in health and wellness. The differences highlighted do not constitute a new intended use, are accompanied by information that demonstrates that the device is as safe and effective as the predicate and do not raise different questions of safety and effectiveness than the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.