EFAI Neurosuite CT Midline Shift Assessment System (MLS-CT-100)
Applicant
Ever Fortune.Ai, Co., Ltd.
Product Code
QAS · Radiology
Decision Date
Dec 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K241923 · Dec 6, 2024
EFAI Neurosuite CT Midline Shift Assessment System (MLS-CT-100)
Ever Fortune.Ai, Co., Ltd.
Retrospective clinical CT scans from multiple U.S. clinical sites
A retrospective, multisite clinical validation study was conducted to evaluate the performance of the EFAI MLSCT in identifying midline shift (MLS) on a dataset of 300 clinical cases.
Retrospective study; Clinical validation; Multisite; Head CT
Individuals aged 18 years and above undergoing non-contrast head CT scans; Sample Size: 300 cases (102 positive, 198 negative); Number of Sites: Multiple clinical sites across the United States
Not applicable for this study
Sensitivity, specificity, and AUROC for identifying midline shift (MLS)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Midline shift
Deep learning algorithms
Lower bounds of 95% confidence intervals of sensitivity and specificity > 0.8
EFAI NEUROSUITE CT MIDLINE SHIFT ASSESSMENT SYSTEM (EFAI MLSCT) is a software workflow tool designed to aid in prioritizing the clinical assessment of non-contrast head CT cases with features suggestive of midline shift (MLS) in individuals aged 18 years and above. EFAI MLSCT analyzes cases using deep learning algorithms to identify suspected MLS findings. It makes case-level output available to a PACS/workstation for worklist prioritization or triage. EFAI MLSCT is not intended to direct attention to specific portions of an image or to anomalies other than MLS. Its results are not intended to be used on a stand-alone basis for clinical decision-making nor is it intended to rule out MLS or otherwise preclude clinical assessment of CT studies.
Device Story
EFAI MLSCT is a radiological computer-assisted triage and notification software; analyzes non-contrast head CT scans using deep learning algorithms to identify suspected midline shift (MLS). Device operates on-premise; integrates with PACS/workstation to provide case-level notifications for worklist prioritization. Does not mark or highlight specific image regions; does not provide preview images. Intended to assist radiologists by flagging cases for earlier review; does not replace standard of care or clinical assessment. Benefits include improved workflow efficiency for time-sensitive neurological cases. Clinicians receive notifications and must review full original images for final diagnosis.
Clinical Evidence
Retrospective, blinded, multisite clinical validation study (U.S. sites) using 300 cases (102 positive, 198 negative). Reference standard established by majority agreement of three board-certified radiologists. Results: Sensitivity 0.961 (95% CI: 0.903-0.985), Specificity 0.955 (95% CI: 0.916-0.973), AUROC 0.983 (95% CI: 0.967-0.996). Average processing time 62.04 seconds. Subgroup analysis confirmed consistent performance across diverse demographics, CT manufacturers, and slice thicknesses, as well as in the presence of confounders like intracranial hemorrhage or artifacts.
Technological Characteristics
Software-only radiological triage system; deployed on-premise. Uses deep learning algorithms for image analysis. Compatible with non-contrast head CT scans from various manufacturers (Siemens, Toshiba, GE, Philips, Canon). Compliant with IEC 62304:2006/A1:2016 for software life cycle and ISO 14971:2019 for risk management.
Indications for Use
Indicated for prioritizing clinical assessment of non-contrast head CT cases with features suggestive of midline shift (MLS) in individuals aged 18 years and above.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Ever Fortune.AI. Co., Ltd. Dr. Ti-Hao Wang Chief Technology Officer 8F. No. 360, Sec.1 Jingmao Rd., Beitun Dist. Taichung City, 406040 Taiwan
December 6, 2024
## Re: K241923
Trade/Device Name: EFAI Neurosuite CT Midline Shift Assessment System (MLS-CT-100) Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological Computer Aided Triage And Notification Software Regulatory Class: Class II Product Code: QAS Dated: June 21, 2024 Received: November 18, 2024
Dear Dr. Ti-Hao Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241923
Device Name
EFAINEUROSUITE CT MIDLINE SHIFT ASSESSMENT SYSTEM (MLS-CT-100)
Indications for Use (Describe)
EFAI NEUROSUITE CT MIDLINE SHIFT ASSESSMENT SYSTEM (EFAI MLSCT) is a software workflow tool designed to aid in prioritizing the clinical assessment of non-contrast head CT cases with features suggestive of midline shift (MLS) in individuals aged 18 years and above. EFAI MLSCT analyzes cases using deep learning algorithms to identify suspected MLS findings. It makes caselevel output available to a PACS/workstation for worklist prioritization or triage.
EFAI MLSCT is not intended to direct attention to specific portions of an image or to anomalies other than MLS. Its results are not intended to be used on a stand-alone basis for clinical decision-making nor is it intended to rule out MLS or otherwise preclude clinical assessment of CT studies.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
0ver-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for a company called "EVER FORTUNE.AI". The logo consists of a stylized human figure with a circular head containing a network of interconnected dots, all in shades of green and teal. To the right of the figure, the company name is written in teal, with "EVER" on top and "FORTUNE.AI" below, the "O" in fortune is replaced with the same circular head containing a network of interconnected dots.
# 510(k) Summary
#### General Information 1.
| 510(k) Sponsor | Ever Fortune.AI Co., Ltd. |
|-----------------------|---------------------------------------------------------------------------------------|
| Address | 8F., No.360, Sec. 1, Jingmao Rd.,<br>Beitun Dist.,<br>Taichung City 406040,<br>Taiwan |
| Applicant | Joseph Chang |
| Contact Information | 886-04-23213838 #216<br>joseph.chang@everfortune.ai |
| Correspondence Person | Ti-Hao Wang |
| Contact Information | 886-04-23213838 #168<br>thothwang@gmail.com<br>tihao.wang@everfortune.ai |
| Date Prepared | November, 2024 |
#### Proposed Device 2.
| Proprietary Name | EFAI NEUROSUITE CT MIDLINE SHIFT ASSESSMENT SYSTEM (MLS-CT-100) |
|---------------------|-----------------------------------------------------------------|
| Common Name | EFAI MLSCT |
| Classification Name | Radiological computer-assisted triage and notification software |
| Regulation Number | 21 CFR 892.2080 |
| Product Code | QAS |
| Regulatory Class | II |
#### 3. Predicate Device
| Proprietary Name | qER |
|------------------------|-----------------------------------------------------------------|
| Premarket Notification | K200921 |
| Classification Name | Radiological computer-assisted triage and notification software |
| Regulation Number | 21 CFR 892.2080 |
| Product Code | QAS |
| Regulatory Class | II |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Ever Fortune AI. The logo consists of a stylized figure of a person with a head made of a network of interconnected dots. The figure is teal in color. To the right of the figure is the text "EVER FORTUNE.AI", also in teal.
#### Device Description 4.
EFAI NEUROSUITE CT MIDLINE SHIFT ASSESSMENT SYSTEM (EFAI MLSCT) is a radiological computer-assisted triage and notification software system. The software uses deep learning techniques to automatically analyze non-contrast head CTs and alerts the PACS/RIS workstation once images with features suggestive of MLS are identified.
Through the use of EFAI MLSCT, a radiologist is able to review studies with features suggestive of MLS earlier than in standard of care workflow.
The device is intended to provide a passive notification through the PACS/workstation to the radiologists indicating the existence of a case that may potentially benefit from the prioritization. It does not mark, highlight, or direct users' attention to a specific location on the original non-contrast head CT. The device aims to aid in prioritization and triage of radiological medical images only.
#### ર. Intended Use / Indications for Use
EFAI NEUROSUITE CT MIDLINE SHIFT ASSESSMENT SYSTEM (EFAI MLSCT) is a software workflow tool designed to aid in prioritizing the clinical assessment of non-contrast head CT cases with features suggestive of midline shift (MLS) in individuals aged 18 years and above. EFAI MLSCT analyzes cases using deep learning algorithms to identify suspected MLS findings. It makes case-level output available to a PACS/workstation for worklist prioritization or triage.
EFAI MLSCT is not intended to direct attention to specific portions of an image or to anomalies other than MLS. Its results are not intended to be used on a stand-alone basis for clinical decision-making nor is it intended to rule out MLS or otherwise preclude clinical assessment of CT studies.
#### Comparison of Technological Characteristics with Predicate Device 6.
| Feature/<br>Function | Proposed Device:<br>EFAI MLSCT | Predicate Device:<br><i>qER</i><br>(K200921) |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/Indication<br>for Use | EFAI NEUROSUITE CT MIDLINE<br>SHIFT ASSESSMENT SYSTEM<br>(EFAI MLSCT) is a software<br>workflow tool designed to aid in | <i>qER</i> is a radiological computer aided<br>triage and notification software<br>indicated for use in the analysis of<br>non-contrast head CT images. |
| | prioritizing the clinical assessment of<br>non-contrast head CT cases with<br>features suggestive of midline shift<br>(MLS) in individuals aged 18 years<br>and above. EFAI MLSCT analyzes<br>cases using deep learning algorithms<br>to identify suspected MLS findings. It<br>makes case-level output available to a<br>PACS/workstation for worklist<br>prioritization or triage.<br>EFAI MLSCT is not intended to direct<br>attention to specific portions of an<br>image or to anomalies other than<br>MLS. Its results are not intended to be<br>used on a stand-alone basis for clinical<br>decision-making nor is it intended to<br>rule out MLS or otherwise preclude<br>clinical assessment of CT studies. | The device is intended to assist<br>hospital networks and trained medical<br>specialists in workflow triage by<br>flagging the following suspected<br>positive findings of pathologies in<br>head CT images: intracranial<br>hemorrhage, mass effect, midline<br>shift and cranial fracture.<br>qER uses an artificial intelligence<br>algorithm to analyze images on a<br>standalone cloud-based application in<br>parallel to the ongoing standard of care<br>image interpretation. The user is<br>presented with notifications for cases<br>with suspected findings.<br>Notifications include non-diagnostic<br>preview images that are meant for<br>informational purposes only. The<br>device does not alter the original<br>medical image and is not intended to<br>be used as a diagnostic device.<br>The results of the device are intended<br>to be used in conjunction with other<br>patient information and based on<br>professional judgment, to assist with<br>triage/prioritization of medical images.<br>Notified clinicians are responsible for<br>viewing full images per the standard of<br>care. |
| Classification/<br>Product Code | 21 CFR 892.2080/QAS | 21 CFR 892.2080/QAS |
| Anatomical<br>region of interest | Head | Head |
| Data acquisition<br>protocol | Non contrast CT scan of the head | Non contrast CT scan of the head |
| Segmentation of<br>region of interest | No; device does not mark, highlight,<br>or direct users' attention to a<br>specific location in the original<br>image. | No; device does not mark, highlight,<br>or direct users' attention to a<br>specific location in the original<br>image. |
| Algorithm | Artificial intelligence algorithm with<br>database of images. | Artificial intelligence algorithm with<br>database of images. |
| Notification/<br>Prioritization | Yes, with midline shift as the only<br>output, there are no controls for the<br>user to select the finding or<br>combination of findings for triage | Yes, with controls for the user to select<br>the finding or combination of findings<br>for triage |
| Preview images | No | Presentation of a preview of the study<br>for initial assessment not meant for<br>diagnostic purposes. The device<br>operates in parallel with the standard<br>of care, which remains the default<br>option for all cases. |
| Alteration of<br>original image | No | No |
| Removal of cases<br>from worklist<br>queue | No | No |
| Abnormalities<br>triaged | finding: midline shift | Four findings: Intracranial<br>haemorrhage, Mass effect, midline<br>shift, cranial fracture and ICH |
| User control over<br>triage | Triages non-contrast CT scan only<br>when midline shift is identified. No<br>user control for configuring the triage<br>finding. | Triages non-contrast CT scan when<br>any of the 4 abnormalities are<br>identified.<br>Notification/ prioritization with user<br>provided control for configuring the<br>triage finding or combination of<br>finding(s) |
| Device<br>Deployment | deployed only on-premise | Performs de-identification of studies<br>on-premise and pushes them to the<br>cloud module for analysis. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Ever Fortune AI. The logo consists of a stylized human figure in teal, with a green circle containing a network of white dots at the head. To the right of the figure, the words "EVER" and "FORTUNE.AI" are written in teal. The word "FORTUNE.AI" is written in a smaller font than the word "EVER", and the dot in the "O" of "FORTUNE" is replaced with a green circle containing a network of white dots.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for a company called "EVER FORTUNE.AI". The logo consists of a stylized figure of a person with a green head and a turquoise body. The head is a green sphere with a network of white lines and dots on it. The text "EVER" is in large, turquoise letters, and the text "FORTUNE.AI" is in smaller, turquoise letters below it. The logo is simple and modern, and it conveys a sense of innovation and technology.
#### 7. Performance Data
Performance of the EFAI MLSCT has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Additionally, the software validation activities were performed in accordance with IEC 62304:2006/A1:2016 - Medical device software - Software life cycle processes, in addition to the FDA Guidance documents, "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions."
Ever Fortune.AI conducted a retrospective, blinded, multisite clinical validation study with the proposed device EFAI MLSCT with a pre-determined primary and secondary endpoint and
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains a logo for a company called "EVER FORTUNE.AI". The logo consists of a stylized human figure with a circular head and a body that resembles the letter "T". The head is green and has a network of interconnected dots, suggesting a connection or technology theme. To the right of the figure, the company name is written in a teal color, with "EVER" stacked above "FORTUNE.AI".
performance goals to evaluate the performance of the EFAI MLSCT in identifying midline shift (MLS) from non-contrast head computed tomography (CT) scans on a validation dataset of 300 cases (102 positives and 198 negatively consecutively collected from multiple clinical sites across the United States (U.S.). Each case included only one CT study. None of the cases was used as part of the EFAI MLSCT model development or analytical validation testing, as the U.S. cases were solely collected for this study, while the model development and validation utilized cases from Taiwan.
The study population contained 51.67% females and 48.33% males, and the mean age of cases was 59.46 years. The CT scanner manufacturers of images were acquired from Siemens, Toshiba, GE Medical Systems, Philips, and Canon Medical Systems. Potential confounding cases in the dataset include artifact, poor patient position issue, intracranial hemorrhage, ischemic-stroke and related features, space-occupying lesions, and others.
The presence of MLS in each case was determined independently by three U.S. board-certified radiologists, and the reference standard (ground truth) was generated by the majority agreement between the three experts. The performance acceptance criteria were set such that the lower bounds of 95% confidence intervals (Cls) of both sensitivity and specificity should exceed 0.8.
The observed results of the standalone performance validation study demonstrated that EFAI MLSCT by itself, in the absence of any interaction with a clinician, can provide case-level notifications with features suggestive of MLS with satisfactory results. The EFAI MLSCT was able to demonstrate sensitivity and specificity of 0.961 (95% CI=0.903-0.985) and 0.955 (95% CI=0.916-0.973) respectively, along with an AUROC of 0.983 (95% CI=0.967-0.996), which is substantially equivalent to the predicate device (qER, K200921). The secondary endpoint of the observed system processing time per study is 62.04 seconds (95% Cl=60.65-63.44) on average and was significantly less than the pre-specified performance goal.
In addition, the results of the subgroup analysis, which included different genders, age groups, race or ethnicity groups, CT manufacturer groups, and CT slice thickness groups, demonstrated that EFAI MLSCT consistently performed high performance, underscoring its reliability and effectiveness across diverse subgroups. We also evaluated the device's performance in cases with image quality issues (including artifact and poor patient position issue) and radiologic findings (including intracranial hemorrhage, ischemic-stroke and related features, space-occupying lesions, and others) to assess the impact of these potential confounders. The device consistently performed reliably across these circumstances. Furthermore, our analysis of the device's ability to identify mild to severe levels of MLS, revealed that EFAI MLSCT maintains stable performance. In conclusion, the results demonstrate that the EFAI MLSCT device is determined to be substantially equivalent in safety and effectiveness to the predicate device.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Ever Fortune AI. The logo consists of a stylized human figure with a green network-like design in place of the head. To the right of the figure is the text "EVER FORTUNE.AI" in a teal color. The logo appears to be for a company that specializes in artificial intelligence.
#### Safety & Effectiveness 8.
EFAI MLSCT has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with ISO 14971:2019 Medical devices - Application of risk management to medical devices. The EFAI MLSCT performance has been validated using retrospective data from case data and through the use of Reader comparison analysis.
#### 9. Substantial Equivalence
The EFAI MLSCT has minor differences compared to the predicate device. These differences are related to the Intended Use, user control for triage configuration and inclusion of preview images in notifications.
The indications for use statement for the subject device is substantially similar to the cleared indications for Qure Ai's qER device. The overall effect of both the subject and predicate devices-namely, the workflow triaging of CT images by flagging suspected findings of pathologies in head CT images-remains the same. Although the subject device is intended to flag midline shift only without user control for configuring the triage finding, whereas the predicate flags mass effect, midline shift, cranial fractures, and ICH with user control for configuring the triage finding or combination of findings, this difference does not raise new questions of safety or effectiveness.
Both the subject and predicate devices use an artificial intelligence algorithm to analyze images and highlight cases with detected pathologies, alongside the ongoing standard of care image interpretation. The subject device provides notifications for cases with suspected findings without preview images, whereas the predicate provides notifications that include preview images for informational purposes only, not intended for diagnostic use beyond notification. Since clinicians can and should view the original CT before making a final determination on a case, with or without the sent preview images, the differences in notification outputs do not introduce new risks. Thus, the proposed device does not raise new questions of safety or efficacy, as the risks and technology used by both devices are the same and are mitigated similarly.
The predicate devices can be deployed either on client premises (on-premise) or on cloud servers (on-cloud). For on-cloud mode, an on-premise gateway interfaces with HIPAA-compliant cloud servers. The subject device, however, is exclusively deployed on-premise and does not interface with cloud servers, thereby no new risks or safety issues arise.
#### 10. Conclusion
Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics, and performance testing, the EFAI MLSCT raises no new questions of safety and effectiveness and demonstrates substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.