Neonav ECG Tip Location System

K241910 · Navi Medical Technologies · LJS · Jan 19, 2025 · General Hospital

Device Facts

Record IDK241910
Device NameNeonav ECG Tip Location System
ApplicantNavi Medical Technologies
Product CodeLJS · General Hospital
Decision DateJan 19, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5970
Device ClassClass 2
AttributesPediatric

Indications for Use

Neonav is indicated for navigation and positioning of central venous access devices (CVADs) of at least 1Fr in size. Neonav provides real-time catheter tip location information by using the patient's cardiac electrical activity. Neonav is indicated for use as an alternative method to chest X-ray and fluoroscopy for all central venous access device tip placement confirmation. In adult patients and in adolescents (greater than 12 through 21 years of age), Neonav can be used with CVADs such as peripherally inserted central catheters (PICCs), centrally inserted central catheters (CICCs), implantable ports, and hemodialysis catheters; in children (greater than 2 to 12 years of age), Neonav can be used with CICCs; in infants (greater than 1 month to 2 years of age) and in neonates (from birth to 1 month of age), Neonav can be used with PICCs and CICCs. In each specific age group, the CVAD type and size must be chosen and the CVAD must be used according to the CVAD's indications and instructions for use.

Device Story

Non-invasive system for real-time tracking of CVAD tip location during placement. Inputs: intravascular ECG (I-ECG) signals via catheter tip (conductive fluid or guidewire/stylet method) and surface ECG reference/noise reduction electrodes. Device transforms electrical impulses into visual display on console; highlights QRS complex; displays real-time surface/intravascular ECG, heart rate, and insertion depth. Used in clinical settings by healthcare providers. Output allows clinician to manually determine optimal catheter tip position based on P-wave morphology. Benefits: provides real-time guidance, potentially reducing reliance on post-procedural X-ray or fluoroscopy.

Clinical Evidence

Bench testing only. No clinical data presented. Verification and validation performed per recognized consensus standards (ISO 10993, ISO 14971, IEC 60601 series, IEC 62304, IEC 62366-1). Testing confirmed performance of 1Fr catheter detection and system safety/usability.

Technological Characteristics

Materials: biocompatible polycarbonate and stainless steel (patient-contacting components). Sensing: I-ECG via catheter tip and surface electrodes. Connectivity: wired USB connection between acquisition unit and console; USB port for data export. Form factor: console with touchscreen, separate acquisition unit and remote. Sterilization: ethylene oxide. Software: standalone medical application.

Indications for Use

Indicated for navigation/positioning of CVADs (≥1Fr) in adults, adolescents (>12-21y), children (>2-12y), infants (>1mo-2y), and neonates (birth-1mo). Used as alternative to X-ray/fluoroscopy for tip placement confirmation. Contraindicated/limited in patients with atrial fibrillation, atrial flutter, severe tachycardia, or pacemaker-driven rhythms where P-wave presentation is altered.

Regulatory Classification

Identification

A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.

Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 19, 2025 Navi Medical Technologies Zorana Mayooran QA/RA Specialist 700 Swanston Street Melbourne, VIC 3053 Australia Re: K241910 Trade/Device Name: Neonav ECG Tip Location System Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long-term intravascular catheter Regulatory Class: Class II Product Code: LJS Dated: June 30, 2024 Received: July 1, 2024 Dear Zorana Mayooran: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Aneesh S. Deoras -S Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K241910 Device Name Neonav ECG Tip Location System #### Indications for Use (Describe) Neonav is indicated for navigation and positioning of central venous access devices (CVADs) of at least 1Fr in size. Neonav provides real-time catheter tip location information by using the patient's cardiac electrical activity. Neonav is indicated for use as an alternative method to chest X-ray and fluoroscopy for all central venous access device tip placement confirmation. In adult patients and in adolescents (greater than 12 through 21 years of age), Neonav can be used with CVADs such as peripherally inserted central catheters (PICCs), centrally inserted central catheters (CICCs), implantable ports, and hemodialysis catheters; in children (greater than 2 to 12 years of age), Neonav can be used with CICCs; in infants (greater than 1 month to 2 years of age) and in neonates (from birth to 1 month of age), Neonav can be used with PICCs and CICCs. In each specific age group, the CVAD type and size must be chosen and the CVAD must be used according to the CVAD's indications and instructions for use. Limiting but not contraindicated situations for use of the Neonav are in patients where alterations of cardiac rhythm change the presentation of the P wave as in atrial flutter, severe tachycardia, and pacemaker driven rhythm. In such patients, who are easily identifiable prior to central venous access device insertion, the use of an additional method is required to confirm catheter tip location. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="display:inline-block; margin-right: 5px;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-right: 5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------| |--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ | 510(k) #: | K241910 | | 510(k) Summary | Prepared on: 2024-12-20 | |-----------|---------|--|----------------|-------------------------| |-----------|---------|--|----------------|-------------------------| Contact Details | Applicant Name | Navi Medical Technologies | |-----------------------------|--------------------------------------------------| | Applicant Address | 700 Swanston Street Melbourne VIC 3053 Australia | | Applicant Contact Telephone | +61481882530 | | Applicant Contact | Dr. Zorana Mayooran | | Applicant Contact Email | zorana@navitechnologies.com | Device Name | Device Trade Name | Neonav ECG Tip Location System | |---------------------|----------------------------------------------------------------------| | Common Name | Percutaneous, implanted, long-term intravascular catheter | | Classification Name | Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days | | Regulation Number | 880.5970 | | Product Code(s) | LJS | Legally Marketed Predicate Devices | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |-------------|----------------------------------------------------------|--------------| | K141634 | Nautilus Delta | LJS | Device Description SummaryThe Neonav ECG Tip Location System (Neonav) is a non-invasive medical device that analyzes electrocardiographic signals (ECG) to track the location of central venous access devices (CVADs) in patients' blood vessels during placement in real time via a display. Catheter position in the vasculature is analyzed based on the patient's intravascular ECG (I-ECG). An I-ECG is recorded via the catheter tip from within the blood vessels of a patient. For the conductive fluid method, the sterile Neonav ECG Adapter Cable are used whereas for guidewire/stylet method, the sterile Neonav ECG clip cable is used. These components allow the electrical impulses detected at the catheter tip to be recorded by a non-sterile Acquisition Unit and analyzed by the Neonav Console to assist in determining the catheter tip position. The Neonav Controlled via a touch screen interface or a Remote enclosed in a sterile cover. The Neonav Console displays real-time surface ECG and I-ECG signals, highlighted QRS complex, max measured P-wave, the patient's heart rate, and other device performance indicators. A surface electrode is placed on the subject to act as a reference electrode. A second surface electrode is placed on the subject for noise reduction. Neonav components are designed to attach easily to standard catheter and IV therapy components via standard luer lock connections. The only patient contacting component of the ECG Adapter. It has indirect contact with the patient via fluids and is composed of biocompatible polycarbonate and stainless steel. ### Intended Use/Indications for Use #### Neonav is indicated for navigation and positioning of central venous access devices (CVADs) of at least 1Fr in size. Neonav provides real-time catheter tip location information by using the patient's cardiac electrical activity. Neonav is indicated for use as an alternative method to chest X-ray and fluoroscopy for all central venous access device tip placement confirmation. In adult patients and in adolescents (greater than 12 through 21 years of age), Neonav can be used ### 21 CFR 807.92(a)(5) {5}------------------------------------------------ with CVADs such as peripherally inserted central catheters (PICCs), centrally inserted central catheters (CICCs), implantable ports, and hemodialysis catheters; in children (greater than 2 to 12 years of age), Neonav can be used with PICCs and with CICCs; in infants (greater than 1 month to 2 years of age) and in neonates (from birth to 1 month of age), Neonav can be used with PICCs and CICCs. In each specific age group, the CVAD type and size must be chosen and the CVAD must be used according to the CVAD's indications and instructions for use. Limiting but not contraindicated situations for use of the Neonav are in patients where alterations of cardiac rhythm change the presentation of the P wave as in atrial fibrillation, atrial flutter, severe tachycardia, and pacemaker driven rhythm. In such patients, who are easily identifiable prior to central venous access device insertion, the use of an additional method is required to confirm catheter tip location. # Indications for Use Comparison ### 21 CFR 807.92(a)(5) 21 CFR 807.92(a)(6) Neonav has equivalent indications for use to the predicate with the following adjustments: a) Neonav is indicated to be used with PICCs in infants and neonates (in addition to CICCs for this patient population). b) Neonav allows for use of catheter sizes of 1Fr and above, instead of 3Fr and above. These changes are supported by the most current literature and effectiveness of PICCs in infants and neonates and non-clinical bench tests demonstrating that Neonav is able to capture data from lines down to 1Fr. They are still within the scope of the current intended use which states "The intenavis to support navigation and tip positioning of central venous access devices. Neonav can be used as an alternative method to fluoroscopy and chest X-ray for central venous access device tip placement confirmation". The intended use of the Neonav is equivalent to the intended use of the predicate device. c) To align with recent standards and guidance documents the terminology CVAD has been adopted in place of the generic term of CVC to refer to the family of vascular access devices terminating in a central location (INS Infusion Therapy Standard of Practice 9th Ed. ISSN 1533-1458, INS Policies and Procedure for Infusion Therapy: Neonate to Adolescent 3rd Ed). CICC has been adopted to refer to specific devices that are inserted in the jugular or subclavian vein and terminate in the SVC (Vascular Access in Neonates and Children ISBN 978-3-030-94708-8). ## Technological Comparison The Neonav has features, materials and technological characteristics equivalent to the predicate device. Notable differences that do not raise new questions about safety and effectiveness are: - a) Neonav does not have a printer. It allows patient summary reports to a USB rather than sent directly to a printer. b) Neonav's remote and acquisition unit are separate items, rather than a single item c) Neonav Acquisition Unit is wired (USB cable) rather than wireless (Bluetooth) d) Neonav medical application is standalone, instead of mobile. e) Neonav does not provide computation of ECG signal for navigation. Neonav displays the intravascular ECG to the user makes their own decision regarding catheter navigation of the real-time intravascular ECG signal. f) Neonav highlights the complete QRS complex instead of placing red marks on the R-wave. g) Neonav can display multiple frozen I-ECG waveforms simultaneously with insertion depth readings to aid in finding the depth with the max measured P-wave. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Testing was conducted to verify and validate performants and determine substantial equivalence. The following guidance documents or FDA recognized consensus standards were used/referenced for testing: · ISO 10993-1: Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process • 1099-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterlization residuals [Including: Technical Corrigendum 1 (2009), AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)] · ISO 11135-7 Sterilization of health-care products — Requirements for the development, validation and routine control of a sterilization process for medical devices • ISO 11135-2 Sterilization of health care products — Ethylene oxide — Part 2: Guidance on the application of ISO 11135-1 · ISO 11607-1 Second edition 2019-02 Packaging for terminally sterilized medical devices — Part 1: Requirements for materials, sterile barrier systems and packaging systems (including all applicable ASTM standards to verify the characteristics of the packaging material) • ISO 11607-2: Second edition 2019-02 Packaging for terminally sterilized medical devices — part 2: validation requirements for forming, sealing and assembly processes • 15223-1 Fourth edition 2021-07 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied. • ISO 14971: 2019 Medical devices - Applications of risk management to medical devices. · EN ISO 13485 Medical devices. QMS. Requirements for Regulatory Purposes. • ANSI AAMI IEC 60601-1: ES60601-1:2005/(R)2012 & A1:2010/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance {6}------------------------------------------------ · EC 60601-1-2: Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. · IEC/TR 60601-4-2, IEC / TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems « EC 60601-1-6: 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability • IEC 62366-1: 62366-1 Edition 1.1 2020-06 CONSOLIDATEDVERSION Medical devices - Application of usability engineering to medical devices • IEC 62304: 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software life cycle processes • IEC 62133-2: 62133-2 Edition 1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems · IEC 60601-2-27: 60601-2-27 Edition 3.0 2011-03 Medical electrical equipment - Part 2-27: Particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment • ISO 20417: First edition 2021-04 Corrected version 2021-12 Information supplied by the manufacturer of medical devices • IEC 60529 Degree of Protection Provided By Enclosures (IP Code) • ISO 80369-7:2016 80369-7 Second edition 2021-05 Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications • ANSI/AAMI EC53:2013 (R2020) ECG Trunk Cables And Patient Leadwires • ASTM D543 Standard Practices for Evaluating the Resistance of Plastics to Chemical Reagents · ASTM F88/F88M-21 Standard Test Method for Seal Strength of Flexible Barrier Materials · ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems • ASTM D7386-16 Standard Practice for Performance Testing of Packages for Single Parcel Delivery Systems • ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) • ISTA 3A 2018 General Simulation Performance Test - Parcel Delivery System Shipments 70 kg (150lb) or Less · ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices • AAMI TIR57 AAMI TIR57:2016, Principles for medical device security - Risk management • Off-The-Shelf Software Use in Medical Devices –Guidance for Industry and Food and Drug Administration Staff (August 11, 2023) • Content of Premarket Submissions for Device Software Functions – Guidance for Industry and FDA Staff, June 14, 2023 • Cybersecurity in Medical Devices: Quality System Content of Premarket Submissions Guidance for Industry and Food and Drug Administration Staff, September 27, 2023 · Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff, September 2023 · Applying Human Factors and Usability Engineering to Medical Devices, Guidance for Industry and Food and Drug Administration Staff, February 2016 The subject device, the Neonav ECG Tip Location System, met all predetermined acceptance criteria as required from the listed standards and has demonstrated substantial equivalence to the predicate device. Based on intended use, technological characteristics, and performance testing it is concluded that the subject device, the Neonav ECG Tip Location System, is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...