K241888 · Leo Medical Co., Ltd. · FGE · Dec 20, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K241888
Device Name
Single-use Balloon Dilatation Catheter
Applicant
Leo Medical Co., Ltd.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Dec 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract. It is also indicated in adults for endoscopic dilatation of Sphincter of Oddi with or without prior sphincterotomy.
Device Story
Single-use balloon dilatation catheter; available in Rx, OTW, and OTW with stainless wire-guided configurations. Device consists of tip, balloon, marker band, shaft, sleeve, stress diffusion tube, and hub. Used endoscopically by physicians to dilate gastrointestinal strictures or the Sphincter of Oddi. Operates by inflating a balloon to apply radial force to strictures. Input includes manual manipulation by the clinician; output is mechanical dilation of the target tissue. Benefits include relief of strictures and improved luminal patency. Device is provided sterile (EO) for single use.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including dimension, appearance, compatibility, delivery/retrieval force, peak tensile force, burst pressure, kink stability, corrosion resistance, leakage, radiopacity, balloon compliance, deflation time, and fatigue testing. Biocompatibility was verified per ISO 10993-1.
Technological Characteristics
Materials: Pebax 3533 SA01 Med (tip), Pebax 7233 SA01 Med (shaft), Nylon 12 ML21 (balloon), SUS 304 (guide wire). Energy: Mechanical (balloon inflation). Dimensions: 180cm/240cm length; 6-20mm diameter; 30/55/80mm balloon length. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Connectivity: None. Standards: ISO 10555-4, ISO 10555-1, ISO 80369-7/20, ASTM F640, ASTM F88, ASTM F1929, ASTM D4169.
Indications for Use
Indicated for adult and adolescent populations for endoscopic dilation of gastrointestinal tract strictures; also indicated for adults for endoscopic dilation of the Sphincter of Oddi, with or without prior sphincterotomy.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2024
Leo Medical Co., Ltd. % Diana Hong Official Correspondent Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai, Shanghai 200120 China
Re: K241888
Trade/Device Name: Single-use Balloon Dilatation Catheter Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary Catheter And Accessories Regulatory Class: Class II Product Code: FGE Dated: November 25, 2024 Received: November 25, 2024
Dear Diana Hong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S
Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241888
Device Name
Single-use Balloon Dilatation Catheter
Indications for Use (Describe)
This device is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract. It is also indicated in adults for endoscopic dilatation of Sphincter of Oddi with or without prior sphincterotomy.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K241888
- 1. Date of Preparation: 06/21/2024
- Sponsor Identification 2.
## Leo Medical Co., Ltd.
No.10 Building, 23 Middle Huashan Road, Xinbei District, 213022, Changzhou City, Jiangsu Province, PEOPLE'S REPUBLIC OF CHINA
Contact Person: Guojun Zhang Position: Senior RA Manager Tel: +86-15961165897 Email: gjzhang@leomed.cn
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)
## Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
- 4. Identification of Proposed Device
Trade Name: Single-use Balloon Dilatation Catheter Common Name: Biliary catheter and accessories
Regulatory Information Classification Name: stents, drains and dilators for the biliary ducts Classification: II,
{6}------------------------------------------------
Product Code: FGE, Regulation Number: 21 CFR 876.5010. Review Panel: Gastroenterology/Urology
Indications for Use
This device is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract. It is also indicated in adults for endoscopic dilatation of Oddi with or without prior sphincterotomy.
#### Device Description
According to the structure of the device, it can be divided into three types, RX type, OTW type and OTW with Stainless Wire-guided type. The Rx proposed device mainly consists of tip, balloon, marker band, shaft, sleeve, stress diffusion tube and hub. The OTW proposed device mainly consists of tip, balloon, marker band, shaft, stress diffusion tube and hub. The structure of the OTW with Stainless Wire-guided model same with the structure of OTW model. The difference between OTW with Stainless Wire-guided type catheter and OTW type catheter is that OTW with Stainless Wire-guided type catheter is preloaded with a guide wire and a guide wire locking component. The difference between the RX type catheter and the OTW series catheter is that the guidewire exit position is different. The guide wire of OTW series catheter passed through the guide wire lumen of the catheter. The guidewire of the Rx type catheter will pass through the guidewire exchange port on the shaft.
Single-use Balloon Dilatation Catheter is divided into different specifications. Shafts are available in two different diameters. The effective length of the catheter is 180cm and 240cm. Balloon diameters are available in six different diameters 6-7-8mm, 8-9-10 mm, 12-13.5-15 mm, 15-16.5-18 mm, 18-19-20 mm, and the balloon lengths are available in 30 mm,55 mm and 80 mm three different length.
- Identification of Predicate Devices and Reference Device న.
Predicate Device 1 510K Number: K112994 Trade Name: CRE Dilatation Balloon
Predicate Device 2 510K Number: K172520 Trade Name: CRE™ RX Biliary Balloon Dilatation Catheter
Reference device 510K Number: K122924 Trade Name: CRE Fixed Wire Balloon Dilatation Catheter
{7}------------------------------------------------
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A ISO 10993-7:2008 Biological Evaluation Of Medical Devices-Part 7: Ethylene Oxide Sterilization Residuals
- A ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for In Vitro Cytotoxicity
- A ISO 10993-10:2021 Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization
- A ISO 10993-11:2017 Biological Evaluation of Medical Devices-Part 11: Tests for systemic toxicity
- > ISO 10993-23:2021 Biological evaluation of medical devices — Part 23: Tests for irritation.
- > ASTM F88/F88M-21 Standard test method for seal strength of flexible barrier materials
- A ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dve Penetration
- A ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
- > USP<151> Pyrogen Test
- > ISO10555-4:2013 Intravascular catheters -- Sterile and single-use catheters -- Part 4: Balloon dilatation catheters.
- A ISO 10555-1:2013/A1:2017 Intravascular catheters -Sterile and single-use catheters -Part 1: General requirements.
- > ISO20696:2018(E) Sterile urethral catheters for single-use.
- ASTM F640-20 Standard Test Methods for Determining Radiopacity for Medical Use A
- A ASTM F1886/F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- A USP<85> Bacterial Endotoxins Test
- > ISO 80369-7:2023 Small-bore connectors for liguids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications
- > ISO 80369-20:2023 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
- A ASTM F3172 Standard Guide for Design Verification Device Size and Sample Size Selection for Endovascular Devices
The performance testing conducted on subject device and predicate device are listed below. All the test results demonstrate proposed device meet the requirements of its pre-defined acceptance criteria and intended uses.
- . Dimension Test
- o Appearance
- o Compatibility Test
- o Delivery and Retrieval Force
{8}------------------------------------------------
- Peak Tensile Force
- . Burst Pressure
- . Kink Stability
- Corrosion Resistance Test ●
- Air Leakage
- Liquid Leakage
- . Luer connector
- Radiopacity
- o Balloon Burst Pressure
- . Balloon Compliance
- . Balloon Deflation Time
- . Balloon Fatigue
The proposed device sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 10 6). EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 20 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 3-year shelf-life.
- Package Integrity Test after environmental conditioning, simulated transportation testing in ● accordance to ASTM D4169-22 on final, packaged, and sterile device.
- . Sterile Barrier Packaging performed on the proposed device:
- o Visual Inspection per ASTM F1886/F1886M-16
- 0 Seal Strength per ASTM F88/F88-21
- o Dye Penetration per ASTM F1929-15
- Shelf-life of 3 years is validated using FDA recognized standard ASTM F1980-16 Standard Guide for ● Accelerated Aging of Sterile Barrier Systems for Medical Devices.
The biocompatibility evaluations were conducted in accordance with the 2020 FDA Guidance document Use of International Standard ISO-10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The cytotoxicity, sensitization, intracutaneous irritation, system toxicity and pyrogen tests were performed to demonstrate the biocompatibility of the device.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{9}------------------------------------------------
#### 8. Substantially Equivalent (SE) Comparison
| Table 1. General Comparison | | | | | |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Item | Proposed Device | Predicate Device 1<br>K112994 | Predicate Device 2<br>K172520 | Reference Device<br>K122924 | Remark |
| Device Name | Single-use Balloon<br>Dilatation Catheter | GRE Dilatation Balloon | CRE™ RX Biliary Balloon<br>Dilatation Catheter | CRE Fixed Wire Balloon<br>Dilatation Catheter | / |
| Product Code | FGE | FGE | FGE | KNQ | Same |
| Regulation No. | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 876.5365 | Same |
| Class | II | II | II | II | Same |
| Indications for<br>Use | This device is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract. It is also indicated in adults for endoscopic dilatation of Sphincter of Oddi with or without prior sphincterotomy. | The CRE™ Wireguided<br>Balloon Dilatation Catheter<br>is indicated for use in adult<br>and adolescent populations<br>to endoscopically dilate<br>strictures of the alimentary<br>tract.<br>It is also indicated in adults<br>for endoscopic dilatation of<br>Sphincter of Oddi with or<br>without prior<br>sphincterotomy. | CRE™ RX Biliary Balloon<br>Dilatation Catheter is<br>indicated for use in adults<br>for endoscopic dilatation of<br>the Sphincter of Oddi with<br>or without prior<br>sphincterotomy.<br>The CRE™ RX Biliary Balloon<br>Dilatation Catheters may<br>be used for injection of<br>contrast medium for<br>fluoroscopic visualization<br>of the bile ducts. | The CRE™ Balloon<br>Dilatation Catheter is<br>intended for use in adult and<br>adolescent populations to<br>endoscopically dilate<br>strictures of the esophagus. | Similar |
| Catheter Type | Rx, OTW,<br>OTW<br>(Stainless Wire-guided) | OTW(Stainless<br>Wire-guided) | RX | OTW (Fixed wire) | Different |
| Components | Rx: Tip, Balloon, Marker<br>band, Shaft, Sleeve, Stress<br>diffusion tube, Hub | Tip, balloon, marker band,<br>handle junction, and<br>guidewire | Tip, balloon, marker band<br>and handle junction | Tip, balloon, marker band,<br>tube and handle junction. | Different |
| Table 1. | General Comparison |
|----------|--------------------|
|----------|--------------------|
{10}------------------------------------------------
| | OTW(Stainless<br>Wire-guided):<br>Tip,<br>balloon, marker band,<br>Shaft, Sleeve, Stress<br>diffusion tube , Hub,<br>Guide wire locking<br>device, Stainless Guide<br>wire<br><br>OTW: Tip, balloon,<br>marker band, Shaft, Stress<br>diffusion tube, Hub, | | | | |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-------------|-------------|-----------|
| Marker<br>Bands<br>Present | Yes | Yes | Yes | Yes | Same |
| Single Use | Single Use | Single Use | Single Use | Single Use | Same |
| | 6-7-8, | 6-7-8, | 6-7-8, | 6-7-8, | |
| | 8-9-10, | 8-9-10, | 8-9-10, | 8-9-10, | |
| Balloon<br>Diameter(mm) | 10-11-12, | 10-11-12, | 10-11-12, | 10-11-12, | Same |
| | 12-13.5-15, | 12-13.5-15, | 12-13.5-15, | 12-13.5-15, | |
| | 15-16.5-18, | 15-16.5-18, | 15-16.5-18, | 15-16.5-18, | |
| | 18-19-20 | 18-19-20 | 18-19-20 | 18-19-20 | |
| Balloon<br>length(mm) | 30, 55, 80 | 55 | 30,55 | 80 | Different |
| Working<br>length(mm) | 1800,2400 | 1800,2400 | 1800 | 1800 | Same |
| Patient-contact Material | | | | | |
| Tip | Pebax 3533 SA01 Med | Guide head (Pebax) | Unknown | Unknown | Different |
{11}------------------------------------------------
| Shaft | Pebax 7233 SA01 Med | Catheter (Pebax) | | | |
|----------------------------------------------------------------|---------------------------------|-------------------------------------|-------------------------------------|-------------------------------------|------|
| Balloon | Nylon 12 ML21 | Balloon (Pebax) | | | |
| Stainless Guide<br>wire (OTW with<br>Stainless<br>Wire-guided) | SUS 304 | Guidewire (stainless steel) | | | |
| Sterilization | | | | | |
| Sterile/non-sterile: | Sterile | Sterile | Sterile | Sterile | Same |
| Method | EO sterilized | EO sterilized | EO sterilized | EO sterilized | Same |
| SAL | 10-6 | 10-6 | 10-6 | 10-6 | Same |
| Biocompatibility | | | | | |
| Cytotoxicity | No cytotoxicity | | | | |
| Intracutaneous<br>Reactivity | No intracutaneous<br>reactivity | | | | |
| Sensitization | No sensitization | Conform with ISO 10993<br>standards | Conform with ISO 10993<br>standards | Conform with ISO 10993<br>standards | Same |
| Acute System<br>Toxicity | No acute system Toxicity | | | | |
| Pyrogen | No pyrogen | | | | |
{12}------------------------------------------------
# Similar - Indications for Use
The indication for use of the proposed device and the indication for use of the predicate device 1 are exactly the same. The indication for use of the proposed device 2 are slightly different, but the intend use "indicated in adults for endoscopic dilatation of Sphincter of Oddi with or without prior sphincterotomy." can be covered by predicate device 2. The clinical application site and applicable population of the proposed device can be completely covered by two predicate devices. Therefore, the similar indication for use will not affect the safety and effectiveness of the proposed device.
#### Different - Catheter Type
The catheter type of the proposed device is different to that of the predicate device 1 and predicate device 2. The proposed device includes Rx, OTW and OTW (Stainless Wire-guided). The Rx type catheter and OTW (Stainless Wire-guided) type catheter can be covered by the predicate device 1 and predicate device 2. The OTW type catheter is very similar to OTW (Stainless Wire-guided) catheter, and the difference with OTW (Stainless Wire-guided) catheter is only the lack of preinstalled guide wire and the structure of OTW (Stainless Wire-guided) catheter same with OTW catheter. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
### Different – Components
The component of the proposed device is different to that of the predicate device 1 and predicate device 2. Different manufacturers have different claims on the name of the device components, the difference in the component is only the difference in the name. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
#### Different - Balloon length
The balloon length of the predicate device 1 is 55mm, the balloon length of the predicate device 2 is 30mm and 55mm, the 30mm and 50mm balloon length of the proposed device is same as the balloon length of the predicate device 2. The 80 mm balloon length of proposed device can be covered by the reference device. Besides, the 80 mm balloon length proposed device was tested for performance and the test results meet the acceptance criteria. Therefore, this difference will not affectiveness of the proposed device.
#### Different - Material
The materials of the proposed device are different from those of the predicate device. The biocompatibility tests were also performed on the proposed device and the test results showed that there was no adverse effect. Therefore, this difference does not raise new safety and efficacy issues.
#### 9. Substantially Equivalent (SE) Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and perform as well as or better than the legally marketed predicate device 1 K112994 and
{13}------------------------------------------------
predicate device 2 K172520.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.