K241873 · Zimmer, Inc. · PKC · Dec 11, 2024 · Orthopedic
Device Facts
Record ID
K241873
Device Name
OsseoFit Stemless Shoulder System
Applicant
Zimmer, Inc.
Product Code
PKC · Orthopedic
Decision Date
Dec 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Anatomic Total Shoulder Applications: · Osteoarthritis • The patient must be anatomically and structurally suited, as evident by scapula and proximal humerus closure, to receive the implants. OsseoFit™ Stemless Shoulder humeral components have a porous surface and are indicated for uncemented biological fixation applications. Compatible Glenoid components are intended to be implanted with bone cement. The porous posts may be inserted without bone cement. Compatible Convertible Glenoid Baseplate components are intended for cementless applications with the addition of screw fixation.
Device Story
OsseoFit Stemless Shoulder System; anatomic total shoulder arthroplasty implant. Humeral component: asymmetric anchor with six fins; additive manufactured; porous surface for uncemented biological fixation. Available in Onlay (sits on bone) and Inlay (sits within bone) variants; includes anterior suture holes for soft tissue attachment. Assembled to compatible Identity Shoulder System humeral head via locking taper using disposable impactor. Used with compatible Alliance Glenoid Implants or Comprehensive Convertible Baseplate liners. Intended for use by orthopedic surgeons in clinical settings. Provides stable humeral fixation; restores shoulder joint function; benefits patients with osteoarthritis.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including cortical perforation analysis, construct and corrosion fatigue testing, micromotion, pull-out, lever-out, and torque-out fixation characterization, static tensile/torsional modular junction testing, range of motion analysis, MRI compatibility, and additive manufacturing process validation.
Technological Characteristics
Additive manufactured humeral implant; OsseoTi porous technology for biological fixation. Asymmetric design with six fins; Onlay and Inlay variants. Modular assembly via locking taper. Materials and sterilization methods are similar to predicate devices. Standalone orthopedic implant; no software or energy source.
Indications for Use
Indicated for patients with osteoarthritis requiring anatomic total shoulder arthroplasty; patients must have anatomically and structurally suitable scapula and proximal humerus closure.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 11, 2024
Zimmer, Inc. Alexandria Irwin Regulatory Affairs Sr Specialist 1800 W. Center Street Warsaw, Indiana 46850
Re: K241873
Trade/Device Name: OsseoFit Stemless Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PKC Dated: November 5, 2024 Received: November 5, 2024
Dear Alexandria Irwin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Digitally signed by Farzana Farzana Sharmin -S Sharmin -S Date: 2024.12.11 16:43:06 -05'00'
Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241873
Device Name
OsseoFit Stemless Shoulder System
Indications for Use (Describe)
Anatomic Total Shoulder Applications:
· Osteoarthritis
• The patient must be anatomically and structurally suited, as evident by scapula and proximal humerus closure, to receive the implants.
OsseoFit™ Stemless Shoulder humeral components have a porous surface and are indicated for uncemented biological fixation applications.
Compatible Glenoid components are intended to be implanted with bone cement. The porous posts may be inserted without bone cement.
Compatible Convertible Glenoid Baseplate components are intended for cementless applications with the addition of screw fixation.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
| K241873 | | Page 1 of 3 | | | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------|----------------|--|-------------------------|
| 510(k) #: | K241873 | | 510(k) Summary | | Prepared on: 2024-12-11 |
| | | | | | 21 CFR 807.92(a)(1) |
| Contact Details | | | | | |
| Applicant Name | Zimmer, Inc. | | | | |
| Applicant Address | 1800 W. Center Street Warsaw IN 46850 United States | | | | |
| Applicant Contact Telephone | 1 574-373-0167 | | | | |
| Applicant Contact | Mrs. Alexandria Irwin | | | | |
| Applicant Contact Email | alexandria.irwin@zimmerbiomet.com | | | | |
| | | | | | 21 CFR 807.92(a)(2) |
| Device Name | | | | | |
| Device Trade Name | OsseoFit Stemless Shoulder System | | | | |
| Common Name | Prosthesis, Total Anatomic Shoulder, Uncemented Metaphyseal Humeral Stem with No Diaphyseal Incursion, Semi-Constrained | | | | |
| Classification Name | Shoulder joint metal/polymer semi-constrained cemented prosthesis | | | | |
| Regulation Number | 888.3660 | | | | |
| Product Code(s) | PKC | | | | |
| Legally Marketed Predicate Devices | | | | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | Product Code | | |
| K171858 | Sidus Stem-Free Shoulder | | PKC | | |
| K182516 | Comprehensive Nano Stemless Shoulder | | PKC | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| Device Description Summary | | | | | 21 CFR 807.92(a)(4) |
The OsseoFit™ Stemless Shoulder System consists of a stemless humeral head adapter for anatomic total shoulder applications. The OsseoFit humeral implant is an anatomical (left/right specific), asymmetric humeral anchor with six fins whose location, length, and height vary. The OsseoFit humeral implants are available in Onlay and Inlay variants. The Onlay implants sit on top of the prepared bone while the Inlay implants have a dished superior portion with a shorter fin length to sit within the bone. The OsseoFit Humeral Implant also includes anterior suture holes for soft tissue attachment.
{5}------------------------------------------------
21 CFR 807.92(a)(5)
The humeral implants are manufactured through additive manufacturing which includes areas of OsseoTi porous technology to allow biological fixation.
The OsseoFit Stemless Shoulder System is intended for cementless applications and is designed to be used with a compatible ldentity Shoulder System (K213856) humeral head. The Identity Humeral Head is assembled to the OsseoFit Humeral Head Adapter via a locking taper by a disposable impactor that is packaged together with the head adapter. The ldentity heads can be used with a compatible Alliance Glenoid Implant (K191814 and K193180) or Comprehensive Convertible Baseplate liner (K211729).
## Intended Use/Indications for Use
Anatomic Total Shoulder Applications:
· Osteoarthritis
• The patient must be anatomically suited, as evident by scapula and proximal humerus closure, to receive the implants.
OsseoFit™ Stemless Shoulder humeral components have a porous surface and are indicated for uncemented biological fixation applications.
Compatible Glenoid components are intended to be implanted with bone cement. The porous posts may be inserted without bone cement.
Compatible Convertible Glenoid Baseplate components are intended for cementless applications with the addition of screw fixation.
# Indications for Use Comparison
The subject device has the same intended use as the predicate device(s) and has the similar indications for use as the predicate(s). The OsseoFit Stemless Shoulder System narrowed the indications to Osteoarthritis.
## Technological Comparison
The subject device has similar technological characteristics as the designated predicates. The rationale for substantial equivalence is based on consideration of the following characteristics:
• Intended Use: Identical to predicates
• Indications for Use: Similar to predicates
• Materials: Similar to predicates
· Design Features: Similar to predicates
• Packaging: Similar to predicates
· Sterilization: Identical to predicates
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-Clinical Tests
- Analysis of Potential Cortical Perforation
- · Construct Fatigue Testing
- · Corrosion Fatigue Testing
- Initial Fixation Evaluation in physiological loading condition: Micromotion testing
- · Initial Fixation Characterization: Pull out testing
- Initial Fixation Characterization: Lever out testing
- Initial Fixation Characterization: Torque out testing
- Static Tensile and Torsional Modular Junction Testing
- Range of Motion Analysis
- · Magnetic Resonance Imaging (MRI) Environment Compatibility Evaluation
- · Porous coating/metal characterization
- · Additive Manufacturing Process Validation Report
- Mechanical and chemical characterization of Additive Manufacturing
- Cleaning Validation
Clinical Tests None Provided
21 CFR 807.92(a)(5)
{6}------------------------------------------------
Conclusion
Non-clinical testing demonstrated that the new device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.