K241858 · Ovesco Endoscopy AG · PKL · Aug 26, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K241858
Device Name
BARS Set (100.60)
Applicant
Ovesco Endoscopy AG
Product Code
PKL · Gastroenterology, Urology
Decision Date
Aug 26, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BARS Set is indicated for clip placement within the context of flexible endoscopy in the GI tract for the purpose of: - Hemostasis for: - Mucosal/submucosal defects < 3 cm - Arteries < 2 mm - Treatment of luminal perforations (< 20 mm) and / or anastomoses (< 50 mm) - Lumen reduction NOTE: The BARS Set is not indicated for stent fixation, For stent fixation, use the stentfix OTSC System Set.
Device Story
BARS Set is an over-the-scope clip (OTSC) application system for flexible endoscopy. Device consists of a memory-shape alloy clip mounted on a transparent plastic cap, which attaches to the distal end of an endoscope. Includes endoscopic instruments (anchors, balloons, guide wire) to mobilize target tissue into the cap lumen. Physician operates a thread and hand-wheel mechanism to deploy the clip onto the tissue. Used in clinical GI endoscopy settings. Output is the mechanical closure of tissue defects or lumen reduction. Benefits include effective hemostasis and closure of perforations/anastomoses. Device design ensures compatibility with endoscope maneuverability and vision.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by performance bench testing, including functional capability, robustness, endoscope compatibility, clip application ring strength, balloon performance, usability, and MRI suitability, compared against established reference devices.
Technological Characteristics
Over-the-scope clip system; memory-shape alloy clip; transparent plastic cap; thread and hand-wheel deployment mechanism; includes application ring and additional working channels. Designed for compatibility with flexible endoscopes. Biocompatible materials used. Non-powered, mechanical device.
Indications for Use
Indicated for patients requiring endoscopic clip placement in the GI tract for hemostasis of mucosal/submucosal defects (<3 cm) or arteries (<2 mm), treatment of luminal perforations (<20 mm) or anastomoses (<50 mm), and lumen reduction. Not indicated for stent fixation.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it.
August 26, 2024
Ovesco Endoscopy AG % Amko Groeneveld Senior Consultant novineon CRO GmbH Friedrich-Miescher-Strasse 9 Tuebingen, 72076 Germany
Re: K241858
Trade/Device Name: BARS Set (100.60) Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: Class II Product Code: PKL Dated: June 27, 2024 Received: June 27, 2024
Dear Amko Groeneveld:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Anthony Lee -S
Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn,
{3}------------------------------------------------
General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K241858
Device Name
BARS Set (100.60)
Indications for Use (Describe)
The BARS Set is indicated for clip placement within the context of flexible endoscopy in the GI tract for the purpose of:
- Hemostasis for:
- Mucosal/submucosal defects < 3 cm
- Arteries < 2 mm
- Treatment of luminal perforations (< 20 mm) and / or anastomoses (< 50 mm)
- Lumen reduction
NOTE: The BARS Set is not indicated for stent fixation, For stent fixation, use the stentfix OTSC System Set.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
ver-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
| K241858 -- Page 1/2 | | | | |
|---------------------|----------------|--|--|-------------------------|
| 510(k) #: | 510(k) Summary | | | Prepared on: 2024-08-21 |
| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------------------|------------------------------------------------------|
| Applicant Name | Ovesco Endoscopy AG |
| Applicant Address | Friedrich-Miescher-Strasse 9 Tuebingen 72076 Germany |
| Applicant Contact Telephone | +49707196528154 |
| Applicant Contact | Mr. Antonio Caputo |
| Applicant Contact Email | antonio.caputo@ovesco.com |
| Device Name | 21 CFR 807.92(a)(2) |
|---------------------|----------------------------------------|
| Device Trade Name | BARS Set (100.60) |
| Common Name | Hemostatic Metal Clip for the GI Tract |
| Classification Name | Hemorrhoidal ligator |
| Regulation Number | 876.4400 |
| Product Code(s) | PKL |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K101428 | OTSC System Set | PKL |
| | | |
| | | |
| | | |
### Device Description Summary
21 CFR 807.92(a)(4)
Device Description - Explanation of how the device functions:
The subject device constitutes an over-the-scope clip application device used within the context of flexible endoscopy in the gastrointestinal tract. The over-the-scope clip application capabilities center around a memory-shape alloy clip mounted on a transparent plastic cap. Cap (and clip) are mounted on the distal end of an endoscope. Endoscopic instruments are provided in the BARS Set to facilitate insertion and for mobilization of target tissue into the cap lumen. With the target tissue inside the cap lumen, the memory-shape alloy clip is applied onto the target tissue using a thread mechanism. Contents of the BARS Set are: - BARS (readily assembled unit of memory-shape alloy clip, transparent plastic cap, thread deployment components, working channels, and guide elements)
- Endoscopic instruments (BARS Anchor silver, BARS Anchor black, Insertion Balloon, Space Keeper Balloon, Guide Wire)
Scientific Concepts forming the basis for the device:
Established memory-shape alloy clips used for over-the-scope clip in flexible endoscopy in the gastrointestinal tract.
Significant Physical and Performance Characteristics: - Device Design:
{6}------------------------------------------------
Device design is guided by compatibility with flexible endoscopes and device influence on endoscopic functions, e.g., vision and maneuverability.
- Material Used:
Materials used need to be sufficiently biocompatible considering the duration and nature of contact to the human body.
- Physical Properties:
Dimensions need to allow insertion via flexible endoscope into the gastrointestinal tract.
-Reference Devices: K200684: gastroduodenal FTRD Set K170867: FTRD System Set K183309: stentfix OTSC System Set
## Intended Use/Indications for Use
The BARS Set is indicated for clip placement within the context of flexible endoscopy in the Gl tract for the purpose of: - Hemostasis for:
- Mucosal/submucosal defects < 3 cm
- Arteries < 2 mm
- Treatment of luminal perforations (< 20 mm) and / or anastomoses (< 50 mm)
- Lumen reduction
NOTE: The BARS Set is not indicated for stent fixation, use the stentfix OTSC System Set.
## Indications for Use Comparison
Indications for use differ slightly in specification. These are found to not raise different questions of safety and effectiveness (decision making flowchart point 4). Methods for investigating differences effectiveness are applied. Data from these methods support substantial equivalence.
#### Technological Comparison
Technological comparison shows that the subject device, as well as leveraged reference devices share common technological aspects. I.e., all are caps that are attached to an endoscope tip for the purpose of clip application in the gastrointestinal tract. In each case, a thread and hand wheel mechanism is application. For the subject device, an application ring is present which is not present in the predicate device. However, reference devices illustrate that presence of an application ring is established for these clip bearing cap device also provides additional working channels that have an established use in endoscopy of the GI-tract.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807-92(b)
The sponsor manufactures a number of devices similar to the subject device. Some are leveraged as reference devices. Performance bench testing settings investigated key aspects of device safety and effectiveness. As these performance bench testings were largely applied to the reference devices, these tests are considered to be acceptable (510(k) Flowchart decision 5a). Data from these acceptable methods support and demonstrate substantial equivalence (510(k) Flowchart decision 5b).
Device was tested for functional capability and robustness, endoscope compatibility, clip application ring strength, space keeper balloon capability and robustness, usability, and MRI suitability.
Not applicable - no clinical testing submitted for demonstration of substantial equivalence.
The subject device was subjected to performance bench testing previously applied to other devices that constitute a cap attached to an endoscope tip for astrointestinal clip application. The established precedent from these devices were met as evidenced by application of acceptance criteria and / or comparison to their respective performance bench testing substantial equivalence.
## 21 CFR 807.92(a)(5)
plied. Data from these
21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.