MedtFine Safety Lancet

K241848 · Ningbo Caremed Medical Products Co., Ltd. · FMK · Aug 20, 2024 · General, Plastic Surgery

Device Facts

Record IDK241848
Device NameMedtFine Safety Lancet
ApplicantNingbo Caremed Medical Products Co., Ltd.
Product CodeFMK · General, Plastic Surgery
Decision DateAug 20, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4850
Device ClassClass 2

Indications for Use

The safety lancet is intended for capillary blood sampling.

Device Story

MedtFine Safety Lancet is a sterile, single-use, disposable blood lancet featuring an integral sharps injury prevention mechanism. It consists of an outer shell, middle sleeve, needle holder, spring, and needle. The device is operated by the user (medical professional or patient) to obtain capillary blood samples. Upon activation, the spring-loaded mechanism drives the needle to a specific depth to puncture the skin, after which the needle is retracted into the housing to prevent accidental sharps injury. The device is provided sterile via radiation (SAL 10^-6) and is intended for single-person use. It is available in various needle gauges (18G to 32G) and puncture depths to accommodate different sampling requirements. The device is used in clinical or home environments; it does not require complex training or external processing.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including performance testing per ISO 9626:2016, sharps injury protection testing per ISO 23908, simulated clinical use studies, and biocompatibility testing per ISO 10993-1. Sterility, packaging, and shelf-life (5 years) were validated per ISO 11137-2, ASTM F1980-16, and ASTM F4169-22.

Technological Characteristics

Materials: PP (outer shell, small cover), ABS (middle sleeve), PE (needle holder, upper shell, back cover), carbon steel (spring), stainless steel (needle). Sensing/Actuation: Spring-loaded mechanical puncture. Dimensions: Needle gauges 18G-32G; exposure lengths 1.6mm-3.0mm. Sterilization: Radiation (SAL 10^-6). Connectivity: None. Software: None.

Indications for Use

Indicated for capillary blood sampling in both clinical and home settings for medical professionals and lay persons.

Regulatory Classification

Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (b) *Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (c) *Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that: (A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and (B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected. (vi) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use. (C) Handwashing instructions for the user before and after use of the device. (D) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device. (F) Instructions for the safe disposal of the device. (G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades. (H) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.” (B) “Used lancet blades must be safely discarded after a single use.” (C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.” (d) *Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class III (premarket approval).(3) *Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. August 20, 2024 Ningbo Caremed Medical Products Co. Ltd. % Evan Hu Technical and Regulatory Director Shanghai Mind-link Consulting Co., Ltd. 377 Tianzhu Road, Jiading Shanghai, 201801, China Re: K241848 Trade/Device Name: MedtFine Safety Lancet Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: June 13, 2024 Received: June 27, 2024 Dear Evan Hu: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by Long H. Chen -Long H. Chen-S Date: 2024.08.20 13:09:43 -04'00' Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K241848 Device Name MedtFine Safety Lancet Indications for Use (Describe) The safety lancet is intended for capillary blood sampling. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary ### 1. Preparation date: Jun 12, 2024 ### 2. Submitter Manufacturer: Ningbo Caremed Medical Products Co. Ltd. Address: No.79 Jiutang Road, South Side Hangzhou Bay New Zone, Ningbo 315336, China. Contact person: Cen Wei, General Manager, +86057458582551, wei-cen@caremed-nb.com Submission correspondent: Evan Hu, Technical and Regulatory Director, +86-18616124827, evan.ww.hu@outlook.com ### 3. Device Trade name: MedtFine Safety Lancet Common name: Blood Lancet Regulation Name: Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature Regulation: 21 CFR § 878.4850 Product code: FMK Classification: Class II Review Pannel: General & Plastic Surgery ### 4. Predicate device Predicate device: K220370 Trade name: Safety Lancet (Model: XIII, XVII, XXI, XXIII, XXIV, XXV, XXVI) Common name: Blood Lancet Regulation Name: Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature Regulation: 21 CFR § 878.4850 Product code: FMK Classification: Class II Review Pannel: General & Plastic Surgery ### 5. Device description Sterile safety lancet consists of outer shell, middle sleeve, needle holder, spring, upper shell, back cover, small cover, needle. The sterile part of the safety lancet is the needle tip. The sterile barrier is the needle holder and sterilized to a SAL of 10-6 by radiation {5}------------------------------------------------ #### K241848 sterilization. It is intended for single use and single person use only. The shelf-life of the product is 5 years. The device is for medical professionals and lay person use as an OTC use. | NO. | Gauge | Needle exposure length<br>(±0.3mm) | Color | |-----|-----------|------------------------------------|-------------| | 1 | 18G | 1.8 | Orange | | 2 | 18G | 2.2 | Rose red | | 3 | 18G Blade | 3.0 | Deep green | | 4 | 21G | 2.2 | Grass green | | 5 | 21G | 2.4 | Light Blue | | 6 | 23G | 2.2 | Deep blue | | 7 | 26G | 1.8 | Yellow | | 8 | 28G | 1.8 | Lake blue | | 9 | 30G | 1.6 | Pink | | 10 | 32G | 1.6 | Purple | #### Table 1. Device specifications ### 6. Indications for use/Intended use The safety lancet is intended for capillary blood sampling. ## 7. Comparison of technological characters between proposed and predicate devices Table 2. Characters comparison | Element of<br>Comparison | MedtFine Safety Lancet | Safety Lancet<br>(K220370) | Comment | |-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------| | Product Code | FMK | FMK | Same | | Indication for Use | The safety lancet is<br>intended for capillary<br>blood sampling. | The safety lancet is<br>intended<br>for capillary blood<br>sampling. | Same | | OTC and Rx Use | OTC use | OTC use | Same | | Element of<br>Comparison | MedtFine Safety Lancet | Safety Lancet<br>(K220370) | Comment | | Specifications | Needle: 18G, 18G blade,<br>21G, 23G, 16G, 28G,<br>30G, 32G<br>House Type: XXV (below<br>graph) | Needle: 18G, 18G blade,<br>21G, 23G, 16G, 28G, 30G,<br>32G<br>House Type: XIII, XVII,<br> | Same,<br>within range. | | Safety protection<br>features | Yes | Yes | Same | | Reuse durability | Single use for single<br>person | Single use for single<br>person | Same | | User environments | Home, clinical | Home, clinical | Same | | Components and<br>materials | Outer shell: PP<br>Middle sleeve: ABS<br>Needle holder: PE<br>Spring: carbon steel<br>Upper Shell: PE<br>Back cover: PE<br>Small cover: PP<br>Needle: Stainless Steel | Needle core (contain<br>Needle): PE, PP<br>Calcium powder (main<br>ingredient: calcium<br>carbonate)<br>stainless steel (needle)<br>silicone oil (needle)<br>Housing, Button, Bottom,<br>Protective cap, Small lid,<br>Depth adjuster ring: ABS,<br>PS | #1 | | Sterilization | Radiation<br>SAL=10-6 | Radiation<br>SAL=10-6 | Same | | Shelf life | 5 years | 5 years | Same | | Activation force | 4~7N | 4~15N | Same, within<br>range. | | Puncture force | 18-26G: 0.6~1.3N <2N<br>28~32G: 0.38~0.71N <1N | 18~26G: 0.7~1.4N <2N<br>28~32G: 0.7~0.9N <1N | Same, meet<br>requirements | | Puncture depth<br>(needle exposure<br>length) | 18~32G: 1.6~2.2mm<br>18G blade: 3.0mm | 18~32G: 1.6~2.2mm<br>18G blade: 3.0mm | Same | | Biocompatibility | Meet requirements of ISO<br>10993 serial standards | Meet requirements of ISO<br>10993 serial standards | Same | {6}------------------------------------------------ {7}------------------------------------------------ Comments: #1: The proposed and predicate devices used different materials for manufacturing. The device structures kept the same. The performance testing and biocompatibility testing results demonstrated the proposed device met the requirements of various standards ## 8. Non-clinical testing ## PERFORMANCE TESTING The following tests were performed per ISO 9626:2016 and internal requirements to demonstrate substantial equivalence: - Appearance - - -Needle size - Firmness of needle and plastic handle - - Puncture force - - Safety performance - - Limits for acidity or alkalinity - - -Resistance to corrosion Besides, the following tests were performed to determine the safe usage of devices. - Sharps injury protection per ISO 23908 - - Simulated clinical use study per FDA Guidance: Medical Devices with Sharps Injury -Prevention Features ## BIOCOMPATIBILITY TESTING: The proposed device was tested in compliance with the 2020 FDA Guidance document Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing within a Risk Management Process", as the Externally Communicating Device, Blood Path Indirect, Limited Contact (< 24hrs). The items listed below underwent testing in accordance with ISO 10993-1. - -Cytotoxicity - Intracutaneous Reactivity - - -Sensitization - -Systemic toxicity - Hemolysis - - Pyrogen - - Endotoxin - ### STERILE, PACKAGE AND SHELF-LIFE: The sterilization process of the proposed device has been validated in compliance with ISO 11137-2:2013. The Shelf-Life validation study was conducted under accelerated aging conditions in compliance with ASTM F1980-16 to verify the claimed shelf-life of 5 years. {8}------------------------------------------------ Package integrity testing under simulated shipping conditions was conducted to satisfy the requirements in ASTM F4169-22. All packaging was deemed acceptable for protection of product and sterility maintenance. Sterile barrier testing was conducted in compliance with the following FDA recognized consensus standards. - Cap removal force: ASTM F88/F88-15 - - Dye penetration: ASTM F1929-23. - - Sterility: USP <71> #### 9. Clinical testing Not applicable for this submission. #### 10. Conclusion The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use and technological characteristics.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...