ATMOS C 051 Thorax (317.0200.0); Secretion canister 800ml (317.1300.0); Hose system (312.1177.0); Hose system with connector small (312.1206.0); Hose system with connector medium (312.1207.0); Hose system with connector large (312.1208.0); Hose system with Y-connector medium (312.1209.0); Hose system with Y-connector large (312.1210.0); Universal bracket for ATMOS C 051 Thorax (316.0200.0); Bracket for ATMOS C 051 Thorax - Standard rail (317.1160.0); Charger Storage for bracket ATMOS C
K241799 · Atmos Medizintechnik GmbH & Co. KG · BTA · Mar 7, 2025 · General, Plastic Surgery
Device Facts
Record ID
K241799
Device Name
ATMOS C 051 Thorax (317.0200.0); Secretion canister 800ml (317.1300.0); Hose system (312.1177.0); Hose system with connector small (312.1206.0); Hose system with connector medium (312.1207.0); Hose system with connector large (312.1208.0); Hose system with Y-connector medium (312.1209.0); Hose system with Y-connector large (312.1210.0); Universal bracket for ATMOS C 051 Thorax (316.0200.0); Bracket for ATMOS C 051 Thorax - Standard rail (317.1160.0); Charger Storage for bracket ATMOS C
Applicant
Atmos Medizintechnik GmbH & Co. KG
Product Code
BTA · General, Plastic Surgery
Decision Date
Mar 7, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ATMOS C 051 Thorax is indicated for aspiration and removal of surgical fluids, tissue, gases, bodily fluids, or infectious material from a patient's respiratory system after surgery. The general indication of the ATMOS C 051 Thorax is thoracic drainage.
Device Story
ATMOS C 051 Thorax is a portable, battery-operated digital thoracic drainage system; creates vacuum in pleural cavity to drain air/secretion. Inputs: pleural cavity air/fluids via hose system; outputs: vacuum-assisted drainage into 800ml secretion canister. Features electronic monitoring with optical/acoustic displays; includes air-rinsing of hoses to prevent debris; bacterial/viral filters prevent contamination. Used in clinical or home settings by patients or healthcare providers. Provides mobility via carrying handle/strap. Healthcare providers use output data to monitor drainage progress; aids clinical decision-making regarding patient recovery. Benefits include portable, mains-independent operation for post-surgical thoracic management.
Clinical Evidence
No clinical data. Evidence consists of bench testing (electrical/mechanical/thermal safety per IEC 60601 series), biocompatibility (ISO 10993-1), and human factors validation (IEC 62366-1). Performance metrics included flow rate accuracy, vacuum control, battery operation, and alarm functionality. Risk management performed per ISO 14971.
Technological Characteristics
Portable powered suction pump; rechargeable battery; 800ml secretion canister; hose system with bacterial/viral filters. Vacuum control system; optical/acoustic display. Complies with IEC 60601-1, 60601-1-2, 60601-4-2. Sterilization via ISO 11135. Software-controlled monitoring.
Indications for Use
Indicated for post-surgical aspiration and removal of surgical fluids, tissue, gases, bodily fluids, or infectious material from the respiratory system (thoracic drainage) in patients.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
ATMOS E 201 Thorax and ATMOS S 201 Thorax (K103042)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
March 7, 2025
ATMOS MedizinTechnik GmbH & Co. KG Abdulaziz Abdunasimov Regulatory Affairs Manager Ludwig-Kegel-Str. 16 Lenzkirch, BW 79853 Germany
## RE: K241799
Trade/Device Name: ATMOS C 051 Thorax (317.0200.0); Secretion canister 800 ml (317.1300.0); Hose system (312.1177.0); Hose system with connector small (312.1206.0); Hose system with connector medium (312.1207.0); Hose system with connector large (312.1208.0); Hose system with Y-connector medium (312.1209.0); Hose system with Y-connector large (312.1210.0); Universal bracket for ATMOS C 051 Thorax (316.0200.0); Bracket for ATMOS C 051 Thorax - Standard rail (317.1160.0); Charger Storage for bracket ATMOS C 051 Thorax (317.1170.0); Disposable strap for ATMOS C 051 Thorax (316.1200.0): Hose clamp (061.0079.0) Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: February 7, 2025
Dear Abdulaziz Abdunasimov:
Received: February 10, 2025
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
{1}------------------------------------------------
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James H. Digitally signed by
James H. Jang -S
Jang -S Date: 2025.03.07
07:04:37 -05'00'
For
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241799
Device Name ATMOS C 051 Thorax
Indications for Use (Describe)
The ATMOS C 051 Thorax is indicated for aspiration and removal of surgical fluids, tissue, gases, bodily fluids, or infectious material from a patient's respiratory system after surgery. The general indication of the ATMOS C 051 Thorax is thoracic drainage.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|--|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for a company called ATMOS Medizintechnik. The logo consists of the word "ATMOS" in blue, with the letters arranged in a circular shape. Below the word "ATMOS" is the word "MedizinTechnik" in black. The logo is simple and modern, and it is likely used to represent the company's brand.
# 510(k) Summary – K241799
#### l. General Information
| ■ Applicant Name: | ATMOS MedizinTechnik GmbH & Co. KG |
|-------------------|------------------------------------|
|-------------------|------------------------------------|
- " Applicant Address: Ludwig-Kegel-Str. 16 Lenzkirch BW 79853 Germany
- Applicant Contact Telephone: +49 7653 689236
- Applicant Contact: Mr. Abdulaziz Abdunasimov
- . Applicant Contact Email:
- regulatory@atmosmed.de
- March 5, 2025 Date Prepared:
#### II. Device Name
.
.
.
- Device Trade Name: ATMOS C 051 Thorax
- . Accessories Trade Name:
Secretion canister 800ml; Hose system; Hose system with connector small; Hose system with connector medium; Hose system with connector large; Hose system with Y-connector medium; Hose system with Y-connector large; Universal bracket for ATMOS C 051 Thorax; Bracket for ATMOS C 051 Thorax - Standard rail; Charger Storage for bracket ATMOS C 051 Thorax; Disposable strap for ATMOS C 051 Thorax; Hose clamp
| ■ | Common Name: | Powered suction pump |
|---|----------------------|------------------------------------------------|
| ■ | Classification Name: | Pump, Portable, Aspiration (Manual Or Powered) |
| ■ | Regulation Number: | 878.4780 |
| ■ | Product Code(s): | BTA |
| ■ | Class: | Class II |
| ■ | Review Panel: | General and Plastic Surgery |
#### Legally Marketed Predicate Device(s) lll.
| ▪ 510(k) Number: | K103042 |
|------------------|-------------------------------------------|
| ▪ Trade Name: | ATMOS E 201 Thorax and ATMOS S 201 Thorax |
| ▪ Product Code: | BTA |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image features the logo for ATMOS MedizinTechnik. The logo consists of the word "ATMOS" in a bold, sans-serif blue font, positioned within a blue circle. Below the circle and the word "ATMOS", the words "MedizinTechnik" are written in a smaller, bold, sans-serif black font. The overall design is clean and professional, suggesting a company in the medical technology sector.
#### IV. Device Description
The ATMOS C 051 Thorax (317.0200.0) drainage system is a device for mobile, digital thoracic drainage. It is portable, mains-independent and has an electronic monitoring system with optical and acoustic display. The device takes medical effect as a system in combination with a secretion canister and a hose system. The system creates the (natural) vacuum in the pleural cavity by draining off air and secretion.
The ATMOS C 051 Thorax is equipped with a rechargeable battery. A charging unit which is located within the suction device guarantees the secure charging of the battery.
The hose system is rinsed with air at regular intervals to prevent the collection of debris in the hose. Bacterial and viral filters in the secretion canister and measuring hose prevent contaminated secretions from entering the device. The device is equipped with a disposable strap and a carrying handle. These enable mobility. A universal bracket or the standard rail bracket can be ordered separately as accessories.
The ATMOS C 051 Thorax is a reusable device delivered in non-sterile state. The secretion canister and hose systems are single-use devices and delivered sterile. All other accessories subject to this submission are reusable and delivered non-sterile.
#### V. Indications for Use
The ATMOS C 051 Thorax is indicated for aspiration and removal of surgical fluids, tissue, gases, bodily fluids, or infectious material from a patient's respiratory system after surgery. The general indication of the ATMOS C 051 Thorax is thoracic drainage.
#### VI. Indications for Use Comparison
The subject ATMOS C 051 Thorax and the predicate ATMOS S 201 Thorax are both devices for mobile, digital thoracic drainage. Both devices have the same indications for use - aspiration and removal of surgical fluids, tissue, gases, bodily fluids, or infectious material from a patient's respiratory system after surgery. The general indication is thoracic drainage.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image features the logo of ATMOS MedizinTechnik. The logo consists of the word "ATMOS" in a bold, sans-serif font, stacked above the words "MedizinTechnik" in a smaller, bolder font. The word "ATMOS" is encircled by a blue ring, giving the logo a clean and modern appearance. The overall design is simple and professional, reflecting the company's focus on medical technology.
#### VII. Technological Comparison
The ATMOS C 051 Thorax has the same technological characteristics regarding operation principle, use and type of accessories, vacuum control and maximum vacuum compared to the predicate device ATMOS S 201 Thorax.
The following differences exist:
- . Size and weight: The subject secretion canister has only 800 ml compared to 2 L of the predicate canister, and the subject device weighs only 1.16 kg compared to 3.27 kg of the predicate device.
- " Home use: Home care use is possible with the subject device but not for the predicate device.
- . Pump performance: Free flow rate is lower for the subject device (5 +/- 0.5 L/min) compared to the predicate device (18 +/- 2 L/min).
The completed verification and validation tests, as well as human factors validation and risk management activities, show that these differences do not raise new or different questions of safety and/or effectiveness.
#### VIII. Summary of Non-Clinical and/or Clinical Tests
Performance tests were carried out to ensure that the subject device functions as intended and meet design specifications. The following performance data were provided in support of substantial equivalence determination:
- . Electrical, Mechanical, and Thermal Safety
The ATMOS C 051 Thorax has been successfully tested for electrical, mechanical, and thermal safety in accordance with AAMI/ANSI and IEC 60601 standards and complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-4-2.
- l Software
The ATMOS C 051 Thorax contains software. The documentation level for the device software is considered to be "enhanced." Software validation activities were performed in accordance with the FDA guidance "Content of Premarket Submissions for Device Software Functions," issued on June 14, 2023. To fulfill the requirements with regard to cybersecurity, the FDA guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions," issued on September 27, 2023, was followed.
- 트 Biocompatibility Testing
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for ATMOS MedizinTechnik. The logo consists of the word "ATMOS" in blue, surrounded by a blue circle. Below the circle, the words "MedizinTechnik" are written in black.
A biological evaluation was performed according to ISO 10993-1 and followed the recommendations in the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," issued September 8, 2023, including Attachment G: Biocompatibility of Certain Devices in Contact with Intact Skin.
#### l Performance Bench Testing
The completed tests include tests with regard to design and performance, as well as human factors validation. Risk analysis was carried out based on acceptance criteria established in accordance with ISO 14971.
The following bench performance tests of ATMOS C 051 Thorax were performed with no unexpected results or significant deviations:
- -Maximum flow rate
- -Function and accuracy of flow display
- -Minimum to maximum vacuum
- Function, accuracy and control delay of vacuum display -
- -Battery operation at full power mode
- -Battery charge time
- -Over suction protection
- -Weight
- Alarm sound and noise level -
- -Length of hose system
### Human factors validation:
A usability engineering process has been performed in accordance with IEC 62366-1:2015+Cor1:2016+A1:2020, IEC 60601-1-6:2010+A1:2013+A2:2020, and FDA guidance document "Applying Human Factors and Usability Engineering to Medical Devices," issued on February 3, 2016. FDA draft guidance "Content of Human Factors Information in Medical Device Marketing Submission," issued on December 9, 2022, has been followed as well. The use-related risk analysis lists all identified risks associated with the use of the device. The human factors (HF) engineering report summarizes the conducted HF testing. The performed human factors validation shows that the use of the device is safe, especially with regard to the use by lay persons. The residual risks that
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for Atmos Medizintechnik. The logo features the word "ATMOS" in a bold, sans-serif font, stacked on top of the words "MedizinTechnik" in a smaller, sans-serif font. The word "ATMOS" is enclosed in a blue circle. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.
remain after HF validation testing for the ATMOS Thorax C 051 have been reviewed and determined to be acceptable.
A risk management process is routinely performed by ATMOS as part of the product development in accordance with ISO 14971:2019. The Risk Management Documentation specifies all identified risks and summarizes all conducted risk management activities. Risks specifically related to the home use of the subject device have been considered in the risk management. The benefit-risk analysis shows that the overall benefits of using the device outweigh the risks. The device is safe and effective for use in the intended patient population.
#### " Reprocessing, Sterility, and Shelf-Life
Required cleaning and disinfecting procedures for the ATMOS C 051 Thorax are described in the instructions for use. A sterile validation was performed for the secretion canister and hose systems. Sterilization was performed according to ISO 11135.
#### IX. Conclusion
The performance testing demonstrates that the ATMOS C 051 Thorax is as safe and effective as, and substantially equivalent to, the predicate device for the same intended use. Any differences between the subject and predicate devices do not raise new or different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.