Archived surgical pathology medical records; Routine clinical FFPE tissue slides
A retrospective, multi-center study compared the diagnostic performance of the Epredia E1000 Dx Digital Pathology Solution (Manual Digital) against traditional optical light microscopy (Manual Optical) using archived clinical cases to demonstrate non-inferiority.
Retrospective, multi-center study of Epredia E1000 Dx Digital Pathology Solution; Retrospective, multi-center, non-inferiority study; Follow-up/Duration: Not applicable; Study Period: Cases were at least one year old
Surgical pathology cases (FFPE tissue) from routine clinical practice; Sample Size: 1,299 cases (1,796 individual slides); Number of Sites: 3 enrollment sites
Manual Optical (MO) microscopy review of the same clinical slides
Major discordance rate between digital review and original sign-out diagnosis
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Surgical pathology slide interpretation
Machine learning model for tissue detection
Lower bound of 95% CI of overall agreement >= 85%
96.9% (95% CI: 96.1%, 97.6%)
—
—
Precision study: 279 study features from 257 glass slides
9 (US board-certified reading pathologists)
Surgical pathology diagnosis
—
Major discordance rate < 7.0% and upper limit of 95% CI of difference < 4.0%
Major discordance rate 2.54% (MD) vs 2.65% (MO); Difference -0.15% (95% CI: -0.40%, 0.41%)
—
—
Retrospective, multi-center study: 1,299 cases
>1 (pathologists) + 5 (Adjudication Pathologists)
Indications for Use
The Epredia E1000 Dx Digital Pathology Solution is an automated digital slide creation, viewing, and management system. The Epredia E1000 Dx Digital Pathology Solution is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Epredia E1000 Dx Digital Pathology Solution is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The Epredia E1000 Dx Digital Pathology Solution consists of a Scanner (E1000 Dx Digital Pathology Scanner), which generates in MRXS image file format, E1000 Dx Scanner Software, Image Management System (E1000 Dx IMS), E1000 Dx Viewer Software, and Display (Barco MDPC-8127). The Epredia E1000 Dx Digital Pathology Solution is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using Epredia E1000 Dx Digital Pathology Solution.
Device Story
Automated whole slide imaging system; processes up to 1,000 FFPE tissue slides. Scanner captures glass slides; generates digital images in proprietary MRXS format. System includes scanner, scanner software, Image Management System (IMS), viewer software, and Barco MDPC-8127 display. Used in clinical pathology settings; operated by laboratory technicians and pathologists. Pathologists view digital images on high-resolution display as alternative to conventional brightfield microscopy. Output aids in diagnostic interpretation; supports clinical decision-making by providing digital access to slide content. Benefits include high-capacity automated workflow and remote/digital review capabilities.
Clinical Evidence
Retrospective, multi-center study (1,299 cases, 7,778 reading pairs) compared digital WSI review (MD) to traditional optical microscopy (MO). Primary endpoint: difference in major discordance rates relative to reference sign-out diagnosis. MD major discordance rate: 2.51% (95% CI: 2.26%, 2.79%); MO major discordance rate: 2.59% (95% CI: 2.29%, 2.82%). Difference: -0.15% (95% CI: -0.40%, 0.41%). Results met non-inferiority criteria. Precision studies (within-system, between-system, between-site, between-pathologist, within-pathologist) showed agreement rates with lower 95% CI bounds >85%.
Indicated for use by pathologists to review and interpret digital images of surgical pathology slides prepared from FFPE tissue. Not indicated for frozen section, cytology, or non-FFPE hematopathology specimens.
Regulatory Classification
Identification
The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.
Special Controls
A whole slide imaging system must comply with the following special controls: (1) Premarket notification submissions must include the following information: (i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system. (ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate: (A) Slide feeder; (B) Light source; (C) Imaging optics: (D)Mechanical scanner movement; (E) Digital imaging sensor; (F) Image processing software; (G)Image composition techniques; (H)Image file formats; (I) Image review manipulation software; (J) Computer environment; (K)Display system. (iii)Detailed bench testing and results at the system level, including for the following, as appropriate: (A)Color reproducibility; (B) Spatial resolution; (C) Focusing test; (D) Whole slide tissue coverage; (E) Stitching error: (F) Turnaround time. (iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate: (A)Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (e.g., main sign-out diagnosis). (D) A detailed human factors engineering process must be used to evaluate the whole slide imaging system user interface(s). (2) Labeling compliant with 21 CFR 809.10(b) must include the following: The intended use statement must include the information described in paragraph (i) (1)(i) of this section, as applicable, and a statement that reads, "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device." (ii) A description of the technical studies and the summary of results, including those that relate to paragraph (1)(ii) and (1)(iii) of this section, as appropriate. (iii) A description of the performance studies and the summary of results, including those that relate to paragraph (1)(iv) of this section, as appropriate. (iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
*e.g.,* main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K241717
B Applicant
Shandon Diagnostics Limited
C Proprietary and Established Names
Epredia E1000 Dx Digital Pathology Solution
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PSY | Class II | 21 CFR 864.3700 | 88-Pathology |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Type of Test:
Digital pathology whole slide imaging
## III Intended Use/Indications for Use:
A Intended Use(s):
The Epredia E1000 Dx Digital Pathology Solution is an automated digital slide creation, viewing, and management system. The Epredia E1000 Dx Digital Pathology Solution is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Epredia E1000 Dx Digital Pathology Solution is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
The Epredia E1000 Dx Digital Pathology Solution consists of a Scanner (E1000 Dx Digital Pathology Scanner), which generates images in MRXS image file format, E1000 Dx Scanner Software, Image Management System (E1000 Dx IMS), E1000 Dx Viewer Software, and Display (Barco MDPC-8127). The Epredia E1000 Dx Digital Pathology Solution is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using Epredia E1000 Dx Digital Pathology Solution.
## B Indication(s) for Use:
Same as Intended Use.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For In vitro diagnostic (IVD) use only.
## IV Device/System Characteristics:
### A Device Description:
The E1000 Dx Digital Pathology Solution is an automated whole slide imaging system for the creation, viewing, and management of digital images of surgical pathology slides, which consists of the following three components:
Scanner component:
- E1000 Dx Digital Pathology Scanner with E1000 firmware version 2.0.3
- E1000 Dx Scanner Software version 2.0.3
Viewer component:
- E1000 Dx Image Management System (IMS) Server version 2.3.2
- E1000 Dx Viewer Software version 2.7.2
Display component:
- Barco MDPC-8127
The E1000 Dx Digital Pathology Solution creates digital whole slide images (WSIs) by scanning FFPE tissue slides. The user loads prepared glass slides into commercially available (Sakura Tissue-Tek® 20-Slide Basket) slide baskets which holds a maximum of 20 slides. The total capacity of the system is 50 baskets, resulting in a capacity of 1,000 slides. The E1000 Dx Scanner Software (EDSS) runs on the scanner workstation and controls the operation of the E1000 Dx Digital Pathology Scanner. The scanner workstation, provided with the E1000 Dx Digital Pathology Solution, includes a PC, monitor, keyboard, and mouse. The EDSS user interface allows the user to select baskets containing slides for scanning and generation of WSIs. The system will only scan areas of the glass slide where tissue sample is present using a machine learning model trained on annotated samples to automatically detect the presence and location of the tissue sample on the glass slide. Once the sample has been located, the system generates a scan-map from the data. FFPE tissue mounted on glass slides are scanned at a resolution of at least $0.24\mu \mathrm{m}$ per pixel using a $20\mathrm{x}$ objective. The system acquires images in a square matrix pattern. The device uses a proprietary MRXS format to store and transmit images between the
K241717 - Page 2 of 17
{2}
E1000 Dx Digital Pathology Scanner and the E1000 Dx Image Management System (IMS). MRXS images displayed in E1000 Dx Viewer are compressed to 80% of the source MRXS file, or JPEG Q = 80.
The E1000 Dx IMS is a software only component which contains two modules hosted on the IMS server: E1000 Dx IMS (allows the pathologist to manage patient cases from digitization through to pathologist assignment and viewing of images) and E1000 Dx Storage (server software for storing and managing WSIs created by the E1000 scanner and provides background processes for storing and managing the digital slides to be used within E1000 Dx IMS and E1000 Dx Viewer). IMS local server is connected to the Scanner Workstation on a high-speed local LAN network. The server runs all IMS services and applications as well as authentication and logging services. A user can access the IMS through a web browser via the E1000 Dx IMS web application user interface (UI) and open slides in the E1000 Dx Viewer software from a customer provided computer environment.
The E1000 Dx Viewer is a Windows application that allows viewing and annotating MRXS digital slides created by the E1000 Whole Slide Imaging System. It is integrated with the E1000 Dx IMS and it cannot be used as a standalone application. The E1000 Dx Viewer can only be launched from the E1000 Dx IMS web application by clicking on a digital slide attached to a case. The main features of the viewing software are:
- Display image with seamless zooming and panning
- Record and recall image annotations
- Export image snapshots
- Display multiple slides from the same case side by side
The Barco MDPC-8127 display (cleared under K203364) allows the WSIs images to be viewed.
## B Instrument Description Information:
1. Instrument Name:
Epredia E1000 Dx Digital Pathology Solution
2. Specimen Identification:
Glass slides and scanned images are identified based on the previously assigned specimen identifiers such as patient identifiers, barcodes, etc. Digital images of surgical pathology slides prepared from FFPE tissue. The preview image captures the barcode information on the slide, which links the scanned slides to the correct entry in the E1000 Dx IMS.
3. Specimen Sampling and Handling:
Specimen sampling and handling are performed independently of the system. Specimen sampling includes surgical pathology specimens such as biopsy or resection specimens which are processed using standard histology techniques. The FFPE tissue sections are stained using the Hematoxylin and Eosin (H&E) staining procedure. Then digital images are obtained from these glass slides using the Epredia E1000 Dx Digital Pathology scanner.
4. Calibration:
The E1000 Dx Digital Pathology Scanner is factory calibrated. No calibration is required by the operator of the instrument. The spectral properties of components in the pixel pathway, including the light source, are stable by design, so no regular color recalibration is needed.
K241717 - Page 3 of 17
{3}
Image intensity might vary from slide to slide depending on the slide thickness, and the system automatically adjusts for it. The E1000 Dx IMS subsystem does not need calibration. The Barco MDPC-8127 8MP 27” Pathology Monitor Display is automatically calibrated.
5. Quality Control:
Before loading the slide to the instrument, the operator ensures the slide is within the specifications of the instrument and laboratory standards such as slide size, coverslip size and placement, barcode placement, staining quality, etc. After scanning a lab technician checks if the WSI quality and rescans the slide(s) as needed.
Before reading the scanned WSIs, the pathologist should ensure that all scanned slide images have been imported for every case and the images are of acceptable quality for diagnostic purposes. The pathologist reviews scanned images of all slides associated with a case before rendering a diagnosis.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Philips IntelliSite Pathology Solution (PIPS)
B Predicate 510(k) Number(s):
DEN160056
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K241717 | DEN160056 |
| --- | --- | --- |
| Device Trade Name | Epredia E1000 Dx Digital Pathology Solution | Philips IntelliSite Pathology Solution (PIPS) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Epredia E1000 Dx Digital Pathology Solution is an automated digital slide creation, viewing, and management system. The Epredia E1000 Dx Digital Pathology Solution is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Epredia E1000 Dx Digital Pathology Solution is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. | The Philips IntelliSite Pathology Solution (PIPS) is an automated digital slide creation, viewing, and management system. The PIPS is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The PIPS is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The PIPS comprises the Image |
K241717 - Page 4 of 17
{4}
K241717 - Page 5 of 17
| | The Epredia E1000 Dx Digital Pathology Solution consists of a Scanner (E1000 Dx Digital Pathology Scanner), which generates images in MRXS image file format, E1000 Dx Scanner Software, Image Management System (E1000 Dx IMS), E1000 Dx Viewer Software, and Display (Barco MDPC-8127). The Epredia E1000 Dx Digital Pathology Solution is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using Epredia E1000 Dx Digital Pathology Solution. | Management System (IMS), the Ultra-Fast Scanner (UFS) and Display. The PIPS is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using PIPS. |
| --- | --- | --- |
| Device Components | WSI Scanner (E1000 Dx Digital Pathology Scanner), Image Management System (E1000 Dx IMS), Image Viewer (E1000 Dx Viewer) and color monitor display. | PIPS Ultra Fast Scanner, Image Management System and color monitor display. |
| Scanning Magnification | 0.25 μm per pixel | 0.24 μm per pixel |
| Principles of Operation | The technician performs quality control procedures per the laboratory’s standard and loads surgical pathology slides, prepared from FFPE tissue, into the Epredia Dx Digital Scanner. Subsequently, the scanner scans the slides, generating WSIs from each slide. The resulting WSI images are stored in the image storage system included in the device. During the review process, the pathologist uses the IMS to open the E1000 Dx Viewer to view the images | Prior to scanning the slide on the Ultra Fast Scanner (UFS), the technician conducts quality control on the slides prepared from FFPE tissue per the laboratory’s standards. The technician places the slides then into slide racks, loads the racks into the UFS. The images scanned in the UFS are transmitted to the IMS subsystem. The pathologist accesses WSI images acquired through the UFS from the image |
{5}
| | from the image storage and proceeds to analyze the WSI images of the slides. | storage and proceeds to analyze the WSI images of the slides. |
| --- | --- | --- |
| General Device Characteristic Differences | | |
| Image Review Manipulation Functionality | Continuous panning and pre-fetching, continuous zooming capability to compare multiple slides simultaneously on multiple windows, image enhancement and color manipulation, and annotation tools. Sharpening is done on the scanner side. | Continuous panning and pre-fetching, continuous zooming, discrete Z-axis displacement, ability to compare multiple slides simultaneously on multiple windows, image enhancement and sharpening functions, color manipulation, annotation tools, digital bookmarks, and virtual multi-head microscope. |
| Image File Format | Proprietary format MRXS is used to store and transmit images between the E1000Dx Scanner and the IMS. | Philips’s proprietary format, iSyntax, is used to store and transmit the images between the UFS and the IMS. |
| Slide feeder | 1000 slides (50 glass slide racks with up to 20 slides per rack). | 300 slides (15 glass slide racks with up to 20 slides per rack). |
| Display | Barco MDPC-8127 | Philips PP27QHD (Barco MMPC-4227F1) |
VI Standards/Guidance Documents Referenced:
1. FDA guidance “Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices,” dated April 20, 2016.
2. Applying Human Factors and Usability Engineering to Medical Devices: Guidance for Industry and Food and Drug Administration Staff February 3, 2016.
3. ISO 13485 Third Edition 2016-03 - Medical devices — Quality management systems — Requirements for regulatory purposes.
4. ISO 14971 Third edition 2019-12 - Medical devices — Application of risk management to medical devices.
5. IEC 62304 Edition 1.1 2015-06 - Medical device software — Software life cycle processes.
6. IEC 61010-1 Edition 3.1 2017-01 - Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements.
7. IEC 61010-2-101 Third Edition 2018-10 - Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment.
8. IEC 61326-1: 2020 - Electrical equipment for measurement, control, and laboratory use - EMC requirements - Part 1: General requirements.
K241717 - Page 6 of 17
{6}
9. IEC 61326-2-6: 2020 - Electrical equipment for measurement, control and laboratory use-EMC Requirements. Part 2-6: Particular requirements- In vitro diagnostic (IVD) medical equipment.
10. IEC 62366-1:2015 +A1:2020 - Medical devices — Part 1: Application of usability engineering to medical devices — Amendment 1
11. IEC 60825-1:2014 - Safety of laser products - Part 1: Equipment classification and requirements.
12. IEC 62471:2006-07 - Photobiological safety of lamps and lamp systems (including LEDs)
13. IEC 63000:2018 - Technical documentation for the assessment of electrical and electronic products with respect to the restriction of hazardous substance.
14. ISO 15223-1 Fourth edition 2021-07 - Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements.
15. ISO 18113-1:2011- In vitro diagnostic medical devices Information supplied by the manufacturer (labelling) — Part 1: Terms, definitions, and general requirements.
16. ISO 18113-3:2011 In vitro diagnostic medical devices Information supplied by the manufacturer (labelling) — Part 3: In vitro diagnostic instruments for professional use.
17. ISO 20417: 2021 - Medical devices - Information to be supplied by the manufacturer.
## VII Performance Characteristics (if/when applicable):
### A. Analytical Performance:
#### 1. Precision/Reproducibility:
The objective of this study was to evaluate the repeatability (within- and between- system) and reproducibility (between-site) of the E1000 Dx Digital Pathology Solution.
The precision of the device was evaluated based on the review and identification of specific histopathologic "features" that are observed in FFPE H&E stained slides. Features selected for the study included tissue level features, cellular level features and sub-cellular level features. The study included 279 "study" features and 42 additional "wild card" features at different magnifications (10x, 20x, 40x) from 257 glass slides as shown in Table 1 below. Wild card features were used to reduce the risk of reader recall bias and were not used in the analysis. A minimum of 5 different features were evaluated at each relevant magnification level. Three scanning sites, 3 reading sites and a total of 9 US board-certified reading pathologists (RPs) were used in this study.
Table 1: Primary Histologic Study Features in Precision Study
| 10x Magnification (n) | 20x Magnification (n) | 40x Magnification (n) |
| --- | --- | --- |
| Tissue level Feature | Cellular level Feature | Subcellular level Feature |
| Adipocytes (16) | Clear Cells (18) | Pigment (17) |
| Small Artery (16) | Duct (17) | Macrophages/Histocytes (17) |
K241717 - Page 7 of 17
{7}
| Nerve (16) | Lymphoid Aggregate (16) | Intercellular Bridges (16) |
| --- | --- | --- |
| Necrosis (15) | Myxoid Stroma (16) | Lobular Carcinoma (15) |
| Squamous Epithelium (16) | Goblet Cells (16) | Nucleoli (14) |
| Gland (16) | Calcification (12) | Cilia (13) |
| Muscle (16) | - | Nuclear Pleomorphism (12) |
| - | - | Keratin Pearl (11) |
| Study Feature =111
Wild Card Feature = 15 | Study Feature =95
Wild Card Feature =12 | Study Feature =115
Wild Card Feature = 15 |
Consecutive cases were selected from the pathology laboratory using the laboratory information system (LIS) by an enrollment pathologist (EP). A different validating enrollment pathologist (VEP) confirmed whether the feature was present on the glass slide. Slide scanning of the same set of enrolled slides was performed sequentially at all three scanning sites. Three rescans, as needed, of each slide was allowed. Once the slides were scanned, the EP reviewed the image and defined an area (bookmark) containing the selected feature(s) at the appropriate magnification. Then a static full resolution extraction image of the bookmark was created and defined as the Field of View (FOV). The VEP confirmed whether the feature(s) was present in the FOV. After VEP confirmation, the FOV was considered enrolled into the study. These FOV images were randomly rotated in 90-degree increments and randomly assigned to the reading pathologists (RPs) to minimize recall bias. The reading pathologist (RP) at each site recorded the presence of each observed feature on a checklist. For each magnification, a separate checklist containing features was developed with a multiple answer list of up to 9 possible answers. There was a minimum washout period of 14 days between pathologist reading sessions. Only the primary features were used in the data analysis.
The precision of the Epredia E1000 Dx Digital Pathology Solution was assessed in the following studies:
- Intra-System (Within-System) precision was assessed using 3 independent systems. Overall within system precision was also assessed.
- Inter-System (Between-System) precision was assessed using 3 independent systems at 3 different sites.
- Within- and Between-pathologist precision was assessed using images generated from a single system.
Study acceptance criteria: The precision was considered acceptable if the lower bounds of the 2-sided $95\%$ confidence interval (CI) of the overall agreements for each precision component (within-system, between-system/site, within-pathologist, and between-pathologist) were $\geq 85\%$ .
K241717 - Page 8 of 17
{8}
Within-System Precision
In total, 321 features (3 x 93 study features + 42 wild card features) were used. Each of 279 study slides had at least one primary feature. The study slide set was divided equally (n = 93 study slides + 42 wild card slides) over 3 systems. Each slide set was scanned three times with at least six hours between scanning iterations on each system. From these scans, a total of 405 FOVs (279 study + 126 wild card) were extracted.
Randomly selected FOVs originating from three different systems and three different iterations were read by each of three RPs for a total of 3,645 [(93+42 FOVs) x 3 systems x 3 scans x 3 Pathologists] reads. The reading pathologist at the site recorded the presence of each observed feature on a checklist for each pre-determined magnification level. Data analysis included 2,511 (93 study FOVs x 3 systems x 3 scans x 3 Pathologists) reads out of 3,645 reads.
Study results are shown in Table 2 below. The data showed that the acceptance criteria of a lower limit of the 95% Confidence interval (CI) greater than 85%, with an Average Positive Agreement (APA) of 96.9% (lower limit of 96.1%) was met.
Table 2: Within Systems Precision Study Results
| System | Number of Pairwise Agreements* | Number of Comparison Pairs | Agreement Rate | |
| --- | --- | --- | --- | --- |
| | | | % | 95% CI |
| System 1 | 820 | 837 | 98.0 | (96.8, 98.8) |
| System 2 | 803 | 837 | 95.9 | (94.5, 97.3) |
| System 3 | 810 | 837 | 96.8 | (95.4, 97.9) |
| Total | 2,433 | 2,511 | 96.9 | (96.1, 97.6) |
Between-System Precision Study
The between-system precision study utilized the full set of glass slides from the within-system study, including the same extracted FOVs, primary study features and wildcard features. Each slide was scanned once on each system. Three reading pathologists then evaluated each enrolled FOV once on each system. The pathologists recorded the presence of each observed feature on a checklist for each magnification level. Each RP read 321 FOVs (279 Study FOVs, 42 wild card FOVs) in each of three (3) different reading sessions. There was a total of 2,889 reads [(279 study slides + 42 wild card slides) x 3 systems x 1 times x 3 Pathologists] and 378 of wild-card FOVs readings were excluded. There were 2,511 pairwise comparisons used for calculating the agreement rates.
Study results are shown in Table 3 below. The data showed that the acceptance criteria of a lower limit of 95% CI greater than 85%, with an APA of 95.1% (lower limit of 94%) was met.
Table 3: Between-System Precision Study Results
| System | Number of Pairwise Agreements | Number of Comparison Pairs | Agreement Rate | |
| --- | --- | --- | --- | --- |
| | | | % | 95% CI |
| System 1 vs 2 | 788 | 837 | 94.1 | (91.3, 96.0) |
| System 1 vs 3 | 793 | 837 | 94.7 | (92.6, 96.5) |
| System 2 vs 3 | 808 | 837 | 96.5 | (94.7, 97.9) |
| Total | 2,389 | 2,511 | 95.1 | (94.1, 96.1) |
K241717 - Page 9 of 17
{9}
Between-Site Reproducibility
For the between-site reproducibility study, the FOVs from the within-system and between-system precision studies were used as the study FOVs. The study involved 3 different RPs, each located at one of 3 different sites, each equipped with its own E1000 Dx Digital Pathology Solution. The order of FOV evaluation at each site was randomly determined. Each pathologist evaluated each FOV once and recorded the presence of each observed feature on a checklist for each magnification level.
Study results are shown in Table 4 below. The data showed that the acceptance criteria of a lower limit of 95% CI greater than 85%, with an APA of 95.4% (lower limit of 93.5%) was met.
Table 4: Between-Site Reproducibility Agreement Rates
| System | Number of Pairwise Agreements | Number of Comparison Pairs | Agreement Rate | |
| --- | --- | --- | --- | --- |
| | | | % | 95% CI |
| Site 1 vs 2 | 261 | 279 | 93.5 | (89.4, 96.7) |
| Site 1 vs 3 | 268 | 279 | 96.1 | (92.5, 98.4) |
| Site 2 vs 3 | 270 | 279 | 96.8 | (93.3, 98.8) |
| Total | 799 | 837 | 95.4 | (93.6, 97.2) |
Between-Pathologist Reproducibility
Each RP read 321 FOVs (279 Study FOVs, 42 wild card FOVs) in each of the 3 different reading sessions. There were a total 2,889 reads (963 reads per RP), and 378 of wild-card FOVs readings were excluded from the analysis. A total of 2,511 pairwise comparisons were used for calculating the agreement rates.
Study results are shown in Table 5 below. The data show that the study acceptance criterion of the lower limit of the 95% confidence interval of the % of agreement rate exceeding 85% was met.
Table 5: Between-Pathologist Reproducibility
| System | Number of Pairwise Agreements | Number of Comparison Pairs | Agreement Rate | |
| --- | --- | --- | --- | --- |
| | | | % | 95% CI |
| Pathologist 1 vs 2 | 818 | 837 | 97.7 | (96.6, 98.7) |
| Pathologist 1 vs 3 | 813 | 837 | 97.1 | (95.9, 98.3) |
| Pathologist 2 vs 3 | 816 | 837 | 97.5 | (96.4, 98.5) |
| Total | 2,447 | 2,511 | 97.5 | (96.9, 98.1) |
Within-Pathologist Reproducibility
Each RP read 321 FOVs (279 Study FOVs, 42 wild card FOVs) in each of 3 different reading sessions. There was a total of 2,889 reads (963 reads per RP), and 378 wild-card FOVs readings were excluded from the analysis. Agreement pairs were scored for each pathologist between repeated scans (Scan 1 vs 2, Scan 1 vs 3, and Scan 2 vs 3) across all 3 systems. Totals for each pathologist as well as aggregated data from all pathologists was also calculated.
Study results are shown in Table 6 below. The data show that the study acceptance criterion of the lower limit of the 95% confidence interval of the % of agreement rate exceeding 85% was met.
K241717 - Page 10 of 17
{10}
Table 6: Between-Pathologist Reproducibility
| System | Number of Pairwise Agreements | Number of Comparison Pairs | Agreement Rate | |
| --- | --- | --- | --- | --- |
| | | | % | 95% CI |
| RP 1, System 1 | 269 | 279 | 96.4 | (93.6, 98.4) |
| RP 1, System 2 | 269 | 279 | 96.4 | (93.6, 98.4) |
| RP 1, System 3 | 265 | 279 | 95.0 | (91.9 97.3) |
| RP 1 (Total) | 803 | 837 | 95.9 | (94.4, 97.2) |
| RP 2, System 1 | 271 | 279 | 97.1 | (94.5, 98.8) |
| RP 2, System 2 | 269 | 279 | 96.4 | (93.7, 98.4) |
| RP 2, System 3 | 267 | 279 | 95.7 | (92.6, 97.9) |
| RP 2 (Total) | 807 | 837 | 96.4 | (94.9, 97.7) |
| RP 3, System 1 | 271 | 279 | 97.1 | (94.5, 98.8) |
| RP 3, System 2 | 267 | 279 | 95.7 | (92.6, 97.8) |
| RP 3, System 3 | 269 | 279 | 96.4 | (93.6, 96.5) |
| RP 3 (Total) | 807 | 837 | 96.4 | (95.0, 97.6) |
| Overall | 2417 | 2,511 | 96.3 | (95.5, 97.2) |
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Accuracy (Instrument):
Not applicable
5. Carry-Over:
Not applicable
B. Technical Studies:
Multiple studies were conducted to evaluate the performance of the E1000 Dx Digital Pathology Solution as recommended in FDA Guidance, "Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices".
a. Slide Feeder
The slide feeder mechanism was described, including a physical description of the slide, the number of slides in queue (carrier), and the class of automation. The user interaction with the slide feeder was also described, including hardware, software, and feedback mechanisms.
b. Light Source
Descriptive information about the lamp and the condenser was provided. Testing information was provided to verify the spectral distribution of the light source as part of the color reproduction capability of the E1000 Dx Scanner subsystem.
K241717 - Page 11 of 17
{11}
c. Imaging Optics
An optical schematic was provided, showing all optical elements from the slide (object plane) to the digital image sensor (image plane). Descriptive information about the microscope objective, the auxiliary lenses, and the magnification of the imaging optics was also provided. Testing information about the relative irradiance, image distortions, and lateral chromatic aberrations was also provided.
d. Mechanical Scanner Movement
Information and specifications about the configuration of the stage, the method of movement, and the control of movement of the stage were provided. Test data to verify the repeatability of the stage movement and to verify that the stage movement stays within limits during operations was also provided.
e. Digital Imaging Sensor
Information and specifications about the sensor type, pixel information, responsivity specifications, noise specifications, readout rate, and digital output format were provided. Vendor provided test data to determine the correct functioning of the digital image sensor, which converts optical signals of the slide to digital signals that consist of a set of numerical values corresponding to the brightness and color at each point in the optical image, was also provided.
f. Image Processing Software
Information and specifications on the exposure control, white balance, color correction, and flat-field correction was provided.
g. Image Composition
Information and specifications on the scanning method was provided. Test data to analyze the image composition performance was provided.
h. Image Files Format
Information and specifications on the compression method, compression ratio, file format, and file organization were provided.
i. Image Review Manipulation Software
Information and specifications on continuous panning and pre-fetching, continuous zooming, discrete Z-axis displacement, ability to compare multiple slides simultaneously on multiple windows, image enhancement and sharpening functions, color manipulation, annotation tools, digital bookmarks, and virtual multihead microscope was provided.
j. Computer Environment
Information and specifications on the computer hardware, operating system, graphics card, graphics card driver, color management settings, color profile, and display interface was provided.
k. Display
Information and specifications on the technological characteristics of the display device, physical size of the viewable area and aspect ratio, backlight type and properties, frame rate and refresh rate, pixel pitch, and supported color spaces, color depth, display interface, user
K241717 - Page 12 of 17
{12}
controls of brightness, contrast ratio, gamma, color space, etc., via the on-screen display menu, ambient light adaptation, and color calibration tools was provided.
1. Color Reproducibility
Test data to evaluate the color reproducibility of the system was provided.
m. Spatial Resolution
Test data to evaluate the composite optical performance of all components in the image acquisition phase was provided.
n. Focusing Test
Test data to evaluate the technical focus quality of the system was provided.
o. Whole Slide Tissue Coverage
Test data to demonstrate that the entire tissue specimen on the glass slide is detected by the tissue detection algorithms and that all the tissue specimens are included in the digital image file was provided.
p. Stitching Error
Test data to evaluate the stitching errors and artifacts in the reconstructed image was provided.
q. Turnaround Time
Test data to evaluate the turnaround time of the system was provided.
C. Clinical Study:
A retrospective, multi-center study was conducted to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology FFPE tissue slides using the Epredia E1000 Dx Digital Pathology Solution is non-inferior to using traditional optical light microscopy. The primary endpoint was the difference in agreement rates between diagnoses rendered using Epredia E1000 Dx Digital Pathology Solution’s digital WSI review modality [Manual Digital (MD)] and the manual microscopy slide review modality [Manual Optical (MO)] when each was compared to the reference diagnosis, which is based on the original main sign-out diagnosis (SD) rendered at the study sites using an optical light microscope. The study consisted of reviewing archived, de-identified and previously “signed-out” slides representing main organ systems within surgical pathology. Cases included retrospective H&E stained FFPE tissue, special stains and/or immunohistochemical stains (IHC) from the pathology practice, but did not include frozen sections, or cytological and hematological cases.
The study included 1,299 consecutively enrolled cases containing 1796 individual slides. These cases were at least one year old for which a sign-out diagnosis is available at 3 enrollment sites. These slides were scanned in one site using the Epredia E1000 Dx slide scanner. Slides corresponding to each case were selected by the Enrollment Pathologist (EP) from the original slide(s) used for the main (sign-out) diagnosis and then given a case identifier using a barcode label and randomized. The EP reviewed the pathology report for each case and determined the main diagnosis for each case. The EP then matched the case to the list of types of cases to be evaluated and selected the representative slide(s) that reflected the main diagnosis.
K241717 - Page 13 of 17
{13}
Cases were randomly distributed to 3 reading sites. At site one, 24 pathologists were divided into 3 groups with each group consisting of pathologists representing 10 subspecialties. Some pathologists had expertise in two subspecialties. Cases within each subspecialty were read three times by the corresponding specialists at site one, distributed across the three groups for their respective subspecialties. At sites two and three, 3 RPs read every case assigned to the site. Each case was read 3 times in an alternating and randomized order, with a washout period of at least 30 days between reads, resulting in a case total of 3,897 WSI reads and 3,881 Glass reads.
## Study Inclusion Criteria
- The case is at least one year for which a sign-out diagnosis is available.
- Formalin Fixed Paraffin Embedded (FFPE) tissue slides.
- All glass slides, with human tissue obtained via surgical pathology of original case are available.
- The original sign-out diagnosis and ancillary supporting information is available.
- The selected slide or slides for the main diagnosis and the control slide(s) fulfill quality checks according to general practice.
- Case, including cases previously referred to another site, for which Hematoxylin and Eosin (H&E), Immunohistochemistry (IHC) or special stains slides used for the original sign-out diagnosis is available at the site.
- The selected slide (s) for the main diagnosis match any subtype of the organ for which the case is selected.
## Study Exclusion Criteria
- Cases that were signed out less than one year prior to the date of curation.
- Case with an amended report resulting in a change in diagnosis.
- Case with slides containing indelible markings.
- Case with slides that are damaged.
- Glass slide that is broken, has many pen markings or dirt that cannot be removed, has abnormal size/thickness, beveled edges, poor coverslip (cracks, waviness, scratches), is sticky, contains air bubbles and overhanging labels that can't be corrected, and if the stain is severely faded.
- More than one case selected (only one case may be enrolled per patient).
- Case consists of frozen section(s) only.
- Case consists of gross specimen(s) only.
An independent adjudication panel consisting of 5 Adjudication Pathologists (APs) independent of the study were utilized for this study to determine if the study diagnosis to consistent with the reference (sign-out) diagnosis. Four of the 5 APs were divided into 2 Adjudication Groups (AGs) of 2 APs each. The remaining 1 AP was assigned as tiebreaker Adjudication Pathologist (TAP). Each CRF was independently reviewed by both 2 APs of 1 of the 2 AGs to determine whether the diagnosis is consistent with the sign-out (main) diagnosis. A major discordance was defined as a difference in diagnosis that would be associated with a clinically important difference in patient management. A minor discordance was defined as a difference in diagnosis that would not be associated with a clinically important difference in patient management. In the event the APs in an AG disagreed, the TAP reviewed the case to achieve majority vote (tiebreaker). In cases where all APs and TAP have a different opinion, consensus was achieved in an adjudication panel meeting consisting of the same three APs.
K241717 - Page 14 of 17
{14}
The observed major discordance rates were observed and calculated using a Generalized Linear Mixed Model (GLIMMIX) logistic regression. For each reading result, the dependent variable was the major discordance status and independent variables included modality as a fixed effect (MD vs MO) and site, reader, and case as random effects. A two-sided 95% Confidence Interval (CI) for the overall discordance between reading modalities (MD-MO) was calculated from this analysis.
Out of the 1,299 cases included in the study, 5 cases were removed from the study and therefore did not undergo MO review. Due to the following reasons: Two cases were recalled to their original clinics as the patients were readmitted and the slides were requested by the primary physicians and slides from 3 cases were damaged during transport and were not able to be read. The reading sessions resulted in a total of 7,778 reading pairs (3,897 WSI and 3,881 Glass). In total, 3,897 (1299 cases x 3 pathologist) readings of WSI cases and 3,881(*) readings of glass slide cases were adjudicated [*final cases (glass) reviewed (1299-5)=1294; 1 case was reviewed by 2 out of 3 pathologists due to a slide breaking during transit, which resulted in 3881 glass slide case reads [(1293 cases x 3 pathologists) + (1 case x 2 pathologists)] = 3881].
## Study Results
The major discordance rate between WSI and SD (MD) was observed to be 2.54% (99/3897) and the major discordance rate between Glass and SD (MO) was observed to be 2.65% (103/3881). The overall major discordance rates estimated by the generalized linear model were 2.51% (95% CI: 2.26%; 2.79%) for MD and 2.59% (95% CI: 2.29%, 2.82%) for MO. The estimated difference in major discordance between the MD and MO rates was calculated to be -0.15% (95% CI: -0.40%, 0.41%). These results met the acceptance criteria for the study. Specifically, the major discordance rate for WSI (MD), which was 2.5%, was required to be <7.0% and the upper limit of the 95% confidence interval of the difference in major discordance rate between WSI and Glass (MD-MO), which was 0.41%, was required to be <4.0%.
The study results are shown in Table 7 below.
Table 7: Overall Major Discordance Rate for MD and MO
| | MD: Major Discordance Rate | | | MO: Major Discordance Rate | | | Difference MD - MO | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | % | 95% CI | N | % | 95% CI | % | 95% CI |
| Observed | 3897 | 2.54% | | 3881 | 2.65% | | -0.11% | |
| Modeled | | 2.51% | (2.26%; 2.79%) | | 2.59% | (2.29%, 2.82%) | -0.15% | (-0.40%, 0.41%) |
The detailed list of the major discordance rate for each organ system and anatomical region are shown in Table 8 below.
Table 8: Major Discordance Rates Observed Across Organ Systems
| Organ System | MD Major Discordance | | MO Major Discordance | | MD - MO Difference |
| --- | --- | --- | --- | --- | --- |
| | N | % | N | % | % Difference |
| Breast | 360 | 0.6% | 357 | 1.1% | -0.6% |
K241717 - Page 15 of 17
{15}
| Dermatological | 297 | 3.4% | 297 | 1.7% | 1.7% |
| --- | --- | --- | --- | --- | --- |
| Endocrine | 270 | 3.0% | 270 | 3.3% | -0.4% |
| Gastrointestinal | 1047 | 1.5% | 1037 | 1.3% | 0.3% |
| Genitourinary | 777 | 2.2% | 777 | 1.4% | 0.8% |
| Gynecological | 357 | 7.3% | 357 | 10.1% | -2.8% |
| Hematopoietic | 222 | 1.8% | 222 | 3.2% | -1.4% |
| Pulmonary | 210 | 2.4% | 207 | 4.3% | -2.0% |
| Musculoskeletal | 150 | 1.3% | 150 | 2.0% | -0.7% |
| Neurological | 207 | 4.3% | 207 | 2.9% | 1.4% |
An additional analysis was performed on the 3881 pairs in which diagnoses were recorded using both reading modalities (MD and MO) by the same pathologist. The reported diagnoses from each case were directly compared between the two reading modalities (without comparison to SD) for each pathologist. Each diagnosis deemed "Major Discordance" to SD by the adjudication panel was compared to that same pathologist's diagnosis using the other reading modality. If the diagnoses for both reading modalities were the same, that pair result was "agree". Otherwise, the pair was marked "disagree". All pairs not containing a "Major Discordance" were assumed to be "agree". Using this methodology, all available pairs were scored and the summary of the agreement rate per site is listed below in Table 9 below.
Table 9: Agreement Rate Between MD and MO (without comparing to SD)
| Study sites | Number of cases | Number of reads | Agree | | Disagree | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | N | % | N | % |
| Site 1 | 569 | 1706 | 1671 | 97.9% | 35 | 2.1% |
| Site 2 | 429 | 1287 | 1243 | 96.6% | 44 | 3.4% |
| Site 3 | 296 | 888 | 876 | 98.6% | 12 | 1.4% |
| Total | 1294 | 3881 | 3790 | 97.7% | 91 | 2.3% |
# D. Human Factor Study:
Human factors studies for E1000 Dx Digital Pathology Solution to assess performance of critical user tasks and use scenarios by representative users, including anatomic pathology lab technicians and pathologists were conducted. The studies were designed around critical user tasks and use scenarios performed by representative users from histotechnicians and pathologists (15 pathologists and 15 laboratory technicians) to assess software interface, hardware interface, and product labeling. All participants were able to perform all tasks (including the critical tasks) and no critical task failures were observed. Critical tasks were identified internally by conducting a detailed uFMEA (user Failure Modes and Effects Analysis). Both user groups perceived that the risk mitigation steps for critical tasks were successfully implemented and that the E1000 Dx Digital Pathology Solution can be used safely and effectively by its intended users, for its intended purpose, and in its intended use environment.
# VIII Proposed Labeling:
The labeling is sufficient, and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable, and the special controls for this device type under 21 CFR 864.3700.
K241717 - Page 16 of 17
{16}
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K241717 - Page 17 of 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.