FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).
Device Story
FemChec is a dual-barrel contrast media syringe used during sono-hysterosalpingogram (Sono HSG) procedures. Operated by a clinician, the device is connected to an intrauterine catheter. Retracting the plunger fills the barrels with air and saline; depressing the plunger delivers a consistent alternating stream of saline-air contrast media into the uterus and fallopian tubes. The device features a ratcheting plunger and a viewing window for monitoring contrast bubbles. It is provided sterile via ethylene oxide and is intended for single-use. The delivered contrast media allows for ultrasound visualization of the fallopian tubes and uterine cavity, aiding in clinical assessment.
Clinical Evidence
Bench testing only. Performance data included sterilization validation (ISO 11135-1, AAMI TIR 28, ISO 10993-7), package integrity (ASTM F2096-11, ASTM F88/F88M-15), transportation simulation (ASTM D4169-14), and biocompatibility (ISO 10993-5, ISO 10993-10). Bench performance studies, including saline-air quantification, were conducted before and after accelerated aging (ASTM F1980-16), meeting all acceptance criteria.
Technological Characteristics
Dual-barrel manual syringe; ratcheting plunger mechanism; spin Luer connector; integrated contrast viewing window. Materials evaluated for biocompatibility per ISO 10993-1. Sterilization: Ethylene Oxide (EO). Single-use. Dimensions: 8.7"L x 3.2"W x 1.7"H.
Indications for Use
Indicated for patients undergoing sono-hysterosalpingogram (Sono HSG) procedures for assessment of the uterine cavity and fallopian tubes.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
September 6, 2024
Femasys Inc. Christine Thomas SVP Regulatory and Clinical Affairs 3950 Johns Creek CT. Suite 100 Suwanee, Georgia 30024
Re: K241693
> Trade/Device Name: FemChec Controlled Saline-Air Device (FCD-250) Regulation Number: 21 CFR 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF Dated: June 12, 2024 Received: June 12, 2024
Dear Christine Thomas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241693
Device Name
FemChec Controlled Saline-Air Device (FCD-250)
Indications for Use (Describe)
FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
510(k) Summary - K241693
#### I. SUBMITTER
| Applicant: | Femasys Inc. |
|--------------------|-----------------------------------------------------------|
| Applicant Address: | 3950 Johns Creek Court, Suite 100<br>Suwanee, GA 30024 |
| Phone: | +770-500-3910 x137 |
| Email: | CThomas@femasys.com |
| Contact Person: | Christine Thomas<br>Sr. VP, Regulatory & Clinical Affairs |
| Date Prepared: | September 3, 2024 |
| II.<br>DEVICE | |
| Trade Name: | FemChec® Controlled Saline-Air Device |
| Common Name: | Contrast media syringe |
| Regulation Name: | Obstetric-gynecologic specialized manual instrument |
| Regulation Number: | 21 CFR 884.4530 |
| Regulatory Class: | II |
| Product Code: | LKF (Cannula, Manipulator/Injector, Uterine) |
### PREDICATE DEVICE III.
FemVue® Saline-Air Device (K110288), manufactured by Femasys Inc.
The predicate device has not been subject to a design-related recall.
## IV. DEVICE DESCRIPTION
FemChec® is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity.
The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterile via EO sterilization and is intended for single-use.
### INDICATIONS FOR USE V.
{6}------------------------------------------------
FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).
### COMPARISON OF THE INTENDED USE AND TECHNOLOGICAL VI. CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICES
The intended use and key technological characteristics of the subject and predicate device are compared in the table below.
| Comparison<br>Item | FemChec® Controlled Saline-<br>Air Device Subject<br>Device (K241693) | FemVue® Saline-Air Device<br>Predicate Device (K110288) | Comparison |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR 884.4530 | 21 CFR 884.4530 | Same |
| Device Class | Class 2 | Class 2 | Same |
| Product Code | LKF | LKF | Same |
| Device<br>Classification<br>Name | Cannula, Manipulator/Injector,<br>Uterine | Cannula, Manipulator/Injector,<br>Uterine | Same |
| Regulation<br>Description | Obstetric-gynecologic specialized<br>manual instrument | Obstetric-gynecologic specialized<br>manual instrument | Same |
| Regulation<br>Medical<br>Specialty | Obstetrics/Gynecology | Obstetrics/Gynecology | Same |
| Indications for<br>Use | FemChec is intended to instill a<br>consistent alternating pattern of<br>saline and air as a continuous<br>stream of contrast media into the<br>uterus and fallopian tubes to be<br>used in conjunction with an<br>intrauterine catheter for<br>performance of sono-<br>hysterosalpingogram (Sono<br>HSG). | The FemVue Saline-Air Device<br>instills a consistent alternating<br>pattern of saline and air as a<br>continuous stream of contrast<br>media into the uterus and fallopian<br>tubes to be used in conjunction<br>with an intrauterine catheter for<br>performance of sono-<br>hysterosalpingogram<br>(Sono HSG). | Same |
| Design<br>Features | Dual-barrel contrast media<br>syringe that can be connected to<br>an intrauterine catheter. | Dual-barrel contrast media syringe<br>that can be connected to an<br>intrauterine catheter. | Same |
| Principle of<br>Operation | Sterile dual-barrel contrast media<br>syringe that is connected to an<br>intrauterine catheter to instill<br>saline-air contrast media during<br>sono-hysterosalpingogram (Sono<br>HSG) procedures. | Sterile dual-barrel contrast media<br>syringe that is connected to an<br>intrauterine catheter to instill<br>saline-air contrast media during<br>sono-hysterosalpingogram (Sono<br>HSG) procedures. | Same |
| Prescription<br>Only | Yes | Yes | Same |
| Target Patient | Patient undergoing Sono HSG<br>procedure. | Patient undergoing Sono HSG<br>procedure. | Same |
| Device<br>Dimensions | 8.7"L x 3.2"W x 1.7"H | 9.3"L x 3.5"W x 1.7"H | Different: The differences do<br>not raise different questions<br>of safety and effectiveness. |
| Catheter Luer<br>Connection | Spin Luer | Fixed Luer | Different: The different luer<br>connectors do not raise<br>different questions of safety<br>and effectiveness. |
| Contrast<br>Viewing<br>Window | Contrast bubbles viewed through<br>window in device | Contrast bubbles viewed through<br>neck of luer | Different: The different<br>viewing location of the<br>contrast bubbles does not<br>raise different questions of<br>safety and effectiveness. |
Table 1. Intended use and technological characteristics comparison.
{7}------------------------------------------------
The subject and predicate device have the same intended use - to deliver contrast media into the uterus and fallopian tubes in conjunction with an intrauterine catheter for the performance of Sono-HSG procedures.
The subject and predicate device have the same principle of operation (dual-barrel syringe device that delivers saline and air contrast through an intrauterine catheter). However, the devices differ in dimensions, type of Luer connection, inclusion of a viewing window, and materials. In addition, the subject device includes a ratcheting plunger, while the predicate device has a smooth plunger. These differences do not raise different questions of safety and effectiveness.
### VII. SUMMARY OF NON-CLINICAL PERFORMANCE DATA
The following studies have been performed to support substantial equivalence to the predicate device:
- Ethylene Oxide Sterilization Validation testing per: .
- > ISO 11135-1:2014
- A AAMI TIR 28:2016
- A ISO 10993-7: 2008
- Package Integrity testing: .
- A Visual inspection
- A Bubble Leak test per ASTM F2096-11
- A Seal Strength testing per ASTM F88/ F88M-15
- Transportation Simulation testing per ASTM D4169-14 ●
- Biocompatibility studies conducted in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" for devices in indirect contact with mucosal tissues. Testing included the following assessments:
{8}------------------------------------------------
- Cytotoxicity per ISO 10993-5: 2009 A
- A Sensitization ISO 10993-10: 2021
- Irritation per ISO 10993-10: 2021 >
Testing showed the device material to be non-cytotoxic, non-sensitizing, and non- irritating.
- Bench performance studies before and after accelerated aging to the equivalent of 12-. months of real-time aging in accordance with ASTM F1980-16 demonstrated that all predetermined acceptance criteria were met in the following tests:
- > Visual Inspection
- > Functional/Cycle Testing
- > Saline-Air Quantification
# VIII. CONCLUSIONS
The results of the testing described above demonstrate that the FemChec Controlled Saline-Air Device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.