FemChec Controlled Saline-Air Device (FCD-250)

K241693 · Femasys, Inc. · LKF · Sep 6, 2024 · Obstetrics/Gynecology

Device Facts

Record IDK241693
Device NameFemChec Controlled Saline-Air Device (FCD-250)
ApplicantFemasys, Inc.
Product CodeLKF · Obstetrics/Gynecology
Decision DateSep 6, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.4530
Device ClassClass 2

Intended Use

FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

Device Story

FemChec is a dual-barrel contrast media syringe used during sono-hysterosalpingogram (Sono HSG) procedures. Operated by a clinician, the device is connected to an intrauterine catheter. Retracting the plunger fills the barrels with air and saline; depressing the plunger delivers a consistent alternating stream of saline-air contrast media into the uterus and fallopian tubes. The device features a ratcheting plunger and a viewing window for monitoring contrast bubbles. It is provided sterile via ethylene oxide and is intended for single-use. The delivered contrast media allows for ultrasound visualization of the fallopian tubes and uterine cavity, aiding in clinical assessment.

Clinical Evidence

Bench testing only. Performance data included sterilization validation (ISO 11135-1, AAMI TIR 28, ISO 10993-7), package integrity (ASTM F2096-11, ASTM F88/F88M-15), transportation simulation (ASTM D4169-14), and biocompatibility (ISO 10993-5, ISO 10993-10). Bench performance studies, including saline-air quantification, were conducted before and after accelerated aging (ASTM F1980-16), meeting all acceptance criteria.

Technological Characteristics

Dual-barrel manual syringe; ratcheting plunger mechanism; spin Luer connector; integrated contrast viewing window. Materials evaluated for biocompatibility per ISO 10993-1. Sterilization: Ethylene Oxide (EO). Single-use. Dimensions: 8.7"L x 3.2"W x 1.7"H.

Indications for Use

Indicated for patients undergoing sono-hysterosalpingogram (Sono HSG) procedures for assessment of the uterine cavity and fallopian tubes.

Regulatory Classification

Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable. September 6, 2024 Femasys Inc. Christine Thomas SVP Regulatory and Clinical Affairs 3950 Johns Creek CT. Suite 100 Suwanee, Georgia 30024 Re: K241693 > Trade/Device Name: FemChec Controlled Saline-Air Device (FCD-250) Regulation Number: 21 CFR 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF Dated: June 12, 2024 Received: June 12, 2024 Dear Christine Thomas: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2}------------------------------------------------ the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Monica D. Garcia -S Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K241693 Device Name FemChec Controlled Saline-Air Device (FCD-250) Indications for Use (Describe) FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG). Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ 510(k) Summary - K241693 #### I. SUBMITTER | Applicant: | Femasys Inc. | |--------------------|-----------------------------------------------------------| | Applicant Address: | 3950 Johns Creek Court, Suite 100<br>Suwanee, GA 30024 | | Phone: | +770-500-3910 x137 | | Email: | CThomas@femasys.com | | Contact Person: | Christine Thomas<br>Sr. VP, Regulatory & Clinical Affairs | | Date Prepared: | September 3, 2024 | | II.<br>DEVICE | | | Trade Name: | FemChec® Controlled Saline-Air Device | | Common Name: | Contrast media syringe | | Regulation Name: | Obstetric-gynecologic specialized manual instrument | | Regulation Number: | 21 CFR 884.4530 | | Regulatory Class: | II | | Product Code: | LKF (Cannula, Manipulator/Injector, Uterine) | ### PREDICATE DEVICE III. FemVue® Saline-Air Device (K110288), manufactured by Femasys Inc. The predicate device has not been subject to a design-related recall. ## IV. DEVICE DESCRIPTION FemChec® is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity. The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterile via EO sterilization and is intended for single-use. ### INDICATIONS FOR USE V. {5}------------------------------------------------ FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG). ### COMPARISON OF THE INTENDED USE AND TECHNOLOGICAL VI. CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICES The intended use and key technological characteristics of the subject and predicate device are compared in the table below. | Comparison<br>Item | FemChec® Controlled Saline-<br>Air Device Subject<br>Device (K241693) | FemVue® Saline-Air Device<br>Predicate Device (K110288) | Comparison | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Classification | 21 CFR 884.4530 | 21 CFR 884.4530 | Same | | Device Class | Class 2 | Class 2 | Same | | Product Code | LKF | LKF | Same | | Device<br>Classification<br>Name | Cannula, Manipulator/Injector,<br>Uterine | Cannula, Manipulator/Injector,<br>Uterine | Same | | Regulation<br>Description | Obstetric-gynecologic specialized<br>manual instrument | Obstetric-gynecologic specialized<br>manual instrument | Same | | Regulation<br>Medical<br>Specialty | Obstetrics/Gynecology | Obstetrics/Gynecology | Same | | Indications for<br>Use | FemChec is intended to instill a<br>consistent alternating pattern of<br>saline and air as a continuous<br>stream of contrast media into the<br>uterus and fallopian tubes to be<br>used in conjunction with an<br>intrauterine catheter for<br>performance of sono-<br>hysterosalpingogram (Sono<br>HSG). | The FemVue Saline-Air Device<br>instills a consistent alternating<br>pattern of saline and air as a<br>continuous stream of contrast<br>media into the uterus and fallopian<br>tubes to be used in conjunction<br>with an intrauterine catheter for<br>performance of sono-<br>hysterosalpingogram<br>(Sono HSG). | Same | | Design<br>Features | Dual-barrel contrast media<br>syringe that can be connected to<br>an intrauterine catheter. | Dual-barrel contrast media syringe<br>that can be connected to an<br>intrauterine catheter. | Same | | Principle of<br>Operation | Sterile dual-barrel contrast media<br>syringe that is connected to an<br>intrauterine catheter to instill<br>saline-air contrast media during<br>sono-hysterosalpingogram (Sono<br>HSG) procedures. | Sterile dual-barrel contrast media<br>syringe that is connected to an<br>intrauterine catheter to instill<br>saline-air contrast media during<br>sono-hysterosalpingogram (Sono<br>HSG) procedures. | Same | | Prescription<br>Only | Yes | Yes | Same | | Target Patient | Patient undergoing Sono HSG<br>procedure. | Patient undergoing Sono HSG<br>procedure. | Same | | Device<br>Dimensions | 8.7"L x 3.2"W x 1.7"H | 9.3"L x 3.5"W x 1.7"H | Different: The differences do<br>not raise different questions<br>of safety and effectiveness. | | Catheter Luer<br>Connection | Spin Luer | Fixed Luer | Different: The different luer<br>connectors do not raise<br>different questions of safety<br>and effectiveness. | | Contrast<br>Viewing<br>Window | Contrast bubbles viewed through<br>window in device | Contrast bubbles viewed through<br>neck of luer | Different: The different<br>viewing location of the<br>contrast bubbles does not<br>raise different questions of<br>safety and effectiveness. | Table 1. Intended use and technological characteristics comparison. {6}------------------------------------------------ The subject and predicate device have the same intended use - to deliver contrast media into the uterus and fallopian tubes in conjunction with an intrauterine catheter for the performance of Sono-HSG procedures. The subject and predicate device have the same principle of operation (dual-barrel syringe device that delivers saline and air contrast through an intrauterine catheter). However, the devices differ in dimensions, type of Luer connection, inclusion of a viewing window, and materials. In addition, the subject device includes a ratcheting plunger, while the predicate device has a smooth plunger. These differences do not raise different questions of safety and effectiveness. ### VII. SUMMARY OF NON-CLINICAL PERFORMANCE DATA The following studies have been performed to support substantial equivalence to the predicate device: - Ethylene Oxide Sterilization Validation testing per: . - > ISO 11135-1:2014 - A AAMI TIR 28:2016 - A ISO 10993-7: 2008 - Package Integrity testing: . - A Visual inspection - A Bubble Leak test per ASTM F2096-11 - A Seal Strength testing per ASTM F88/ F88M-15 - Transportation Simulation testing per ASTM D4169-14 ● - Biocompatibility studies conducted in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" for devices in indirect contact with mucosal tissues. Testing included the following assessments: {7}------------------------------------------------ - Cytotoxicity per ISO 10993-5: 2009 A - A Sensitization ISO 10993-10: 2021 - Irritation per ISO 10993-10: 2021 > Testing showed the device material to be non-cytotoxic, non-sensitizing, and non- irritating. - Bench performance studies before and after accelerated aging to the equivalent of 12-. months of real-time aging in accordance with ASTM F1980-16 demonstrated that all predetermined acceptance criteria were met in the following tests: - > Visual Inspection - > Functional/Cycle Testing - > Saline-Air Quantification # VIII. CONCLUSIONS The results of the testing described above demonstrate that the FemChec Controlled Saline-Air Device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%