K241656 · Termosalud · GEX · Aug 12, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241656
Device Name
Eneka Selectif
Applicant
Termosalud
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ENEKA SELECTIF is intended for use in dermatologic and general surgical procedures.
Device Story
Eneka Selectif is a non-invasive diode laser system for hair removal, PFB treatment, and benign pigmented/vascular lesion treatment. It utilizes AlGaAs laser diode arrays at 755nm, 810nm, 1060nm, or a blended 810-1060nm wavelength. The device features seven handpieces with varying spot sizes and two treatment modes: Dynamic (DHR) and Fast Dynamic (FDHR). Operators (clinicians) configure pulse width, frequency, and fluence via an LCD touchscreen. The system includes a contact cooling system (3°C) and a contact detection system to prevent laser firing without skin contact. The device is used in clinical settings for dermatologic procedures. Output affects clinical decision-making by providing controlled thermal energy to target tissues, benefiting patients through permanent hair reduction and lesion treatment.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by bench testing, including verification of output power, electromagnetic compatibility (IEC 60601-1-2), and electrical safety (IEC 60601-1, IEC 60601-2-22). Biocompatibility is established by reference to identical materials used in the Eneka Pro (ISO 10993-1).
Indicated for hair removal (permanent reduction in hair regrowth), treatment and prevention of Pseudofolliculitis barbae (PFB), and treatment of benign pigmented or vascular lesions. Applicable to all skin types (Fitzpatrick I-VI) using 755nm, 810nm, 1060nm, or 810-1060nm laser applicators.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Underneath the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Augusst 12, 2024
TermoSalud % Aubrey Thompson Regulatory Consultant Hoy & Associates Regulatory Consulting 1830 Bonnie Way Sacramento, California 95825
Re: K241656
Trade/Device Name: Eneka Selectif Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 8, 2024 Received: June 10, 2024
Dear Aubrey Thompson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{3}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
TANISHA TANISHA L. HITHE L. HITHE -S Date: 2024.08.1
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical & Infection Control Devices, Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241656
Device Name Eneka Selectif
Indications for Use (Describe) Intended Use ENEKA SELECTIF is intended for use in dermatologic and general surgical procedures.
Indications for Use
ENEKA SELECTIF System with 755nm Laser applicator is intended for:
- Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth.
defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime.
- Treatment and prevention of Pseudofolliculitis barbae (PFB).
- Use on all skin types (Fitzpatrick I-VI).
- Treatment of benign pigmented lesions.
ENEKA SELECTIF System with 810nm Laser applicator is intended for:
- Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime.
- Treatment and prevention of Pseudofolliculitis barbae (PFB).
- Use on all skin types (Fitzpatrick I-VI).
- Treatment of benign pigmented lesions.
ENEKA SELECTIF System with 1060nm Laser applicator is intended for:
- Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime.
- Treatment and prevention of Pseudofolliculitis barbae (PFB).
- Use on all skin types (Fitzpatrick I-VI).
- Treatment of benign vascular lesions.
ENEKA SELECTIF System with 810 - 1060mm Laser applicator is intended for:
-Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime.
- Treatment and prevention of Pseudofolliculitis barbae (PFB).
- Use on all skin types (Fitzpatrick I-VI).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
This 510(K) Summary of safety and effectiveness for the ENEKA SELECTIF is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.
| Applicant | TermoSalud |
|----------------------------------------------|----------------------------------------------------------------|
| Address | Ataulfo Friera Tarfe, 8 - 33211 Gijón, Spain |
| Contact Person | Aubrey Thompson, Regulatory Consultant |
| Contact Information | Aubreythompson@hoyregulatory.com |
| Preparation Date | August 6, 2024 |
| Device Trade Name<br>K Number<br>Common Name | ENEKA SELECTIF<br>K241656<br>Powered Laser Surgical Instrument |
| Regulation Number | 21 CFR 878.4810 |
| Product Code | GEX |
| Regulatory Class | II |
| Legally Marketed Predicate<br>Device | Primelase Excellence (K201594), ENEKA PRO (K223680) |
#### Device Description:
The ENEKA SELECTIF is a non-invasive diode laser system used for hair removal, prevention and treatment of Pseudofolliculitis barbae and treatment of vascular and pigmented lesions.
The system operates at different wavelengths (755, 810 and 1060 nm) separately plus a blended handpiece combining three wavelengths (810, 940 and 1060 nm).
The ENEKA SELECTIF is a diode laser device with 7 handpieces (spot sizes 10x9mm, 20x9mm and 30x12mm) and it allows two modes of treatment, DHR and FDHR. Both of them can be applied with every handpiece. Mode DHR allows the configuration of pulse width, frequency and fluency and mode FDHR allows the configuration of frequency and fluency.
This submission is based on substantial equivalence between the Eneka Selectif and its predicate device, the Primelase Excellence diode laser. In addition, this device is an expansion of capabilities for the previously cleared Eneka Pro (K223680), which is used as a reference device. The following tables describe the similarities between the three devices.
## Intended Use and Indications for Use:
#### Intended Use
ENEKA SELECTIF is intended for use in dermatologic and general surgical procedures.
{7}------------------------------------------------
## Indications for Use
ENEKA SELECTIF System with 755nm Laser applicator is intended for:
- Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime.
- Treatment and prevention of Pseudofolliculitis barbae (PFB).
- . Use on all skin types (Fitzpatrick I-VI).
- Treatment of benign pigmented lesions.
ENEKA SELECTIF System with 810nm Laser applicator is intended for:
- Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime.
- . Treatment and prevention of Pseudofolliculitis barbae (PFB).
- Use on all skin types (Fitzpatrick I-VI).
- Treatment of benign pigmented lesions. ●
ENEKA SELECTIF System with 1060nm Laser applicator is intended for:
- Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction . in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime.
- Treatment and prevention of Pseudofolliculitis barbae (PFB).
- . Use on all skin types (Fitzpatrick I-VI).
- . Treatment of benign vascular lesions.
ENEKA SELECTIF System with 810 - 1060nm Laser applicator is intended for:
- . Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime.
- Treatment and prevention of Pseudofolliculitis barbae (PFB).
- Use on all skin types (Fitzpatrick I-VI).
{8}------------------------------------------------
Substantial Equivalence—Indications for use comparison:
| Subject Device: Eneka Selectif | Predicate Device: Primelase Excellence<br>K201594 | Reference Device: Eneka Pro<br>K223680 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ENEKA SELECTIF System with 755nm Laser applicator is intended for:<br>Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime. Treatment and prevention of Pseudofolliculitis barbae (PFB). Use on all skin types (Fitzpatrick I-VI). Treatment of benign pigmented lesions. ENEKA SELECTIF System with 810nm Laser applicator is intended for:<br>Hair removal with Dynamic and Fast Dynamic (FDHR) modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regime. Treatment and prevention of Pseudofolliculitis barbae (PFB). Use on all skin types (Fitzpatrick I-VI). | PRIMELASE Excellence System with 755nm Laser applicator is intended for:<br>Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime. Treatment of Pseudofolliculitis barbae (PFB). Use on all skin types (Fitzpatrick I-VI). Treatment of benign pigmented lesions PRIMELASE Excellence System with 810nm Laser applicator is intended for:<br>Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime. Treatment of Pseudofolliculitis barbae (PFB). Use on all skin types (Fitzpatrick I-VI). Treatment of benign pigmented lesions | Indications for use for Eneka Pro diode laser hair removal system with 808nm applicators include:<br>Hair Removal with Dynamic (DHR) and Fast Dynamic (FDHR) mode intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime. Treatment of Pseudofolliculitis barbae (PFB). Use on all skin types (Fitzpatrick I-VI) |
| Treatment of benign pigmented lesions. | | |
| ENEKA SELECTIF System with 1060nm Laser<br>applicator is intended for: | PRIMELASE Excellence System with 1060nm<br>Laser applicator is intended for: | |
| Hair removal with Dynamic and Fast Dynamic<br>modes intended for permanent reduction in<br>hair regrowth, defined as a long term, stable<br>reduction in the number of hairs re-growing<br>when measured at 6, 9 and 12 months after<br>the completion of a treatment regime. Treatment and prevention of<br>Pseudofolliculitis barbae (PFB). Use on all skin types (Fitzpatrick I-VI). Treatment of benign vascular lesions. | Hair Removal with Static and Dynamic<br>modes intended for permanent<br>reduction in hair regrowth, defined as a<br>long term, stable reduction in the<br>number of hairs re-growing when<br>measured at 6, 9, and 12 months after<br>the completion of a treatment regime. Treatment of Pseudofolliculitis barbae<br>(PFB). Use on all skin types (Fitzpatrick I-VI). Treatment of benign vascular lesions,<br>including angiomas, hemangiomas,<br>telangiectasia, and other benign<br>vascular lesions and leg veins. Treatment of wrinkles | |
| ENEKA SELECTIF System with 810 – 1060nm Laser<br>applicator is intended for: | PRIMELASE Excellence System with 810-1060nm<br>Laser applicator is intended for: | |
| Hair removal with Dynamic and Fast Dynamic<br>modes intended for permanent reduction in<br>hair regrowth, defined as a long term, stable<br>reduction in the number of hairs re-growing<br>when measured at 6, 9 and 12 months after<br>the completion of a treatment regime. Treatment and prevention of<br>Pseudofolliculitis barbae (PFB). Use on all skin types (Fitzpatrick I-VI). | Hair Removal with Static and Dynamic<br>modes intended for permanent<br>reduction in hair regrowth, defined as a<br>long term, stable reduction in the<br>number of hairs re-growing when<br>measured at 6, 9, and 12 months after<br>the completion of a treatment regime. Treatment of Pseudofolliculitis barbae<br>(PFB). Use on all skin types (Fitzpatrick I-VI). | |
{9}------------------------------------------------
{10}------------------------------------------------
Substantial Equivalence: Technical Specifications Comparison
| Specification | Subject Device<br>ENEKA SELECTIF | Predicate Device<br>Primelase Excellence –<br>K201594 | Reference Device ENEKA PRO –<br>K223680 | Comparison |
|--------------------------------|---------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>operation | AlGaAs Laser diode array | AlGaAs Laser diode array | AlGaAs Laser diode array | Same |
| Laser<br>Wavelength | 755 nm<br>810 nm<br>1060 nm<br>810-1060 nm | 755 nm<br>810 nm<br>1060 nm<br>810 – 1060 nm | 808nm | Same. This submission includes 7<br>handpieces with 755, 810, 1060nm<br>and 810-1060nm wavelengths. The<br>predicate device contains handpieces<br>with the same wavelengths, including<br>the wavelength with multiple<br>wavelength options. The ENEKA PRO<br>includes 2 handpieces at 808nm,<br>both of which are included in the<br>current submission. |
| Laser Contact | Sapphire, Al2O3 | Sapphire, Al2O3 | Sapphire, Al2O3 | Same |
| Spot Sizes | 10x9mm (0.9cm2),<br>30x12mm (3.6cm2)<br>20x9mm (1.8cm2) | 10x10mm (1.0cm2)<br>20x9mm (1.8cm2)<br>30x9mm (2.7cm2)<br>30x17mm (5.1cm2) | 20x9mm (1.8cm2),<br>34x14mm (4.76cm2) | The spot size for ENEKA SELECTIF L<br>handpiece (10x9mm) is very similar<br>to the spot size of Primelase<br>(10x10mm). The spot size of ENEKA<br>SELECTIF XL handpiece (20x9mm) is<br>identical to Primelase (20x9mm) and<br>Eneka Pro (20x9mm) handpieces. The<br>spot size dimension of the 2XL<br>handpiece for ENEKA SELECTIF<br>(30x12mm) is between the two<br>largest Primelase handpieces<br>(30x9mm and 30x17mm). Safety and<br>effectiveness are not impacted by the<br>minor differences in spot size. |
| Fluence Max | 100J/cm2 | 120J/cm2 | 43J/cm2 | The fluence of the ENEKA SELECTIF is<br>very similar to the fluence of the<br>predicate device. The difference with |
| K241656 | | | | |
| Frequency | 1 – 4Hz (Dynamic)<br>5 – 10Hz (Fast Dynamic) | Up to 3 Hz (Static)<br>5 – 10 Hz (Dynamic) | 1 – 4Hz (Dynamic)<br>5 – 10Hz (Fast Dynamic) | ENEKA PRO is that this device is used<br>only for hair removal and this<br>treatment does not require high<br>fluence values. Fluence for Hair<br>removal treatments with the Eneka<br>Selectif is limited to 43J/cm².<br>Same. The frequency of the ENEKA<br>SELECTIF is identical to ENEKA PRO,<br>and is within the range of frequencies<br>of the predicate device. |
| Pulse Duration | 5 – 400ms | 3 – 400 ms | 5 – 400ms | The pulse duration of ENEKA<br>SELECTIF is identical to the reference<br>device pulse duration and very<br>similar to the predicate's pulse<br>duration. |
| Tissue Cooling | Contact cooling system | Contact continuous,<br>thermo-electrical | Contact cooling system | Same. |
| Cooling<br>Temperature | 3 °C | 5 °C | 3 °C | Same. The contact cooling<br>temperature is identical to ENEKA<br>PRO predicate device, and almost<br>identical (±2 °C) to Primelase. |
| Contact<br>detection<br>system | Yes | No | No | Different. There are no contact<br>detection systems on the predicate<br>or reference devices. Contact<br>detection system is a requirement to<br>comply with the last revision of IEC<br>60601-2-22. |
| User Interface | LCD Touchscreen | LCD Touchscreen | LCD Touchscreen | Same. |
| Pulsing<br>Control | Finger Switch | Finger Switch | Finger Switch | Same. |
| Configuration | Main unit and handpiece | Main unit, handpiece, and<br>foot control (optional) | Main unit and handpiece | Same |
| Laser Classification | Class 1C (Module class 4) | Class 4 | Class 4 | Same. |
| Power Supply | 115 VAC 50/60 Hz<br>230 VAC 50/60 Hz | Single Phase, 100-240V, 50-60 Hz | 115 VAC 50/60 Hz<br>230 VAC 50/60 Hz | Different from Predicate, same as reference device |
| Dimension | 620 x 430 x 630 mm | 1140 x 480 x 550 mm | 620 x 430 x 630 | Different from Predicate, same as reference device. |
| Weight | 32 kg | 75 kg | 32 kg | Different from Predicate, same as reference device |
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
### Performance Testing
Verification and validation activities were successfully completed and establish that the ENEKA SELECTIF performs as intended. Testing included the following:
IEC 60601-1:2005 + A1:2012 + A2:2020 Medical electrical equipment −Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014+A1:2020 Medical electrical equipment, Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility Requirements and tests
IEC 60601-2-22:2019 : Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
ISO 10993-1:2018: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
Software verification and validation testing was conducted, and documentation provided in accordance with FDA's Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices.
Performance testing was conducted to verify the output power of the laser at each power level with all handpieces.
Clinical Evidence - N/A. No clinical studies were conducted as part of this submission.
Biocompatibility – The patient contacting materials in the Eneka Selectif are identical to those used in the Eneka Pro. No modifications to materials have been made, and no additional testing was required.
#### Conclusion
Based on the comparison and analysis above, the ENEKA SELECTIF and PRIMELASE are nearly identical in the specifications and the hair removal handpieces and indications are identical to the previously cleared Eneka Pro. The indications for use are identical to the PRIMELASE with the exception of wrinkle treatment, which is not included in this submission. The Eneka Selectif and Primelase have small differences in the design that do not impact the safety or effectiveness of the subject device, but only impact the user experience. To increase the safety of the treatment, the ENEKA SELECTIF includes a contact detection system to avoid firing the laser when no contact with skin is established. The main difference between them resides in their accessories (wavelengths and spot sizes). The Eneka Seletif is lacking one spot size that is included in the predicate device, and the spot size for the Eneka Selectif are slightly different from the Eneka Pro.
The treatment modes only differ in the naming; the modes from Primelase Excellence, Eneka Pro and Eneka Selectif are nearly identical. The Primelase Excellence refers to its lower frequency as "static" and "dynamic." Eneka Selectif and Eneka Pro refer to the lower frequency and "dynamic" and the higher frequency as "fast dynamic." The frequency ranges for the different modes are nearly identical. The devices can be considered substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.